Ratings and duty cycles of CT subsystems apply if site environment meets the standards of this
section. Maintain environmental conditions listed below at all times – including, for example,
overnight, weekends and holidays. Shut down the CT system if air conditioning is not working.
When system is shut down for major repair, air conditioning may be shut down also.
Section 1.0
Temperature and Humidity Specifications
Environmental specifications apply to the Table, Gantry, Power Distribution Unit, and Operator
Console.
NOTICE System Operation and Image Quality may be affected, if environmental specifications are
exceeded.
5 – Environmental
Minimum allowable ambient room temperature: 18º C (64º F)
Conditions
Maximum allowable ambient room temperature rate of change: 3º C per hour (5º F per hour)
Note: Any cooling equipment cycle control range must be taken into account, such that the maximum and
minimum ambient room temperatures shown above are not exceeded, during room thermal cycling.
For example, if the HVAC is capable of ± 2º C control, then the limits would 20º C - 24º C, to maintain
absolute limits.