Agilent 7890A Network Gas Chromatograph: Data Sheet
Agilent 7890A Network Gas Chromatograph: Data Sheet
Chromatograph
Data Sheet
Chromatographic extremely precise retention time repro- guarantee provides greater assurance
Performance* ducibility, which is the basis for all for a low-cost of ownership throughout
• Retention time repeatability chromatographic measurement. the GC’s life.
< 0.008% or < 0.0008 min Agilent’s proprietary Capillary Flow
• Area repeatability < 1%RSD Technology provides a new dimension System Capabilities
The Agilent 7890A is a state-of-the-art in chromatography with reliable, leak- • Supports simultaneously:
gas chromatograph that provides supe- free, in-oven capillary connections that - Two inlets
rior performance for all applications. stand up to repeated GC oven cycling - Three detectors (third detector
Key to its performance is the use of over time. The 7890A GC has enhanced as TCD)
advanced electronic pneumatic control firmware to extend Capillary Flow capa- - Four detector signals
(EPC) modules and high performance bilities and enhanced datasystem soft- • State-of-the-art detector electronics
GC oven temperature control. Each EPC ware to simplify set-up and operation of and the full-range digital data path
unit is optimized for its intended use backflush. These new tools make it enable peaks to be quantified over
with a specific inlet and detector easier to analyze complex matrices and the entire concentration range of the
option. unknowns, and provide gains in produc- detector (107 for the FID) in a single
tivity and data integrity for routine run.
GC oven temperature control of the analyses via 2-dimensional heart cut-
7890A oven allows for fast and precise ting, detector splitting, and column • Full EPC is available for all inlets and
temperature ramping. Overall thermal backflushing. detectors. Control range and resolu-
performance provides optimal chro- tion are optimized for the specific
matography including peak symmetry, The 7890A GC has advanced built-in inlet or detector module.
retention time repeatability, and reten- capabilities to monitor system • Up to six EPC modules can be
tion index accuracy. resources (counters, electronic logs and installed, providing control of up to
diagnostics). Agilent GC systems are 16 channels of EPC.
The combination of precise pneumatic known for their reliability, ruggedness,
and temperature control leads to and long life. The Agilent 10-year use
3
PPIP headspace, purge and trap, or ther- • Flameout detection and automatic
• Direct injection onto packed and mal desorption samplers. reignition
wide-bore capillary columns. • Three modes for optimized sample • 450 °C maximum operating tempera-
• Electronic flow/pressure control: introduction: split (up to 100:1 split ture
0 to 100 psig pressure range, 0.0 to ratio), splitless, and direct.
200.0 mL/min flow range. Ranges • Optimized EPC (H2 or He carrier, TCD
are chosen to provide optimum per- 0.00 to 100 psig pressure control, • Thermal conductivity detector (TCD),
formance over normal packed col- 0.0 to 100 mL/min flow control). a universal detector that responds to
umn setpoint ranges. all compounds, excluding the carrier
• Electronic septum purge flow con- gas.
• Electronic septum purge flow con- trol.
trol. • Minimum detectable level: 400 pg
• Silcosteel® treated flow path pro- tridecane/mL with He carrier. (This
• 400 °C maximum operating vides inert surface for minimum
temperature. value may be affected by laboratory
component adsorption. environment).
• Adapters included for 1/4-in. and • Maximum temperature: 400 °C.
1/8-in. packed columns, and • Linear dynamic range: > 105 ± 5%
0.530-mm capillary columns. • Unique fluidic switching design pro-
Detectors vides rapid stabilization from turn-
• Electronic pneumatics control and
PTV on, low-drift performance.
electronic on/off for all detector
• Supports hot/cold split and splitless • Signal polarity can be run-pro-
gases.
modes as well as large volume injec- grammed for components having
tions. • EPC compensated for atmospheric
higher thermal conductivity than the
pressure and temperature variation.
• Temperature control: either LN2 (to carrier gas.
–160 °C) or LCO2 (to –65 °C) cooling. • Maximum temperature: 400 °C
Detectors available:
Temperature programming of up to
3 ramps at up to 720 °C/min. • Standard EPC for 2 gases (He, H2, or
FID
Maximum temperature: 450 °C. N2 matched to carrier gas type)
• Flame ionization detector (FID) that
• EPC pressure range 0 to 100 psig. responds to most organic com- • Make-up gas: 0 to 12 mL/min
pounds. • Reference gas: 0 to 100 mL/min
• Split ratio up to 7,500:1. Setting split
ratios (particularly low split ratios) is • Minimum detectable level (for tride- • The 7890A GC can accommodate a
limited by column parameters and cane): < 1.5 pg C/s third detector as TCD located on the
control of system flows (particularly • Linear dynamic range: >107 (± 10%). left-hand side of the GC.
low system flows). Full-range digital data path enables
• Electronic septum purge flow con- peaks to be quantified over the Micro-ECD
trol. entire 107 concentration range in a • Micro-electron capture detector
single run. (micro-ECD), a very sensitive detec-
• Choice of Gerstel septumless head tor for electrophilic compounds such
or Merlin Microseal® septum head. • Data rates up to 500 Hz accommo-
as halogenated organic compounds.
date peaks as narrow as 10 msec at
• 450 °C maximum operating • Minimum detectable level:
half height.
temperature. < 5.5 fg/mL lindane
• Standard electronic pneumatic con-
• Total flow setting range: At standard checkout conditions,
trol for three gases:
with a detector temperature of
- 0 to 200 mL/min N2 - Air: 0 to 800 mL/min 300 °C and flow to the detector
- 0 to 1,250 mL/min H2 or He - H2 : 0 to 100 mL/min (makeup plus column) of
30 mL/min, this is equivalent to
VI - Makeup gas (N2 or He): 0 to
5.5 fg/sec.
• Very low volume (32 µL) interface 100 mL/min
suitable for gas or prevaporized sam- • Proprietary signal linearization.
• Available in two versions: capillary
ples. Recommended for use with Linear dynamic range: > 5 × 104 with
column optimized or adaptable for
lindane
either packed or capillary columns.
• Data acquisition rate: up to 50 Hz
4
• Uses ß emission of < 15 mCi 63Ni as - H2: 0 to 30 mL/min • MDL: < 3 pg N/s, in both N and
the electron source. - Makeup gas: 0 to 100 mL/min nitrosamine modes, 25 ppm N as
nitrobenzene in toluene
• Unique micro-cell design minimizes • Available for packed/capillary
contamination and optimizes sensi- columns or optimized for capillary • Linear dynamic range: > 104
tivity. columns • Selectivity: > 2 x 107 g N/g C (selec-
• 400 °C maximum operating tempera- • 400 °C maximum operating tempera- tivity in nitrosamine mode is matrix
ture ture dependent)
• Standard EPC makeup gas types: See Agilent Sulfur Chemiluminescence
argon/5% methane or nitrogen; FPD Detector and Nitrogen Chemilumin-
0 to 150 mL/min • Single-wavelength flame photomet- escence Detector Specification Guide
ric detector (FPD), or dual-wave- for additional information regarding per-
NPD length flame photometric detector formance and physical and environ-
• Nitrogen-phosphorus detector (DFPD) – a sensitive, specific detec- mental specifications.
(NPD), a detector specific to nitro- tor to sulfur- or phosphorus-contain-
gen or phosphorus-containing com- ing compounds.
MSD
pounds. • MDL: < 60 fg P/s, < 3.6 pg S/s with See 5975 Series MSD specifications.
• NPD available with one of two methylparathion See 7000A Triple Quadrupole GC/MS
beads, Blos (glass) bead or white • Dynamic range: > 103 S, 104 P with specifications.
ceramic bead (legacy offering) methylparathion
Specialized detectors are available
Compared to the legacy white • Selectivity: 106 g S/g C, 106 g P/g C through Agilent Channel Partners
ceramic bead, the Blos bead offers: including: atomic emission, helium
• Data acquisition rate: up to 200 Hz
- Longer lifetime onization, and pulsed discharge
• Standard EPC for three gases: ionization.
- More stable operation during the
- Air: 0 to 200 mL/min
bead’s lifetime
- H2: 0 to 250 mL/min Auxiliary EPC Devices
• MDL: < 0.1 pg N/s, < 0.03 pg P/s
with azobenzene/malathion/ - Makeup gas: 0 to 130 mL/min The 7890A GC has two positions for
octadecane mixture with Blos bead • Available in single- or dual-wave- auxilary EPC devices located on the
length versions. back of the GC. Each position can be
• MDL: < 0.4 pg N/s, < 0.2 pg P/s
any combination of auxiliary EPC or
with azobenzene/malathion/ • 250 °C maximum operating tempera- pneumatics control module.
octadecane mixture with white ture
ceramic bead Note: The communication for a third
• Agilent 7890A GC’s ability to handle
• Dynamic range: > 105
N, > 105P 4 signals allows simultaneous use of detector as TCD EPC module (located
with azobenzene/malathion mixture DFPD, top-mounted GC detector, and on the left side of the GC) interfaces via
with Blos or white ceramic bead TCD. one of these auxiliary EPC module posi-
tions. If a third detector (TCD) is
• Selectivity: 25,000 to 1 g N/g C,
SCD (Model 355) installed, one of these auxiliary posi-
200,000 to 1 g P/g C with azoben-
• Highest sensitivity and selectivity for tions is thus taken.
zene/malathion/octadecane mixture
with Blos bead sulfur-containing compounds.
• MDL: Typical < 0.5 pg/s, dimethyl Auxiliary EPC Module
• Selectivity: 25,000 to 1 g N/g C,
sulfide in toluene • Three channels of pressure
75,000 to 1 g P/g C with azoben-
control
zene/malathion/octadecane mixture • Linear dynamic range: > 104
with white ceramic bead • EPC compensated for atmospheric
• Selectivity: > 2 x 107 g S/g C pressure and temperature variation
• Data acquisition rate: up to
when connected to a user-defined
200 Hz NCD (Model 255) capillary column
• Standard EPC for three gases: • High selectivity for nitrogen-
containing compounds. • Psig (gauge) and psia (absolute)
- Air: 0 to 200 mL/min pressure control
5
• Forward pressure regulated in the sample path for inertness QuickSwap
• Maximum of 2 auxiliary EPC mod- All of the following purged Capillary The QuickSwap device, for GC/MS,
ules per GC Flow devices require one channel from allows you to change a column or per-
an auxiliary EPC or PCM module. form inlet maintenance without venting
Pneumatics Control Module (PCM) the MSD, saving a considerable
• 2 channels for operation Purged capillary flow devices, such as amount of downtime.
the Deans switch, purged effluent split-
• EPC compensated for atmospheric • Dimensions:
ters, and QuickSwap, introduce an addi-
pressure and temperature variation 31 mm x 16 mm x 1 mm
tional flow in the sample stream. For
when connected to a user-defined (31 mm x 16 mm x 22 mm, including
detectors that operate at low flow
capillary column weldment connectors)
rates, like the MSD and TCD, some
• First channel: decrease in sensitivity will occur. • Weight: 10 grams, not including con-
- Pressure or flow control nector tubing.
6
ALS Interface Module operation after delivery and long-term • Two external 24-volt connections (up
• 7693A ALS Interface standard. usage. See http//www.chem.agilent. to 150 mA)
Provides power and communications com/cag/aboutapg/aboutQuality.html • Two on/off contact closures
for up to two 7693A automatic injec- for further information and typical (48 V, 250 mA max)
tors, one automatic sampler tray, product testing.
• 550 timed events via data system.
and one heater/mixer/bar code
• Ambient operating temperature: 50 timed events via GC keyboard.
reader.
15 °C to 35 °C • Support for up to 8 valves.
• 7683 ALS Interface standard.
• Ambient operating humidity: - Valves 1 to 4, 12V DC 13 watt in
Provides power and communications
5% to 95% a heated valve box
for up to two 7683 automatic injec-
tors, one automatic sampler tray, • Storage extremes: –40 °C to - Valves 5 to 6, 24 V DC 100 mA
and one bar code reader. 70 °C unheated, for low power valve
• Injector and tray install easily with- • Line voltage requirements: applications
out the need for alignment. ± 10% of nominal - Valves 7 to 8, externally powered
• Conforms to the following safety as a remote event from separate
Data Communications standards: contact closure
• LAN - Canadian Standards Association • Independent heated zones, not
• Two analog output channels (CSA): C22.2 No. 1010 including oven: six (two inlets, two
(1-mV, 1-V, and 10-V output avail- - CSA/Nationally Recognized Test detectors, and two auxillary). Third
able) as standard Laboratory (NRTL): UL 3101 detector as TCD can utilize any avail-
able zone from inlet or auxillary
• Remote start/stop - International Electrotechnical zones.
• Keyboard control of the Agilent Commission (IEC): 61010-1
• Maximum operating teperatures for
Automatic Liquid Sampler (ALS) - EuroNorm (EN): 61010-1 auxiliary zones: 400 °C
• Storage of 10 methods • Conforms to the following
• Storage of five ALS sequences regulations on Electromagnetic References
Compatibility (EMC) and Radio 1. A Guide to Interpreting Detector
• Binary-coded decimal input for a
Frequency Interference (RFI): Specifications for Gas
stream selection valve
- CISPR 11/EN 55011: Group 1 Chromatography. Agilent
Class A Technologies, publication
Maintenance and Support - IEC/EN 61326
5989-3423EN
Services 2. The Importance of Area and
• Remote diagnostics • Designed and manufactured under
Retention Time Precision in Gas
a quality system registered to ISO
• Performance verification Chromatography. Agilent
9001, Declaration of Conformity
services Technologies, publication
available.
5989-3425EN
7
www.agilent.com/chem
Silcosteel® is a registered trademark of the Restek
Corporation.
Merlin Microseal® is a registered trademark of Gerstel
GmbH & Co. KG
Agilent shall not be liable for errors contained herein or
for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.