You are on page 1of 8

Electrophoresis 2003, 24, 1253–1260 1253

Lung-Ming Fu Low-voltage driven control in electrophoresis


Ruey-Jen Yang
microchips by traveling electric field
Department of
Engineering Science, This paper presents the use of a physical model and numerical simulation in the inves-
National Cheng Kung University, tigation of traveling electric fields on capillary electrophoresis (CE) chips. The principal
Tainan, Taiwan material transport mechanisms of electrokinetic migration, ionic concentration, fluid
flow, and diffusion are all taken into consideration. Traditionally, the high electric field
strength required for the separation of biological samples by microfluidic devices has
involved the application of high external voltages. In contrast, this study presents a
proposal for samples separation by means of a moving electric field within a low volt-
age-driven CE chip. Under this proposal, the separation channel is partitioned into a
series of smaller separation zones by means of electrode pairs. This paper considers
two different electrode configurations, namely arranged along a single side of the
separation channel, and arranged on two sides of the separation channel. The quality
of the separation achieved with these two configurations is then compared with the
traditional straight separation channel approach. The results confirm that the pro-
posed method is successful in maintaining an adequate field strength for separation
purposes in a low-voltage driven CE chip. Furthermore, it is determined that the best
separation results are obtained using electrodes arranged along both sides of the
separation channel.

Keywords: Double-banked electrode channel / Capillary electrophoresis chips / Miniaturization /


Traveling electric field EL 5343

1 Introduction ments. Generally speaking, DNA separation requires an


electric field strength in the range of 300–800 V/cm, and
an applied voltage in the order of 1–3 kV.
Recent years have seen an increasing interest in the
development of microfluidic devices for chemical or bio- When AC-type traveling electric fields (dielectrophoresis)
logical applications, including the separation of DNA for are used in the separation of cells and dielectric particles,
genetic engineering purposes. Separation techniques

Miniaturization
they have the ability to drive the cells at a rate of around
for the microdevices used in such applications typically 25–30 mm/s [15–17]. The use of DC-type moving electric
include capillary electrophoresis (CE) [1–5], synchronized fields in the CE process was also considered [18]. As
cyclic electrophoresis (SCE) [6], free-flow electrophoresis stated above, the conventional separation process using
[7, 8], gel electrophoresis [9, 10], and electroosmotic flow CE microchips involves the application of a voltage in the
[11–14]. These techniques exploit the fact that dielectric range of 1–3 kV to the separation channel. Recently,
particles move in the opposite direction to an applied external voltage flow control has been extended to micro-
electrical potential under the influence of the resulting fluidic devices [19–23]. They show measurable flow con-
electric field. Furthermore, different dielectric particles trol with the application of the voltage about 100–150 V.
exhibit different mobility characteristics. Regarding DNA Though the driving potential is lower, the necessary
applications, the traditional CE chip-based approach field strength for separation purposes is maintained. This
has involved establishing a potential difference across is achieved by partitioning the separation length with
the two ends of the separation channel. This voltage arrayed electrodes into a series of smaller separation
must be high enough to create an electric field of suffi- zones. The principle of the separation approach is to use
cient intensity to bring about separation of the DNA frag- a DC voltage to energize particular sets of these elec-
trodes in a controlled sequence such that the sample is
driven along the channel and becomes separated.
Correspondence: Prof. Ruey-Jen Yang, Department of Engi-
neering Science, National Cheng Kung University, Tainan, 70101
We propose to partition the separation channel into many
Taiwan
E-mail: rjyang@mail.ncku.edu.tw smaller separation zones using arrayed electrodes. Fig-
Fax: +886-6-276-6549 ure 1 provides a schematic view of the separation channel

 2003 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim 0173-0835/03/07-0804–1253 $17.501.50/0
1254 L.-M. Fu and R.-J. Yang Electrophoresis 2003, 24, 1253–1260

Figure 2. Schematic diagram of an arrayed-electrode


design and the mechanism for driving and separating
DNA fragments using a moving electric field with (a) sin-
gle-sided electrodes; (b) double-banked electrodes.
Figure 1. Schematic layout of the subdivision of the
separation channel showing the principal design param-
eters. In the present investigation, two electrode configurations
are considered, namely electrodes arrayed along one side
of the separation channel, and electrodes arrayed along
subdivision. The principal design parameters are clearly opposite sides of the channel. These two configurations
shown, namely the interval between two successive elec- are presented in Figs. 2a and b, respectively. The current
trodes, D, and the span of the separation zone, which lies study provides a detailed description of the design of low-
between any two active electrodes, S. Figure 2 provides voltage driven CE microchips using the traveling electric
two schematic views of the arrayed electrode design, and field method, and compares the results with those ob-
indicates how sample driving and DNA fragment separa- tained from the traditional straight separation channel
tion are achieved by moving the electric field sequentially approach. With the moving electric field technique, the
along the channel. Switching the electrode pairs on and applied voltage for generating the necessary electric field
off successively causes the sample to be driven pro- to drive and separate samples can therefore be lowered.
gressively through the individual separation zones, which
are established within the currently active electrode pair.
When a particular electrode pair is active, an electric field 2 Formulation and numerical method
is established in the interval between the two electrodes.
The intensity of the field within this zone is sufficient to Pressure-driven flow through a separation channel re-
bring about DNA separation. Therefore, the proposed quires the use of a micropump. The production cost of
method both drives and separates the sample. such a separation system is expensive, and the operation
process is difficult to control. A more convenient alterative
It is known that the velocities of the DNA fragments are is to drive fluid motion by applying an electric field to the
determined by the electric field strength and by certain net fluid charge in order to produce a body force. This
buffer solution properties. By considering the charge and method is commonly referred to as electroosmotic flow.
weight of the DNA fragments, the electric field strength, Although most previous studies regarding eletroosmotic
and the required detected signal resolution of the sepa- flows [24–28] assumed the velocity profile to be fully
rated DNA fragments, it is possible to determine an ap- developed in the microchannels, and considered the
propriate design for the separation subdivisions, and the charge density to conform to the Boltzmann equilibrium
optimal arrayed-electrode configuration. distribution, these assumptions are not strictly valid since
Electrophoresis 2003, 24, 1253–1260 Traveling electric field within CE chips 1255

electroosmotic flow involves an entrance region within investigation, the value of Sc in the equations given
which neither condition is satisfied. To take account of above is of the order of 105, while Re is of the order of
this entrance effect, the current authors have previously one. Although this implies that the diffusion term is very
developed physical models based on (i) the Poisson small in comparison to the convection term, diffusion is
equation for the electric potential and zeta potential, one of the primary mechanisms for sample dispersion,
(ii) the Nernst-Planck equations for ionic concentration, and accordingly, the term is retained in the numerical
and (iii) the full Navier-Stokes equations modified to simulation. Since the sample concentration is carried
include the effects of the body force due to the electric passively by the flow field, Eq. (6) can be solved sepa-
and the charge density [11, 12, 29–31]. rately once the flow field has been determined.

In order to simulate the distribution of a sample plug, it In most electrophoresis experiments, the microchannel is
is necessary to determine an equation which expresses filled with an aqueous electrolyte solution. The sample is
its concentration. The computational domain considered injected into the channel, and an electric potential is then
in the current simulation extends from the entry region to activated in order to carry out sample separation. Conse-
the fully developed region. Since the flow rate is very quently, the assumed initial conditions are that the veloc-
small, the entrance length is very short, and has there- ity is static throughout the system, and that the zeta
fore only a marginal influence upon the separation re- potential of the wall is given by Cuwall = z. Furthermore,
sults. By introducing the following reference quantities, the initial ionic concentration and potential are assumed
Lref, Uref, pref, and rref and defining re ˆ re =n0 ze ˆ to conform to the Boltzmann equilibrium condition. The
…n‡ n †=n0 , c ˆ zec=k b T, t ˆ tUref =Lref , u ˆ u=Uref , detailed expressions of the initial conditions and the

v ˆ v=Uref , x ˆ x=Lref , y ˆ y=Lref , p ˆ …p pref †= rU2ref , boundary conditions are provided in [29].
r ˆ rf =rref , n‡ ˆ …n‡ n0 †=n0 , and n ˆ …n n0 †=n0 ,
the dimensionless forms of the governing equations after
dropping the head symbols can be written as: 3 Results and discussion
k2
r2 c ˆ r (2) In the present investigation, a CE chip was fabricated
2 e
qni * 1 1 onto a glass substrate of dimensions 7562562 mm,
‡ u rni ˆ r2 ni ‡ ‰r…ni rc†Š (3) and was composed of two layers of micromachined glass
qt ScRe ScRe
bonded together. The microchannels were fabricated
*
r u ˆ 0 (4) using standard microfabrication technology [31, 32]. The
* lengths of the injection channel and the separation chan-
qu * * 1 2*
‡ u ru ˆ rp ‡ r u Gx re rc (5) nels were 10 mm and 40 mm, respectively, and the cross-
qt Re
section of each channel was 80 mm wide by 40 mm deep.
qC * 1 The numerical study assumed that the microchannel was
‡ u rC ˆ r2 C (6)
qt ScRe made of silica glass and that the buffer fluid was sodium
borate. The physical and electrical properties of the liquid
In these expressions, k = WK, where K is the Debye-
were taken to be as follows: dielectric constant e = 80,
Huckel parameter and is given by K = (2n0z2e2/ee0kbT)1/2,
zeta potential of the channel wall z = 275 mV, electro-
1/K is the characteristic thickness of the charge density,
kinetic diameter k = 32, fluid viscosity m = 1023 Ns/m2,
re = (n12 n2)ze is the charge density, n1 and n2 are the
Schmidt number of the charge density Sc = 105, and diffu-
concentrations of the positive and negative ions, e is the
sion coefficient of the sample D = 6.9610211 m2/s.
dielectric constant of the medium, e0 is the permittivity
of a vacuum, n0 is the bulk concentration of the ions,
kb is the Boltzmann’s constant, T is the absolute temper- 3.1 Traditional straight separation channel
ature, Lref = W, where W is the channel height, Uref =
Cinletee0uzu/mL, Cinlet is the activated potential at the inlet, A traditional electrophoresis separation employs a straight
Re is the Reynolds    and is given by Re ˆ
numbers separation channel and a large driving voltage since this
rf Uref Lref =m ˆ rf cinletmLee0 jzj Lmref , Sc is the Schmidt number arrangement provides a better sample band shape in the
and is given by Sc = m/rfDi, m is the liquid viscosity, detection area for resolution purposes. In this type of
rf is the fluid density, Di is the diffusion coefficient of microchannel flow, an applied electric field acts on the
sample, z is the surface zeta potential, p is the pressure, net fluid charge near the wall to produce a body force
Gx is the ratio of the electrical double layer (EDL) energy which drives fluid motion. Figure 3 shows typical plots of
to the mechanical kinetic energy and is given by, Gx = the electric potential field and the velocity vector profile
2n0kbTrfW2/mRe2 and finally, C is the sample concentra- in the traditional straight separation channel. It can be
tion. For the flow conditions considered in the current seen that the electric field is uniformly distributed in the
1256 L.-M. Fu and R.-J. Yang Electrophoresis 2003, 24, 1253–1260

was 1.2% HPMC (hydroxypropylmethylcellulose) in TBE


(Tris-borate-EDTA) with 1% Cy5 fluorescence dye. The
injection voltage was 300 V for a duration of 30 s. The
separation voltage was 1.2 kV for a duration of 2 min.
The resulting migration velocity served later as a bench-
mark for the design of the traveling electric field chips.
The separated DNA fragments were detected using a
fluorescent microscope (BX60; Olympus, Japan). The
excitation light was generated using a mercury lamp,
which was filtered by a 530–550 bandpass filter, and
then focused on the detection spot of the CE chip through
the fluorescent microscope. The induced fluorescence
from the DNA sample was filtered using a 580 nm high-
pass filter, and then detected through a fluorescence
microscopy technique using a photomultiplier tube (R928;
Hamamatsu, Japan). Figure 5 presents electrophero-
grams generated from an image collected at the detec-
tion area of a straight separation channel. It can be seen
that the fX-174 DNA separation has 11 obvious peaks.
Figure 3. (a) Electric potential distribution; (b) velocity These experimental results provided in this figure serve
vector profile in the traditional straight separation chan- as a benchmark against which to compare the results of
nel. the two traveling electric field CE chips which are de-
scribed in the following subsections.

separation channel, and that the velocity vectors have a


flat profile. These characteristics provide excellent condi-
tions for sample flow and separation. Figure 4 shows the
typical distribution of a sample band as it flows along a
straight separation channel, i.e., from location (a) to the
detection area at location (e). Despite a slight dispersion
of the sample at the detection area, it can be seen that its
shape remains suitable for detection purposes.

Figure 5. Electropherograms of the separation result for


the fX-174 DNA (HaeIII digested) sample generated from
an image collected at the detection area of a traditional
straight separation channel with a voltage of 1.2 kV.

3.2 Single-sided electrode separation channel


Figure 4. Sample band concentration distribution along
a traditional straight separation channel with a voltage of A numerical simulation was conducted in order to analyze
1.2 kV. the electric potential distribution, the velocity vector pro-
file, and the travel characteristics of a single sample band
within a low-voltage driven CE chip having a single-sided
The DNA sample used in the current straight separation electrode separation channel. The simulation considered
channel investigation was a HaeIII-digested fX-174 DNA a separation channel of 40 mm length and 40 mm height.
consisting of eleven fragments, i.e., 72, 118, 194, 224, Twenty-one electrode pairs were arranged along the bot-
271, 281, 310, 603, 872, 1078, and 1353 base pairs (bp). tom of the separation channel with a pitch, D, of 2 mm.
A 0.5 mL DNA sample with a concentration of 1 mg/mL was The span between active electrodes, S, was taken to be
injected into a conventional cross-type CE chip. CE buffer 6 mm. Accordingly, the voltage required to drive and
Electrophoresis 2003, 24, 1253–1260 Traveling electric field within CE chips 1257

separate the DNA fragments was reduced to about 30% sample band concentration is already skewed. It can be
of the voltage required for the straight separation channel seen that the band becomes increasingly distorted as it
case. flows through the subsequent separation zones within
the partitioned separation channel, i.e., towards the
Figure 6 shows the resulting electric potential distribution
detection area at location (e).
and velocity vector profiles between two active electrodes
separated by a span of 6 mm with a voltage of 360 V. Figure 8 compares the typical sample band shape and
The electric potential distribution is nonuniform at the peak intensity at the detection area for the cases of a
two extremities of the span (Fig. 6a). The nature of the straight separation channel and a single-sided electrode
field distribution would be expected to result in an asym- separation channel. It is possible to define a shape error
metric velocity distribution since it is the effect of an value, Ê, of the sample in the single-sided electrode chan-
electric field acting on the net fluid charge near the wall nel by comparing it with the benchmark result obtained in
which produces the necessary body force to drive fluid the straight separation channel case, i.e., Ê = [(detection
motion. This is confirmed by the results presented in shape area – maximum overlap area)/the shape area
Fig. 6b, which indicates a skewed velocity distribution in of straight separation channel]100%. A small value of Ê
the region of the two active electrodes. Figure 7 shows implies that the shapes of the two samples are similar,
the computed sample band concentration distribution and that detection results will be satisfactory. However,
along a single-sided electrode separation channel. At the results in Fig. 8a suggest that the band shape ob-
location (a), which indicates the first separation zone, the tained with the single-sided electrode separation channel
is unsuitable for detection purposes, and indeed, analysis
shows the shape error value, Ê, to be approximately
80.3%. Figure 8b compares the peak intensity of the
sample band within a single-sided electrode separation
channel with the intensity in the straight separation chan-
nel. While the peak in concentration intensity achieved

Figure 6. (a) Electric potential distribution; (b) velocity


vector profile within an interval of one span S = 6 mm in a
single-sided electrode separation channel with a voltage
of 360 V.

Figure 8. Comparison of (a) band shape and (b) peak


intensity at the detection area between a single-sided
Figure 7. Sample band concentration distribution along electrode separation channel and a traditional straight
a single-sided electrode separation channel. separation channel.
1258 L.-M. Fu and R.-J. Yang Electrophoresis 2003, 24, 1253–1260

with the straight channel is obvious, it can be seen that 3.3 Double-banked electrodes separation
the intensity within the single-sided electrode channel is channel
extremely weak and widely dispersed.
The effect of employing double-banked electrodes along
Figure 9 presents electrophoerograms generated from the separation channel is to increase the electric potential
images collected at the detection area of the traditional strength in the upper regions of the channel. Figure 10
straight separation channel and numerical simulation re- shows the electric potential distribution and velocity
sults for a single-sided electrode separation channel. The vector profiles which result when using double-banked
sample consists of 1024 M rhodamine B and 6.561023 M electrodes separated by span S = 6 mm and a voltage of
Cy3 fluorescent dye. From experimental results, the sam- 180 V. The current simulation considered a separation
ple separation exhibits four obvious peaks, which corre- channel of 40 mm length, a span, S, of 6 mm, a pitch be-
spond to the three fragments of Cy3 in the hydrolysate tween successive electrodes, D, of 2 mm, and a total of
and the single fragment of rhodamine B within the sam- 21 electrode pairs located along the upper and lower
ple. However, from numerical simulations, for a single- regions of the separation channel. The voltage required
sided electrode separation channel, Fig. 9 (dashed line) to drive and separate the sample is reduced to approxi-
shows the peak concentration intensity at the detection mately 15% of the voltage required within the straight
area by simulation for sample fragments with concentra- separation channel. The simulated results in Fig. 10a
tion values of 2 (rhodamine B), 0.45 (Cy3(a)), 1.5 (Cy3(b)), show that the double-banked electrodes create a sym-
and 1.1 (Cy3(c)), respectively. The electroosmotic mobility metric electric potential distribution in the separation
ratio of four fragments are rhodamine B : Cy3(a) : Cy3(b) : channel between the two active electrode pairs. There-
Cy3(c) = 1 : 0.75 : 0.54 : 0.47. Injecting the fragments at fore, as would be expected, Fig. 10b indicates that the
the same location and with the different electroosmotic resulting velocity profiles are also flat and symmetric.
mobility simulates the convection of the fragments. The
peak intensity of the sample concentration in detection Figure 11 shows the typical concentration distribution of a
PN sample as it passes along a double-banked electrode
area is defined as C ˆ Ci =N, where C is the average
iˆ1
separation channel. It can be seen that the sample con-
concentration intensity in the detection area, Ci is the
local concentration intensity in the detection area, and N
is the total number of points in the detection area. Figure 9
(dashed line) indicates that the computed concentration
value of the rhodamine B peak is about 1.07, and the
Cy3(a) is unobvious. This corresponds to a dispersive
ratio of 46.5% (i.e.,(221.07)/26100%). It is noted that
this ratio is very high, and therefore the detection results
are unlikely to be satisfactory from the current single-
sided electrode separation channel.

Figure 10. (a) Electric potential distribution; (b) velocity


vector profile within an interval of one span S = 6 mm in a
double-banked electrode separation channel with a volt-
age of 180 V.

Figure 9. Comparison of electropherograms generated


from an image collected in a traditional straight separa-
tion channel by experimental (solid line) and intensity of
concentration fragments in a single-sided electrode
separation channel with numerical simulation (dashed Figure 11. Sample band concentration distribution along
line) at the detection area. a double-banked electrode separation channel.
Electrophoresis 2003, 24, 1253–1260 Traveling electric field within CE chips 1259

centration distribution is uniform and flat within each of Figure 13 presents electrophoerograms generated from
the intermediate separation zones, i.e., there is very little images collected at the detection area of the traditional
tilting or distortion of the sample at locations (a)–(d). When straight separation channel and numerical simulation
the sample arrives at the detection area, i.e., at location results for a double-banked electrode separation chan-
(e), its band shape is sufficiently good to ensure satis- nel. The peak intensities of the four samples are very
factory detection results. Figure 12 shows the band shape clearly distinguishable. The dispersive ratio is determined
and peak intensity of the sample at the detection area of a to be less than 5.2%. The results indicate that a good
double-banked electrode separation channel, and com- quality of separation detection will be provided by this
pares the results to those achieved with the traditional particular arrayed-electrode configuration. Figure 14
straight separation channel. Figure 12a compares the presents the relationship between the applied voltage
two band shapes at the detection area. It is observed required for driving and separation purposes and the
that there is good correlation between the two results, active electrode interval. The figure considers the case of
and in fact the shape error value of the sample in the a double-banked electrode separation channel of 40 mm
double-banked separation channel is very low, i.e., Ê =
6.5%. Figure 12b compares the peak intensity of a sam-
ple band in the double-banked separation channel with
the results observed for the straight separation channel.
Compared to the previous case of the single-sided
electrode channel, it is noted that there is now a distinct
peak intensity within the double-banked electrodes sepa-
ration channel, and that it corresponds very closely to the
results obtained in the benchmark straight channel case.
The results presented in Figs. 12a and b suggest that the
use of a double-banked electrode separation channel will
yield a sample of satisfactory shape and intensity for
detection purposes.

Figure 13. Comparison of electropherograms generated


from an image collected in a traditional straight separa-
tion channel by experimental (solid line) and intensity of
concentration fragments in a double-banked electrode
separation channel with numerical simulation (dashed
line) at the detection area.

Figure 12. Comparison of (a) band shape and (b) peak


intensity at the detection area between a double-banked Figure 14. Relationship between voltage ratio and active
electrode separation channel and a traditional straight electrode span for a double-banked electrode separation
separation channel. channel.
1260 L.-M. Fu and R.-J. Yang Electrophoresis 2003, 24, 1253–1260

length with 21 electrode pairs separated by a pitch of [4] Russel, W. B., Saville, D. A., Schowalter, W. R., Colloidal Dis-
2 mm. The results show a direct relationship between persion, Cambridge University Press, New York 1989.
[5] Culbertson, C., Ramsey, R., Ramsey, J. M., Anal. Chem.
the electrode span and the required driving voltage. The
2000, 72, 2285–2291.
voltage can be as low as 10% of the original value if the [6] Burggraf, N., Manz, A., Verpoorte, E., Effenhauser, C. S.,
interval of the span S is long enough for the sample plug Sens. Actuators B 1994, 80, 103–110.
to be separated. [7] Hymer, W., Kirshnan, K., Kraemer, W., Welsch, J., Lanham,
W., Electrophoresis 2000, 21, 311–317.
[8] Junkers, J., Kopplin, P. S., Munch, J., Kettrup, A., Electro-
phoresis 2002, 23, 2872–2879.
4 Concluding remarks [9] Rizzi, C., Rossini, K., Bruson, A., Sandri, M., Peruffo, A. D. B.,
Carraro, U., Electrophoresis 2002, 23, 3266–3269.
This paper has presented an investigation into the gen- [10] Compton, M. M., Lapp, S. A., Pedemonte, R., Electrophore-
eration of traveling electric fields within CE chips. This sis 2002, 23, 3262–3265.
study provides valuable information in designing elec- [11] Yang, R.-J., Fu, L.-M., Lin, Y.-C., J. Colloid Interface Sci.
2001, 239, 98–105.
trode layouts which maintain a sufficient electric field
[12] Yang, R.-J., Fu, L.-M., Hwang, C.-C., J. Colloid Interface Sci.
strength to carry out separation, while requiring driving 2001, 244, 173–179.
voltages which are orders of magnitude lower than those [13] Jacobson, S. C., Hergenroder, R., Moore, A. W., Ramsey,
used in conventional straight separation channel CE J. M., Anal. Chem. 1994, 66, 4127–4132.
chips. The simulated results have shown that the use of [14] Jacobson, S. C., Koutny, L. B., Hergenroder, R., Moore,
A. W., Ramsey, J. M., Anal. Chem. 1994, 66, 3472–3476.
a single-sided electrode separation channel generates
[15] Xu, C., Wang, Y., Cao, M., Lu, Z., Electrophoresis 1999, 20,
the nonuniform electric field distribution and asymmetric 1829–1831.
velocity profile within the channel. This results in tilting [16] Schnelle, T., Müller, T., Gradl, G., Shirley, S. G., Fuhr, G.,
and distortion of the sample band as it is driven along Electrophoresis 2000, 21, 66–73.
the channel, which results in poor detection of the DNA [17] Cui, L., Holmes, D., Morgan, H., Electrophoresis 2001, 22,
separation. The use of a double-banked electrode chan- 3893–3901.
nel greatly improves the uniformity of the electric field and [18] Ying, F., Mastrangelo, C. H., Burkes, D. T., Burns, M. A.,
Electrophoresis 1998, 19, 1388–1393.
the symmetry of the velocity profile. The resulting sample
[19] Schasfoort, R. B. M., Schlautmann, S., Hendrikse, J., van
band distribution is very similar to that achieved in the den Berg, A., Science 1999, 286, 942–945.
traditional straight separation channel, and is ideal for [20] Polson, N. A., Hayes, M. A., Anal. Chem. 2000, 72, 1088–
detection purposes. Furthermore, this electrode config- 1092.
uration allows the original driving voltage to be reduced [21] McKinght, T. E., Culbertson, C. T., Jacobson, S. C., Ramsey,
J. M., Anal. Chem. 2001, 73, 4045–4049.
by 85%.
[22] Hartley, N. K., Hayes, M. A., Anal. Chem. 2002, 74, 1249–
1255.
The authors thank Prof. Che-Hsin Lin of Southern Taiwan
[23] Heller, C., Electrophoresis 2001, 22, 629–643.
University of Technology for comments on the experi-
[24] Grabtree, H. J., Cheong, C. S., Tilroe, D. A., Backhouse,
ment. This project is partially supported by the Taiwan C. J., Anal. Chem. 2001, 73, 4079–4086.
National Research Council under NSC-90-2212-E006- [25] Patankar, N. A., Hu, H. H., Anal. Chem. 1998, 70, 1870–
109. 1881.
[26] Arulanandam, S., Li, D., Colloids Surf. A 2000, 161, 89–102.
Received September 11, 2002 [27] Griffiths, S., Nilson, R. H., Anal. Chem. 2001, 73, 272–278.
[28] Dutta, P., Beskok, A., Anal. Chem. 2001, 73, 1979–1986.
[29] Fu, L.-M., Yang, R.-J., Lee, G.-B., Electrophoresis 2002, 23,
602–612.
5 References [30] Yang, R.-J., Fu, L.-M., Lee, G.-B., J. Sep. Sci. 2002, 25, 996–
1010.
[1] Behnke, B., Metzger, J. W., Electrophoresis 1999, 20, 80–83. [31] Fu, L.-M., Yang, R.-J., Lee, G.-B., Liu, H.-H., Anal. Chem.
[2] Xin, B., Lee, M. L., Electrophoresis 1999, 20, 67–73. 2002, 74, 5084–5091.
[3] Mistry, K., Krull, I., Grinberg, N., J. Sep. Sci. 2002, 25, 935– [32] Lin, C. H., Lee, G. B., Lin, Y. H., Chang, G. L., J. Micromech.
958. Microeng. 2001, 11, 726–732.

You might also like