DISCUSSION
An aqueous sample is injected onto a high capacity
anion exchange column where Cr(VI) as CrO42- is
retained and then eluted with an alkaline sulfate eluent.
After this separation, a diphenylcarbazide color reagent
is added to the eluent stream, which then flows through
a photometric detector. The regent forms a color
complex with Cr(VI), which is detected by absorbance
at 530 nm. An example of hexavalent chromium in a Figure 1 Area response for colorimetric chromium detection
aqueous sample is shown in Figure 2.
This method is not applicable to the determination
of total chromium in environmental samples. For
regulatory purposes, total chromium is traditionally
defined as the sum of the free and combined chromium
from both the liquid and solid portions of the sample.
Accordingly, the sample must undergo a strong acid
digestion to release the chromium for analysis. This
step converts all chromium to the Cr(III) form. Other
methods can be used to determine the total chromium
in the digest solution.
Dionex Corporation Dionex Corporation Dionex U.S. Regional Offices Dionex International Subsidiaries
1228 Titan Way Salt Lake City Technical Center Sunnyvale, CA (408) 737-8522 Austria (01) 616 51 25 Belgium (015) 203800 Canada (905) 844-9650 France 01 39 46 08 40 Germany 06126-991-0
P.O. Box 3603 1515 West 2200 South, Suite A Westmont, IL (630) 789-3660 Italy (06) 66 51 50 52 Japan (06) 6885-1213 The Netherlands (0161) 434303 Switzerland (062) 205 99 66 United Kingdom (01276) 691722
Sunnyvale, CA Salt Lake City, UT Houston, TX (281) 847-5652 * Designed, developed, and manufactured under an NSAI registered IO 9001 Quality System.
94088-3603 84119-1484 Atlanta, GA (770) 432-8100 LPN 034685-01 10M 9/00
(408) 737-0700 (801) 972-9292 Marlton, NJ (856) 596-0600 © 2000 Dionex Corporation