Service Manual 1-Channel

Model CD500-98

Service check

Introduction:

The 1-Channel is a analyzer system which needs only less service and maintenance actions.

So the following checklist will give suggested preventive checks for a proper function.

According the national requirements we suggest to do a service check once a year / every two years.

Vers. 0 - Rev. 0 9.5


Service Manual 1-Channel
Model CD500-98

9.6 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Service check

Type: Serial-No.:

Software Version: Parameter-ID.:

Pos. OK nOK done


1 Identification   
Check the rating plate on the bottom site (Type, Serial number)  --- ---
write down in protocol
2 Known Problems
Are there any known problems reported by the user?   
3 Visual control
Is the outside of the analyzer in good conditions?   
Are there any parts to be replaced?   
Is the measuring channel clean 
(free from samples, reagents, particles)
Is the light protection cap clean? can it move freely?   
Is the ext. power supply in good condition?   
Is the cable free from damages?   
Does the DC plug fit into DC socket of analyzer   
4 Parameterprotocol
Print parameter protocol w. external printer, check printout  --- ---
Does type and serial number of printout match to the rating plate   
Actual date and time on printout   
Software version   
Parameter-ID   
5 Replacements
Replace Li-Battery after 4 years (± 1 year)   
6 Technical check (use service menu)
Optic test <optic >   
Mixer test <mixer > / mixer spinning   
Keyboard test <keyboa>   
Check / Calibrate thermal block temperature (37,4°C ±0,4°C)   
7 Analytical check
Run QC-Test with aPTT in double determination with
Control plasma N and Control plasma P
   9

Service Check: Passed  Not passed 

Date Name Sign

Vers. 0 - Rev. 0 9.7


Service Manual 1-Channel
Model CD500-98

Service check

Pos. Quantity Order-No. Description

Remarks:

9.8 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Adjust LCD contrast

Preparation:

Remove screws from bottom housing.

And remove bottom housing.

Switch Analyzer ON

Wait until Analyzer has warmed up and


System is ready to use.

Vers. 0 - Rev. 0 9.9


Service Manual 1-Channel
Model CD500-98

Adjust LCD contrast contd.

Adjustment:

Find potentiometer from backside whole of


PCB.

Use small screwdriver for adjusting the


contrast.

Example:

Adjustment ok.

Goal is, to find the best combination


between good visible characters with less
grey shadows in the background.

9.10 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Speed sensor distance

Adjustment:

For proper speed sensor function make


sure that the space between magnet and
≥ 3 mm speed sensor is ≥ 3mm.

≤ 3mm speed sensor malfunction


3 – 4 mm optimal distance
≥ 4mm speed sensor malfunction

Vers. 0 - Rev. 0 9.11


Service Manual 1-Channel
Model CD500-98

9.12 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Mixer spinning

Introduction:

For a save function a proper function of the mixer motor and the mixer is required.

Check procedure:

Prepare a new cuvette with 200µl distilled


water. **1)

Make sure that a new mixer is inside.

Make sure that there are no air bubbles


inside.

Start analyzer in service mode

Select <mixer >

Start motor with 200 rpm

Put prepared cuvette into measuring


channel.

Make a visual check:

does the mixer spin in the center of the


cuvette

does the spinning mixer seems to be a


star like 

in case it seems to be a triangle


like 
then check or replace the magnet / mixer
motor unit.

**1) It is required to use a new cuvette with a new mixer to get a objective visual picture of the
spinning mixer.
9

Vers. 0 - Rev. 0 9.13


Service Manual 1-Channel
Model CD500-98

9.14 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Calibrate thermal block temperature

Preparation:

For preparation prefill a defined cuvette


position (1) of incubation block ¾ with
distilled water.

Put the distilled water direct into the


block. Do NOT use any cuvette therefore.
1

Place external temperature sensor


(probe) in the water filled position.

Switch on analyzer.
Select softwarefunction <calib.>
In service- or UTILITIES menu.
Refer to chapter “software structure” or
chapter “Service mode”

Press arrow key until


9
<calib.>

is displayed.

Vers. 0 - Rev. 0 9.15


Service Manual 1-Channel
Model CD500-98

Calibrate thermal block temperature Contd.

<calib.>

Pressing Enter key will give access to this function

Int 37.4

Calibration

ext 37.4

1. Wait until the temperature has stabilized to 37.4 on the int display. (int = internal)
2. Press Enter key.
3. Read the actual temperature of the thermal block from the thermometer
4. Key in this value via the numerical keyboard.
It will be displayed in the ext field automatically. (ext = external)
The decimal point can be entered by pressing the „>„ key.
Press Enter to confirm. The calibration value will be set in the analyzer
automatically and will be displayed in the int line.
5. Wait again until the temperature has stabilized to 37,4 in the int line.

6. If the value in the internal line has stabilized to 37,4 it must be compared to the
digital multimeter. If the values deviate repeat this procedure from item 2 above.
If both values are identical finish calibration and save by pressing the ESC key.

For detailed procedure see flow chart next page.

When calibration process finished:

Remove probe

Remove distilled water from incubation


position.

9.16 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Calibrate thermal block temperature Contd.

Start of procedure: Switch ON analyzer

Call Service menu by


keying in Service Code

Select:
<calib.>

Press Enter

Wait until:

is displayed
Example:

Check temperature at
external sensor

Is external No
Temp = 37,4°C Press Enter key

Yes

Press ESC key ext 37.4


is displayed

Key in:
ext 36.1
value.

Press ESC key There for press:

3 6 > 1

Press Enter key

int 36.1
is displayed

End of procedure:

Vers. 0 - Rev. 0 9.17


Service Manual 1-Channel
Model CD500-98

9.18 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Requirements for updating

Hardware requirements:

Desktop Computer or Notebook


RS232c serial interface COM1 or COM2 or COM3 or COM 4 *1) *2)
Data cable

Software requirements:

Operating System *3)


MS-DOS
Windows 95
Windows 98
Windows ME
Windows 2000
Windows XP
Uploading Software CD5_MIN.EXE and DEFAULT.C5A

*3) Windows Vista and Windows 7 will not work

Vers. 0 - Rev. 0 9.19


Service Manual 1-Channel
Model CD500-98

Requirements for updating contd.

Hardware explanations:

*1)
USB to Serial converter does not work.

If a Notebook does not support serial interface converter should be used:

*2)

PCMCIA to Serial Converter


Cardbus to Serial Converter

For example:
Delock 61622 or 61603

www.delock.de / www.delock.com

Also compatible devices could be used.

*2)

Express Card to Serial Converter

For example:
Delock 66211 or 66217
9 www.delock.de / www.delock.com

Also compatible devices could be used.

9.20 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Requirements for updating contd.

If a Desktop computer does not support serial interface should be used:

*2)

Serial interface card

For example:
Delock 89003

www.delock.de / www.delock.com

or:
Longshine LCS-6021
www.longshine.de

Also compatible devices could be used.

Vers. 0 - Rev. 0 9.21


Service Manual 1-Channel
Model CD500-98

9.22 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Software update

Preparation:

Connect 9pin SUB-D connector from

Programming-cable

to Serial-COM-interface of computer

Connect 6pin Mini-DIN connector from

Programming cable

to serial interface connector of analyzer.

Switch ON analyzer by supporting 12VDC


power as usual.

The analyzer will start in Boot-Mode

By displaying BOOTMODE the analyzer


is ready for receiving update file.

Vers. 0 - Rev. 0 9.23


Service Manual 1-Channel
Model CD500-98

Software update contd.

Create folder

Software

In root of your C:\ drive (HDD)

Select folder and copy

CD500.MX
FLASH.BAT
FLASH.CMD
FLASH2.BAT
FLASH2.CMD
LOADER.EXE

Into folder

9.24 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Software update contd.

Update process:

Depending on COM interface Port

Start
FLASH.BAT

or

FLASH2.BAT

By doubleclick

The uploading software will start in a


DOS-Box automatically.

The main screen of the uploading


programm appears.

The new softwareversionD will be shown

The present analyzer with Boot-Loader


Version will be shown.

While updating process an address


counter will increase.

The analyzer will indicate the datatransfer


by a flashing * - Symbol

Vers. 0 - Rev. 0 9.25


Service Manual 1-Channel
Model CD500-98

Software update contd.

The software update will be finished after


163 seconds.

Depending on computers operating


system taking a longer time may be
possible.

Press any key to close window.

The analyzer will return to

BOOTMODE

screen, then.

Furter steps:

- switch OFF analyzer


- remove serial cable connection
- switch ON analyzer

- load default parameters if required


- install valid parameters, then.
Refer to chapter “Parameter update”

9.26 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Software update contd.

Error messages:

Possible Errormessage

Boot-Loader meldet sich nicht!


Bitte RS232-Verbindung prüfen.

= no answer from Boot-Loader

- check COM Port


- check cable connection
- check BOOTMODE in display

Press any key to close window.

Vers. 0 - Rev. 0 9.27


Service Manual 1-Channel
Model CD500-98

9.28 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Parameter update

Preparation:

Connect 9pin SUB-D connector from

Data-cable

to Serial-COM-interface of computer

Connect 6pin Mini-DIN connector from

Data cable

to serial interface connector of analyzer.

Switch ON analyzer by supporting 12VDC


power as usual.

Vers. 0 - Rev. 0 9.29


Service Manual 1-Channel
Model CD500-98

Parameter update contd.

A correct setup of analyzers serial interface


is required.

Select <UTILITI>
Press Mode key
Key in PIN Code (11111 = default)

Select <printr>
Press Enter key
Select < OFF >
Press Enter key

Press Enter = more parameters

Select <comput>
Press Enter key
Select < OFF >
Press Enter key

Press ESC = go to work

Select <1 TST > / <1 PT >

Connect Analyzer with Computer

Use DATA-Cable 105-93-170-00

Create folder

Parameter

In root of your C:\ drive (HDD)

9
Select folder and copy

CD05_MIN.EXE
DEFAULT.C5A

Into folder

9.30 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Parameter update contd.

Update process:

Select and start file

CD05_MIN.EXE

By doubleclick

The uploading software will start in a


DOS-Box automatically.

Select Serial Interface Port by pressing


key:

1 = COM 1
2 = COM 2
3 = COM 3
4 = COM 4

The software will read the parameter file


from harddisk.

The main screen of the uploading


programm appears.

In Parameter-ID will be shown the


specific file-number.

Verify this number at first.

“Demo-Para” = example
9
Press F2 – key at computers keyboard to
transmit setup data to the analyzer.

Vers. 0 - Rev. 0 9.31


Service Manual 1-Channel
Model CD500-98

Parameter update contd.

Sende zum Geraet…

Indicates the transfer process.

While datatransfer the following Messages


on analyzer display will appear:

If transfer was successful no error


message will appear and counter has
increased.

Further steps:

- remove serial cable connection


- check / calibrate thermalblock temperature
- enter calibration values matching to
reagent lot

9.32 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Parameter update contd.

Error messages:

Possible Errormessage

!!! keine Antwort !!!

= no answer

- check Analyzer Setup


- check COM Port
- check cable connection

Possible Errormessage

!!! Kommunikationsfehler !!!

= communication error

- check Analyzer Setup


- check COM Port
- check cable connection

Vers. 0 - Rev. 0 9.33


Service Manual 1-Channel
Model CD500-98

9.34 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Parameter protocol

Introduction:

The parameter protocol printout is an overview about present analyzer´s setup (example).

Printout (example):

==========================
==========================
PARAMETER-PROTOCOL
( 980 Bytes)

1-channel
V2.14, 06/07/06 System-Family
math-vers V02.99 Software Version and Release Date

actual date & time:


22.03.11, 10:53:38 Printout date & time
[dd.mm.yy, hh:mm:ss]
--- device-specific: -----
ser.no. A1234567 Instrument serial no.
secret no. = 11111 Secret no (PIN code)
ntc_soll = 606 Calibrated target NTC ( xxx = 37,4°C)
--- global: --------------
Analyzer Instrumentname
distributor Manufacturer / distributor
parameter-ID: Demo-Para Parameter-ID (Demo-Para = example)
printer OFF Present printer setup
computer OFF Present computer setup
header OFF
beeper ON
cuv detection ON Ins. cuvette into channel auto. detected
method 1 PT Actual method PT
--- method store 1 -------
PT Test Setup for PT
cuv remove detection OFF Removing cuvette not tetected
automatically, keypress reqired = default
filt.no. = 0

mode SINGLE Single determination selected

Incubation = 60 sec.
incubat. = 60 s
9
start-reagent:
Lot 1 Startreagentvolume 100µl (displayed)
reagent = 100 ul

1st convers INTERPOLAT. Conversion Referencecurve selected


2nd convers INR Conversion INR selected
ISI = 1.05

100.0 % 11.6 s
50.0 % 17.7 s Conversion curve
25.0 % 29.9 s
10.0 % 66.6 s

Vers. 0 - Rev. 0 9.35


Service Manual 1-Channel
Model CD500-98

Parameter protocol contd.

--- method store 2 -------


aPTT Test Setup for aPTT
--- method store 3 -------
Fib.1 Test Setup for Fib.1 = g/l
--- method store 4 -------
Fib.2 Test Setup for Fib.2 = mg/dl
--- method store 5 -------
Thrmb Test Setup for Thrombin Time
--- method store 6 -------
Intr. Test Setup for Intrinsic Factor
--- method store 7 -------
Extr. Test Setup for Extrinsic Factor
--- I/O: ----------------- Present system values
ntc = 764 ntc < NTC_soll = heating up
photo = 4 photo (sensor) value
rpm = 0 mixermotor speed
==========================
==========================

9.36 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Printer connection

Introduction:

The 1-Channel can be connected to an


external serial thermal printer.

LABiTec suggests to use a Seiko Epson


DPU-414 thermal printer.

In some countries the DPU 414 Printer and


it´s power supply have to be ordered
separately.

Contact your local distributor for details.

To run the DPU-414 printer with a


1-Channel 504 analyzer a special printer
cable will be needed.

Please refer to chapter 4.

Vers. 0 - Rev. 0 9.37


Service Manual 1-Channel
Model CD500-98

Printer connection contd.

Introduction:

Before using the printer with the 1-Channel


the serial interface of the printer has to be
configurated.

DPU 414 printer setup 9600 baud

Print a list of current setting:

Slide the power switch to ON while pressing the ON/OFFLINE/PAPER END button.
Red and green lamps blink alternately.
Release the button after a list of current settings starts printing out.

[ DIP SW setting mode ]

Dip SW-1
1 (OFF) : Input = Serial
2 (ON ) : Printing Speed = High
3 (ON ) : Auto Loading = ON
4 (OFF) : Auto LF = OFF
5 (ON ) : Setting Command = Enable
6 (OFF) : Printing
7 (ON ) : Density
8 (ON ) : = 100 %
Dip SW-2
1 (ON ) : Printing Colums = 40
2 (ON ) : User Font Back-up = ON
3 (ON ) : Character Select = Normal
4 (ON ) : Zero = Normal
5 (ON ) : International
6 (ON ) : Cahracter
7 (ON ) : Set
8 (OFF) : = U.S.A.
Dip SW-3
1 (ON ) : Data Length = 8 bits
2 (ON ) : Parity Setting = No
3 (ON ) : Parity Condition = Odd
4 (OFF) : Busy Control = XON/XOFF
9 5
6
(OFF)
(ON )
:
:
Baud
Rate
7 (ON ) : Select
8 (ON ) : = 9600 bps

Note:

For changing the setting of the Dip Switches, please refer to the printers user´s manual.

9.38 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Printer connection contd.

Connect the printers cable at the serial


interface port of the 1-Channel analyzer

Switch analyzer ON

Connect the printers cable at the serial


interface port of the DPU-414 printer.

Switch Printer ON

The printout at the analyzer has to be


enabled at first.

start up the analyzer as usual

select <UTILIT>
press Mode Key

key in security no (default is 11111)

select <printr>
press Enter Key
select <AUTO>
press Enter Key

press ESC Key (= work)

Vers. 0 - Rev. 0 9.39


Service Manual 1-Channel
Model CD500-98

9.40 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Printer configuration of Thermal Printer DPU-414

This section describes how to change the settings of the Dip switches of the Thermal Printer DPU-414
with the use of the buttons ONLINE and FEED

Button: ONLINE Dip Switch = ON


Button: FEED Dip Switch = OFF

Set the status of a Dip switch to “ON” by pressing ONLINE button or to “OFF” by pressing FEED button
as indicated in the tables below.

After the ONLINE button has been pressed the GREEN lamp indicates that the button operation has
been confirmed. When the FEED button has been pressed, the button operation is confirmed by the
RED lamp.

Alternately blinking of both lamps indicates that the new settings are written to memory.

WARNING: Never turn the printer OFF while it is writing new settings to memory.

After each Dip switch setting via button, the status of the actual Dip switch is printed out automatically.

Wait until “DIP SW setting complete!!” is printed out and the printer returns to the ONLINE mode
before turn off the power.

Vers. 0 - Rev. 0 9.41


Service Manual 1-Channel
Model CD500-98

Printer configuration contd.

Procedure

Step Action Lamps Printer prints:


1 Slide the power switch OFF

2 Slide the power switch ON RED and GREEN Starting to print out
while pressing ONLINE lights blinking
button alternately
3 Release the button after a
list of current settings starts
printing out
3 The print out of the current
settings is followed by:
Continue ?: Push “On-line-SW”
Write ?: Push “Paper feed SW”
4 Move to chart Dip SW-1
Next page

9.42 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Printer configuration contd.

DIP SW -1

Step Action Lamps Printer prints:


5 Press RED lamp lights Dip Switches-1
ONLINE
button.

6 Press GREEN lamp lights 1 (OFF): Input = Serial


FEED
button
7 Press RED lamp lights 2 (ON): Printing Speed = High
ONLINE
button.

8 Press RED lamp lights 3 (ON): Auto Loading = ON


ONLINE
button.

9 Press GREEN lamp lights 4 (OFF): Auto LF = OFF


FEED
button.

10 Press RED lamp lights 5 (ON): Setting Command =


ONLINE Enable
button.

11 Press GREEN lamp lights 6 (OFF): Printing


FEED
Button
12 Press RED lamp lights 7 (ON): Density
ONLINE
button.

13 Press RED lamp lights 8 (ON): 100%


ONLINE
button.

Continue ?: Push “On-line-SW”


Write ?: Push “Paper feed SW”
14 Press
ONLINE
button.
9
15 Move to chart Dip SW-2
Next page

Vers. 0 - Rev. 0 9.43


Service Manual 1-Channel
Model CD500-98

DPU 414 printer setup contd.

DIP SW -2

Step Action Lamps Printer prints:


With Dip Switches-2
showing on the printer paper
16 Press RED lamp lights 1 (ON): Printing Colums = 40
ONLINE
button.

17 Press RED lamp lights 2 (ON): User Font Back-up =


ONLINE ON
button.

18 Press RED lamp lights 3 (ON): Character Select =


ONLINE Normal
button.

19 Press RED lamp lights 4 (ON): Zero =Normal


ONLINE
button.

20 Press RED lamp lights 5 (ON): International


ONLINE
button.

21 Press RED lamp lights 6 (ON): Character


ONLINE
button.

22 Press RED lamp lights 7 (ON): Set


ONLINE
button.

23 Press GREEN lamp lights 8 (OFF): = USA


FEED
button
Continue ?: Push “On-line-SW”
Write ?: Push “Paper feed SW”
24 Press
ONLINE
button.
9 25 Move to chart Dip SW-3
Next page

9.44 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

DPU 414 printer setup contd.

DIP SW -3

Step Action Lamps Printer prints:


With Dip Switches-3
showing on the printer paper
26 Press RED lamp lights 1 (ON): Data Length = 8 bits
ONLINE
button.

27 Press RED lamp lights 2 (ON): Parity Setting = No


ONLINE
button.

28 Press RED lamp lights 3 (ON): Parity Condition =


ONLINE odd
button.

29 Press RED lamp lights 4 (ON): Busy Control =


ONLINE XON/XOFF
button.

30 Press GREEN lamp lights 5 (OFF): Baud


FEED
button.

31 Press RED lamp lights 6 (ON): Rate


ONLINE
button.

32 Press RED lamp lights 7 (ON): Select


ONLINE
button.

33 Press RED lamp lights 8 (ON): = 9600 bps


ONLINE
button.

Continue ?: Push “On-line-SW”


Write ?: Push “Paper feed SW”
34 Press RED and GREEN
FEED lights blinking
button alternately
Dip Switch settings complete! 9
GREEN lamp lights
permanent
35 Slide the power switch OFF

Vers. 0 - Rev. 0 9.45


Service Manual 1-Channel
Model CD500-98

9.46 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

LIS connection

Introduction:

The 1-Channel analyzer supports no direct


LIS interface.

In case a LIS communication is needed,


the result data could sent out by the serial
printer interface to a computer instead to a
printer.

Vers. 0 - Rev. 0 9.47


Service Manual 1-Channel
Model CD500-98

LIS connection contd.

Preparation:

Connect the 1-channel analyzer with a


Computer system by using the RS232
interface and a data cable.

Please refer to chapter 4.

The interface parameters of the 1-


Channel analyzer are fixed:

9600 baud
8 bit
no parity
1 stop bit
no handshake

=> 9600, 8, N, 1

Enable the dataoutput at the analyzer at


first.

start up the analyzer as usual

select <UTILIT>
press Mode Key

key in security no (default is 11111)

select <printr>
press Enter Key
9 select <AUTO>
press Enter Key

press ESC Key (= work)

The analyzer will send out the data in


ASCII format after the measurement
result appeared once a time.

The LIS system has to pick up this data by


interrupt or polling.

9.48 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

LIS connection contd.

Example:

Method (test) header


Informations about test setup and
conversion parameters

--- method store 1 ------- Method Store: 1


PT Method: PT
actual date 21.03.11 Actual date: dd.mm.yy
cuv remove detection OFF

mode SINGLE

incubat. = 60 s Incubation time: 60sec.

start-reagent:
Lot 1 Lot-No startreagent
reagent = 100 ul startreagent volume

1st convers INTERPOLAT. st


1 conversion = interpolation (enabled)
2nd convers INR nd
2 conversion = INR (enabled)
ISI = 1.05 ISI = 1.05
100.0 % 11.6 s
50.0 % 17.7 s 4 point calibration curve
25.0 % 29.9 s
10.0 % 66.6 s

--------------------------

results: Measurement result

PT Method: PT
patient ____________
21.03.11, 15:38:04 Patient: <ID> **1)
time = 14.3 s Date time: dd.mm.yy hh:mm:ss
INR = 1.25 Result time: 14.3 sec (example)
% = 69.3 INR: 1.25 (example)
% 69.3 (example)
Measurement result

PT
patient ____________
Method:
Patient:
PT
<ID> **1)
9
21.03.11, 15:45:13
time = 12.0 s Date time: dd.mm.yy hh:mm:ss
INR = 1.04 Result time: 12.0 sec (example)
% = 93.8 INR: 1.04 (example)
% 93.8 (example)

**1) <ID must entered in the


LIS manually>

Vers. 0 - Rev. 0 9.49


Service Manual 1-Channel
Model CD500-98

9.50 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Service assistance

Important
Guidelines for getting service assistance

Contact
Contact

10

Vers. 0 - Rev. 0 10.1


Service Manual 1-Channel
Model CD500-98

10

10.2 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Guidelines for getting service assistance

For getting service assistance please note that we need at least a minimum of important informations.

1. Customer identification

Customer
Name
Adress
Email
Telephone
Fax
Distributor / Enduser

2. Analyzer

Type of analyzer Type:


Serial-Number SN:
Reference number REF

Softwareversion

3. Discribe the problem as detailed as possible

Display Messages
Printouts
Screenshots
Photos

May clarify the situation.

4 In case of analytical problems further informations will be needed

Test type eg. PT, aPTT, Fibrinogen, ….


Sample (Patient sample, Control N, Control P, dilution, ….)
Reagent Manufacturer, Name of reagent, Order number, Lot-No, date of expire
Results rated results, actual measured results, raw value in seconds, converted values
Conversion Conversion table, ISI, …

Describe the problem as detailed as possible.

Note:

Supporting our service department with the basic informations as detailed as possible will help us to 10
give you the needed assistance.

Thank you in advance.

Vers. 0 - Rev. 0 10.3


Service Manual 1-Channel
Model CD500-98

Contact

In case of needed service assistance please contact us at:

LABiTec
LAbor BioMedical Technologies GmbH
An der Strusbek 6
22926 Ahrensburg, Germany

Email: service@labitec.de

Phone: +49 4102 4795 0


Fax: +49 4102 4795 35

www.labitec.de or www.labitec.com

10

10.4 Vers. 0 - Rev. 0


Service Manual 1-Channel
Model CD500-98

Document release notes

Revision Date Note


Rev. 0 October, 08 1st release of Service manual
2015 biobas 10

Vers. 0 - Rev. 0
Service Manual 1-Channel
Model CD500-98

Vers. 0 - Rev. 0