Professional Documents
Culture Documents
Disposal
Please observe the applicable local or national regulations concerning
the disposal of “waste electrical and electronic equipment”.
Knick
Elektronische Messgeräte GmbH & Co. KG
Beuckestraße 22
14163 Berlin
Germany
2
Table of Contents
Safety Information.......................................................................... 5
Intended Use.....................................................................................................7
Registered Trademarks..................................................................................7
Quickstart Guides............................................................................................8
Overview of Stratos Eco 2405 Oxy............................................... 9
Assembly.........................................................................................10
Package Contents..........................................................................................10
Mounting Plan................................................................................................11
Pipe Mounting, Panel Mounting..............................................................12
Installation and Connection........................................................14
Installation Instructions..............................................................................14
Terminal Assignments.................................................................................14
Wiring Example..............................................................................17
Sensors with Connection via VP Cable .................................................17
Protective Wiring of Relay Outputs........................................................18
User Interface and Display..........................................................20
Operation: Keypad........................................................................22
Safety Functions............................................................................23
Sensocheck, Sensoface Sensor Monitoring.........................................23
GainCheck Device Self-Test.......................................................................23
Automatic Device Self-Test........................................................................23
Hold Mode.......................................................................................................24
Configuration.................................................................................26
Menu Structure of Configuration............................................................27
Overview of Configuration Steps............................................................28
Output 1...........................................................................................................30
Output 2...........................................................................................................38
Correction........................................................................................................44
Calibration Mode...........................................................................................46
Alarm Settings................................................................................................48
3
Table of Contents
Limit Function.................................................................................................50
Controlling a Rinsing System....................................................................52
Connecting a Rinsing System...................................................................53
Parameters......................................................................................54
Factory Settings of Parameters................................................................54
Parameters – Individual Settings.............................................................56
Calibration......................................................................................58
Calibration to Percent Saturation (SAT), in Water..............................60
Calibration to Concentration (Conc), in Air ........................................62
Zero Calibration ............................................................................................64
Product Calibration .....................................................................................66
Temp Probe Adjustment..............................................................69
Measurement.................................................................................69
Diagnostics Functions..................................................................70
Error Messages (Error Codes)......................................................72
Calibration Error Messages........................................................................74
Operating States............................................................................75
Sensoface........................................................................................77
Appendix.........................................................................................81
Product Line and Accessories...................................................................81
Specifications..................................................................................................82
Approvals – Canada......................................................................................87
Index................................................................................................89
Passcodes........................................................................................93
4
Safety Information
Safety information –
Be sure to read and observe the following instructions!
The device has been manufactured using state of the art technology
and it complies with applicable safety regulations.
When operating the device, certain conditions may nevertheless lead
to danger for the operator or damage to the device.
CAUTION!
Commissioning must be carried out by trained experts.
Whenever it is likely that protection has been impaired, the device
shall be made inoperative and secured against unintended operation.
CAUTION!
Before commissioning, make sure that the transmitter may be
connected with the other equipment.
5
6
Intended Use
Stratos Eco 2405 Oxy is used for dissolved oxygen and temperature
measurement in biotechnology, pharmaceutical industry, as well as in
the field of environment, food processing, and sewage treatment.
The sturdy molded enclosure can be fixed into a control panel or
mounted on a wall or at a post.
The protective hood provides additional protection against direct
weather exposure and mechanical damage.
The device has been designed for application with amperometric
sensors, e.g. Knick SE 703 / SE 706. It provides two current outputs
(for transmission of measured value and temperature, for example),
two contacts, and a universal power supply 24 ... 230 V AC/DC,
AC: 45 ... 65 Hz.
Registered Trademarks
7
Provided Documentation
Safety Instructions
In official EU languages and others.
• FM and Control Drawings
• EC Declarations of Conformity
Quickstart Guides
More languages on our website: www.knick.de
• Installation and Commissioning
• Operation
• Menu structure
• Calibration
• Error messages and recommended actions
8
Overview
Overview of Stratos Eco 2405 Oxy
Do not connect
terminal ! 1 Output 1 9 + Output 1
Anode 5 13 Relay 1
14 Not in use
RTD E
Temp Alarm
input 15 Alarm
RTD D
16 Alarm
Shield C
17 Cleaning
9
Assembly
Package Contents
Check the shipment for transport damage and completeness.
The package should contain:
• Front unit
• Rear unit
• Bag containing small parts
• CD-ROM with documentation
• Specific test report
• Passcode sticker
1
11
10 2
3
9
7 6 5 4
1 Jumper (2 x) 6 Sealing insert (1 x)
2 Washer (1 x), for conduit mounting: 7 Rubber reducer (1 x)
Place washer between enclosure and 8 Cable gland (3 x)
nut 9 Filler plug (3 x)
3 Cable tie (3 x) 10 Hexagon nut (5 x)
4 Hinge pin (1 x), insertable from either 11 Sealing plug (2 x), for sealing in case
side of wall mounting
5 Enclosure screw (4 x)
10
Assembly
Mounting Plan
144 105
27
144
15
42 21
84 43
1 2
32 80 1 Cable gland (3 x)
2 Knockouts for cable gland
or 1/2“ conduit, ø 21.5 mm
(2 knockouts) Conduits not
included!
72
(2 x)
11
Assembly
Pipe Mounting, Panel Mounting
132
40 60
1
2
3
4
165 132
1 1
173
Fig.: Z
U 0276 protective hood for wall and pipe mounting
(All dimensions in mm!)
12
Assembly
max. 25 1 Screw (4 x)
78 27 2 Gasket (1 x)
3 Control panel
4 Span piece (4 x)
5 Threaded sleeve (4 x)
Panel cut-out
1
138 x 138 mm (DIN 43700)
5 4
2
1...22 3
13
Installation and Connection
Installation Instructions
CAUTION!
• Installation of the Stratos must be carried out by trained experts in
accordance with this user manual and as per applicable local and
national codes.
• Be sure to observe the technical specifications and input ratings
during installation.
• Be sure not to notch the conductor when stripping the insulation.
• Before connecting the device to the power supply, make sure that
its voltage lies within the range 20.5 to 253V AC/DC.
• All parameters must be set by a system administrator prior to
commissioning.
The terminals are suitable for single wires and flexible leads up to
2.5 mm2 (AWG 14).
Terminal Assignments
14
Installation and Connection
4 3 1
1 ESD shield covering the signal inputs (Screw off for assembly)
Note: The cable shield must end under the ESD shield. (Cut lines if required.)
2 Terminals for temperature probe
3 Terminals for sensor
4 Power supply connection
Division 2 Wiring
The connections to the device must be installed
in accordance with the National Electric Code
(ANSI NFPA 70) Division 2 hazardous (classified)
location non-incendive wiring techniques.
15
16
Wiring Example
Sensors with Connection via VP Cable
Stratos Eco 2405 Oxy
Transparent
Yellow/Green
Red
Green
White
Gray
Blue
Please note:
The polarization voltage is factory set to 675 mV.
When using a sensor with a different polarization volt-
age, you must change this setting before connecting
the sensor! Otherwise the sensor may be damaged!
17
Protective Wiring of Relay Outputs
Protective Wiring of Relay Contacts
1
1 2
3 2
1 Load
2 RC combination, e.g. RIFA PMR 209
Typical RC combinations for 230 V AC:
Capacitor 0.1 µF / 630 V Resistor 100 ohms / 1 W
3 Contact
18
Protective Wiring of Relay Outputs
Typical Protective Wiring Measures
A1 Inductive load
A2 Free-wheeling diode, e.g. 1N4007 (Observe polarity)
A3 Contact
B1 Capacitive load
B2 Resistor, e.g. 8 Ω / 1 W at 24 V / 0.3 A
B3 Contact
C1 Incandescent lamp, max 60 W / 230 V, 30 W / 115 V
C3 Contact
WARNING!
Make sure that the maximum ratings of the relay contacts are not
exceeded even during switching!
19
User Interface and Display
User Interface
20
User Interface and Display
Display
1 2 3 4 5 6 7 8 9 10
20
11
19
18 12
13
17 16 15 14
21
User Interface and Display
Operation: Keypad
22
Safety Functions
Sensocheck, Sensoface Sensor Monitoring
Sensocheck continuously monitors the sensor and its wiring.
Sensocheck can be switched off (Configuration, page 48).
23
Safety Functions
Hold Mode
Display:
24
Safety Functions
Alarm
25
Configuration
In the Configuration mode you set the device parameters.
HOLD icon
Input errors The configuration parameters are
checked during the input. In the case
of an incorrect input ”Err” is displayed
for approx. 2 sec. The incorrect param-
eters cannot be stored. Input must be
repeated.
Exit conf Exit by pressing conf. The measured
value and Hold are displayed alternately,
“enter” blinks.
enter Press enter key to exit the Hold mode.
The measured value is displayed.
The output current remains frozen
for another 20 sec (HOLD icon on,
“hourglass” blinks).
26
Configuration
Menu Structure of Configuration
Select Select
Menu group Code Display
menu group menu item
Output 1
enter
Menu item 1
Menu item 2 enter
...
enter
Menu item ...
enter
Output 2
Correction
Calibration mode
Alarm settings
Previous
Relay menu group:
Rinsing probes
27
Configuration
Overview of Configuration Steps
28
Configuration
29
Configuration
Output 1
Select sensor type. Process variable
3
4
enter enter
Select sensor type*
Select process variable
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for error
HOLD mode
5
conf enter
Note: Stratos Eco 2405 Oxy has as device a resolution of 0.01 ppm.
30
Configuration
Code Display Action Choices
Select sensor type A / B Type A
(see table on left-hand side) (SE 703/
Select using. SE 706)
Press enter to proceed. Type B
(sensors with
higher current)
Select process variable (valid %
for all following settings): mg/l
• SAt: Saturation (%) ppm
• Conc: Concentration
(mg/l or ppm)
Select using.
Press enter to proceed.
31
Configuration
Output 1
Output current range, current start, current end
3
4
enter enter
Select sensor type
Select process variable
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for error
HOLD mode
5
conf enter
32
Configuration
Code Display Action Choices
Set output current range 4-20 mA
Select usingkey, (0 - 20 mA)
press enter to proceed.
33
Configuration
Output 1
Time constant of output filter
3
4
enter enter
Select sensor type
Select process variable
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for error
HOLD mode
5
conf enter
34
Configuration
Code Display Action Choices
Time constant of output filter 0 sec
Default setting: 0 s (inactive). 0 ... 120 sec
To specify a time constant:
Select usingkey,
edit number using key.
Press enter to proceed.
Please note:
The filter only acts on the current output, not on the display or the
limit value!
35
Configuration
Output 1
Output current during Error and HOLD
3
4
enter enter
Select sensor type
Select process variable
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for error
HOLD mode
5
conf enter
36
Configuration
Code Display Action Choices
22 mA signal for error OFF
message (OFF / ON)
Select usingkey,
press enter to proceed.
37
Configuration
Output 2
Temperature unit and probe, output current
Output 2:
3
4
enter enter
Select °C/°F
Select temp probe
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for temp error
HOLD mode
5
conf enter
38
Configuration
Code Display Action Choices
Specify temperature unit °C
Select usingkey, (°C / °F)
press enter to proceed.
40
20 Output current Output current
0 20
4 20 [mA] 4 20 [mA]
39
Configuration
Output 2
Time constant of output filter
Output 2:
3
4
enter enter
Select °C/°F
Select temp probe
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for temp error
HOLD mode
5
conf enter
40
Configuration
Code Display Action Choices
Time constant of output filter 0 sec
Default setting: (0 ... 120 sec)
0 sec (inactive). To specify a
time constant: Select using
key, edit number using
key. Press enter to proceed.
Please note:
The filter only acts on the current output, not on the display!
41
Configuration
Output 2
Temperature error, output current during HOLD
Output 2:
3
4
enter enter
Select °C/°F
Select temp probe
Select 0-20 / 4-20 mA
Enter current start
Enter current end
Set output filter
22 mA for temp error
HOLD mode
5
conf enter
42
Configuration
Code Display Action Choices
22 mA signal for error
message
Select usingkey,
press enter to proceed.
43
Configuration
Correction
Polarization voltage / Membrane temperature compensation /
Process pressure / Salinity correction
Correction
3
4
enter enter
Polarization voltage
Membrane temperature
compensation
Meas. unit (pressure)
Process pressure
Salinity correction
5
conf enter
44
Configuration
Code Display Action Choices
Enter polarization voltage
Select usingkey,
edit number usingkey.
Press enter to proceed.
Membrane temperature 01.00
compensation
Select position using
key and edit number using
key.
Press enter to proceed.
Select pressure unit
Select usingkey,
press enter to proceed.
Please note:
When using a sensor with a polarization voltage different from 675 mV
(factory setting), you must enter the correct voltage before connect-
ing the sensor!
Otherwise the sensor may be damaged!
45
Configuration
Calibration Mode
Calibration mode:
3
4
enter enter
Calibration mode
Cal timer interval
5
conf enter
46
Configuration
Code Display Action Choices
Specify calibration mode
(calibration to saturation or
concentration)
Select usingkey,
press enter to proceed.
Please note:
When calibrating in air-saturated water (standard practice for biotech-
nological processes), you should select calibration to saturation (SAT).
47
Configuration
Alarm Settings
Alarm settings:
3
4
enter
Select Sensocheck
5
conf enter
48
Configuration
Code Display Action Choices
Select Sensocheck
(continuous monitoring of
sensor)
Select usingkey,
press enter to proceed.
49
Configuration
Limit Function
Relay
4
enter
Contact function
Contact response
Enter setpoint
Enter hysteresis
Limit function: Delay
3
5
enter conf enter
50
Configuration
Code Display Action Choices
Contact function
(see below for function
principle)
Select usingkey,
press enter to proceed.
Contact response
N/C: normally closed contact
N/O: normally open contact
Select usingkey,
press enter to proceed.
Setpoint 000.0 %
Select usingkey, (xxx.x %)
edit number using key.
Press enter to proceed.
Hysteresis 001.0 %
Select usingkey, (xxx.x %)
edit number using key.
Press enter to proceed.
Limit Lo Limit Hi
Signal Signal
Hysteresis +
Setpoint Setpoint
Hysteresis -
1 1
Contact Contact
0 0
51
Configuration
1 conf Controlling a Rinsing System
“Clean” contact
2
1 Press conf key.
2 Enter passcode 1200.
3 Select Rinsing probes menu group using
arrow keys. All items of this menu group are
indicated by the “Pb.” code.
4 Press enter to select menu, edit using arrow
keys (see next page).
Confirm (and proceed) using enter.
5 Exit: Press conf, then enter.
4
enter
Rinsing interval
Rinse duration
Contact response
Rinse contact:
3
5
enter conf enter
52
Configuration
Code Display Action Choices
Rinsing interval 0.000 h
Select using key, enter (x.xxx h)
number using ,
press enter to proceed.
Rinse duration 0060 s
Select using key, (xxxx s)
enter number using ,
press enter to proceed.
Rinsing system
Clean 17
18
Power supply
53
Parameters
Factory Settings of Parameters
Activation:
Simultaneously press conf + right arrow key and
enter passcode “4321“.
The lower display line reads “Clear“. To prevent accidental resetting,
“NO“ is set as default (blinking in the main display). Press one of the
arrow keys to select “YES“ and confirm by pressing enter.
CAUTION!
Your data (also calibration data) will be overwritten by the factory
settings!
54
Parameters
Code Parameter Factory setting
Polarization voltage
Membrane temp. compensation
Pressure unit
Pressure
Salinity
Calibration mode
Cal interval
Sensocheck
Contact function
Contact response
Setpoint
Hysteresis
Delay
Rinsing interval
Rinse duration
Contact type
Please note:
Fill in your configuration data on the following pages.
Please note:
Factory settings for the calibration data are 60.0 nA (slope) and
0.000 nA (zero).
55
Parameters
Parameters – Individual Settings
%, mg/l, ppm
0/4-20 mA
Current start
Current end
Filter time
22mA signal
HOLD response
Fix current
Unit °C / °F
Temp probe
Current start
Current end
Filter time
22mA signal
HOLD response
56
Parameters
Polarization voltage
Membrane temp.
compensation
Pressure unit
Pressure
Salinity
Calibration mode
Cal interval
Sensocheck
Contact function
Contact response
Setpoint
Hysteresis
Delay
Rinsing interval
Rinse duration
Contact type
57
Calibration
Calibration adjusts the device to the sensor.
Activation cal Activate by pressing cal
Enter passcode:
• Zero point 1001
• Water/Air 1100
Edit parameter using and.
Confirm and proceed by pressing enter.
(Exit by pressing cal, then enter.)
HOLD During calibration the device remains in
the Hold mode for reasons of safety. The
During configu- output current is frozen (at its last value
ration the device or at a preset fixed value, depending
remains in the on the configuration), limit and alarm
Hold mode. contacts are inactive. Sensoface is off,
“Calibration” mode indicator is on.
HOLD icon
Input errors The calibration parameters are checked
during the input. In the case of an incor-
rect input ”Err” is displayed for approx.
3 sec. The incorrect parameters cannot
be stored. Input must be repeated.
Exit enter Exit by pressing enter (abort using cal).
The measured value and Hold are dis-
played alternately, “enter” blinks.
enter Press enter to exit the Hold mode.
The measured value is displayed.
The output current remains frozen
for another 20 sec (HOLD icon on,
“hourglass” blinks).
58
Calibration
Information on Calibration
It is always recommended to calibrate in air.
Compared to water, air is a calibration medium which is easy to
handle, stable, and thus safe. In the most cases, however, the sensor
must be dismounted for a calibration in air. When dealing with bio-
technological processes which require sterile conditions, the sensor
cannot be removed for calibration. Here, calibration must be per-
formed with aeration directly in the process medium (e.g. after ster-
ilization). In the field of biotechnology, for example, often saturation
is measured and calibration is performed in the medium for reasons
of sterility. For other applications where concentration is measured
(water control etc.), calibration in air has proved to be useful.
Common Combination:
Process Variable / Calibration Mode / Calibration Medium
Process Cal Calibration Default rel. Default cal
variable mode humidity pressure
Saturation (%) Water 100 % Process
pressure
Concentration Air 50 % 1.013 bar
(mg/l, ppm)
Please note:
When a 2-point calibration is required, the zero calibration should be
performed prior to saturation or concentration calibration, resp.
All calibration procedures must be performed by trained personnel.
59
Calibration
Calibration to Percent Saturation (SAT), in Water
Display Action Remark
Press cal key, enter code 1100. SAT or Conc calibra-
Select using key, tion is selected during
edit number using key. configuration.
Press enter to proceed. Device switches to
Hold mode.
If an invalid code is
entered, the device
returns to measuring
mode.
Immerse sensor in cal medium
60
Calibration
61
Calibration
Calibration to Concentration (Conc), in Air
Display Action Remark
Press cal key, enter code 1100. SAT or Conc calibra-
Presskey to select position, tion is selected during
enter number usingkey. configuration.
Press enter to proceed. Device switches to
Hold mode.
If an invalid code is
entered, the device
returns to measuring
mode.
Place sensor in air
Press enter to start.
62
Calibration
63
Calibration
Zero Calibration
The sensor models SE 703 and SE 706 have a low zero current.
Therefore, you should not perform a zero calibration with Stratos Eco
2405. If you still wish to perform a zero calibration, the sensor should
remain for at least 10 to 30 minutes in an oxygen-free calibration
medium in order to obtain stable, non-drifting values. During zero
calibration, a drift check is not performed. Zero current of a properly
functioning sensor is notably less than 0.5 % of air current. The display
(secondary: measured value, main: entered value) does not change
until an input current is entered for the zero point, see page 65. When
measuring in an oxygen-free medium, the displayed current can be
taken directly.
64
Calibration
Display Action Remark
Select calibration (press cal key). Device is in the Hold
mode. If an invalid
code is entered, the
Enter passcode 1001.
device returns to
Select using key,
measuring mode.
edit number using key.
Press enter to proceed.
65
Calibration
Product Calibration
Calibration by comparison
66
Calibration
Display Action Remark
Save currently measured value. Perform reference
Press enter to proceed. measurement.
67
68
Temp Probe Adjustment
Display Action Remark
Select calibration Wrong settings
Press cal key, enter code 1015. change the measure-
Presskey to select position, ment properties! If
enter number usingkey, an invalid code is
confirm by pressing enter. entered, the device
returns to measuring
mode.
Ready for calibration Device is in Hold
Measure the temperature of the mode.
process medium using an exter- Display approx. 3 sec
nal thermometer.
Measurement
Display Action
In the measuring mode the main display shows the
configured process variable (%, mg/l, or ppm) and the
lower display the temperature.
During calibration you can return to measuring mode by
pressing the cal key, during configuration by pressing
conf (waiting time for signal stabilization approx.
20 sec).
69
Diagnostics Functions
Display Action
Display of output currents
Press enter while in measuring mode.
The current at output 1 is shown in the main display,
the current at output 2 in the secondary display.
After 5 sec the device returns to measuring mode.
Display of calibration data (Cal Info)
Press cal while in measuring mode and enter code 0000.
The slope is shown in the main display, the zero point in
the secondary display.
After 20 sec the device returns to measuring mode
(immediate return at pressing enter).
Sensor monitor (display of sensor current)
Press conf while in measuring mode and enter code
2222. The sensor current (without temperature com-
pensation) is shown in the main display, the measuring
temperature in the secondary display.
Press enter to return to measurement.
Display of last error message (Error Info)
Press conf while in measuring mode and enter code
0000.
The last error message is displayed for approx. 20 sec.
After that the message will be deleted (immediate return
to measurement at pressing enter).
70
Diagnostics Functions
These functions are used for testing the connected peripherals.
Display Action
Specify current for output 1
Press conf while in measuring mode
and enter code 5555.
The main display indicates the current provided by
output 1. This value can be edited.
To do so, select the position usingkey,
edit number using key.
Confirm entry by pressing enter. The entered value will
be shown in the secondary display.
The device is in Hold mode.
The specified current is output.
Press conf, then enter to return to measurement
(Hold remains active for another 20 sec).
Specify current for output 2
Press conf while in measuring mode
and enter code 5556.
The main display indicates the current provided by
output 2. This value can be edited.
To do so, select the position usingkey,
edit number using key.
Confirm entry by pressing enter. The entered value will
be shown in the secondary display.
The device is in Hold mode.
The specified current is output.
Press conf, then enter to return to measurement
(Hold remains active for another 20 sec).
71
Error Messages (Error Codes)
Alarm contact
Red LED
Possible causes
72
Error Messages (Error Codes)
Alarm contact
Red LED
(blinks) Possible causes
ERR 33 Sensocheck:
Sensor: Connecting cable defective
73
Calibration Error Messages
Problem
Icon blinks:
Possible causes
Slope out of range
Wrong calibration values specified
(relative humidity, pressure, saturation,
concentration)
Calibration aborted after 12 minutes
Sensor defective or dirty
• No electrolyte in the sensor
• Sensor cable insufficiently shielded or
defective
• Strong electric fields influence the
In addition “CAL Err” measurement
blinks. • Temperature fluctuation of calibration
solution
74
Operating States
limit value
Timeout
Cleaning
contact
Relay 1
Operating status
contact
Alarm
Out 1
Out 2
Measure
Cal Info
20 s
(cal) 0000
Error Info
20 s
(conf ) 0000
Calibration
(cal) 1100
Temp adjustment
(cal) 1015
Product calibration
(cal) 1105
Configuration
20 min
(conf ) 1200
Sensor monitor
20 min
(conf ) 2222
Current source 1
20 min
(conf ) 5555
Current source 2
20 min
(conf ) 5556
Rinsing function
active
75
76
Sensoface
(Sensocheck must have been activated during configuration.)
Notice
The worsening of a Sensoface criterion leads to the devaluation of the
Sensoface indicator (Smiley becomes “sad”). To reset the Sensoface
indicator, the defect must be remedied.
77
78
Sensoface
Type B Sensors (Sensors with Higher Current)
Response
Slope Zero point Cal timer
time
Permitted 200 ... 550 nA -2 ... +2 nA max. 720 s
range
> 250 ... < 500 nA > -0.5 ... < 0.5 nA < 300 s < 80 %
225 ... 250 nA -1.0 ... -0.5 nA or 300 ... 600 s 80 ... ≤ 100 %
or +0.5 ... +1.0 nA
500 ... 525 nA
Sensocheck
Continuously monitors the sensor and leads for short circuits or
open circuits. Critical values make the Sensoface “sad” and the
corresponding icon blinks:
79
80
Appendix
Product Line and Accessories
Mounting Accessories
Pipe-mount kit ZU 0274
Panel-mount kit ZU 0275
Protective hood ZU 0276
For more information concerning our sensors and fittings product line,
please refer to our website:
www.knick.de
81
Specifications
DO input
Measuring current –2 ... +1800 nA
Resolution 0.05 nA
(with Vpol ≤ 800 mV and
Vref ≤ 200 mV)
Saturation 0... 200%
(-10 ... 80 °C)
Meas. error1,2,3) 0.5 % meas.val. + 0.5 %
Concentration 0,00 ... 20.00 mg/l
(-10 ... 80 °C) 0.00 ... 20.00 ppm
Meas. error1,2,3) 0.5 % meas.val.
+ 0.05 mg/l or 0.05 ppm
Permitted guard current ≤ 20 µA
*
Polarization voltage 0 ... 1000 mV
Factor for membrane 00.50 ... 03.00
temp. compensation *
Process pressure* 0.000 ... 9.999 bars
( ... 999.9 kPa / ... 145.0 psi)
Salinity correction* 00.00 ... 45.00 g/kg
Sensor standardization
Operating modes * • O2 saturation (automatic)
• O2 concentration (automatic)
• Product calibration
• Zero calibration
Calibration range Zero point ± 2 nA
Sensor Type A Slope 25 ... 130 nA
(at 25 °C, 1013 mbars)
Calibration range Zero point ± 2 nA
Sensor Type B Slope 200 ... 550 nA
(at 25 °C, 1013 mbars)
Calibration timer * 0000 ... 9999 h
*
Pressure correction 0.000 ... 9.999 bars / 999.9 kPa / 145.0 psi
82
Specifications
Sensor monitoring
Sensocheck Monitoring for short circuits / open circuits
(can be disabled)
Sensoface Provides information on the sensor condition
(evaluation of zero/slope, response time,
calibration interval, Sensocheck)
83
Specifications
Alarm contact Relay contact, floating
Contact ratings AC< 250 V / < 3 A / < 750 VA
DC< 30 V / < 3 A / < 90 W
Contact response N/C (fail-safe type)
Response delay 10 s
84
Specifications
Service functions
Current source Current specifiable for output 1 and 2
(00.00 ... 22.00 mA)
Device self-test Automatic memory test (RAM, FLASH, EEPROM)
Display test Display of all segments
Last Error Display of last error occurred
Sensor monitor Display of direct, uncorrected sensor signal
Power supply 24 (–15%) ... 230 V AC/DC (+10%); approx. 5 VA, 2.5 W
AC: 45 ... 65 Hz
Overvoltage category II, protection class II
Explosion protection
FM NI Class I Div 2 Group A, B, C & D, T4
Ta = 55 °C; Type 2
NI Class I Zone 2 Group IIC, T4 Ta = 55°C; Type 2
85
Specifications
Enclosure Molded enclosure made of PBT,
glass bead reinforced
Color Black
Mounting • Wall mounting
• Pipe mounting: Ø 40 ... 60 mm 30 ... 45 mm
• Panel mounting,
cutout to DIN 43 700,
sealed against panel
Dimensions H 144 mm, W 144 mm, D 105 mm
Ingress protection IP 65 / NEMA 4X
Cable glands 3 knockouts for cable glands M20x1.5
2 knockouts for NPT 1/2" or
rigid metallic conduit
Weight Approx.1 kg
* User-defined
1) To IEC 746 Part 1, at nominal operating conditions
2) ± 1 count
3) Plus sensor error
86
Approvals – Canada
Warnings and Notes to Ensure Safe Operation
WARNING!
Do not disconnect equipment unless power has been switched off.
CAUTION!
Clean only with antistatic moistened cloth.
CAUTION!
Substitution of components may impair suitability for hazardous
locations.
87
Approvals – Canada
CAUTION!
Use supply wires suitable for 30 °C above ambient and rated at least
250 V.
CAUTION!
Use signal wires suitable for at least 250V.
88
Index
22 mA signal for error message 37, 43
A
Accessories 81
Alarm contact 49
Alarm setings 25, 48
Approvals 87
Assembly 10
C
Calibration 58
Calibration mode selection 46
Calibration to concentration (Conc) 62
Calibration to percent saturation (SAT) 60
Cal timer interval 46
Error messages 74
Product calibration 66
Zero calibration 64
Cal Info 70
“Clean” contact 52
Concentration calibration 62
Configuration 26
Menu structure 27
Configuration: Alarm settings 48
Configuration: Calibration mode 46
Configuration: Correction 44
Configuration: Limit function 50, 52
Configuration: Output 1 30
Output current during Error and HOLD 36
Output current range 32
Process variable 30
Sensor type 30
Time constant of output filter 34
Configuration: Output 2 38
Output current 38
Output current during HOLD 42
89
Index
Temperature 38
Temperature error 42
Time constant of output filter 40
Configuration: Rinsing probes 52
Connection 14
CSA warnings and notes 87
Current start / end 33, 39
D
Device self-test 23
Diagnostics functions 70
Display of calibration data 70
Display of last error message 70
Display of output currents 70
Display of sensor current 70
Specifying the output current 71
Display 21
Disposal 2
Division 2 wiring 15
Documentation 8
E
Error Info 70
Error messages 72
Display of last error message 70
Explosion protection 85
F
Factory settings of parameters 54
Front panel 20
H
HOLD mode 24
Output signal during HOLD 37, 43
Hysteresis 51
90
Index
I
Installation 14
Intended use 7
K
Keypad 22
M
Measurement 69
Mounting plan 11
O
Operating states 75
Output filter 34, 40
Overview 9
Overview of configuration steps 28
P
Package contents 10
Panel mounting 13
Panel-mount kit 13
Parameters
Factory settings 54
Individual settings 56
Passcodes 93
Pipe mounting 12
Pipe-mount kit 12
Polarization voltage 45
Pressure correction 45
Product calibration 66
Product line 81
Protective hood 12
Protective wiring 18
Q
Quickstart guides 8
91
Index
R
Relay 50, 52
Relay contacts 18
Rinse contact 52
Rinsing interval 53
Rinsing probes 52
Rinsing system 53
S
Safety information 5, 87
Salinity correction 45
Self-test 23
Sensocheck 23, 48, 79
Configuration 49
Sensoface 23, 77, 79
Sensor connection 17
Sensor monitor 70
Short instructions 8
Specifications 82
T
Temperature measurement, configuration 39
Temperature probe adjustment 69
Terminal assignments 14
Time constant of output filter 35, 41
Trademarks 7
U
User interface 20
W
Wiring example 17
Z
Zero calibration 64
92
Passcodes
Calibration
Key + passcode Menu item Page
CAL info (display of zero,slope) 70
Zero calibration 64
Calibration (water / air) 60
Product calibration 66
Temp probe adjustment 69
Configuration
Key + passcode Menu item Page
Error info (display of last error, erase) 70
Configuration 26
Sensor monitor (sensor current) 70
Current source 1 (specify output current) 71
Current source 2 (specify output current) 71
Factory setting 54
93
Knick
Elektronische Messgeräte
GmbH & Co. KG
Headquarters
Beuckestraße 22 • 14163 Berlin
Germany
Phone: +49 30 80191-0
Fax: +49 30 80191-200
info@knick.de
www.knick.de
Local Contacts
www.knick-international.com
096964
TA-194.433-KNEN05 20200406 Software version: 2.x