Professional Documents
Culture Documents
4072-An113 LPN01736
4072-An113 LPN01736
Pressurizable Analytical
Reservoir Column
Chamber
Sample
Loading Valve
Waste 11738
1/4-28
1/4-28
Eluent IN
Gas Separator
Waste Tube
12284
Figure 4 AutoSuppression™ external water mode using the pressurized water delivery system.
ganisms, and particulate matter larger than 0.2 µm. Soak For the best performance at trace levels it is critical
containers for at least 24 hours with deionized water and that baseline noise be kept to a minimum. From startup, a
rinse several times prior to use. Wear disposable gloves (for system typically requires about 5 hours to establish a stable
cleanroom electronics applications) when handling ap- background conductivity. For this reason it is best to keep
paratus that makes contact with eluent, standard, or samples. the system running continuously. Ensure that there is an
The total baseline shift between the beginning and the adequate supply of the eluents, external water, and waste
end of the gradient analysis should be no more than 1 µS. containers to allow the system to run unattended. All
To ensure this, sodium hydroxide eluents should be pre- moving parts of the gradient pump need to be in good
pared with minimal carbonate contamination. In addition, working order (check valves, piston seals, and pistons).
the ATC should be regenerated at the beginning and end of The DX 500 instrumentation and PeakNet software
each operational day to remove any contaminants that may provides the analyst with the ability to monitor baseline
have collected on it. This can be readily done by pumping noise. In the diagnostic menu under “pressure statistics”,
at least 30 mL of 100 mM sodium hydroxide (eluent no. 3) the GP40 Gradient Pump displays the measured pressure in
through the ATC. psi from the pressure transducer as well as the pressurization
point values from the left and right pistons. In a 1-minute
segment, the pump pressure reading difference should be
Peaks: less than 20 psi and the pressure point difference should
1. Acetate 0.50 µg/L (ppb) be 5 units or less. In the PeakNet Optimize program, the
2. Formate 0.20
3. Carbonate — Auto Threshold function lets the user select a baseline
3 4. Sulfate 0.10 region for noise measurement. In a representative 1-minute
1.0
section, an equilibrated system in good working order
should be able to consistently deliver a noise reading less
4 than 10 nS peak to peak.
µS
Prepare standards using the information in Table 1.
It is recommended to make up a 100-mL final volume of
1 2 1000-mg/L stock standards in 125-mL high-density
0.0 polyethylene HDPE containers. From this stock standard, a
0 5 10 15 20
1-mg/L dilute standard is made. Take aliquots from this
Minutes 12283 dilute standard to make working standards at the low-µg/L
(ppb) down to the high-ng/L (ppt) range.
Figure 5 A representative blank for trace analysis.
Stock standards are stable for at least 1 month when
stored in a refrigerator at 4 °C. Dilute stock standards at the Peaks:
1.0 1. Fluoride 1.4 µg/L (ppb) 7. Nitrate 1.8 µg/L (ppb)
low-mg/L (ppm) levels should be prepared fresh weekly. 2. Acetate 3.0 8. Unidentified —
Working standards at the low-µg/L (ppb) range should be 3. Formate 0.88 9. Carbonate —
4. Chloride 1.2 10. Sulfate 0.81
made fresh daily. 5. Nitrite 0.71 11. Oxalate 1.3
The IonPac AS11 gradient separation was chosen for 6. Bromide 1.6 12. Phosphate 3.2
µs 9 12
this method because of its high resolving power for inor-
ganic anions and organic acids. A representative low-level 1
1011
7
standard for 10 anions in high purity water is shown in 2 4
5
6 8
3
Figure 6. The large peak at 14 minutes is carbonate, the
size of which will vary with the amount of dissolved 0.0
elute after the void volume and resolves acetate and for-
mate as well. Strongly retained polyvalent ions such as Figure 6 Trace anion determination.
sulfate and phosphate elute later as the gradient concen-
tration rises to a final concentration of 26 mM sodium
Peaks:
hydroxide. To regenerate the system, the gradient begins — µg/L (ppb)
1. Fluoride 0.51 µg/L (ppb) 7. Unidentified
with 2 minutes of 38 mM sodium hydroxide. 2. Acetate 3.2 8. Carbonate —
1.0
This method is also applicable to high-purity power 3. Formate 1.7 9. Sulfate 1.1
4. Chloride 3.3 10. Oxalate 0.47
plant waters containing corrosion inhibitors. Figure 7 5. Nitrite 0.26 11. Unidentified —
shows a sample of 8-mg/L morpholine containing trace- 6. Nitrate 2.1
8
level anions. No significant difference in peak efficiency 4
µs 11
or retention time is observed compared to the analysis of 9
deionized water. Method detection limits were established 2 6 10
Dionex Corporation Dionex Corporation Dionex U.S. Regional Offices Dionex International Subsidiaries
1228 Titan Way Salt Lake City Technical Center Sunnyvale, CA (408) 737-8522 Belgium (015) 203800 Canada (905) 844-9650 France (1) 39 46 08 40 Germany (06126) 991-0 Italy (06) 30895454
P.O. Box 3603 1515 West 2200 South, Suite A Westmont, IL (708) 789-3660 Japan (06) 885-1213 The Netherlands (076) 57 14 800 Switzerland (062) 205 99 66 United Kingdom (01276) 691722
8 Determination of Trace Anions in High-Purity Waters
Sunnyvale, CA
94088-3603
Salt Lake City, UT
84119-1484
Smyrna, GA
Houston, TX
(770) 432-8100
(713) 847-5652
* Designed, developed, and manufactured under an NSAI registered ISO 9001 Quality System.