Professional Documents
Culture Documents
CAPILLARY
ELECTROPHORESIS
1
Course Learning Outcomes
Students should be able to:
1.Understand the concepts & principles of
separation in CE.
2.Compare the separation occurs in CE with
chromatographic separation.
2
Introduction
+ -
4
Capillary Electrophoresis
Is a new separation technique to separate minute
quantities of ionic species (based on charge) in
relatively short time with high resolution.
5
Basic design of CE instrumentation
6
CE instrument
7
Main components in CE :
✔ A fused silica capillary.
✔ 2 buffer reservoirs; source vial & destination
vial.
✔ 2 electrodes.
✔ A high-voltage power supply.
✔ A detector.
✔ A data output & handling device.
8
Capillaries
The capillaries used are normally narrow bore (10-100
µm I.D., 30-100 cm long) fused silica capillaries
covered with an external polyimide.
9
Capillaries are typically
30-100 cm long with 10-100
µm I.D..
10
Capillary cartridge
Detection
window
11
Sample introduction
12
Basic design of CE instrumentation
13
The migration of the analytes is then initiated by an electric
field that is applied between the source & destination vials
& is supplied to the electrodes by the high-voltage power
supply (5-30 kV dc).
14
The analytes separate as they migrate due to their
electrophoretic mobility & are detected near the outlet
end of the capillary.
15
Detector are similar to HPLC (spectrometry :
fluorescence, UV or UV-Vis absorbance;
electrochemical : conductivity, amperometry).
16
Electropherogram
17
Movement of ions in the capillary as a result of :
1. Electrophoretic flow
2. Electroosmosis flow (EOF)
18
Electrophoretic flow
19
Cations migrate toward the negatively charged
electrode (cathode).
+ -
#
20
Electrophoretic flow velocity is proportional to :
21
Migration rates
The migration of solutes is determined by their charge
to mass ratio (the larger the ratio, the faster an
ion migrate in the electric field).
✔ Small highly charged solutes will migrate more quickly then
large less charged solutes.
✔ If 2 ions are the same size, the one with greater charge will
move the fastest.
✔ For ions of the same charge, the smaller particle has less
friction & overall faster migration rate.
22
The length of the arrow indicates the magnitude of the
velocity; the direction of the arrow indicates of the
motion. The positive electrode in the right & the negative
electrode in the left.
23
Electroosmotic flow (EOF)
24
EOF caused by the electric double layer that develops at
the silica/solution interface.
25
Buffer cations congregate in an electric double layer
adjacent to the negative surface of the silica capillary
(i.e. fixed & mobile layer).
26
The cations in the mobile layer are attracted to
cathode.
27
Depiction of the interior of a fused silica
capillary in the presence of buffer
solution
28
Solvated cations drag solvent molecules during the
migration, hence there is net solution movement from
anode to cathode.
29
Since the EOF of the buffer solution is generally greater
than that of the electrophoretic flow of the analytes, all
analytes are carried along with the buffer solution
towards the cathode.
30
The order of migration will be cations, neutrals &
anions.
31
Diagram of the separation of charged & neutral
analytes (A) in capillary according to their respective
electrophoretic & EOF mobilities.
Electroosmotic flow
EOF
Electrophoretic flow
+ -
Electrophoretic flow + EOF
EOF
+ -
tm (min)
0
Sample problem
35
Cations eluted first, the greater the charge to mass ratio, the
greater the EOF, the faster the elution for 20X+ compared to
25 +
X.
36
The significance of conducting electrophoresis at high
pH is to establish EOF.
37
Effect of pH on the EOF
38
Velocity of EOF depends on :
1. Charge density of the capillary wall (charge density α
pH of the buffer solution).
The degree of ionization of silica is controlled by the
pH of the buffer.
EOF will increase with pH until all the available Si-OH
lining the wall of the capillary are fully ionized.
If the pH is increases, charge density increases,the
magnitude of EOF increased & thus decrease the
migration time.
39
2. Ionic strength of buffer.
40
Viscosity of the running buffer effect on the migration
time :
41
To reduce velocity of EOF :
42
Advantages of EOF :
43
Sample problem 2
EOF is pH dependent. Explain why & how does this affect
the migration time.
44
Sample problem 3
Consider a CE expt. designed to separate 5
components with similar masses. At pH 6.7, the
components are A+, B2+, C-, D & E (neutral species).
The electrophoresis was run with the injection end
positive & the detection end negative. The EOF was
greater than the electrophoretic flow.
45
i. With reasons, describe the order of elution.
46
ii. Explain what would happen to the separation if the
pH is decreased to 4 (assume that the charges of
the components do not change).
47
iii. Explain what would happen to the separation if the
silanols on the capillary wall were partially ionized.
48
iv. The pH was decreased to 2.0 (assume that the
charges of the components do not change).
49
Efficiency of CE
The efficiency of CE separations is typically
much higher than other separation techniques
like HPLC.
Cross-section of a capillary :
50
In CE, a very narrow open-tubular capillary is
used : Lower H
51
Flow profile
In electrically driven systems (CE), the driving force
of the EOF is uniformly distributed along the entire
length of the capillary (except right at the wall where
the double layer is fixed).
52
Capillary flow profiles
53
Result : EOF does not contribute significantly
to band broadening like pressure-driven flow
in LC & related techniques.
54
In pressure-driven flow :
55
Applications of CE
Due to its high efficiency & resolution, CE can be
used to separate & determine a wide variety of
comps :
✔ Simple inorganic ions, metal ions,
oligosacharides, nucleic acids (RNA & DNA) &
proteins.
Commonly used to analyze larger, water-soluble
biomolecules.
56
CE is a attractive technique :
✔ Requires small amounts of buffer & sample to
perform several analysis.
✔ Capable of separating comps with different size to
charge ratios.
57
Modes of capillary electrophoresis
58
Capillary Zone Electrophoresis (CZE)
Known as capillary electrophoresis.
59
Fundamental of CZE are constant field strength &
buffer solution throughout the length of the capillary.
60
Completely resolved zones have regions of
buffer between them.
61
Separation of cations in CZE :
62
Separation of anions in CZE :
✔ The EOF is reversed by treating capillary wall with a
cationic surfactant (alkyl ammonium salt such as
ethyl ammonium bromide)..
63
Charge reversal created by a cationic surfactant coated
on the capillary wall. The diffuse layer contains excess
anions & EOF is in the opposite direction.
64
Another approach for the separation of anions
is to operate at a low pH where the silanol
groups are not ionized & electrophoretic
migration of anions dominates towards the
anode.
65
IMPORTANT
NOTES
If the capillary wall is negative, EOF is toward the
cathode & the order of elution is cations >
neutrals > anions.
66
Sample problem 4
67
Describe how the original CE expt. would be different if the
capillary wall is treated with ethyl ammonium bromide.
Include the order of elution.
68
Sample problem 5
You are required to separate a sample mixture of anions at
pH4 using CE. Describe how you would position the major
components of the instrument (include treatment of column,
if required).
69
Capillary Gel Electrophoresis (CGE)
70
This sieving action is particularly helpful in
separating macromolecules (large Mw) such as
proteins & DNA fragments where they have the
same charge but differ in size.
71
+ -
73
Micellar Electrokinetic Capillary
Chromatography (MEKC or MECC)
74
MEKC is a form of chromatography in which surfactants
are added to the buffer solution at concentrations that
form micelles.
E.g. of anionic surfactant is sodium dodecyl sulfate
(SDS).
Hydrophobic Hydrophilic
75
Above a certain concentration, called the “critical
micelle concentration” (CMC), the surfactant
molecules will self-aggregate, forming micelles in
aqueous solution.
76
Even though these anionic micelles are attracted toward the anode
(since the micelles are negatively charged) which gives a large
electrophoretic mobility, they will still migrate toward the cathode
because of EOF.
78
If the concentration of SDS is too low (below the CMC
level), micelles will not form.
79
MEKC is a form of chromatography because the
micelles behave as a “pseudo-stationary phase” in
the capillary because their concentration is uniform
throughout the capillary.
80
Analyte will partition between the mobile phase (buffer)
& micelles (pseudo-stationary phase) as the analyte
travels through the capillary.
81
E.g. for SDS, migration order will be :
1. Anions – electrostatic repulsions from micelle.
i. More negatively charged anionic species.
ii. Less negatively charged anionic species.
2. Neutrals – hydrophobicity.
(Less hydrophobic, migration time faster, more
hydrophobic, spend more time in the micelle, so
migrate slowly).
3. Cations – attraction to micelle.
ii. Less positively charged species.
iii. More positively charged species.
82
Sample problem 7
Explain how micelles able to separate neutral compounds
in MEKC.
The interior of the micelles is non-polar, therefore capable
to absorb non-polar (neutral) analytes into the
hydrocarbon interior of the particles & solubilizing the
non-polar analytes. Partitioning of the analytes based on
their varying hydrophobicities. The exterior of the micelle is
polar making it soluble in water.
83
Sample problem 8
Analyte species (A: polar neutral, B: non polar neutral,
20 + 25 + 20 - 25 -
X , X , Y , Y ) in an aqueous sample were
separated by capillary electrophoresis (CE) in 15 mM
borate buffer (pH 5.8). The CE instrument was set up
with the detection end at cathode.
84
With the addition of anionic surfactant, micelles were
formed and moved towards cathode but slower than EOF.
Neutrals can be separated based on their varying
hydrophobicities. A was a hydrophilic neutral molecule,
thus spend almost no time inside the micelle and migrate
essentially at the same rate as the EOF and elute earlier.
Micelles are capable to absorb B as the hydrophobic
neutral molecule into the hydrocarbon interior of the
particles and move with the micelles and therefore eluted
later.
85
Sample problem 9
Four types of water soluble compounds; A and B
(neutral compounds whereby A is slightly more
hydrophobic), C (an anionic compound) and D (a
cationic compound) are separated by micellar
electrokinetic capillary chromatography (MEKC) in
15 mM borate buffer (pH 8) with 50 mM sodium
dodecyl sulfate.
86
i. With justification, predict the order of elution of the
above compounds.
87
ii. What would the order of elution have been in
the absence of sodium dodecyl sulfate?
1. Compound D
2. Compound A & B
3. Compound C
88
MECC is more versatile than CE because it can
separate neutral molecules & ionic species.
89
Advantages of MECC compared to HPLC.
90