Liquichek Cardiac Markers Plus Control


Levels 1, 2 and 3
180 Trilevel 6 x 3 mL
Level 1 29891
181 Level 1 6 x 3 mL
2021-11-30 29890 Level 2 29892
182 Level 2 6 x 3 mL
Level 3 29893
183 Level 3 6 x 3 mL
180X MiniPak 3 x 3 mL

http://www.myeinserts.com/29890

Revision Date 2020-02-12 Indicates Revised Information

METHOD (1)
Level 1 - 29891 Level 2 - 29892 Level 3 - 29893 (2)
Units Mean Range Mean Range Mean Range
TROPONIN I
Abbott ARCHITECT iSystems (STAT) (Troponin I) ng/mL 1.32 0.897 – 1.74 10.2 7.73 – 12.7 32.4 21.8 – 43.1
Abbott ARCHITECT iSystems (STAT) (Troponin I hs) (REF 3P25) pg/mL 410 279 – 542 3717 2821 – 4613 9263 7062 – 11464
Beckman Coulter Access/2/2i (AccuTnI+3 REF A98143) (TnIA2) ng/mL 0.231 0.131 – 0.331 1.69 1.17 – 2.20 5.24 3.67 – 6.81
Beckman Coulter UniCel DxI Series (AccuTnI REF A78803) ng/mL 0.238 0.137 – 0.338 1.77 1.23 – 2.32 5.62 3.93 – 7.32
Beckman Coulter UniCel DxI Series (AccuTnI+3 REF A98264)
(TnIDxI) ng/mL 0.212 0.112 – 0.313 1.74 1.20 – 2.27 6.23 4.35 – 8.11
Monobind AccuBind ELISA (4) ng/mL <0.500 1.95 1.31 – 2.60 5.80 3.89 – 7.71
Monobind AccuLite CLIA (4) ng/mL <0.500 2.07 1.39 – 2.75 6.18 4.14 – 8.22
Ortho Vitros MicroWell Series (4) ng/mL 1.19 <0.012 – 3.63 3.96 2.86 – 5.06 12.9 9.30 – 16.5
Ortho Vitros MicroWell Series (hsTnI) (4) ng/L 385 270 – 499 3225 2267 – 4183 11660 7778 – 15540
Siemens ADVIA Centaur CP (TnI Ultra)(Rgt. Lot 010110 and
above) ng/mL 0.522 0.310 – 0.735 4.43 3.15 – 5.71 15.0 10.9 – 19.2
Siemens ADVIA Centaur Systems (TnI Ultra)(Rgt Lot 010110
and above) ng/mL 0.602 0.395 – 0.809 4.93 3.68 – 6.19 16.7 12.6 – 20.8
Siemens Dimension EXL (LOCI) ng/mL 0.464 0.319 – 0.609 2.77 2.13 – 3.41 6.94 5.37 – 8.50
Siemens Dimension Series (CTNI) ng/mL 0.222 0.092 – 0.353 1.76 1.17 – 2.34 4.39 3.02 – 5.76
Siemens Dimension Vista Systems (CTNI) (LOCI) ng/mL 0.415 0.275 – 0.556 2.52 2.01 – 3.04 6.47 5.25 – 7.69
Siemens IMMULITE / IMMULITE 1000 ng/mL
Siemens IMMULITE / IMMULITE 1000 Turbo ng/mL
Siemens IMMULITE 2000 / 2000 XPi ng/mL 0.789 0.511 – 1.07 4.89 3.52 – 6.26 18.1 13.2 – 22.9
Siemens Stratus CS ng/mL § § §
Snibe MAGLUMI ng/mL
TOSOH AIA Systems (cTnI 2nd Gen) (4) ng/mL 1.46 1.02 – 1.90 8.71 6.10 – 11.3 27.7 19.4 – 36.0
TOSOH AIA Systems (cTnI 3rd Gen) (4) ng/mL 1.44 0.860 – 2.01 10.8 8.12 – 13.5 32.7 24.5 – 40.8

FOOTNOTES
(1) All footnotes may not apply to your custom selected data chart.
(2) A dilution should be made prior to running those tests where the value listed exceeds the highest standard.
(3) Values were obtained at 37°C.
(4) The assigned values were determined using the reagent and/or instrument manufacturer’s protocol and may not represent ±3SD ranges.
(5) This product should be run with automatic dilution, the same as patient samples.
Data is not available at this time. Please inquire.
§ The data required to establish the means and acceptable ranges for this assay were not obtained due to limited assignment participation. If your facility is
interested in participating in the Value Assignment Program for this assay, please contact your local Bio-Rad office.