Professional Documents
Culture Documents
ARTICLE 6
LIQUID PENETRANT EXAMINATION
163
No further reproduction or distribution is permitted.
ARTICLE 6 2013 SECTION V
Table T-621
Requirements of a Liquid Penetrant Examination Procedure
Nonessential
Requirement Essential Variable Variable
Identification of and any change in type or family group of penetrant materials including X ...
developers, emulsifiers, etc.
Surface preparation (finishing and cleaning, including type of cleaning solvent) X . . .
Method of applying penetrant X . . .
Method of removing excess surface penetrant X . . .
Hydrophilic or lipophilic emulsifier concentration and dwell time in dip tanks and agitation time X . . .
for hydrophilic emulsifiers
Hydrophilic emulsifier concentration in spray applications X . ..
Method of applying developer X . ..
Minimum and maximum time periods between steps and drying aids X . ..
Decrease in penetrant dwell time X . ..
Increase in developer dwell time (Interpretation Time) X . ..
Minimum light intensity X . ..
Surface temperature outside 40°F to 125°F (5°C to 52°C) or as previously qualified X . ..
Performance demonstration, when required X . ..
Personnel qualification requirements ... X
Materials, shapes, or sizes to be examined and the extent of examination ... X
Post examination cleaning technique ... X
164
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 6
air inlet to preclude contamination of the penetrant by oil, the mixture shall be removed by immersing in or rinsing
water, dirt, or sediment that may have collected in the with water. The temperature and pressure of the water
lines. shall be as recommended by the manufacturer.
NOTE: Additional information may be obtained from SE 165.
T-672 PENETRATION (DWELL) TIME
T-673.3 Solvent Removable Penetrants. Excess sol-
Penetration (dwell) time is critical. The minimum pene- vent removable penetrants shall be removed by wiping
tration time shall be as required in Table T-672 or as qua- with a cloth or absorbent paper, repeating the operation
lified by demonstration for specific applications. until most traces of penetrant have been removed. The re-
maining traces shall be removed by lightly wiping the sur-
T-673 EXCESS PENETRANT REMOVAL face with cloth or absorbent paper moistened with
After the specified penetration (dwell) time has elapsed, solvent. To minimize removal of penetrant from disconti-
any penetrant remaining on the surface shall be removed, nuities, care shall be taken to avoid the use of excess sol-
taking care to minimize removal of penetrant from vent. Flushing the surface with solvent, following the
discontinuities. application of the penetrant and prior to developing,
is prohibited.
T-673.1 Water-Washable Penetrants. Excess water-
washable penetrant shall be removed with a water spray. T-674 DRYING AFTER EXCESS PENETRANT
The water pressure shall not exceed 50 psi (350 kPa), and REMOVAL
the water temperature shall not exceed 110°F (43°C). (a) For the water washable or post-emulsifying techni-
T-673.2 Post-Emulsification Penetrants. que, the surfaces may be dried by blotting with clean ma-
(a) Lipophilic Emulsification. After the required pene- terials o r by using circulating air, provided the
trant dwell time, the excess surface penetrant shall be temperature of the surface is not raised above 125°F
emulsified by immersing or flooding the part with the (52°C).
emulsifier. Emulsification time is dependent on the type (b) For the solvent removable technique, the surfaces
of emulsifier and surface condition. The actual emulsifica- may be dried by normal evaporation, blotting, wiping, or
tion time shall be determined experimentally. After emul- forced air.
sification, the mixture shall be removed by immersing in
or rinsing with water. The temperature and pressure of T-675 DEVELOPING
the water shall be as recommended by the manufacturer. The developer shall be applied as soon as possible after
(b) Hydrophilic Emulsification. After the required pene- penetrant removal; the time interval shall not exceed that
trant dwell time and prior to emulsification, the parts shall established in the procedure. Insufficient coating thick-
be pre-rinsed with water spray using the same process as ness may not draw the penetrant out of discontinuities;
for water-washable penetrants. Pre-rinsing time shall not conversely, excessive coating thickness may mask
exceed 1 min. After pre-rinsing, the excess surface pene- indications.
trant shall be emulsified by immersing in or spraying with With color contrast penetrants, only a wet developer
hydrophilic emulsifier. Bath concentration shall be as re- shall be used. With fluorescent penetrants, a wet or dry
commended by the manufacturer. After emulsification, developer may be used.
Table T-672
Minimum Dwell Times
Dwell Times [Note (1)]
(minutes)
Material Form Type of Discontinuity Penetrant
Aluminum, magnesium, steel, brass and bronze, Castings and welds Cold shuts, porosity, lack of fusion, 5
titanium and high temperature alloys cracks (all forms)
Wrought materials Laps, cracks 10
extrusions, forgings, plate
Carbide tipped tools Brazed or welded Lack of fusion, porosity, cracks 5
Plastic All forms Cracks 5
Glass All forms Cracks 5
Ceramic All forms Cracks 5
NOTE:
(1) For temperature range from 50°F to 125°F (10°C to 52°C). For temperatures from 40°F (5°C) up to 50°F (10°C), minimum penetrant
dwell time shall be 2 times the value listed.
165
No further reproduction or distribution is permitted.
ARTICLE 6 2013 SECTION V
T-675.1 Dry Developer Application. Dry developer ensure adequate sensitivity during the examination and
shall be applied only to a dry surface by a soft brush, hand evaluation of indications. The light source, technique used,
powder bulb, powder gun, or other means, provided the and light level verification is required to be demonstrated
powder is dusted evenly over the entire surface being one time, documented, and maintained on file.
examined.
T-676.4 Fluorescent Penetrants. With fluorescent
T-675.2 Wet Developer Application. Prior to apply- penetrants, the process is essentially the same as in
ing suspension type wet developer to the surface, the de- T-676.3, with the exception that the examination is per-
veloper must be thoroughly agitated to ensure adequate formed using an ultraviolet light, called black light. The ex-
dispersion of suspended particles. amination shall be performed as follows:
(a) Aqueous Developer Application. Aqueous developer (a) It shall be performed in a darkened area.
may be applied to either a wet or dry surface. It shall be (b) Examiners shall be in a darkened area for at least
applied by dipping, brushing, spraying, or other means, 5 min prior to performing examinations to enable their
provided a thin coating is obtained over the entire surface eyes to adapt to dark viewing. Glasses or lenses worn by
being examined. Drying time may be decreased by using examiners shall not be photosensitive.
warm air, provided the surface temperature of the part (c) B l a c k l i g h t s s h a l l a c h i e v e a m i n i m u m o f
is not raised above 125°F (52°C). Blotting is not 1000 μW/cm2 on the surface of the part being examined
permitted. throughout the examination.
(b) Nonaqueous Developer Application. Nonaqueous de- (d) Reflectors and filters should be checked and, if ne-
velopers shall be applied by spraying, except where safety cessary, cleaned prior to use. Cracked or broken filters
or restricted access preclude it. Under such conditions, de- shall be replaced immediately.
veloper may be applied by brushing. For water-washable
(e) The black light intensity shall be measured with a
or post-emulsifiable penetrants, the developer shall be ap-
black light meter prior to use, whenever the light’s power
plied to a dry surface. For solvent removable penetrants,
source is interrupted or changed, and at the completion of
the developer may be applied as soon as practical after ex-
the examination or series of examinations.
cess penetrant removal. Drying shall be by normal
evaporation.
T-677 POST-EXAMINATION CLEANING
T-675.3 Developing Time. Developing time for final
interpretation begins immediately after the application When post-examination cleaning is required by the pro-
of a dry developer or as soon as a wet developer coating cedure, it should be conducted as soon as practical after
is dry. Evaluation and Documentation using a process that does
not adversely affect the part.
T-676 INTERPRETATION
T-676.1 Final Interpretation. Final interpretation
shall be made not less than 10 min nor more than
T-680 EVALUATION
60 min after the requirements of T-675.3 are satisfied. If (a) All indications shall be evaluated in terms of the ac-
bleed-out does not alter the examination results, longer ceptance standards of the referencing Code Section.
periods are permitted. If the surface to be examined is (b) Discontinuities at the surface will be indicated by
large enough to preclude complete examination within bleed-out of penetrant; however, localized surface irregu-
the prescribed or established time, the examination shall larities due to machining marks or other surface condi-
be performed in increments. tions may produce false indications.
T-676.2 Characterizing Indication(s). The type of (c) Broad areas of fluorescence or pigmentation which
discontinuities are difficult to evaluate if the penetrant dif- could mask indications of discontinuities are unaccepta-
fuses excessively into the developer. If this condition oc- ble, and such areas shall be cleaned and reexamined.
curs, close observation of the formation of indication(s)
during application of the developer may assist in charac-
terizing and determining the extent of the indication(s). T-690 DOCUMENTATION
T-676.3 Color Contrast Penetrants. With a color con- T-691 RECORDING OF INDICATIONS
trast penetrant, the developer forms a reasonably uniform
T-691.1 Nonrejectable Indications. Nonrejectable in-
white coating. Surface discontinuities are indicated by
dications shall be recorded as specified by the referencing
bleed-out of the penetrant which is normally a deep red
Code Section.
color that stains the developer. Indications with a light
pink color may indicate excessive cleaning. Inadequate T-691.2 Rejectable Indications. Rejectable indica-
cleaning may leave an excessive background making inter- tions shall be recorded. As a minimum, the type of indica-
pretation difficult. A minimum light intensity of 100 fc tions (linear or rounded), location and extent (length or
(1000 lx) is required on the surface to be examined to diameter or aligned) shall be recorded.
166
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 6
167
No further reproduction or distribution is permitted.
ARTICLE 6 2013 SECTION V
MANDATORY APPENDIX I
GLOSSARY OF TERMS FOR LIQUID PENETRANT EXAMINATION
(a) The following SE-1316 terms are used in conjunc- solvent removable penetrant: a type of penetrant used
tion with this Article: black light; bleedout; blotting; clean; where the excess penetrant is removed from the surface
contaminant; contrast; developer; developer, aqueous; of the part by wiping using a nonaqueous liquid.
168
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 6
MANDATORY APPENDIX II
CONTROL OF CONTAMINANTS FOR LIQUID PENETRANT
EXAMINATION
169
No further reproduction or distribution is permitted.
ARTICLE 6 2013 SECTION V
block “A” and the other block “B” for identification in sub-
3 in. (75 mm)
170
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 6
To qualify a procedure for temperatures above 125°F (a) When the single comparator block and photographic
(52°C), the upper and lower temperature limits shall be technique is used, the processing details (as applicable)
established and the procedure qualified at these tempera- described in III-641.1 and III-641.2 apply. The block shall
tures. [As an example, to qualify a procedure for the tem- be thoroughly cleaned between the two processing steps.
perature range 126°F (52°C) to 200°F (93°C), the Photographs shall be taken after processing at the non-
capability of a penetrant to reveal indications on the com- standard temperature and then after processing at the
parator shall be demonstrated at both temperatures.] standard temperature. The indication of cracks shall be
III-641.3 Alternate Techniques for Color Contrast compared between the two photographs. The same criter-
Penetrants. As an alternate to the requirements of ia for qualification as III-641.1 shall apply.
III-641.1 and III-641.2, when using color contrast pene- (b) Identical photographic techniques shall be used to
trants, it is permissible to use a single comparator block make the comparison photographs.
for the standard and nonstandard temperatures and to
make the comparison by photography.
171
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24
ARTICLE 24
LIQUID PENETRANT STANDARDS
543
No further reproduction or distribution is permitted.
INTENTIONALLY LEFT BLANK
544
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-129
SD-129
545
No further reproduction or distribution is permitted.
ARTICLE 24, SD-129 2013 SECTION V
SD-129
546
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-129
quantitative recovery of the liquids from the bomb may stable and representative of the samples of interest. These
be achieved readily. The inner surface of the bomb may QC samples can be used to check the validity of the testing
be made of stainless steel or any other material that will process as described in Section 10.
not be affected by the combustion process or products.
Materials used in the bomb assembly, such as the head
gasket and lead-wire insulation, shall be resistant to heat 6. Procedure
and chemical action, and shall not undergo any reaction 6.1 Preparation of Bomb and Sample — Cut a piece of
that will affect the sulfur content of the liquid in the bomb. firing wire 100 mm in length. Coil the middle section
4.2 Sample Cup, platinum, 24 mm in outside diameter (about 20 mm) and attach the free ends to the terminals.
at the bottom, 27 mm in outside diameter at the top, 12 mm Arrange the coil so that it will be above and to one side
in height outside, and weighing 10 to 11 g. of the sample cup. Insert between two loops of the coil a
wisp of cotton or nylon thread of such length that one end
4.3 Firing Wire, platinum, No. 26 B & S gage, 0.41 mm will extend into the sample cup. Place about 5 mL of
(16 thou), 27 SWG, or equivalent. (Warning — The switch Na2CO3 solution in the bomb (Note 2) and rotate the bomb
in the ignition circuit shall be of a type which remains in such a manner that the interior surface is moistened by
open, except when held in closed position by the operator.) the solution. Introduce into the sample cup the quantities
4.4 Ignition Circuit, capable of supplying sufficient cur- of sample and white oil (Note 3 and Note 4) specified in
rent to ignite the cotton wicking or nylon thread without the following table, weighing the sample to the nearest
melting the wire. The current shall be drawn from a step- 0.2 mg (when white oil is used, stir the mixture with a
down transformer or from a suitable battery. short length of quartz rod and allow the rod to remain in
the sample cup during the combustion).
4.5 Cotton Wicking or Nylon Sewing Thread, white.
NOTE 2 — After repeated use of the bomb for sulfur determinations, a
film may be noticed on the inner surface. This dullness can be removed
5. Reagents and Materials by periodic polishing of the bomb. A satisfactory method for doing this
5.1 Purity of Reagents — Reagent grade chemicals shall is to rotate the bomb in a lathe at about 300 rpm and polish the inside
surface with emery polishing papers Grit No. 2⁄0, or equivalent paper,
be used in all tests. Unless otherwise indicated, it is coated with a light machine oil to prevent cutting, and then with a paste
intended that all reagents shall conform to the specifications of grit-free chromic oxide and water. This procedure will remove all but
of the Committee on Analytical Reagents of the American very deep pits and put a high polish on the surface. Before the bomb is
Chemical Society, where such specifications are available. used it shall be washed with soap and water to remove oil or paste left
Other grades may be used, provided it is first ascertained from the polishing operation.
that the reagent is of sufficiently high purity to permit its 6.1.1 Warning — Do not use more than 1.0 g total
use without lessening the accuracy of the determination. of sample and white oil or other low sulfur combustible
5.2 Purity of Water — Unless otherwise indicated, refer- material or more than 0.8 g if the IP 12 bomb is used.
ences to water shall mean water as defined by Type II or Weight of Weight of
III of Specification D 1193. Sulfur Content, % Sample, g White Oil, g
5 or under 0.6 to 0.8 0.0
5.3 Barium Chloride Solution (85 g/litre) — Dissolve Over 5 0.3 to 0.4 0.3 to 0.4
100 g of barium chloride dihydrate (BaCl2 W 2H2O) in
distilled water and dilute to 1 liter. NOTE 3 — Use of sample weights containing over 20 mg of chlorine
5.4 Bromine Water (saturated). may cause corrosion of the bomb. To avoid this, it is recommended that
for samples containing over 2% chlorine, the sample weight be based on
5.5 Hydrochloric Acid (sp gr 1.19) — Concentrated the chlorine content as given in the following table:
hydrochloric acid (HCl). Weight of Weight of
5.6 Oxygen, free of combustible material and sulfur Chlorine Content, % Sample, g White Oil, g
compounds, available at a pressure of 41 kgf/cm2 (40 atm). 2 to 5 0.4 0.4
Over 5 to 10 0.2 0.6
5.7 Sodium Carbonate Solution (50 g/litre) — Dissolve Over 10 to 20 0.1 0.7
135 g of sodium carbonate decahydrate (Na2CO3 W 10H2O) Over 20 to 50 0.05 0.7
or its equivalent weight in distilled water and dilute to
1 litre. NOTE 4 — If the sample is not readily miscible with white oil, some
other low sulfur combustible diluent may be substituted. However, the
5.8 White Oil, USP, or Liquid Paraffin, BP, or equiv- combined weight of sample and nonvolatile diluent shall not exceed 1.0 g
alent. or more than 0.8 g if the IP 12 bomb is used.
5.9 Quality Control (QC) Samples, preferably are por- 6.2 Addition of Oxygen — Place the sample cup in posi-
tions of one or more liquid petroleum materials that are tion and arrange the cotton wisp or nylon thread so that
547
No further reproduction or distribution is permitted.
ARTICLE 24, SD-129 2013 SECTION V
the end dips into the sample. Assemble the bomb and or dropwise. Stir the solution during the addition and for
tighten the cover securely. (Warning — Do not add oxygen 2 min thereafter. Cover the beaker with a fluted watch
or ignite the sample if the bomb has been jarred, dropped, glass and continue boiling slowly until the solution has
or tilted.) Admit oxygen slowly (to avoid blowing the oil evaporated to a volume approximately 75 mL as indicated
from the cup) until a pressure is reached as indicated in by a mark on the beaker. Remove the beaker from the hot
the following table: plate (or other source of heat) and allow it to cool for 1 hr
before filtering. Filter the supernatant liquid through an
Capacity of Minimum Gage Maximum Gage ashless, quantitative filter paper (Note 5). Wash the precipi-
Bomb, ml Pressure,A kgf/cm2 (atm) Pressure,A kgf/cm2 (atm) tate with water, first by decantation and then on the filter,
300 to 350 39 (38) 41 (40) until free from chloride. Transfer the paper and precipitate
350 to 400 36 (35) 38 (37) to a weighed crucible and dry (Note 6) at a low heat until
400 to 450 31 (30) 33 (32)
the moisture has evaporated. Char the paper completely
450 to 500 28 (27) 30 (29)
without igniting it, and finally ignite at a bright red heat
A
The minimum pressures are specified to provide sufficient oxygen for until the residue is white in color. After ignition is com-
complete combustion and the maximum pressures represent a safety plete, allow the crucible to cool at room temperature, and
requirement. weigh.
6.3 Combustion — Immerse the bomb in a cold distilled- NOTE 5 — A weighed porcelain filter crucible (Selas type) of 5 to 9-m
water bath. Connect the terminals to the open electrical porosity may be used in place of the filter paper. In this case the precipitate
circuit. Close the circuit to ignite the sample. (Warning — is washed free of chloride and then dried to constant weight at 500 ±25°C.
Do not go near the bomb until at least 20 s after firing.) NOTE 6 — A satisfactory means of drying, charring, and igniting the
Remove the bomb from the bath after immersion for at paper and precipitate is to place the crucible containing the wet filter
paper in a cold electric muffle furnace and to turn on the current. Drying,
least 10 min. Release the pressure at a slow, uniform rate
charring, and ignition usually will occur at the desired rate.
such that the operation requires not less than 1 min. Open
the bomb and examine the contents. If traces of unburned 6.6 Blank — Make a blank determination whenever new
oil or sooty deposits are found, discard the determination reagents, white oil, or other low-sulfur combustible mate-
and thoroughly clean the bomb before again putting it in rial are used. When running a blank on white oil, use 0.3
use (Note 2). to 0.4 g and follow the normal procedure.
6.4 Collection of Sulfur Solution — Rinse the interior
of the bomb, the oil cup, and the inner surface of the bomb
cover with a fine jet of water, and collect the washings in 7. Calculation
a 600-mL beaker having a mark to indicate 75 mL. Remove 7.1 Calculate the sulfur content of the sample as
any precipitate in the bomb by means of a rubber police- follows:
man. Wash the base of the terminals until the washings
are neutral to the indicator methyl red. Add 10 mL of Sulfur, weight percent p (P − B)13.73/W (1)
saturated bromine water to the washings in the beaker.
where:
(The volume of the washings is normally in excess of
300 mL.) Place the sample cup in a 50-mL beaker. Add P p grams of BaSO4 obtained from sample,
5 mL of saturated bromine water, 2 mL of HCl, and enough B p grams of BaSO4 obtained from blank, and
water just to cover the cup. Heat the contents of the beaker W p grams of sample used.
to just below its boiling point for 3 or 4 min and add to
the beaker containing the bomb washings. Wash the sample
cup and the 50-mL beaker thoroughly with water. Remove 8. Report
any precipitate in the cup by means of a rubber policeman.
Add the washings from the cup and the 50-mL beaker, 8.1 Report the results of the test to the nearest 0.01%.
and the precipitate, if any, to the bomb washings in the
600-mL beaker. Do not filter any of the washings, since
filtering would remove any sulfur present as insoluble 9. Precision and Bias
material. 9.1 The precision of this test is not known to have been
6.5 Determination of Sulfur — Evaporate the combined obtained in accordance with currently accepted guidelines.
washings to 200 mL on a hot plate or other source of heat. 9.1.1 Repeatability — The difference between two
Adjust the heat to maintain slow boiling of the solution test results, obtained by the same operator with the same
and add 10 mL of the BaCl2 solution, either in a fine stream apparatus under constant operating conditions on identical
548
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-129
test material, would in the long run, in the normal and (a) Results of duplicate tests should not differ by more than the follow-
correct operation of the test method, exceed the following ing amounts:
Repeatability Reproducibility
values only in one case in twenty:
0.016 x + 0.06 0.037 x + 0.13
9.1.2 Reproducibility — The difference between two where x is the mean of duplicate test results.
single and independent results obtained by different opera- (b) These precision values were obtained in 1960 by statistical exami-
tors working in different laboratories on identical test mate- nation of interlaboratory test results. No limits have been established for
rial would, in the long run, in the normal and correct additive concentrates.
operation of the test method, exceed the following values 9.2 Bias — Results obtained in one laboratory by Test
only in one case in twenty: Method D 129 on NIST Standard Reference Material Nos.
1620A, 1621C, and 1662B were found to be 0.05 mass %
Sulfur, higher than the accepted reference values.
Weight % Repeatability Reproducibility
0.1 to 0.5 0.04 0.05
0.5 to 1.0 0.06 0.09 10. Quality Control
1.0 to 1.5 0.08 0.15 10.1 Confirm the performance of the instrument or the
1.5 to 2.0 0.12 0.25
2.0 to 5.0 0.18 0.27
test procedure by analyzing a QC sample (see 5.9).
10.1.1 When QC/Quality Assurance (QA) protocols
are already established in the testing facility, these may be
NOTE 7 — The precision shown in the above table does not apply to used to confirm the reliability of the test result.
samples containing over 2% chlorine because an added restriction on the
amount of sample which can be ignited is imposed. 10.1.2 When there is no QC/QA protocol established
in the testing facility, Appendix X1 can be used as the
NOTE 8 — This test method has been cooperatively tested only in the QC/QA system.
range of 0.1 to 5.0% sulfur.
NOTE 9 — The following information on the precision of this method 11. Keywords
has been developed by the Institute of Petroleum (London): 11.1 bomb; sulfur
549
No further reproduction or distribution is permitted.
ARTICLE 24, SD-129 2013 SECTION V
APPENDIX
(Nonmandatory Information)
X1. QUALITY CONTROL on the criticality of the quality being measured, the demon-
strated stability of the testing process, and customer
X1.1 Confirm the performance of the instrument or the
requirements. Generally, a QC sample is analyzed each
test procedure by analyzing a quality control (QC) sample.
testing day with routine samples. The QC frequency should
X1.2 Prior to monitoring the measurement process, the be increased if a large number of samples are routinely
user of the test method needs to determine the average analyzed. However, when it is demonstrated that the testing
is under statistical control, the QC testing frequency may
value and control limits of the QC sample (see Practice
be reduced. The QC sample precision should be checked
D 6299 and MNL 7).
against the ASTM method precision to ensure data quality.
X1.3 Record the QC results and analyze by control X1.5 It is recommended that, if possible, the type of
charts or other statistically equivalent techniques to ascer- QC sample that is regularly tested be representative of the
tain the statistical control status of the total testing process material routinely analyzed. An ample supply of QC sam-
(see Practice D 6299 and MNL 7). Any out-of-control data ple material should be available for the intended period
should trigger investigation for root cause(s). of use, and must be homogenous and stable under the
anticipated storage conditions. See Practice D 6299 and
X1.4 In the absence of explicit requirements given in MNL 7 for further guidance on QC and control charting
the test method, the frequency of QC testing is dependent techniques.
550
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-516
SD-516
551
No further reproduction or distribution is permitted.
ARTICLE 24, SD-516 2013 SECTION V
SD-516
2. Referenced Documents
5. Significance and Use
2.1 ASTM Standards:
5.1 The determination of sulfate is important because
D 1066 Practice for Sampling Steam it has been reported that when this ion is present in excess
of about 250 mg/L in drinking water, it causes a cathartic
D 1129 Terminology Relating to Water
action (especially in children) in the presence of sodium
D 1192 Guide for Equipment for Sampling Water and and magnesium, and gives a bad taste to the water.
Steam in Closed Conduits
D 1193 Specification for Reagent Water 6. Interferences
D 2777 Practice for Determination of Precision and Bias 6.1 Insoluble suspended matter in the sample must be
of Applicable Methods of Committee D19 on Water removed. Dark colors that can not be compensated for in
the procedure interfere with the measurement of suspended
D 3370 Practices for Sampling Water from Closed Con- barium sulfate (BaSO4).
duits
6.2 Polyphosphates as low as 1 mg/L will inhibit barium
D 5847 Practice for Writing Quality Control Specifications sulfate precipitation causing a negative interference. Phos-
for Standard Test Methods of Water Analysis phonates present in low concentrations, depending on the
552
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-516
type of phosphonate, will also cause a negative interfer- of the test method. Type II water was specified at the time
ence. Silica in excess of 500 mg/L may precipitate along of round robin testing of this test method.
with the barium sulfate causing a positive interference.
8.3 Barium Chloride — Crystals of barium chloride
Chloride in excess of 5000 mg/L will cause a negative
(BaCl2W2H2O) screened to 20 to 30 mesh. To prepare in
interference. Aluminum, polymers, and large quantities of
the laboratory, spread crystals over a large watch glass,
organic material present in the test sample may cause the
desiccate for 24 h, screen to remove any crystals that are
barium sulfate to precipitate nonuniformly. In the presence
not 20 to 30 mesh, and store in a clean, dry jar.
of organic matter certain bacteria may reduce sulfate to
sulfide. To minimize the action of sulfate reducing bacteria, 8.4 Conditioning Reagent — Place 30 mL of concen-
samples should be refrigerated at 4°C when the presence trated hydrochloric acid (HCl, sp gr 1.19), 300 mL reagent
of such bacteria is suspected. water, 100 mL 95% ethanol or isopropanol and 75 g sodium
chloride (NaCl) in a container. Add 50 mL glycerol and
6.3 Although other ions normally found in water do mix.
not appear to interfere, the formation of the barium sulfate
suspension is very critical. Determinations that are in doubt 8.5 Sulfate Solution, Standard (1 mL p 0.100 mg
may be checked by a gravimetric method in some cases, SO4− −) — Dissolve 0.1479 g of anhydrous sodium sulfate
or by the procedure suggested in Note 2. (Na2SO4) in water, and dilute with water to 1 L in a volu-
metric flask. A purchased stock solution of adequate purity
is also acceptable.
7. Apparatus
7.1 Photometer — One of the following which are 9. Sampling
given in order of preference. 9.1 Collect the sample in accordance with Practice
7.1.1 Nephelometer or turbidimeter; D 1066, Specification D 1192, and Practices D 3370, as
applicable.
7.1.2 Spectrophotometer for use at 420 nm with light
path of 4 cm to 5 cm;
7.1.3 Filter photometer with a violet filter having a 10. Calibration
maximum near 420 nm and a light path of 4 cm to 5 cm. 10.1 Follow the procedure given in Section 11, using
appropriate amounts of the standard sulfate solution pre-
7.2 Stopwatch, if the magnetic stirrer is not equipped
pared in accordance with 8.5 and prepare a calibration
with an accurate timer.
curve showing sulfate ion content in milligrams per litre
7.3 Measuring Spoon, capacity 0.2 mL to 0.3 mL. plotted against the corresponding photometer readings
(Note 1). Prepare standards by diluting with water 0.0, 5.0,
7.4 Filter photometers and photometric practices pre-
10.0, 15.0, 20.0, 30.0, and 40.0 mL of standard sulfate
scribed in this test method shall conform to Practice E 60;
solution to 100-mL volumes in volumetric flasks. These
spectrophotometer practices shall conform to Practice
solutions will have sulfate ion concentrations of 0.0, 5.0,
E 275.
10.0, 15.0, 20.0, 30.0, and 40.0 mg/L (ppm), respectively.
NOTE 1: A separate calibration curve must be prepared for each photome-
ter and a new curve must be prepared if it is necessary to change the
8. Reagents
cell, lamp, or filter, or if any other alterations of instrument or reagents
8.1 Purity of Reagents — Reagent grade chemicals are made. Check the curve with each series of tests by running two or
shall be used in all tests. Unless otherwise indicated, it is more solutions of known sulfate concentrations.
intended that all reagents shall conform to the specifications
of the Committee on Analytical Reagents of the American
Chemical Society. Other grades may be used, provided it 11. Procedure
is first ascertained that the reagent is of sufficiently high 11.1 Filter the sample if it is turbid through a 0.45-m
purity to permit its use without lessening the accuracy of membrane and adjust the temperature to between 15°C
the determination. and 30°C.
8.2 Purity of Water — Unless otherwise indicated, 11.2 Pipette into a 250-mL beaker 100 mL or less of
reference to water shall be understood to mean reagent the clear sample containing between 0.5 mg and 4 mg of
water conforming to Specification D 1193, Type I. Other sulfate ion (Note 2). Dilute to 100 mL with water if
reagent water types may be used provided it is first ascer- required, and add 5.0 mL of conditioning reagent (Note 1).
tained that the water is of sufficiently high purity to permit NOTE 2: The solubility of BaSO4 is such that difficulty may be experi-
its use without adversely affecting the precision and bias enced in the determination of sulfate concentrations below about 5 mg/L
553
No further reproduction or distribution is permitted.
ARTICLE 24, SD-516 2013 SECTION V
TABLE 1 TABLE 2
OVERALL (ST) AND SINGLE-OPERATOR (SO) OVERALL (ST) AND SINGLE-OPERATOR (SO)
STANDARD DEVIATIONS AGAINST MEAN STANDARD DEVIATIONS AGAINST MEAN
CONCENTRATION FOR INTERLABORATORY RECOVERY CONCENTRATION FOR INTERLABORATORY RECOVERY
OF SULFATE FROM REAGENT WATERA OF SULFATE FROM DRINKING, GROUND, AND
SURFACE WATERA
Mean Concentration (x–), Standard Deviation, mg/L
mg/L ST SO Mean Concentration (–x), Standard Deviation, mg/L
mg/L ST SO
6.6 0.5 0.1
20.4 1.0 0.4 6.9 0.7 0.5
63.7 2.5 1.3 20.2 2.2 1.8
63.3 4.5 1.6
NOTE:
A
The test method is linear to 40 mg/L. Testing at the 63.9 level was NOTE:
accomplished through dilution as described in 11.2. A
The test method is linear to 40 mg/L. Testing at the 63.9 level was
accomplished through dilution as described in 11.2.
11.5 Stir exactly 1.0 min at constant speed. Reagent water 20.8 20.4 -0.4 -1.9 % no
63.9A 63.7A -0.2 -0.2 % no
NOTE 3: The stirring should be at a constant rate in all determinations. 7.0 6.6 -0.4 -5.3 % no
The use of a magnetic stirrer has been found satisfactory for this purpose. Drinking, ground 20.8 20.2 -0.6 -2.7 % no
and surface water 63.9A 63.3A -0.6 -0.9 % no
11.6 Immediately after the stirring period has ended, 7.0 6.9 -0.1 -1.8 % no
pour solution into the cell and measure the turbidity at
30-s intervals for 4 min. Record the maximum reading NOTE:
A
obtained in the 4-min period. The test method is linear to 40 mg/L. Testing at the 63.9 level was
accomplished through dilution as described in 11.2.
11.7 If the sample contains color or turbidity, run a
sample blank using the procedure 11.2 through 11.6 with-
out the addition of the barium chloride.
and Table 2 for drinking, ground, and surface waters.
11.8 If interferences are suspected, dilute the sample
13.2.1 Seven laboratories participated in the round
with an equal volume of water, and determine the sulfate
robin at three levels in triplicate, making a total of 21
concentration again. If the value so determined is one-half
observations at each level for reagent water and for matrix
that in the undiluted sample, interferences may be assumed
water (drinking, ground, and surface water).
to be absent.
13.3 Recoveries of known amounts of sulfate from
reagent water and drinking, ground, and surface waters are
12. Calculation as shown in Table 3.
12.1 Convert the photometer readings obtained with
13.3.1 A table for estimating the bias of the test
the sample to milligrams per litre sulfate ion (SO4− −) by
method through its applicable concentration range can be
use of the calibration curve described in Section 10. found in Table 4.
13.3.2 These collaborative test data were obtained on
13. Precision and Bias reagent grade water and natural waters. For other matrices,
13.1 The precision and bias data presented in this test these data may not apply.
method meet the requirements of Practice D 2777–86.
13.4 Precision and bias for this test method conforms
13.2 The overall and single-operator precision of the to Practice D 2777–86, which was in place at the time of
test method, within its designated range, varies with the collaborative testing. Under the allowances made in 1.4 of
quantity being tested according to Table 1 for reagent water D 2777–06, these precision and bias data do meet existing
554
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-516
555
No further reproduction or distribution is permitted.
ARTICLE 24, SD-516 2013 SECTION V
APPENDIX
(Nonmandatory Information)
556
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-808
SD-808
557
No further reproduction or distribution is permitted.
ARTICLE 24, SD-808 2013 SECTION V
SD-808
558
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-808
559
No further reproduction or distribution is permitted.
ARTICLE 24, SD-808 2013 SECTION V
560
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SD-808
561
No further reproduction or distribution is permitted.
ARTICLE 24, SD-808 2013 SECTION V
APPENDIX
(Nonmandatory Information)
X1. QUALITY CONTROL on the criticality of the quality being measured, the demon-
strated stability of the testing process, and customer
X1.1 Confirm the performance of the instrument or the
requirements. Generally, a QC sample is analyzed each
test procedure by analyzing a QC sample.
testing day with routine samples. The QC frequency should
X1.2 Prior to monitoring the measurement process, the be increased if a large number of samples are routinely
user of the method needs to determine the average value analyzed. However, when it is demonstrated that the testing
is under statistical control, the QC testing frequency may
and control limits of the QC sample (see Practice D 6299
be reduced. The QC sample precision should be checked
and MNL 7).
against the ASTM method precision to ensure data quality.
X1.3 Record the QC results and analyze by control X1.5 It is recommended that, if possible, the type of
charts or other statistically equivalent techniques to ascer- QC sample that is regularly tested be representative of the
tain the statistical control status of the total testing process material routinely analyzed. An ample supply of QC sam-
(see Practice D 6299 and MNL 7). Any out-of-control data ple material should be available for the intended period
should trigger investigation for root cause(s). of use, and must be homogenous and stable under the
anticipated storage conditions. See Practice D 6299 and
X1.4 In the absence of explicit requirements given in MNL 7 for further guidance on QC and Control Charting
the test method, the frequency of QC testing is dependent techniques.
562
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SE-165
SE-165
563
No further reproduction or distribution is permitted.
ARTICLE 24, SE-165 2013 SECTION V
SE-165
564
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SE-165
2.2 ASNT Documents: NOTE 2: Fluorescent penetrant examination shall not follow a visible
penetrant examination unless the procedure has been qualified in accor-
SNT-TC-1A Recommended Practice for Nondestructive dance with 10.2, because visible dyes may cause deterioration or quench-
Testing Personnel Qualification and Certification ing of fluorescent dyes.
ANSI/ASNT CP-189 Standard for Qualification and Certi- 4.2 Processing parameters, such as surface precleaning,
fication of Nondestructive Testing Personnel penetration dwell time and excess penetrant removal meth-
2.3 Military Standard: ods, are dependent on the specific materials used, the nature
of the part under examination (that is, size, shape, surface
MIL-STD-410 Nondestructive Testing Personnel Qualifi- condition, alloy), and type of discontinuities expected.
cation and Certification
2.4 APHA Standard: 5. Significance and Use
429 Method for the Examination of Water and Wastewater 5.1 Liquid penetrant examination methods indicate the
presence, location and, to a limited extent, the nature and
2.5 AIA Standard: magnitude of the detected discontinuities. Each of the vari-
NAS-410 Certification and Qualification of Nondestructive ous methods has been designed for specific uses such as
Test Personnel critical service items, volume of parts, portability or local-
2.6 SAE Standards: ized areas of examination. The method selected will depend
accordingly on the design and service requirements of the
AMS 2644 Inspection Material, Penetrant parts or materials being tested.
QPL-AMS-2644 Qualified Products of Inspection Materi-
als, Penetrant 6. Classification of Penetrant Materials and
Methods
3. Terminology 6.1 Liquid penetrant examination methods and types
3.1 The definitions relating to liquid penetrant examina- are classified in accordance with MIL-I-25135 and
tion, which appear in Terminology E 1316, shall apply to AMS 2644 and listed in Table 1.
the terms used in this standard. 6.2 Fluorescent Penetrant Testing (Type 1) — Fluores-
cent penetrant testing utilizes penetrants that fluoresce bril-
4. Summary of Test Method liantly when excited by black light (UVA). The sensitivity
of fluorescent penetrants depends on their ability to be
4.1 Liquid penetrant may consist of visible or fluores-
retained in the various size discontinuities during pro-
cent material. The liquid penetrant is applied evenly over
cessing, then to bleed out into the developer coating and
the surface being examined and allowed to enter open
produce indications that will fluoresce. Fluorescent indica-
discontinuities. After a suitable dwell time, the excess sur-
tions are many times brighter than their surroundings when
face penetrant is removed. A developer is applied to draw
viewed under black light illumination.
the entrapped penetrant out of the discontinuity and stain
the developer. The test surface is then examined to deter- 6.3 Visible Penetrant Testing (Type 2) — Visible pene-
mine the presence or absence of indications. trant testing uses a penetrant that can be seen in visible
565
No further reproduction or distribution is permitted.
ARTICLE 24, SE-165 2013 SECTION V
light. The penetrant is usually red, so that the indications 7.3 Emulsifiers:
produce a definite contrast with the white background of 7.3.1 Lipophilic Emulsifiers are oil-miscible liquids
the developer. Visible penetrant indications must be viewed used to emulsify the post-emulsified penetrant on the sur-
under adequate white light. face of the part, rendering it water-washable. The individ-
ual characteristics of the emulsifier and penetrant, and the
7. Types of Materials geometry/surface roughness of the part material contribute
to determining the emulsification time.
7.1 Liquid penetrant examination materials consist of
fluorescent and visible penetrants, emulsifiers (oil-base and 7.3.2 Hydrophilic Emulsifiers are water-miscible liq-
water-base), removers (water and solvent) and developers uids used to emulsify the excess post-emulsified penetrant
(dry powder, aqueous and nonaqueous). A family of liquid on the surface of the part, rendering it water-washable.
penetrant examination materials consists of the applicable These water-base emulsifiers (detergent-type removers) are
penetrant and emulsifier, as recommended by the manufac- supplied as concentrates to be diluted with water and used
turer. Any liquid penetrant, remover, and developer listed as a dip or spray. The concentration, use, and maintenance
in QPL-25135/QPL-AMS2644 can be used, regardless of shall be in accordance with manufacturer’s recommenda-
the manufacturer. Intermixing of penetrants and emulsifiers tions.
from different manufacturers is prohibited. 7.3.2.1 Hydrophilic emulsifiers function by dis-
placing the excess penetrant film from the surface of the
NOTE 3: Refer to 9.1 for special requirements for sulfur, halogen and
alkali metal content. part through detergent action. The force of the water spray
or air/mechanical agitation in an open dip tank provides
NOTE 4: While approved penetrant materials will not adversely affect the scrubbing action while the detergent displaces the film
common metallic materials, some plastics or rubbers may be swollen or of penetrant from the part surface. The individual charac-
stained by certain penetrants.
terisitics of the emulsifier and penetrant, and the geometry
7.2 Penetrants: and surface roughness of the part material contribute to
7.2.1 Post-Emulsifiable Penetrants are insoluble in determining the emulsification time. Emulsification con-
water and cannot be removed with water rinsing alone. centration shall be monitored weekly using a suitable
They are formulated to be selectively removed from the refractometer.
surface using a separate emulsifier. Properly applied and 7.4 Solvent Removers. Solvent removers function by
given a proper emulsification time, the emulsifier combines dissolving the penetrant, making it possible to wipe the
with the excess surface penetrant to form a water-washable surface clean and free of excess penetrant.
mixture, which can be rinsed from the surface, leaving the
surface free of excessive fluorescent background. Proper 7.5 Developers. Developers form a translucent or white
emulsification time must be experimentally established and absorptive coating that aids in bringing the penetrant out
maintained to ensure that over-emulsification does not of the surface discontinuities through blotting action, thus
occur, resulting in loss of indications. increasing the visibility of the indications.
7.2.2 Water-Washable Penetrants are formulated to 7.5.1 Dry Powder Developers. Dry powder devel-
be directly water-washable from the surface of the test opers are used as supplied, that is, free-flowing, non-caking
part, after a suitable penetrant dwell time. Because the powder (see 8.8.2). Care should be taken not to contaminate
emulsifier is “built-in,” water-washable penetrants can be the developer with fluorescent penetrant, as the contami-
washed out of discontinuities if the rinsing step is too long nant developer specks can appear as penetrant indications.
or too vigorous. It is therefore extremely important to 7.5.2 Aqueous Developers. Aqueous developers are
exercise proper control in the removal of excess surface normally supplied as dry powder particles to be either
penetrant to ensure against overwashing. Some penetrants suspended (water suspendable) or dissolved (water soluble)
are less resistant to overwashing than others, so caution in water. The concentration, use, and maintenance shall
should be exercised. be in accordance with manufacturer’s recommendations.
Water soluble developers shall not be used with Type 2
7.2.3 Solvent-Removable Penetrants are formulated
penetrants or Type 1, Method A penetrants.
so that excess surface penetrant can be removed by wiping
until most of the penetrant has been removed. The NOTE 5: Aqueous developers may cause stripping of indications if not
remaining traces should be removed with the solvent properly applied and controlled. The procedure should be qualified in
remover (see 8.6.4). To prevent removal of penetrant from accordance with 10.2.
discontinuities, care should be taken to avoid the use of 7.5.3 Nonaqueous Wet Developers. Nonaqueous wet
excess solvent. Flushing the surface with solvent to remove developers are supplied as suspensions of developer parti-
the excess penetrant is prohibited as the penetrant indica- cles in a nonaqueous solvent carrier ready for use as sup-
tors could easily be washed away. plied. Nonaqueous, wet developers are sprayed on to form
566
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SE-165
a thin coating on the surface of the part when dried. The be clean and dry before the penetrant is applied. If only a
thin coating serves as the developing medium. section of a part, such as a weld, including the heat affected
zone is to be examined, all contaminants shall be removed
NOTE 6: This type of developer is intended for application by spray only.
from the area being examined as defined by the contracting
7.5.4 Liquid Film Developers are solutions or colloi- parties. “Clean” is intended to mean that the surface must
dal suspensions of resins/polymer in a suitable carrier. be free of rust, scale, welding flux, weld spatter, grease,
These developers will form a transparent or translucent paint, oily films, dirt, and so forth, that might interfere
coating on the surface of the part. Certain types of film with the penetrant process. All of these contaminants can
developer may be stripped from the part and retained for prevent the penetrant from entering discontinuities (see
record purposes (see 8.8.4). Annex or Cleaning of Parts and Materials).
8.4.1 Drying After Cleaning — It is essential that the
8. Procedure surface of parts be thoroughly dry after cleaning, since any
8.1 The following processing parameters apply to both liquid residue will hinder the entrance of the penetrant.
fluorescent and visible penetrant testing methods. Drying may be accomplished by warming the parts in
drying ovens, with infrared lamps, forced hot air, or expo-
8.2 Temperature Limits — The temperature of the pene- sure to ambient temperature.
trant materials and the surface of the part to be processed
shall be between 40°F and 125°F (4°C and 52°C) or the NOTE 9: Residues from cleaning processes such as strong alkalies, pick-
ling solutions and chromates, in particular, may adversely react with the
procedure must be qualified at the temperature used as
penetrant and reduce its sensitivity and performance.
described in 10.2.
8.5 Penetrant Application — After the part has been
8.3 Examination Sequence — Final penetrant examina-
cleaned, dried, and is within the specified temperature
tion shall be performed after the completion of all opera-
range, the penetrant is applied to the surface to be examined
tions that could cause surface-connected discontinuities or
so that the entire part or area under examination is com-
operations that could expose discontinuities not previously
pletely covered with penetrant. Application methods
open to the surface. Such operations include, but are not
include dipping, brushing, flooding, or spraying. Small
limited to, grinding, welding, straightening, machining, and
parts are quite often placed in suitable baskets and dipped
heat treating. Satisfactory inspection results can usually be
into a tank of penetrant. On larger parts, and those with
obtained on surfaces in the as-welded, as-rolled, as-cast,
complex geometries, penetrant can be applied effectively
as-forged, or ceramics in the densified condition.
by brushing or spraying. Both conventional and electro-
8.3.1 Surface Treatment — Final penetrant examina- static spray guns are effective means of applying liquid
tion may be performed prior to treatments that can smear penetrants to the part surfaces. Not all penetrant materials
the surface but not by themselves cause surface discontinu- are suitable for electrostatic spray applications, so tests
ities. Such treatments include, but are not limited to, vapor should be conducted prior to use. Electrostatic spray appli-
blasting, deburring, sanding, buffing, sandblasting, or lap- cation can eliminate excess liquid build-up of penetrant on
ping. Performance of final penetrant examination after such the part, minimize overspray, and minimize the amount
surface treatments necessitates that the part(s) be etched of penetrant entering hollow-cored passages which might
to remove smeared metal from the surface prior to testing serve as penetrant reservoirs, causing severe bleedout prob-
unless otherwise agreed by the contracting parties. Note lems during examination. Aerosol sprays are conveniently
that final penetrant examination shall always precede sur- portable and suitable for local application.
face peening.
NOTE 10: With spray applications, it is important that there be proper
NOTE 7: Sand or shot blasting can close discontinuities so extreme ventilation. This is generally accomplished through the use of a properly
care should be taken to avoid masking discontinuities. under certain designed spray booth and exhaust system.
circumstances, however, grit blasting with certain air pressures and/or
8.5.1 Penetrant Dwell Time — After application,
mediums may be acceptable without subsequent etching when agreed by
the contracting parties. allow excess penetrant to drain from the part (care should
be taken to prevent pools of penetrant from forming on
NOTE 8: Surface preparation of structural or electronic ceramics for the part), while allowing for proper penetrant dwell time
penetrant testing by grinding, sand blasting, and etching is not recom- (see Table 2). The length of time the penetrant must remain
mended because of the potential for damage.
on the part to allow proper penetration should be as recom-
8.4 Precleaning — The success of any penetrant exami- mended by the penetrant manufacturer. Table 2, however,
nation procedure is greatly dependent upon the surrounding provides a guide for selection of penetrant dwell times for
surface and discontinuity being free of any contaminant a variety of materials, forms, and types of discontinuity.
(solid or liquid) that might interfere with the penetrant Unless otherwise specified, the dwell time shall not exceed
process. All parts or areas of parts to be examined must the maximum recommended by the manufacturer.
567
No further reproduction or distribution is permitted.
ARTICLE 24, SE-165 2013 SECTION V
TABLE 2
RECOMMENDED MINIMUM DWELL TIMES
Aluminum, magnesium, steel, brass castings and welds cold shuts, porosity, lack of fusion, 5 10
and bronze, titanium and high- cracks (all forms)
temperature alloys
wrought materials — extrusions, laps, cracks (all forms) 10 10
forgings, plate
Carbide-tipped tools lack of fusion, porosity, cracks 5 10
Plastic all forms cracks 5 10
Glass all forms cracks 5 10
Ceramic all forms cracks, porosity 5 10
A
For temperature range from 50 to 100°F (10 to 38°C) for fluorescent penetrants and 50 to 125°F (10 to 52°C) for visible penetrant.
B
Maximum penetrant dwell time in accordance with 8.5.2.
C
Development time begins as soon as wet developer coating has dried on surface of parts (recommended minimum). Maximum development
time in accordance with 8.8.6.
8.6 Penetrant Removal emulsifier shall not be agitated while being immersed. After
8.6.1 Water Washable (Method A): application of the emulsifier, the parts shall be drained and
positioned in a manner that prevents the emulsifier from
8.6.1.1 Removal of Excess Penetrants — After the
pooling on the part(s).
required penetrant dwell time, the excess penetrant on the
surface being examined must be removed with water. It 8.6.2.2 Emulsification Time — The emulsification
can be removed manually by a coarse spray or wiping the time begins as soon as the emulsifier is applied. The length
aprt surface with a dampened rag, automatic or of time that the emulsifier is allowed to remain on a part
semi-automatic water-spray equipment, or by water immer- and in contact with the penetrant is dependent on the type
sion. For immersion rinsing, parts are completely immersed of emulsifier employed and the surface roughness. Nominal
in the water bath with air or mechanical agitation. emulsification time should be as recommended by the man-
(a) The temperature of the water shall be maintained ufacturer. The actual emulsification time must be deter-
within the range of 50°F to 100°F (10°C to 38°C). mined experimentally for each specific application. The
(b) Spray-rinse water pressure should not exceed 40 psi surface finish (roughness) of the part is a significant factor
(275 kPa). When hydro-air pressure spray guns are used, in the selection of and in the emulsification time of an
the air pressure should not exceed 25 psi (172 kPa). emulsifier. Contact time shall be kept to the minimum time
to obtain an acceptable background and shall not exceed
NOTE 11: Overwashing should be avoided. Excessive washing can cause three minutes.
penetrant to be washed out of discontinuities. With fluorescent penetrant
methods perform the rinsing operation under black light so that it can 8.6.2.3 Post Rinsing — Effective post rinsing of the
be determined when the surface penetrant has been adequately removed. emulsified penetrant from the surface can be accomplished
8.6.1.2 Removal by Wiping (Method C) — After using either manual, semi-automated, or automated water
the required penetrant dwell time, the excess penetrant is immersion or spray equipment or combinations thereof.
removed by wiping with a dry, clean, lint-free cloth/towel. 8.6.2.4 Immersion — For immersion post rinsing,
Then use a clean lint-free cloth/towel lightly moistened parts are completely immersed in the water bath with air
with water or solvent to remove the remaining traces of or mechanical agitation. The amount of time the part is in
surface penetrant as determined by examination under the bath should be the minimum required to remove the
black light for fluorescent methods and visible light for emulsified penetrant. In addition, the temperature range of
visible methods. the water should be 50°F to 100°F (10°C to 38°C). Any
8.6.2 Lipophilic Emulsification (Method B): necessary touch-up rinse after an immersion rinse shall
8.6.2.1 Application of Lipophilic Emulsifier — meet the requirements of 8.6.2.5.
After the required penetration dwell time, the excess pene- 8.6.2.5 Spray Post Rinsing — Effective post rins-
trant on the part must be emulsified by immersing or flood- ing following emulsification can also be accomplished by
ing the parts with the required emulsifier (the emulsifier either manual or automatic water spray rinsing. The water
combines with the excess surface penetrant and makes the temperature shall be between 50°F and 100°F (10°C and
mixture removable by water rinsing). Lipophililc emulsifier 38°C). The water spray pressure shall not exceed 40 psi
shall not be applied by spray or brush and the part or (275 kPa) when manual spray guns are used. When
568
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SE-165
hydro-air pressure spray guns are used, the air pressure emulsify the residual penetrant on part surfaces to render
should not exceed 25 psi (172 kPa). it water-washable. The concentration of the emulsifier for
8.6.2.6 Rinse Effectiveness — If the emulsification spray application should be in accordance with the manu-
and final rinse step is not effective, as evidenced by exces- facturer’s recommendations, but it shall not exceed 5%.
sive residual surface penetrant after emulsification and rins- The water spray pressure should be less than 40 psi
ing, thoroughly reclean and completely reprocess the part. (275 kPa). Contact with the emulsifier shall be kept to the
minimum time to obtain an acceptable background and
8.6.3 Hydrophilic Emulsification (Method D): shall not exceed two minutes. The water temperature shall
8.6.3.1 Application of Hydrophilic Remover — be maintained between 50°F and 100°F (10°C and 38°C).
Following the required penetration dwell time, the parts 8.6.3.6 Post-Rinsing of Hydrophilic Emulsified
may be be prerinsed with water prior to application of the Parts — Effective post-rinsing of emulsified penetrant from
hydrophilic emulsifier. This prerinse allows for the removal the surface can be accomplished using either manual or
of excess surface penetrant from the parts prior to emulsifi- automated water spray, water immersion, or combinations
cation so as to minimize penetrant contamination in the thereof. The total rinse time shall not exceed two minutes
hydrophilic emulsifier bath, thereby extending its life. It regardless of the number of rinse methods used.
is not necessary to prerinse a part if a spray application of
8.6.3.7 Immersion Post-Rinsing — If an agitated
emulsifier is used.
immersion rinse is used, the amount of time the part(s) is
8.6.3.2 Prerinsing Controls — Effective prerinsing (are) in the bath shall be the minimum required to remove
is accomplished by manual, semi-automated, or automated the emulsified penetrant and shall not exceed two minutes.
water spray rinsing of the part(s). The water spray pressure In addition, the temperature range of the water shall be
shall not exceed 40 psi (275 kPa) when manual or hydro within 50°F and 100°F (10°C and 38°C). Be aware that a
air spray guns are used. When hydro-air pressure spray touch-up rinse may be necessary after immersion rinse,
guns are used, the air pressure shall not exceed 25 psi but the total wash time still shall not exceed two minutes.
(172 kPa). Water free of contaminants that could clog spray 8.6.3.8 Spray Post-Rinsing — Effective post-rins-
nozzles or leave a residue on the part(s) is recommended. ing following emulsification can also be accomplished by
8.6.3.3 Application of Emulsifier — The residual manual, semi-automatic, or automatic water spray. The
surface penetrant on part(s) must be emulsified by immers- water spray pressure shall not exceed 40 psi (275kPa) when
ing the part(s) in an agitated hydrophilic emulsifier bath manual or hydro air spray guns are used. When hydro-air
or by spraying the part(s) with water/emulsifier solutions pressure spray guns are used, the air pressure shall not
thereby rendering the remaining residual surface penetrant exceed 25 psi (172 kPa). The water temperature shall be
water-washable for the final rinse station. The emulsifica- between 50°F and 100°F (10°C and 38°C). The spray rinse
tion time begins as soon as the emulsifier is applied. The time shall be less than two minutes, unless otherwise spec-
length of time that the emulsifier is allowed to remain on ified.
a part and in contact with the penetrant is dependent on 8.6.3.9 Rinse Effectiveness — If the emulsification
the type of emulsifier employed and the surface roughness. and final rinse steps are not effective, as evidenced by
The emulsification time should be determined experimen- excessive residual surface penetrant after emulsification
tally for each specific application. The surface finish and rinsing, thoroughly reclean, and completely reprocess
(roughness of the part is a significant factor in determining the part.
the emulsification time necessary for an emulsifier. Contact 8.6.4 Removal of Solvent-Removable Penetrant
emulsification time should be kept to the least possible (Method C) — After the required penetrant dwell time, the
time consistent with an acceptable background and shall
excess penetrant is removed by wiping with a dry, clean,
not exceed two minutes. lint-free cloth/towel. Then use a clean, lint-free cloth/towel
8.6.3.4 Immersion — For immersion application, lightly moistened with solvent remover to remove the
parts shall be completely immersed in the emulsifier bath. remaining traces of surface penetrant. Gently wiping must
The hydrophilic emulsifier concentration shall be as recom- be used to avoid removing penetrant from any discontinu-
mended by the manufacturer and the bath or part shall be ity. On smooth surfaces, an alternate method of removal
gently agitated by air or mechanically throughout the cycle. can be done by wiping with a clean, dry cloth. Flushing
The minimum time to obtain an acceptable background the surface with solvent following the application of the
shall be used, but the dwell time shall not be more than penetrant and prior to developing is prohibited.
two minutes unless approved by the contracting parties. 8.7 Drying — Regardless of the type and method of
8.6.3.5 Spray Application — For spray applica- penetrant used, drying the surface of the part(s) is necessary
tions, all part surfaces should be evenly and uniformly prior to applying dry or nonaqueous developers or follow-
sprayed with a water/emulsifier solution to effectively ing the application of the aqueous developer. Drying time
569
No further reproduction or distribution is permitted.
ARTICLE 24, SE-165 2013 SECTION V
will vary with the size, nature, and number of parts being surface has been dried, apply nonaqueous wet developer
processed. by spraying in such a manner as to ensure complete part
8.7.1 Drying Parameters — Components shall be air coverage with a thin, even film of developer. The developer
dried at room temperature or in a drying oven. Oven tem- shall be applied in a manner appropriate to the type of
peratures shall not exceed 160°F (71°C). Drying time shall penetrant being used. For visible dye, the developer must
only be that necessary to adequately dry the part. Compo- be applied thickly enough to provide a contrasting back-
nents shall be removed from the oven after drying. Compo- ground. For fluorescent dye, the developer must be applied
nents should not be placed in the oven with pooled water thinly to produce a translucent covering. Dipping or flood-
or pooled aqueous solutions/suspensions. ing parts with nonaqueous developers is prohibited,
because the solvent action of these types of developers can
8.8 Developer Application — There are various modes flush or dissolve the penetrant from within the discontinu-
of effective application of the various types of developers ities.
such as dusting, immersing, flooding or spraying. The
developer form, the part size, configuration, and surface NOTE 12: The vapors from the volatile solvent carrier in the developer
may be hazardous. Proper ventilation should be provided at all times,
roughness will influence the choice of developer appli-
but especially when the developer is applied inside a closed area.
cation.
8.8.4 Liquid Film Developers — Apply by spraying
8.8.1 Dry Powder Developer (Form A) — Dry pow- as recommended by the manufacturer. Spray parts in such
der developers should be applied immediately after the a manner as to ensure complete part coverage of the area
part is dry in such a manner as to ensure complete coverage being examined with a thin, even film of developer.
of the area of interest. Parts can be immersed in a container
8.8.5 Developing Time — The length of time the
of dry developer or in a fluid bed of dry developer. They
developer is to remain on the part prior to inspection shall
can also be dusted with the powder developer through a
be not less than ten minutes. Developing time begins imme-
hand powder bulb or a conventional or electrostatic powder
diately after the application of dry powder developer or as
gun. It is common and effective to apply dry powder in
soon as the wet (aqueous and nonaqueous) developer coat-
an enclosed dust chamber, which creates an effective and
ing is dry (that is, the water or solvent carrier has evaporated
controlled dust cloud. Other means suited to the size and
to dryness). The maximum permitted developing times
geometry of the specimen may be used, provided the pow-
shall be four hours for dry powder developer (Form A),
der is applied evenly over the entire surface being exam-
two hours for aqueous developer (Forms B and C), and
ined. Excess developer powder may be removed by shaking
one hour for nonaqueous developer (Forms D and E).
or tapping the part, or by blowing with low-pressure dry,
clean, compressed air not exceeding 5 psi (34 kPa). Dry 8.9 Inspection — After the applicable development
developers shall not be used with Type II penetrant. time, perform inspection of the parts under visible or ultra-
8.8.2 Aqueous Developers (Forms B and C) — Water violet light as appropriate. It may be helpful to observe
soluble developers (Form B) are prohibited for use with the bleed out during the development time as an aid in
Type 2 penetrants or Type 1, Method A penetrants. Water interpreting indications.
suspendable developers (Form C) can be used with noth 8.9.1 Ultraviolet Light Examination — Examine parts
Type 1 and Type 2 penetrants. Aqueous developers shall tested with Type 1 fluorescent penetrant under black light
be applied to the part immediately after the excess penetrant in a darkened area. Ambient light shall not exceed 2 fc
has been removed and prior to drying. Aqueous developers (21.5 lx). The measurement shall be made with a suitable
shall be prepared and maintained in accordance with the visible light sensor at the inspection surface.
manufacturer’s instructions and applied in such a manner
NOTE 13: Because the fluorescent constituents in the penetrant will
as to ensure complete, even, part coverage. Aqueous devel- eventually fade with direct exposure to ultraviolet light, direct exposure
opers may be applied by spraying, flowing, or immersing of the part under test to ultraviolet light should be minimized when not
the part in a prepared developer bath. Immerse the parts removing excess penetrant or evaluating indications.
only long enough to coat all of the part surfaces with the 8.9.1.1 Black Light Level Control — Black lights
developer since indications may leach out if the parts are shall produce a minimum light intensity of 1000 W/cm2,
left in the bath too long. After the parts are removed from at a distance of 15 in. (38.1 cm). The intensity should be
the developer bath, allow the parts to drain. Drain all excess checked daily to ensure the required output. Reflectors and
developer from recesses and trapped sections to eliminate filters should also be checked daily for cleanliness and
pooling of developer, which can obscure discontinuities. integrity. Cracked or broken ultraviolet filters shall be
Dry the parts in accordance with 8.7. The dried developer replaced immediately. Since a drop in line voltage can
coating appears as a translucent or white coating on the part. cause decreased black light output with consequent incon-
8.8.3 Nonaqueous Wet Developers (Forms D and sistent performance, a constant-voltage transformer should
E) — After the excess penetrant has been removed and the be used when there is evidence of voltage fluctuation.
570
No further reproduction or distribution is permitted.
2013 SECTION V ARTICLE 24, SE-165
NOTE 14: Certain high-intensity black lights may emit unacceptable 9. Special Requirements
amounts of visible light, which can cause fluorescent indications to disap-
pear. Care should be taken to use only bulbs suitable for fluorescent
9.1 Impurities:
penetrant testing purposes. 9.1.1 When using penetrant materials on austenitic
8.9.1.2 Black Light Warm-Up — Unless otherwise stainless steels, titanium, nickel-base or other high-temper-
ature alloys, the need to restrict certain impurities such as
specified by the manufacturer, allow the black light to
sulfur, halogens and alkali metals must be considered.
warm up for a minimum of five minutes prior to use or
These impurities may cause embrittlement or corrosion,
measurement of its intensity.
particularly at elevated temperatures. Any such evaluation
8.9.1.3 Visual Adaptation — Personnel examining should also include consideration of the form in which the
parts after penetrant processing shall be in the darkened impurities are present. Some penetrant materials contain
area for at least one minute before examining parts. Longer significant amounts of these impurities in the form of vola-
times may be necessary under some circumstances. Photo- tile organic solvent that normally evaporate quickly and
chromic or tinted lenses shall not be worn during the pro- usually do not cause problems. Other materials may contain
cessing and examination of parts. impurities, which are not volatile and may react with the
part, particularly in the presence of moisture or elevated
8.9.2 Visible Light Examination— Inspect parts tested
temperatures.
with Type 2 visible penetrant under either natural or artifi-
cial visible light. Proper illumination is required to ensure 9.1.2 Because volatile solvents leave the surface
adequate sensitivity of the examination. A minimum light quickly without reaction under normal examination proce-
intensity at the examination surface of 100 fc (1076 lx) is dures, penetrant materials are normally subjected to an
required. evaporation procedure to remove the solvents before the
materials are analyzed for impurities. The residue from
8.9.3 Housekeeping — Keep the examination area this procedure is then analyzed in accordance with Test
free of interfering debris, including fluorescent residues Method D 1552 or Test Method D 129 decomposition
and objects. followed by Test Method D 516, Method B (Turbidimetric
8.9.4 Indication Verification — For Type 1 inspec- Method) for sulfur. The residue may also be analyzed by
tions only, it is common practice to verify indications by Test Method D 808 or Annex A2 on Methods for Measur-
wiping the indication with a solvent-dampened swab or ing Total Chlorine Content in Combustible Liquid Pene-
brush, allowing the area to dry, and redeveloping the area. trant Materials (for halogens other than fluorine) and Annex
Redevelopment time shall be a minimum of ten minutes, A3 on Method for Measuring Total Fluorine Content in
except nonaqueous redevelopment time should be a mini- Combustible Liquid Penetration Materials (for fluorine).
mum of three minutes. If the indication does not reappear, An alternative procedure, Annex A4 on Determination of
the original indication may be considered false. This proce- Anions by Ion Chromatography, provides a single instru-
dure may be performed up to two times for any given mental technique for rapid sequential measurement of com-
original indication. mon anions such as chloride, fluoride, and sulfate. Alkali
metals in the residue are determined by flame photometry,
8.9.5 Evaluation — All indications found during atomic absorption spectrophotometry, or ion chromatogra-
inspection shall be evaluated in accordance with acceptance phy (see ASTM D 4327).
criteria as specified. Referenced Photographs of indications
NOTE 15: Some current standards require impurity levels of sulfur and
are noted in E 433.
halogens to not exceed 1% of any one suspect element. This level,
however, may be unacceptable for some applications, so the actual maxi-
8.10 Post Cleaning — Post cleaning is necessary when
mum acceptable impurity level must be decided between supplier and
residual penetrant or developer could interfere with subse- user on a case by case basis.
quent processing or with service requirements. It is particu-
larly important where residual penetrant testing materials 9.2 Elevated-Temperature Testing — Where penetrant
might combine with other factors in service to produce testing is performed on parts that must be maintained at
corrosion and prior to vapor degreasing or heat treating elevated temperature during examination, special penetrant
the part as these processes can bake the developer onto materials and processing techniques may be required. Such
the part. A suitable technique, such as a simple water rinse, examination requires qualification in accordance with 10.2
water spray, machine wash, solvent soak, or ultrasonic and the manufacturer’s recommendations shall be
cleaning may be employed (see Annex A1 for further infor- observed.
mation on post cleaning). It is recommended that if devel-
oper removal is necessary, it should be carried out as 10. Qualification and Requalification
promptly as possible after examination so that the devel- 10.1 Personal Qualification — When required by cus-
oper does not adhere to the part. tomer, all penetrant testing personnel shall be qualified/
571
No further reproduction or distribution is permitted.
ARTICLE 24, SE-165 2013 SECTION V
certified in accordance with a written procedure conform- 10.3 Nondestructive Testing Agency Qualification — If
ing to the applicable edition of recommended Practice a nondestructive testing agency as described in Practice
SNT-TC-1A, ANSI/ASNT CP-189, NAS-410, or E 543 is used to perform the examination, the agency
MIL-STD-410. should meet the requirements of Practice E 543.
10.2 Procedure Qualification — Qualification of proce-
dures using times, conditions, or materials differing from 10.4 Requalification may be required when a change or
those specified in this general practice or for new materials substitution is made in the type of penetrant materials or
may be performed by any of several methods and should in the procedure (see 10.2).
be agreed upon by the contracting parties. A test piece
containing one or more discontinuities of the smallest rele-
vant size is used. When agreed upon by the contracting 11. Keywords
parties, the test piece may contain real or simulated discon-
tinuities, providing it displays the characteristics of the 11.1 fluorescent liquid penetrant testing; hydrophilic
discontinuities encountered in product examination. emulsification; lipophilic emulsification; liquid penetrant
10.2.1 Requalification of the procedure to be used testing; nondestructive testing; solvent removable; visible
may be required when a change is made to the procedure liquid penetrant testing; water-washable; post-emulsified;
or when material substitution is made. black light; ultraviolet light; visible light
572