Bi-directional Communication
Specifications
1. Preface -------------------------------------------------------------------------------------- 1
2. Connection Condition -------------------------------------------------------------------- 2
3. Protocal of Data Link Layer ------------------------------------------------------------ 3
3.1. Frame-------------------------------------------------------------------------------- 3
3.2. Transition of States -------------------------------------------------------------- 4
3.2.1. Idle State --------------------------------------------------------------------- 4
3.2.2. Establishment Phase ----------------------------------------------------- 4
3.2.3. Transfer Phase ------------------------------------------------------------- 6
3.2.4. Termination Phase -------------------------------------------------------- 7
4. Record -------------------------------------------------------------------------------------- 8
4.1. Delimiter---------------------------------------------------------------------------- 9
4.2. Usable Character Code --------------------------------------------------------- 9
4.3. The maximum Length of Record--------------------------------------------- 10
4.4. The Record to be transmitted from analyzer to Host ------------------ 10
4.4.1. Message Header Record (From analyzer to Host) ---------------- 10
4.4.2. Patient Information Record (From analyzer to Host) ----------- 11
4.4.3. Measurement Order Record (From analyzer to Host) ---------- 12
4.4.4. Enquiry Record (From analyzer to Host) --------------------------- 13
4.4.5. Comment Record (From analyzer to Host) ------------------------- 14
4.4.6. Measurement Result Record (From analyzer to Host) ---------- 15
4.4.7. Message Terminator Record (From analyzer to Host) ---------- 16
4.5. The Record to be transmitted from Host to analyzer ------------------ 17
4.5.1. Message Header Record (From Host to analyzer) ---------------- 17
4.5.2. Patient Information Record (From Host to analyzer) ----------- 18
4.5.3. Measurement Order Record (From Host to analyzer) ---------- 19
4.5.4. Enquiry Record (From Host to analyzer) --------------------------- 21
4.5.5. Comment Record (From Host to analyzer) ------------------------- 22
4.5.6. Measurement Result Record (From Host to analyzer) ---------- 22
4.5.7. Message Terminator Record (From Host to analyzer) ---------- 22
i
V1.06
5. Message ------------------------------------------------------------------------------------- 23
5.1. Enquiry Message (From analyzer to Host) -------------------------------- 23
5.1.1. Transmission Timing------------------------------------------------------ 23
5.1.2. Record Configuration ----------------------------------------------------- 24
5.2. Measurement Order Message (From Host to analyzer) --------------- 25
5.2.1. Transmission Timing------------------------------------------------------ 25
5.2.2. Record Configuration ----------------------------------------------------- 26
5.2.3. Test order cancellation------------------------------------------------------27
5.2.4. Order entry for control sample-------------------------------------------27
5.2.5. Order entry for STAT sample---------------------------------------------28
5.2.6. Measurement Order Refusal (From analyzer to Host) ---------- 29
5.3. Measurement Result Message (From analyzer to Host) -------------- 30
5.3.1. Transmission Timing------------------------------------------------------ 30
5.3.2. Record Configuration ----------------------------------------------------- 30
5.3.3. Measuring Failure --------------------------------------------------------- 31
5.3.4. Measurement by Simple Operation Mode -------------------------- 32
ii
1. Preface
This manual specifies the Bi-directional communication between analyzer and the
(Host) based on ASTM1381-91 and ASTM1394-91
There are the following Sample ID mode and Requast setting for analyzer.
Sample ID mode
By Sample No. mode: This is a method to inquire to Sample No.
By Sample ID mode: This is a method to inquire to barcode attached with sample.
Request
All (Batch): enquire by “ALL” orders of host.
Appoint (individually) : In case enquire by barcode.
Note:
The trace file of communication log between the analyzer and host is automatically
created. The file is named “astm_tushin.log” at following place.
C:¥BiOLiS48i¥PC_Prog¥file¥astm_tushin.log
Do not open the file during communication.
Create a copied file when viewing the communication log. Original file should not be
opened.
2. Connection Condition
The connection for the serial communication port requires a RS-232C cross cable
(Interlink) and standard 9 pin D-SUB type connector.
Assignment of pins
1 1
Rx 2 2 Rx
Tx 3 3 Tx
DTR 4 4 DTR
SG 5 5 SG
DSR 6 6 DSR
RTS 7 7 RTS
CTS 8 8 CTS
9 9
1
V1.06
Please set the communication port manually. The analyzer is not find the port
automatically. Refer to Appendix B.
2
3. Protocol of Data Link Layer
The protocol of data link layer uses the following transmission control codes.
3.1. Frame
[STX]3O|1|01040218001|0^1^1|^^^1^T-BIL^・・・¥^^^18^CPK^0¥^^^19[ETB]??[CR][LF]
[STX]4^AMY^0|R||||||N||||Serum||||||||||F[CR][ETX]??[CR][LF]
Checksum is the least 8 bits of the value, that is gotten when the sum of character codes
from [FN] to <ETB>, <ETX>. (Modulo 256).
[C1] and [C2] are ASCII alphanumeric hexadecimal notations of the upper 4 bits and the
lower 4 bits of checksum, respectively.
3
V1.06
Idle state
Establishment Termination
phase phase
Transfer phase
The idle state is the state that transmission is not performed between analyzer and
Host.
The system with information to send transmits the <ENQ>, after transforming the state
into the establishment phase.
1) A reply of <ACK>
The state is transformed from the establishment phase to the transfer phase.
4
2) A reply of <NAK>
If the receiving system is not able to received the <ENQ> it transmit a <NAK>.
In case that the system received a reply of <NAK> six times successively, the state is
transformed into the termination phase.
3) No response/response delay
In case that there is no response within 15 seconds after transmitting of <ENQ>, the
state is transformed into the termination phase.
4) Conflict of <ENQ>
Upon receiving a reply of <ENQ> to its transmitted <ENQ>, Host waits for a reply of
<ENQ> from analyzer.
5
V1.06
5) Invalid response
Upon receiving a code except <ACK>, <NAK>, <EOT> and <ENQ>, the state is
transformed into the termination phase.
The transfer phase transmits the data from the sender to the receiver.
The data are transmitted in frames.
2) A reply of <ACK>
Upon receiving the data adequately, the system replies with <ACK>.
3) A reply of <NAK>
Upon not receiving the data adequately, the system replies with <NAK>.
In case of receiving a reply of <NAK> to its transmitted data, the system re-transmits
the same data (frames).
6
The conditions that analyzer replies with <NAK> are as follows.
5) Invalid response
Upon receiving a reply except <ACK>, <NAK> and <EOT> to the transmitted data,
the system re-transmits the same frame.
The system transmits the same frame six times, and then transforms the state into
the termination phase.
Note: Process that each one sample transmits result within 20 seconds.
(For “ACK” response, rerurn within 0.1 second (100msec).)
Transmitting <EOT>, the system transforms the state into the idle state.
7
V1.06
4. Record
The relationship between the hierarchical level supported by analyzer and the
hierarchical level of record is as follows.
* Level supported
-- Level not supported
8
4.1. Delimiter
The setting of the delimiter used by the record that is transmitted from analyzer can be
changed.
The character codes, 20h – 7Eh and 80h – FFh, can be used in the records.
When the records include the character codes that cannot be shown, they are likely to
disrupt the screen of analyzer
In case that the records include the same codes as the codes defined for the delimiters,
use the escape characters.
For analyzer escape characters may only be used in the following fields
No. Field name Notes
1 Patient name (Family name, First name) Patient Information Record
2 Doctor in charge Patient Information Record
3 Patient ID Patient Information Record
4 Sample ID Measurement Order Record
5 Comment Text Comment Record
9
V1.06
This chapter explains the definition of each record sent from the analyzer to Host
In the field of each record, the necessity of transmission means as follows.
Example of transmission
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
10
4.4.2. Patient Information Record (From analyzer to Host)
When the patient information is in analyzer, the optional items of patient information
record are set.
11
V1.06
Example of transmission
By sample No.
O|1||0^123^12|^^^1^GOT^0¥^^^2^GPT^1¥^^^3^ALP^0|R||||||N||||Serum||||||||||F<CR>
By sample ID
O|1|12345|0^1^1|^^^1^GOT^0¥^^^2^GPT^1¥^^^3^ALP^0|R||||||N||||Serum||||||||||F<CR>
12
4.4.4. Enquiry Record (From analyzer to Host)
The enquiry timing for “ALL” is when press the “Request” button or “Stat start”
button.
The enquiry timing for “Appoint” is when the analyzer reads the sample barcode.
13
V1.06
Example of transmission
C|1|I|1|I<CR>
The qualitative record of Hemolytic, Lipemic and Icteric for serum information
are set the field 4 as comment text in comment record.
Hemolytic :H
Lipemic :L
Icteric :I
Note:
When qualitative serum measured, the qualitative results are sent at the end of the
sequences of records of each samples (such as measurememnt result record or
comment record )
14
4.4.6. Measurement Result Record (From analyzer to Host)
15
V1.06
Transmission at STAT
STAT sample will cut in between the items of the sample measurement.
The result data will be transmited twice: at the time when all the item
measurement results are obtained before STAT sample cuts in and at the time
when all the measurement results are obtained after STAT sample has cut in.
16
4.5. The Record to be transmitted from Host to analyzer
There is no discrimination between capital letters and small letters in Sender ID and
Receiver ID.
Example of transmission
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
17
V1.06
In case there is no patient information to transmit, leave the optional field blank.
In case the patient ID is blank, other optional field are ignored.
18
4.5.3. Measurement Order Record (From Host to analyzer)
As test item number and item name are set or modified by each laboratory, they
are not defined in the specifications. Host system should have compatible item
number and item name list registered in analyzer. Please get the list from the
person in the laboratory.
Note:
In case of measurement of urine sample, ISE (Na, K and Cl) and other items
cannot be measured at the same time.
The reason is that the uring samples for ISE are measured after six times dilution
by uring diluent. Therefore, the result is automatically multiplied by six and the
transmitted.
Also, in case of dialysis sample, please select “Dialysis” as a specimen in the order
screen.
20
Example of discrimination of test item
(The following three records mean the same meaning.)
O|1|12345|^1^1|^^^22^^0|R||||||N||||Serum||||||||||O<CR> *Omitted test item name
O|1|12345|^1^1|^^^^PTT^0|R||||||N||||Serum||||||||||O<CR> *Omitted test item No.
O|1|12345|^1^1|^^^22^PTT^0|R||||||N||||Serum||||||||||O<CR>
The following record is null in case that the test item name of test item No. 78 is “Na”.
O|1|12345|^1^1|^^^40^Na^0|R||||||N||||Serum||||||||||O<CR>
In case that Host rejects an enquiry from analyzer, this record is transmitted.
If the condition code of the enquiry record is not “X”, this record is ignored.
21
V1.06
Example of transmission
C|1|L|Not in Service|G<CR>
Example of transmission
L|1|N<CR>
22
5. Message
Message is a aggregation of records from the message header record to the message
terminator record.
The enquiry message is the message that analyzer inquires of Host about the
measurement order.
Analyzer will transmit inquiry message at the timing of following three cases.
1) When analyzer inquiry button is pressed.
When an operator presses inquiry button, analyzer sends an inquiry message
to host.
Note:
Do not transmit all the measurement orders every time analyzer inquires to Host.
Except the case of changing the contents of measurement order, you do not
have to transmit the measurement order which has been once transmitted.
In case of “All” (batch) request, please transmit measurement order enquiry when
order is decided on host side. The order enquiry transmit when press “Request”
button. In case of “All” (batch) request, please transmit no measurement
order enquiry when order is indetermination. ( Refer to example of response
No.3 : In case of no measurement order.)
23
V1.06
Example of response No. 2 :in case of designating measurement items using multiple
measurement order records
(Example of transmission : by Sample ID)
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1<CR>
O|1|123456|0|^^^1^GOT^0|R||||||N||||Serum||||||||||O<CR>
O|2|123456|0|^^^11^LDH^0|R||||||N||||Serum||||||||||O<CR>
O|3|123456|0|^^^42^Ca^0|R||||||N||||Serum||||||||||O<CR>
L|1|N<CR>
Note:
In this example of transmission, only the first measurement order record
is effective.The measurement item of sample ID 123456 is only 1 (GOT).
When requesting measurement items 1 (GOT), 11 (LDH) and 42 (Ca),
request them by a single measurement order record using repeat delimiters.
The followings are the timings for Host to transmit the measurement order messages.
When Host transmits a measurement request to analyzer under the same sample ID for
the second time , analyzer will take different registration action according to the action
code specified in the measurement order record.
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|1234|^1^1|^^^1^GOT^0¥^^^79^K^0|R||||||N||||Serum||||||||||O<CR>
L|1|N<CR>
If transmitted under action code “N”, the registration is over written with the second
measurement request transmitted.
The measurement items of the sample ID 1234 transmitted first are changed into “GOT”,
“K” from “GOT”, “GPT” by the measurement request transmitted for the second time.
However if designated sample is under testing or has already been tested,and
measurement request is refused.
25
V1.06
If transmitted under action code “A”, the measurement request of the second
transmittion is registered additionally.
The measurement items of the sample ID 1234 transmitted for the first time are
changed into “GOT”, “GPT”, “K” from “GOT”, “GPT” by the measurement request
transmitted for the second time. (“GOT” is over written, “K” is new registeration.)
If the measurement requested first has finished when second request is transmitted, the
second request is registered as the rerun order. The measurement request items, that
will be registered as rerun order items, are “GOT” and “K” of the second transmission.
If the measurement request of the first transmission is under testing, the second
measurement request is refused.
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|123|||Guillen^Carlos||20000101|M<CR>
O|1|1234|^1^1|^^^1^GOT^0¥^^^2^GPT^0¥^^^3^ALP^0|R||||||N||||Serum||||||||||O<CR>
P|2|124|||Martinez^Edgar||19991231|M|||||Mark McLemore<CR>
O|1|1235|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||O<CR>
O|2|1236|^1^1|^^^1^GOT^0|R||||||N||||Urine||||||||||O<CR>
L|1|N<CR>
26
Example of transmission No. 3 : the measurement orders not gathered for each patient
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|1234|^1^1|^^^1^GOT^0¥^^^2^GPT^0¥^^^3^ALP^0|R||||||N||||Serum||||||||||O<CR>
P|2|<CR>
O|1|1235|^1^1|^^^1^GOT^0¥^^^2^GPT^0¥^^^3^ALP^0|R||||||N||||Serum||||||||||O<CR>
P|3|123|||Olerud^John||20000101|M<CR>
O|1|1236|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||O<CR>
P|4|123|||Olerud^John||20000101|M<CR>
O|1|1237|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||O<CR>
P|5|123|||Olerud^John||20000101|M<CR>
O|1|1238|^1^1|^^^1^GOT^3|R||||||||N|||||Serum|||||||||||O<CR>
L|1|N<CR>
Example of transmittion:
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|QC001|^1^C7|^^^1^GOT^0¥^^^2^GPT^0|R||||||Q||||Serum||||||||||O<CR>
O|2|QC002|^1^C8|^^^1^GOT^0¥^^^2^GPT^0|R||||||Q||||Serum||||||||||O<CR>
O|3|QC003|^1^20|^^^1^GOT^0¥^^^2^GPT^0|R||||||Q||||Serum||||||||||O<CR>
L|1|N<CR>
27
V1.06
Example of transmittion:
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|STAT_001|^1^E1|^^^1^GOT^0¥^^^2^GPT^0|S||||||N||||Serum||||||||||O<CR>
O|2|STAT_002|^1^E2|^^^1^GOT^0¥^^^2^GPT^0|S||||||N||||Serum||||||||||O<CR>
O|3|STAT_003|^1^8|^^^1^GOT^0¥^^^2^GPT^0|S||||||N||||Serum||||||||||O<CR>
L|1|N<CR>
28
5.2.6. Measurement Order Refusal (From analyzer to Host)
Example of response
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1|SAMPLE_1|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||X<CR>
C|1|I|ASTM_9_4_1|I<CR> … SAMPLE_1 message level abnormal
O|2|SAMPLE_3|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||X<CR>
C|1|I|ASTM_9_4_2|I<CR> … SAMPLE_3 sequence number abnormal
O|3|SAMPLE_4|^1^C1|^^^1^GOT^0|R||||||N||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_4 sample position abnormal
O|4|SAMPLE_5|^1^C1|^^^1^GOT^0|S||||||Q||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_5 sample position abnormal
O|5|SAMPLE_6|^1^E1|^^^1^GOT^0|R||||||N||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_6 sample position abnormal
O|6|SAMPLE_7|^1^E1|^^^1^GOT^0|S||||||Q||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_7 sample position abnormal
L|1|N<CR>
29
V1.06
The timing for analyzer to transmit the measurement result message is the followings.
Example of transmission
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0¥^^^27^TG^0|R||||||N||||Serum||||||||||O<CR>
R|1|^^^1^GOT^0|21.5143|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^0|8.5793|IU/L|4 TO 44|N||F||||20010530192517<CR>
C|1|I|Operator's Comment|G<CR>
R|3|^^^27^TG^0|381.596|mg/dl|50 TO 130|H||F||||20010530192520<CR>
L|1|N<CR>
30
Example of transmission: in case that there are remeasured data.
H|¥^&|||BIOLIS NEO^System1|||||Host^PC1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0¥^^^27^TG^0|R||||||N||||Serum||||||||||F
<CR>
R|1|^^^1^GOT^0|21.5143|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^0||IU/L|4 TO 44|N||X||||20010530192517<CR>
R|3|^^^2^GPT^1||IU/L|4 TO 44|N||X||||20010530192931<CR>
R|4|^^^2^GPT^2|8.6212|IU/L|4 TO 44|N||F||||20010530193551<CR>
R|5|^^^27^TG^0|381.596|mg/dl|50 TO 130|H||F||||20010530192520<CR>
L|1|N<CR>
“State of result” of the measurement result record for the item which is in the waiting
state is transmitted as the waiting result of “I”. (Without measurement result)
When the measurement is completed, the measurement result will be retransmitted
with the state of result, “F”.
H|¥^&|||BIOLIS NEO^System1|||||Host^PC1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0¥^^^27^TG^0|R||||||N||||Serum||||||||||F<CR>
R|1|^^^1^GOT^0|21.5143|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^1|230.5687|IU/L|4 TO 44|N||F||||20010530192931<CR>
R|3|^^^2^GPT^2||IU/L|4 TO 44|N||I||||<CR>
R|4|^^^27^TG^0|381.596|mg/dl|50 TO 130|H||F||||20010530192520<CR>
L|1|N<CR>
In case of the measuring failure of sample, the following measurement result comes on.
And the cause of the measuring failure is included in the comment record.
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0|R||||||N||||Serum||||||||||F<CR>
R|1|^^^1^GOT^0|0.02|IU/L|8 TO 38|N||P||||20010530192515<CR>
C|1|I|1|I<CR>
R|2|^^^2^GPT^0||IU/L|4 TO 44|N||X||||20010530192517<CR>
C|1|I|2|I<CR>
L|1|N<CR>
31
V1.06
Sample which measured by simple operation mode is transferred as sample round No.
991-999.
Sample No. is transferred as sample No. 1-72.
The result is not transferred automatically.
You have to do transfer (Batch) from R&E screen in normal operation mode.
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^1601|^^^1^GOT^0¥^^^2^GPT^0|R||||||N||||Serum||||||||||F<CR>
R|1|^^^1^GOT^0|30.0224|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^0|25.2268|IU/L|4 TO 44|N||F||||20010530192517<CR>
L|1|N<CR>
32
Appendix A Error Code List
33
V1.06
Click here.
Click
Delimiter definition
ComponentDelimiter : ComponentDelimiter example: ^
EscapeDelimiter : EscapeDelimiter example :&
FieldDelimiter : FieldDelimiter example :|
RepeatDelimiter : RepeatDelimiter example : ¥
Sender ID
HostName : Sender name example: BIOLIS 50i
System ID
HostSystemID : Receive name example: HOST
35
V1.06
Within ± 10°
・Label direction
Not specified.
Note:
This specification is based on barcode reader specification.
Please note that barcode reader may not read if you edit the barcode to be widened or
narrowed. This barcod reader is a fixed type.
When thermal transfer printer is used to print the barcode, barcode reader may not
read affected by the condition of the printing paper and printer head temperature.
36
C-2 Label paste position
1. 5 mL 12.4 x 75 mm
2. 7 mL 12.4 x 100 mm
3. 7 mL 15.4 x 75 mm
4. 10 mL 15.4 x 100 mm
13
64 Lower limit
55
35
20
14
100mm 75mm
Note:
When affixing the barcode label ensure the bottom of the barcode label
is20mm above the base of the tube.
In case of using a 12.4 x 100 mm tube, the bottom of the barcode labe
should be 35mm from the the base of the tube to avoid turning of tube.
37
V1.06
When transmitting the past data to your host computer, you need to change the
system date on analyzer.
Procedures:
Click System
Click Date/Print
Select the date which data is to be transmitted a date of the data you wish to
transmit on the displayed calendar.
Click HOST Send button at upper left position of the screen. The following screen
shows.
Send to Host “by Sample No.” screen Send to Host “bySample ID(Barcode)” screen
38
Input the lowest sample number or first sample ID of data that you wish to transmit
in the Start Sample No. or Start Sample ID box.
Input the highest sample number or last sample ID of data that you wish to transmit
in the End Sample No. or Start Sample ID box.
The results which are assigned by sample No. or sample ID will be sent to the Host.
Please note that it takes some time to transmit data to the host computer. It takes
about 3minutes to transmit data for ten samples that have ten test items.
If the system is under operation, it might take more time.
Inportant Note:
・ After the transmission of the past date is finished, follow the same procedure and
ensure the system date is changed back to the current date.
・ In case you transmit the results which measured by simple mode, input No.
991~999 to “Round No.”( Input used round No.)
39
V1.06
Tansmission procedures:
Click R & E button.
Send to Host “by Sample No.” screen Send to Host “bySample ID(Barcode)” screen
Input the lowest sample number or sample ID that you wish to transmit to the Start
Sample No. or Start Sample ID box.
Input the highest sample number or sample ID that you wish to transmit to the End
Sample No. or Start Sample ID box.
The results which are assigned by Sample No. send to the Host.
Please note that it takes time to transmit data to the host computer. It takes about
3minutes to transmit data for ten samples that have ten test items.
If the system is under operation, it might take more time.
Inportant Note:
・ After the transmission of the current date is finished.
In case you transmit the past results, please refer to the transmission of
previus data.
40
Appendix E ISE Error Code List
Code Contents
At measurement of sample or
100000 Noise error of Na electrode :
Calibrator 2
At measurement of sample or
200000 Noise error of K electrode :
Calibrator 2
Noise error of Na and K At measurement of sample or
300000 :
electrodes Calibrator 2
At measurement of sample or
400000 Noise error of Cl electrode :
Calibrator 2
Noise error of Na and Cl At measurement of sample or
500000 :
Electrodes Calibrator 2
Noise error of K and Cl At measurement of sample or
600000 :
electrodes Calibrator 2
Noise error of Na, K and Cl At measurement of sample or
700000 :
electrodes Calibrator 2
S00000 Air detected by the sensor : At measurement of sample
B00000 Air detected by the sensor : At measurement Calibrator 2
Defective drainage detected by At measurement of sample or
E00000 :
the flow senosor Calibrator 2
010000 Noise error of Na electrode : At measurement of Calibrator 1
020000 Noise error of K electrode : At measurement of Calibrator 1
Noise error of Na and K
030000 : At measurement of Calibrator 1
electrodes
040000 Noise error of Cl electrode : At measurement of Calibrator 1
Noise error of Na and Cl
050000 : At measurement of Calibrator 1
Electrodes
Noise error of K and Cl
060000 : At measurement of Calibrator 1
electrodes
Noise error of Na, K and Cl
070000 : At measurement of Calibrator 1
electrodes
0A0000 Air detected by the sensor : At measurement of Calibrator 1
Defective drainage detected by the
0E0000 : At measurement of Calibrator 1
flow sensor
Electric potential drift error at
001000 : At measurement of Calibrator 1
Na electrode.
Electric potential drift error at
002000 : At measurement of Calibrator 1
K electrode.
Electric potential drift error at
003000 : At measurement of Calibrator 1
Na and K electrodes.
Electric potential drift error at
004000 : At measurement of Calibrator 1
Cl electrode.
Electric potential drift error at
005000 : At measurement of Calibrator 1
Na and Cl electrodes.
Electric potential drift error at
006000 : At measurement of Calibrator 1
K and Cl electrodes.
Electric potential drift error at
007000 : At measurement of Calibrator 1
Na, K and Cl electrodes.
Electric potential over range of Na
000100 : Out of range
electrode
41
V1.06
Code Contents
Electric potential over range of
000200 : Out of range
K electrode
Electric potential over range of
000300 : Out of range
Na and K electrodes
Electric potential over range of
000400 : Out of range
Cl electrodes
Electric potential over range of
000500 : Out of range
Na and Cl electrodes
Electric potential over range of
000600 : Out of range
K and Cl electrodes
Electric potential over range of
000700 : Out of range
Na, K and Cl electrodes
000010 Slope error of Na electrode : Out of slope range, slope drift
42
Appendix F ASTM Communication Setting and Analyzer movement
43
V1.06
↓
Detect sample container. If sample cup or tube detected, measurement will
start. If not, error message will be displayed
↓
Resutlts are transmitted to the host each time when all the results for one
sample has been obtained
Note:
Enquiry is ”ALL” inquiry. Please transmit measurement order enquiry
when order is decided on host side. Please transmit no measurement order
enquiry when order is indetermination. ( Refer to example of response No.3 :
In case of no measurement order.)
Note:
Enquiry is “ALL” inquiry. Transmit text without request to analyzer.
When the request has been confirmed, transmit to the analyzer. It can be
received at any time.
44
●In case of by Sample ID (barcode) and “Appoint”.
Click “Start” button.
↓
Sample ID (barcode) enquiry
* If the ID can not be read 5 times in a row or the End barcode is read,
barcode reading will be stop.
↓
Transmit ID (barcode) requested from Host to analyzer.
↓
Detect sample container. If sample cup or tube detected, measurement will
start. If not, error message will be displayed
↓
Resutlts are transmitted to the host each time when all the results for one
sample has been obtained
Note:
In case of start button pressed after “sampling stop” status , it inquires from
the next sample number. If you press start button after “End of run”,
it inquires from the first number of sample tray.
45
V1.06