You are on page 1of 11

Micron 42 (2011) 175–185

Contents lists available at ScienceDirect

Micron
journal homepage: www.elsevier.com/locate/micron

Review

Low voltage high-resolution SEM (LVHRSEM) for biological structural and


molecular analysis
Heide Schatten ∗
Department of Veterinary Pathobiology, University of Missouri, 1600 E Rollins Street, Columbia, MO 65211, USA

a r t i c l e i n f o a b s t r a c t

Article history: High-resolution scanning electron microscopy (HRSEM) is being used increasingly to gain new insights
Received 1 April 2010 into three-dimensional organization of biological structure, macromolecular complexes and interactions
Received in revised form 22 August 2010 of cellular components as well as isolated cell organelles. Modern scanning electron microscopes (SEMs)
Accepted 23 August 2010
combined with adequate sample preparation can now provide resolution comparable with that achieved
using transmission electron microscopes (TEMs) down to 2–5 nm for biological material. The versatility
Keywords:
of the instrument and new sample preparation techniques have allowed detailed analysis of chromo-
High-resolution scanning electron
somes, cytoskeletal components, virus and other biological material that has not been possible with TEM.
microscopy (HRSEM)
Resolution
The present review addresses resolution and specific specimen preparations for HRSEM, and highlights
Specimen preparation the importance of specimen preparation and choice of methods to achieve optimal results for proteins,
Cytoskeleton macromolecular complexes and subcellular structures using low voltage HRSEM (LVHRSEM).
Cell structure © 2010 Published by Elsevier Ltd.
Virus

Contents

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
2. Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
3. General sample preparation for SEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
3.1. Conventional sample preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
3.2. Freezing methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
4. Specific sample preparations for specific structural and macromolecular analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
4.1. Chromosome structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
4.2. Isolated mitotic spindles, centrosomes, and nuclear envelopes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
4.3. De-embedding of Epon-embedded thick sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
5. Future perspectives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

1. Introduction

The appreciation for scanning electron microscopy (SEM) has Maconnachie, 1979) caused by ethanol dehydration and inade-
increased significantly in recent years with the recognition that quate critical point drying have been overcome by several refined
modern SEMs combined with adequate sample preparation can preparation methods, as well as cryofixation to remove unbound
provide resolution comparable with that achieved using trans- or extracellular water from the specimen while preserving bound
mission electron microscopes (TEMs). Furthermore, the versatility water to retain structural and molecular integrity (reviewed in
of the instrument and sample preparation resulted in multiple Erlandsen, 2008).
new approaches to utilize SEM for molecular and structural imag- The present review addresses resolution and specific speci-
ing. Early preparation techniques associated with shrinkage in men preparations for LVHRFESEM. It highlights the importance
cell dimensions and collapse of molecular structure (Boyde and of adequate specimen preparation and the choice of meth-
ods required to achieve optimal results for specific analyses of
proteins, macromolecular complexes and subcellular structures
∗ Corresponding author. Tel.: +1 573 882 2396; fax: +1 573 884 5414. using low voltage high-resolution field-emission scanning electron
E-mail address: SchattenH@missouri.edu. microscopy (LVHRFESEM).

0968-4328/$ – see front matter © 2010 Published by Elsevier Ltd.


doi:10.1016/j.micron.2010.08.008
176 H. Schatten / Micron 42 (2011) 175–185

2. Resolution gold particles (reviewed in detail by Pawley, 2008) that was one
of the keys to reliably detect immuno-labeled particles associated
The limitations of spatial resolution for biological samples in the with biological structure without steric interference that otherwise
SEM has been defined by Joy and Pawley (1992) as follows: “The oftentimes affects correct analysis of immuno-stained molecular
spatial resolution of the scanning electron microscope is limited by components (reviewed in Albrecht and Meyer, 2008). Analysis
at least three factors: the diameter of the electron probe, the size of immuno-labeled structures can either be achieved by using
and shape of the beam/specimen interaction volume with the solid different voltages and overlay of separate images (for example,
for the mode of imaging employed, and the Poisson statistics of the small gold particles have been detected at 5 kV while detection
detected signal.” Enormous progress in instrumentation and speci- of biological structure utilized 20 kV (Pawley and Albrecht, 1988;
men preparation development has moved SEM in close competition reviewed in Pawley, 2008)), or by employing a combination of
to TEM, achieving resolution of biological material down to the BSE and SE recording simultaneously (reviewed in Hermann et al.,
2–5 nm range and in some instances down to 1 nm. Improved reso- 1996).
lution has allowed gaining new knowledge on biological structure, The high resolution available with LVFESEM has also allowed
macromolecular complexes, and cellular interactions of intracel- more detailed immunolabeling of biological structure as detailed by
lular components, cell–matrix interactions, and isolated biological Albrecht and Meyer (2008). Small metal labels as small as 2–3 nm
material. However, there is no universal answer to the question on can be imaged on very thinly coated or completely uncoated spec-
resolution as it depends on a number of different factors and there imen.
is no standard method of measuring SEM resolution despite the Taken together, the development of electron sources, lenses,
search for an ideal bulk SEM test specimen (reviewed in Pawley, vacuum systems and magnetic shielding considerably contributed
2008; McMullan, 2008). to high-resolution LVSEM (reviewed by Pawley, 2008). Instrument
The past two decades have seen significant improvements in development in 1986 included an FE source, a short focal-length
resolving power of commercially available SEMs as a result of lens, and a good vacuum system that resulted in achieving res-
several developments including the use of field-emission cath- olution down to 3 nm at 1.5 kV and provided the long-desired
odes as electron sources (Crewe et al., 1968) and improved lens opportunity to use LVSEM for high-resolution topographic imaging.
designs (Pawley, 1992). The primary beam diameter of modern As mentioned above, it is now possible to use in-lens instruments
high-resolution SEMs is less than 1 nm (Nagatani et al., 1987). Spec- to achieve resolution of 1 nm at 1 kV and instruments in which the
imen preparation is still one of the challenging components in specimen is mounted below the final lens to achieve resolution of
achieving optimal resolution as will be discussed in Section 3. As 1.5 nm. The Hitachi S-900 at the Integrated Microscopy Resource
has been discussed by Walther (2008), field-emission SEMs of the in Madison, WI, was the first instrument to offer such high reso-
in-lens-type offer the smallest available probe size of less than 1 nm lution afforded by then important instrument modifications and
at V0 = 30 kV and about 1 nm at V0 = 10 kV (Nagatani et al., 1987). At thorough evaluation of sample preparations. This instrument in an
1 kV, however, the diameter increases and reaches about 3 nm at NIH-funded resource environment has served numerous investiga-
1 kV (Nagatani et al., 1987). tors involved in cutting-edge research (reviewed in Pawley, 2008)
Several excellent articles have addressed the various fac- offering ca. 5× better resolution than had been possible previously
tors affecting resolution in LVFESEM (reviewed in Pawley, 2008; (reviewed in Pawley, 2008). Several images obtained on this instru-
Erlandsen, 2008; Walther, 2008, and others). The history of SEM ment are shown in the present review paper. All manufacturers
development and improvement in resolution has been excellently of SEM instruments now offer capabilities that allow resolutions
reviewed by McMullan (2008) and progress has been remarkable down to 3–5 nm or even 1.5 nm under favorable specimen prepa-
since 1960 when a resolution of 200 nm at 1 keV was considered ration conditions using below the objective lens specimen imaging.
excellent. The development of ultra-high vacuum technology as Modern high-resolution SEMs allow 3 nm resolution at any V0
well as corrections for aberrations played a critical role in allowing above 1.5 kV. It has to be stressed that specimen preparation is
SEMs to be marketed (reviewed in Joy, 2008a,b; McMullan, 2008; crucially important to obtain high-resolution image results; the
Newbury, 2008) and overcome setbacks that had been experienced best instrumentation is only as good as the best specimen and
up to the middle 1960s. As mentioned above, by now the practi- lack of expertise may affect the results significantly. Therefore, a
cal resolution of LVSEM rivals that obtained with thin-section TEM major part of the present review is dedicated to specimen prepa-
which has yielded three-dimensional new information in stud- ration.
ies of cell structure; an additional advantage of using LVFESEM
includes detailed structural information with high spatial resolu-
3. General sample preparation for SEM
tion.
The early SEM instruments developed by the Oatley group
The importance of adequate sample preparation cannot be
around 1950 (Oatley, 1982; reviewed by Pawley, 2008) used
overemphasized, as the best instrumentation is only as good as the
hot tungsten filaments and were operated at 10–30 kV, although
sample and the best instrumentation cannot compensate for poor
Thornley used low V0 to avoid charging when using uncoated,
specimen quality. Maintaining structural integrity is an absolute
frozen biological samples (Thornley, 1960; Thornley and Cartz,
requirement for specimen preparation and subsequent analysis
1962). In the 1970s, LaB6 Schottky thermionic cathodes (Broers,
in the vacuum of the microscope. A variety of different sample
1974) and cold field-emission (FE) cathodes (Crewe et al., 1968;
preparation techniques are available that have been elaborated by
Crewe, 1973; Hainfeld, 1977) were introduced for SEM use. Further
various investigators. In this section, the most common specimen
developments included using a TEM-type objective lens allow-
preparation techniques are discussed. Specialized sample prepa-
ing collection of secondary electrons (SEs) and backscattered
ration techniques that had been worked out for specific research
electrons (BSEs) from a specimen mounted inside the lens field
applications are featured in Section 4.
which reduced lens-aberration coefficients by a factor of up to
20 (reviewed in Pawley, 2008). Later instrument improvements
allowed FESEM performance below the lens which is now used 3.1. Conventional sample preparation
in most modern FESEMs. Improvement in SE and BSE detectors
followed in subsequent years and the design of an improved A typical protocol includes fixing the biological sample with
YAG/BSE detector (Autrata, 1990) allowed signal collection from appropriate aldehydes such as 2% paraformaldehyde/2% glu-
H. Schatten / Micron 42 (2011) 175–185 177

Fig. 1. Actin filament displaying helical twist 5 nm subunits of in vitro preparations


after dilute glutaraldehyde and cryofixation. Bar = 60 nm.
Reprinted from Erlandsen (2008).

taraldehyde in 0.1 M phosphate buffer (PBS) followed by PBS rinses,


postfixation in 1% osmium tetroxide in 0.1% PBS, rinses in PBS,
dehydration in increasing series of ethanol or acetone, critical point
drying, mounting on aluminum stubs, and coating. Fig. 3. Stereo-pair of 3T3 cell after fresh-freezing in propane, freeze-fracture and
Chemical fixation is adequate for many samples but, depending thawing into fixative followed by critical point drying and coating with ion-
on the questions addressed, it is important to remember that chem- beam sputtered Pt. Nucleus (top), nuclear membrane and peri-nuclear space are
clearly discernable. Image provided by Haggis (Agriculture Canada, Ottawa). Field
ical fixation may destroy structural integrity of some structures of
width = 2.8 ␮m.
interest. For example, glutaraldehyde cross-links the free amino
Reprinted from Pawley (2008).
groups of polypeptides and amino acids and inactivates enzymes. It
is therefore inadequate for enzyme localization and related studies.
Dehydration is being used routinely to prepare samples for Maconnachie, 1979, 1981). However, large macromolecules and
viewing in a vacuum environment. However, dehydration may structural components associated with them or covalently attached
be damaging to biological structures because the shape of a are preserved (Ris, 1985, 1988, 1990, 1991).
macromolecule or membrane is produced and maintained by its Critical point drying (CPD) is one of the preparation procedures
interactions with water which may be destroyed by removing requiring several precautions to avoid artifacts (Ris, 1985). Traces
water during the dehydration process. As much as 30–60% shrink- of water contaminating the intermediate liquid used for drying
age has been reported after preparing soft tissue for SEM (Boyde and (ethanol or acetone) or in the liquid CO2 transition fluid may distort
ultrastructure and induce artifacts which has clearly been shown
by Ris (1985) in a thorough series of elaborate control experiments.
In this case, studies clearly showed that the previously termed
microtrabecular structure was the result of residual water arti-
facts and not a newly discovered structure (Ris, 1985). This study
has been critically important to raise awareness of facts and arti-
facts and again, it needs to be stressed that the most impressive
instrumentation cannot compensate for inferior or improper sam-
ple preparation. Thorough CPD is one of the critical aspects for
producing artifact-free samples.
Coating is important for biological structure, as most biologi-
cal samples have insulating characteristics and need to be made
more conductive. Typically, a thin metal coating is applied to reduce
the effects of charging, e.g. glaring on the sample. The coating also
helps that secondary electrons from the very surface of the sample

Fig. 2. Spread human platelets after light fixation in glutaraldehyde followed by


immunolabeling for cell-adhesion molecules and cryofixation. (A) Shows Low mag- Fig. 4. The importance of adequate specimen preparation is explained in this TEM
nification of the platelet while (B) shows high-magnification of dimeric P-selectin. image of bacteriophage T4 polyheads. While fixation by conventional methods and
(C) Displays gold-labeled GPIb in the GPI-IX complex; (D) reveals gold-labeling critical point drying (A) does not preserve substructure, cryo-immobilized and par-
of individual integrin molecules for GPIIbIIIa. Bars A = 1 ␮m; B = 130 nm; C and tially freeze-dried samples show easily discernible subunits (B). Bar = 100 nm.
D = 100 nm. Reprinted from Erlandsen (2008); originally published in Hermann and Müller
Reprinted from Erlandsen (2008); originally published in Erlandsen et al. (2001). (1991).
178 H. Schatten / Micron 42 (2011) 175–185

(from the coating layer) only are released, whereas on uncoated electrical conductivity (Walther, 2008). These samples are best
biological samples secondary electrons can also escape from a cer- imaged using the backscattered electron signal, since then con-
tain depth, reducing surface resolution (discussed in Hermann and trast formation occurs at the thin heavy metal layer, which is in
Müller, 1991). Heavy metals or heavy metal compounds are used close contact with the biological structure of interest, whereas
for coating; however, it is oftentimes difficult to determine the cor- the secondary electron signal would image the overlaying carbon
rect amount of coating which eliminates charging, but also allows layer.
reliable viewing of biological structures without obscuring the Charging (e.g. glaring on the sample) may occur more often
areas of interest. Coating may affect biologically relevant resolution on flat specimens, as more electrons accumulate at the sur-
(reviewed in Pawley, 2008). For LVFESEM, 1–2 nm gold–palladium face while fewer leave the surface which may especially occur
applied by sputter coating has been optimal for a large variety when viewing samples at 5–20 kV (reviewed by Pawley, 2008).
of samples (reviewed in Schatten, 2008). Optimizing the coating Although it is oftentimes advertised that no coating is neces-
thickness may be required for delicate biological structure. sary under specific instrument conditions such as low voltage
A variety of coating procedures have been described and dis- (LV) or environmental (ESEM), charging oftentimes does occur
cussed in the literature (Peters, 1980; Peters, 1982, 1985, 1986a,b, and a thin layer of coating may still be required to avoid charg-
1988; Peters and Fox, 1990) offering several excellent alternative ing.
methods which include coating with a uniform film of Ta or Cr Good sample preparation and importantly, the investigator’s
deposited as a 1–2 nm layer by Penning sputtering before view- expertise are among the most important criteria for reliable results
ing with 30–40 kV. Fine details below 5 nm have been observed using EM.
with this method. Others have used Pt, W, and Ta by DC-ion
sputtering to view cells growing on EM grids (Bell et al., 1989; 3.2. Freezing methods
Lindroth et al., 1988; Lindroth and Sundgren, 1989). For biological
samples with charging problems, Walther and Hentschel (1989) Ultrarapid freezing is oftentimes used to provide superior
developed a double-coating technique by which the sample is ultrastructure and to preserve molecules in a more native state
first coated with a thin layer of heavy metal (platinum–carbon compared to chemical fixation. Ultrarapid freezing typically is
or tungsten with an average thickness of 2–3 nm) followed by a achieved at a rate of 104 –105 ◦ C/s to avoid formation of damaging
5–10 nm carbon layer to improve mechanical stability and enhance ice crystals in cells or tissue. At this temperature rate, cellular water

Fig. 5. SEM image using backscattered electrons (BSE) shows high-pressure frozen and cryofractured pancreas tissue displaying porous structures in the ER membranes
(arrows). (a) A perpendicular fracture to the ER membrane (center) and fracture parallel to the membranes (right). (b) and (c) Higher magnifications displaying ribosomes
and filamentous structures. The smooth extraplasmic fracture face (EF) faces the ER lumen; the rough protoplasmic fracture face (PF) faces the cytosol.
Reprinted from Walther (2008).
H. Schatten / Micron 42 (2011) 175–185 179

Fig. 6. High-resolution cryo-SEM image using secondary electrons (SE) shows ade-
novirus at high magnification.
Reprinted from Walther (2008).

becomes vitrified rather than forming ice crystals. Commonly used


freezing methods include the following.
Plunge freezing requires plunging a specimen into a liquid cryo-
gen such as supercooled liquid nitrogen or supercooled ethane
or propane which allows an average depth of vitrification of ca.
1–2 ␮m with minimal ice crystal artifacts.
Slam freezing (cold metal block freezing) is achieved by slam- Fig. 8. Cryo-FESEM images of the outer nuclear membrane (o) of mock-infected (A)
ming a specimen onto a copper or silver block that has been chilled and HSV-1-infected (B) HeLa cells after 10 h of incubation. (A) Nuclear pores appear
as round, small depressions with a distinct NPC structure, and a few appear as small
to −196 to −269 ◦ C with liquid nitrogen or liquid helium, allowing buttons (arrows). (B) Pores appear either as small but deep depressions without
an average depth of vitrification of ca. 10–15 ␮m with minimal ice obvious NPC structure or as buttons. Some of them have larger diameters and are
crystal artifacts. over top the nuclear membrane (arrows). In addition, there are holes with distinct
Propane jet freezing is achieved by sandwiching a 200–500 ␮m borders (arrowheads) and numerous budding capsids. The outer nuclear membrane
is locally broken away, giving view to the inner nuclear membrane (i). Bars = 200 nm.
thick specimen between two metal plates that are clamped into a
Reprinted from Wild et al. (2009).
device that directs jets of liquid propane cooled with liquid nitrogen
against both sides of the specimen plates; this allows an average
depth of vitrification of ca. 40 ␮m with minimal ice crystal artifacts. mens (Pawley et al., 1991) has provided somewhat better results
High-pressure freezing follows propane jet freezing or liquid regarding resolution (better than 3 nm) compared to freezing fol-
nitrogen and is achieved by pressurizing the specimen to 2100 lowed by freeze-drying. Freeze-fracture (Haggis and Pawley, 1988)
atmospheres to suppress or reduce growth of ice crystals at the is an excellent method to view intracellular structures which has
moderate freezing rates that can be achieved in the depth of the been reviewed by Pawley (2008). Dry fracture of tissue culture cells
sample. High pressure lowers the freezing point of water as well has been achieved by touching intact cells to the surface of adhesive
as the rate of ice crystal formation, allowing an average depth of tape (Ris, 1988, 1989; Lim et al., 1987; Ris and Pawley, 1989) which
vitrification of ca. 500 ␮m with minimal ice crystal artifacts. allows excellent insights into intracellular structure. This method
Several variations of these basic freezing methods have been may also be useful to view incorporation of labels that decorate
applied to various specimens, some of which will be described internal cell structure. The cell surface can be viewed first before
below. Freezing followed by freeze-drying has been used suc- removing it for viewing the internal cell structure.
cessfully (Pawley and Ris, 1987) and freezing followed by Cryo-SEM of chemically fixed cells has been described by
freeze-substitution is one of the methods that has gained increasing Erlandsen (2008). As mentioned above, several cryo-preparation
popularity (Erlandsen, 2008). Direct observation of frozen speci- methods are available to arrest cells in a “life-like” state (Mueller,

Fig. 7. High-resolution cryo-SEM image using secondary electrons (SE) shows freeze-dried and tungsten-coated actin filaments in which actin subunits are directly visible.
Reprinted from Walther (2008).
180 H. Schatten / Micron 42 (2011) 175–185

Fig. 9. Scanning electron micrograph of plant chromosomes in metaphase (top image; barley), early prometaphase (lower left image; barley) and prometaphase (lower right
image; rye) isolated with the drop/cryo technique. In metaphase the chromatids exhibit numerous chromomeres (top image; circle) and begin to separate. The centromere
(top image, arrow) is characterized by exposed parallel fibers. Parallel fibers (lower left and right image; arrows) and chromomeres of different sizes (lower left and right
image; circles) are seen in all stages of condensation.
Lower right image from Zoller et al. (2004); reproduced by permission of S. Karger AG, Basel. Entire figure reprinted from Wanner and Schroeder-Reiter (2008).

1990). Cryo-immobilization occurs in milliseconds compared to tion methods have been used for the analysis of macromolecular
chemical fixation which may take seconds or even minutes. complexes and resulted in preserving remarkable detail of actin
Freeze-substitution in acetone, methanol or other solvents can be filaments (Erlandsen, 2008) after coating with chromium. The heli-
used for subsequent processing and permanent fixation. In some cal twists of two polypeptide chains in the filament and 5 nm
cases, chemical fixation with low concentrations of glutaralde- subunits are clearly seen in Fig. 1. Other studies of macromolecu-
hyde (0.1–1.0%) is used for 10–15 min to stabilize macromolecules lar complexes revealed biological resolution of 2–3 nm. Erlandsen
prior to cryo-immobilization (Chen et al., 1995). Such prepara- et al. (2001) used cryo-methods to study details of the glyco-

Fig. 10. Stereomicrographs of barley metaphase chromosomes at different stages of artificial decondensation (fixation with glutaraldehyde and treatment with proteinase
K). During stretching, chromomeres loosen and can be recognized as bundles of solenoid fibers. The architecture of the centromere also loosens, showing three-dimensional
arrangement of parallel matrix fibers (A). After ultimate stretching, the number of matrix fibers is decreased and the chromomeres are stretched and decondensed to the
level of the individual solenoid fibers (B, arrow).
Wanner and Formanek, 2000; reproduced by permission of Elsevier. Partial figure reprinted from Wanner and Schroeder-Reiter (2008).
H. Schatten / Micron 42 (2011) 175–185 181

Fig. 11. Isolated mitotic apparatus of dividing sea urchin cells after critical point drying displaying microtubules and associated substructures observed at 1.5 kV without
charging artefacts after fracturing and light coating.
Image from collaborations (Thompson-Coffe et al., 1996) and reprinted from Pawley (2008).

calyx on the extracellular surface of human platelets that were is currently perhaps the best method to preserve cells and tissue
labeled with colloidal-gold markers and then visualized with high- in a relative native state. As has been proposed by Boyde (2008),
resolution in-lens cryo-SEM. All three cell-adhesion molecules in live-time stereo-imaging may be one step further to obtain a better
the glycocalyx that were selected for detection could be detected understanding of native cell features.
by using this method as seen in Fig. 2. In this case, after stabiliza- Other applications to view internal cell structure using cryo-
tion by fixation followed by rinses in distilled water, the samples SEM include (i) cryo-microtomy of cryo-immobilized plant and
were plunge-frozen into propane chilled with liquid nitrogen. After animal cells (Nusse and Van Aelsi, 1999; Walther and Müller, 1999)
partial freeze-drying at −85 ◦ C, the double-coating method devel- that includes examining the surface of the tissue block rather than
oped by Walther et al. (1995) was used and involved cryo-coating sections by cryo-SEM; and (ii) use of a cryo-dual beam instrument
samples by evaporation of 2 nm TaW at 45◦ by electron-beam depo- that incorporates both focusing electrons (SEM) and focusing ion
sition and 7–10 nm carbon at 90 ◦ C. This example demonstrates beam (FIB) columns which has been applied by Mulders (2003)
that complex combination methods may be used to obtain opti- to biological samples including yeast, bacteria, and gut epithelial
mal results for specific projects. The choice of sample preparation cells.
requires significant knowledge of the science project and technical As emphasized before, adequate sample preparation is critical
expertise, as so well demonstrated by several outstanding cell biol- and several methods may be employed for comparison studies,
ogists who have used microscopy as their primary research tool and
advanced science as well as techniques. The double-coating tech-
nique has been used successfully in numerous other applications
including double-layer coating of yeast cells after high-pressure
freezing and freeze-fracture (reviewed in Erlandsen, 2008).
Cryotechniques have been used with great success to stabi-
lize biological tissue before viewing them with LVFESEM. Frozen
specimen can be viewed directly in the SEM by using a cold stage
which had become available when side-entry eucentric goniometer
stages were employed that can accept high-stability, cryotransfer
stage rods as typically used for TEM electron crystallography. Sev-
eral excellent results have been reported (Müller, 1992; Pawley
et al., 1991; Herter et al., 1991; Boyde, 2008; Walther, 2008).
Uncoated and slightly coated specimens have been viewed in this
mode. Cryo-coating has been applied successfully and revealed
comparable images as seen after freeze-fracture TEM analysis and
included stereo-imaging of frozen-hydrated mitochondria. Modifi- Fig. 12. Stereoimage of the nuclear envelope (NE) isolated from an oocyte of the
cations of this technique have revealed stunning new insights into newt Notophthalamus viridescens. The nucleus was dissected into low-salt buffer
(LSB), transferred onto a cover-glass chip, extracted with 0.1% Triton X-100 in LSB,
intracellular structure as shown in images obtained by applying a
opened with fine glass needles, fixed in 2% glutaraldehyde and 0.2% tannic acid
freeze-fracture, thaw-fix technique developed by Haggis (Haggis, in LSB, postfixed in 0.1 aqueous osmium tetroxide, dehydrated with ethanol, and
1987; Haggis and Pawley, 1988). In this case, vitreous water that critical point dried. The cytoplasmic surface of the NE is seen on the lower right,
is sensitive to electron radiation is no longer present when speci- closely packed with nuclear pore complexes (NPCs). In the upper half, we look down
on a structure consisting of an annulus with eight cylindrical particles. A central
mens are imaged. The technique involves fresh-freezing in propane,
particle is visible in most NPCs. In the lower half, we look down on the intranuclear
freeze-fracture and thawing into fixative, critical point drying and surface of the NE. The fishtraps, representing the intranuclear segments of the NPCs,
ion-beam sputter coating with Pt before imaging as shown in Fig. 3. are clearly visible. Between the fishtraps, are scattered remnants of the lamin fibers,
Stereo-imaging of such samples allows a better understanding of disupted during swelling of the NE in LSB. Bar = 200 nm.
complex structural interactions of cell components. Cryofixation Reprinted with permission from Ris and Malecki (1993).
182 H. Schatten / Micron 42 (2011) 175–185

Fig. 13. (a) Stereoimage of a tangential section (250 nm thick) through an isolated NE of a Xenopus laevis oocyte, after Epon extraction. This image shows the cytoplasmic
surface with many NPCs. The NPCs have the same appearance as in whole mounts of the NE (Fig. 1). Bar = 200 nm. (b) Stereo image of a tangential section through an isolated
NE from a X. laevis oocyte. The intranuclear surface is exposed here. Several polar views of the fishtraps are visible (arrows), resembling the fishtraps seen in whole mounts.
Of special interest are the side views of fishtraps (black arrowheads), which can be obtained only with the method described in this article. Bar = 200 nm.
Reprinted with permission from Ris and Malecki (1993).

taking into consideration that some biological specimen are more using high-resolution cryoscanning electron microscopy of biolog-
fragile than others. The very different results obtained through cor- ical samples are described in Walther (2008).
rect versus inadequate specimen preparation are demonstrated in Recently, cryo-LVFESEM has been used to study nuclear
the example shown in Fig. 4 displaying cryo-TEM of polyheads of envelope alterations in Herpes Simplex Virus 1-infected cells
bacteriophage T4. Detail of polyhead polygonal subunit structure (Wild et al., 2009) which allowed remarkable three-dimensional
is only seen after cryo-immobilization and partial freeze-drying insights into virus-induced nuclear pore alterations. The inner
while no detail is seen after conventional fixation and critical point and outer parts of the nuclear membrane could be analyzed
drying. Controlled freeze-drying coupled with high-resolution clearly in infected and mock-infected (control) cells as seen in
cryo-coating of heavy metals and cryo-FESEM has demonstrated Fig. 8.
that a resolution in the range of 2–3 nm is possible in these cases
(Chen et al., 1995; Centonze and Chen, 1995; Chen et al., 1997; 4. Specific sample preparations for specific structural and
Hermann and Müller, 1991; Wepf, 1994). macromolecular analysis
LVFESEM has been used for a variety of investigations resulting
in new insights that had not been possible previously. Fig. 5 is an This section highlights specific applications that have been
example taken from Walther (2008) of high-pressure frozen and worked out by individual investigators to address specific research
cryofractured pancreas tissue images with SEM in the BSE mode. questions yielding results that could only have been obtained
Seen here are different fracture faces of the ER membranes show- with LVFESEM and no other microscopy technologies. It further
ing the smooth extraplasmic fracture face (EF) facing the lumen highlights the importance of understanding the specimen itself,
and rough protoplasmic fracture face (PF) facing the cytosol. Aden- the science, and the capabilities/limitations of instrument and
ovirus particles imaged with high-resolution cryo-SEM of samples methods. The projects presented here have been possible as a
imaged with secondary electrons are shown in Fig. 6. Fig. 7 dis- result of the investigators’ unique expertise. Three main examples
plays freeze-dried and tungsten-coated actin filaments with actin are highlighted which are (1) chromosome structure, (2) isolated
subunits clearly visible. Further examples and detailed methods mitotic spindles, centrosoms, and nuclear envelopes, and (3) de-
H. Schatten / Micron 42 (2011) 175–185 183

Fig. 14. (a) Stereo image of a cross-section (250 nm thick) through an isolated NE from an oocyte of Xenopus laevis. A fishtrap in profile is visible at the arrow. Bar = 200 nm.
(b) Stereo image of a cross-section (250 nm thick) through an isolated NE from a X. laevis oocyte. A side view of a fishtrap is seen at the arrows. The tips of the arrows point to
the small ring at the top of the fishtrap. A cable-like structure is attached to this ring. The nature of this ring is better seen in whole mounts. Bar = 200 nm. (c) Whole mount
of an isolated oocyte nucleus from X. laevis with the intranuclear surface exposed. The nucleus was isolated in LSB, attached to a glass chip, extracted with 0.1% Triton X-100
and 1% glutaraldehyde in LSB, postfixed with 1% aqueous osmium tetroxide, and critical point dried. The tops of adjacent fishtraps are connected to each other through a
complex cable system. Bar = 200 nm.
Reprinted with permission from Ris and Malecki (1993).

embedding of Epon-embedded thick sections featuring the nuclear studies with the 10 nm elementary fiber (Thoma et al., 1979), the
pore complex (NPC). 30 nm solenoid (Finch and Klug, 1976) and the nucleosome (Arents
et al., 1991). However, three-dimensional information was diffi-
4.1. Chromosome structure cult to obtain with TEM although it has been successfully applied
by Ris and Witt (1981) using stereo-imaging with high-voltage
The structural details of chromosomes has been of interest for TEM. High-resolution SEM provided a breakthrough in analyzing
over 120 years (Strasburger, 1888; Boveri, 1902; Sutton, 1902) three-dimensional chromatin organization (reviewed in Wanner
and excellent results have been reported for chromatin structural and Schroeder-Reiter, 2008) and has also made it possible to dis-
184 H. Schatten / Micron 42 (2011) 175–185

tinguish DNA and proteins which had not been possible with TEM progress as well as other instrument developments related to SEM
because of the TEM staining requirements including osmium and to allow imaging of cellular structures that are closer to live-cell
uranyl acetate. By using experimental manipulations and various conditions than currently possible and require modified versions of
labeling techniques details in chromosome substructure could be SEM. So far, excellent combination methods have been employed
visualized and analyzed with FESEM (reviewed in Wanner and to assess live conditions observed with light microscopy followed
Schroeder-Reiter, 2008). High-resolution SEM provided for the first by analysis with electron microscopy (Knott et al., 2009). Stud-
time details that had not been visualized previously, including an ies by these authors and others identified structures of interest
unusual nuclear organizing structure (NOR). Two examples of this with live confocal microscopy followed by analysis with focused
work are shown in Figs. 9 and 10 taken from Wanner and Schroeder- ion beam/scanning electron microscopy (FIB/SEM) and serial block
Reiter (2008). face/scanning electron microscopy (SBF/SEM). Serial section SEM
of adult brain tissue using focused ion beam milling allowed visu-
4.2. Isolated mitotic spindles, centrosomes, and nuclear envelopes alization of the ultrastructure of a time-limited biological event
within the context of a whole organism (Knott et al., 2008). These
As for chromosomes described above, isolated cell components studies and others offer new directions into an area of future appli-
with a complex surface topography must be coated with metals cations aimed at studying live-like events at high resolution. It can
without obscuring structural details, presenting a challenge for be expected that new developments in SEM will play a major role
many samples including the isolated mitotic apparatus (MA) and in such studies.
centrosomes. As for chromosomes, the isolated MA had previously
been visualized with high-voltage TEM or negative staining but References
details in surface structure could not be seen with these techniques.
By using LVFESEM, it was possible to visualize microtubules ema- Albrecht, R., Meyer, D., 2008. Molecular labeling for correlative microscopy: LM,
nating from a central core, decorated with components that are LVSEM, TEM, EF-TEM and HVEM. In: Schatten, H., Pawley, J. (Eds.), Biological
Low-Voltage Scanning Electron Microscopy. Springer, pp. 171–196.
tightly attached to microtubules. Vesicles were co-isolated along Arents, G., Burlingame, R.W., Wang, B.C., Love, W.E., Moudrianakis, E.N., 1991. The
microtubules (Fig. 11). An isolated nuclear envelope is shown in nucleosomal core histone octamer at 3.1 Å resolution: a tripartite protein assem-
Fig. 12. bly and a left-handed superhelix. Biochemistry 88, 10148–10152.
Autrata, R., 1990. New configurations of single-crystal scintillator detectors in SEM.
In: Proc XIIth ICEM Mtg, Seattle, pp. 376–377.
4.3. De-embedding of Epon-embedded thick sections Bell, P.B., Lindroth, M., Fredriksson, B.A., 1989. Problems associated with the prepara-
tion of cytoskeletons for high resolution electron microscopy. Scanning Microsc.
De-embedding of thick-sectioned cells and tissue offers unique (Suppl. 3), 117–135.
Boveri, T., 1902. Űber mehrpolige Mitosen als Mittel zur Analyse des Zellkerns.
insights into internal cell structures. The methods for this very dif- Verhandlungen der physikalisch-medizinischen Gesellschaft zu Würzburg 35,
ferent specimen preparation approach are described in more detail 67–90.
by Schatten (2008) reviewing various applications including the Boyde, A., Maconnachie, E., 1979. Volume changes during preparation of mouse
embryonic tissue for scanning electron microscopy. Scanning 2, 149–163.
nuclear pore complex (NPC), insect flight muscle (Ris and Malecki, Boyde, A., Maconnachie, E., 1981. Morphological correlations with dimensional
1993), Ascaris (Sepsenwol, 2006) and internal structures of the change during SEM specimen preparation. Scan. Electron Microsc. IV, 27–34.
apicomplexan parasite Toxoplasma gondii (Schatten and Ris, 2002, Boyde, A., 2008. Low kV and video-rate, beam-tilt stereo for viewing live-time exper-
iments in the SEM. In: Schatten, H., Pawley, J. (Eds.), Biological Low-Voltage
2004; Schatten et al., 2003). In the present review, one representa- Scanning Electron Microscopy. Springer, pp. 197–214.
tive application is chosen featuring the NPC that had originally been Broers, A.N., 1974. Recent advances in SEM with lanthanum hexaboride cathodes.
elaborated with great care by the late Hans Ris (Ris and Malecki, In: Johari, 0, Corvin, I. (Eds.), Scanning Electron Microscopy ll 974, vol. I. IITRI,
Chicago, pp. 10–18.
1993). Centonze, V.E., Chen, Y., 1995. Visualization of individual reovirus particles by low-
The study of the nuclear pore complex (NPC) had been facilitated temperature high-resolution scanning electron microscopy. J. Struct. Biol. 115,
greatly by using HRLVFESEM of thick-sectioned material coupled 215–225.
Chen, Y., et al., 1995. Imaging of cytoskeletal elements by low-temperature high-
with stereo-imaging. Previous studies had used negative staining
resolution scanning electron microscopy. J. Microsc. 179, 67–76.
and thin-section TEM which only revealed part of this complex Chen, Y., et al., 1997. High-resolution cryo-scanning electron microscopy study of
structural component of the nuclear envelope. LVFESEM analysis macromolecular structure of fibronectin fibrils. Scanning 19, 349–355.
showed for the first time intranuclear components that had not Crewe, A.V., Eggenberger, D.N., Wall, J., Welter, L.M., 1968. Electron gun using field
emission sources. Rev. Sci. Inst. 39, 576–583.
been possible with any other method applied so far. The images Crewe, A.V., 1973. Production of electron probes using a field emission source. In:
showed details of the cytoplasmic part and the intranuclear part in Wolf, E. (Ed.), Progress in Optics XI. North Holland, pp. 225–246.
three dimensions revealing that the cytoplasmic ring consisted of Erlandsen, S.L., et al., 2001. High resolution cryo-FESEM and detection of individual
cell adhesion molecules by stereo-imaging in the glycocalyx of human platelets:
short rods twice as long in height than in width. The nuclear pore immunogold localization of P-selectin (CD62P), integrin GpIIb/IIIa (CD41/CD61),
complex structure was termed fishtrap by Hans Ris who showed by and GpI-IX (CD42a,b). J. Histochem. Cytochem. 49, 809–819.
using LVFESEM that the inside of the NPC consists of a ring 120 nm Erlandsen, S.L., 2008. Cryo-SEM of chemically fixed animal cells. In: Schatten, H.,
Pawley, J. (Eds.), Biological Low-Voltage Scanning Electron Microscopy. Springer,
in diameter with eight thin filaments projecting from it into the pp. 215–228.
nuclear space. The ends of these filaments are attached to a smaller Finch, J.T., Klug, A., 1976. Solenoidal model for superstructure in chromatin. Proc.
ring of about 60 nm in width. The comparison with TEM showed Natl. Acad. Sci. U.S.A. 73, 1897–1901.
Knott, G.W., Marchman, H., Wall, D., Lich, B., 2008. Serial section scanning electron
that TEM was only partially able to capture these delicate struc- microscopy of adult brain tissue using focused ion beam milling. J. Neurosci. 28
tural features. Figs. 13 and 14 are representative images of nuclear (12), 2959–2964.
pore details as viewed with HRFESEM after de-embedding of thick Knott, G.W., Holtmaat, A., Trachtenberg, J.T., Svoboda, K., Welker, E., 2009. A proto-
col for preparing GFP-labeled neurons previously imaged in vivo and in slice
sections.
preparations for light and electron microscopic analysis. Nat. Protoc. 4 (8.),
1145–1156.
5. Future perspectives Haggis, G.H., 1987. Freeze-fracture of cell nuclei for high-resolution SEM and deep-
etch TEM. Proc. Electron Microsc. Soc. Am. 45, 560–564.
Haggis, G.H., Pawley, J.B., 1988. Freeze-fraacture of 3T3 cells for high-resolution
As emphasized in this review, the preservation of native cellular scanning electron microscopy. J. Microsc. 150, 211–218.
events afforded by controlled chemical fixation and/or cryo- Hainfeld, J., 1977. Understanding and using field emission sources. In: Johari, 0,
preservation of delicate cellular structures is among the aspects Corvin, I (Eds.), Scanning Electron Microscopy lZ977, vol. I. IITRI, Chicago, pp.
591–604.
important for future developments in HRLVFESEM. The devel- Hermann, R., Müller, M., 1991. Prerequisites of high resolution scanning electron
opment of high-resolution variable pressure instruments are in microscopy. Scan. Electron Microsc. 5, 653–664.
H. Schatten / Micron 42 (2011) 175–185 185

Hermann, R., Walther, P., Müller, M., 1996. Immunogold labeling in scanning elec- Ris, H., Witt, P.L., 1981. Structure of the mammalian kinetochore. Chromosoma 82,
tron microscopy. Histochem. Cell Biol. 106, 31–39. 153–170.
Herter, P., et al., 1991. High-resolution scanning electron microscopy of inner sur- Ris, H., 1985. The cytoplasmic filament system in critical point dried whole mounts
faces and fracture faces of kidney tissue using cryo-preparation methods. J. and plastic-embedded sections. J. Cell Biol. 100, 1474–1487.
Microsc. 161, 375–385. Ris, H., 1988. Application of LVSEM in the analysis of complex intracellular struc-
Joy, D.C., Pawley, J.B., 1992. High resolution scanning electron microscopy. Ultrami- tures. In: ProcEMSA46, pp. 212–213.
cros 47, 80–100. Ris, H., 1989. Three-dimensional imaging of cell ultrastructure with high resolution
Joy, D., 2008a. The aberration-corrected SEM. In: Schatten, H., Pawley, J. (Eds.), Bio- low voltage SEM. Inst. Phys. Conf. Ser. 98, 657462.
logical Low-Voltage Scanning Electron Microscopy. Springer, pp. 107–128. Ris, H., Pawley, J.B., 1989. Analysis of complex three-dimensional structures involved
Joy, D., 2008b. Noise and its effects on the low-voltage SEM. In: Schatten, H., Paw- in dynamic processes by high voltage electron microscopy and low voltage high
ley, J. (Eds.), Biological Low-Voltage Scanning Electron Microscopy. Springer, pp. resolution scanning electron microscopy. In: Pattner, H (Ed.), Microscopy of
129–144. Subcellular Dynamics. CRC Press, Boca Raton, FL, pp. 309–323.
Lindroth, M., Bell, P.B., Fredriksson, B.A., 1988. Comparison of the effects of critical Ris, H., 1990. Application of low voltage high resolution SEM in the study of complex
point drying and freeze-drying on cytoskeletons, and microtubules. J. Microsc. intracellular structures. In: Proc Xllth ICEM Mtg, Seattle, pp. 18–19.
151 (2), 103–114. Ris, H., 1991. The three-dimensional structure of the nuclear pore complex as seen
Lindroth, M., Sundgren, J.E., 1989. Ion beam-sputtered and magnetron-sputtered by high voltage electron microscopy and high resolution low voltage scanning
thin films on cytoskeletons: a high resolution TEM study. Scanning I 1, 243–253. electron microscopy. EMSA Bull Spring/9 1 2 I, 54–56.
Lim, S.S., Ris, H., Schnasse, B., 1987. Pigment granules in goldfish xanthophores are Ris, H., Malecki, M., 1993. High resolution field emission scanning electron micro-
attached to intermediate filaments. J. Cell Biol. 105 (37a). scope imaging of internal cell structures after Epon extraction from sections: a
McMullan, D., 2008. The early development of the scanning electron microscope. new approach to correlative ultrastructural and immunocytochemical studies.
In: Schatten, H., Pawley, J. (Eds.), Biological Low-Voltage Scanning Electron J. Struct. Biol. 11, 148–157.
Microscopy. Springer, pp. 1–26. Schatten, H., Ris, H., 2002. Unconventional specimen preparation techniques using
Mulders, H., 2003. The use of a SEM/FIB dual beam applied to biological samples. high resolution low voltage field emission scanning electron microscopy to
GIT Imag. Microsc. 2, 8–10. study cell motility, host cell invasion, and internal structures in Toxoplasma
Müller, M., 1992. The integrating power of cryo-fixation based electron microscopy gondii. Microsc. Microanal. 8, 94–103.
in biology. Acta Microscopica 1, 37–44. Schatten, H., Sibley, D., Ris, H., 2003. Structural evidence for actin filaments in Toxo-
Nagatani, T., et al., 1987. Development of an ultra high-resolution scanning electron plasma gondii using high resolution low voltage field emission scanning electron
microscope by means of field-emission source and in-lens system. Scan. Microsc. microscopy. Microsc. Microanal. 9, 330–335.
1, 901–909. Schatten, H., Ris, H., 2004. Three-dimensional imaging of Toxoplasma gondii–host
Newbury, D., 2008. Developments in instrumentation for microanalysis in low- cell membrane interactions. Microsc. Microanal. 10, 580–585.
voltage scanning electron microscopy. In: Schatten, H., Pawley, J. (Eds.), Schatten, H., 2008. High-resolution, low voltage, field-emission scanning electron
Biological Low-Voltage Scanning Electron Microscopy. Springer, pp. 263–304. microscopy (HRLVFESEM) applications for cell biology and specimen prepara-
Nusse, J., Van Aelsi, A.C., 1999. Cryo-planning for cryo-scanning electron microscopy. tion protocols. In: Schatten, H., Pawley, J. (Eds.), Biological Low-Voltage Scanning
Scanning 21, 372–378. Electron Microscopy. Springer, pp. 145–169.
Oatley, C.W., 1982. The early history of the scanning electron microscope. J. Appl. Sepsenwol, S., 2006. Revealing internal cell structure for high-resolution scanning
Phys. 7532, Rl–R13. electron microscopy. Proc. Microsc. Soc. Am..
Pawley, J.B., Ris, H., 1987. Structure of the cytoplasmic filament system in freeze- Strasburger, E., 1888. Űber Kern- und Zelltheilung im Pflanzenbereich, nebst einem
dried whole mounts viewed by HVEM. J. Microsc. 13, 319–332. Anhang über Befruchtung. Gustav Fischer, Jena.
Pawley, J.B., Albrecht, R.M., 1988. Imaging colloidal gold labels in LVSEM. Scanning Sutton, W.S., 1902. The chromosomes in heredity. Biol. Bull. 4, 231–251.
10, 184–189. Thoma, F., Koller, T., Klug, A., 1979. Involvement of histone H1 in the organization
Pawley, J.B., Walther, P., Shih, S.J., Malecki, M., 1991. Early results using high resolu- of the nucleosome and the salt-dependent superstructures of chromatin. J. Cell
tion low voltage low temperature SEM. J. Microsc. 162, 327–335. Biol. 83, 402–427.
Pawley, J.B., 1992. LVSEM for high resolution topographic and density contrast imag- Thornley, R.F.M., 1960. Recent developments in scanning electron microscopy. In:
ing. In: Hawkes, P.W., Kazan, B. (Eds.), Advances in Electronic Electron Physics, Proc. EUREM, pp. 173–176.
vol. 83. Academic Press, New York, pp. 203–274. Thornley, R.F.M., Cartz, L., 1962. Direct examination of ceramic surfaces with the
Pawley, J.B., 2008. LVSEM for biology. In: Schatten, H., Pawley, J. (Eds.), Biological scanning electron microscope. J. Am. Ceram. Soc. 45, 425–428.
Low-Voltage Scanning Electron Microscopy. Springer, pp. 27–106. Walther, P., Hentschel, J., 1989. Improved representation of cell surface structures by
Peters, K.R., 1980. Penning sputtering of ultra thin metal films for high resolution freeze substitution and backscattered electron imaging. Scan. Microsc. 3 (Suppl.
electron microscopy. In: Johari, 0, Corvin, I (Eds.), Scanning Electron Microscopy 3), 201–211.
lY80, vol. 1. SEM Inc., Chicago, pp. 143–154. Walther, P., et al., 1995. Double layer coating for high-resolution low temperature
Peters, K.R., 1982. Conditions required for high quality high magnification images in SEM. J. Microsc. 179, 229–237.
secondary electron scanning electron microscopy. In: Johari, 0, Corvin, 1 (Eds.), Walther, P., Müller, M., 1999. Biological ultrastructure as revealed by high-resolution
Scanning Electron Microscopy/lY82, vol. IV. SEM Inc., Chicago, pp. 1359–1372. cryo-SEM of blockfaces after cryo-sectioning. J. Microsc. 196 (3), 279–287.
Peters, K.R., 1985. Working at higher magnifications in scanning electron microscopy Walther, P., 2008. High-resolution cryoscanning electron microscopy of biologi-
with secondary and backscattered electrons on metal coated biological spec- cal samples. In: Schatten, H., Pawley, J. (Eds.), Biological Low-Voltage Scanning
imens and imaging macromolecular cell membrane structures. In: Johari, 0, Electron Microscopy. Springer, pp. 245–262.
Corvin, I (Eds.), Scanning Electron Microscopy/l985, vol. IV. SEM Inc., Chicago, Wanner, G., Schroeder-Reiter, E., 2008. Scanning electron microscopy of chromo-
pp. 1519–1544. somes. Methods Cell Biol. 88, 451–474.
Peters, K.R., 1986a. Rationale for the application of thin, continuous metal films in Wepf, R., 1994. High-resolution SEM of biological macromolecular complexes. In:
high magnification electron microscopy. J. Microsc. 142, 25–34. Bailey, G.W., Garratt-Reed, A.J. (Eds.), Proc. 82 Ann Meet Microsc. Soc. Am. San
Peters, K.R., 1986b. Metal coating thickness and image quality in scanning electron Francisco Press, San Francisco, pp. 1026–1027.
microscopy. Proc. EMSA 44, 664–667. Wild, P., Senn, C., Manera, C.L., Sutter, E., Schraner, E.M., Tobler, K., Ackermann, M.,
Peters, K.R., 1988. Current state of biological high resolution scanning electron Ziegler, U., Lucas, M.S., Kaech, A., 2009. Exploring the nuclear envelope of herpes
microscopy. Proc. EMSA 46, 180–181. simplex virus 1-infected cells by high-resolution microscopy. J. Virol., 408–419.
Ultra-high resolution cinematic digital 3D imaging of the cell surface by field emis- Zoller, J.F., Herrmann, R.G., Wanner, G., 2004. Chromosome condensation in mitosis
sion scanning electron microscopy Peters, K.R., Fox, M.D. (Eds.), 1990. Proc. XIIth and meiosis of rye (Secale cereale L.). Cytogen. Gen. Res. 105, 134–144.
ICEM Mtg., vol. 1, pp. 12–13.

You might also like