Professional Documents
Culture Documents
Cardio Aid 360B Manual de Servicio
Cardio Aid 360B Manual de Servicio
Cardio Aid 360B Manual de Servicio
®
Cardio-Aid 360-B
It has been prooved about the STAR biphasic waveform during clinical researches made
according to FDA guidelines, that it is one of the most effective waveforms for stopping
ventricular fibrillation.
STAR (BTE) biphasic waveform characteristics: the period of first phase is depending on
patient impedance, 3 – 11 ms, the period of second phase is fixed 3,2ms. The amplitude of the
signal depends on the set energy and on patient impedance.
2.2. Classifications
The CA360-B defibrillator is classified as II.b. according to the rule no. 9 described in the
point III./3.1 of the supplement no. IX of the 93/42/EEC board directive amended by
Directive 2007/47/EC.
From the aspect of electromagnetic compatibility the classification of the CA360-B
defibrillator, according to the EN 55011 standard is: group no. 1, class B.
2.7. Battery
Type NiMH battery 3000mAh
2.12. Pacer
Mode „Demand” (VVI) or „Fixed” (VOO)
Pulse type Rectangular, constant current
Pulse duration 40 ms ± 5%
Pulse amplitude 20 – 200 mA ± 5%.
Pulse amplitude step ± 1 mA
Pace rhythm 30 – 200 bpm ± 5%.
Pace rhythm step ± 1 bpm
Output voltage Max. 250V
2.15. Printing
Printing mode Manual or automatic
Number of simultaneously printable 1
curves
Printing time Selectable: 15, 20, 30s or 2 minutes
Speed 25 mm/s
Paper Thermal, non-rastered 58mm x 30m roll
Printer head resolution In the direction of amplitude: 8 dots/mm
On time axis: 8 dots/mm
Preprinting of curves 8s
U [V]
U1
U2
Phase 1
t3
t1 t2 T [ms]
Phase 2
-U3
-U2
Clean the defibrillator and the cables with soft cloth or sponge, not with aggressive detergent.
During cleaning, make sure that no liquid should enter into the connectors of the device. Pay
special attention when cleaning the tube of the blood pressure measurement cuff, because if
any water gets into the tube, entering inside the device during the operation of the blood
pressure measurement device, it can damage the valve and the compressor.
For cleaning you can use alcoholic solution, ammonium solution with water, peroxide
solution or any non-aggressive domestic detergents dropped on a dampened sponge.
Warning!
Even if the defibrillator is out of use, it still has to be connected to the mains, once
in every 3 month, to avoid complete run down of the battery.
The lifetime of the battery, in case of normal use, is minimum 2 years. During operation and
with time the battery capacity reduces, under a certain critical level the defibrillator indicates
the necessity of battery replacement.
The CA 360 B is dismantled in steps braking the unit down into its major
components.
Note! The necessary precautions against Electro Static Discharge (ESD) must be
taken. This applies to all handling of electrical components as well as printed
circuit boards and all service involving dismantling the CA 360 B.
Note! Before starting dismantle operation you have to shoot down the device,
disconnect power cable and wait at least 2 minutes. Before handling High
Voltage parts such as HV-capacitor, HV-circuit etc., ensure that there is no
voltage using a high voltage probe (1000:1). High-voltage parts are identified
with the “Danger High-voltage” symbol. Be very careful when connecting the
high voltage probe- avoid touching the terminals and use proper insulation If
voltage is present, leave the unit until the HV-capacitor is fully drained. If the
capacitor remains charged it can be discharged by connecting a drain resistor
e.g. 100kohm, 17W, directly over the HV capacitor terminals for min. 30 sec.
Be very cautious when connecting the resistor and use proper insulation. When
storing an HV-capacitor, always leave it in a short-circuit state. Re-assembling
the CA 360 B taking care that the various rubber seals are seated correctly.
Tools needed:
• Torx screwdriver kit.
• Hexagon screwdriver kit
• Phillips screwdriver kit.
• Wire cutter
Note! Do not remove the smaller screw holding the metal plate on the Paddle holders
Note! The screws are secured and can not fall out.
Note! To avoid damaging the front panel, place the CA 360 B on a soft fabric or
something similar.
5. Gently lift up the rear cover assembly until the connected printer cable
(optional) do not scretch.
Printer cable
2. Carefully remove the front assembly from the black internal chassis until
connected cables do not stretch.
3. Gently disconnect the keyboard cable, and the internal ECG cable.
keyboard
cable
internal ECG
cable
3. To romove the NIBP module carefully dislock the four pcb supports and
lift the NIBP module while disconnect the electrical plug.
4. To remove the ECG board, unscrew the four screws in the board.
pcb supports
Assembled
connector rack
NIBP
module
ECG
module
5. Remove the nine screws marked with arrows to dismantle the keyboard
circuit.
6. To remove the LCD, unscrew the four hexagon nut on the LCD fixing rail,
lift the rails and remove the LCD.
1. To remove the power supply module, unscrew the seven screws on power
supply circuit board, marked with arrows.
2. To remove the battery connector board, remove the one screw as shown,
and carefully disconnect the electrical plug on the board
3. To remove the DEFI test board unscrew the marked bolt, and pull out the
board from it’s slot.
Note ! The defi board and the power supply board are connected to each other with a
rigid multipin socket and plug. During remove operation you have to carefully
disconnect this socket and plug.
Defi test board and
fixing screw
5.
HV
capacitors
Battery connector
board and fixing
screw
Rigid
connection to
defi board
Note ! The defi board and the power supply board are connected to each other with a
rigid multipin socket and plug. If the power supply board is not removed before
this step you have to carefully disconnect this socket and plug to remove defi
board, see steps above!
Please send this servicelog directory in compressed form with a note of the failure time and
description to Innomed Medical. The Innomed Medical will give the solution in very short
time.
To read out the servicelog file the first step is to enter the service menu.
This sdata directory shall be zipped and sent to Innomed with fault description.
Before deleting error flags, save the service data and error flags!
• Enter configuration menu and then protected service menu (See 7.1 )
• Select "Clear Flag" in Calibrations Menu/Defibrillator Calibration Menu within
Service Settings menu by OK^. Now the color of text Off changes green. Push the
] key, then the green text changes to On. Press OK^ which deletes the error flag on
the defi board.
• Select "Clear error flags" in "Calibrations Menu/Battery Calibration Menu" within
Service Settings menu by OK^. Now the color of text Off changes green. Push the
] key, then the green text changes to On. Press OK^ which deletes error flag on the
power board.
• Leave the Service settings menu, switch off the defibrillator, and switch it on again.
• Check whether red flashing ceases.
Important! Switch the Online data to ’Off’ state before You leave this menu point.
Enter the software update menu and connect the pendrive to the USB-A connector.
Select Software Packet Update with OK^.
The device copies the software from the pendrive. Do not touch the defibrillator until
the process ends and the device automatically enters the configuration menu.
Select Software Update Menu with OK^ again.
Select Load Licence File with OK^.
The device copies the licence file from the pendrive. The process ends if “Loading
state: Done” message appears in the last line of the window.
Quit this menu with a key
In Software Update Menu upgrade the software where “BAD” message can be read by
pushing OK^ key.
Defi Panel Software update starts, it is indicated by Download bar.
During downloading please ignore error messages on the display. Do not touch the
defibrillator in this time.
At the end of download the service menu appears on screen.
Select Software Update Menu again and the next software item indicated by BAD.
Continue the process until the four BAD messages disappear.
At the end of Power Panel Update, “Please wait for the switching off …” message
appears. It takes approximately 15 seconds.
After switching off, remove the pendrive then turn the device on.
Enter configuration menu.
Set the timer.
Switch the device off.
At the next switching the defibrillator will work on the new software.
8.8. DEVICE ID
This is only an information about the device identification number, this is not a selectable
menu item.