1.

5T superconducting MRI Configuration


Imaging System
Magnet system
Type:Superconductive
Magnet cooling system: 4K cold head
Magnet monitoring system
RF transmitting &receiving system
Type: All-digital transmitting &receiving systems
RF amplifier
Transmitting coil: Cage quadrate transmitting
RF pre-endincluded
RF receiving coils: Phased array head coil 1pc
Phased array neck coil 1pc
Phased array body coil1pc
Phased array shoulder coil 1pc
Phased array knee coil1pc
Phased array ankle coil1pc
Phased array multi-function coil1pc
Phased array head and neck combined coil 1pc
(either choose head and neck coil each, or one piece of head and neck combined coil)
Gradient system
Gradient Amplifier
Gradient powersupply
Gradient coil
Spectrometer
Multi-channel all digital spectrometer
Computer
Computer: brand name computer
Intel Xeon processor quad core
Main frequency: ≥3.0 G
Number of CPU: ≥2
Memory: ≥16GB
Hard disk capacity: ≥1TB
Support for double-layer burning, DVD/CDRW burner.
High resolution high lighting TFTdisplayer(24')
Standard keyboard
Mouse
Doctor’s table
Designed for MRI usage
Patient control system
Two-dimensional electric tableand control system
Gating system (ECG & respiratory)
Earphoneused in a MRI room
Patient button & speaking system
Music system
Standard mattress, pillow kit
MRIsoftware
XGY MRI software: MEDViewlatest verion
Imaging Sequence Software Package
System Control Package
Data Processing Software Package
Image reconstruction package
Advanced image post-processing software package
Failure Analysis Package
Quality Control Software Package
Laser Camera Interface Package
RF shielding room(Faraday cage)
Faraday cage included
Shielding doors, shielding observation window, filter plate, waveguide plate andinterior decoration, all
included
Power supply, cooling&air conditioning
AC power supply
Air conditioningof Faraday cage room included
Water-chiller included
Accessories
Testing Phantoms
TestingRF loads
Replacing fuses
Safety signs
Documents accompanied
User manual & Maintenance manual
No. Parameters
1 Magnet
1.1 Field strength: 1.5T
1.2 Type: Superconductive (4K cold
head)
1.3 Shielding mode: active shielding
+ anti-interference
1.4 shimming mode: active
shimming + passive shimming +
dynamic shimming
1.5 Patient personalized shimming:
available
1.6 Advanced high-order shimming:
available
1.7 Automatic shimming time (3D)
≤ 3 seconds
1.8 magnet length ≤ 157cm
1.9 Magnet inner diameter ≥ 850mm
1.10 Patient bore ≥600mm
1.11 Magnet material: niobium
titanium (NbTi) alloy
1.12 Magnet weight (including liquid
helium) ≤ 4.3 tons
1.13 Magnetic field stability
≤0.04ppm/h
1.14 Homogeneity (V-RMS)
1.15 45cmDSV≤0.50 ppm (Vrms)
1.16 40cmDSV≤0.2 ppm (Vrms)
1.17 30cmDSV≤0.06 ppm (Vrms)
1.18 20cmDSV≤0.02 ppm (Vrms)
1.19 10cmDSV≤0.002 ppm (Vrms)
1.20 Liquid helium consumption rate:
zero volatilization
1.21 Liquid helium filling interval: no
need to add liquid helium
1.22 Cold head warranty time: ≥2.5
years
1.23 5 Gauss line axial range: axial ≤
4.0m
1.23 5 Gauss line radial range: radial
≤ 2.5m
2 Sub-systems on Patient
scanning
2.1 Music System in the
scanning room
2.2 Patient surrounding
systems: lighting,
ventilation, speaking &
headsets
2.3 Patient’s call button
2.4 Patient table controlling
system on both sides of
the table
2.5 Material of the table:
Fiber Reinforced
Plastics
2.6 Table length ≥210cm
2.7 Table is electronically
moveable
2.8 Table automatic
stepping available
2.9 Moving direction: Up
&Down, In & Out
2.10 Lowest position (to the
ground) ≤ 60cm
2.11 Table moving range (up
& down) ≥25cm
2.12 Table moving range (in
& out) ≥180cm
2.13 Table moving speed
max. (in & out) ≥
20mm/s
2.14 Patient moving accuracy
±0.5mm
2.15 Table emergency brake
available
2.16 Table manual
movement available
2.17 Mode of patient feet
entering first available
2.18 Video Surveillance
System checking up on
3 Gradient system
3.1 Gradient amplifier
3.2 Maximum single-axis
gradient field
strength≥35mT/m
3.3 Maximum single-axis
gradient slew rate (non-
effective) ≥134mT/m/s
3.4 Gradient rise time (98%)
≤0.27ms
3.5 Gradient control technology:
all-digital real-time control
3.6 Gradient shielding method:
active shielding
3.7 Gradient working mode: non-
resonant
3.8 Noise Reduction Mode:
Hardware Noise Reduction
3.9 Gradient cooling method:
water cooling
3.10 Gradient duty cycle: 100%
3.11 Non-linearity: 50cm × 50cm
× 40cm ellipsoid <5% (X / Y
direction) <3% (Z direction)
4 RF system and coil
4.1 Multi-channel spectrometer
≥16 channels
4.2 Resonance frequency: 63.86
MHz ± 0.5 MHz
4.3 System Type: Fully digital
real-time control of the RF
system (including digitization
of transmitting & receiving)
4.4 Transmitter and receiver
frequency range:
14~400MHz
4.5 Transmit Coil Type: Cage
Orthogonal Emission
4.6 Receive coil type: multi-
channel phased array
reception
4.7 Number of receiving
channels: ≥16 channels
4.8 Phased Array Coil Tuning:
Tuning Free
4.9 Parallel Acquisition
Technology
4.10 Type of receiving
coil (standard 7)
4.10.1 Phased array head
and neck combined
coil
4.10.2 Phased array
abdominal coil
4.10.3 Phased array
flexible body coil
4.10.4 Phased Array Knee
Coil
4.10.5 Phased shoulder
coil
4.10.6 Phased Array Ankle
Coil
4.10.7 Phased array small
joint coil
4.10.8 Phased Array Head
Coil (optional)
4.10.9 Phased array breast
coil (optional)
4.11 Preamplifier gain
30 dB
4.12 RF Amplifier
4.13 RF Power
Amplifier type:
Pulsed
4.14 Maximum output:
≥18kW
4.15 Working range:
63.86
MHz±0.5MHz
4.16 Noise figure
≤0.5dB
4.17 Fixing pad,
fastening tape, etc
for each coil
5 Computer