Professional Documents
Culture Documents
GC-MS
LC-MS/MS LC-HRMS
Rapid Test
3
Sample Preparation
4
Objectives of Sample Preparation
5
Comparison of LLE with HyperSep Verify CX SPE
6
HyperSep Silica Based SPE
7
ISQ-7000 GC-MS
Surpass competition in the GCMS market with leading performances and scalability
Column connection
(inside the oven)
GC oven
Dual Filament
assembly
Vacuum Probe
Interlock (VPI)
(Optional)
Step 1. Insert removal tool Step 2. Remove source Step 3. Hot source is held in tool Step 4. Push source out of tool
Extends the capability of the Vacuum Probe Interlock (VPI) design with
the newly introduced source plug, V-Lock
4 hours
98% 4.5 hours
87%
Time Time
98% Saving 87%
Saving
Time Time
Saving Saving
15
Technical Note 10592
• 30 μL -glucuronidase (Merck 5000 I.U.) were added • The cartridge was dried after interference elution with
to 3 mL of urine and incubated for 60 min at 56 °C. strong vacuum.
• The Thermo Scientific™ HyperSep™ Verify CX cartridge, • Elution was performed with a mixture of methanol and
6 mL/200 mg, was conditioned with 3 mL MeOH 5% ammonia solution (95:5), pH 9, two times 0.5 mL.
followed by 3 mL 0.1% formic acid. • The sample was evaporated under nitrogen until
• Urine was mixed with 3 mL of 2M acetate buffer, pH dryness at 65 °C.
4.8. Urine was checked and adjusted, where • The sample was reconstituted with 50 μL MeOH and
necessary, for accurate pH. placed in 50 μL MS certified vials.
• The sample was added to the HyperSep Verify CX
cartridge and a slight vacuum was applied to achieve,
for example, approximately one drop per second
elution rate. The vials were subsequently centrifuged for precipitating
• Interference elution was done with a mixture of 1 mL the particles before putting them in the autosampler.
water + 0.1% formic acid, total volume 3 mL, followed
by a mixture of 1 mL MeOH/water 50:50 + 0.1% formic
• acid, total volume 3 mL.
16
Technical Note 10592
• Thermo Scientific™ HyperSep™ Verify CX Cartridges, 200 mg/3 mL; 50 pack (P/N 60108-777)
• 24-port Vacuum Manifold (P/N 60104-233)
• Vacuum Pump (P/N 60104-233)
• Analyte Elution 4 mL Vial
• Thermo Scientific™ Reacti-Vap™ Evaporator (P/N TS-18826)
• Thermo Scientific™ SureStop™ MS Certified Vials, 100 μL reservoir (P/N MSCERT5000-36LVW)
• Thermo Scientific™ Marathon™ Injection Port Septa (P/N 313P3240)
• Thermo Scientific™ LinerGOLD™ GC Liners (P/N 453A-1255-UI)
• Thermo Scientific™ TRACE™ TR-DoA 35MS column, 15 m, 0.25 mm ID, 0.25 μm (P/N 26AF130P)
17
Technical Note 10592
Compound separation and detection was achieved
using a Thermo Scientific™ TRACE™ 1310 GC coupled
with a Thermo Scientific™ ISQ™ 7000 mass spectrometer
Sample introduction was performed using a Thermo
Scientific™ TriPlus™ 100 LS autosampler, injecting 1 μL into
the Thermo Scientific™ Instant Connect Split/Splitless
(SSL) injector module.
Tables 1 and 2 list the method parameters.
18
19
20
ISQ 7000 single quadrupole GC-MS system
21
Thermo Scientific GC Sampling Systems
Thermo Scientific™
Thermo Scientific™ Thermo
Scientific™ TriPlus™ RSH™
TriPlus™ 100 LS
TriPlus™ 500 Autosampler
Headspace
Autosampler
Sample Throughput
Automation
120/240 vial
configuration
Thermo Scientific™
AI/AS 1310 Series
Autosampler 12 vial TriPlus 500 HS
configuration + AI/AS 1310
22
Thermo Scientific Chromeleon CDS software
• Control more than 350 modules from Thermo Fisher Scientific™ and many
other vendors
techniques and MS variants, all using the same intuitive user interface
23
Expanded Sympathomimetic Amines by GC-MS
24
THC Metabolite in Urine by GC/MS
25
Codeine and Morphine
26
Codeine and Morphine by GC-MS
27
Ephedrine and Pseudoephedrine by GC-MS
28
Benzoylecgonine in Urine by GC-MS
29
Testing Drugs in Hair
30
Hair Sample Preparation by GC-MS (UNODC General Method)
Decontamination Clean-up
Procedure with organic solvent Treat 30-50 mg of washed hair by extraction or digestion/extraction as
1. Take a hair strand (~ 100 mg). described above. Extensively centrifuge the extracts
2. Wash with 5 ml dichloromethane for 2 min.
3. Dry with adsorbent paper. SPE clean-up
4. Wash again in 5 ml dichloromethane for 2 min. 1. Condition an SPE column with methanol (2 ml) and phosphate
5. Dry again. buffer 0.1 M, pH 6 (2 ml).
2. Transfer the supernatant hair extract to the column, taking care to
Extraction avoid transferring the precipitate.
With methanol or aqueous acids or buffered solutions. Incubation in 3. Wash with 2 ml of water.
aqueous 0.01-0.50 M HCl or in 1 M phosphate buffer at pH 6.4-7.6 4. Wash with 3 ml of 0.1 M HCl.
is usually performed at 56°C or 60°C overnight. 5. Dry the column at maximum vacuum for 10 minutes.
6. Wash with 5 ml of methanol.
Digestion 7. Elute with 2 ml CH2Cl2/Isopropanol (80:20) at 2 % of NH4OH.
With diluted NaOH, 1 M NaOH is added to hair and incubation is typically 8. Dry the organic eluate under nitrogen at 40°C; if gas
performed for one hour at 80°C, or overnight at 60°C. It is suitable only for chromatographymass spectrometry (GC-MS) is planned (see
drugs that are stable under alkaline conditions. below), derivatize by adding N-Methyl-N-
(trimethylsilyl)trifluoroacetamide (MSTFA) + 1 % of
trimethylchlorosilane (TMCS) at 75°C for 20 minutes).
31
Application Note 10493
32
Application Note 10493
33
Application Note 10493
34
Technical Note 10592
Over the course of several weeks, over 700 urine samples were analyzed
35
Technical Note 10592
36
Technical Note 10592
37
Technical Note 10592
38
Thermo Scientific HyperSep SPE Applications Example
39
GC-MS for Forensic Analysis
• Used extensively for decades in analytical forensic toxicology laboratories, “gold standard”
• All analytes eluting from column are hit with a beam of electrons at 70 eV so fragmentation pattern is
identical on all instruments
• Published in a searchable full scan GCMS libraries (NIST, SWGDRUS, AAFS, Maurer Wiley)
• Analysis performed in a gas phase
• To maximize sensitivity, targeted analysis can be conducting using SIM (selected ion monitoring)
• Straight forward, simple operation
• Sensitivity ug/mL to ng/mL
• Cheaper than LC-MS/MS or LC-HRMS
40
Comparison of GC-MS, LC-MS/MS, and LC-HRMS
41
Library (GC-MS and LC-HRMS)
LC-HRMS (Orbitrap)
GC-MS
42