You are on page 1of 4

Medical Devicelink Page 1 of 4

Print t

Originally Published MDDI M


Cover Story: Pumps & Valves
Declaring War on Wear in Pumps and Valves
Manufacturers and users can take steps to lengthen the lives of air- and fluid-moving components

William Leventon

Pumps and valves do vital work in many different medical devices and manufacturing operations. Of
course, their work wears them down—and eventually wears them out. But in many cases, they’re able
to fend off wear and tear for many years and millions of cycles.
What’s being done to extend the lives of these crucial components? Using innovative designs and
hardy materials, manufacturers are equipping pumps and valves for lengthy battles against wear and
tear. For their part, users are promoting component longevity by making the work that pumps and The small s
valves do less damaging and less stressful. And when they work together, manufacturers and users of valve
pumps and valves can come up with long-lasting solutions for even the most wearing fluid- and air- Hannifin re
moving problems. travel d
mov
Wear-Fighting Valve Designs
Ask valve manufacturers how they minimize product wear and you get a variety of answers. O
technique is to shorten the travel distance of moving parts. Inside the miniature valves made b
Hannifin’s Pneutronics Div. (Hollis, NH), parts usually move no more than 0.02 in., reports ma
development manager Mark Garland.
Another wear-reducing strategy is to use simple designs. For example, most of the dispensing
made by I&J Fisnar Inc. (Fair Lawn, NJ) have only two or three moving parts, says Vladimir Si
president of the company. These products are primarily used in device manufacturing operatio
In addition, some valve makers try to minimize the number of parts that come in contact with t
medium flowing through the valve. These wetted parts are particularly prone to wear, accordin
Victoria, application development manager for EFD Inc. (Lincoln, RI), which also makes dispe
valves used in medical device manufacturing. In EFD’s diaphragm valves, Victoria notes, only
This valve, by I&J diaphragm and sealing head come in contact with the fluid being dispensed.
Fisnar Inc., applies
seals to housings of Like these EFD valves, Parker Hannifin’s Liquid X valve has only two wetted parts. Pneutronic
medical devices. designed this miniature valve so that the diaphragm protects the stainless-steel actuator from
Designing valves with medium. The diaphragm isolation design helps the valve last for 10 million cycles, according t
only one or two Garland.
moving parts can
reduce wear. Some of EFD’s valves include packings that separate the fluid and air chambers. Made of soft
material, the packings form a tight seal around the piston to prevent leakage from one chamber to the other. Teflon i
compatible with more chemicals than most plastics, Victoria adds, lessening the risk of deterioration when the mater
exposed to fluids in dispensing operations.
Wear can also be minimized by a valve design featuring tight tolerances and carefully balanced loads, says Ralph B
electrofluidic systems product manager for The Lee Co., a valve and pump manufacturer in Westbrook, CT. Valves t
lot of force to open and close can overcome high levels of friction, so the design does not have to be highly engineer
explains. But valves wear faster when excessive force is used to drive the valve seat into
the elastomer seal. “It’s like hitting a nail with a sledgehammer,” he says.

http://www.devicelink.com/grabber.php3?URL=http://www.devicelink.com/mddi/archive/... 7/13/2006
Medical Devicelink Page 2 of 4

The Right Material


A key tool in the battle against wear is material selection. Manufacturers can use a variety of metals,
plastics, and rubbers to make valve components. In part, the choice depends on which material holds
up best when exposed to a particular medium. For example, Siroky notes, manufacturers might choose
ultra-high-molecular-weight plastics for the diaphragms of valves used to dispense aggressive
cyanoacrylates, or so-called crazy glues.
Elec
Or an application may involve anaerobic fluids, which crystallize or harden when they come in contact comm
with metal ions. Valves used to dispense such fluids must have wetted parts made of plastic, Siroky moto
says. Neub
prov
Then there are UV resins, which are used in the manufacture of a number of medical devices. control, all
According to Siroky, these resins cause a reaction if they contact aluminum, so they’re often dispensed pumps to b
using valves with stainless-steel bodies. up to spe

Polycarbonate is a popular choice for many medical device components. But lipids take a toll on many polycarbonate
formulations, warns Jim Pisula, vice president of marketing for Value Plastics Inc. (Fort Collins, CO), which makes p
tubing connectors. So for applications involving lipids, valve manufacturers must be sure to use a lipid-resistant vers
polycarbonate. Another alternative, Pisula notes, is to eschew polycarbonates in favor of acrylics or other materials
characterized by native resistance to lipids.
For many valves, leakage is a common failure mode. Valve leakage can often be traced to the seal, says Buck. So t
lengthen valve life, manufacturers can opt for a durable seal material. For example, they might select Viton rather tha
silicone.
Manufacturers can also boost seal durability with a coating. Halkey-Roberts Corp. (St. Petersburg, FL) makes valves
into sterilizers for colonoscopes. These valves are exposed to peracetic acid, which makes silicone seals sticky and
over time, explains Lew Lecceardone, the company’s vice president of medical marketing. So the seals are coated w
special oil that protects them from peracetic acid. This helps to keep the valves operating efficiently in sterilizers for y
Some EFD valves also have coated components. For example, Victoria notes, the company reduces corrosion insid
spool valve by coating wetted parts with titanium. Similarly, Garland says, Pneutronics uses passivation treatments t
corrosion of stainless-steel valve components.
In addition to valves, some suppliers offer extras that help OEMs reduce wear in fluid- and air-carrying systems. For
EFD provides feed lines and fittings made of materials that are compatible with customers’ media. The company has
introduced a device that controls the temperature inside dispensing equipment to minimize fluctuation of fluid viscos
According to Victoria, the ProcessMate 6500 reduces the frequency of manual valve adjustments to compensate for
changes, thereby reducing valve damage and wear caused by operator handling of the components.
Pump Makers Battle Wear
Like valve manufacturers, pump makers use a variety of techniques to battle wear. Inside WO
pumps made by Rietschle Thomas (Sheboygan, WI), the longer the stroke, the more a pump’s
connecting rod will wobble. This makes the pump work harder, which increases wear and tear
piston seal, explains David Droege, the company’s miniature-products support manager. Mos
KNF Neuberger Inc.
uses stainless-steel
manufactured at the firm’s Sheboygan facility are custom made. So, depending on the applica
or aluminum heads to company could design a WOB-L pump with a shorter stroke length to reduce wear, Droege sa
make corrosion-
resistant pumps. At KNF Neuberger Inc., based in Trenton, NJ, pump diaphragms are designed using finite-ele
analysis. The computer program “helps us develop the most robust design possible,” says Da
Vanderbeck, the company’s business development manager. For example, Vanderbeck says, the program showed d
where to add structural elements on the underside of a diaphragm to boost strength without too much of an effect on
flexibility of the component.
Reinforced diaphragms are also used in miniature pumps made by Parker Hannifin’s T Squared Pumps unit (Fairfiel
some cases, the firm molds webbing or matting into rubber diaphragms to boost their resistance to tensile loads. In h
pressure applications, the reinforced elastomer diaphragms last longer than a standard diaphragm would, notes Len
engineering manager at T Squared Pumps. But in low-pressure applications, he adds, a standard diaphragm might l
as long as a reinforced one and also operate more efficiently because it would produce less internal friction.
Besides reinforcing diaphragms when necessary, Prais and his colleagues try to eliminate restrictions in their pumps
cause parasitic losses. These losses can add loads to the diaphragm and other components, thereby shortening pum
According to Prais, one way to reduce harmful restrictions is to use special valve designs that minimize backpressur
the pumps.

http://www.devicelink.com/grabber.php3?URL=http://www.devicelink.com/mddi/archive/... 7/13/2006
Medical Devicelink Page 3 of 4

In many cases, pump life can also be lengthened by effective cooling. This is true in oxygen concentrators that use R
Thomas WOB-L pumps. Enclosed in a case, these WOB-L units are cooled by two fans, one on either side of the pu
application is different in how and where the pump is mounted, as well as in the amount of room available to move a
the pump. So Rietschle Thomas engineers work with each OEM to produce a cooling design that will maintain the pr
ambient temperature around the pump, Droege says.
Material Considerations
In addition, Droege notes, the company’s engineers pay attention to the material used to make piston
seals in the pumps. The seals include a Teflon substrate that can be supplemented with proprietary
polymer materials that lessen wear and boost seal life.
To make pump heads, KNF uses a variety of materials, with the choice depending in part on the
medium. Stainless steel is fine for many applications, but others might call for a plastic material. KNF
has had good luck with Ryton, an expensive plastic that offers good chemical compatibility,
Rietschle Th
Vanderbeck says. work wit
design a p
According to Droege, heat generated by pumps produces moisture that will quickly degrade aluminum this
pump heads and valve plates. So many manufacturers treat these parts with dichromate. However, diap
Teflon-impregnated parts last longer in high-humidity environments than those treated with dichromate, compressor
Droege explains. So sometimes the company impregnates metal components with Teflon. Some firms maintain t
also use Teflon to help preserve pump diaphragms. For example, diaphragm pumps are used in temperatur
diagnostic equipment that handles many different reagents. To prevent these reagents from attacking
rubber diaphragms, Vanderbeck says, the elastomeric components can be coated with a protective Teflon layer.
The User’s Role
Broadly speaking, there are two major factors that affect the life span of pumps and valves: how they’re made and h
they’re used. No matter how well a component is made, its life will be short if users don’t do their part to minimize we
tear.
For starters, Victoria says, users should read all manuals and instructions that come with a component and follow th
manufacturer’s recommendations. This may seem obvious, but many users simply don’t do it, Victoria maintains. Fa
follow instructions “is probably one of the major reasons for wear and tear in our equipment,” he says. In many cases
notes, rapid and excessive wear is caused by easily avoidable mistakes such as incorrect setup and inappropriate u
Some users who don’t read and follow instructions make disastrous mistakes even when they are trying to do the rig
For instance, Victoria points to people who try to clean a valve used to dispense cyanoacrylate by flushing it with wa
Instead of cleaning the valve, Victoria says, flushing with water causes the glue to harden inside the component.
Although improper maintenance can lead to disaster, proper maintenance can be crucial in extending component life
silicone-dispensing systems, for example, small amounts of material may get cured inside a valve. Eventually, Siroky
this buildup of cured material will impede the action of the valve. So he recommends that users periodically take the
apart, clean it, and make sure all the components are moving freely before reassembling it.
As for pneumatic systems, air containing rust or dirt particles will eventually break down valve seals. To prevent such
problems, Siroky recommends filtration of air flowing into pneumatic components. Filters should be installed at sever
along a pneumatic line, including the point just before air flows into a valve.
According to Garland, filtration is particularly important in systems that include miniature valves because of the vulne
the small components and sealing surfaces in these valves. Generally speaking, Pneutronics recommends a minimu
filtration level at the source. But Garland says that a micron rating of 10 or less is even better at reducing the risk of
contamination, which is one of the main causes of valve failure. On the downside, he notes, a 10-µm filter may have
significantly larger than its 40-µm counterpart to prevent unacceptable flow restriction in a system.
In liquid-moving systems, debris can get trapped between the valve seat and seal, Buck says. To prevent this and th
resulting valve leakage, Lee provides last-chance safety screens that filter liquids flowing into valve components. Be
harming valves, Buck says, liquid-borne debris can scratch the seal or piston inside a pump, causing leakage. So us
also filter the liquid flowing into pumps. The filters must be changed regularly to prevent dirt buildup on them that will
the load on the pump, Vanderbeck notes.
Another way pump users can reduce wear and tear is to pay careful attention to mounting. For example, Droege say
Rietschle Thomas pumps should only be mounted in a horizontal plane. Vertical mounting of these pumps adversely
the motion of the internal components, which can result in damage that will shorten pump life.
In many cases, pumps are mounted on shock mounts that lessen the vibration of the unit. According to Droege, the
shock mounts a user selects can have a major effect on how much a pump will shake while in operation. So Rietsch

http://www.devicelink.com/grabber.php3?URL=http://www.devicelink.com/mddi/archive/... 7/13/2006
Medical Devicelink Page 4 of 4

Thomas helps customers through the process of determining which shock mounts are right for their applications.
System Issues
To minimize pump and valve wear and extend component life, users should look beyond the
components themselves and consider other issues. For instance, what type of fluid is in the sy
it’s prone to crystallization, it’s important not to let the system dry out, Buck warns. Drying can
the formation of crystals that lodge in components and damage them. In pinch valves, for exam
crystals can make small cuts in the rubber, creating weak spots that will be prone to failure, Bu
To prevent damaging crystallization, Buck recommends that OEMs either keep fluids in these
at all times or flush them out with water or some kind of cleaning solution following each use. S
can remain dry once the crystal-precipitating liquids are cleaned out of them, he says. But don
the mistake of trying to clean out a system with air. This will evaporate the liquid, but leave be
mischief-making crystals.
Temperature control Another issue to consider is how an air- or fluid-moving system is powered. According to Vand
units, like the
ProcessMate from
one option worth considering is voltage shaping. As the name implies, this involves shaping th
EFD Inc., can reduce of the voltage applied to the motor. Voltage shaping requires a logic-control device that varies
valve wear and tear supplied voltage, which in turn changes the speed of the pump.
by keeping a material
at optimum Instead of turning on the system voltage all at once, Vanderbeck notes, voltage shaping allow
dispensing ramp up the voltage from zero to a maximum operating level, so pumps can be brought up to
temperature. slowly. Voltage shaping also allows control of pump speed so that output matches the require
the system in a particular situation. “If you don’t need the full output of the pump, you can slow it down so you’re just
out what the system requires—no more,” he explains.
Since the pump isn’t constantly running at full speed, the pump and other components are subjected to less pressure
flow, thereby reducing wear and tear throughout the system. As a result, Vanderbeck says, voltage shaping can resu
“tremendous increase” in component life.

William Leventon is a freelance writer who contributes frequently to MD&DI.

Copyright ©2006 Medical Device & Diagnostic Industry


Comments about this article?
Post them in our Members' Discussion Forums.

Copyright © 2005 Canon Communications, LLC About Us | Customer Care | Advertising Information | FAQ | Site Guide

http://www.devicelink.com/grabber.php3?URL=http://www.devicelink.com/mddi/archive/... 7/13/2006

You might also like