Instruction Manual STE 8 D 01F50600 v2.5
Instruction Manual STE 8 D 01F50600 v2.5
Prior to operating this instrument, please read the operations manual carefully and follow all
installation instructions.
IMPORTANT NOTICE:
If you can't open the door, please unlock the door according to the instruction “How to open the
door in the case of power outage” in the manual.
Need Maintenance
If this picture appears on the screen when power on or E88 appears on the
report, please call your dealer or local service maintenance. Your steam
sterilizer need a regular maintenance.
Instructions Manual
Icanclave STE-8-D
Australian Sponsor
Laboratory Equipment Pty Ltd
26-30 Farr Street
Marrickville NSW 2204 Australia
www.labec.com.au
Specifications
Electrical supply 240VAC, 50Hz
Chamber Φ170mm X 320mm
Circuit breaker F25A
Power 2200VA
Sterilization temperature 121 °C / 134 °C
Capacity of the distilled water tank Approx. 2.5l
Working temperature 5 °C / 40 °C
Working relative Humidity Max. 80%, non condensing
Overall dimensions W 420mm / H 370mm / D525mm
Net weight 34.5 kg
Max. noise Level <70 dB
Atmospheric pressure 76kPa - 106 kPa
Accesory Quantity
Tray 2
Tray rack 1
Draining hose 2
Tray holder 1
Documents
Declaration of conformity CE Autoclave 1
3 Instructions manual
Appendix 2
Holding Time
Total Time
Max. Load
Pressure
(min)
(min)
kPa
(kg)
C°
(kg)
Programs Type
Drying
- - - 1-20 - - -
(optional)
The time required for sterilizer to be ready for routine use after the power is switched on is less than 5 minutes.
The max. Temperature of the 134°C sterilization cycle is 137 °C The max.
Temperature of the 121°C sterilization cycle is 124 °C
17 Instructions manual
Solid
Instructions manual 18