INSTRUCTION MANUAL
MULTI POWER TRANSDUCER MODEL LSMT4
n ENVIRONMENT
BEFORE USE .... • Indoor use
• Do not install the unit where it is directly exposed to rain,
Thank you for choosing M-System. Before use, please check
water droplets or sunlight.
contents of the package you received as outlined below.
• When heavy dust or metal particles are present in the
If you have any problems or questions with the product,
air, install the unit inside proper housing and ventilate it.
please contact M-System’s Sales Office or representatives.
• Do not install the unit where it is subjected to continuous
vibration. Do not apply physical impact to the unit.
■ PACKAGE INCLUDES:
• Environmental temperature must be within -10 to +55°C
Multi Power Transducer.............................................. (1)
(14 to 131°F) with relative humidity within 30 to 90% RH
Output label for users ................................................. (2)
in order to ensure adequate life span and operation.
Mounting slider ........................................................... (2)

n WIRING
■ MODEL NO.
• Do not install cables (power supply, input and output)
Confirm that the model number described on the product is
close to noise sources (relay drive cable, high frequency
exactly what you ordered.
line, etc.).
• Do not bind these cables together with those in which
■ INSTRUCTION MANUAL
noises are present. Do not install them in the same duct.
This manual describes necessary points of caution when
you use this product, including installation, connection and
n AND ....
basic maintenance procedures.
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
POINTS OF CAUTION sheet.

■ GENERAL PRECAUTION
• Before you remove or mount the unit, turn off the power
supply and the input signal for safety.
LIGHTNING SURGE PROTECTION
In order to protect the unit from lightning surges entering
n AUXILIARY POWER INPUT RATINGS through signal and power supply cables, use of appropriate
• Power input ratings are specified by the model number lightning surge protectors are recommended. Please con-
suffix code. tact M-System.
100 – 240V AC rating: 85 – 264V AC, 50/60 Hz, <20VA
110 – 240V DC rating: 99 – 264V DC, <9W
24V DC rating: 24V±10%, <9W
48V DC rating: 48V±10%, <9W

COMPONENT IDENTIFICATION
n FRONT PANEL CONFIGURATION

Output Terminals

DATA Display

Configuration Jack

Power Input Terminals Control Buttons

Input Terminals

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 1 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4

INSTALLATION
n WALL MOUNTING n DIN RAIL MOUNTING
Insert the provided slider into upper side of the unit, and Set the unit so that its DIN rail adaptor is at the bottom.
pull out the slider on the lower side. Fasten the M4 screw to Position the upper hook at the rear side of the unit on the
the hole (4.5 mm dia.) of the slider. (torque 1.4 N­m) DIN rail and push in the lower. When removing the unit,
To remove the slider, pull it out while pushing the latch to push down the DIN rail adaptor utilizing a minus screw-
inside as shown in the figure below. driver and pull.

DIN Rail
35mm wide

push push
Slider

■ EXTERNAL DIMENSIONS unit: mm (inch)

7.2 (.28) 24–M3.5 TERMINAL SCREW

4–4.5 (.17) DIA. MTG HOLE


TERMINAL COVER

DIN RAIL
35mm wide
119 (4.69) (DIN rail mounting)
140 (5.52)
126 (4.96)
115 (4.53)

75 (2.95) 6.5 (.26)


125 (4.92) 116 (4.57) [3.3 (.13)]

13–M4 TERMINAL SCREW


8.6 (.34)

■ MOUNTING REQUIREMENTS unit: mm (inch)

75 (2.95) 4-M4
126 (4.96)

130 (5.12) min.

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 2 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4

TERMINAL CONNECTIONS
■ TERMINAL ASSIGNMENT ■ ELECTRICAL CONNECTION

Output Terminals CH1+


ANALOG OUTPUT 1
CH1– .
.
.
.
CH10+
ANALOG OUTPUT 10
CH10–

CH11+
Power Input Terminals CONTACT OUTPUT 1
CH11–
Input Terminals
CH12+
CONTACT OUTPUT 2
CH12–

U(+)
POWER
V(–)

FE

JACK CONFIGURATOR

System /
Application
Terminal

P1 P2 P3 1S 1L 3S 3L P1 P2 P3 1S 1L 3S 3L

source load source k l load


L1 L1
K L k l
L2 L2
Three phase / L3 L3
3-wire, K L

unbalanced load
P1 P2 P3 1S 1L 3S 3L

u v v u

source k l load
U V V U
L1
K L k l
L2
System / L3
Application
Terminal K L

P1 N 1S 1L P1 N 1S 1L P1 N 1S 1L P1 N 1S 1L

source load source k l load source load source k l load


L1 L1 L1 L1
K L K L
N N L2 L2
Single phase / Three phase / L3 L3
2-wire 4-wire, N N

P1 N 1S 1L balanced load P1 N 1S 1L

u v u v

source U V k l load k l
source U V load
L1 L1
K L K L
N L2
L3
N

P1 P2 P3 1S 1L P1 P2 P3 1S 1L

P1 P2 P3 N 1S 1L 2S 2L 3S 3L
source load source k l load
L1 L1
K L
L2 L2
Three phase / L3 L3 source load
3-wire, L1
L2
balanced load P1 P2 P3 1S 1L L3
N
u v v u

source k l load
U V V U
L1 P1 P2 P3 N 1S 1L 2S 2L 3S 3L
K L
L2
L3
Three phase /
4-wire, source k l load
k l
P1 N P3 1S 1L 3S 3L P1 N P3 1S 1L 3S 3L unbalanced load L1
K L k l
L2
K L
L3
K L
N
source load source k l load
L1 L1
K L k l
N N
P1 P2 P3 N 1S 1L 2S 2L 3S 3L
Single phase / L2 L2
K L

3-wire u v u v u v
P1 N P3 1S 1L 3S 3L
source k l load
U V U V U V k l
L1
u v v u K L k l
L2
K L
source k l load L3
U V V U K L
L1 N
K L k l
N
L2
K L

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 3 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4

OPERATING MODE
The measurement display items
can be chosen with 4 buttons.

Phase voltage, Phase 1 Active power Power factor 1THD

Phase voltage, Phase 2 Reactive power Frequency 2THD

Phase voltage, Phase 3 Apparent power Power factor, Phase 1 3THD

Phase to phase, 1 – 2 Active power, Phase 1 Power factor, Phase 2 Phase to phase THD, 1 – 2

Phase to phase, 2 – 3 Active power, Phase 2 Power factor, Phase 3 Phase to phase THD, 2 – 3

Phase to phase, 3 – 1 Active power, Phase 3 Phase to phase THD, 3 – 1

Current, Line 1 Reactive power, Phase 1 Current THD, Line 1

Current, Line 2 Reactive power, Phase 2 Current THD, Line 2

Current, Line 3 Reactive power, Phase 3 Current THD, Line 3

Neutral current Apparent power, Phase 1 Neutral current THD

Apparent power, Phase 2

Apparent power, Phase 3

CHARACTER SET
0 1 2 3 4 5 6 7 8 9 – A B C D E F G H I J K L MN O P Q R S T U VWX Y Z

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 4 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4

SETTING MODE
■ BUTTON OPERATION
Button Menu/Item Numeric Input

IU ↑ / PQS ↓ Move between Menu/Item Increment/decrement at selected digit

Select Menu/Item Enter the value

Escape from Menu/Item Return without entering the value

■ SETTING MENU/SETTING GROUP MENU


• Flow chart

Operating Mode

Press and hold Incorrect Passcode

Enter Passcode
Enter the passcode. Passcode range: 0000 to 9999.
When entered passcode is correct, the unit enters to setting menu.
When entered passcode is incorrect, the unit returns to operating mode with error message.

Incorrect Passcode

Setting Menu
Version
DEV.I
Each time is pressed, the firmware version and the
model number are alternately displayed.

Reboot
REBT
Item Operation
NO Return to Setting Menu
YES Reboot the unit

Change passcode
PASS
The passcode can be changed within 0000 to 9999.
Factory default is 0000.

Loop test
LOOP
By selecting output item of loop test; output selection can be changed.
Item Output Loop test operation
AO1 Analog output 1 → Select the digit in the engineering unit value
↑ Increase the value at selected digit
• •
• •

AO10 Analog output 10 ↓ Decrease the value at selected digit


PO1 Pulse output 1 Select ON/OFF with ↑/↓
PO2 Pulse output 2

Setting Group Menu

SETT INP To Input Setting Menu

ENER To Pulse Output Setting Menu

AOUT To Analog Output Setting Menu

OPT To Option Setting Menu

MNTE To Maintenance Menu

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 5 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
■ INP INPUT SETTING MENU
• Flow chart

Input Setting Menu


To Setting System Configuration
SYS
Group Menu
Item Operation
1P2W : Single phase/2-wire
1P3W : Single phase/3-wire
3P3B : 3-phase/3-wire balanced
3P3U : 3-phase/3-wire unbalanced
3P4B : 3-phase/4-wire balanced *1
3P4U : 3-phase/4-wire unbalanced *1
*1. Available only with LSMT4-2xxx-x

CT rating, primary
CT1
1 to 20 000 A [5A]

CT rating, secondary
CT2
1 to 5 A [5A]

VT rating, primary
VT1
0.05 to 400.00 kV [0.11 kV] *2

VT rating, secondary
VT2
50 to 480 V [110 V] *2
When this setting is changed, voltage range is automatically
set as shown in table below.

Setting Automatic voltage range


110 150
220 300
Others Same as the setting value

Voltage range
VRNG
50 to 480 (V) [150 V]

AC frequency
FREQ
Item Operation
U1N : Voltage 1 (*)
I1 : Current 1

Low-end cutout, current


CO−I
0.0 to 99.9 (%) [1.0 %]
The current value below ( rated current × setting value )
is regarded as 0 A.

Low-end cutout, voltage


CO−V
0.0 to 99.9 (%) [1.0 %]
The voltage value below ( rated voltage × setting value ) is
regarded as 0 V.
(*): Factory default setting
*2. Set rated voltage for VT according to the table shown below.
System Configuration Voltage
Single phase/2-wire Voltage between V1 – N 50 – 277 V
3-phase/3-wire Line-to-line voltage 50 – 480 V
Single phase/3-wire Voltage between V1 – N 50 – 277 V
3-phase/4-wire Phase voltage 50 – 277 V

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 6 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
■ ENTER PULSE OUTPUT SETTING MENU
• Flow chart

Select Output Pulse Output


Setting Menu
To Setting PO1 Energy count type
PO2 CNT
Group Menu
ID PARAMETER
EP Active energy, incoming
EQ Reactive energy, LAG
ES Apparent energy
EP– Active energy, outgoing
EQ– Reactive energy, LEAD
EQ G Reactive energy, incoming, LAG
EQ D Reactive energy, incoming, LEAD
EQ–G Reactive energy, outgoing, LAG
EQ–D Reactive energy, outgoing, LEAD
EQ P Reactive energy, incoming
EQ–P Reactive energy, outgoing

Pulse weight
WEI
ID PARAMETER
0.001(*) p × 0.001 kWh/pulse
0.01 p × 0.01 kWh/pulse
0.1 p × 0.1 kWh/pulse
1.0 p × 1.0 kWh/pulse

p is determined by n calculated from the formula shown


below and the table shown below.
n= a × primary rated voltage × primary rated current ÷ 1000
Where a is 1: Single phase/2-wire, 2: Single phase/3-wire,
√3: 3-phase, 3: 3-phase/4-wire.
n p
< 10 1
10 ≤ n < 100 10
100 ≤ n < 1000 100
1000 ≤ n < 10 000 1000
10 000 ≤ n < 100 000 10 000
≥ 100 000 100 000

Pulse width
DUR
Selectable within 100 to 2000 msec. by 100 msec. steps.
[100 msec.]

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 7 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
■ AOUT ANALOG OUTPUT SETTING MENU
• Flow chart

Select Output Analog Output


Setting Menu
To Setting AO1 Assigned measurand
AO2 INP
Group Menu
Refer to the table, Parameters to be assigned to analog outputs.
AO3
AO4
AO5 Proportional output
SCAL
AO6 Set the parameters for four items shown below.
AO7
Item Description Setting range
AO8 ISPN Input range Refer to Analog output parameters table
AO9 IZER Input 0 % Refer to Analog output parameters table
AO10 OSPN Output 100 % Refer to Analog output range table
OZER Output 0 % Refer to Analog output range table

With Proportional output, the input range (Input 0 % to Input range)


is converted to output range (Output 0 % to Output 100 %.)
After Input range is specified, Input 0 % is automatically assigned.
Refer to the table, Parameters to be assigned to analog outputs for
the value automatically assigned.

Linearization
LINA
Item Description
SCAL (*) Set 0 % and 100 % for input and output to convert
with Proportional output. Refer to the Proportional
output for settings.
TBL Set the linearization table to convert with the
linearization table. Refer to the linearization
table for settings.

Linearization table
TBL Set input/output of 10 points in the linearization table as shown
below.
Item Description Setting range
IN0 Input 1st point Refer to the Analog
: : output range
IN9 Input 10th point
OUT0 Output 1st point Refer to the Analog
: : output range
OUT9 Output 10th point

Adjustment
ADJ
By selecting item in the table below, output adjustment is
performed in the adjustable range.

Item Adjustable range


ZERO -5.00 to +5.00 (%) [0.00 %]
SPAN 95.00 to 105.00 (%) [100.00 %]

•Analog output range


Model number Output rating Available setting range
LSMT4-xxAx-x 4.00 to 20.00 (mA) 3.20 to 20.80 (mA)
LSMT4-xx4x-x 0.000 to 10.000 (V) -0.500 to +10.500 (V)
LSMT4-xx5x-x 0.000 to 5.000 (V) -0.250 to +5.250 (V)
LSMT4-xx6x-x 1.000 to 5.000 (V) 0.800 to 5.200 (V)

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 8 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
• Analog output parameters
Input range Available setting Auto setting
Item Description
-100% -75% -50% -25% 0% 25% 50% 75% 100% range for input range value for input 0 %
NONE Not assigned – –
I1 Current, Line 1 0.00 to 120.00% 0.00
I2 Current, Line 2 0 CT secondary rating [0.00 to 100.00%]
I3 Current, Line 3
IN Neutral current
U12 Phase to phase, Line 1 – 2 0.00 to 120.00% 0.00
U23 Phase to phase, Line 2 – 3 [0.00 to 100.00%]
U31 Phase to phase, Line 3 – 1 0 Voltage range
U1N Phase voltage, Phase 1
U2N Phase voltage, Phase 2
U3N Phase voltage, Phase 3
P Active power -120.00 to 120.00% 0.00
P1 Active power, Phase 1 –P (outgoing) 0 (incoming) P [0.00 to 100.00%]
P2 Active power, Phase 2
P3 Active power, Phase 3
Q Reactive power -120.00 to 120.00% Sign inversion
Q1 Reactive power, Phase 1 –P (LEAD) 0 (LAG) P [-100.00 to 100.00%] value of the input
Q2 Reactive power, Phase 2 range
Q3 Reactive power, Phase 3
S Apparent power 0.00 to 120.00% 0.00
S1 Apparent power, Phase 1 0 P [0.00 to 100.00%]
S2 Apparent power, Phase 2
S3 Apparent power, Phase 3
PF Power factor -100.00 to 100.00% Sign inversion
PF1 Power factor, Phase 1 0 (LEAD) 1 (LAG) 0 [-50.00 to 50.00%] value of the input
PF2 Power factor, Phase 2 range
PF3 Power factor, Phase 3
F Frequency 0.00 to 100.00% –
45 55 65
[0.00 to 100.00%]

T–Q Reactive power for -100.00 to 100.00% Sign inversion


bidirectional current (outgoing) (incoming) [-100.00 to 100.00%] value of the input
–P (LEAD) 0 (LAG) P –P (LEAD) 0 (LAG) P
range

When LEAD/LAG range of power rating is set, the unit outputs 0 to 50 %


for outgoing and outputs; 50 to 100 % for incoming.

T–PF Power factor for -100.00 to 100.00% Sign inversion


(outgoing) (incoming)
bidirectional current 0 (LEAD) 1 (LAG) 0 (LEAD) 1 (LAG) P [-50.00 to 50.00%] value of the input
range

When LEAD/LAG range of power factor is set, the unit outputs 0 to 50 %


for outgoing; 50 to 100 % for incoming.

• Calculation of P (power rating) • Inversion of output 0 % to 100 %


P (power rating) is automatically determined by rounding to the unit When negative value is set to input range, output is inverted.
(CT secondary rating × 100 (W)) after calculated with following formula by For example, -50.00 % is set to input range of reactive power,
system configuration and secondary rating of CT/VT. 50.00 % is automatically set to input 0 % and then output
It is available not to round with option setting. characteristic is the figure shown below.
Power rating = (CT secondary rating) × (VT secondary rating) × a
Output (mA)
Calculation example
20
System configuration a Rating Power rating
Single phase/2-wire 1 110V/5A 500W
12
220V/5A 1000W
Single phase/3-wire 2 110V/5A 1000W 4
3-phase/3-wire 2 110V/5A 1000W Input (var)
1000 0 -1000
220V/5A 2000W (LAG) (LEAD)
3-phase/4-wire 3 220V/5A 3500W

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 9 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
■ OUTPUT EXAMPLES
• Reactive power for bidirectional current
Output (mA)
20

12

4
Input (var)
1000 0 1000 0 1000
(LEAD) (LAG) (LEAD) (LAG)
Outgoing Incoming

• Power factor for bidirectional current


Range-100 to 100% Range-50 to 50%
Output (mA) Output (mA)
20 20

12 12

4 4
input input
0 1 0 1 0 0 0.5 1 0.5 0 0.5 1 0.5 0
(LEAD) (LAG) (LEAD) (LAG) (LEAD) (LAG) (LEAD) (LAG)
Outgoing Incoming Outgoing Incoming

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 10 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
■ OPT OPTION SETTING MENU
• Flow chart

Option Setting
Menu
To Setting Analog limit
Group Menu ALIM
Select ON/OFF of Analog limit.
Item Description
ON Limit at -1 % and +101 %
OFF (*) No limit (output -5 to +105 %)

Power rating round


RPR
Item Description
ON (*) After power rating is calculated by the formula below, it is
rounded to nearest the integral multiple of (secondary rating × 100 (W)).
OFF Power rating is calculated by the formula below (not rounded)

Power rating = VT secondary rating × CT secondary rating × a

System a
configuration ON OFF
Single phase/2-wire 1 1
Single phase/3-wire 2 2
3-phase/3-wire 2 √3
3-phase/4-wire 3 3

Power factor sign


PF–S
Item
IIEC (*) Active power is incoming: positive, outgoing: negative

IEEE Phase direction is LAG: positive, LEAD: negative

Reactive power sign


Q–S
Item Description
IEC (*) The range from power factor 1.0 to -180° is positive,
other is negative.
SPC When incoming same as IEC. When outgoing,
sign reversed.

Reactive power calculation


QN–C
Item Description
VEC (*) Vector calculation method: Qn= Sn2–Pn2
SIGM Reactive power meter method:
1 Nsmp
Qn= ∑ (Uni–Nni) Ii + (Nsmp/4)
Nsmp i=1

Apparent power calculation


S–C
Item Description
(*): Factory default setting
VEC (*) Vector calculation method: S = P2 + Q2
SUM Sum calculation: S = S1 + S2 + S3

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 11 / 12


Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
LSMT4
■ MNTE MAINTENANCE MENU
• Flow chart

Maintenance
Menu
To Setting Backup setting
BKUP
Group Menu
Item Description
NO Return to Maintenance Menu.
YES Backup present settings into internal memory.

Restore setting
REST
Item Description
NO Return to Maintenance Menu.
YES Restore settings in the internal memory.

Initialization
CLR.A
Item Description
NO Return to Maintenance Menu.
YES Initialize settings with default.

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-1957 Rev.2 P. 12 / 12