Perfusor ® compact S

The handy syringe pump


for paediatrics and neonatology

Wide selection of syringes for greater


flexibility in applications
Very low delivery rates for administration
of the smallest volumes
Perfusor ® compact S
Wide selection of syringes and high delivery
accuracy to match the requirements

Simple and safe Selection of syringe sizes Secure syringe fixation;


both in operation and from 2 to 50 ml meets fastest reaction times in
application individual requirements the event of an occlusion
With its performance profile, Up to 6 different syringe types Electronic monitoring of the
the Perfusor compact S fulfils can be used in the Perfusor syringe fixation and innovative
the special requirements in compact S. This selection can pressure sensors ensure greater
paediactrics and neonatology. be configured to meet specific safety when syringes are
– 10 self-explanatory keys allow user needs. exchanged and whenever
simple parameter entries – Syringe size freely selectable a pressure alarm occurs.
– Infusion rates from 0.01 to according to the particular – Automatic recognition of the
200.0 ml/h possible medication and the dose to syringe size used and of
be administered whether a syringe is correctly
– Automatic calculation of the
The advantages provided by inserted
infusion rate based on the
volume and time pre-selection the use of smaller syringes for – Fastest possible alarm reaction
Perfusor compact S – Stackable paediatrics and neonatology time for any syringe size in
– Either bolus administration
for building mobile therapy units can be optimally utilised. the event of an occlusion
that lock together
with simultaneous display
Battery operation up to 60 h of the delivery volume or ad- – Enhanced delivery – Automatic bolus reduction,
with inexpensive, easy-to-change ministration of pre-selected characteristics at infusion for re-starts with accurate
standard batteries
bolus volumes, simply at rates below 1.0 ml/h dosage following a pressure
Optional rechargeable battery pack alarm
for up to 10 hours of operation the touch of a key – Faster alarm reaction times
– Reducible occlusion in the event of an occlusion;
pressure limits for faster lower bolus volume within
alarm reaction times the system
– Data lock function; prevents
unauthorised changes to
pre-selected values

Performance Data
Classification: CF, defibrillation-proof, protection class II
Delivery performance
Delivery rates: 0.01 to 200.0 ml/h selectable Bolus delivery rate: 1,200 ml/h selectable
Accuracy of the set delivery rate: typ. ±2.5% (meas. time: >1 h, and delivery vol. >2 ml)
Function spectrum
Syringe selection configurable in service menu: 2/3, 5, 10, 20, 30 and 50/60 ml syringes;
automatic recognition of syringe size. Up to 6 syringe types can be selected by menu code.
Automatic calculation of the infusion rate: based on volume and time pre-selection
Bolus administration: as long as key is pressed or choice of pre-selectable volumes
Stand-by infusion pause: without time limit; pre-set values remain stored
Battery/rechargeable battery capacity display: in h and min Ordering Data Code No./REF
Safety system Perfusor ® compact S 110/120 V 8714886
Occlusion alarm pressure limits: 3 levels; 0.3, 0.6, 1.2 bar Perfusor ® compact S 220/240 V 8714843
Automatic reduction of the bolus volume within system following pressure alarm
Pre-alarms: 3 min. before syringe is empty/30 min. before battery/rech. battery is discharged MFC connecting lead for staff call 8711682
Data lock function: keys are blocked Staff call: max. 24 V/1 A/24 VA (VDE 0834) MFC connecting lead for ambulances (12 V) 8711674
Mains/battery operation, dimensions
___ MFC connecting lead
Rated vol.: 110–120 V/220–240 V, 50/60 Hz ~ External extra-low voltage: 12 V - - - DC
Battery type: 4 x 1.5 V DC alkaline-manganese Dimensions: (WxHxD): 190x100x120 mm for interface use, RS-232 with electrical isolation 8711681
Batt. oper. time: >80 h at a del. rate of ≤10 ml/h Weight: approx. 1.6 kg Rechargeable battery pack 8714991
Interfaces
Syringes that can be used: Omnifix® with lock fitting; 2, 5, 10, 20, 30, 50 ml
RS-232/MFC plug-in contact for fm anaesthesia/fm intensive workstations: Original-Perfusor®-Syringes 20 and 50 ml, Proinjekt syringe 50 ml
Data transmission, staff call, power supply and potential equalisation Further types chart of available syringes available on request
Warranty, technical audits
Warranty: 24 months as of the date of delivery Technical audits: 24-month intervals

C 0123

HOSPITAL CARE B. Braun Melsungen AG


P.O.Box 11 20
D-34209 Melsungen
Tel (0 56 61) 71-0
B. 04. 05. 01/4 Nr. 879 1759 Fax (0 56 61) 71-45 55
Printed on chlorine-free bleached cellulose using 50 % recycled paper www.bbraun.com