Professional Documents
Culture Documents
Capnography Application in
Acute and Critical Care
Tom Ahrens, DNS, RN, CCRN, CS; Carrie Sona, MSN, RN, CCRN, CS
123
124 AHRENS AND SONA AACN Clinical Issues
frared light shown through a filter that allows advantages and disadvantages. The advan-
infrared light in the 4.3 m wavelength to tage of the mainstream analysis is that it im-
pass into a chamber containing exhaled air mediately samples the exhaled CO2 and pro-
with CO2 present. By measuring the amount vides real-time information (Figure 2). It also
of light that passes through the chamber and is less affected by factors, such as water va-
how much is absorbed by the CO2, a quantifi- por, that interfere with the measurement of
able amount of CO2 can be measured.2 This CO2. The major disadvantage of the main-
technique is best performed if the measure- stream analysis is that the weight of the ana-
ment occurs immediately upon inspiration lyzer rests on the patient’s ventilator tubing.
and exhalation. In a patient on a ventilator, This can be uncomfortable, or it can pull on
the measurement should take place within the endotracheal tube. In addition, main-
the endotracheal tube. On a nonintubated pa- stream analysis can be used only on intu-
tient, the measurement should take place bated patients.
near the mouth or nares (Figure 1). The sidestream analyzer has the advan-
The sampling of exhaled air can be mea- tage of being lightweight. It also can be used
sured immediately (mainstream analysis), or on both intubated and nonintubated patients
the air can be drawn through a sampling (Figure 3). The major disadvantage of this
tube to an analyzer distant from the patient technique is its likelihood of becoming ob-
(sidestream analysis). Each technique has structed by water because the sampling tube
has a very small internal diameter. It also pro- portant as a CO2 value reading. For example,
vides a waveform slightly delayed by a few if the waveform shows a climbing phase III,
milliseconds. Advances in sidestream tech- then there may be obstruction in the way of
nology have reduced some of the problems reaching the alveolar plateau phase, a condi-
previously seen, such as obstructions in the tion common in reactive airways (eg, asthma)
tubing.3 The choice for mainstream or side- (Figure 6). Any condition that alters the rela-
stream analysis depends on the clinical appli- tion of ventilation to lung perfusion can
cation. Both techniques, however, work well change the slope and value of phase III. For
as long as their limitations are understood. example, a condition that decreases blood
flow to the lungs (eg, pulmonary emboli or
Capnography Waveforms and cardiac arrest) while breathing is maintained
Values will create an effective increase in physiologic
dead space. This can change either the angle
When CO2 is measured as the patient or the value of phase III (Figure 7).
breathes, a characteristic waveform is created In addition, the absence of a waveform or
(Figure 4). The components of the waveform the presence of a steady CO2 value indicates
have different meanings. Figure 5 provides a a loss of effective breathing (Figure 8). This
description of the capnogram waveform.4 may occur with a misplaced endotracheal
The presence of a capnogram wave indi- tube or correct placement of nasogastric
cates that the sensor is detecting CO2 from the tubes, which would be in the stomach
lungs. The presence of a waveform is as im- where CO2 is not detected.
Figure 5. Capnogram Waveform Phases. Phase 0: Inspiration. Phase I: Early expiration. This represents the
anatomic dead-space of ventilation that is either absent or has small amounts of CO2 present. Phase II: Expira-
tory phase, in which CO2 is being emptied from initial gas exchanging parts of the lung. This level is slightly re-
duced due to mixing with the airways, which has small amounts or absence of CO2. Phase III: Alveolar emptying
phase, noted by a sudden increase and stabilization in the CO2 wave. The point of the sudden change is often
called the alpha angle. A sharp drop in CO2 values marks the point where Phase III ends and Phase 0 begins.
The beta angle is the angle between Phase III and Phase 0. The end of Phase III is also called the end-tidal CO2
value (PetCO2).
126 AHRENS AND SONA AACN Clinical Issues
authors found that a PetCO2 level lower than the assessment of dead space ventilation.
10 after 20 minutes corresponded to 100% Minute ventilation is the amount of air (mea-
mortality. Similar findings were reported by sured in liters/minute) determined by the res-
Ahrens et al21-22 for an in-hospital setting. piratory rate times the tidal volume. The nor-
Because of its ability to detect changes in mal response to increased minute ventilation
pulmonary blood while maintaining alveolar is a decrease in PetCO2. Because PaCO2 is
ventilation (dead space ventilation), cap- determined by the relation between alveolar
nography can be helpful in the diagnosis of ventilation and CO2 production, a high
other conditions involving altered pulmonary minute ventilation in the presence of hyper-
blood flow (ie, pulmonary embolism).23-24 capnia indicates either increased dead space
Pulmonary embolism diminishes pulmonary ventilation or increased CO2 production.34
blood flow while not affecting alveolar venti- This information can assist in the determina-
lation. An obstructed pulmonary artery leads tion of clinically relevant changes such as
to areas of the lung being ventilated but not pulmonary embolus and decreased perfusion
perfused. This leads to a decrease of the states when used with an understanding of
PetCO2 level and a widening of the arterial to the relation between the CO2 gradient and
end-tidal gradient (PaCO2-PetCO2). Normally, minute ventilation in the assessment of dead
the PaCO2-PetCO2 gradient is less than 5 mm space ventilation.
Hg. In a study by Chopin et al, the CO2 gradi- Another ventilatory pattern easily identi-
ent considered to be diagnostic for pul- fied with PetCO2 is incomplete neuromuscu-
monary embolism in one study was identi- lar blockade. Patients who have been chem-
fied as 12 ⫾ 6.9 mm Hg, which was correct ically paralyzed for operative procedures or
as correlated with pulmonary angiogram for who remain chemically paralyzed in the in-
74% of the patients.25 In this study, the pa- tensive care unit for ventilatory management
tients with negative results for pulmonary can benefit from capnography. When a
embolism via pulmonary angiogram had a chemically paralyzed patient initiates a spon-
CO2 gradient of 1 ⫾ 2.4 mm Hg, which was taneous breath, which would indicate in-
correct for 100% of the patients. Other re- complete chemical paralysis, a characteristic
searchers have shown that patients with pul- cleft (dip) appears in the PetCO2 waveform.
monary embolism have PetCO2 values lower This can be helpful for postoperative pa-
than those without pulmonary embolism. tients who may need a reversal agent or for
patients that need an additional dose of neu-
romuscular blocker.
Assessment of Ventilation Many intensive care unit nurses use the
benefit of capnography to help them inter-
Continuous PetCO2 monitoring offers useful pret hemodynamic information. Hemody-
information in the assessment of arterial CO2, namic waveforms are most reliable when
particularly in patients with normal ventila- read at the end of the patient’s expiratory cy-
tion-perfusion relations. The relation between cle.22 With the use of PetCO2, end expiration
alveolar and arterial CO2 (PaCO2-PetCO2) is is easily identified (Figure 10). This makes
close in healthy patients (eg, less than 5 mm bedside interpretation of hemodynamic in-
Hg of difference). For example, if the arterial formation more reliable and accurate.
CO2 is 35 mm Hg and the PetCO2 is 30, the The benefits of capnography in a variety
CO2 gradient is 5 mm Hg. Exhaled CO2 is al- of clinical settings and scenarios are shown
most always slightly lower than the PaCO2. in the following examples.
In patients with normal dead space, it is
safe to use PetCO2 as a substitute for
PaCO2.27-33 When the baseline CO2 gradient Case Studies
is assessed and trends are followed this can Case 1. Prehospital: Malpositioned
greatly reduce the necessity for an assess- Endotracheal Tube
ment of arterial blood gases using pulse
oximetry and capnography. Emergency medical system workers respond
The continuous monitoring of PetCO2 to a single car motor vehicle crash at 0200
with minute ventilation also can be helpful in during a snowstorm. The unrestrained driver
Vol. 14, No. 2 May 2003 CAPNOGRAPHY APPLICATION 129
was ejected from her vehicle, and according Case 2. Emergency Department: Full Arrest
to an eyewitness account, was thrown 20 feet
from the vehicle. The emergency medical An 80-year-old man is on a stretcher in the
technician finds the victim with only shallow, emergency department reporting chest pain
agonal respiratory effort and unresponsive. and shortness of breath. He is in the process
He immediately intubates the patient after se- of ruling out myocardial ischemia via 12-lead
curing the cervical spine and transports the ECG and cardiac enzymes. A routine my-
patient via air evacuation to the closest level ocardial ischemia algorithm is followed, and
1 trauma center. The patient is monitored the patient is placed on a monitor. Oxygen is
with electrocardiography (ECG) and pulse applied via a nasal cannula. An intravenous
oximetry. She requires 100% fraction of in- line is started and morphine given. Sud-
spired oxygen (FiO2) by bag mask with per- denly, the ECG monitor alarms and shows
sistently low O2 saturations in route. the development of ventricular fibrillation.
On the patient’s arrival in the emergency Immediate defibrillation at 200 joules is per-
department, an ETCO2 monitor is hooked up. formed, and advanced cardiac life support
No waveform or number is visible on the protocol is followed. The patient receives
monitor. The emergency department nurse three successive countershocks without con-
auscultates only distant breath sounds. The version of the rhythm. Cardiopulmonary re-
tube is removed and replaced with an imme- suscitation (CPR) is initiated. The patient is
diate capnogram detected on the PetCO2 intubated, and drug therapies are started.
monitor. The tube placement had been The capnogram shows a waveform and a
esophageal. PetCO2 reading of 14 mmHg. With CPR, the
The emergency medical system staff used PetCO2 reading increases to 18 mm Hg. Ad-
the monitoring equipment available to them vanced cardiac life support protocol contin-
en route and confirmed endotracheal tube ues, and an increase in the PetCO2 to 25 mm
placement quickly by hearing distant breath Hg is seen. Cardiopulmonary resuscitation is
sounds in a noisy environment in the middle stopped. A rhythm is detected on the moni-
of a snowstorm. The availability of capnog- tor, and a pulse is palpable.
raphy to this crew could have been life sav- This example shows how an early sign of
ing for this patient. She received 100% FiO2 restored spontaneous circulation and im-
via a bag mask en route, but did not have proved blood flow was detected with
adequate ventilation because of the tube capnography. Capnography is easy to read
malposition. This could have resulted in hy- and not interfered with by manual chest
poxic brain injury for this patient. compressions. The increase in the PetCO2
130 AHRENS AND SONA AACN Clinical Issues
alerted the clinicians to stop CPR and check tended abdomen. The surgeon is notified,
the rhythm and pulse. and immediate rapid transfusion is initiated
with a stat return to the operating room suite
Case 3. Intensive Care Unit: Pulmonary
for repair of an anastomotic leak.
Embolism
In this case, the low PetCO2 value warned
of the reduced blood flow, consistent with
A 50-year-old woman with end-stage cervi- other clinical symptoms.
cal cancer is admitted to the intensive care
unit with respiratory distress. She has no ad- Case 5. Outpatient Surgery Center:
vanced directive or durable power of attor- Oversedation
ney. She is not alert. Her pulse oximetry
reading shows 88% saturation, and she is A 57-year-old man is admitted to a local out-
tachypneic, with a respiratory rate of 40. Im- patient surgery center for arthroscopic knee
mediate intubation is performed. Capnogra- surgery. He is very fearful and requests gen-
phy shows a waveform and a PetCO2 of 18, eral anesthesia with the use of a laryngeal
with an immediate increase in the patient’s mask airway and a local anesthetic during
pulse oximetry to 93%. She is placed on a the case. The nurse anesthetist (CRNA) has
ventilator, with persistently low PetCO2 read- no difficulty inserting the tube, and a capno-
ings ranging from 18 to 22 mm Hg. Arterial gram with a reading of 36 mm Hg is immedi-
blood gas analysis is performed after intuba- ately present on the portable capnography
tion, with a PaCO2 result of 40. Because the unit. The CRNA, wanting to ensure that the
large CO2 gradient of 22 suggests a pul- patient is pain free throughout the proce-
monary embolism, bilateral lower extremity dure, administers 200 mcg of fentanyl and 50
Dopplers and a portable ventilation-perfu- mg of propofol. The patient is being venti-
sion scan are ordered. The Dopplers show a lated at a rate of 14 via ambu bag, but has
large superficial femoral vein thrombosis. shallow spontaneous efforts, with a PetCO2
The ventilation-perfusion scan shows a large reading of 36 mm Hg. The CRNA stops bag-
perfusion defect, and heparin therapy is ging to administer additional anxiolytic, and
started for pulmonary embolism. the PetCO2 reading climbs to 50 mm Hg. The
The use of capnography helped to iden- CRNA notes that the patient is not breathing
tify the large CO2 gradient and to move the spontaneously and resumes ventilation. The
case toward early anticoagulation therapy. PetCO2 decreases to 40 mm Hg. The CRNA
holds the administration of more analgesics
and anxiolytics until the patient resumes
Case 4. Postanesthesia Recovery Room:
Massive Exsanguination
spontaneous ventilatory efforts.
In this case, the PetCO2 was a clear warn-
A 68-year-old man is in the recovery room ing of inadequate ventilation.
after open repair of a 10-cm infrarenal ab-
dominal aortic aneurysm. He remains se-
dated and mechanically ventilated. The Summary
postanesthesia recovery staff is obtaining
postoperative laboratory work, a chest radi- Capnography is one of the emerging new vi-
ograph, and an ECG. The patient’s blood tal signs in acute and critical care. There are
pressure via arterial line was 140/90 initially, at least 10 uses of capnography in clinical
but currently is 80 systolic. His PetCO2 on settings, ranging from confirmation of tra-
admission was 35 mm Hg, and now is 20. A cheal and esophageal tube placement to as-
fluid bolus is given while the clinicians await sessment of blood flow. Expansion of
the results of the complete blood count. capnography use beyond the operating
The patient’s hypotension persists, and room and intensive care units will continue,
the PetCO2 level continues to decrease. The particularly as the need for ventilation as-
nurse knows the patient is sedated and not sessment grows in outpatient settings. The
overbreathing the set ventilator rate of 10. noninvasive nature of capnography serves
She pulls back the sheets to examine the pa- only to make its appeal greater. Capnogra-
tient and notices mottled feet and a dis- phy likely will expand to every area of
Vol. 14, No. 2 May 2003 CAPNOGRAPHY APPLICATION 131
healthcare that requires monitoring of venti- 15. Salen P, O’Connor R, Sierzenski P, et al. Can
lation and blood flow, making it one of the cardiac sonography and capnography be
most important new technologies in patient used independently and in combination to
monitoring. predict resuscitation outcomes? Acad Emerg
Med. 2001;8(6):610-615.
16. Grmec S, Klemen P. Does the end-tidal carbon
dioxide (EtCO2) concentration have prognos-
References tic value during out-of-hospital cardiac arrest?
Eur J Emerg Med. 2001;8(4):263-269.
1. Barnes TA, MacDonald D, Nolan J, et al. Air- 17. Asplin BR, White RD. Prognostic value of
way devices. Ann Emerg Med. 2001;37:148- end-tidal carbon dioxide pressures during out
150. of hospital cardiac arrest. Ann Emerg Med.
2. Paulus DA. Capnography. Int Anesthesiol. 1995;25:756-760.
1989;27:167-175. 18. Cantineau JP, Lambert Y, Merckx P, et al. End-
3. Singh S, Venkataraman ST, Saville A, et al. tidal carbon dioxide during cardiopulmonary
NPB-75: a portable quantitative microstream resuscitation in humans presenting mostly
capnometer. J Emerg Med. 2001;19(3):208- with asystole: a predictor of outcome. Crit
210. Care Med. 1996;24:791-796.
4. Bhavani-Shankar K, Philip JH. Defining seg- 19. Kalenda Z. The capnogram as a guide to the
ments and phases of a time capnogram. efficacy of cardiac massage. Resuscitation.
Anesth Analg. 2000;91(4):973-977. 1978;6:259-263.
5. Webb RK, van der Walt JH, Runciman WB, et 20. Levine RL, Wayne MA, Miller CC. End-tidal
al. The Australian Incident Monitoring Study. carbon dioxide and outcome of out of hospi-
Which monitor? An analysis of 2000 incident tal cardiac arrest. N Engl J Med. 1997;337:301-
reports. Anaesth Intensive Care. 1993;21:529- 306.
542. 21. Ahrens T, Schallom L, Bettorf K, et al. End-
6. Joint Commission on Accreditation of Health- tidal carbon dioxide measurements as a prog-
care Organizations. Sentinel event alert. Issue nostic indicator of outcome in cardiac arrest.
25. February 26, 2002. Am J Crit Care. 2001;10(6):391-398.
7. Burns SM, Carpenter R, Truwit JD. Report on 22. Ahrens TS, Wiejera H, Ray SE. Capnography.
the development of a procedure to prevent Crit Care Nurs Clin North Am. 1999;11:49-62.
placement of feeding tubes into the lungs us- 23. Weg JG. A new nich for end-tidal CO2. Crit
ing end-tidal CO2 measurements. Crit Care Care Med. 2000;28(11):3752-3754.
Med. 2001;29(5):936-939. 24. Johanning JM, Veverka TJ, Bays RA, et al.
8. Kindopp AS, Drover JW, Heyland DK. Evaluation of suspected pulmonary embolism
Capnography confirms correct feeding tube utilizing end-tidal CO2 and D-dimer. Am J
placement in intensive care unit patients. Can Surg. 1999;178(2):98-102.
J Anaesth. 2001;48(7):705-710. 25. Chopin C, Fesard P, Mangalaboyi J, et al. Use
9. Idris AH, Staples ED, O’Brien DJ, et al. End- of capnography in diagnosis of pulmonary
tidal carbon dioxide during extremely low embolism during acute respiratory failure of
cardiac output. Ann Emerg Med. 1994;23:568- chronic obstructive pulmonary disease. Crit
572. Care Med. 1990;18;353-357.
10. Shibutani K. Do changes in end-tidal PCO2 26. Kline JA, Arunachlam M. Preliminary study of
quantitatively reflect changes in cardiac out- the capnogram waveform area to screen for
put? Anesth Analg. 1994;79(5):829-833. pulmonary embolism. Ann Emerg Med.
11. Isserles SA, Breen PH. Can changes in end- 1998;32(3 Pt 1):289-296.
tidal PCO2 measure changes in cardiac out- 27. Bongard F, Wu Y, Lee TS, Klein S. Capno-
put? Anesth Analg. 1991;73:808-814. graphic monitoring of extubated postopera-
12. Weil MH, Bisera J, Trevino RP, et al. Cardiac tive patients. J Invest Surg. 1994;7:259-264.
output and end-tidal carbon dioxide. Crit 28. Hess D, Scholttag A, Levin B, et al. An evalua-
Care Med. 1985;13:9079. tion of the usefulness of end-tidal PCO2 to aid
13. Gazmuri RJ, vonPlanta M, Weil MH, et al. Ar- weaning from mechanical ventilation follow-
terial PCO2 as an indicator of systemic perfu- ing cardiac surgery. Respir Care. 1991;36:837-
sion during cardiopulmonary resuscitation. 843.
Crit Care Med. 1989;17:237-240. 29. Morley TF, Giaimo J, Maroszan E, et al. Use of
14. Jin X, Weil MH, Tang W, et al. End-tidal CO2 capnography for assessment of the adequacy
as a noninvasive indicator or cardiac index of alveolar ventilation during weaning from
during circulatory shock. Crit Care Med. mechanical ventilation. Am Rev Respir Dis.
2000;28(7):2415-2419. 1993;148:339-344.
132 AHRENS AND SONA AACN Clinical Issues
30. Thrush DN, Mentis SW, Downs JB. Weaning during transport. Crit Care Med. 1996;24:
with end-tidal CO2 and pulse oximetry. J Clin 608-611.
Anesth. 1991;3:456-460. 33. Cook D, Meade M, Guyatt G, et al. Trials of
31. Smarma SK, McGuire GP, Cruise CJ. Stability miscellaneous interventions to wean from
of the arterial to end-tidal carbon dioxide dif- mechanical ventilation. Chest. 2001;120(6S):
ference during anaesthesia for prolonged 438S-444S.
neurosurgical procedures. Can J Anesth. 34. Jubran A, Tobin MJ. Monitoring gas exchange
1995;42:498-503. during mechanical ventilation. In: Tobin MJ,
32. Palmon SC, Liu M, Moore LE, et al. Capnog- ed. Principles and practice of mechanical
raphy facilitates tight control of ventilation ventilation. New York: McGraw-Hill; 1994.