Specification R1:60ml×5,R2:6ml×1 / /
Semi-Auto or Auto Chemistry
Batch Applicable
Sample Size 3 boxes 20191022 Analyzer produced by Rayto Life
No. Instrument
and Analytical Sciences Co., Ltd.
Shelf Date of
Date of Inspection 2019.10.28 2021.04.21 2019.10.22
Life Production
Inspection Standard (Technical Requirements of
Inspection Item Inspection Results
Product)
R1 and R2 are both colorless, transparent
1 Appearance liquid, without cloudiness or undissolved ☑Conformable □Inconformable
materials.
5 Accuracy The relative deviation should not exceed ±10%. Result: -0.39%
☑Conformable □Inconformable
Result: Low-level: 1.77%
6 Repeatability Coefficient of Variation (CV) ≤ 5% High-level: 0.78%
☑Conformable □Inconformable
When testing the 65umol/L calibrator or sample,
Sensitivity for Result: 0.3272A
7 at the optic path length of 10mm, the
analysis
absorbance difference (△A) should be ≥ 0.29A. ☑Conformable □Inconformable
Within the range of 0.8umol/L ~ 342 umol/L
Linearity Result: 0.9993
8 (37℃), the correlation coefficient (r) should be ≥
range
0.990. ☑Conformable □Inconformable
Conclusion Conforming