You are on page 1of 4

[Downloaded free from http://www.ijamhrjournal.org on Friday, January 3, 2020, IP: 185.130.147.

68]

Original Article

Diagnostic Accuracy of Apex-Pulse Deficit for Detecting Atrial


Fibrillation
Anjali Rajkumar, Aditya Bhattacharjee, Raja J. Selvaraj
Department of Cardiology, Jawaharlal Institute of Postgraduate Medical Education and Research, Puducherry, India

Abstract
Background: Screening for asymptomatic atrial fibrillation (AF) can identify patients at risk of stroke and help initiate treatment. Apex-pulse
deficit, the difference between apex beat rate and peripheral pulse rate, has been described as a clinical sign to identify AF. However, the
accuracy of this measure to identify AF has not been studied before. Aims: The primary aim of this study was to determine the sensitivity
and specificity of apex-pulse deficit more than 10, measured over 1 min, to identify AF using 12‑lead electrocardiogram (ECG) as the gold
standard. Methods: This was a prospective cross‑sectional study. Subjects were those above 30 years of age with known AF (cases) or not
in AF (controls). Apex-pulse deficit was measured in each of them and correlated with rhythm detected in 12‑lead ECG. Results: A total of
70 patients were studied, 35 cases and 35 controls. Apex-pulse deficit was significantly larger for cases as compared to controls and was a
good discriminant to identify AF. Receiver operating characteristic curve analysis showed an area under the curve of 0.86. With a cutoff of 10,
sensitivity and specificity to identify AF were 62.8% and 85.7%, respectively. Using a cutoff of 5 increased the sensitivity to 80%. Counting
over 30 s was significantly less accurate than counting over one full minute. Conclusion: Apex-pulse deficit is a low‑cost method to identify
AF and may be useful for screening. A cutoff of 5 may enhance the sensitivity of measurement as compared to the traditional cutoff of 10.

Keywords: Atrial fibrillation, pulse deficit, screening

Introduction Various approaches have been used to screen for AF at a


single time point. Devices have been developed to detect
Atrial fibrillation (AF) is the most common sustained disorder
AF by identifying pulse irregularity, and these include blood
of heart rhythm.[1] Normally, the atria are activated by regular
pressure monitors with AF detection function.[4] Opportunistic
pacemaker impulses generated by the sinoatrial node. In AF,
detection of AF using such blood pressure monitors
there is rapid, chaotic activation of the atria. This results in an
has been proved to have reasonable sensitivity and
irregular and often rapid heartbeat.
specificity.[5] Smartphone applications with pulse detection
AF is present in 1%–4% of the population in the West[2] by finger photoplethysmography have also been used.[6]
and becomes more prevalent with increasing age. AF often These are, however, expensive, especially if required in large
produces symptoms from the fast heart rate or heart failure, numbers for community screening, and require training for
resulting in the patients seeking treatment. However, AF can use. Low technology methods such as palpation of pulse by a
also be asymptomatic. AF is associated with a risk of stroke trained nurse to detect irregularity have been employed[7] but
due to the formation of clots in the heart, and this risk is the require training and are subjective. AF can be detected in an
same in asymptomatic patients also. About 25% of stroke electrocardiogram (ECG) and this remains the gold standard.
patients have been diagnosed with asymptomatic AF. [3]
Although the evidence is not clear at present if screening Address for correspondence: Dr. Raja J. Selvaraj,
for asymptomatic patients with AF will provide a net Department of Cardiology, Jawaharlal Institute of Postgraduate Medical
benefit, asymptomatic patients are also at risk of stroke. If these Education and Research, Dhanvantri Nagar, Puducherry ‑ 605 006, India.
E‑mail: rajajs@gmail.com
asymptomatic patients are identified by screening, treatment
with antiplatelet or anticoagulant drugs, especially in patients
with additional risk factors can reduce the risk of stroke. This is an open access journal, and articles are distributed under the terms of the Creative
Commons Attribution‑NonCommercial‑ShareAlike 4.0 License, which allows others to
Access this article online remix, tweak, and build upon the work non‑commercially, as long as appropriate credit
is given and the new creations are licensed under the identical terms.
Quick Response Code:
Website: For reprints contact: reprints@medknow.com
www.ijamhrjournal.org
How to cite this article: Rajkumar A, Bhattacharjee A, Selvaraj RJ.
Diagnostic accuracy of apex-pulse deficit for detecting atrial fibrillation.
DOI:
Int J Adv Med Health Res 2019;6:52-5.
10.4103/IJAMR.IJAMR_48_19
Received: 19‑04‑2019, Accepted: 02‑11‑2019, Published: 02-01-2020

52 © 2020 International Journal of Advanced Medical and Health Research | Published by Wolters Kluwer - Medknow
[Downloaded free from http://www.ijamhrjournal.org on Friday, January 3, 2020, IP: 185.130.147.68]

Rajkumar, et al.: Apex-pulse deficit in atrial fibrillation

For the purposes of screening, a 3‑lead or single‑lead ECG Analysis


has been tried; however, the need for equipment and trained The ECG was read by a cardiologist to identify the rhythm as
personnel for the interpretation has precluded its use for AF or not AF. The use of an apex-pulse deficit measurement
widespread community screening purposes in the background of more than 10 to identify AF was compared with this
of insufficient evidence for screening. gold standard method. Based on these results, sensitivity
An ideal screening technique must satisfy three important and specificity of the apex-pulse deficit to identify AF were
calculated. For the secondary endpoints,  (1) the apex-pulse
criteria – it must be rapid and preferably noninvasive, it must
deficit calculated by doubling the difference at 30 s was
be available at low cost, and minimal training is desirable
compared with the measurement at 1  min and  (2) receiver
with a view to cover large populations, including remote
operating characteristic (ROC) curve analysis was performed
areas. Apex-pulse deficit has been described as a clinical sign
to identify the optimal cut‑off of apex-pulse deficit to identify
in patients with AF. This is the difference between the heart
AF.
rate counted from heart sounds (in terms of apex beat) and
peripheral pulse palpated at the radial artery. The deficit is seen With a case‑to‑control ratio of 1:1, we estimated that a
in AF because some of the heartbeats which are audible do not minimum sample size of 66 patients will be required to provide
produce a palpable pulse owing to the irregularity. A deficit a power of 0.9 to detect an area under the curve of 0.7 using
of 10 or more has been described to indicate AF.[8] However, ROC curve analysis.
the sign has not been clinically validated, and the accuracy to
identify AF is not known. We designed this study to determine Results
the diagnostic accuracy of apex-pulse deficit to identify AF as
A total of 70 patients were studied, 35 in sinus rhythm and 35
compared to the gold standard of a 12‑lead ECG.
in AF. The baseline characteristics of the patients are shown
in Table 1.
Methods
The mean heart rate (counted from apex beat over 1 min) was
This was a prospective cross‑sectional study among patients 81 ± 15 bpm in cases and 90 ± 20 bpm in the controls. Among
attending the cardiology outpatient department in a tertiary the cases, eight patients had a heart rate of more than or equal
care center in southern India. The study was approved by to 100 bpm, while this was present in five patients from the
the Institute Ethics Committee (no: JIP/lEC/2018/0203) and control group.
informed consent was obtained from all participants.
Apex pulse deficit measured at 30 s and at 60 s was significantly
Inclusion criteria larger for cases as compared to controls. Irregular pulse by
1. Patients above the age of 30 years, with known AF (cases) palpation or auscultation was also more frequently noted in
2. Patients above the age of 30 years, attending the cardiology patients with AF [Table 2]. None of the patients in the control
outpatient clinic with conditions such as hypertension or group had premature atrial or ventricular beats on the ECG.
coronary artery disease but not in AF (controls).
The analysis of apex-pulse deficit measured at 60 s to identify
For each patient, informed consent was first obtained. AF showed an area under the ROC curve of 0.86. The ROC
Apex-pulse deficit was measured as described below by two
operators. Immediately following this, a 12‑lead ECG with a
rhythm strip was recorded. Care was taken to avoid a delay Table 1: Baseline characteristics of the patients
longer than 5 min between the calculation of apex-pulse deficit Characteristics Non‑AF (n=35) AF (n=35)
and recording of the ECG. Age, years 55.1±14.6 47.8±11
Male Sex, n (%) 32 (91.4) 18 (51.4)
Measurement of apex-pulse deficit Diabetes mellitus, n (%) 10 (28.6) 3 (8.6)
Two operators, both 3rd year medical students, performed the Hypertension, n (%) 11 (31.4) 6 (17.1)
measurements for all the patients after brief training. With Smoking, n (%) 9 (25.7) 2 (5.7)
the patient seated, operator 1 used a stethoscope to listen to Coronary artery disease, n (%) 22 (62.9) 4 (11.4)
the heart sounds, while operator 2 palpated the radial pulse. Rheumatic heart disease, n (%) 2 (5.7) 12 (34.3)
Once both the operators were ready, a timer was used to start AF=Atrial fibrillation
counting simultaneously. Heart rate counted by each operator
was recorded at the end of 30 s. The procedure was then
repeated and the heart rate was recorded at the end of 1 min. Table 2: Comparison of apex-pulse deficit in the control
The two operators interchanged their roles in different subjects group (sinus) versus atrial fibrillation
to avoid bias. Apex-pulse deficit was calculated as the difference Sinus (n=35) AF (n=35)
between the rates counted by the two operators at the end of Apex-pulse deficit at 30 s 0 (−2, 3) 14 (2, 25)
1 min. The apex-pulse deficit was also derived from the 30‑s (median, 25th and 75th percentiles)
recording by multiplying the difference by 2. In addition, each Apex-pulse deficit at 60 s 1 (0, 2.5) 16 (8, 26)
operator also designated the rhythm as regular or irregular. (median, 25th and 75th percentiles)

International Journal of Advanced Medical and Health Research  ¦  Volume 6  ¦  Issue 2  ¦  July‑December 2019 53
[Downloaded free from http://www.ijamhrjournal.org on Friday, January 3, 2020, IP: 185.130.147.68]

Rajkumar, et al.: Apex-pulse deficit in atrial fibrillation

Table 3: Irregular pulse detection


Sinus (n=35), AF (n=35),
n (%) n (%)
Irregular pulse (palpation) 4 (11.4) 32 (91.4)
Irregular heart beat (auscultation) 4 (11.4) 33 (94.3)

Table 4: Sensitivity and specificity of apex-pulse deficit


and irregular pulse on auscultation and palpation
Measures Sensitivity (%) Specificity (%)
Apex-pulse deficit >10 62.8 85.7
Apex-pulse deficit >5 80 80
Irregular pulse (palpation) 91.4 88.6
Irregular heartbeat (auscultation) 94.3 88.6

Figure 1: Receiver operating characteristic curve of apex-pulse deficit more than 10, whereas in the highest tertile  (heart rate
to identify atrial fibrillation >90 bpm), 100% had an apex-pulse deficit more than 10.
While this assumes significance in a hospital setting where the
curve is shown in Figure 1. The analysis of apex-pulse deficit patients are already diagnosed with AF and are on treatment
measured at 30 s to identify AF showed an area under the to control the ventricular rate, it may be less significant in a
ROC curve of 0.75. The sensitivity and specificity for pulse community screening of previously undiagnosed patients. If
irregularity and different cutoffs of apex-pulse deficit are used for screening, an additional method of screening may be
listed in Tables 3 and 4 respectively. The positive predictive considered in patients with a controlled ventricular rate.
value and negative predictive value of apex-pulse deficit of 10
or more to identify AF were 81.5% and 69.8%, respectively. The sensitivity and specificity of irregular pulse/heartbeat
detection seem to be better than the apex-pulse deficit. Similar
studies using trained nurses have shown sensitivity ranging
Discussion from 91% to 100%, while specificity ranged from 70% to
In this study, we assessed the diagnostic accuracy of apex-pulse 77%.[9] However, this is a subjective measure, and it is unclear
deficit counted over a minute to identify AF. We found that an if the accuracy in the identification of irregular pulse achieved
apex-pulse deficit of 10 or more counted over 1 min identified in this study by medical students and in other studies by nurses
AF with reasonable accuracy. Although apex-pulse deficit has would apply to other health‑care workers.
been described as a clinical sign to identify AF, its diagnostic
accuracy has not been studied before. This assumes importance Smartphone applications have been developed to screen for
in the current era where screening for asymptomatic AF is AF. A smartphone application using photoplethysmography to
assuming importance given the potential to reduce strokes. The identify the pulse can help in screening for AF, according to
sensitivity of 62.8% and specificity of 85.7% suggest that it late‑breaking results from the DIGITAL‑AF study presented
is a reasonably accurate method to identify AF. This makes it at European Society of Cardiology (ESC) congress.[10] Blood
a promising approach for screening of AF in the community pressure monitors have also been used for opportunistic
since it requires minimal training, is cheap, and takes only a detection and screening for AF and have generally shown
couple of minutes to perform. sensitivity >85% and specificity >90%.[4] Our study looked at
For a screening test, it is preferable to have a high sensitivity, a low‑cost, low‑technology option that may be more applicable
even at the cost of loss of some specificity to reduce false for large‑scale screening.
negatives. Our ROC curve analysis suggests that a cutoff of In addition to evaluating apex-pulse deficit as a screening tool,
5 will improve the sensitivity to 80%, while the specificity our study also has value in being the first time, this classically
is still 80%. This cutoff may be preferable for the use for described clinical sign in AF is evaluated in an evidence‑based
screening. fashion. Our findings suggest that while this could be used as
We explored if counting for 30 s may be sufficient instead of for a clinical sign to identify AF with reasonable accuracy, the
a full minute as this will significantly reduce the time taken for detection of an irregular pulse itself may have equal or better
a measurement. However, this was less accurate to identify AF. accuracy.

Apex-pulse deficit is likely to have reduced sensitivity to Limitations


identify AF when the heart rate is well controlled. In our A significant limitation of the study is the use of a two‑gate
patients with AF, among the lowest tertile of heart rate design where patients with previously diagnosed AF were used
(heart rate ≤78 bpm), only 15.4% had an apex-pulse deficit as cases, while those without diagnosed AF were controls.

54 International Journal of Advanced Medical and Health Research  ¦  Volume 6  ¦  Issue 2  ¦  July‑December 2019
[Downloaded free from http://www.ijamhrjournal.org on Friday, January 3, 2020, IP: 185.130.147.68]

Rajkumar, et al.: Apex-pulse deficit in atrial fibrillation

Sensitivity and specificity measured from the study may not References
accurately reflect the sensitivity and specificity of apex-pulse 1. Carlsson  J, Miketic  S, Windeler  J, Cuneo A, Haun  S, Micus  S, et al.
deficit when used for screening in the community. However, Randomized trial of rate‑control versus rhythm‑control in persistent
it does provide a preliminary estimate of the utility as a atrial fibrillation: The strategies of treatment of atrial fibrillation (STAF)
diagnostic test. study. J Am Coll Cardiol 2003;41:1690‑6.
2. Friberg L, Bergfeldt L. Atrial fibrillation prevalence revisited. J Intern
Similarly, the prevalence of AF in the general population is Med 2013;274:461‑8.
expected to be much lower than 50%. The positive predictive 3. Healey JS, Connolly SJ, Gold MR, Israel CW, Van Gelder IC, Capucci A,
value of the test in a population with a low prevalence of the et al. Subclinical atrial fibrillation and the risk of stroke. N Engl J Med
condition would be much lower. As with any screening test, it 2012;366:120‑9.
4. Wiesel  J, Wiesel  D, Suri  R, Messineo  FC. The use of a modified
would be preferable to use a prespecified population at higher
sphygmomanometer to detect atrial fibrillation in outpatients. Pacing
risk, older age, for example, to reduce false positives. Clin Electrophysiol 2004;27:639‑43.
5. Kane  SA, Blake  JR, McArdle  FJ, Langley  P, Sims AJ. Opportunistic
Conclusion detection of atrial fibrillation using blood pressure monitors:
A systematic review. Open Heart 2016;3:e000362.
Apex-pulse deficit is a viable tool to identify AF. Using a 6. Chan  PH, Wong  CK, Poh  YC, Pun  L, Leung  WW, Wong  YF, et al.
cutoff of 5 may improve the sensitivity of AF detection. This Diagnostic performance of a smartphone‑based photoplethysmographic
raises the possibility of using apex-pulse deficit as a low‑cost application for atrial fibrillation screening in a primary care setting.
tool to screen for AF without the need for extensive training. J Am Heart Assoc 2016;5. pii: e003428.
7. Fitzmaurice DA, Hobbs FD, Jowett S, Mant J, Murray ET, Holder R,
However, an irregular pulse on palpation may be more sensitive
et al. Screening versus routine practice in detection of atrial fibrillation
and specific. in patients aged 65 or over: cluster randomised controlled trial. Br Med
J (Clin Res Ed) 2007;335:383.
Financial support and sponsorship 8. Longo D, Fauci A, Kasper D, Hauser S, Jameson JL, Loscalzo J.
One of the authors (A.R.) obtained funding from the Indian Harrison’s Principles of Internal Medicine, 18th Edition. Published by
Council of Medical Research for this study as a Short‑Term McGraw Hill Medical; 2011.
Studentship project in the year 2018. 9. Cooke  G, Doust  J, Sanders  S. Is pulse palpation helpful in detecting
atrial fibrillation? A systematic review. J Fam Pract 2006;55:130‑4.
Conflicts of interest 10. Digital AF, Vandervoort  PM. Late‑Breaking Science in Telemedicine.
There are no conflicts of interest. Munich: European Society of Cardiology Congress; 2018. p. 25‑9.

International Journal of Advanced Medical and Health Research  ¦  Volume 6  ¦  Issue 2  ¦  July‑December 2019 55

You might also like