Professional Documents
Culture Documents
MR Procedures: Biologic Effects, Safety, and Patient Care: Reviews
MR Procedures: Biologic Effects, Safety, and Patient Care: Reviews
Radiology
635
mune responsiveness, and other biologic the signal, the polarity of the signal, the
ESSENTIALS processes (20 –38). In the majority of distribution of current in the body, the
these studies, the authors concluded that electric properties, and the sensitivity of
● Most reported cases of MR-related inju- exposures to static magnetic fields pro- the particular cell membrane (8,39 – 48).
ries and the few fatalities that have duce no substantial harmful biologic ef- Several investigations have been con-
occurred have apparently been the re- fects. Although there have been some re- ducted to characterize MR system–related
Radiology
sult of failure to follow safety guidelines ports of potentially injurious effects of gradient magnetic field–induced stimula-
or of use of inappropriate or outdated static magnetic fields on isolated cells or tion in human subjects (41– 48). At suffi-
information related to the safety as- organisms, none of these effects have cient exposure levels, peripheral nerve
been verified or firmly established as a stimulation is perceptible as a “tingling”
pects of biomedical implants and de-
scientific fact (1,9). The relatively few or “tapping” sensation. At gradient mag-
vices.
documented injuries that have occurred netic field exposure levels of 50%–100%
in association with MR system magnets above perception thresholds, patients may
● To prevent accidents and injuries in the were attributed to the inadvertent pres- become uncomfortable or experience pain
MR environment, it is necessary to re- ence or introduction of ferromagnetic ob- (8). At extremely high levels, cardiac stim-
vise information on biologic effects and jects (eg, oxygen tanks, aneurysm clips) ulation is of concern. However, the in-
safety according to changes that have into the MR environment (1,5–7,9). duction of cardiac stimulation requires
occurred in MR technology and with With regard to the effects of long-term exceedingly strong and/or rapid gradient
exposure to static magnetic fields, there magnetic fields—more than an order of
regard to use of current guidelines for
are interactions between tissues and static magnitude greater than those used in com-
biomedical implants and devices.
magnetic fields that could theoretically mercially available MR systems (8,39,40).
lead to pathologic changes in human sub- Fortunately, current safety standards for
● The preservation of a safe MR environ- jects (1,9,16). However, quantitative anal- gradient magnetic fields associated with
ment requires constant attention to the ysis of these mechanisms indicates that present-day MR systems appear to pro-
care of patients and individuals with they are below the threshold of impor- vide adequate protection from potential
metallic implants and devices, because tance with respect to long-term adverse hazards or injuries in patients (2,8,16,39).
the variety and complexity of these ob- biologic effects (1,9,16). Of interest, results of studies per-
At present, the pertinent literature does formed in human subjects indicate that
jects constantly changes.
not contain carefully controlled studies anatomic sites of peripheral nerve stimu-
that demonstrate the absolute safety of lation vary depending on the activation
chronic exposure to powerful magnetic of a specific gradient (ie, x, y, or z gradi-
(1,9,16). Most MR systems in use today fields. With the increased clinical use of ent) (8). Stimulation sites for x gradients
operate with magnetic fields ranging interventional MR procedures, there is a included the bridge of the nose, the left
from 0.2 to 3.0 T. In the research setting, critical need for such investigations. side of the thorax, the iliac crest, the left
an exceptionally powerful MR system op- However, it may be virtually impossible thigh, the buttocks, and the lower back.
erating at 8.0 T is located at Ohio State to demonstrate “absolute safety,” given Stimulation sites for y gradients included
University (Columbus). According to the the various difficulties in conducting the scapula, the upper arms, the shoul-
latest guidelines from the FDA, clinical such a study. In addition, although there der, the right side of the thorax, the iliac
MR systems that use a static magnetic is no evidence for a cumulative effect of crest, the hip, the hands, and the upper
field up to 8.0 T are considered a “non- magnetic field exposures on health, fur- back. Stimulation sites for z gradients in-
significant risk” for patients. The expo- ther studies of the exposed populations cluded the scapula, the thorax, the xy-
sure of research subjects to fields stronger (eg, MR health care workers, patients phoid, the abdomen, the iliac crest, and
than 8.0 T requires approval of the re- who undergo repeated studies) will be the upper and lower back (8). Typically,
search protocol by an institutional re- helpful in establishing rational guide- peripheral nerve stimulation sites were at
view board and the informed consent of lines for occupational and patient expo- bony prominences. According to Schaefer
the subjects. sures to static magnetic fields (1,9,16). et al (8), because bone is less conductive
Schenck (1,9) conducted comprehen- than the surrounding tissue it may in-
sive reviews of biologic effects associated crease current densities in narrow regions
with exposure to static magnetic fields. BIOLOGIC EFFECTS OF of tissue between bone and skin, result-
With regard to short-term exposures (eg, GRADIENT MAGNETIC FIELDS ing in lower nerve stimulation thresholds
limited exposures or those associated with than expected.
the clinical use of MR systems), the avail- During MR procedures, gradient mag-
able information for effects of static mag- netic fields may stimulate nerves or mus-
netic fields on biologic tissues is exten- cles by inducing electric fields in pa- ACOUSTIC NOISE
sive (1,9,20 –38). Investigations include tients. This topic has been thoroughly
studies on alterations in cell growth and reviewed by Schaefer et al (8), Nyenhuis Various forms of acoustic noise are pro-
morphology, cell reproduction and ter- et al (39), and Bourland et al (40). The duced in association with the operation
atogenicity, DNA structure and gene ex- potential for interactions between gradi- of an MR system (49). The primary source
pression, pre- and postnatal reproduction ent magnetic fields and biologic tissue is of acoustic noise, however, is the gradi-
and development, blood-brain barrier dependent on a variety of factors, includ- ent magnetic field activated during the
permeability, nerve activity, cognitive ing the fundamental field frequency, the MR procedure. This noise occurs during
function and behavior, cardiovascular maximum flux density, the average flux rapid alterations of current within the
dynamics, hematologic indexes, temper- density, the presence of harmonic fre- gradient coils that, in the presence of the
ature regulation, circadian rhythms, im- quencies, the waveform characteristics of powerful static magnetic field of the MR
BIOLOGIC EFFECTS OF RF
Problems associated with acoustic noise FIELDS
MR-related Acoustic Noise and
for patients and health care workers in-
Permissible Limits
clude simple annoyance, difficulties in The majority of the RF power transmitted
verbal communication, heightened anx- The FDA indicates that MR-related acous- for MR imaging or spectroscopy (eg, car-
iety, temporary hearing loss, and, poten- tic noise levels must be below the level of bon decoupling, fast spin-echo pulse se-
tially, permanent hearing impairment concern established by pertinent federal quences, magnetization transfer contrast
(49 – 61). Acoustic noise may pose a par- regulatory or other recognized standards- pulse sequences) is transformed into heat
ticular hazard to specific patient groups setting organizations (2). If the acoustic within the patient’s tissues as a result of
who are at increased risk. Patients with noise is not below this level, the sponsor resistive losses (11,67). Not surprisingly,
psychiatric disorders, the elderly, and pedi- (ie, the manufacturer of the MR system) the primary biologic effects associated
atric patients may be confused or experi- must recommend steps to reduce or alle- with exposure to RF radiation are related
ence heightened anxiety (49,51). Sedated viate the noise perceived by the patient. to the thermogenic qualities of this elec-
patients may experience discomfort due A single upper limit of 140 dB is applied tromagnetic field (11,67–77).
to high noise levels. Certain drugs are to peak acoustic noise (2). However, the Prior to 1985, there were no published
known to increase hearing sensitivity instructions for use of MR systems must reports concerning thermal or other
(52). Neonates with immature anatomic advise the MR system operator to provide physiologic responses of human subjects
development may have an increased re- hearing protection to patients for opera- exposed to RF radiation during MR pro-
action to acoustic noise, as has been re- tion above an acoustic noise level of 99 cedures. Since then, many investigations
ported by Philbin et al (53). dB (2). have been conducted to characterize the
In general, acoustic noise levels re- thermal effects of MR procedure–related
corded by various researchers in associa- heating (68 –74,78). This topic has been
Characteristics of MR-related
tion with conventional or routine MR reviewed by Schaefer (67,76) and Shel-
Acoustic Noise
procedures have been below the maxi- lock (11).
Variations in MR-related acoustic noise mum limit permissible by the U.S. Occu-
occur with alterations in the gradient out- pational Safety and Health Administra-
MR Procedures and Specific
put (rise time or amplitude) associated tion (2). Notably, when one considers
Absorption Rate of RF Radiation
with different MR parameters (49,54 – 64). that the duration of exposure is one of
Noise levels, pitch, and frequency char- the more important physical factors that Thermoregulatory and other physio-
acteristics are predominantly increased determine the effect of noise on hearing, logic changes that a human subject ex-
when section thickness, field of view, rep- then acoustic noise levels associated with hibits in response to exposure to RF radi-
etition time, and echo time are decreased. MR procedures do not tend to be prob- ation are dependent on the amount of
The physical features of the MR system, lematic because of the relative short pe- energy that is absorbed. The dosimetric
especially the presence or absence of spe- riods of exposure (65,66). term used to describe the absorption of
cial sound insulation, and the material RF radiation is the specific absorption
and construction of gradient coils and rate (SAR) (11,67,76,79). The SAR is the
Prevention of Acoustic Noise
support structures also affect the trans- mass normalized rate at which RF power
Problems
mission of acoustic noise and its percep- is coupled to biologic tissue and is typi-
tion by the patient. Various techniques have been de- cally expressed in watts per kilogram. The
The patient’s presence and the pa- scribed to attenuate noise and, thus, pre- relative amount of RF radiation that an
tient’s size also affect the level of acoustic vent problems or hazards associated with individual encounters during an MR pro-
noise. An increase in acoustic noise has exposure to MR-related acoustic noise cedure is usually characterized with re-
been reported with a patient or volunteer (49,64). The simplest and least expensive spect to the whole-body averaged and
present in the bore of the MR system means is to use disposable earplugs or peak SAR levels (ie, the SAR averaged in
(63); this may be due to pressure dou- commercially available noise-abatement 1 g of tissue).
bling (ie, an increase in sound pressure) headphones (49). Earplugs, when prop- Measurements or estimates of SAR are
close to an object, as sound waves reflect erly used, can decrease noise by 10 –30 not trivial, particularly in human sub-
and undergo in-phase enhancement. dB, which usually affords adequate pro- jects. There are several methods of deter-
Noise characteristics also have a spatial tection for MR environments with rela- mining this parameter for the purpose of
dependence. For example, noise levels tively loud MR systems. Regardless of the RF energy dosimetry in association with
have been found to vary by as much as 10 technique used, facilities operating with MR procedures (67,76,79,80). The SAR
dB as a function of patient position along MR systems that generate substantial that is produced during an MR procedure
the magnet bore (63). acoustic noise should require all patients is a complex function of numerous vari-
MR-related acoustic noise levels have undergoing an examination to wear a ables, including the frequency (ie, deter-
been measured during a variety of pulse protective hearing device. Exposure of mined by the strength of the static mag-
sequences for MR systems with static staff members, health care workers, and netic field of the MR system), the
magnetic field strengths ranging from 0.2 other individuals (eg, relatives, visitors) to repetition time, the type of RF coil used,
to 4.7 T (54 –56,61– 64). Recent studies per- loud MR systems is also of concern (49,56). the volume of tissue contained within
formed with MR parameters that included Therefore, these individuals should like- the coil, the configuration of the ana-
MR Procedure–related Heating
quences related to these exposures to RF doubling of field strength (eg, 1.5 vs 3.0
Thermophysiologic responses to MR radiation (87). T), the RF power deposition increases four
procedure–related heating depend on Several studies were subsequently con- times for a given MR imaging pulse se-
multiple physiologic, physical, and envi- ducted with volunteer subjects and pa- quence. Therefore, investigations are
ronmental factors (11,67,76,77). These tients undergoing clinical MR procedures needed for characterization of thermal
include the duration of exposure, the rate with the intent of obtaining information responses in human subjects to deter-
at which energy is deposited, the status that would be applicable to patient pop- mine potential thermogenic hazards as-
of the patient’s thermoregulatory system, ulations typically encountered in the MR sociated with the use of these powerful
the presence of an underlying health setting (68–75). These investigations dem- MR devices. To date, however, with the
condition, and the ambient conditions onstrated that changes in body tempera- exception of work conducted at 8 T by
within the MR system. ture were relatively minor (ie, ⬍0.6°C). Kangarlu et al (74), there has been virtu-
With regard to the thermoregulatory While there was a tendency for statisti- ally no investigation of MR procedure–
system, when subjected to a thermal cally significant increases in skin temper- related heating with regard to very high
challenge the human body loses heat by atures to occur, these were of no serious field strength MR systems.
means of convection, conduction, radia- physiologic consequence.
tion, and evaporation. Each of these Of interest, various studies reported a
mechanisms is responsible to a varying poor correlation between body or skin MR SAFETY AND PATIENT
degree for heat dissipation as the body temperature changes versus whole-body CARE
attempts to maintain thermal homeosta- averaged SARs during clinical MR proce-
Screening Patients for MR
sis (11,67,77,79). If the thermoregulatory dures (69,73). These findings are not sur-
Procedures and Individuals
effectors are not capable of totally dissi- prising considering the range of thermo-
for the MR Environment
pating the heat load, then there is an physiologic responses possible to a given
accumulation, or storage, of heat along SAR that are dependent on the individu- The establishment of thorough and ef-
with an elevation in local and/or overall al’s thermoregulatory system and the fective screening procedures for patients
tissue temperatures (11,76,77). presence of one or more underlying con- and other individuals is one of the most
Various underlying health conditions dition(s) that can alter or impair the abil- critical components of a program to
may affect an individual’s ability to tol- ity to dissipate heat. guard the safety of all those preparing to
erate a thermal challenge, including car- An extensive investigation was con- undergo MR procedures or to enter the
diovascular disease, hypertension, diabe- ducted in volunteer subjects exposed to a MR environment (5,13,15–17,89). An im-
tes, fever, old age, and obesity (81– 85). In 1.5-T 64-MHz MR procedure with a portant aspect of protecting individuals
addition, medications such as diuretics, whole-body averaged SAR of 6.0 W/kg from MR system–related accidents and
-blockers, calcium blockers, amphetamines, (75), which, to our knowledge, is the high- injuries involves an understanding of the
muscle relaxants, and sedatives can also est level of RF energy to which human risks associated with the various im-
greatly alter thermoregulatory responses subjects have ever been exposed with an plants, devices, accessories, and other ob-
to a heat load. In fact, certain medica- MR system. This excessive amount of RF jects that may cause problems in this set-
tions have a synergistic effect with re- radiation was achieved by using non- ting (5,13,15–17). This requires obtaining
spect to tissue heating if the heating is clinical MR imaging parameters (75). information and documentation about
specifically caused by exposure to RF ra- Tympanic membrane temperature, six these objects in order to provide the saf-
diation (86). different skin temperatures, heart rate, est MR setting possible. In addition, be-
The environmental conditions that ex- blood pressure, oxygen saturation, and cause MR-related incidents have been
ist in and around the MR system will also skin blood flow were monitored (75). The due to deficiencies in screening methods
affect the tissue temperature changes as- findings indicated that an MR procedure and/or a lack of proper control of access
sociated with RF-induced heating. Dur- performed at a whole-body averaged SAR to the MR environment (especially with
ing an MR procedure, the amount of tis- of 6.0 W/kg can be physiologically toler- regard to preventing personal items and
sue heating that occurs and the concomitant ated by an individual with normal ther- other potentially problematic objects
exposure to RF energy that is tolerable are moregulatory function (75). from entering the MR room) (3,4), it is
dependent on environmental factors that crucial to set up procedures and guide-
include ambient temperature, relative lines to prevent such incidents from oc-
MR Procedure–related Heating and
humidity, and airflow. curring. Various guidelines and recommen-
Very High Field Strength MR
dations have been developed to facilitate the
Systems
screening process (15,17,88,89).
MR Procedure–related Heating and
There are over 200 MR systems operat- Screening patients for MR.—Certain as-
Human Subjects
ing with a static magnetic field strength pects of screening patients for MR proce-
To our knowledge, the first study of of 3 T, several operating at 4 T, a few dures may take place during the schedul-
human thermal response to RF radiation– operating at 7 T, one operating at 8 T ing process. This must be conducted by a
induced heating during an MR procedure (74), and at least one MR unit that oper- health care worker who is specially
was conducted by Schaefer et al (87). Tem- ates at a field strength higher than 8 T is trained in MR safety (17,88,89). That is,
perature changes and other physiologic pa- in the final stage of installation (likely this individual should be (a) trained to
patient has any implant that may be con- ify the information on the form and to
traindicated for the MR procedure (eg, fer- allow discussion of any question or con-
Metallic Orbital Foreign Bodies
romagnetic aneurysm clip, pacemaker) or cern that the patient may have before
and Screening
if there is any condition that requires care- undergoing the MR procedure. This al-
ful consideration (eg, patient is pregnant lows for clarification or confirmation of The single case report in 1986 by Kelly
or has a disability). Preliminary screening the answers to the questions posed to the et al (90) about a patient who sustained
helps to prevent scheduling of patients patient so that there is no miscommuni- an ocular injury from a retained metallic
who may be inappropriate candidates for cation regarding important MR safety is- foreign body has led to controversy re-
MR examinations. sues. In addition, because the patient garding the procedure required to screen
At the facility, it is advisable for every may not be fully aware of the medical individuals prior to their entry to the MR
patient to undergo comprehensive screen- terminology used for a particular implant environment (91–93). To date, this inci-
ing in preparation for the MR examina- or device, it is imperative that this partic- dent is the only serious eye-related injury
tion. Comprehensive patient screening ular information on the form be dis- that has occurred in association with the
involves the use of a printed form to doc- cussed during the oral interview. MR setting, according to recent review of
ument the screening procedure, a review It should be noted that having under- the peer-reviewed literature and review of
of the information on the screening gone a previous MR procedure without data files from the Manufacturer and
form, and an oral interview to verify the incident does not guarantee a safe subse- User Facility Device Experience Database
information and allow discussion of any quent MR examination. Various factors (MAUDE; available at www.fda.gov/cdrh
question or concern that the patient may (eg, static magnetic field strength of the /maude.html) and the Medical Device Re-
have (15,88,89). A health care worker MR system, orientation of the patient, port (available at www.fda.gov/CDRH
trained in MR safety must conduct this orientation of a metallic implant or ob- /mdrfile.html), both from the FDA Center
aspect of patient screening. Various ject) can substantially change the sce- for Devices and Radiological Health.
forms have been developed for screening nario (17,88,89). Therefore, a compre- In the past, any individual or patient
patients in preparation for MR proce- hensive screening procedure must be suspected of having an orbital foreign
dures (5,15,17–19,88,89). An example of conducted each time a patient prepares body typically underwent screening with
a recently developed form for this use is to undergo an MR procedure. This is not conventional radiography of the orbits to
shown in Figure 1 (18,19). an inconsequential matter, because a sur- determine whether a metallic object was
With the use of any type of written gical intervention or accident involving a present. Thus, screening radiographs of
questionnaire, limitations exist related to metallic foreign body may have occurred the orbits were obtained routinely not
incomplete or incorrect answers pro- that could affect the safety of the patient only in individuals who had a history of
vided by the patient (18,19,88,89). For entering the MR environment. injury from a foreign body but also in
example, there may be difficulties associ- Screening individuals for the MR environ- those who simply had a history of expo-
ated with patients who are impaired with ment.—Similar to the procedure con- sure to metallic objects, such as welders,
respect to their vision, language fluency, ducted for screening patients, all other grinders, metal workers, sculptors, and
or level of literacy. Therefore, an appro- individuals (eg, MR technologists, pa- others. Obviously, conventional radio-
priate accompanying family member or tient’s family members, visitors, allied graphs of the orbits may have been ob-
other individual (eg, referring physician) health professionals, maintenance work- tained unnecessarily in many individuals
should be involved in the screening pro- ers, custodial workers, firefighters, secu- because of this policy.
cess to verify any information that may rity officers) should undergo screening by Seidenwurm et al (93) presented research
affect patient safety. Versions of this using appropriate guidelines before being and a new set of guidelines for radiographic
form should also be available in other allowed into the MR environment (17– screening of individuals suspected of
languages, as needed (ie, specific to the 19). This involves the use of a printed having metallic foreign bodies. Their in-
demographics of the population served form to document the screening proce- vestigation addressed the cost-effective-
by the MR facility) (17,88). dure, a review of the information on the ness of the use of a clinical versus a radio-
In the event that the patient is coma- form, and an oral interview to verify the graphic technique to screen individuals for
tose or unable to communicate, the form information and allow discussion of any orbital foreign bodies before an MR proce-
should be completed by the most quali- question or concern that the individual dure (93). The costs of screening were
fied individual (eg, physician, family may have before entry to the MR envi- determined on the basis of published
member) with knowledge of the patient’s ronment is permitted. data, disability rating guides, and results
medical history and present condition. If In general, MR screening forms were of a practice survey. A sensitivity analysis
the screening information is inadequate, developed with patients in mind and, was performed for each variable. For their
it is advisable to look for surgical scars therefore, contain many questions that analysis, the benefit of screening was pre-
on the patient and/or to obtain conven- are inappropriate or confusing to other vention of immediate, permanent, nona-
tional radiographs of the skull and/or individuals who may need to enter the meliorable, or unilateral blindness. Sei-
chest to search for implants that may be MR environment. Therefore, a screening denwurm et al (93) implemented the
particularly hazardous in the MR environ- form was recently created for individuals following policy: “If a patient reports in-
ment (eg, aneurysm clip, cardiac pace- who need to enter the MR environment jury from an ocular foreign body that was
maker). and/or MR system room (Fig 2) (18,19). subsequently removed by a doctor or
640
䡠
Radiology
䡠
September 2004
Figure 1. Example of an MR procedure screening form for patients. (Reprinted, with permission, from the Institute for Magnetic Resonance Safety, Education, and Research.)
the infinite possibilities of pacing sys- verification of lead position, evaluation the bilateral neurostimulation systems to
tems and cardiac and lead geometry, as of patients with poor or worsening out- characterize worst-case and clinically rel-
well as variable RF and gradient magnetic come, and examination of patients with evant positioning scenarios (106,107).
fields, absolute safety with regard to pace- other pathologic abnormalities unrelated MR imaging procedures were performed
maker and MR interactions cannot be as- to DBS neurostimulation, such as stroke, on a gel-filled phantom designed to ap-
sured under all operational conditions. tumor, or hemorrhage (107). proximate the head and upper torso of a
Nevertheless, on the basis of information As with all electronically activated de- human subject. Temperature changes
in the peer-reviewed literature it appears vices in the MR environment, it is gener- were studied in association with MR ex-
that with appropriate patient selection, as ally recommended that patients with a aminations conducted at 1.5 T and 64
well as continuous monitoring and pre- neurostimulation system should not un- MHz at various levels of RF energy by
paredness for resuscitation efforts, perfor- dergo MR imaging because of the poten- using the transmit-receive RF body coil
mance of MR procedures in patients with tial for serious consequences, including and transmit-receive RF head coil. The
an implanted cardiac pacemaker but who movement of the leads or implantable findings from these studies indicated that
are not pacemaker dependent may be pulse generator, excessive MR imaging– substantial heating occurs under certain
achieved with reasonable safety, even at related heating, induced electric currents, conditions, while other conditions pro-
static magnetic field strengths of 1.5 T. and functional disruption of the opera- duced relatively minor physiologically
In the past, the presence of any elec- tional aspects of the device (5–7). Thus, inconsequential temperature increases.
tronically activated implant was consid- before performing MR in a patient with a Furthermore, factors that strongly influ-
ered a strict contraindication for an indi- DBS system, it is essential to collect in enced local temperature increases at the
vidual in the MR environment. Over the vitro experimental data to define MR electrode tip included the positioning of
years, however, various studies have been conditions that may permit imaging to the neurostimulation system (especially
performed to define safety criteria for be performed safely (106,107). the electrode), the type of RF coil used,
electronic devices (104,106 –108). There- From an MR safety point of view, the and the SAR used for the MR procedure.
fore, if highly specific guidelines are fol- greatest concern for electronically acti- According to the study by Rezai et al
lowed, MR procedures may be conducted vated or electrically conductive implants (106), MR-related heating does not ap-
safely in patients with various electroni- in the brain is excessive MR imaging– pear to present a major safety concern for
cally activated implants, including neuro- related heating, which can cause irrevers- patients with the bilateral neurostimula-
stimulation systems, cochlear implants, ible tissue damage (106,107). Results tion systems that underwent testing, as
and programmable drug infusion pumps from studies conducted to date (106,107) long as highly specific guidelines pertain-
(5–7,104,106 –108). In fact, some of these and a recent report (195) revealed that ing to the positioning of these neuro-
electronically activated devices have re- there is a realistic potential for injury due stimulation devices and to the parame-
ceived approval from the FDA for “MR to excessive MR imaging–related heating ters used for MR imaging are carefully
safe” labeling claims. of neurostimulation systems used for adhered to. Finelli et al (107) reported
In consideration of the findings for DBS. that MR imaging sequences commonly
conducting safe MR procedures in pa- Recently, investigators have evaluated used for clinical procedures can be per-
tients with electronically activated de- MR-related heating for the only neuro- formed safely with the use of a transmit-
vices that have been published in the stimulation system (Activa Tremor Con- receive RF head coil at 1.5 T in patients
peer-reviewed literature, it is hoped that trol System; Medtronic) approved by the with a bilateral DBS system.
cardiac pacemaker manufacturers will FDA for use in chronic DBS (106,107). It should be noted that most present-
be encouraged to proactively support This neurostimulation system is a fully day high-field-strength MR systems are
and/or conduct investigations directed implantable multiprogrammable device used with a body coil to transmit RF and
toward identifying safety criteria for their designed to deliver electric stimulation to a receive-only head coil. Therefore, addi-
respective devices. This will ultimately the thalamus or other brain structures. tional studies are required to characterize
have a substantial effect on patient care The basic implantable system is com- the effect of the use of this transmit-re-
and the overall health care of patients posed of the neurostimulator (or im- ceive RF coil combination with regard to
with pacemakers who may require MR plantable pulse generator), the DBS lead, MR imaging–related heating of neuro-
procedures. and an extension that connects the lead stimulation systems used for DBS.
Neurostimulation system for deep brain to the implantable pulse generator. This It is important to note that the exact
stimulation.—Because of the increased in- neurostimulation system delivers high- safety recommendations for the particu-
terest in the use of deep brain stimulation frequency electric stimulation to a multi- lar neurostimulation system, with regard
(DBS) of the thalamus, globus pallidus, ple-contact electrode placed in the ven- to the pulse generator, leads, electrodes,
and subthalamic nucleus for treatment of tral intermediate nucleus of the thalamus operational conditions for the device, po-
medically refractory movement disorders or another anatomic site. sitioning of these components, and MR
and other types of neurologic conditions, In their studies on neurostimulation system conditions, must be carefully fol-
the number of patients receiving im- systems, Rezai et al (106) and Finelli et al lowed for MR imaging (106,107,195). As
plantable pulse generators and DBS elec- (107) indicated that MR safety for neuro- highlighted by two recent serious acci-
trodes is rapidly growing (106,107,192– stimulation systems is highly dependent dents (195), the failure to follow safety
194). The use of MR imaging in patients on a number of critical factors. To simu- recommendations strictly may result in
cialized neurostimulation device (Enterra magnetic become firmly incorporated mended thickness, 1-cm) should be placed
Therapy Gastric Electrical Stimulation into tissue 6 – 8 weeks after placement. In between the patient’s skin and transmit RF
coil that is used for the MR procedure (al-
System; Medtronic), is indicated for treat- these cases, retentive forces, or counter-
ternatively, the RF coil itself should be pad-
ment of patients with chronic intractable forces, provided by tissue ingrowth, scar-
ded). For example, position the patient so
nausea and vomiting secondary to gas- ring, or granulation essentially serve to
that there is no direct contact between the
troparesis of diabetic or idiopathic origin. prevent these objects from presenting
patient’s skin and the body RF coil of the
Gastric electric stimulation uses mild hazards to individuals in the MR envi- MR system. This may be accomplished by
electric pulses to stimulate the stomach ronment. Those implants or devices that having the patient place his/her arms over
to help control symptoms associated may be weakly magnetic but that are rig- his/her head or by using elbow pads or foam
with gastroparesis. idly fixed in the body, such as a bone padding between the patient’s tissue and
The gastric electric stimulation device screw, may be studied immediately in the the body RF coil of the MR system. This is
is composed of a neurostimulator, an im- postoperative period (19). Typically, spe- especially important for those MR examina-
plantable intramuscular lead, and an ex- cific information pertaining to the rec- tions that use the body coil or other large RF
ternal programming system. Currently, ommended postoperative waiting period coils for transmission of RF energy.
the use of MR procedures in patients with may be found in the labeling or product 4. Use only electrically conductive devices,
this device is contraindicated owing to insert for a weakly magnetic implant or equipment, accessories (eg, ECG leads, elec-
possible hazards related to dislodgment device. trodes, etc), and materials that have been
or heating of the neurostimulator and/or If there is any concern regarding the thoroughly tested and determined to be
the leads used for stimulation. In addi- integrity of the tissue with respect to its safe and compatible for MR procedures.
tion, the voltage induced through the ability to retain the implant or object in 5. Carefully follow specific MR safety cri-
lead and neurostimulator may cause un- place or if the implant cannot be prop- teria and recommendations for implants
comfortable “jolting” or “shocking” lev- erly identified, the individual in such made from electrically conductive materials
els of stimulation (5). cases should not be exposed to the MR (eg, bone fusion stimulators, neurostimula-
environment. Specific information per- tion systems, etc).
6. Before using electrical equipment,
taining to the recommended postopera-
Postoperative MR Procedures check the integrity of the insulation and/or
tive waiting period may be found in the
housing of all components including sur-
Because confusion exists regarding the labeling or product insert for a weakly
face RF coils, monitoring leads, cables, and
issue of performing an MR procedure magnetic implant or device. wires. Preventive maintenance should be
during the postoperative period in a pa- practiced routinely for such equipment.
tient with a metallic implant or device, CONCLUSIONS 7. Remove all non-essential electrically
guidelines have been developed pertain- conductive materials from the MR system
ing to this MR safety topic (19). Study With the continued advances in MR (ie, unused surface RF coils, ECG leads, ca-
results have supported that, if a metallic technology and the development of bles, wires, etc).
object is a passive implant (ie, there is no more sophisticated implants and devices, 8. Keep electrically conductive materials
electronically or magnetically activated there is an increased potential for hazard- that must remain in the MR system from
component associated with the opera- ous situations to occur in the MR envi- directly contacting the patient by placing
tion of the object) and it is made from a ronment. Therefore, to prevent incidents thermal and/or electrical insulation be-
nonferromagnetic material (eg, titanium, and accidents, it is necessary to be aware tween the conductive material and the pa-
titanium alloy, nitinol), the patient with of the latest information pertaining to tient.
the object may undergo an MR procedure 9. Keep electrically conductive materials
MR biologic effects, to use current evi-
at 1.5 T or lower immediately after im- that must remain within the body RF coil or
dence-based guidelines to ensure safety
plantation (5–7,135,145,149,159). In fact, other transmit RF coil of the MR system
for patients and staff members, and to
there are several reports that describe from forming conductive loops. Note: The
follow proper recommendations pertain- patient’s tissue is conductive and, there-
placement of vascular stents and other im- ing to biomedical implants and devices.
plants with MR-guided procedures that in- fore, may be involved in the formation of a
conductive loop, which can be circular, U-
clude the use of high-field-strength (1.5-T)
APPENDIX A shaped, or S-shaped.
MR systems (164,167,168). In addition, a
10. Position electrically conductive mate-
patient or individual with a nonferromag-
The following guidelines, reprinted, with rials to prevent “cross points.” For example,
netic passive implant would be allowed to
permission, from the Institute for Magnetic a cross point is the point where a cable
enter the MR environment associated with Resonance Safety, Education, and Research crosses another cable, where a cable loops
a 1.5-T or lower-strength MR system imme- (19), pertain to the prevention of excessive across itself, or where a cable touches either
diately after implantation of such an ob- heating and burns in association with MR the patient or sides of the transmit RF coil
ject. Currently, there are few data to pro- procedures. more than once. Even the close proximity
vide guidelines for MR environments with 1. Prepare the patient for the MR proce- of conductive materials with each other
imagers operating at 3 T or higher. dure by ensuring that there are no unnec- should be avoided because some cables and
For an implant or device that exhibits essary metallic objects contacting the pa- RF coils can capacitively couple (without
weakly magnetic qualities, it is typically tient’s skin (eg, metallic drug delivery patches, any contact or crossover) when placed close
necessary to wait for 6 – 8 weeks after im- jewelry, necklaces, bracelets, key chains, etc). together.
33. Schwartz J, Crooks L. NMR imaging pro- 51. Quirk ME, Letendre AJ, Ciottone RA, 68. Morvan D, Leroy-Willig A, Jehenson P,
duces no observable mutations or cyto- Lingley JF. Anxiety in patients undergo- Cuenod CA, Syrota A. Temperature
toxicity in mammalian cells. AJR Am J ing MR imaging. Radiology 1989; 170: changes induced in human muscle by
Roentgenol 1982; 139:583–585. 463– 466. radiofrequency H-1 coupling: measure-
34. Tyndall DA, Sulik KK. Effects of mag- 52. Laurell GF. Combined effects of noise ment with an MR imaging diffusion
netic resonance imaging on eye devel- and cisplatin: short- and long-term fol- technique. Radiology 1992; 185:871–
opment in the C57BL/6J mouse. Teratol- low-up. Ann Otol Rhinol Laryngol 1992; 874.
ogy 1991; 43:263–275. 101:969 –976. 69. Shellock FG, Crues JV. Temperature,
35. Heinrichs WL, Fong P, Flannery M, et al. 53. Philbin MK, Taber KH, Hayman LA. Pre- heart rate, and blood pressure changes
Midgestational exposure of pregnant liminary report: changes in vital signs of associated with clinical MR imaging at
BALB/c mice to magnetic resonance im- term newborns during MR. AJNR Am J 1.5 T. Radiology 1987; 163:259 –262.
aging. Magn Reson Imaging 1988; Neuroradiol 1996; 17:1033–1036. 70. Shellock FG, Crues JV. Temperature
6:305–313. 54. Counter SA, Olofsson A, Borg E. Bjelke B, changes caused by clinical MR imaging
36. McRobbie D, Foster MA. Pulsed mag- Haggstrom A, Grahn HF. Analysis of of the brain at 1.5 tesla using a head coil.
netic field exposure during pregnancy magnetic resonance imaging acoustic AJNR Am J Neuroradiol 1988; 9:287–
and implications for NMR foetal imag- noise generated by a 4.7 T experimental 291.
ing: a study with mice. Magn Reson Im- system. Acta Otolaryngol 2000; 120: 71. Shellock FG, Rothman B, Sarti D. Heat-
aging 1985; 3:231–234. 739 –743. ing of the scrotum by high-field-
37. Brockway J, Bream P. Does memory loss 55. Price DL, De Wilde JP, Papadaki AM, strength MR imaging. AJR Am J Roent-
occur after MR imaging? J Magn Reson Curran JS, Kitney RI. Investigation of genol 1990; 154:1229 –1232.
Imaging 1992; 2:721–728. acoustic noise on 15 MRI scanners from 72. Shellock FG, Schatz CJ. Increases in cor-
38. Besson J, Foreman E, Eastwood L, Smith 0.2 T to 3 T. J Magn Reson Imaging neal temperature caused by MR imaging
F, Ashcroft G. Cognitive evaluation fol- 2001; 13:288 –293. of the eye with a dedicated local coil.
lowing NMR imaging of the brain. 56. Shellock FG, Ziarati M, Atkinson D, Radiology 1992; 185:697– 699.
J Neurol Neurosurg Psychiatry 1984; 47: Chen DY. Determination of gradient 73. Shellock FG, Schaefer DJ, Crues JV. Al-
314 –316. magnetic field-induced acoustic noise terations in body and skin temperatures
39. Nyenhuis JA, Bourland J, Kildishev AV, associated with the use of echo planar caused by MR imaging: is the recom-
Schaefer DJ. Health effects and safety of and three-dimensional fast spin echo mended exposure for radiofrequency ra-
intense gradient fields. In: Shellock FG, techniques. J Magn Reson Imaging diation too conservative? Br J Radiol
ed. Magnetic resonance procedures: 1998; 8:1154 –1157. 1989; 62:904 –909.
health effects and safety. Boca Raton, 57. Strainer JC, Ulmer JL, Yetkin FZ, Haugh- 74. Kangarlu A, Shellock FG, Chakeres D.
Fla: CRC, 2001; 31–54. ton VM, Daniels DL, Millen SJ. Func- 8.0-tesla MR system: temperature changes
40. Bourland JD, Nyenhuis JA, Schaefer tional MR of the primary auditory cor- associated with radiofrequency-induced
DJ. Physiologic effects of intense MRI tex: an analysis of pure tone activation heating of a head phantom. J Magn Reson
gradient fields. Neuroimaging Clin N and tone discrimination. AJNR Am J Imaging 2003; 17:220 –226.
Am 1999; 9:363–377. Neuroradiol 1997; 18:601– 610. 75. Shellock FG, Schaefer DJ, Kanal E. Phys-
41. Cohen MS, Weisskoff R, Kantor H. Sen- 58. Ulmer JL, Bharat BB, Leighton PM, et al. iologic responses to MR imaging per-
sory stimulation by time varying mag- Acoustic echo planar scanner noise and formed at an SAR level of 6.0 W/kg. Ra-
netic fields. Magn Reson Med 1990; 14: pure tone hearing thresholds: the effects diology 1994; 192:865– 868.
409 – 414. of sequence repetition times and acous- 76. Schaefer DJ. Health effects and safety of
42. Budinger TF, Fischer H, Hentshel D, Rein- tic noise rates. J Comput Assist Tomogr radiofrequency power deposition associ-
flder HE, Schmitt F. Physiological effects 1998; 22:480 – 486. ated with magnetic resonance proce-
of fast oscillating magnetic field gradi- 59. Bandettini PA, Jesmanowicz A, Van Ky- dures. In: Shellock FG, ed. Magnetic res-
ents. J Comput Assist Tomogr 1991; 15: len J, Birn RA, Hyde J. Functional MRI of onance procedures: health effects and
909 –914. brain activation induced by scanner safety. Boca Raton, Fla: CRC, 2001; 55–
43. Mansfield P, Harvey PR. Limits to neural acoustic noise. Magn Reson Med 1998; 74.
stimulation in echo-planar imaging. 39:410 – 416. 77. Shellock FG, Schaefer DJ. Radiofre-
Magn Reson Med 1993; 29:746 –758. 60. Goldman AM, Gossman WE, Fried- quency energy-induced heating during
44. Irnich W, Schmitt F. Magnetostimula- lander PC. Reduction of sound levels magnetic resonance procedures: labora-
tion in MRI. Magn Reson Med 1995; 33: with antinoise in MR imaging. Radiol- tory and clinical experiences. In: Shel-
619 – 623. ogy 1989; 173:549 –550. lock FG, ed. Magnetic resonance proce-
45. Ehrhardt JC, Lin CS, Magnotta VA, 61. Hurwitz R, Lane SR, Bell RA, Brant- dures: health effects and safety. Boca
Fisher DJ, Yuh WT. Peripheral nerve Zawadzki MN. Acoustic analysis of gra- Raton, Fla: CRC, 2001; 75–96.
stimulation in a whole-body echo-pla- dient-coil noise in MR imaging. Radiol- 78. Shuman WP, Haynor DR, Guy AW,
nar imaging system. J Magn Reson Im- ogy 1989; 173:545–548. Wesby GE, Schaefer DJ, Moss AA. Super-
aging 1997; 7:405– 409. 62. Counter SA, Olofsson A, Grahn, Borg E. ficial and deep-tissue increases in anes-
46. Abart J, Eberhardt K, Fischer H, et al. MRI acoustic noise: sound pressure and thetized dogs during exposure to high
Peripheral nerve stimulation by time- frequency analysis. J Magn Reson Imag- specific absorption rates in a 1.5-T MR
varying magnetic fields. J Comput Assist ing 1997; 7:606 – 611. imager. Radiology 1988; 167:551–554.
Tomogr 1997; 21:532–538. 63. Hedeen RA, Edelstein WA. Characteris- 79. Gordon CJ. Normalizing the thermal ef-
47. Ham CL, Engels JM, van de Weil GT, tics and prediction of gradient acoustic fects of radiofrequency radiation: body
Machielsen A. Peripheral nerve stimula- noise in MR imagers. Magn Reson Med mass versus total body surface area. Bio-
tion during MRI: effects of high gradient 1997; 37:7–10. electromagnetics 1987; 8:111–118.
amplitudes and switching rates. J Magn 64. Cho ZH, Park SH, Kim JH, et al. Analysis 80. National Council on Radiation Protec-
Reson Imaging 1997; 7:933–937. of acoustic noise in MRI. Magn Reson tion and Measurements. Biological ef-
48. Nyenhuis JA, Bourland JD, Schaefer DJ. Imaging 1997; 15:815– 822. fects and exposure criteria for radiofre-
Analysis from a stimulation perspective 65. Miller LE, Keller AM. Regulation of oc- quency electromagnetic fields. Report
of magnetic field patterns of MR gradi- cupational noise. In: Harris CM, ed. no. 86. Bethesda, Md: National Council
ent coils. J Appl Phys 1997; 81:4314 – Handbook of noise control. New York, on Radiation Protection and Measure-
4316. NY: McGraw-Hill, 1979; 1–16. ments, 1986.
49. McJury M. Acoustic noise and magnetic 66. Melnick W. Hearing loss from noise ex- 81. Rowell LB. Cardiovascular aspects of hu-
heat intolerance. Sports Med 1985; 2: 102. Smith CD, Nyenhuis JA, Kildishev AV. MRI interaction with tattoo pigments:
279 –286. Health effects of induced electrical cur- case report, pathophysiology, and man-
84. Barany FR. Abnormal vascular reaction rents: implications for implants. In: agement. Plast Reconstr Surg 1997; 99:
in diabetes mellitus. Acta Med Scand Shellock FG, ed. Magnetic resonance 1717–1720.
Suppl 1955; 304:556 – 624. procedures: health effects and safety. 120. Kanal E, Shellock FG. MRI interaction
85. Buskirk ER, Lundegren H, Magnusson L. Boca Raton, Fla: CRC, 2001; 393– 413. with tattoo pigments (letter). Plast Re-
Heat acclimation patterns in obese and 103. Shellock FG. MR safety update 2002: im- constr Surg 1998; 101:1150 –1151.
lean individuals. Ann N Y Acad Sci 1965; plants and devices. J Magn Reson Imag- 121. Weiss RA, Saint-Louis LA, Haik BG, Mc-
131:637– 653. ing 2002; 16:485– 496. Cord CD, Taveras JL. Mascara and eye-
86. Jauchem JR. Effects of drugs on thermal 104. Chou CK, McDougall JA, Can KW. Ab- lining tattoos: MRI artifacts. Ann Oph-
responses to microwaves. Gen Pharma- sence of radiofrequency heating from thalmol 1989; 21:129 –131.
col 1985; 16:307–310. auditory implants during magnetic res- 122. Colletti PM. Magnetic resonance proce-
87. Schaefer DJ, Barber BJ, Gordon CJ. Ther- onance imaging. Bioelectromagnetics dures and pregnancy. In: Shellock FG,
mal effects of magnetic resonance imag- 1995; 16:307–316. ed. Magnetic resonance procedures:
ing (abstr). In: Book of abstracts: Society 105. Nyenhuis JA, Kildishev AV, Foster KS, health effects and safety. Boca Raton,
of Magnetic Resonance in Medicine Graber G, Athey W. Heating near im- Fla: CRC, 2001; 149 –182.
1985. Berkeley, Calif: Society of Mag- planted medical devices by the MRI RF- 123. McCarthy SM, Filly RA, Stark DD, et al.
netic Resonance in Medicine, 1985; 925. magnetic field. IEEE Transactions on Obstetrical magnetic resonance imag-
88. Sawyer-Glover A, Shellock FG. Pre-mag- Magnetics 1999; 35:4133– 4135. ing: fetal anatomy. Radiology 1985; 154:
netic resonance procedure screening. In: 106. Rezai AR, Finelli D, Nyenhuis JA, et al. 427– 432.
Shellock FG, ed. Magnetic resonance Neurostimulator for deep brain stimula- 124. McCarthy SM, Stark DD, Filly RA, Callen
procedures: health effects and safety. tion: ex vivo evaluation of MRI-related PW, Hricak H, Higgins CB. Obstetrical
Boca Raton, Fla: CRC, 2001; 271–326. heating at 1.5-tesla. J Magn Reson Imag- magnetic resonance imaging: maternal
89. Sawyer-Glover A, Shellock FG. Pre-MRI ing 2002; 15:241–250. anatomy. Radiology 1985; 154:421– 425.
procedure screening: recommendations 107. Finelli DA, Rezai AR, Ruggieri PM, et al. 125. Tempany CM, Fielding JR. Female pel-
and safety considerations for biomedical MR imaging-related heating of deep vis. In: Edelman RR, Hesselink JR, Zlat-
implants and devices. J Magn Reson Im- brain stimulation electrodes: in vitro kin MB, eds. Clinical magnetic reso-
aging 2000; 12:92–106. study. AJNR Am J Neuroradiol 2002; 23: nance imaging. 2nd ed. Philadelphia,
90. Kelly WM, Paglen PG, Pearson JA, San Pa: Saunders, 1996; 1432–1466.
1795–1802.
Diego AG, Soloman MA. Ferromag- 126. Tyndall DA. MRI effects on the teratoge-
108. Shellock FG, Hatfield M, Simon BJ, et al.
netism of intraocular foreign body nicity of X-irradiation in the C57BL/6J
Implantable spinal fusion stimulator:
causes unilateral blindness after MR mouse. Magn Reson Imaging 1990;
assessment of MRI safety. J Magn Reson
study. AJNR Am J Neuroradiol 1986; 8:423– 433.
Imaging 2000; 12:214 –223.
7:243–245. 127. Yip YP, Capriotti C, Yip JW. Effects of
109. Knopp MV, Essig M, Debus J, Zabel HJ,
91. Otto PM, Otto RA, Virapongse C, et al. MR exposure on axonal outgrowth in
van Kaick G. Unusual burns of the lower
Screening test for detection of metallic the sympathetic nervous system of the
extremities caused by a closed conduct-
foreign objects in the orbit before mag- chick. J Magn Reson Imaging 1995;
ing loop in a patient at MR imaging.
netic resonance imaging. Invest Radiol 5:457– 462.
Radiology 1996; 200:572–575.
1992; 27:308 –311. 128. Shellock FG. Biomedical implants and
92. Jarvik JG, Ramsey JG. Radiographic 110. Becker H. The use of intradermal tattoo devices: assessment of magnetic field in-
screening for orbital foreign bodies prior to enhance the final result of nipple- teractions with a 3-tesla MR system. J
to MR imaging: is it worth it? AJNR Am J areola reconstruction. Plast Reconstr Magn Reson Imaging 2002; 16:721–732.
Neuroradiol 2000; 21:245–246. Surg 1986; 77:673– 676. 129. Shellock FG, Tkach JA, Ruggieri PM,
93. Seidenwurm DJ, McDonnell CH 3rd, 111. Halder RM, Pham HN, Hreadon JY, Masaryk T, Rasmussen P. Aneurysm
Raghavan N, Breslau J. Cost utility anal- Johnson HA. Micropigmentation for the clips: evaluation of magnetic field inter-
ysis of radiographic screening for an or- treatment of vitiligo. J Dermatol Surg actions using “long-bore” and “short-
bital foreign body before MR imaging. Oncol 1989; 15:1092–1098. bore” 3-tesla MR systems. AJNR Am J
AJNR Am J Neuroradiol 2000; 21:426 – 112. Carr JJ. Danger in performing MR imag- Neuroradiol 2003; 24:463– 471.
433. ing on women who have tattooed eye- 130. Shellock FG, Tkach JA, Ruggieri PM,
94. Dempsey MF, Condon B, Hadley DM. liner or similar types of permanent cos- Masaryk TJ. Cardiac pacemakers, ICDs,
Investigation of the factors responsible metic injections. AJR Am J Roentgenol and loop recorder: evaluation of trans-
for burns during MRI. J Magn Reson Im- 1995; 165:1546 –1547. lational attraction using conventional
aging 2001; 13:627– 631. 113. Jackson JG, Acker H. Permanent eyeliner (“long-bore”) and “short-bore” 1.5- and
95. A new MRI complication? Health De- and MR imaging (letter). AJR Am J 3-tesla MR systems. J Cardiovasc Magn
vices Alert, May 27, 1988; 1. Roentgenol 1987; 149:1080. Reson 2003; 5:387–397.
96. Thermal injuries and patient monitor- 114. Lund G, Wirtschafter JD, Nelson JD, 131. Kangarlu A, Shellock FG. Aneurysm
ing during MRI studies. Health Devices Williams PA. Tattooing of eyelids: mag- clips: evaluation of magnetic field inter-
1991; 20:362–363. netic resonance imaging artifacts. Oph- actions with an 8.0 T MR system. J Magn
97. Jones S, Jaffe W, Alvi R. Burns associated thalmic Surg 1986; 17:550 –553. Reson Imaging 2000; 12:107–111.
with electrocardiographic monitoring 115. Tope WD, Shellock FG. Magnetic reso- 132. Yuh WT, Hanigan MT, Nerad JA, et al.
during magnetic resonance imaging. nance imaging and permanent cosmet- Extrusion of a magnetic eye implant af-
Burns 1996; 22:420 – 421. ics (tattoos): survey of complications ter MR examination: a potential hazard
98. Keens SJ, Laurence AS. Burns caused by and adverse events. J Magn Reson Imag- to the enucleated eye. J Magn Reson Im-
ECG monitoring during MRI imaging. ing 2002; 15:180 –184. aging 1991; 1:711–713.
Anaesthesia 1996; 51:1188 –1189. 116. U.S. Food and Drug Administration, 133. Fagan LL, Shellock FG, Brenner RJ, Roth-
99. Kugel H, Bremer C, Puschel M, et al. Center for Food Safety and Applied Nu- man B. Ex vivo evaluation of ferromag-
Hazardous situation in the MR bore: in- trition. Office of Cosmetics and Colors netism, heating, and artifacts of breast
duction in ECG leads causes fire. Eur fact sheet: tattoos and permanent tissue expanders exposed to a 1.5 T MR
Radiol 2003; 13:690 – 694. makeup. Rockville, Md: U.S. Food and system. J Magn Reson Imaging 1995;
100. Nakamura T, Fukuda K, Hayakawa K, et Drug Administration, 2000. 5:614 – 616.
al. Mechanism of burn injury during 117. Vahlensieck M. Tattoo-related cutane- 134. Shellock FG. MR imaging and cervical
magnetic resonance imaging (MRI): ous inflammation (burn grade I) in a fixation devices: evaluation of ferromag-
1998; 207:822– 825. 153. Soulen RL. Reply (letter). Radiology 1997; 46:424 – 429.
136. Shellock FG, Shellock VJ. Cardiovascular 1986; 158:279. 170. Shellock FG. New metallic implant used
catheters and accessories: ex vivo testing 154. Soulen RL, Budinger TF, Higgins CB. for permanent female contraception:
of ferromagnetism, heating, and arti- Magnetic resonance imaging of pros- evaluation of MR safety. AJR Am J
facts associated with MRI. J Magn Reson thetic heart valves. Radiology 1985; 154: Roentgenol 2002; 178:1513–1516.
Imaging 1998; 8:1338 –1342. 705–707. 171. Erlebacher JA, Cahill PT, Pannizzo F,
137. Becker RL, Norfray JF, Teitelbaum GP, et 155. Edwards MB, Taylor KM, Shellock FG. Knowles RJ. Effect of magnetic reso-
al. MR imaging in patients with intra- Prosthetic heart valves: evaluation of nance imaging on DDD pacemakers.
cranial aneurysm clips. AJNR Am J Neu- magnetic field interactions, heating, Am J Cardiol 1986; 57:437– 440.
roradiol 1988; 9:885– 889. and artifacts at 1.5-tesla. J Magn Reson 172. Hayes DL, Holmes DR, Gray JE. Effect of
138. Brothers MF, Fox AJ, Lee DH, Pelz DM, Imaging 2000; 12:363–369. 1.5 tesla nuclear magnetic resonance
Deveikis JP. MR imaging after surgery 156. Frank H, Buxbaum P, Huber L, et al. In imaging scanner on implanted perma-
for vertebrobasilar aneurysm. AJNR vitro behavior of mechanical heart nent pacemakers. J Am Coll Cardiol
Am J Neuroradiol 1990; 11:149 –161. valves in 1.5 T superconducting magnet. 1987; 10:782–786.
139. Brown MA, Carden JA, Coleman RE, et Eur J Radiol 1992; 2:555–558. 173. Holmes DR Jr, Hayes DL, Gray JE, Meri-
al. Magnetic field effects on surgical li- 157. Pruefer D, Kalden P, Schreiber W, et al. deth J. The effects of magnetic reso-
gation clips. Magn Reson Imaging 1987; In vitro investigation of prosthetic heart nance imaging on implantable pulse
5:443– 453. valves in magnetic resonance imaging: generators. Pacing Clin Electrophysiol
140. Burtscher IM, Owman T, Romner B, evaluation of potential hazards. J Heart 1986; 9:360 –370.
Stahlberg F, Holtas S. Aneurysm clip MR Valve Dis 2001; 10:410 – 414. 174. Zimmermann BH, Faul DD. Artifacts
artifacts: titanium versus stainless steel 158. Shellock FG. Prosthetic heart valves and and hazards in NMR imaging due to
and influence of imaging parameters. annuloplasty rings: assessment of mag- metal implants and cardiac pacemakers.
Acta Radiol 1998; 39:70 –76. netic field interactions, heating, and ar- Diagn Imaging Clin Med 1984; 53:53–
141. Kanal E, Shellock FG. Aneurysm clips: tifacts at 1.5 tesla. J Cardiovasc Magn 56.
effects of long-term and multiple expo- Reson 2001; 3:317–324. 175. Achenbach S, Moshage W, Diem B,
sures to a 1.5-T MR system. Radiology 159. Teitelbaum GP, Bradley WG, Klein BD. Bieberle T, Schibgilla V, Bachman K. Ef-
1999; 210:563–565. MR imaging artifacts, ferromagnetism, fects of magnetic resonance imaging on
142. Kanal E, Shellock FG, Lewin JS. Aneu- and magnetic torque of intravascular fil- cardiac pacemakers and electrodes. Am
Heart J 1997; 134:467– 473.
rysm clip testing for ferromagnetic ters, stents, and coils. Radiology 1988;
176. Peden CJ, Collins AG, Butson PC, Whit-
properties: clip variability issues. Radiol- 166:657– 664.
man J, Young I. Induction of microcur-
ogy 1996; 200:576 –578. 160. Shellock FG, Shellock VJ. Stents: evalua-
rents in critically ill patients in magnetic
143. Klucznik RP, Carrier DA, Pyka R, Haid tion of MRI safety. AJR Am J Roentgenol
resonance systems. Crit Care Med 1993;
RW. Placement of a ferromagnetic intra- 1999; 173:543–546.
21:1923–1928.
cerebral aneurysm clip in a magnetic 161. Hennemeyer CT, Wicklow K, Feinberg
177. Fontaine JM, Mohamed FB, Gottlieb C,
field with a fatal outcome. Radiology DA, Derdeyn CP. In vitro evaluation of
Callans DJ, Marchlinski FE. Rapid ven-
1993; 187:855– 856. platinum Guglielmi detachable coils at
tricular pacing in a pacemaker patient
144. New PF, Rosen BR, Brady TJ, et al. Po- 3-T with a porcine model: safety issues
undergoing magnetic resonance imag-
tential hazards and artifacts of ferromag- and artifacts. Radiology 2001; 219:732–
ing. Pacing Clin Electrophysiol 1998;
netic and nonferromagnetic surgical 737. 21:1336 –1339.
and dental materials and devices in nu- 162. Hug J, Nagel E, Bornstedt A, Schacken- 178. Fetter J, Aram G, Holmes DR, Gray JE,
clear magnetic resonance imaging. Radi- burg B, Oswald H, Fleck E. Coronary ar- Hayes DL. The effects of nuclear mag-
ology 1983; 147:139 –148. terial stents: safety and artifacts during netic resonance imagers on external and
145. Pride GL, Kowal J, Mendelsohn DB, MR imaging. Radiology 2000; 216:781– implantable pulse generators. Pacing
Chason DP, Fleckenstein JL. Safety of 787. Clin Electrophysiol 1984; 7:720 –727.
MR scanning in patients with nonferro- 163. Kiproff PM, Deeb DL, Contractor FM, 179. Vahlhaus C, Sommer T, Lewalter T, et al.
magnetic aneurysm clips. J Magn Reson Khoury MB. Magnetic resonance charac- Interference with cardiac pacemakers by
Imaging 2000; 12:198 –200. teristics of the LGM vena cava filter: magnetic resonance imaging: are there
146. Shellock FG, Kanal E. Yasargil clips: eval- technical note. Cardiovasc Intervent Ra- irreversible changes at 0.5 tesla? Pacing
uation of interactions with a 1.5-T MR diol 1991; 14:254 –255. Clin Electrophysiol 2001; 24(4 pt 1):489 –
system. Radiology 1998; 207:587–591. 164. Manke C, Nitz WR, Djavidani B, et al. 495.
147. Romner B, Olsson M, Ljunggren B, et al. MR imaging– guided stent placement in 180. Gimbel JR, Johnson D, Levine PA,
Magnetic resonance imaging and aneu- iliac arterial stenoses: a feasibility study. Wilkoff BL. Safe performance of mag-
rysm clips. J Neurosurg 1989; 70:426 – Radiology 2001; 219:527–534. netic resonance imaging on five patients
431. 165. Marshall MW, Teitelbaum GP, Kim HS, with permanent cardiac pacemakers.
148. Kato Y, Sano H, Katada K, et al. Effects of et al. Ferromagnetism and magnetic res- Pacing Clin Electrophysiol 1996; 19:913–
new titanium cerebral aneurysm clips onance artifacts of platinum emboliza- 919.
on MRI and CT images. Minim Invasive tion microcoils. Cardiovasc Intervent 181. Duru F, Luechinger R, Scheidegger MB,
Neurosurg 1996; 39:82– 85. Radiol 1991; 14:163–166. Luscher TF, Boesiger P, Candinas R. Pac-
149. Lawton MT, Heiserman JE, Prendergast 166. Rutledge JM, Vick GW, Mullins CE, ing in magnetic resonance imaging en-
VC, Zabramski JM, Spetzler RF. Tita- Grifka RG. Safety of magnetic resonance vironment: clinical and technical con-
nium aneurysm clips. III. Clinical appli- immediately following Palmaz stent im- siderations on compatibility. Eur Heart J
cation in 16 patients with subarachnoid plant: a report of three cases. Catheter 2001; 22:113–124.
hemorrhage. Neurosurgery 1996; 38: Cardiovasc Interv 2001; 53:519 –523. 182. Pavlicek W, Geisinger M, Castle L, et al.
1170 –1175. 167. Buecker A, Neuerburg JM, Adam GB, et The effects of nuclear magnetic reso-
150. Piepgras A, Guckel F, Weik T, Schmiedek al. Real-time MR fluoroscopy for MR- nance on patients with cardiac pace-
P. Titanium aneurysm clips and their guided iliac artery stent placement. J makers. Radiology 1983; 147:149 –153.
advantages in diagnostic imaging. Ra- Magn Reson Imaging 2000; 12:616 – 183. Luechinger R, Duru F, Scheidegger MB,
diologe 1995; 35:830 – 833. 622. Boesiger P, Candinas R. Force and
151. Wichmann W, Von Ammon K, Fink U, 168. Spuentrup E, Ruebben A, Schaeffter T, torque effects of a 1.5 tesla MRI scanner
Weik T, Yasargil GM. Aneurysm clips Manning WJ, Gunther RW, Buecker A. on cardiac pacemakers and ICDs. Pacing
made of titanium: characteristics and ar- Magnetic resonance-guided coronary ar- Clin Electrophysiol 2001; 24:199 –205.
Cardiac pacemakers and implantable nium. In: Kanal E, ed. Syllabus: special tion with implanted electrodes. J Neuro-
cardioverter defibrillators are unaffected cross-specialty categorical course in di- surg 1999; 90:583–590.
by operation of extremity MR imaging agnostic radiology—practical MR safety 194. diPierro CG, Francel PC, Jackson TR, Ka-
system. AJR Am J Roentgenol 1999; 172: considerations for physicians, physi- miryo T, Laws ER Jr. Optimizing accuracy
165–170. cists, and technologists. Oak Brook, Ill: in magnetic resonance imaging-guided
186. Inbar S, Larson J, Burt T, Mafee M, Ezri Radiological Society of North America, stereotaxis: a technique with validation
MD. Case report: nuclear magnetic res- 2001; 69 –76. based on the anterior commissure-pos-
onance imaging in a patient with a 190. Gimbel JR. Interference in implanted terior commissure line. J Neurosurg
pacemaker. Am J Med Sci 1993; 305:174 – cardiac devices (letter). Pacing Clin Elec- 1999; 90:94 –100.
175. trophysiol 2003; 26(4 pt 1):922–923. 195. Rezai AR, Phillips M, Baker K, et al.
187. Garcia-Bolao I, Albaladejo V, Benito A, 191. Martin ET, Coman JA, Shellock FG, Pull- Neurostimulation system used for deep
Alegria E, Zubieta J. Magnetic resonance ing CC, Fair R, Jenkins K. Magnetic res- brain stimulation (DBS): MR safety is-
imaging in a patient with a dual cham- onance imaging and cardiac pacemaker sues and implications of failing to fol-
ber pacemaker. Acta Cardiol 1998; 53: safety at 1.5-tesla. J Am Coll Cardiol low guidelines. Invest Radiol 2004; 39:
33–35. 2004; 43:1315–1324. 300 –303.
188. Sommer T, Vahlhaus C, Lauck G, et al. 192. Limousin P, Krack P, Pollak P, et al. Elec-
MR imaging and cardiac pacemakers: in trical stimulation of the subthalamic