Agilent 7890A Network Gas

Chromatograph
Data Sheet

Chromatographic extremely precise retention time repro- guarantee provides greater assurance
Performance* ducibility, which is the basis for all for a low-cost of ownership throughout
• Retention time repeatability chromatographic measurement. the GC’s life.
< 0.008% or < 0.0008 min Agilent’s proprietary Capillary Flow
• Area repeatability < 1%RSD Technology provides a new dimension System Capabilities
The Agilent 7890A is a state-of-the-art in chromatography with reliable, leak- • Supports simultaneously:
gas chromatograph that provides supe- free, in-oven capillary connections that - Two inlets
rior performance for all applications. stand up to repeated GC oven cycling - Three detectors (third detector
Key to its performance is the use of over time. The 7890A GC has enhanced as TCD)
advanced electronic pneumatic control firmware to extend Capillary Flow capa- - Four detector signals
(EPC) modules and high performance bilities and enhanced datasystem soft- • State-of-the-art detector electronics
GC oven temperature control. Each EPC ware to simplify set-up and operation of and the full-range digital data path
unit is optimized for its intended use backflush. These new tools make it enable peaks to be quantified over
with a specific inlet and detector easier to analyze complex matrices and the entire concentration range of the
option. unknowns, and provide gains in produc- detector (107 for the FID) in a single
tivity and data integrity for routine run.
GC oven temperature control of the analyses via 2-dimensional heart cut-
7890A oven allows for fast and precise ting, detector splitting, and column • Full EPC is available for all inlets and
temperature ramping. Overall thermal backflushing. detectors. Control range and resolu-
performance provides optimal chro- tion are optimized for the specific
matography including peak symmetry, The 7890A GC has advanced built-in inlet or detector module.
retention time repeatability, and reten- capabilities to monitor system • Up to six EPC modules can be
tion index accuracy. resources (counters, electronic logs and installed, providing control of up to
diagnostics). Agilent GC systems are 16 channels of EPC.
The combination of precise pneumatic known for their reliability, ruggedness,
and temperature control leads to and long life. The Agilent 10-year use

*Using 7890A with EPC (splitless), ALS and Agilent Data


System for analysis of tetradecane (2 ng to the column).
Results may vary with other samples and conditions.
• Pressure setpoint and contol preci- • A run time deviation log is created • Maximum run time: 999.99 min
sion to 0.001 psi provides more for each analysis to ensure that all (16.7 h).
retention time locking precision for method parameters were achieved • Oven cool down (22 °C ambient) 450
low-pressure applications. and maintained. to 50 °C in 4.0 min (3.5 min with
• EPC with capillary columns provides • A full array of traditional gas sam- oven insert accessory).
four column flow control modes: pling and column switching valves
are available • Ambient rejection: < 0.01 °C per
constant pressure, ramped pressure 1 °C.
(3 ramps), constant flow, or ramped • 550 timed events
flow (3 ramps). Column average lin- • Display of all GC and ALS setpoints
ear velocity is calculated. at the GC or data system. Electronic Pneumatics
• Atmospheric pressure and tempera- • Context-sensitive online help Control (EPC)
ture compensation is standard, so • Compensation for barometric pres-
results do not change, even when
the laboratory environment does.
Column Oven sure and ambient temperature
• Dimensions: 28 × 31 × 16 cm. changes is standard.
• Low Thermal Mass (LTM) system Accommodates up to two 105 m × • Pressure setpoints may be adjusted
can be added to achieve fastest 0.530 mm id capillary columns or two by increments of 0.001 psi, with typi-
cycle times via rapid LTM capillary 10-ft glass packed columns (9 in. coil cal control ± 0.001 for the range
column module heating and cooling. diameter, 1/4 in. od), or two 20-ft 0.000 to 99.999 psi; 0.01 psi for the
• The LAN interface allows real-time stainless steel packed columns range 100.00 psi to 150.00 psi
monitoring of the GC when it is con- (1/8 in. od).
• User may select pressure units as
nected with Lab Monitor & • Operating temperature range suit- psi, kPa, or bar.
Diagnostic Software, even when able for all columns and chromato-
also connected to a data system. • Pressure/flow ramps: Three
graphic separations. Ambient tem-
maximum.
• One-button access to maintenance perature +4 °C to 450 °C.
and service modes from the key- • Carrier and makeup gas settings
- With LN2 cryogenic cooling: selectable for He, H2, N2, and
board –80 to 450 °C. argon/methane.
• Preprogrammed leak tests - With CO2 cryogenic cooling: • Flow or pressure setpoints for each
• Automatic Liquid Sampling is fully –40 to 450 °C. inlet or detector parameter with both
integrated into mainframe control. • Temperature setpoint resolution: Agilent 7890A and Agilent
• Setpoint and automation control can 0.1 °C. ChemStations.
be done from the local keyboard or • Supports 20 oven ramps with 21 • Constant flow mode is available
via a networked data system. Clock- plateaus. Negative ramps are when capillary column dimensions
time programming can be initiated allowed. are entered into the 7890A.
from the front panel to initiate • Maximum achievable temperature
events (on/off, method start, etc.) at • Split/splitless, Multimode, and PTV
ramp rate: 120 °C/min (120 V units inlets have flow sensors for the con-
a future date and time. are limited to 75 °C/min, see trol of split ratio.
Table 1).
• Inlet modules
Pressure sensors: Accuracy:
< ± 2% full scale, Repeatability:
Table 1. Typical 7890A GC Oven Ramp Rates
< ± 0.05 psi, Temperature coeffi-
Temperature 120 V Oven* Fast ramp rates** (°C/min) cient: < ± 0.01 psi/°C, Drift:
range (°C) rates (°C/min) Dual-Channel Single-Channel*** < ± 0.1 psi/6 months.
50 to 70 75 120 120
• Flow sensors: Accuracy: < ± 5%
70 to 115 45 95 120
115 to 175 40 65 110
depending on carrier gas,
175 to 300 30 45 80 Repeatability: < ± 0.35% of setpoint,
300 to 450 20 35 65
* Results obtained with line voltage maintained at 120V
** Fast ramp rates require power > 200 volts at > 15 Amps.
***Requires G2646-60500 oven insert accessory.
2
Temperature Coefficient • Gas saver mode to reduce accessible for improved perfor-
< ± 0.20 mL/min (NTP)* per °C for gas consumption without mance.
He or H2; < ± 0.05 mL/min NTP per compromising performance. • Electronic septum purge flow control
°C for N2 or Ar/CH4. • Electronic septum purge flow con- • Compatible with Merlin Microseal
• Detector modules: trol to eliminate “ghost” peaks. septum
Accuracy: < ± 3 mL/min NTP or 7% • Total flow setting range:
of setpoint, Repeatability: • Setup of parameters facilitated with
0 to 200 mL/min N2 Agilent Solvent Elimination
< ± 0.35% of setpoint 0 to 1,250 mL/min H2 or He
*NTP = 25 °C and 1 atmosphere
Calculator
• Turn top inlet sealing system is built • Total flow setting range:
in standard with each 7890A S/SL
Inlets inlet for quick, easy, injector liner - 0 to 200 mL / min N2
• Maximum of two inlets installed changes - 0 to 1,250 ml/min H2 or He
• EPC compensated for atmospheric
• Turn-top inlet sealing system is built
pressure and temperature variation Multimode Inlet
in standard with each 7890A
• Inlets available: • Provides the flexibility of a standard
Multimode inlet for quick, easy injec-
Agilent split/splitless inlet, com-
tor liner changes
- Packed purged injection port bined with temperature programma-
(PPIP) ble capabilities which allow for
PCOC
- Split/splitless capillary inlet large volume injection. Also sup-
• Direct injection onto cool capillary
(S/SL) ports cool injections for improved
column ensures quantitative sample
signal response.
- Multimode inlet transfer with no thermal degradation.
• Temperature control: LN2 (to
- Temperature-programmable cool • Automatic liquid injection sup-
–160 ºC), LCO2 (to –70 ºC), air cool-
on-column (PCOC) ported directly onto columns
ing (to ambient +10 ºC with oven
≥ 0.250 mm id.
- Programmable temperature vapor- temperature < 50 ºC) (due to high
izer (PTV) consumption, air cooling with cylin- • Maximum temperature: 450 °C.
ders is not advised). Temperature Temperature programming in
- Volatiles inlet (VI)
programming of up to 10 ramps at 3 ramps or tracking oven.
S/SL up to 900 ºC/min. Maximum tem- Subambient control to –40 °C
• Suitable for all capillary columns perature: 450 ºC. is optional.
(50 µm to 530 µm id). • Injection modes: • Electronic pressure control range:
• Split ratios up to 7,500:1 to avoid 0 to 100 psig.
- Hot or cold split/splitless
column overload. Setting split ratios • Electronic septum purge flow con-
(particularly low split ratios) is limit- - Pulsed split/splitless
trol.
ed by column parameters and con- - Solvent vent
• Optional solvent vapor exit for large-
trol of system flows (particularly low
- Direct volume injections.
system flows).
• Suitable for all capillary columns - Electronically controlled, inert,
• Splitless mode for trace analysis.
(50 µm to 530 µm) three-way valve allows solvent
Pressure-pulsed splitless is easily
accessible for best performance. • EPC pressure range (psig): 0 to venting.
100 psig - Includes software for method
• Maximum temperature: 400 °C.
• Split ratio: up to 7500 to 1 to avoid optimization.
• EPC available in two pressure
column overload. Setting split - Preassembled retention gaps/vent
ranges: 0 to 100 psig (0 to 680 kPa)
ratios (particularly low split ratios) line/analytical column for easy
for best control for columns
is limited by column parameters installation.
£ 0.200 mm diameter; 0 to 150 psig
and control of system flows (partic-
for columns < 0.200 mm diameter.
ularly low system flows).
• Splitless mode for trace analysis.
Pressure pulsed splitless is easily

3
PPIP headspace, purge and trap, or ther- • Flameout detection and automatic
• Direct injection onto packed and mal desorption samplers. reignition
wide-bore capillary columns. • Three modes for optimized sample • 450 °C maximum operating tempera-
• Electronic flow/pressure control: introduction: split (up to 100:1 split ture
0 to 100 psig pressure range, 0.0 to ratio), splitless, and direct.
200.0 mL/min flow range. Ranges • Optimized EPC (H2 or He carrier, TCD
are chosen to provide optimum per- 0.00 to 100 psig pressure control, • Thermal conductivity detector (TCD),
formance over normal packed col- 0.0 to 100 mL/min flow control). a universal detector that responds to
umn setpoint ranges. all compounds, excluding the carrier
• Electronic septum purge flow con- gas.
• Electronic septum purge flow con- trol.
trol. • Minimum detectable level: 400 pg
• Silcosteel® treated flow path pro- tridecane/mL with He carrier. (This
• 400 °C maximum operating vides inert surface for minimum
temperature. value may be affected by laboratory
component adsorption. environment).
• Adapters included for 1/4-in. and • Maximum temperature: 400 °C.
1/8-in. packed columns, and • Linear dynamic range: > 105 ± 5%
0.530-mm capillary columns. • Unique fluidic switching design pro-
Detectors vides rapid stabilization from turn-
• Electronic pneumatics control and
PTV on, low-drift performance.
electronic on/off for all detector
• Supports hot/cold split and splitless • Signal polarity can be run-pro-
gases.
modes as well as large volume injec- grammed for components having
tions. • EPC compensated for atmospheric
higher thermal conductivity than the
pressure and temperature variation.
• Temperature control: either LN2 (to carrier gas.
–160 °C) or LCO2 (to –65 °C) cooling. • Maximum temperature: 400 °C
Detectors available:
Temperature programming of up to
3 ramps at up to 720 °C/min. • Standard EPC for 2 gases (He, H2, or
FID
Maximum temperature: 450 °C. N2 matched to carrier gas type)
• Flame ionization detector (FID) that
• EPC pressure range 0 to 100 psig. responds to most organic com- • Make-up gas: 0 to 12 mL/min
pounds. • Reference gas: 0 to 100 mL/min
• Split ratio up to 7,500:1. Setting split
ratios (particularly low split ratios) is • Minimum detectable level (for tride- • The 7890A GC can accommodate a
limited by column parameters and cane): < 1.5 pg C/s third detector as TCD located on the
control of system flows (particularly • Linear dynamic range: >107 (± 10%). left-hand side of the GC.
low system flows). Full-range digital data path enables
• Electronic septum purge flow con- peaks to be quantified over the Micro-ECD
trol. entire 107 concentration range in a • Micro-electron capture detector
single run. (micro-ECD), a very sensitive detec-
• Choice of Gerstel septumless head tor for electrophilic compounds such
or Merlin Microseal® septum head. • Data rates up to 500 Hz accommo-
as halogenated organic compounds.
date peaks as narrow as 10 msec at
• 450 °C maximum operating • Minimum detectable level:
half height.
temperature. < 5.5 fg/mL lindane
• Standard electronic pneumatic con-
• Total flow setting range: At standard checkout conditions,
trol for three gases:
with a detector temperature of
- 0 to 200 mL/min N2 - Air: 0 to 800 mL/min 300 °C and flow to the detector
- 0 to 1,250 mL/min H2 or He - H2 : 0 to 100 mL/min (makeup plus column) of
30 mL/min, this is equivalent to
VI - Makeup gas (N2 or He): 0 to
5.5 fg/sec.
• Very low volume (32 µL) interface 100 mL/min
suitable for gas or prevaporized sam- • Proprietary signal linearization.
• Available in two versions: capillary
ples. Recommended for use with Linear dynamic range: > 5 × 104 with
column optimized or adaptable for
lindane
either packed or capillary columns.
• Data acquisition rate: up to 50 Hz
4
• Uses ß emission of < 15 mCi 63Ni as - H2: 0 to 30 mL/min • MDL: < 3 pg N/s, in both N and
the electron source. - Makeup gas: 0 to 100 mL/min nitrosamine modes, 25 ppm N as
nitrobenzene in toluene
• Unique micro-cell design minimizes • Available for packed/capillary
contamination and optimizes sensi- columns or optimized for capillary • Linear dynamic range: > 104
tivity. columns • Selectivity: > 2 x 107 g N/g C (selec-
• 400 °C maximum operating tempera- • 400 °C maximum operating tempera- tivity in nitrosamine mode is matrix
ture ture dependent)
• Standard EPC makeup gas types: See Agilent Sulfur Chemiluminescence
argon/5% methane or nitrogen; FPD Detector and Nitrogen Chemilumin-
0 to 150 mL/min • Single-wavelength flame photomet- escence Detector Specification Guide
ric detector (FPD), or dual-wave- for additional information regarding per-
NPD length flame photometric detector formance and physical and environ-
• Nitrogen-phosphorus detector (DFPD) – a sensitive, specific detec- mental specifications.
(NPD), a detector specific to nitro- tor to sulfur- or phosphorus-contain-
gen or phosphorus-containing com- ing compounds.
MSD
pounds. • MDL: < 60 fg P/s, < 3.6 pg S/s with See 5975 Series MSD specifications.
• NPD available with one of two methylparathion See 7000A Triple Quadrupole GC/MS
beads, Blos (glass) bead or white • Dynamic range: > 103 S, 104 P with specifications.
ceramic bead (legacy offering) methylparathion
Specialized detectors are available
Compared to the legacy white • Selectivity: 106 g S/g C, 106 g P/g C through Agilent Channel Partners
ceramic bead, the Blos bead offers: including: atomic emission, helium
• Data acquisition rate: up to 200 Hz
- Longer lifetime onization, and pulsed discharge
• Standard EPC for three gases: ionization.
- More stable operation during the
- Air: 0 to 200 mL/min
bead’s lifetime
- H2: 0 to 250 mL/min Auxiliary EPC Devices
• MDL: < 0.1 pg N/s, < 0.03 pg P/s
with azobenzene/malathion/ - Makeup gas: 0 to 130 mL/min The 7890A GC has two positions for
octadecane mixture with Blos bead • Available in single- or dual-wave- auxilary EPC devices located on the
length versions. back of the GC. Each position can be
• MDL: < 0.4 pg N/s, < 0.2 pg P/s
any combination of auxiliary EPC or
with azobenzene/malathion/ • 250 °C maximum operating tempera- pneumatics control module.
octadecane mixture with white ture
ceramic bead Note: The communication for a third
• Agilent 7890A GC’s ability to handle
• Dynamic range: > 105
N, > 105P 4 signals allows simultaneous use of detector as TCD EPC module (located
with azobenzene/malathion mixture DFPD, top-mounted GC detector, and on the left side of the GC) interfaces via
with Blos or white ceramic bead TCD. one of these auxiliary EPC module posi-
tions. If a third detector (TCD) is
• Selectivity: 25,000 to 1 g N/g C,
SCD (Model 355) installed, one of these auxiliary posi-
200,000 to 1 g P/g C with azoben-
• Highest sensitivity and selectivity for tions is thus taken.
zene/malathion/octadecane mixture
with Blos bead sulfur-containing compounds.
• MDL: Typical < 0.5 pg/s, dimethyl Auxiliary EPC Module
• Selectivity: 25,000 to 1 g N/g C,
sulfide in toluene • Three channels of pressure
75,000 to 1 g P/g C with azoben-
control
zene/malathion/octadecane mixture • Linear dynamic range: > 104
with white ceramic bead • EPC compensated for atmospheric
• Selectivity: > 2 x 107 g S/g C pressure and temperature variation
• Data acquisition rate: up to
when connected to a user-defined
200 Hz NCD (Model 255) capillary column
• Standard EPC for three gases: • High selectivity for nitrogen-
containing compounds. • Psig (gauge) and psia (absolute)
- Air: 0 to 200 mL/min pressure control

5
• Forward pressure regulated in the sample path for inertness QuickSwap
• Maximum of 2 auxiliary EPC mod- All of the following purged Capillary The QuickSwap device, for GC/MS,
ules per GC Flow devices require one channel from allows you to change a column or per-
an auxiliary EPC or PCM module. form inlet maintenance without venting
Pneumatics Control Module (PCM) the MSD, saving a considerable
• 2 channels for operation Purged capillary flow devices, such as amount of downtime.
the Deans switch, purged effluent split-
• EPC compensated for atmospheric • Dimensions:
ters, and QuickSwap, introduce an addi-
pressure and temperature variation 31 mm x 16 mm x 1 mm
tional flow in the sample stream. For
when connected to a user-defined (31 mm x 16 mm x 22 mm, including
detectors that operate at low flow
capillary column weldment connectors)
rates, like the MSD and TCD, some
• First channel: decrease in sensitivity will occur. • Weight: 10 grams, not including con-
- Pressure or flow control nector tubing.

- Psig (gauge) and psia (absolute) Deans Switch


Deans switching provides additional Backflush
pressure control Each of the above purged Capillary
selectivity using 2-dimensional GC
- Forward pressure regulated analysis. Peaks of interest that may be Flow devices also provides the ability
coeluting on one column are diverted to to backflush. By reversing column flow
• Second channel:
a separate column of different station- immediately after the last compound of
- Pressure control interest has eluted, you can eliminate
ary phase. This technique can also
- Psig (gauge) and psia (absolute) reduce maintenance costs by having long bake-out times for highly retained
pressure control troublesome solvents or other compo- (or high-boiling) contaminants, thereby
nents bypass detectors or columns. shortening cycle times and protecting
- Forward pressure or back pressure
the column and detector. As backflush
regulated
• Dimensions: occurs after peaks of interest have
• PCM can be located in either/both 65 mm x 31 mm x 1 mm eluted, the chromatographic method
inlet EPC positions, and in (65 mm x 31 mm x 11 mm, including for peaks of interest does not need to
either/both auxiliary positions on weldment connectors with tubing to change. Backflush is available when
the back of the 7890A GC reach through top of oven.) the column is attached to a split/split-
• Maximum of 3 PCMs per GC less, volatiles interface, Multimode, or
• Weight: 30 grams, not including con-
PTV inlet.
nector tubing.
Capillary Flow Technology The 7890A GC firmware has been opti-
Agilent’s proprietary Capillary Flow Purged Effluent Splitters mized for backflush operation:
Technology provides devices with reli- A 3-way purged effluent splitter sends
able, leak-free, in-oven capillary con- column effluent to three detectors, • Displays positive and negative flows
nections to help analyze complex sam- even an MSD. More information can be • Inlet/outlet pressures settable to
ples and provide gains in productivity. obtained in a single run to help locate the limits of the controlling EPC
Devices feature: target peaks in unknowns. A 2-way devices.
purged effluent splitter version is also
• Photolithographic chemical milling available. • EPC can be introduced at any col-
for low dead volume flow pathways umn or restrictor connection
• Diffusion bonding to form a single • Dimensions: • Capillary Flow configuration of up to
flow plate 65 mm x 31 mm x 1 mm six columns/restrictors
(65 mm x 31 mm x 11 mm, including
• “Credit card” profile for fast thermal weldment connectors with tubing to Agilent GC Multitechnique
response reach through top of oven.) ChemStation, EZChrom Elite data sys-
• Projection welded connections for tem, and GC/MSD ChemStation now
• Weight: 26 grams, not including con- include user-interface screens to sim-
leak-tight fittings nector tubing. plify backflush set-up and operation
• Deactivation of all internal surfaces with the 7890A GC.

6
ALS Interface Module operation after delivery and long-term • Two external 24-volt connections (up
• 7693A ALS Interface standard. usage. See http//www.chem.agilent. to 150 mA)
Provides power and communications com/cag/aboutapg/aboutQuality.html • Two on/off contact closures
for up to two 7693A automatic injec- for further information and typical (48 V, 250 mA max)
tors, one automatic sampler tray, product testing.
• 550 timed events via data system.
and one heater/mixer/bar code
• Ambient operating temperature: 50 timed events via GC keyboard.
reader.
15 °C to 35 °C • Support for up to 8 valves.
• 7683 ALS Interface standard.
• Ambient operating humidity: - Valves 1 to 4, 12V DC 13 watt in
Provides power and communications
5% to 95% a heated valve box
for up to two 7683 automatic injec-
tors, one automatic sampler tray, • Storage extremes: –40 °C to - Valves 5 to 6, 24 V DC 100 mA
and one bar code reader. 70 °C unheated, for low power valve
• Injector and tray install easily with- • Line voltage requirements: applications
out the need for alignment. ± 10% of nominal - Valves 7 to 8, externally powered
• Conforms to the following safety as a remote event from separate
Data Communications standards: contact closure
• LAN - Canadian Standards Association • Independent heated zones, not
• Two analog output channels (CSA): C22.2 No. 1010 including oven: six (two inlets, two
(1-mV, 1-V, and 10-V output avail- - CSA/Nationally Recognized Test detectors, and two auxillary). Third
able) as standard Laboratory (NRTL): UL 3101 detector as TCD can utilize any avail-
able zone from inlet or auxillary
• Remote start/stop - International Electrotechnical zones.
• Keyboard control of the Agilent Commission (IEC): 61010-1
• Maximum operating teperatures for
Automatic Liquid Sampler (ALS) - EuroNorm (EN): 61010-1 auxiliary zones: 400 °C
• Storage of 10 methods • Conforms to the following
• Storage of five ALS sequences regulations on Electromagnetic References
Compatibility (EMC) and Radio 1. A Guide to Interpreting Detector
• Binary-coded decimal input for a
Frequency Interference (RFI): Specifications for Gas
stream selection valve
- CISPR 11/EN 55011: Group 1 Chromatography. Agilent
Class A Technologies, publication
Maintenance and Support - IEC/EN 61326
5989-3423EN
Services 2. The Importance of Area and
• Remote diagnostics • Designed and manufactured under
Retention Time Precision in Gas
a quality system registered to ISO
• Performance verification Chromatography. Agilent
9001, Declaration of Conformity
services Technologies, publication
available.
5989-3425EN

Environmental Other Specifications


Conditions/Safety and • Height: 49 cm (19.2 in.) For More Information
For more information on our products
Regulatory Certifications • Width: 58 cm (22.9 in.) with EPC and services, visit our Web site at
The instrument is designed and manu- inlet and detectors; 68 cm www.agilent.com/chem.
factured under a quality system regis- (26.8 in.) with third detector as TCD
tered to ISO 9001. The instrument com- or with certain valving options
plies with international regulatory, safe- mounted on left-hand side of GC
ty, and electromagnetic compatibility
• Depth: 51 cm (20.2 in.)
requirements. The specifications are
Typical weight: 49 kg (108 lb)
more conservative than actual test con-
ditions. In addition, further testing was • Four internal 24-volt connections
done under Agilent standards to ensure (up to 150 mA)

7
www.agilent.com/chem
Silcosteel® is a registered trademark of the Restek
Corporation.
Merlin Microseal® is a registered trademark of Gerstel
GmbH & Co. KG
Agilent shall not be liable for errors contained herein or
for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.

Information, descriptions, and specifications in this


publication are subject to change without notice.

© Agilent Technologies, Inc., 2011


Printed in the USA
December 1, 2011
5989-6317EN