WARNING

MRI
100 Gauss
field line

Place the ventilator no closer than the


100 Gauss field line (typically about
2.5 m/ 8.2 ft from the iso-center of the
magnet).
ƒ Use of the ventilator not in accordance
with these instructions may cause the
ventilator to malfunction.
ƒ Connect the potential equalization pin
on the back of the ventilator to ground
during battery (DC power) operation.
ƒ Remove the green oxygen inlet cover
and attached chain from the back of
the ventilator.
ƒ To avoid degrading image quality,
do not use the static mechanics
function during MRI scanning.
ƒ Lock the ventilator cart wheels to
prevent inadvertent movement.
ƒ VersaMed recommends using its
longer (3.8 m/12.5 ft) patient circuit
when operating in the MRI
environment.

VersaMed Medical Systems, Inc.


2 Blue Hill Plaza
Pearl River, NY 10965
USA
1800-475-9239
www.versamed.com

Part number 350D0165-A0 Rev: 05