BD Vacutainer Eclipse™ Blood Collection Needle ®

withFOR
RUCTIONS BD Vacutainer
USE:
®
Pronto™ Quick Release Needle Holder
INSTRUCTIONS FOR USE:
Usage of Product
INSTRUCTIONS FOR USE: BD Vacutainer ® Eclipse Needle
INSTRUCTIONS
™ FOR USE
READY O R US
E: and Pronto Quick Release

NS F Needle Holder
RU CTIO
INST Pink Safety Shield

Thumb Pad
White Needle Cap
INSTRUCTIONS FOR USE:

Bevel Up

1a. Ensure BD Vacutainer® 2. While holding the pink 3a. Gently position pink
Pronto™ Holder is ready to safety shield and the colored safety shield straight back
Colored Needle Cap
accept new collection device. needle cap firmly, screw toward the holder. (green or black)
If not, press white reset holder onto needle by
button on holder. pushing needle into holder 3b. Twist and pull colored
while turning hands in needle cap straight off. Reusable Holder
1b. Holding both pink safety opposite directions 1/4 turn Note: The needle bevel is always in
shield and colored needle until needle is fully seated position for venipuncture when the
cap, twist and remove white onto holder. pink safety shield is facing up.
needle cap. Ordering Information
Do Not twist or rotate the pink
DO NOT OVERTORQUE safety shield. Reference Numbers:
BD Vacutainer® Pronto™ Holder
368871
BD Vacutainer® Eclipse™
Blood Collection Needles
368607  21 G x 1 ¼"

368608  22 G x 1 ¼"
ECLIPSE™ NEEDLE
EEDProduct
LE Storage
PSE ™N
ECLI ECLIPSE™ NEED

4. Perform venipuncture 5. Immediately after removing 6. Dispose of collection device


according to your facility’s needle from vein and with into nearest approved sharps
established procedures. the needle pointed down container:
and away from yourself and Hold device in a downward ECLIPSE™ NEEDLE
the patient, position thumb position over container TORIALS:
squarely on pink safety shield opening, push green
O R button
A G E/ TR ANSFER PIC
STORAGE/TRANSFER PICTORIALS: ST
thumb pad and push pink on holder to release collection It is recommended that product EC
safety shield forward to cover device into container. Ensure PICTORIALS:
STORAGE/TRANSFER be dispensed directlySTORAGE/TRANSFER
from BD
needle. An audible click may white reset button is not packaging. If you choose to
be heard. Lock shield into place obstructed by other fingers. dispense outside of BD packaging,
and inspect. DO NOT attempt please ensure that all green or
To reuse BD Vacutainer ®
black colored caps are facing up
to engage safety shield by
S: Pronto Holder, push white

I A L in container. Fit product loosely
pressing against a hard surface.
CTO
R resetECLIPSE™
button. NEEDLE in container.
PI
FER
/TR ANS
AGE
STOR STORAGE/TRANSFER PICTORIALS:
Precaution exposure to biologic samples
READY RESETSTORAGE/TRANSFER PICTORIALS:
• Precaution recommends that (e.g., through a puncture
reusable holders be cleaned injury) since samples may
afterECLIPSE™ NEEDLE
each use. BD in-house transmit viral hepatitis, HIV
ECLIPSE™ NEEDLE
testing has verified that the (AIDS), or other infectious
BD Pronto™ Holder can be diseases. Utilize any safety-
used for 10 cleaning cycles.† engineered feature if the
It is recommended that the blood collection device
hospital validates its own provides one. Discard all
Note
procedures for cleaning; the blood collection “sharps”
BD Vacutainer Pronto Holder
TORAGE/TRANSFER
® ™
PICTORIALS: above procedure may provide in biohazard containers
BD Technical Services
can be used for 100 uses* 1.800.268.5430
the basis for such procedures. approved for their disposal.
• If the holder is visibly
Procedure for Cleaning contaminated with blood and/
BD Customer Service
Reference
• Soak in 1:10 dilution or there is failure of product • BD has conducted an internal 1.866.979.9408
household bleach integrity, discard the holder. study and has determined
RIALS:solution for 5 minutes** • Handle all biologic samples that product performance up
www.bd.com/ca
• Agitate in the same and blood collection “sharps” to 10 cleaning cycles is not
solution using ultrasonic (lancets, needles, luer adapters, impacted when the holder is
bath for 5 minutes and blood collection sets) in cleaned after each use using
• Rinse with clean water accordance with the policies the above indicated cleaning
• Drain/Air Dry and procedures of your facility. procedure. Autoclaving or the
Obtain appropriate medical use of 100% bleach solution
attention in the event of any may impact the performance
of this device.

* Data on file at BD. Cleaning between uses was not considered in the determination of 100 uses and is recommended. 2771 Bristol Circle
** CLSI Guideline M29-A3. Oakville, ON L6H 6R5
† BD has conducted an internal study.
www.bd.com/ca