No. MPDCT0290EA
2
MPDCT0290EA
3
• SUREStartTM: • Image reconstruction
– Continuous scan time: Max. 100 s – Number of images: Max. 4 images/scan
– Region of interest – Image interval: Reconstruction is possible in
(ROI): Max. 3 ROIs increments of 0.1 s.
– CT number • Reconstruction time: Min. 0.5 s
measurement interval: 0.083 s (12 measurements/s) • Real-time reconstruction
– Scan start delay time: Min. 3 s time: 12 images/s (0.083 s/image)
– Display function: Mean CT number and elapsed (1 slice, 512 × 512 matrix)
time
• Specification of X-ray generation
reconstruction position: By entering the couch-top posi- • X-ray exposure: Continuous
tion or using the scanogram • X-ray tube voltage: 80, 100, 120, and 135 kV
• Reconstruction method: TCOT reconstruction (applicable • X-ray tube current: 10 mA to 300 mA (260 mA for
to 16-row acquisition data) 135 kV) (in increments of 10 mA)
MUSCOT reconstruction (applic- • X-ray tube heat capacity:4.0 MHU
able to 4-row acquisition data) • X-ray tube cooling rate: Max. 864 kHU/min
• Reconstruction mode: Full image • Focal spot size
Half image – IEC 60336: 1993
Detail image nominal : 0.9 mm × 0.7 mm (small)
• The relationships between the slice thickness and the 1.4 mm × 1.4 mm (large)
imaging area scanned in 7.5 s (0.75 s, 10 rotations) in
the longitudinal direction are shown for helical pitches X-ray detection
10 and 20. • Detection system: Solid-state detectors
• Main detector: 800 channels × 28 elements
Setting slice Helical pitch* • Data acquisition: 800 channels × 16 rows
thickness 10 20 • Reference detector: 1 set
0.5 mm 50 mm 100 mm • View rate: Max. 1,200 views/s
1 mm 100 mm 200 mm
Data processing
*: The couch-top traveling distance per rotation is shown rela- • Reconstruction matrix: 512 × 512
tive to the slice thickness. • Picture element (pixel) size
– CT image: Unit: mm
Dynamic scan Scan field SS S M L LL
• Scan time: 0.75, 1, 1.5 s/360° * to * to * to * to * to
• Programmable time: Max. 1 hour Pixel size
0.35 0.47 0.63 0.76 0.98
This refers to the maximum time
within which a series of scans is *: Depending on the Vari-Area or Zoom factor
performed following a predeter-
– Scanogram: Unit: mm
mined eXam Plan.
• Number of Enlargement ratio (area) Standard
programmable scans: Max. 10 LL 4.00
Max. time of one continuous L 2.00
scan is 100 s. Pixel size
M 1.0
• Scan plan
S 0.50
– Scan interval: Min. interval is 1 s
Setting is possible in increments
of 0.1 s in a scan interval of • Reconstruction filter functions
more than 1 s. – Functions for the abdomen with BHC
Note: When a scanning mode with patient couch movement is
– Functions for the abdomen without BHC
used, the minimum scan interval is limited by the time – Functions for the brain with BHC
required for movement. – Functions for the brain without BHC
• Scan start delay time: Min. 0.5 s – Functions for the inner ear and bone
Setting is possible in increments – Functions for the lung
of 0.1 s. – Functions for high-resolution mode
• Scan rate: Max. 133 scans/100 s – Functions for super-resolution mode for the inner ear,
(0.75-s scan, 133 rotations) bone, and lung
– Functions for maintenance
4
MPDCT0290EA
5
• System management IMAGE QUALITY
– Warm-up
– Calibration data acquisition • Noise
– Picture retouching – Standard deviation: Less than 0.5%
– Operating environment setting function – Scan parameters
– Examination record · Tube voltage: 120 kV
• Display of exposure · Tube current: 260 mA
dose: CTDIvol (or CTDIw) /DLP/ · Scan time: 1.5 s
Geometric eff. · Reconstruction
function: FC70
3D color image processing · Slice thickness: 10 mm
High-quality 3D images can be obtained rapidly by easy · Scan field: S
operation. – Phantom: φ24 cm water
• 3D surface rendering • Spatial resolution: 14.5 lp/cm at MTF 2%
– Clipping, texture or non-texture 18 lp/cm at cutoff
• 3D volume rendering (reference)
– Maximum intensity projection (Max-IP) 8.0 lp/cm at MTF 50%
– Minimum intensity projection (Min-IP) (MTF calculation value)
– X-ray volume rendering – Scan parameters
– Intensity volume rendering · Tube voltage: 120 kV
– Shaded volume rendering (an arbitrary opacity curve · Tube current: 200 mA
can be set) · Scan time: 1s
• Display/processing function · Slice thickness: 2 mm
Zooming, panning, measurement (distance, angle), · Scan field: S
annotation, cutting, drilling · Reconstruction
• Cine display function: FC90
• MPR – Phantom: IRIS QA phantom
3 orthogonal planes/oblique image • High contrast resolution
Curved MPR X-Y plane
• Easy accurate bone elimination function – High-resolution
• High resolution mode mode (FC90): 0.35 ± 0.05 mm
– Standard mode
Image transfer/conversion (FC30): 0.55 ± 0.05 mm
• 1000BASE-T, 100BASE-TX, 10BASE-T – Scan parameters
• Toshiba protocol · Tube voltage: 120 kV
• DICOM storage SCU · Tube current: 200 mA
• TIFF conversion · Scan time: 0.75 s
· Slice thickness: 2 mm
Filming · Scan field: S
• Ethernet: Toshiba protocol – Phantom: Toshiba standard
DICOM PRINT SCU
• Sheet editing function using virtual film
• T-mode: Related information items such
as the patient name are dis-
played in the footer area using a
larger font.
Note: To use T-mode, the laser imager must support
2048 pixels x 2404 pixels for a 1 x 1 frame.
• Auto filming in eXam Plans
6
MPDCT0290EA
X-ray generator
This unit supplies stable high voltage to the X-ray tube unit.
The high-frequency inverter method is employed, resulting
in a light and compact design. This unit is incorporated in
the gantry.
• Max. power: 42 kW
X-ray tube
This is a large-capacity, high-cooling-rate X-ray tube that is
able to withstand continuous operation as in helical scan-
ning.
• Heat capacity: 4 MHU
• Cooling rate: Max. 864 kHU/min
Patient couch
The patient couch is positioned in front of the gantry and
supports the patient. The entire unit moves vertically and
the top moves longitudinally. In an emergency, the couch-
top can be pulled out manually with very little effort. The
couch-top can also be lowered to a minimum height of
approx. 310 mm from the floor, facilitating transfer of the
patient from a low bed or stretcher.
7
Console • Real-time helical reconstruction mode makes it possible
The console is provided with a hybrid keyboard, a monitor, to observe the images being scanned in real time at a
and a mouse. maximum at 12 frames per second. This mode shows
• Functions for scanning any shift in the slice position in real time and helps the
– Selection of scan parameters operator to check the scan field on the image, the con-
– Scanoscope control trast study timing, the presence of patient body motion,
– Scan control etc. The patient can therefore be released immediately
– Remote control of couch-top movement after scanning.
– Remote control of gantry tilt • The SUREStart function allows the operator to start helical
• Functions for image processing scanning at the timing of maximum enhancement in con-
– Window level and window width adjustment trast studies. SUREStart monitors the scan from the start of
– Other mouse-operated image processing functions a contrast study at a certain slice position while measur-
ing the changes in CT number on the image being dis-
OPERATING FEATURES played in real time. When the contrast reaches the
predefined threshold, helical scan automatically starts.
Patient handling and positioning This technique ensures optimal contrast enhancement,
• The couch-top can be lowered to approx. 310 mm (at independent of individual differences in blood flow
the center of the couch top) from the floor, making it easi- speed, and at the same time minimizes the dose of con-
er to transfer the patient to and from a bed or stretcher. trast medium.
• Alignment lights are provided in the gantry aperture for
fast and accurate patient positioning. Data processing
• High-precision couch-top positioning is possible from the • Rapid reconstruction increases the reconstruction speed
integrated console or by manual operation from the con- to 0.5 second in 0.75-second scans. (Scan & View
trol panel and clear digital readouts are provided on the mode)
gantry. • A variety of reconstruction algorithms are available and
• The couch-top can be pulled out manually in an emer- can be selected according to the anatomical region to
gency. be examined and the clinical objective of the study.
These include algorithms for the abdomen, head, bone,
Scanning lung, small structures, soft tissues, etc.
• Toshiba's scanoscope function provides a projection
image of the patient for high-precision advance planning Image display and processing
of the slice positions. • Reconstructed images are automatically displayed
• The longitudinal length of the scanning field for the according to the window settings preset in the eXam
scanogram can be adjusted up to 1,750 mm (1,450 mm Plan.
for the short patient couch version). Real-time helical • The window save function allows the user to store an
permits the scan to be aborted at any time. This mini- image with window settings different from the ones set in
mizes the patient exposure dose. the eXam Plan.
• The auto index function allows automatic incremental • Filter parameters can be customized through simple on-
couch-top movement based on the slice positions deter- screen menu selections. These parameters include the
mined through the scanogram. number of filtering passes, matrix size, and filter coeffi-
• The eXam Plan function allows simple selection of pre- cients.
programmed scanning parameters for routine examina- • Images can be rotated and reversed either right/left,
tions, maximizing patient throughput. top/bottom, or black/white.
• Protocol comments can be saved in each eXamPlan • The Multi-frame feature allows up to 16 images to be
provideng interactive onscreen instructions for all studies retrieved and displayed simultaneously on the screen.
reducing the need to refer to a separate protocol book. • The three-dimensional image display function allows
• The Vari-area function allows the user to pre-select a color three-dimensional and MPR images in real-time to
region of interest for zooming using raw data, permitting be generated from the volumetric scan data acquired by
immediate post-scan analysis. Zooming using raw data helical scanning. This results in higher definition and
yields higher resolution than enlarging an image that has image quality than images reconstructed from conven-
already been reconstructed. tional single-slice scanning. This is because helical
• Dynamic and rapid sequence scan modes are provided. scanning provides superior data continuity along the
• Multislice helical scan acquires raw data by rotating the patient axis compared with conventional scanning.
X-ray tube continuously while moving the patient continu-
ously through the scanner. The volume data acquired
can be used to reconstruct slices at any desired axial
positions. This scan mode is best used for rapid patient
scanning during a single breath-hold and for high-defini-
tion three-dimensional and MPR imaging.
8
MPDCT0290EA
Grounding
Grounding must be provided in accordance with local
regulations for medically used electrical equipment.
9
Power distribution board Room layout example
In case of In case of
3-phase, 200 V 3-phase, 400 V
Ground resistance:
(1,194*)
As per applicable
1,494
legal requirements.
NFB NFB
150 A 100 A
1
Ground bar
System System
5,530 (4,930*)
transformer
(option)
7,130 (6,530*)
2
Ambient conditions
Temperature Humidity Heat generation
Scan room
Gantry 20°C to 26°C 40% to 80% Approx.
Tolerance: ±2°C No 9.720 kJ/h (*1)
condensation 32,070 kJ/h (*2)
10
MPDCT0290EA
Breaker box
12
System
transformer*
Patient couch
17 17 Gantry
17
Console
9 (CPU cabinet 2)
(Scan room)
(Operator's room)
Console
Speaker
(CPU cabinet 1)
*: When line voltage is 200 V ±10 V, the system transformer is not required. Unit: m
11
OUTLINE DRAWINGS
12
MPDCT0290EA
OUTLINE DRAWINGS
900
(35.4)
344 556
(13.5) (21.9)
(TILT ANGLE)
(75.2)
1,910
2,070
(81.5)
Gantry
13
OUTLINE DRAWINGS
(18.5)
(24.8)
470
630
2,190 2,187
(86.2) (86.1)
14
MPDCT0290EA
OUTLINE DRAWINGS
(18.5)
(24.8)
470
630
1,890 1,887
(74.4) (74.3)
15
MPDCT0290EA
OUTLINE DRAWINGS
574
(22.6)
414
(16.3)
(19.7)
500
(27.6)
(27.6)
700
700
Console
http://www.toshibamedicalsystems.com
©Toshiba Medical Systems Corporation 2007 all rights reserved.
Design and specifications subject to change without notice.
"Made for Life" is a trademark of Toshiba Medical Systems Corporation.
Activion, SUREFluoro, SUREExposure, SUREStart, SUREPlaque, and Boost3D are trademarks of Toshiba Medical Systems Corporation meets Toshiba Medical Systems Corporation Nasu Works meets
internationally recognized standards for Quality the Environmental Management System standard, ISO 14001.
Toshiba Medical Systems Corporation. Management System ISO 9001, ISO 13485.
This document may include trademarks and registered trademarks of other companies. Printed in Japan
Model TSX-031A 2007-04 TME/KI