Errata: April 2007

ACLS Course and ACLS Renewal Course Video


[ISBN No 87493-498-2, AHA Product Number 80-1090]

Location Error Correction


In the “Megacode This section shows a classroom simulation of a It was not our intent to suggest that 200 Joules is,
and Team code team using a monitor/defibrillator. In in reality, the correct energy for this specific
Resuscitation order to avoid any implication of endorsement make and model of defibrillator. Viewers should
Concept” section of a specific brand, the defibrillator model be aware that the monitor/defibrillator shown in
number, manufacturer’s logo and brand name the ACLS video is a Philips MRx manual
were not shown on camera. defibrillator and the manufacturer’s recommended
energy setting for conversion of VF is 150 J.
The team leader announces, “We've got v-fib.
This biphasic unit uses 200 joules. Let's shock at This issue will be corrected on the next reprint of
200 joules." The close-up shot on the video the video. The team leader will state, “I don't
then briefly shows the monitor/defibrillator dial know the recommended settings for this biphasic
set to 200 J. Unfortunately users familiar with unit.” On-screen text will also appear stating, “If
the make and model of the device that is shown known, use the manufacturer's recommended
in the video may become confused by the settings.”
discrepancy between the energy setting that is
used in the video and the dose recommended by
the device manufacturer.