Professional Documents
Culture Documents
High
Performance (Pressure)
Liquid
Chromatography
1
ether Chromatography
colors
chlorophyll
CaCO3
2
Light stone water flow
Heavy stone
base
3
flow
Stationary
phase
4
Chromato (-graphy) : Method
Chromato (-graph) : Instrument
Chromato (-gram) : Picture
Chromato (-grapher) : Person
5
Mobile phase: Liquid
6
Reservoir Pump Injector Column Detector
Data
Processor
7
Oven
Detector
Column
Pump
Degasser Autosampler
Waste
Data processor
Reservoir
8
Problems caused by dissolved air in mobile phase
o Unstable delivery in pump
o Bigger noise and large baseline-drift in detector cell
9
Aspirator Membrane filter
(Size: 0.45 um)
Aspirator
Check valves
Plunger
Plunger seal
11
from Pump from Pump
to Column
to Column
LOAD
from Pump from Pump
to Column
to Column
INJECT
12
Partial-filling
o The volume of the sample loaded is limited to half the
sample loop volume.
Complete-filling
o In order to replace all the mobile phase in the loop,
excess sample (two to five loop volumes) must be
used.
13
from Pump to Column from Pump to Column
Sample Loop
Sample Loop
LOAD INJECT 14
Isocratic system
o Fixed (un-changeable) mixing ratio during analysis
Gradient system
o Changeable mixing ratio during analysis
• HPGE (High Pressure Gradient)
• LPGE (Low Pressure Gradient)
15
Low pressure
gradient unit
Gradient mixer
HPGE LPGE
16
in isocratic mode
Methanol / water = 6 / 4
Long analysis time
Methanol / water = 8 / 2
Poor separations
95%
Methanol concentration
in mobile phase
30%
Short analysis time
&
Excellent separation 18
Flow mode
o Isocratic flow / Binary gradient
Flow rate
o Isocratic flow: A.Flow
o Binary gradient: T.Flow, B.Conc
Pressure
o P.Max, P.Min
19
Air circulation heating
Block heating
o Aluminum block heater
Jacket heating
20
Oven temperature
T.Max
21
UV/UV-VIS
Fluorescence
Refractive Index
22
Cell C : Concentration
Ein Eout
D2 / W Lamps
l
A
A = e·C·l = –log (Eout / Ein)
(A : Absorbance)
C 23
Cell
Grating
D2 / W Lamps
1 nm / element
512 photodiodes
24
Spectrum
Chromatogram
Absorbance
Retention time 25
Excitation wavelength
+ hn 1 *
* hn 2 +
Emission wavelength
Excitation state
Quasi-excitation state
hn1
hn 2
Fluorescence
Ground state 26
Photodiode
Reference
angle
W Lamp
Sample
27
PC workstation software
Integrator)
System controller
28
tR
Peak
tR : Retention time
A : Area
Signal
h h : Height
A
Time
29
1. Adequate sensitivity
2. Good Stability and Reproducibility
3. A linear response to analytes that extends over several orders
of magnitude
4. A short response time that is independent of flow rate
5. High reliability and ease of use
6. Similarity in response toward all analytes or alternatively a
highly predictable and selective response toward one or more
classes of analytes
7. Nondestructive of sample
8. Minimal internal volume in order to reduce zone broadening
30
Material Size
o Stainless steel (SUS) O.D. (Outer Diameter)
o PEEK (Polyether ether ketone) o 1.6 mm
I.D. (Inner Diameter)
o 0.1 mm
o 0.3 mm
o 0.5 mm
o 0.8 mm etc.
31
Male nut (SUS) Ferrule
Ferrule (SUS)
o Pressure: up to 40 MPa
Male nut
32
Water Organic solvent
o Ultrapure water o HPLC grade
o HPLC grade o Super-high grade may be
used in some application.
o Some solvents such as THF
and chloroform include
Stabiliser, which cause a
problem.
33
Un-dissolved solvents Buffer must not be
must not be used in replaced directly with
replacement. organic solvent.
Water Buffer
2-Propanol Water
35
Polarity Solubility
o (+) and (-) charges o Similar solvents can be
exist in a molecule. easily soluble.
Water: polar o Polar/Nonpolar
Methane: nonpolar molecules are similar to
Water/Oil.
H
–
O C
H H H H
+ H
Water methane 36
Stationary phase: Low polarity
o ODS (Octadecyl silane) column
37
C18 (ODS) OH
weak
strong
CH3
38
Mobile phase: Methanol /Water
Methanol / Water
60 / 40
Methanol / Water
70 / 30
Methanol / Water
80 / 20
39
2
tR
N = 16
W
2
tR
= 5.55
H W1/ 2
W1/2
tR H
2
H1/2
= 2 π
W Area
40
Retention time
Spectrum
41
Peak area / Peak height
42
Concentration Area
A1 Calibration curve
C1
A4
A2
Peak area
A3
C2
A2
A3
C3
A1
A4
C4 C1 C2 C3 C4
Concentration 43
Concentration
Internal Area
Target standard
A1 AIS Calibration curve
A2 AIS A3 /AIS
C2 CIS
A2 /AIS
A3 AIS
C3 CIS
A1/AIS
A4 AIS
C4 CIS C1/CIS C2 /CIS C3 /CIS C4 /CIS
Concentration: Target / Internal standard
44
1. It is sensitive
46
• Thus what areas would you consider to be important in
producing an ACCURATE analysis?
47
• There should be no interference in the final analysis from
other components in the sample.
• Results should be correctly calculated and archived for
future reference.
48
1. Sampling and sample storage procedures which ensure
that the sample is truly representative and that it reaches
the laboratory unchanged
2. Sampling and analysis in duplicate
3. Specifications within the analytical scheme for reagent
purity and apparatus cleanliness
4. Repeated checks on the instrument performance
5. Traceability in any standards used. This means that the
stated concentrations in any standard used must be
traceable back to primary international standards
49
6. Inclusion in each analytical batch of additional of
additional samples of known composition. These will
confirm the RELIABILITY of the method and could include
the ff:
a. Blank Samples – samples made up as close as possible in
composition to the unknown, excluding the compound being
determined. These are introduced before stages in the in the
analysis when contamination is likely. A positive determination of
the analyte in the blank would indicate contamination.
51