Changing CF card

Please kindly follow below procedures to finish.

1. Make sure the Hematology Analyzer is power off.

2. Use screw diver to take out screws of front cover and open the front cover.

Figure 1 Open the front cover

3. Push a button and take out the old CF card from main board.

Main board

Figure 2
Push this button and then take
out the CF card from main
board.
4. Insert the new CF card into main board.

5. Put the front cover back.

6. Restart the machine. The machine can work with new CF card.