You are on page 1of 9

Int. J. Radiation Oncology Biol. Phys., Vol. 67, No. 2, pp.

630 – 638, 2007


Copyright © 2007 Elsevier Inc.
Printed in the USA. All rights reserved
0360-3016/07/$–see front matter

doi:10.1016/j.ijrobp.2006.10.007

PHYSICS CONTRIBUTION

CORRELATION BETWEEN INTERNAL FIDUCIAL TUMOR MOTION AND


EXTERNAL MARKER MOTION FOR LIVER TUMORS IMAGED WITH 4D-CT

A. SAM BEDDAR, PH.D.,* KRISTOFER KAINZ, PH.D.,* TINA MARIE BRIERE, PH.D.,*
YOSHIKAZU TSUNASHIMA, M.S.,* TINSU PAN, PH.D.,† KARL PRADO, PH.D.,* RADHE MOHAN, PH.D.,*
MICHAEL GILLIN, PH.D.,* AND SUNIL KRISHNAN, M.D.‡
Departments of *Radiation Physics, †Diagnostic Imaging, and ‡Radiation Oncology, The University of Texas M. D. Anderson
Cancer Center, Houston, TX

Purpose: We investigated the correlation between the motions of an external marker and internal fiducials
implanted in the liver for 8 patients undergoing respiratory-based computed tomography (four-dimensional CT
[4D-CT]) procedures.
Methods and Materials: The internal fiducials were gold seeds, 3 mm in length and 1.2 mm in diameter. Four
patients each had one implanted fiducial, and the other four had three implanted fiducials. The external marker
was a plastic box, which is part of the Real-Time Position Management System (RPM) used to track the patient’s
respiration. Each patient received a standard helical CT scan followed by a time-correlated CT-image acquisition
(4D-CT). The 4D-CT images were reconstructed in 10 separate phases covering the entire respiratory cycle.
Results: The internal fiducial motion is predominant in the superior–inferior direction, with a range of 7.5–17.5
mm. The correlation between external respiration and internal fiducial motion is best during expiration. For 2
patients with their three fiducials separated by a maximum of 3.2 cm, the motions of the fiducials were well
correlated, whereas for 2 patients with more widely spaced fiducials, there was less correlation.
Conclusions: In general, there is a good correlation between internal fiducial motion imaged by 4D-CT and
external marker motion. We have demonstrated that gating may be best performed at the end of the respiratory
cycle. Special attention should be paid to gating for patients whose fiducials do not move in synchrony, because
targeting on the correct respiratory amplitude alone would not guarantee that the entire tumor volume is within
the treatment field. © 2007 Elsevier Inc.

4D-CT imaging, Tumor motion, Respiratory gating, Internal fiducials, Hepatobiliary Cancer.

INTRODUCTION motion: (1) direct or indirect tumor monitoring with internal


fiducials and fluoroscopic imaging (2, 5) and (2) indirect
Accounting for tumor motion in treatment planning and tumor monitoring with external markers and/or respiratory
delivery is one of the most recent and significant challenges sensors (6, 11). The first technique enables gated treatments
facing radiotherapy for the abdomen and thorax. Respirato- because the tumor motion can be tracked nearly in real time.
ry-gated radiotherapy can be used to achieve a high dose The second method can be used for time-correlated com-
conformity around the target while reducing the overall puted tomography (CT)-image acquisition (four-dimen-
irradiated volume so as to reduce normal tissue complica- sional CT [4D-CT]) imaging (12, 13) as well as gated
tions (1– 4). Initial efforts were directed toward lung tumors therapy but relies heavily on the ability of the respiratory
because of the tumor’s obvious respiratory motion, the poor sensors to predict the tumor location reproducibly.
prognosis for lung cancer, and the deleterious effects of Although several studies have clarified the relationship
radiotherapy on normal lung tissue including pneumonitis between the motion of an external marker and an internal
and fibrosis (3). Recently, however, there is burgeoning fiducial for specific breathing sensors (4, 10), these relation-
interest in respiratory gating for liver tumors as well. Gated ships may differ for each type of sensor. In addition, most
radiotherapy for the liver is still in its infancy in the United studies have been performed with fluoroscopic units and not
States, and to our knowledge this technique has been used in the CT images used for treatment planning. The primary
only a few institutions worldwide (1, 2, 5–10). purpose of this study was to investigate the correlation
Two basic techniques have been used to track tumor between the external respiratory signal obtained using the

Reprint requests to: A. S. Beddar, Ph.D., Department of Radi- mdanderson.org


ation Physics, Unit 94, University of Texas M. D. Anderson Conflict of interest: none.
Cancer Center, 1515 Holcombe Boulevard, Houston, TX 77030. Received Aug 9, 2006, and in revised form Sept 29, 2006.
Tel: (713) 563-2609; Fax: (713) 563-2620; E-mail: abeddar@ Accepted for publication Oct 2, 2006.

630
Correlation between internal fiducial tumor motion and external marker motion in 4D-CT for liver tumors ● A. S. BEDDAR et al. 631

Real-Time Position Management System (RPM, Varian Brief description of the Varian RPM as the “external
Medical Systems, Palo Alto, CA) and tumor motion for marker” system
liver cancer. Our second goal was to determine how the We presume that the motion of an external marker placed on the
motions of multiple fiducials correlate with each other when patient is correlated primarily with the respiratory component of
implanted in the liver. Our third aim was to establish quali- the patient’s internal motion. The marker is an infrared reflective
tative criteria to identify those patients who could benefit from circle mounted on one face of a plastic box, which is taped to the
patient’s abdomen. This box is a component of the Varian RPM
respiratory-gated radiotherapy. Finally, we hoped to demon-
system and has been described before (14 –16). Additional com-
strate the value of 4D-CT imaging for the management of ponents include an infrared-sensitive video camera, which records
tumor motion for gated as well as nongated treatment delivery. the location and motion of the marker, and a program that tracks
and later displays the location and motion of the marker. Data for
METHODS AND MATERIALS the marker’s motion with time (referred to hereafter as the trace)
can be monitored in real time during imaging or treatment proce-
Terminology: markers and fiducials dures or can be saved in ASCII format for offline analysis. The
Throughout the discussion that follows, the following terminol- system is shown in Fig. 1.
ogy is used to distinguish the reference points that are external to
the patient from those that are internal. Traceable reference points
exterior to the patient are referred to as external markers or simply Description of the 4D-CT data acquisition
markers and are described subsequently, whereas reference points A thorough description of the techniques used to obtain and
located inside the patient are referred to as fiducials. reconstruct 4D-CT data are given by Pan et al. (12). The process
Implanted fiducials consisted of gold cylinders used with the used in this study is briefly reviewed here. Before the 4D-CT
ACCULOC image-guided radiation therapy (IGRT) system (Med- acquisition, the RPM box is placed on the patient’s abdomen at a
Tec, Orange City, Iowa); each cylinder had a diameter of 1.2 mm location that exhibits the maximum motion in the AP direction. A
and a length of 3 mm. The high density of gold enables the series of axial CT scans in cine mode are obtained at consecutive
fiducials to be readily discernible within the CT images (albeit couch positions spaced 2 cm apart. The scanning duration at each
with some beam-hardening artifacts), as well as within radiographs couch position must be long enough to encompass at least one full
obtained using 6-MV photon beams for gated treatment. cycle respiratory cycle. The images are reconstructed to produce at
For some of our liver patients who received 4D-CT scans, postop- least 10 CT images that can be binned into 10 phases covering the
erative surgical clips or stents were present. It was possible to localize entire breathing cycle. The cine CT images are subsequently sent,
accurately the surgical clips or the tips of the stents within the CT data along with the RPM trace recorded during the acquisition, to a GE
as well, and we therefore also designated them as fiducials. Advantage Workstation (AW) for processing. The AW software

Fig. 1. The Real-Time Position Management System (RPM): (a) infrared camera, (b) reflective box placed on the
abdomen of a patient. Sample RPM respiratory traces: (c) the default spacing between phases, with the 0% and 50%
phases shown in green, (d) the same trace with phase correction so that the 50% phase corresponds to the end of
expiration, and (e) amplitude gating at the end of expiration.
632 I. J. Radiation Oncology ● Biology ● Physics Volume 67, Number 2, 2007

half of the respiratory cycle (Fig. 1d). Thus, we correct for the
mismatch between the end of expiration and the 50% phase.
Among the reconstructed 4D-CT cine images, those that corre-
sponded best to phases of the marker motion (or to phases of the
respiratory cycle) were identified. The phase-bins were taken in
10-percent increments of the full marker-motion period.
We should note that the 4D images are used only for internal
target volume (ITV) tumor contouring. Contouring of normal tissues
and organs for treatment planning is performed on a normal CT
image set that is obtained from a free-breathing (FB) helical scan
just before the 4D image acquisition.

Technique to locate the centers of each fiducial in the


4D-CT data
First, the 10 4D-CT image sets were exported to the Pinnacle
treatment planning system (Philips Medical Systems, Bothell,
WA). For each fiducial in each CT image set, a point of interest
(POI) was defined. For the first 4 patients in our study, we used a
single fiducial, and for the last 4 patients, we used three fiducials.
A typical example is shown in Fig. 2. The axial CT slice that
Fig. 2. Digitally reconstructed radiograph of Patient 8, showing an corresponded to the center of the fiducial was identified; within
example of the placement of the fiducials within the liver. The this slice, the POI was positioned at the center of the fiducial, along
three fiducials are red, the liver is contoured in yellow, and the
the lateral and anterior–posterior (AP) directions (i.e., anatomic
tumor is contoured in blue.
transverse plane). The sagittal view of the CT image set was then
used to refine the centering of the POI along the superior–inferior
associates each reconstructed image with a point on the RPM trace. (SI) direction. An example is shown in Fig. 3. The position
In this way, the cine CT images are associated with specific phases uncertainty of each POI is greatest along the SI direction, due to
of the marker’s motion. The standard RPM software identifies the the 2.5-mm axial slice thickness. In the lateral and AP directions,
peaks of the respiratory traces and then equally subdivides each beam-hardening artifacts complicate the POI placement; it is ex-
respiratory cycle into 10 phases. The peaks are denoted as the “0% pected that the uncertainty of selecting the location along these
phase” and correspond to the end of inspiration; however, the directions is approximately 0.5 mm.
crests do not necessarily correspond to the “50% phase” because
each cycle is evenly divided in time (Fig. 1c). To improve the RPM trace analyses and evaluation method
correspondence between the phase identification with the actual To analyze the correlation of the external marker’s position to
peak and crest of the patient’s respiratory cycle, for some patients that of the fiducial implanted within the liver, a means is necessary
we have used an internal software package (17) to identify the to normalize the RPM trace data to compare it directly with the
crest as being the 50% phase and then separately subdivide each fiducial location.

Fig. 3. Axial, coronal, and sagittal views of the anatomy of Patient 1 exhibiting the location of the fiducial marker in
the free-breathing computed tomography series (in blue) and in the 50% phase (in red) within the liver.
Correlation between internal fiducial tumor motion and external marker motion in 4D-CT for liver tumors ● A. S. BEDDAR et al. 633

The RPM trace data were processed in the following manner The parameters m and b were used as the scale and offset
with the use of a MATLAB routine. Of the entire RPM trace data terms to convert the “raw” RPM trace amplitudes at each time t,
recorded during the 4D-CT scan, the only section of the RPM trace RPMraw(t), to amplitudes RPMnorm(t) that, in principle, should
considered for analysis consisted of the full free-breathing cycles correspond directly to the SI coordinate of the fiducial within the
during which the 4D-CT scan acquisition took place. A “full patient:
free-breathing cycle” comprises the region of the RPM trace
between consecutive phase values of zero; over the course of a RPMnorm(t) ⫽ m ⫻ RPMraw(t) ⫹ b. (2)
free-breathing cycle, the phase value ranges from zero to 2␲.
Within the ASCII data for the RPM trace, a TTL “on” signal sent
Each full free-breathing cycle of RPMnorm(t) was plotted, and
from the CT scanner console to the RPM computer indicated when
the SI coordinates of the fiducial position and their uncertainties
the CT scanner’s X-ray tube was on, and the RPM-trace ASCII
were overlaid on this graph. Such a plot would be useful in
data were flagged accordingly (1 vs. 0). For each free-breathing
addressing the following issues:
cycle within the RPM trace region considered for analysis, the 0%,
10%, 20%, . . . “time-in-cycle” points (corresponding to phase 1. The correspondence between the marker’s AP motion and the
values of 0, ␲/5, 2␲/5, . . .) were located. The amplitude (y axis fiducial’s SI motion. Better agreement at certain phase values
value) of the RPM trace was recorded at each of these points, and may suggest that the RPM amplitude would more reliably
for each phase value ␾ the average RPM amplitude 具RPMraw典␾ predict the position of the fiducial (and thus the target) at those
was calculated. phase values.
Meanwhile, the SI position of the fiducial was measured, in the 2. The legitimacy of assuming a linear correspondence between
manner described earlier, within the 4D-CT phase image data sets 4DCT␾ and 具RPMraw典␾.
(0%, 10%, 20%, . . .) corresponding to the RPM trace data. For the 3. The consistency of the patient’s AP motion, overall free-
patients with three fiducials, the average SI position of the three breathing cycles recorded during the 4D-CT scan. Consider-
fiducials relative to each midpoint of motion was used. The un- able spread in the amplitude of RPMnorm(t) might suggest
certainty in the phase value is a measure of how “close” a given considerable uncertainty in the SI position of the fiducial.
reconstructed 4D-CT phase image is to a particular “canonical”
phase value (0%, 10%, 20%, . . .). For example, the GE Advantage
Workstation (AW) may associate, with a canonical phase of 50%, RESULTS
a reconstructed cine image whose time stamp is most consistent A single fiducial was first identified within the liver for
with the 50% phase, but which may not correspond exactly to the
each of the first 4 patients considered in this study. Within
50% phase; the cine image could be anywhere between the 45%
each of the 4D-CT image sets for all 10 individual phases as
and 55% phases. In this case the matching tolerance is 5%. In
principle, the uncertainty in the phase can be determined at each well as for the free-breathing helical CT scan, the coordi-
phase value by reviewing the AW results. In practice, we recon- nates of the fiducial (in the SI, AP, and lateral directions)
struct and export enough images to AW to keep the matching were determined as described earlier and plotted in Fig. 5.
tolerance to within 3%. (Note that the scale on the ordinate of the SI displacement
Once the SI coordinate of the fiducial 4DCT␾ and the average is twice as large as the ordinates of the AP and lateral
RPM amplitude 具RPMraw典␾ were determined for each phase value displacements. Therefore, the AP and lateral displacements
␾, the RPM trace data were renormalized in the following manner. are not as large as the SI direction as they may appear in the
First, a plot of 4DCT␾ vs. 具RPMraw典␾ was generated, an example figure.) The locations of the fiducials relative to the dome of
of which is shown in Fig. 4. We assumed that not only would there the patient’s liver and midcoronal and midsagittal planes are
be a one-to-one correspondence between 4DCT␾ and 具RPMraw典␾,
shown in Table 1. Also shown in Table 1 is the range of
but also that the correlation would be linear, as suggested by
motion of the fiducials. The lateral motion of the fiducial
Gierga et al. (18). A linear fit was applied to the data, and the slope
m and intercept b were recorded for the fit: tends to be minimal, less than 3 mm over the course of a full
respiration. This motion also does not show a strong depen-
dence with the phase of the respiratory motion. In the AP
4DCT⌽ ⫽ m ⫻ 具RPMraw典⌽ ⫹ b. (1)
direction, the displacement ranged from 1.2 mm to 3.5 mm
for all 4 patients. For Patients 2 and 4, the motion does
appear to correlate roughly with the respiratory phase. The
SI motion of the fiducial is significant for all 4 patients,
ranging between 7.5 mm and 10.5 mm. Furthermore, the
extrema of the fiducial motion occurs near the 90 – 0% and
the 40 – 60% phase values.
Also plotted in Fig. 5 are the coordinates of the fiducial as
measured within the FB image data set (dashed line). The
fiducial coordinates in the FB image set are mostly within
the range of the 4D data points with a maximum deviation
of 2 mm. Some differences in the positions of the marked
Fig. 4. Example of the fiducial’s superior–inferior (SI) coordinate
as a function of the amplitude of the raw Real-Time Position isocenters determined from three external BBs placed on the
Management System (RPM) trace. A linear fit to this line is used skin of the patients are also observed (Table 2). The largest
to generate the renormalized RPM trace. difference seen between the FB and 4D-CT isocenters is
634 I. J. Radiation Oncology ● Biology ● Physics Volume 67, Number 2, 2007

Fig. 5. Fiducial coordinates in the superior–inferior (SI), anterior–posterior (AP), and lateral (Lat) directions for the first
4 patients in this study. The symbols represent the fiducial’s position in the 10 phases of the time-correlated computed
tomography (CT) series, and the dotted line represents the fiducial’s position in the free-breathing CT series. The
coordinates were normalized to the midpoint of the extrema in each direction. Note the scale for the coordinates in the SI
direction is twice as large as in the AP and lateral directions. Note also that the positive fiducial motion in the SI direction
is inferior to the diaphragm.

Table 1. Patient data, including the locations of the fiducials with respect to the dome of the liver in the superior–inferior direction and
midsagittal and midcoronal planes

Fiducial location (cm) Range of motion (cm)

Patients Dome Midsagittal plane Midcoronal plane SI Lat AP

1 5.8 ⫺8.8 ⫺0.8 0.80 0.16 0.12


2 9.7 ⫺1.4 6.5 1.00 0.23 0.35
3 6.0 ⫺3.5 4.7 0.75 0.25 0.29
4 10.3 ⫺6.9 1.3 1.05 0.20 0.25
5 2.3, 2.8, 3.8 ⫺4.6, ⫺4.7, ⫺2.5 6.3, 3.1, 4.1 1.08 0.11 0.60
6 9.0, 9.0, 9.7 ⫺2.2, ⫺2.1, ⫺0.3 5.4, 6.4, 6.7 0.80 0.16 0.25
7 7.0, 8.9, 13.6 ⫺4.2, ⫺3.7, ⫺8.4 1.0, ⫺3.7, ⫺2.3 1.75 0.50 0.87
8 4.2, 9.6, 10.0 1.1, 2.5, ⫺4.8 6.8, 9.9, 8.4 0.93 0.35 0.84

Abbreviations: SI ⫽ superior-inferior; AP ⫽ anterior-posterior; Lat ⫽ lateral.


Patients 1– 4 each had one fiducial, whereas Patients 5– 8 had three fiducials. The distances were calculated at the 50% phase (end-expiration).
A positive distance from the midsagittal plane indicates placement in the left side of the body, and a negative distance indicates placement in the
right side. A positive distance from the midcoronal plane indicates anterior placement of the fiducial, and a negative distance indicates posterior
placement. All fiducials were inferior to the dome of the liver. The range of fiducial motion in the SI, lateral (Lat), and anterior–posterior (AP)
directions is also given. For Patients 5– 8, this represents the maximum value for the three fiducials in each direction.
Correlation between internal fiducial tumor motion and external marker motion in 4D-CT for liver tumors ● A. S. BEDDAR et al. 635

Table 2. Differences between isocenters in the free-breathing Table 3. The calculated slope (m) and intercept (b) used for the
and respiratory-based CT scans calculation of the renormalized Real-Time Position Management
System trace
Patients ⌬SI (cm) ⌬AP (cm) ⌬Lat (cm)
Patients m b
1 0.00 0.05 0.05
2 0.25 ⫺0.25 ⫺0.32 1 2.06 ⫺3.55
3 0.03 0.04 ⫺0.09 2 0.79 ⫺1.53
4 0.20 0.06 ⫺0.04 3 1.34 ⫺1.05
5 0.00 0.04 ⫺0.07 4 1.51 0.69
6 0.25 ⫺0.08 ⫺0.38 5 1.39 ⫺10.4
7 ⫺0.25 0.15 ⫺0.25 6 1.38 1.95
8 0.00 0.02 0.05 7 1.39 ⫺1.33
8 0.62 ⫺2.85
Abbreviations: SI ⫽ superior–inferior; AP ⫽ anterior–posterior;
Lat ⫽ lateral.

trace amplitudes at near the 100% phase agree well with


about 3 mm for Patient 2 and coincides with a deviation in their amplitudes at the 0% phase, we may suspect that the
the FB fiducial coordinates from the 4D coordinates, par- linear correlation between fiducial position and marker po-
ticularly in the AP and lateral directions. This indicates that sition breaks down in some fashion between end-expiration
the patient probably shifted slightly during simulation. and end-inspiration. A possible explanation for this general
Figure 6 shows the renormalized RPM trace data along trend may be that, during inhalation, the AP motion of the
with the SI coordinates of the fiducial for the first 4 patients. abdomen lags behind the SI motion of the diaphragm and
The slopes and intercepts used in the renormalization equa- the liver. Perhaps, during inhalation, the chest tends to
tion for all patients are shown in Table 3. For these patients, expand before the abdomen expands. Also, there may be
the renormalized RPM trace amplitude tends to correlate components of abdominal motion unrelated to respiration,
best with the fiducial position, within the uncertainty of the such as muscle tightening or loosening, that may be present
fiducial position, from the 0% phase through the 60% phase during either inhalation or exhalation. It should be noted
(from end-inspiration to end-expiration). The differences that if only the phase values corresponding to inhalation
between the RPM trace and the fiducial position are 0.03– (60 –90%) were used in the linear fit described by Eq. 1,
0.06 mm for the 0 – 60% phases and (–1.4)–(– 0.06) mm for then the RPM trace would likely overpredict the fiducial
the 70 –90% phases. This would suggest that for these motion during exhalation.
patients, the RPM trace is a reliable predictor for the fiducial The results for the 4 patients having three fiducials are
position. In particular, gating treatments during phases the similar to those having only one fiducial. From Fig. 7, we
40 – 60% phases, using the RPM trace amplitude as a guide, can see that, again, the amplitude of lateral motion is small
should yield suitable localization of the fiducial and thus and shows little dependence on respiration. There is a stron-
the target volume during treatment. For Patients 1 and 3, the ger dependence for AP motion, particularly for Patients
renormalized RPM amplitude underpredicts the fiducial po- 6 – 8. Finally, the greatest amplitude occurs for SI motion,
sition during inspiration. Given that the renormalized RPM with a range of 8.0 –17.5 mm. This motion is well correlated

Fig. 6. Superior–inferior (SI) coordinates of the fiducials (diamonds) and renormalized Real-Time Position Management
System (RPM) traces (x) for the first 4 patients in this study. The RPM traces were renormalized using the linear fit
discussed in Methods and Materials.
636 I. J. Radiation Oncology ● Biology ● Physics Volume 67, Number 2, 2007

Fig. 7. Fiducial coordinates in the superior–inferior (SI), anterior–posterior (AP), and lateral (Lat) directions for the
second 4 patients in this study. Fiducial 1 (diamond) is the most superior, and Fiducial 3 (circle) is the most inferior.
Note the scale in the SI direction is twice as large as in the AP and lateral directions. The error bars are not shown. Note
also that the positive fiducial motion in the SI direction is inferior to the diaphragm.

with the respiratory phase for all 4 patients. More important, this study, 2 showed shifts in the isocenter of 3 mm or more.
for Patients 5 and 6, there is a strong correlation and At our institution, the tumor is contoured on the 4D image
synchrony in the motion of all three fiducials. However, for set, but actual treatment planning is performed on the FB
Patients 7 and 8, where the fiducials are much farther apart, scan. The main goal of performing a 4D-CT scan is to
we found less correlation. Although beyond the scope of reduce the uncertainty in the location of the tumor resulting
this study, this may be related to deformation of the liver. A from respiration, thereby allowing a reduction on the mar-
study using deformable image registration, for example, gin of the CTV. One should therefore be careful not to
found that the motion of the diaphragm may not be a good introduce additional uncertainty caused by patient shifts
indicator of liver tumor motion (19). For Patient 8, the AP between acquisition of the FB and 4D scans. This is of less
motion of fiducial 2 has the same amplitude as the SI motion concern for patients undergoing gated radiotherapy because
and thus is an exception to the results for other patients and the treatment isocenter is determined solely from the
fiducials. Figure 8 shows the results for the renormalized
positions of the fiducials during each treatment session.
RPM trace data along with the SI coordinates of the fiducial
Although treatment planning would ideally be performed on
for these 4 patients. For all patients, there appears to be a
the 4D image set (or an intensity projection of the data), it
good correlation between the renormalized RPM trace data
is not currently common practice to do so. This is due, in
and the location of the fiducials for all phases.
part, to the large amount of data required to generate an
image set, limiting the scan to an area covering the tumor
DISCUSSION region within the liver but not the entire abdominal region
Analysis of the data for the first 4 patients shows the required for treatment planning calculations. Furthermore,
importance of comparing the FB and 4D isocenters before the 4D images are acquired with high-Z contrast agents
treatment planning (Fig. 5 and Table 2). Of the 8 patients in (Optiray 320, Mallinkrodt, St. Louis, Missouri) to enhance
Correlation between internal fiducial tumor motion and external marker motion in 4D-CT for liver tumors ● A. S. BEDDAR et al. 637

Fig. 8. Average superior–inferior (SI) coordinates of the fiducials (diamonds) and renormalized Real-Time Position
Management System (RPM) traces (x) for the second 4 patients in this study. The RPM traces were renormalized using
the linear fit discussed in Methods and Materials.

the borders of the tumor, whereas the FB scan is acquired the fiducial and thus the target volume during treatment. Be-
without any contrast agent. cause the renormalized RPM amplitude tends to underpredict
As discussed earlier, the motion of the fiducials is pre- the fiducial position during inspiration, end-inspiratory gating
dominantly in the SI direction. The range of motion is consis- would not be as ideal a choice as end-expiratory gating.
tent with earlier studies obtained with fluoroscopy (5, 7, 18) For Patients 2 and 5, there is a significant variation in the
and a most recent 4D-CT study of liver, spleen, and kidney amplitude of the renormalized RPM traces, with a greater
motion (20). Comparison of the relative locations of the spread during both inspiration and expiration when com-
three fiducials implanted in the liver for Patients 5 and 6 pared with the end of expiration. This tends to lead to some
(Fig. 7) shows a good correlation between the motion of the choppiness in the reconstructed 4D image sets, which is
fiducials throughout the respiratory cycle. Considering the typically of only minor concern because the FB scan is used
distances between the fiducials for these 2 patients (3.2 cm for treatment planning. This is an issue, however, when the
max), our data suggest that one should expect the areas patient’s breathing becomes too shallow during acquisition
delimited by closely spaced fiducials within the liver to at the tumors inferior or superior borders, because this could
move in synchrony. However, for Patients 7 and 8, where lead to an underestimation in tumor volume. Thus it is
the fiducials are more widely spaced (8.6 cm max), there is important to examine the respiration trace during image
less correlation. This is an important result for gating. For reconstruction. When more than one image of the appropri-
Patients 5 and 6, a “miss” in the phase of the gate, while still ate phase is available, as is often the case, the image
at the correct RPM amplitude, would likely cover the tumor coinciding with the more representative respiratory ampli-
volume because it would move as a unit and the relative tude should be selected.
orientation of the fiducials will be phase independent. Identifying the patients who may benefit from respirato-
However, for Patients 7 and 8, irradiating at the incorrect ry-gated radiotherapy is a more difficult task. Patients 1, 3,
phase, even if the external marker is at the correct ampli- 4, and 6 appear to be ideal candidates for respiratory gating
tude, would probably lead to some of the tumor lying because of their consistent breathing cycles. Treatment de-
outside the treatment field because the fiducials would only livery would be most efficient for such a patient. However,
be in the correct orientation at the targeted phase. Further- patients with greater variation in their respiratory cycle
more, verification of the correct targeting during treatment could benefit from respiratory-gated radiotherapy if the gate
delivery (with acquisition of cine electronic portal images, is carefully controlled during treatment delivery. For a pa-
for instance) may be more accurate if more than one fiducial tient with inconsistent breathing, the tumor could be better
is visible in the treatment field. targeted if the beam is turned on only at the end of expira-
Analysis of the renormalized respiration trace and the tion and if the amplitude is within a certain range (Fig. 1e).
locations of the fiducials in the SI direction for all 8 patients
shows strongest correlation between external respiration
CONCLUSIONS
and internal motion during expiration, the 40 – 60% phases.
This would suggest that for these patients, the RPM trace is Our study of 4D-CT images of implanted fiducials in the
a reliable predictor for the fiducial position. In particular, liver shows that the motion of the fiducials is predominant
gating treatments during the 40 – 60% phases, using the RPM in the superior–inferior direction. This motion is well cor-
trace amplitude as a guide, should yield suitable localization of related with the respiratory motion measured by the RPM
638 I. J. Radiation Oncology ● Biology ● Physics Volume 67, Number 2, 2007

system with an external marker placed on the patient’s correlation occurs when the fiducials are closely spaced.
abdomen, with the best agreement occurring at expiration. When performing respiratory gating, care should be taken to
Respiratory gating may therefore be most precise if per- irradiate not only at the correct amplitude but also at the
formed at the end of expiration (i.e., the 40 – 60% phases). correct phase. Finally, one should verify that the isocenters
For patients with multiple implanted fiducials, the best remain aligned from the FB to the 4D-CT scans.

REFERENCES
1. Ohara K, Okumura T, Akisada M, et al. Irradiation synchro- 11. Whyte RI, Crownover R, Murphy MJ, et al. Stereotactic
nized with respiration gate. Int J Radiat Oncol Biol Phys radiosurgery for lung tumors: Preliminary report of a Phase I
1989;17:853– 857. trial. Ann Thorac Surg 2003;75:1097–1101.
2. Shirato H, Shimizu S, Kitamura K, et al. Four-dimensional 12. Pan T, Lee T-Y, Rietzel E, et al. 4D-CT imaging of a volume
treatment planning and fluoroscopic real-time tumor tracking influenced by respiratory motion on multi-slice CT. Med Phys
radiotherapy for moving tumor. Int J Radiat Oncol Biol Phys 2004;31:333–340.
2000;48:435– 442. 13. Chi P, Balter P, Luo D, et al. Relation of external surface to
3. Tada T, Minakuchi K, Fujioka T, et al. Lung cancer: Inter- internal tumor studied with cine CT. Med Phys 2006;33:
mittent irradiation synchronized with respiratory motion— 3116 –3123.
results of a pilot study. Radiol 1998;207:779 –783. 14. Keall PJ, Todor AD, Vedam SS, et al. On the use of EPID-
4. Berbeco RI, Nishioka S, Shirato H, et al. Residual motion of based implanted marker tracking for 4D radiotherapy. Med
lung tumours in gated radiotherapy with external respiratory Phys 2004;31:3492–3499.
surrogates. Phys Med Biol 2005;50:3655–3667. 15. Vedam SS, Kini VR, Keall PJ, et al. Quantifying the predict-
5. Dawson LA, Brock K, Kazanjian S, et al. The reproducibility ability of diaphragm motion during respiration with a nonin-
of organ position using active breathing control (ABC) during
vasive external marker. Med Phys 2003;30:505–513.
liver radiotherapy. Int J Radiat Oncol Biol Phys 2001;51:
16. Mageras GS, Pevsner A, Yorke ED, et al. Measurement of
1410 –1421.
lung tumor motion using respiration-correlated CT. Int J Ra-
6. Minohara S, Kanai T, Endo M, et al. Respiratory gated irra-
diation system for heavy-ion radiotherapy. Int J Radiat Oncol diati Oncol Biol Phys 2004;60:933–941.
Biol Phys 2000;47:1097–1103. 17. Pan T. Comparison of helical and cine acquisitions for
7. Shirato H, Seppenwoolde Y, Kitamura K, et al. Intrafractional 4D-CT imaging with multi-slice CT. Med Phys 2005;32:
tumor motion: Lung and liver. Semin Radiat Oncol 2004;14: 627– 634.
10 –18. 18. Gierga DP, Brewer J, Sharp GC, et al. The correlation be-
8. Wagman R, Yorke E, Ford E, et al. Respiratory gating for tween internal and external markers for abdominal tumors:
liver tumors: Use in dose escalation. Int J Radiat Oncol Biol Implications for respiratory gating. Int J Radiat Oncol Biol
Phys 2003;55:659 – 668. Phys 2005;61:1551–1558.
9. Berbeco RI, Neicu T, Rietzel E, et al. A technique for respi- 19. Brock K, Dawson LA, Sharpe MB, et al. Feasibility of a novel
ratory-gated radiotherapy treatment verifcation with an EPID deformable image registration technique to facilitate classifi-
in cine mode. Phys Med Biol 2005;50:3669 –3679. cation, targeting, and monitoring of tumor and normal tissue.
10. Tsunashima Y, Sakae T, Shioyama Y, et al. Correlation be- Int J Radiat Oncol Biol Phys 2006;64:1245–1254.
tween the respiratory waveform measured using a respiratory 20. Brandner ED, Wu A, Chen H, et al. Abdominal organ motion
sensor and 3D tumor motion in gated radiotherapy. Int J measured using 4D CT. Int J Radiat Oncol Biol Phys 2006;
Radiat Oncol Biol Phys 2004;60:951–958. 65:554 –560.

You might also like