You are on page 1of 5

MEDICAL RESEARCH CENTER

7726 Arjons Dr., San Diego, CA 92126


TEL: 858.621.2630 FAX: 858.621.6019

ANMAR METROLOGY.
A TRANSCAT´S
A TRANSCAT´S CALIFORNIA
CALIFORNIA CALIBRATION
CALIBRATION LABORATORY.
LABORATORY.

The reported calibration(s) are accredited by the American Association of Physicist in Medicine (AAPM)
and/or the American Association of Laboratory Accreditation (A2LA) to the ISO/IEC 17025:2005 standard.
Calibrations are directly traceable to System of Units (SI) through the National Institute of Standards and
Technology (NIST)) or other recognized National Metrology Institute (NMI).

Your enclosed calibration certificate displays the appropriated accreditation logo(s).

For additional information about our accreditation status and scope of service, visit transact website:
www.transcat.com

ANMAR, ELECTRONIC CALIBRATION LABORATORY


Report No. EDM
EDM24895

ANMAR METROLOGY.
ELECTRONIC CALIBRATION
CALIBRATION LABORATORY.
LABORATORY.
This calibration is compliant with the ISO/IEC 17025:2005 standard,
and is accredited by the AAPM and the American Association for Laboratory Accreditation.

7726 Arjons Dr., San Diego, CA 92126


TEL: 858.621.2630 FAX: 858.621.6019

Report of Calibration for

RF303 Electrosurgery Analyzer

Submited by: DCG- Electrónica y Biomédica


Plan de Agua Prieta 47-A
47 Misión del Sol.
Hermosillo, México. C.P. 83100.

Calibrated Unit: Electrosurgery Unit Analizer


Model: RF303 Electrosurgery Analyzer.
Make: Flu
Fluke Biomedical
S/N: 9740014

Date Received: 11/10/2017


/2017 Calibration Completed: 1
11/22/2017

This calibration is directly traceable to the National Institute of Standards and Technology (NIST). The expanded uncertainty of
measurement as stated in the data page of this report is the standard uncertainty of measurements
measu multiplied by the coverage factor k=2,
which for a normal distribution corresponds to a coverage probability of approximately 95%. This report shall not be reproducedexcept
in full, without the written permission of the ANMAR METROLOGY-ECL.
METROLOGY

ANMAR, ELECTRONIC CALIBRATION LABORATORY


Report No. EDM24895

ANMAR METROLOGY.
ELECTRONIC CALIBRATION
CALIBRATION LABORATORY.
LABORATORY.
This calibration is compliant with the ISO/IEC 17025:2005 standard,
and is accredited by the AAPM and the American Association for Laboratory Accreditation.

7726 Arjons Dr., San Diego, CA 92126


TEL: 858.621.2630 FAX: 858.621.6019

RF303 Electrosurgery Analyzer Measurement


Measurement Data
Data

Calibration Completed: 11/22/2017


11/22 Report Date: 11/23/2017

Electrosurgery Analizer Unit Settings

Electrosurgery Unit Analizer


Model: RF303 Electrosurgery Analyzer.
Make: Fluke Biomedical
S/N: 9740014

V Bias Main Load (Guard Resistance with respect to connector outer shell)
+100%: 124.3 V 300 ohms.

Unit Calibration Coefficient

Scale Zero Drft Leakage Range Linearity Calibration


Calibration Coefficient
Coefficient
Watts 50 -- 50-750 +/- 0.01% 1.000 W/Rdg
mA -- 0.003 50-750
50 +/- 0.01% 1.000 mA/Rdg
Calibration Uncertainty: 0.12% Refer to the
t appendix for a description of reported calibration coefficients.
Environmental Conditions: T=22.8 C , P=98.56kPa, RH=36%. Calibration Protocol E.5.3

________________________________ _______________________________
Calibrated By: Reviewed By:
Carlos A. Zamora. B.S. Accredited Electronic Calibration Laboratory
Accredited Electronic Calibration Laboratory Advisory Committee Member or Designee

ANMAR, ELECTRONIC CALIBRATION LABORATORY


ANMAR METROLOGY.
ELECTRONIC CALIBRATION LABORATORY.
UNOFFICIAL CALIBRATION SUMMARY
(Refer to Formal Report for Accredited Calibration Results)

Unit: RF303 Electrosurgery Analyzer (9740014) Due for re-cal.: 11/21/2019

Report # Cal. Date Scale Calibration Coefficient


EDM24895 11/22/2017 WATTS 1.000 W/Rdg
11/22/2017 mA 1.000 mA/Rdg

ANMAR, ELECTRONIC CALIBRATION LABORATORY


Calibrated: 11/22/2017
Cal. Due: 11/21/2019
By: CAZ

Calibrated: 12/19/2017
Cal. Due: 21/18/2019
By: CAZ