Professional Documents
Culture Documents
Calculation: Total Alkalinity (T) or Phenolphthalein Alkalinity (P) will be calculated as follows
Volume (mL) of N H 2 SO 4 1000
Alkalinity, mg/L as CaCO 3 50
mL of sample
Caustic Alkalinity : Caustic alkalinity is the alkalinity corresponding to the OH - present in water
and calculated using the following Table. Carbonate and bicarbonate Alkalinity also can be
calculated using the same Table.
P=0 0 0 T
P = T/2 0 2P 0
P > T/2 2P - T 2 (T - P) 0
P=T
P 0 0
Discussion on Result:
Environmental Engineering Laboratory, Civil Engineering Department, Bengal Engineering & Science University
Determination of pH of Water
Introduction: Measurement of pH is one of the most important and frequently used tests in.
Practically every phase of water supply and wastewater treatment, e.g., acid base neutralization,
water softening, precipitation, coagulation, disinfection and corrosion control, is pH dependent. pH
is used in alkalinity and carbon dioxide measurements and many other acid–base equilibria. At a
given temperature, the intensity of the acidic or basic character of a solution is indicated by pH or
hydrogen ion activity. It is expressed on a scale ranging from 0 to 14, which is the common
logarithm of the reciprocal of the hydrogen ion concentration. Completely neutral water measures
7.0 on the scale. The values of pH lower than 7.0 indicate the hydrogen ion concentration is greater
than the hydroxyl ion concentration, and the water is termed acidic. The opposite condition is
implied when the pH exceeds 7.0, and water is termed basic.
Acidic range Alkaline range
|______________________|______________________|
pH 0 7 14
pH may be represented by
pH = log 101/[H+] or, pH = - log10 [H+]
similarly, pOH = - log 10[OH-]
where, [H+] & [OH-] are the molar concentrations of hydrogen and hydroxyl ions,
respectively.
Pure water is slightly ionized and at equilibrium the ion product is
[H+][OH-] = Kw where Kw is the ion product of water,
= 10-14 at 250C
i.e., log10[H+] + log10[OH-] = log10Kw
or, – pH –pOH = -14
or, pH + pOH = 14
or, pOH = 14 - pH
Apparatus & Reagents: pH meter and 7.0 & 4.0 or 9.2 Buffer Solutions
Procedure :
1. Connect the instrument to 230V AC main. Display shown "SYS - 361 Ready".
2. Connect the pH electrodes to back panel.
3. Press “STD” key:
a) If standardization is not in memory; display shows "Put 7.00 pH".
b) If standardization is already in memory, display shows "RE-STD". The system is as
such ready for measurement. [However, if re-standardization is desired, press “STD”
key again and display will show. Put 7.00 pH.]
c) Put /dip electrodes in standard pH 7.00 buffer solution. Press “STD” key again.
Display will show "SYS BUSY", then finally will show -"Put pH 4.00 or pH 9.2".
d) Clean electrodes and dip in either pH 4.00 or pH 9.2 buffer solution. Press “STD”
key
again. Display would show "SYS BUSY" and finally "STD” over.
4. Clean, dry and dip electrodes in unknown solution.
5. Press "pH" key and display will show temperature and pH of the water to be tested.
Result: Date of Analysis: - ___________
Discussion on Result:
Environmental Engineering Laboratory, Civil Engineering Department, Bengal Engineering & Science University
Introduction: Turbidity is defined as an expression of the optical property that causes light to be
scattered and absorbed rather than transmitted in straight lines through the sample. This scattering
and absorption are caused by the interaction of light with suspended particles in the samples.
Turbidity is caused by the suspended solids including silt, clay, algae and other plankton, microbes,
and other fine insoluble particles. The higher the intensity of scattered light, the higher the
turbidity. However, turbidity is not a direct measure of suspended particles in water. Turbidity is
expressed in Nephelometric Turbidity Units (NTU).
Apparatus & Glassware:
Turbiditymeter (Nephelometer), Sample cell.
Reagents:
Preparation of Formazin suspension which is used as standard turbidity suspension is as
follows:
1) Solution I – Dissolve 1.0 g hydrazine sulfate [(NH 2)2.H2SO4], in distilled water and
dilute to 100 ml in a volumetric flask.
3) In a 100 ml volumetric flask, mix 5.0 ml Solution I and 5.0 ml Solution II. Let it stand
for 24 h at 25 0C, dilute to mark, and mix. The turbidity of this suspension is 400 NTU.
Operation of Nephelometer:
a) Switch the instrument ON and allow 10/15 minutes for warm up.
b) Select appropriate RANGE: (insert the CELL RISER for the 0-1000 NTU Range.)
c) Set STANDARDIZE CONTROL(CAL knob)to maximum and insert Sample Cell with
distilled water into cell holder and cover with shield. Using "SET ZERO CONTROL" adjust the
meter reading to zero.
d) Remove the above Cell and replace with another Cell containing standard solution
( Formazin).
e) Adjust "STANDARDIZE CONTROL" to get the desired turbidity reading on the scale.
Repeat steps from (c) to (e) twice.
4
f) Instrument is now ready for test samples. Insert the cell containing samples of unknown
turbidity in the cell holder and note the readings on meter.
g) Standardize the instrument when using other RANGES.
Discussion on Result:
Environmental Engineering Laboratory, Civil Engineering Department, Bengal Engineering & Science University
Light Shield
Scattered Light
Sample Cell
Photo Cell
Lenses
Lamp
Calcium Hardness(Ca
Volume Total Hardness(TH)
Sa H)
of
Date of Volume Remark
mple Location Source Sample Ca H
No.
Collection
taken
of
EDTA
Mean TH
(mg/l as
Volume
of EDTA
Mean (mg/l as
s
(ml) (ml) (ml)
used CaCO3) used (ml) CaCO3)
(ml)
Environmental Engineering Laboratory, Civil Engineering Department, Bengal Engineering & Science University
mL of EDTA 1000
Calculation: Hardness, mg/L as CaCO 3
mL of sample
Discussion on Result: