Model CD500-98
Service check
Introduction:
The 1-Channel is a analyzer system which needs only less service and maintenance actions.
So the following checklist will give suggested preventive checks for a proper function.
According the national requirements we suggest to do a service check once a year / every two years.
Service check
Type: Serial-No.:
Service check
Remarks:
Preparation:
Switch Analyzer ON
Adjustment:
Example:
Adjustment ok.
Adjustment:
Mixer spinning
Introduction:
For a save function a proper function of the mixer motor and the mixer is required.
Check procedure:
**1) It is required to use a new cuvette with a new mixer to get a objective visual picture of the
spinning mixer.
9
Preparation:
Switch on analyzer.
Select softwarefunction <calib.>
In service- or UTILITIES menu.
Refer to chapter “software structure” or
chapter “Service mode”
is displayed.
<calib.>
Int 37.4
Calibration
ext 37.4
1. Wait until the temperature has stabilized to 37.4 on the int display. (int = internal)
2. Press Enter key.
3. Read the actual temperature of the thermal block from the thermometer
4. Key in this value via the numerical keyboard.
It will be displayed in the ext field automatically. (ext = external)
The decimal point can be entered by pressing the „>„ key.
Press Enter to confirm. The calibration value will be set in the analyzer
automatically and will be displayed in the int line.
5. Wait again until the temperature has stabilized to 37,4 in the int line.
6. If the value in the internal line has stabilized to 37,4 it must be compared to the
digital multimeter. If the values deviate repeat this procedure from item 2 above.
If both values are identical finish calibration and save by pressing the ESC key.
Remove probe
Select:
<calib.>
Press Enter
Wait until:
is displayed
Example:
Check temperature at
external sensor
Is external No
Temp = 37,4°C Press Enter key
Yes
Key in:
ext 36.1
value.
3 6 > 1
int 36.1
is displayed
End of procedure:
Hardware requirements:
Software requirements:
Hardware explanations:
*1)
USB to Serial converter does not work.
*2)
For example:
Delock 61622 or 61603
www.delock.de / www.delock.com
*2)
For example:
Delock 66211 or 66217
9 www.delock.de / www.delock.com
*2)
For example:
Delock 89003
www.delock.de / www.delock.com
or:
Longshine LCS-6021
www.longshine.de
Software update
Preparation:
Programming-cable
to Serial-COM-interface of computer
Programming cable
Create folder
Software
CD500.MX
FLASH.BAT
FLASH.CMD
FLASH2.BAT
FLASH2.CMD
LOADER.EXE
Into folder
Update process:
Start
FLASH.BAT
or
FLASH2.BAT
By doubleclick
BOOTMODE
screen, then.
Furter steps:
Error messages:
Possible Errormessage
Parameter update
Preparation:
Data-cable
to Serial-COM-interface of computer
Data cable
Select <UTILITI>
Press Mode key
Key in PIN Code (11111 = default)
Select <printr>
Press Enter key
Select < OFF >
Press Enter key
Select <comput>
Press Enter key
Select < OFF >
Press Enter key
Create folder
Parameter
9
Select folder and copy
CD05_MIN.EXE
DEFAULT.C5A
Into folder
Update process:
CD05_MIN.EXE
By doubleclick
1 = COM 1
2 = COM 2
3 = COM 3
4 = COM 4
“Demo-Para” = example
9
Press F2 – key at computers keyboard to
transmit setup data to the analyzer.
Further steps:
Error messages:
Possible Errormessage
= no answer
Possible Errormessage
= communication error
Parameter protocol
Introduction:
The parameter protocol printout is an overview about present analyzer´s setup (example).
Printout (example):
==========================
==========================
PARAMETER-PROTOCOL
( 980 Bytes)
1-channel
V2.14, 06/07/06 System-Family
math-vers V02.99 Software Version and Release Date
Incubation = 60 sec.
incubat. = 60 s
9
start-reagent:
Lot 1 Startreagentvolume 100µl (displayed)
reagent = 100 ul
100.0 % 11.6 s
50.0 % 17.7 s Conversion curve
25.0 % 29.9 s
10.0 % 66.6 s
Printer connection
Introduction:
Introduction:
Slide the power switch to ON while pressing the ON/OFFLINE/PAPER END button.
Red and green lamps blink alternately.
Release the button after a list of current settings starts printing out.
Dip SW-1
1 (OFF) : Input = Serial
2 (ON ) : Printing Speed = High
3 (ON ) : Auto Loading = ON
4 (OFF) : Auto LF = OFF
5 (ON ) : Setting Command = Enable
6 (OFF) : Printing
7 (ON ) : Density
8 (ON ) : = 100 %
Dip SW-2
1 (ON ) : Printing Colums = 40
2 (ON ) : User Font Back-up = ON
3 (ON ) : Character Select = Normal
4 (ON ) : Zero = Normal
5 (ON ) : International
6 (ON ) : Cahracter
7 (ON ) : Set
8 (OFF) : = U.S.A.
Dip SW-3
1 (ON ) : Data Length = 8 bits
2 (ON ) : Parity Setting = No
3 (ON ) : Parity Condition = Odd
4 (OFF) : Busy Control = XON/XOFF
9 5
6
(OFF)
(ON )
:
:
Baud
Rate
7 (ON ) : Select
8 (ON ) : = 9600 bps
Note:
For changing the setting of the Dip Switches, please refer to the printers user´s manual.
Switch analyzer ON
Switch Printer ON
select <UTILIT>
press Mode Key
select <printr>
press Enter Key
select <AUTO>
press Enter Key
This section describes how to change the settings of the Dip switches of the Thermal Printer DPU-414
with the use of the buttons ONLINE and FEED
Set the status of a Dip switch to “ON” by pressing ONLINE button or to “OFF” by pressing FEED button
as indicated in the tables below.
After the ONLINE button has been pressed the GREEN lamp indicates that the button operation has
been confirmed. When the FEED button has been pressed, the button operation is confirmed by the
RED lamp.
Alternately blinking of both lamps indicates that the new settings are written to memory.
WARNING: Never turn the printer OFF while it is writing new settings to memory.
After each Dip switch setting via button, the status of the actual Dip switch is printed out automatically.
Wait until “DIP SW setting complete!!” is printed out and the printer returns to the ONLINE mode
before turn off the power.
Procedure
2 Slide the power switch ON RED and GREEN Starting to print out
while pressing ONLINE lights blinking
button alternately
3 Release the button after a
list of current settings starts
printing out
3 The print out of the current
settings is followed by:
Continue ?: Push “On-line-SW”
Write ?: Push “Paper feed SW”
4 Move to chart Dip SW-1
Next page
DIP SW -1
DIP SW -2
DIP SW -3
LIS connection
Introduction:
Preparation:
9600 baud
8 bit
no parity
1 stop bit
no handshake
=> 9600, 8, N, 1
select <UTILIT>
press Mode Key
select <printr>
press Enter Key
9 select <AUTO>
press Enter Key
Example:
mode SINGLE
start-reagent:
Lot 1 Lot-No startreagent
reagent = 100 ul startreagent volume
--------------------------
PT Method: PT
patient ____________
21.03.11, 15:38:04 Patient: <ID> **1)
time = 14.3 s Date time: dd.mm.yy hh:mm:ss
INR = 1.25 Result time: 14.3 sec (example)
% = 69.3 INR: 1.25 (example)
% 69.3 (example)
Measurement result
PT
patient ____________
Method:
Patient:
PT
<ID> **1)
9
21.03.11, 15:45:13
time = 12.0 s Date time: dd.mm.yy hh:mm:ss
INR = 1.04 Result time: 12.0 sec (example)
% = 93.8 INR: 1.04 (example)
% 93.8 (example)
Service assistance
Important
Guidelines for getting service assistance
Contact
Contact
10
10
For getting service assistance please note that we need at least a minimum of important informations.
1. Customer identification
Customer
Name
Adress
Email
Telephone
Fax
Distributor / Enduser
2. Analyzer
Softwareversion
Display Messages
Printouts
Screenshots
Photos
Note:
Supporting our service department with the basic informations as detailed as possible will help us to 10
give you the needed assistance.
Contact
LABiTec
LAbor BioMedical Technologies GmbH
An der Strusbek 6
22926 Ahrensburg, Germany
Email: service@labitec.de
www.labitec.de or www.labitec.com
10
Vers. 0 - Rev. 0
Service Manual 1-Channel
Model CD500-98
Vers. 0 - Rev. 0