Professional Documents
Culture Documents
87]
ORIGINAL ARTICLE
ABSTRACT
Aims: Establishing the reference change values (RCVs) and validating the delta check auto-verification in the hospital information
system (HIS). Materials and Methods: This study was conducted in the Hospital Laboratory-Biochemistry. Fifty-one parameters were
analyzed in three phases. Phase I: Delta check reference change values were established. Phase II: Delta check auto-verification was
validated in the hospital information system. Phase III: Calculation of test requiring manual verification, true and false positive rates.
Results: Out of all the test results, 1.35% failed the RCV-delta check thus requiring manual verification, and the remaining 98.65%
were auto-verified. Only 0.12% test results were true positives indicating laboratory error, and 1.23% were false positives and were
correlated clinically. Ten percent simulated data results and 0.37% actual patient results were not identified by the newly introduced
HIS. Conclusions: RCV-delta check is a refined form of the delta checks used to analyze acceptable analytical and biological variation
in laboratories. Majority of tests passed the RCV-delta check auto-verification, implying that very few test reports require manual
verification. True positives can be detected in the laboratory. All HISs should be validated before implementing complete auto-verification.
Key Words: Auto-verification, delta check, reference change values (RCVs), turnaround time (TAT)
42 © 2017 Muller Journal of Medical Sciences and Research | Published by Wolters Kluwer - Medknow
[Downloaded free from http://www.mjmsr.net on Thursday, February 09, 2017, IP: 109.64.154.87]
probability.[5] RCV brings an objective evidence-based CVA — from the IQC program of our laboratory
edge to the interpretation of a patient’s test report when
a previous report is available. CVI — from Westgard database of desired biological
variable[8]
In a large clinical biochemistry lab with an input of more
than 1,000 samples per day, it is absolutely necessary to Total variance in the lab,[5] SD2 = preanalytical Standard
establish the RCV. This present study was done in order Deviation (SD)2 + analytical SD2 + postanalytical SD2
to reduce the TAT for the release of accurate and precise
results with very little unacceptable and unexplainable Instead of SD, CV was used as a measure of
variation. The RCV and validating the delta check auto- dispersion. [5] Since the preanalytical, i.e. sample
verification manually according to the CLSI Auto-10A collection and centrifugation are standardized, the
guidelines,[6] in a clinical biochemistry lab was done. preanalytical variation is minimal and was excluded. Only
Analytical CVA and CVI were used for the calculation of
Materials and Methods RCV according to Harris and Yasaka. In the IQC, all the
analytes were estimated using IQC dispersed across 6
This study was conducted in the Clinical Biochemistry months. This IQC data was used for calculation of CVA.
section of the Hospital Laboratory. Fifty-one parameters IQC controls that are freeze-dried lyophilized powder
were analyzed. stored at –20° were reconstituted with 5 mL distilled water
and restored in aliquots at –20° when required these
Institutional ethical clearance was taken before starting the aliquots were brought to room temperature by keeping
study. No human or animal subjects were directly involved out for 20 min and then the quality control (QC) run in
in this study. This study was carried out in three phases. the instrument, the similar procedure was followed for
all analytes based on types of control used. CVA was
Phase I calculated for all the parameters, using the following
The delta check (RCV)[7] was established for each of the
formula: . The respective values were
chemistry and immunoassay analytes and was done
based on the following equation: computed using the formula for RCV and the results
obtained are shown in Table 1.
RCV = 2½ × Z × [CVA2 + CVI2]½
Phase II
where RCV — Reference Change values[7] Validation of delta check auto-verification alert in the HIS
was done under standard guidelines, CLSI Auto-10A
2½ — Variation in present sample is compared with guidelines,[6] which include:
previous sample 1. Using simulated data of patient test result, artificially
generated data (30 values for each analyte) were
Z — Z-score for 95% confidence interval. entered in HIS and were verified manually.
Muller Journal of Medical Sciences and Research | Vol 8 | Issue 1 | January - June 2017 43
[Downloaded free from http://www.mjmsr.net on Thursday, February 09, 2017, IP: 109.64.154.87]
2. Actual patient values spanning over a month for each Table 2: Validation of RCV (phase II)
chemistry and immunoassay analyte were run in the Validation Total Manual and his His percentage
system. The data generated were verified manually. tests data mismatch error (%)
Simulated patient data 200 20 10
Validation of HIS with manual cross-checking was done Actual patient reports 810 3 0.37
to ensure the systems reliability.[4] Here the patient test HIS = Hospital information system
Out of all the inpatient test results, 1.35% failed the Using these established RCV for checking patient test
RCV-delta check thus requiring manual verification; the report is a sign of good laboratory practice, but for most
44 Muller Journal of Medical Sciences and Research | Vol 8 | Issue 1 | January - June 2017
[Downloaded free from http://www.mjmsr.net on Thursday, February 09, 2017, IP: 109.64.154.87]
tertiary care centers this is not practical with the given on treatment, nephrotic syndrome, Chronic Kidney
sample load. Hence, it is necessary to implement auto- Disease (CKD) patients on dialysis, and hepatic failure.
verification. Auto-verification maintains the quality of test
reports and significantly reduces the TAT. Here we entered RCV-delta check was therefore a better model of the delta
the RCV into the HIS, test site followed by the live site. checks, which was established and validated in the HIS.
All HISs need to be validated before the implementation.
Auto-verification systems dispatch patient test reports Auto-verification drastically reduces the TAT.
after performing appropriate predefined checks without
the need of manual validation. Therefore, before auto- Standardization of patient test result verification with RCV-
verification is implemented, it needs to be validated.[6] delta percentage change and auto-verification under the
The guideline[6] also recommends a two-step validation, CLSI Auto-10A guidelines can produce quality reports
with simulated data and actual patient test results. consistently even as the test load increases with time.
Simulated data help check this system across the
entire analytical measuring range for robustness of Limitations
performance and helps pick up errors not detectable
1. One percent error was present in the HIS with actual
by using patient test results alone. Validation with
patient results were tested.
simulated data revealed 10% error in the HIS, whereas
2. Ten percent error was present with simulated data.
validation with patient test results revealed 0.37%
3. For establishing the RCV-delta check, we have used
errors, This discrepancy was seen because, validation
analytical CV. For analytes where analytical CV is
with simulated data tests this system across it’s
greater than allowable CV, the established RCV may
analytical measuring range (AMR) thereby helps in
be broader and may fail to detect true positives.
detecting minute errors, these may not be seen in day-
4. Large number of actual patient test results need to
to-day practice but need to be identified and corrected.
be checked over a longer duration of time.
The errors identified were taken constructively by the
5. Manual verification for the detection of false
HIS team and corrected.
negatives was not done.
Approximately, 1.35% of test results failed the delta
Financial Support and Sponsorship
check and required manual verification. Remaining
98.65% were auto-verified when the RCV-delta check Nil.
was used alone, further checks of these patient reports
in the form of limit, critical, and consistency checks Conflicts of Interest
are mandatory for complete auto-verification. [4,6] This study was presented as a poster in the South
Nevertheless, this clearly implies that very few test Regional Conference of the Association of Clinical
reports need manual validation, this is supported by Biochemists of India 2014.
a study done by Shih et al., who demonstrated that
80% test reports can be auto-released and the TAT References
is shortened. [4] Their study was conducted using a
1. Southwick K. Expert systems a feast for leaner laboratories. In:
middleware and a dedicated auto-verification system. CAP Today. Available via DIALOG. Available from: http://www.
Their delta checks were based on the four types of cap.org/apps/docs/cap_today/features_stories/systems_features.
delta checks as follows: Delta difference, delta percent html. [Last accessed on 2014 Feb].
change, rate difference, and rate percent change.[5] In 2. Pearlman ES, Bilello L, Stauffer J, Kamarinos A, Miele R, Wolforet
our study, we used RCV as a form as delta check and M. Implications of autoverification for the clinical laboratory. Clin
Leadersh Manag Rev 2002;16:237-9.
observed it to be a form of delta percent change; our
3. Felder R. Laboratory reporting for the future: Linking
validation was performed using the HIS. We observed a
autoverification to the electronic medical record. American
higher auto-verification rate of 98.65% as we performed Association for Clinical Chemistry. AACC Continuing Education.
only the RCV-delta check. The purpose of delta check Available from: http://www.aacc.org/access/record/default.stm.
is to detect true positives. True positives are errors that [Last accessed on 2014 Feb].
occur in the laboratory[9] (for example, specimen mix-up, 4. Shih MC, Chang MH, Tien N, Hsiao CT, Peng CT. Building and
calibration errors, and typing errors). The true positive validating an autoverification system in the clinical chemistry
laboratory. Lab Med 2011;42:668-73.
rate was 0.12%. False positives are changes occurring
5. Burtis CA, Ashwood ER, Bruns DE. Tietz Textbook of Clinical
in patient test report either due to disease or due to Chemistry and Molecular Diagnostics. 4th ed. Missouri: Elsevier
response to treatment; these correlate clinically and is Saunders; 2006. p. 467-71.
not the main purpose of the delta check. The common 6. Clinical and Laboratory Standards Institute. Autoverification of
false positives encountered were cases of hypokalemia Clinical Laboratory Test Results; Approved Guideline. (CLSI
Muller Journal of Medical Sciences and Research | Vol 8 | Issue 1 | January - June 2017 45
[Downloaded free from http://www.mjmsr.net on Thursday, February 09, 2017, IP: 109.64.154.87]
document AUTO10-A). Wayne, Pennsylvania: Clinical and imprecision, bias, derived from intra and inter-individual biological
Laboratory Standards Institute; 2006. variation. Scand J Clin Lab Invest 1999;59:491-500.
7. Kim WJ, Kim QJ, Kim SI. Differential application of rate and delta 9. Park HS, Kim SY, Lee W, Chun S, Min KW. New decision criteria
check on selected clinical chemistry tests. J Korean Med Sci for selecting delta check Methods based on the ratio of the delta
1990;5:189-95. difference to the width of the reference range can be generally
8. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, applicable for each clinical chemistry test item. Ann Lab Med
Jimenez CV, et al. Desirable specification for total error, 2012;32:345-54.
46 Muller Journal of Medical Sciences and Research | Vol 8 | Issue 1 | January - June 2017