Catalogue Viscometry

ViscoSystem® | AVS® | ViscoClock | Viscometers


Content viscometry

Xylem Analytics Germany GmbH / SI Analytics GmbH / 독일

Application for AVS measurement systems


Page 3

자동 점도계 : ViscoClock Plus, AVS470, AVS370, AVSPro


Page 5

점도 측정용 Viscosity Bath : CT72 Transparent Thermostats (Bath & Cooler)


Page 27

모세관 점도계 Viscometers


Page 31

피스톤 뷰렛 Titronic (piston burette)


Page 45

Cannon Instrument Company / 미국

자동 점도계 : PolyVISC, MiniPV, MiniPV-HX


Page 49

점도 측정용 Temperature Bath


Page 57

모세관 점도계 Glass Capillary Viscometers


Page 64

점도표준액 Viscosity Standards


Page 77

Digital Vacuum Regulators


Page 83

Pro-Pak
Page 84

1
Ufit.de / 독일

자동점도계 : UVS Basic


Page 85

Bah & chiiler


Page 91

Tamson Instruments / 네델란드

점도측정용 Viscosity Bath


Page 95

2mag AG / 영국

다점용해조 Stirring Heating Blocks


Page 105

2
®
AVS® measurement systems within

AVS
quality assurance systems

Business sector Product Example

Automotive engineering motor oil (fresh and used)

high polymer plastics bumpers

Brewery original wort beer

hop-wort beer

Electrical engineering and electronics high polymer plastics of all types chips, casings

Power supply turbine oil generators

transformer oil

Plastics manufacturers high polymer plastics of all types

Plastics processors high polymer plastics of all types injection molding

Food industry starch instant flour thickeners

gelatin jelly bears

packaging materials yogurt containers

milk products yogurt drink

fruit and fruit juice concentrates

gelatinizing agents pectin

Aviation high polymer plastics of all types

fuels kerosene

hydraulic fluids horizontal stabilizers and undercarriages

Mechanical engineering mold oil mill trains

hardening emulsions stamp shops

hydraulic fluids

Medicine body fluids blood, bile

injection solutions insulin

tinctures and drops nose, eyes

blood substitute materials blood plasma

Mineral oil light motor oil

turbine oil

liquid fuels of all types gasoline, diesel fuel, kerosene (jet fuel)

Textile high polymer plastics of all types for mixed fibers

cotton

Entertainment high polymer plastics CDs, DVDs

The table on the right illustrates the extensive area of high polymer plastics and the large variety of testing methods.

4 3
Polymer applications for the AVS® measurement systems

Polymers, their applications and utilization of our automatic systems

Suitability of the
AVS® measurement systems
Operating
Type Abbr. Solvent Capillary temperature Standards VC* 370 470 Pro
Cellulose C EWN 20 °C SNV 195598
I Cuen (CED) 20 °C DIN EN 60450
Cuen (CED) 20 °C ASTM D 4243
Cuen (CED) 25 °C ISO 5351
Cuen (CED) 25 °C AST; D 1795
Cuen (CED) 25 °C SCAN CM 15:99
Cuen (CED) 25 °C TAPPI T230-0M99
Cellulose CA Dimethyl- 0c 25 °C ASTM D817
acetate chloride/ I
methanol I Micro
Polyamide PA Sulphuric acid II 25 °C ISO 307
(96 %) IIc
Polyamide PA Formic acid I 25 °C ISO 307
(90 %) Ic
Polyamide PA m-cresol II 25 °C ISO 307
IIc
Polybutylene PBT Phenol/dichloro Ic 25 °C ISO 1628-5
terephthalate benzene (50 : 50) II
Polycarbonate PC Dichloromethane 0c 25 °C ISO 1628-4
I
Polyethylene PE Decahydro I 135 °C ISO 1628-3 ASTM D 1601
-naphthalene Ic
Polyethylene PET m-cresol II 25 °C ISO 1628-5
terephthalate IIc ASTM D 4603
IIc Micro
Polyethylene PET Phenol/dichloro Ic 25 °C ISO 1628-5
terephthalate benzene (50 : 50) II ASTM D 4603
Polyethylene PET Dichloroacetic II 25 °C
terephthalate acid IIc Micro
Polymethyl PMMA Chloroform 0c 25 °C ISO 1628-6
methacrylate I
Polypropylene PP Decahydro- I 135 °C ISO 1628-3
naphtalene Ic
Polystyrene PS Toluene I 25 °C
Ic
Polysulphone PSU Chloroform 0c 25 °C

Polyvinyl chloride PVC Cyclohexanone Ic 25 °C ISO 1628-2 ASTM D 1243

Styrene-acrylo- SAN Ethyl methyl 0c 25 °C


nitrile copolymer ketone I
Styrene-butadiene SB Toluene 0c 25 °C
copolymer I

VC* = ViscoClock

excellent suitability; can be used; limited suitability for application related reasons.

This table makes no claim to completeness.

4 5
ViscoCIock plus - Measurement plus data storage

The ViscoCIock plus is an electronic timing unit for glass capillary viscometers used to determine kinematic
and relative viscosity. Succeeding the well-proven ViscoCIock, the new instrument features data storage
and simpler handling. The ViscoClock plus is especially designed for Ubbelohde type viscometers which
are well-known for highest precision.

The ViscoCIock plus


The ViscoClock plus automatically measures the flow time of temperature-stabilized liquids in capillary
viscometers by means of infrared light barriers: the manual measurement with a stopwatch becomes
obsolete.
The viscometer including a sample is inserted into the ViscoClock plus and immersed into a thermostatic
bath for temperature stabilization. After thermostating, the sample is pumped into the measuring bulb,
and the flow time is detected automatically. The large display enables easy read-off of flow times and
additional information: date, time, sample ID and viscometer ID.

ViscoCloc
ViscoClock plus - head

25
Automatic measurement of flow times
The ViscoCIock plus is designed for SI Analytics® Ubbelohde, Micro Ubbelohde and Micro Ostwald
viscometers. The flow time is measured automatically by two infrared light barriers which detect the
passing liquid meniscus. The repeatability of the automatic time measurement is considerably higher in
comparison to the measurement using a stop watch. Therefore some viscometry standards allow a flow
time reduction in case of automatic flow time measurement.

Properties and materials


The ViscoCIock plus can be used for measuring temperatures ranging from -40 °C to 150 °C. The stand of
the ViscoClock plus ist made of high quality polymer PPA. For temperature stabilization in a thermostatic
bath, the following liquids are suitable: Water, alcohol, glycol, paraffin oil, and silicon oil.
The electronic measuring unit is built-in to a PP casing.

Easy handling
The ventilation of Ubbelohde viscometers is managed by an electromechanic valve which makes handling
easier in comparison to the mechanical mechanism of the previous ViscoClock.

Data storage
The measuring results of the ViscoClock plus can be stored on a USB flash drive including date, time and
sample/viscometer ID. The data are stored as pdf (non-editable) and csv (editable). Alternatively, for data
transfer the ViscoClock plus can be connected to a printer (TZ 3863) or a PC.

ck plus Automatic and precise flow time


measurement for a low price
Suitable for SI Analytics® Ubbelohde,
Micro Ubbelohde and Micro Ostwald
viscometers
Data storage incl. time, date, viscome-
ter and sample ID
Stand is made of high performance
engineering plastic PPA and enables
measuring temperatures up to 150°C
Electromagnetic venting valve for
convenient handling of Ubbelohde
viscometer
Compatible with all SI Analytics®
thermostatic bath types

Benefits
ViscoClock plus
63
ViscoCIock plus - The plus for your measurements

Sample and viscometer identification

ViscoClo
To allocate the stored measuring results, the user can enter 2-digit numbers to the ViscoClock plus
before measurement. These IDs – together with date and time – ensure an unambiguous assignment of
the flow times.

Absolute viscosity
To determine absolute kinematic viscosities, calibrated viscometers have to be used. To guarantee best
accuracy, viscometers which were calibrated by automatic measurement should be used. The constant
of automatic calibration can be slightly different in comparison to manual calibration, as the level of the
light barriers may not be identical to the position of timing marks.

Relative viscosity
For determination of relative viscosities, calibrated as well as non-calibrated viscometers can be used.
For evaluation, the calibration constant is not required in this case.

Well arranged:
The new display

The viscometer stand:


high-performance plastic PPA

47
Technical Data - ViscoCIock plus
Measuring range - Time up to 999.99 s; resolution 0.01 s
Accuracy of time ± 0.01 s / ± 1 digit; however no more precise than 0.1 %;
measurement indicated as measuring uncertainty with a confidence level of 95 %
Measuring range - 0.35 to10,000 mm²/s (cSt)
viscosity the absolute, kinematic viscosity is additionally dependent on the uncertainty of the
numerical value of the viscometer constant and on the measuring conditions, in particular
the measuring temperature.
Display LCD grafic display (FSTN) 128x64 pixel, 51x31mm (w x h)
seconds indication with 2 decimal digits after the decimal point, resolution 0.01 s
Voltage supply DC + 9 V
Power supply socket for low voltage connection: coaxial power connector, inner diameter 2.1 mm, plus
pole at inner contact
for connection of Universal power supply TZ 1858
Power supply in accordance to class of protection III
degree of protection for dust and humidity IP 50 in accordance with DIN 40 050
Universal power supply TZ 1858: 100 – 240 V, 50 – 60 Hz (9 V, 550 mA)
not suitable for use in areas subject to explosion hazards
Plug Connections USB Host to connect USB flash drive or printer (TZ 3863)
USB OTG to connect (PC), printer (TZ 3863) or USB flash drive
Type A USB connector
Type B mini USB connector
Ambient Conditions Ambient temperature + 10 to + 40 °C for storage and transport
Operating temperature stand: – 40 to + 150 °C
electronic measuring unit: +10 to +40 °C
Humidity in accordance with EN 61 010, Part 1;
max. relative humidity 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % of relative humidity at a temperature
of 40 °C
Housing Materials stand: polyphthalamide (PPA)
casing*: polypropylene (PP)
gaskets: silicone
Dimensions ~515 x 90 x 30 mm (H x W x D)
Weight ~450 g (without viscometer)
power supply unit: ~220 g
Country of origin Federal Republic of Germany
CE symbol In accordance with low voltage guideline 2014/35/EU
Test regulation EN 61 010-1:2011-07 for laboratory instruments in accordance with EMC
regulation 2014/30/EU
Test regulation EN 61 326 Part1:2012
In accordance with RoHS regulation 2011/65/EU;
Test regulation EN 50 581:2013-02
FCC Symbol
Viscometer types Ubbelohde (DIN; ISO; ASTM; Micro), Micro-Ostwald, manufactured by SI Analytics®
Transparent The ViscoClock plus can be used in all SI Aanyltics bath types
thermostatic baths

* Use in heat carrier liquids can result in discoloration of the synthetic material. The discoloration does not, however, have any effect on the
function and quality of the ViscoClock. Subject to technical changes.

6 8
Precise Capillary Viscometry –
Easy, Flexible and Independent: AVS® 470

No PC needed: “Suction” and “Pressure” measurements Perfectly equipped for fully


with just one instrument automatic viscosity measurements

The AVS® 470 is the first viscosity measurement device that The AVS® 470 is a measuring system that includes almost
allows “suction” and “pressure” measurements completely everything you need to take precise and ­reproducible mea-
independent of a PC. This allows for maximum indepen- surements. All common types of viscosity calculation are

AVS
dence and flexibility; set up a measuring station that meets integrated into the device, a small PS2 keyboard allows you
the highest requirements even under difficult conditions, to enter additional data. A serial printer can be used to con-
e.g. to monitor production or quality control in the polymers veniently document your results.
and mineral oil industry.
So, in a minimum of space, you can set up a measuring sta-
tion equal in every way to complex measuring installations in
terms of precision and reproducibility.

8 9
“Suction” or “Pressure”?
Preferred applications in comparison

“Pressure” “Suction”
highly viscous samples e.g. oils, polymers
Solvents: highly volatile –
(examples) Dichloromethane –
Chloroform –
Sulfuric acid –
Dichloroacetic acid –
Toluene
Hexafluoro-isopropanol
m-cresol –
Formic acid –
Phenol-dichlorobenzene –
Phenol-Tetrachloroethane –

Simple and updateable Modular Concept

The AVS® 470 is of a modular design and an optional optical


or TC version ViscoPump II module. Automatic and highly precize measurements
– independent of a PC
You can use your e
­ xisting accessories such as thermostats,

®
stands, flow-through coolers or automatic cleaners e. g. "Suction" and "pressure" measurements

S   470
AVS® 26. Also, virtually all customary capillary viscometers with the same system
can be used.
Simple data input and parameterization via
included PS2-mini-keyboard

GLP documentation compliant when


connected to an optional printer

­Advantages
AVS® 470
10 9
AVS® 470 – Precise and Reliable

Working with the AVS® 470 is easy If required, individual parameters may be input via an
included PS 2 keyboard. A serial printer can be used to print
The desired measurement method can be preselected and
measurement logs.
started on the device. The entire measurement is automatic
to eliminate subjective measurement errors. Once the set The connections are on the front panel of the device for easy
pre-heating time is reached, the desired number of mea- control. Over-pumping and oversuction are prevented by
surements are taken and the viscometer automatically the use of a an optional capacitive sensor.
cleaned if required.The status of the measurements is con-
tinuously displayed.

Individually
determined readings

The print-out shows every-


thing you need for reliable
documentation of your test.

Indication of
method set

Charge Number
Designation of
specimen
Bild fehlt
User

Readings used
for evaluation

Set maximum
Set equalization permissible deviation
time from average
Average of running
times
Corrected
average

AVS
running time Viscosimeter
constant
Calculated
Viscosity
Operating temperature,
date and time at time
of test

10 11
Technical data

Clear user guidance, clear Measuring range (time) up to 9,999.99 s; resolution 0.01 s
status – even without PC! Measuring range (viscosity)pressure: 0.35 to 1,800 mm²/s (cSt)
suction: 0.35 to ~5,000 mm²/s (cSt)
Measured parameter flow-through time [s]
Time measuring accuracy ± 0.01 %
Measured value display LC-Display
Display accuracy ± 0.01 s, ± 1 Digit, but not exceeding 0.1%
Pumping pressure fully automatically controlled
suction up to ~-160 mbar, pressure up to ~+160 mbar
Preselectable tempering 0 to 20 min
period
Preselectable no. of 1 to 99 for each sample
measurements
Connections Pneumatic connections threaded connections for viscometers
After switching on the AVS® 470 a self test is
Electrical connections circular connector with bayonet lock
run and then an entry prompt appears.
for viscometer
4-pin DIN socket for TC viscometer
4-pin circular connector for capacitive sensor
7-pin circular connector for AVS® 26,
with bayonet lock
RS232-C interface 9-pin for serial printer
Mains connection connector in acc. with EN 60320
Pump connection socket outlet in accordance with EN 60320
Ambient Conditions Ambient temperature +10 to +40 °C for operation and storage
Air humidity max. 80 % in acc. with EN 61010, Part 1
Housing Material steel aluminium housing;
The parameters can be set in the test mode. with chemically resistant 2-component coating
The t0 value is determined automatically. Dimensions (W x H x D) ~255 x 205 x 320 mm
Weight (incl. pump module) ~5.4 kg
Power supply 90 to 240 V ~, 50 to 60 Hz
Equipment safety EMC in acc. with Council Directive 89/336/EWG;
low-voltage directive

The AVS® 470 allows the use of the following viscometers:


Ubbelohde viscometer to DIN, Ubbelohde viscometer to ASTM, micro Ubbelohde viscometer to DIN,
micro Ostwald viscometer, Cannon-Fenske routine viscometer, TC Ubbelohde viscometer,
TC micro Ubbelohde viscometer.

We reserve the right to make technical changes.


All setup parameters can be preset AVS® is a registered trademark of SI Analytics and stands for: "Automatic Viscosity System".
conveniently, e.g. pressure/suction, velocity,
waiting time between two tests, etc.

®
S   470
The readings can be read off conveniently
on the display regardless of whether or not
a printer is connected.

12
AVS® 370 makes maximum precision …

Well equipped for all viscosity determination Now possible for the first time ever: “suction“ and
“pressure“ measure-ment – with one device
The AVS® 370 is a measuring device, which not only mea-
sures as precisely and c­ onsistently as you expect, but also The AVS® 370 is the first viscosity measuring device, which
offers maximum flexibility and future extensions. Further- can be used for both “suction” and “pressure” measurement.
more, it saves ­laboratory space. This enables simple adjustment of the measurement method
for sample. Significantly reducing investment costs for addi-
tional measuring stations at which pressure and suction
methods are to be used. In most cases, using the AVS® 370

®
also saves set-up time.

AVS   37

12 13
… easier and more flexible, with provision for future
expansion!

Easy modular concept ideal for future expansion “Suction” or “pressure”?


A comparison of preferred applications
The AVS® 370 has a modu-lar design. The basic version is
available with one ViscoPump II module in optical or in TC “pressure” “suction”
version. Up to 3 other ViscoPump II modules can be installed Highly viscous samples e.g. oils, polymers
in the compact 19“ housing. The measuring station can be Solvent: highly volatile –
adapted to increasing requirements at any time. (examples)
Dichlormethane –

Can be expanded from an affordable single Chloroform –

measuring station up to an 8-sample station Sulfuric acid –

370
Dichloroethanoic acid –
The basic version of the AVS® 370 is able to measure high or
Toluene
low viscosity liquids. The TC version viscometers, for example,
Hexafluorisopropanol
it is ideal for measuring opaque and black fluids. If neces-
m-Cresol –
sary, each single measuring station can be expanded to a
Formic acid –
multiple measuring station with PC-controlled multitasking.
Phenol-dichlorobenzene –
The WinVisco 370 software included with the standard equip-
Phenol-tetrachloroethane –
ment enables parallel operation of two fully equipped
AVS® 370, with a total of eight ViscoPump II modules. Each
module can measure a different sample using its own
method. All the results can be quickly and easily evaluated
and documented independently. It could hardly be more
flexible!

Compatible with existing accessories

Existing accessories (thermostats, stands, flow through cooler,


etc.) can ­continue to be used with the AVS® 370. Also, virtu-
ally all customary capillary viscometers can be used.

Automatic and highly precise measurements

"Suction" and "pressure" measurements


with the same module

Modular concept for up to four ViscoPump II


modules in one AVS® 370

Each ViscoPump II module in a AVS® 370 can


measure a different sample using a different
method.

Real multitasking for up to eight parallel


measurements with the software WinVisco 370

TC version for measurement of


nontransparent and black liquids

­Advantages
AVS® 370
14 13
® 
AVS  3
AVS® 370 – the right solution for all situations

Anyone working with the AVS® 370 is Measurement of the flow time of the New: Two working principles
perfectly equipped for all tasks liquid’s meniscus can be scanned with the same device.
involved in determining viscosity optoelectronically or with TC sen-
­
With the AVS®  370 you can use one
using capillary viscometers. sors. During optoelectronic scanning
device to work with either “pressure”
the meniscus is detected by glass
How to automatically achieve or “suction” offering more flexibility
light ­fibers, with TC sensors the sen-
the right results with the liquids to be examined.
sor ­detects the different thermal con-
PC-controlled, the AVS®  370 deter- ductivity of the sample and air. The In the “pressure” method an overpres-
mines the time which the liquid to be AVS® 370 offers an extraordinarily sure is ­applied to the liquid in the cap-
examined requires to flow through the broad range of uses, from viscosity illary, this is particularly advantageous
measuring distance in the capillary vis- measurement of clear fluids to black for ­fluids with a low boiling point. For
cometer with quartz precision. The or fully opaque liquids. the “suction” principle the sample is
time is displayed with a resolution of sucked up into the capillary by a
0.01 s (1 digit). vacuum. Greater reproducibility is
­
achieved using the “suction” method
for higher viscosity samples.

14 15
370
Technical data

Working with AVS® 370 is easy Measuring range (time) up to 9,999.99 s; resolution 0.01 s
Measuring range pressure: 0.35 to 1,800 mm²/s (cSt)
The entire measurement procedure is (viscosity) suction: 0.35 to ~5,000 mm²/s (cSt)
place automatic, subjective measuring Measured parameter flow through time [s]
Accuracy of the time ± 0.01 %
errors are reliably eliminated. The PC measurement
starts the measurement. After the set Measured value display via PC
Display accuracy ± 1 digit (0.1 %)
preconditioning period the selected
Pump pressure automatically controlled
number of measurements processed Preselectable 0 to 20 min
and the measured values saved. tempering period
Preselectable number up to 10
of measurements
The system protects against acciden-
Connections Pneumatic connections threaded connections for viscometers
tal overpumping or oversuction by Electrical connections circular connector with bayonet lock for
means of a capacitive sensor. This pre- measuring stands and TC viscometers
RS232-C interface 9-pin
vents the sample to be measured from Mains connections plug in accordance with EN 60320
getting into the vessel containing the Pump connection socket outlet in accordance with EN 60320
liquid or inside the device. Data Input/Output serial to EIA RS232-C
Ambient conditions Ambient temperature + 10 to + 40 °C
Unique flexibility Air humidity max. 85 % rel.
Housing Material coated aluminum plate
In the PC-controlled multiple mea- Dimensions (for 1 to 4 modules) (W x H x D) ~255 x 205 x 320 mm
Weight (incl. 1 module) ~5.4 kg
surement station, the AVS® 370 offers
Power supply 90 to 240 V ~, 50 to 60 Hz
unique flexibility while working in a Equipment safety EMC-Compatibility according to the Directive 89 / 336 / EEC of the Council;
very small space: Up to eight modules, low-voltage directive according to the Directive 73 / 23 / EEC of the Council,
as amended by the Directive 93 / 68 / EEC of the Council
which equates to two fully equipped
Multi-tasking for 1 to 8 ViscoPump II modules, with WinVisco 370 software
AVS® 370, can be run in parallel with
the WinVisco 370 software. Each The following viscometers can be used with the AVS® 370:
module can measure the same or dif- Ubbelohde viscometer to DIN, Ubbelohde viscometer to ASTM, micro Ubbelohde viscometer to DIN,
micro Ostwald viscometer, Cannon-Fenske routine viscometer, TC-Ubbelohde viscometer,
ferent samples using “pressure” or TC-micro Ubbelohde viscometer.
“suction”, independent of each other.
We reserve the right to make technical changes.
In this way, a series of measurements AVS® is a registered trademark of SI Analytics and stands for: "Automatic Viscosity System".

can be prepared quickly and immedi-


ately evaluated and documented with
the computer. This significantly redu-
ces the time required to carry out vis-
cosity measurements, especially for in
process controls and quality assu-
rance.

16 15
Real multitasking for up to 8 measurements in parallel mode …

Easy to understand and proven: The


WinVisco 370 software

WinVisco 370 is the ideal software for


the AVS® 370*). It is supplied as part of
the standard equipment. Up to eight
viscosity measurement modules can
be controlled with only a few opera-
tional steps. The device parameters
are easy to enter: Constants, t0 flow
time, number of measurements, pre-
conditioning period, type of viscome-
ter, date and sample labeling for each
measurement station.

WinVisco 370 works in a real multitask-


ing mode making it possible for each
All the important parameters required for the measurement are displayed on the “Methods/
measurement to be processed inde-
Results” page. If necessary, the parameter editor can be called up using “Add Parameter”, in
pendently from the others. It also order to enter non-standard or customer specific formulae.
means that time-consuming measure-
ments can be carried out from the
same PC, without hindering the course
of other, faster measurements. During
the measurements you can change
the monitor displays, start or stop
other measurements, print or save
measured values. All data provided by
the software can be passed on to LIMS
system.
All the measurements currently running can be
WinVisco 370 supports three groups monitored in parallel in the overview.
of users. For simple use, access is lim-
ited to: select viscometer, measure,
load and save methods as well as
enter parameters. In the highest level,
users with administrator status can
access all the software’s facilities. Each
user is given a user ID, an access level

A
and a password.
*) The language (English or German) can
be chosen after installation over the
program menu.
The viscometer data required for the evaluation
can be stored in a table. This guarantees perfect
allocation of e.g. the t0 runtime, viscometer
constants, the series number, etc. for each
individual viscometer being used.

17
… with the proven WinVisco 370 software

With AVS® 370 and WinVisco 370 the right connection for rinsing can
be quickly determined

With the daisy chain link of the AVS® 370, additional devices can be inte-
grated with the system and controlled using the WinVisco 370 software. For
example, when working in suction mode the viscometers can be rinsed using
the ­TITRONIC®300 or the TITRONIC® 500 burettes. The TITRONIC®300 is
preferable used for light solvents, the TITRONIC® 500 for solvents with a vis-
cosity >3 mm²/s. For highly aggressive solvents special changeable modules
are available (TA50V and WA50V).

The password protection prevents unwanted A vacuum pump (accessory) integrated with the system can be used to con-
changes to the important measurement param- veniently remove samples and solvents.
eters.

Vacuum pump

PC AVS® 370 TITRONIC® 500 TITRONIC® 300

The parameters can be individually adjusted to


the measurement for each measuring position.

Printer

Transparent thermostat
with four measuring stands

Two basic concepts are available for the rinsing:

•  An AVS® 370 with up to four ViscoPump II modules (max. four measurement
positions) and up to eight burettes, which enables each viscometer to be
rinsed with two solvents. Time consuming removal of the transparent ther-

®
Each rinse/dry step can be individually prese-
lected. Even the application dependent quantity mostat for external rinsing of the viscometer is no longer necessary.

AVS   370
of solvent and the drying time can be separately
determined. •  Two AVS® 370 complete with up to four ViscoPump II modules each (max.
eight measuring positions), which enables semi-automatic rinsing of the vis-
cometer with the next sample or solvent.

18
Ordering information AVS® 470

AVS
Ordering information AVS® 370

®
AVS   370
18 19
®
S   470
The AVS® 470 viscosity test station is composed of individual components.
Please request a detailed quote.

Type no. Order no. Description


AVS® 470 basic unit 285415709 AVS® 470 basic unit, housing incl. one ViscoPump II module for opto-electronic sensing, Keyboard
for opto-electronic Version: 95 V to 230 V/50-60 Hz
sensing

AVS® 470 basic unit 285415708 AVS® 470 basic unit, housing incl. one ViscoPump II module for TC sensing, Keyboard
for TC sensing Version: 95 V to 230 V/50-60 Hz

VZ 8511 1054306 ViscoPump II module for optical sensing

VZ 8512 1054304 ViscoPump II module for TC sensing

The AVS® 370 viscosity test station is composed of individual components.


Please request a detailed quote.

Type no. Order no. Description


AVS® 370 basic unit 1056509 AVS® 370 basic unit, housing incl. one ViscoPump II module and WinVisco 370 software, for opto-electronic sensing
for opto-electronic
sensing

AVS® 370 basic unit 1056515 AVS® 370 basic unit, housing incl. one ViscoPump II module and WinVisco 370 software, for TC sensing
for TC sensing

VZ 8511 1054306 ViscoPump II module for optical sending

VZ 8512 1054304 ViscoPump II module for TC sending

Accessories AVS® 470 and AVS® 370

Type no. Order no. Description


CT 72/P, 230V 285418526 Immersion thermostat 230 V and thermostatic bath (acrylic glass container with two manual gauge slides),
basic configura­tion for the attachment of one flow-through cooler.
CT 72/P, 115V 285418513 Immersion thermostat 115 V and thermostatic bath (acrylic glass container with two manual gauge slides),
basic configura­tion for the attachment of one flow-through cooler.
CT 72/2, 230V 285418547 Immersion thermostat 230 V and thermostatic bath (stainless steel container with one manual gauge slide),
basic configura­tion for the attachment of one flow-through cooler.
CT 72/2, 115V 285418532 Immersion thermostat 115 V and thermostatic bath (stainless steel container with one manual gauge slide),
basic configura­tion for the attachment of one flow-through cooler.
CT 72/4, 230V 285418568 Immersion thermostat 230 V and thermostatic bath (stainless steel container with two manual gauge slides),
basic configuration for the attachment of one flow-through cooler.
CT 72/4, 115V 285418554 Immersion thermostat 115 V and thermostatic bath (stainless steel container with two manual gauge slides),
basic configuration for the attachment of one flow-through cooler.
Z 900 285225620 RS232-C Data printer (230 V)

0
Measuring stand 285410502 Metal measuring stand AVS®/S, preferably for nonaqueous bath fluids
AVS®/S
Measuring stand 285410876 PVDF measuring stand AVS®/SK, corrosion-free, suitable for aqueous and nonaqueous bath fluids
AVS®/SK
Measuring stand 285410892 PVDF measuring stand AVS®/SK-CF, particularly for the use of Cannon-Fenske routine viscometers
AVS®/SK-CF
Measuring stand 285410905 PVDF measuring stand AVS®/SK-V, particularly for the use of dilution viscometers
AVS®/SK-V
CK 300, 115V 285414331 CFC-free flow-through cooler for enhancing the temperature constancy of the bath fluid (according to configuration and
environmental conditions are ± 0.02 K possible) or for measurement at room temperature or below (min. + 5 °C).
CK 300, 230V 285414348 CFC-free flow-through cooler for enhancing the temperature constancy of the bath fluid (according to configuration and
environmental conditions are ± 0.02 K possible) or for measurement at room temperature or below (min. + 5 °C).
05392 285405043 Fixing frame for Ubbelohde viscometers (not TC)

20
Automatic viscosity measurement has been improved …

The AVS® Pro III automatic sampler is a fully automated The ProClean system and the microdosing make routine
instrument for determining the viscosity of Newtonian fluids operation safer. The filtration of solutions, which occasionally
with capillary viscometers. In spite of the high sample may be harmful, may be ommitted. The capacitive sensors in
throughput, the AVS® Pro III provides maximum accuracy the suction pipe effectively prevent any damage of the
and reproducibility. Furthermore, working with the auto- measurment system.
matic sampler is easy and allows unattended 24-hour opera-
The AVS® Pro III automatic sampler works with the capillary

®
tion.
method, which is the most precise method for determining the

AVS  P
Particularly with time consuming measurement runs, the viscosity of Newtonian liquids in terms of physical chemistry.
AVS® Pro III helps to substantially reduce the burden on The great versatility offered by vis-cometers with optical and
qualified users. The fully automatic mode offers an increased TC sensing systems opens up an extremely wide range of
level of safety when handling aggressive media, e.g. sulph- applications. This includes measurements of clear liquids as
uric acid. well as opaque petroleum products.

20 21
… with the AVS® Pro III Automatic Sampler:

Two different sample racks The viscosity measurement requirements of the polymer
are supplied:
and petroleum industries in particular have been incorpo-
a) one rack with 56 positions
for 20 ml sample bottles for rated into the design of the AVS® Pro III. The main feature of
micro-viscometer applica- the automatic unit is the three-axis positioning mechanism
tions
of the sample dosing system. The X-Y-Z positioning mecha-
nism allows operation of up to four ­Micro TC viscometers in
two thermostatic baths, which can be set at two different
­measurement temperatures. This method is used in oil indus-
try in order to ­determine the viscosity index.

The AVS® Pro III allows the operator to select the sample
sequence and which sample is to be filled into which viscom-
eter. The dosing system is available in either normal or micro
construction and operates without a valve. It is thus suitable
for nearly any type of sample.

The AVS® Pro III is equipped with opto-electronic and TC


b) one rack with 16 positions
to accommodate 100 ml
scanning (TC = thermal conductivity method) functions for
sample bottles for normal the meniscus passage in the capillary viscometer. The sam-
volume applications ples are positioned in the easy-to-load sample rack, using
the motorized lifting mechanism. If needed, the rack can be
temperature-regulated.

The electric sample lift


ensures positioning of the
samples in the rack at a Fully automatic and highly precise measuring
convenient and easily
monitored working height. station. Time measurement with a precision of
+/- 0.01s (but less precise than 0.1%)

Ideal for highly aggressive media

Although in combination of optical and thermical

Pro III
sampling of the meniscus channel or different
capillary sizes and types, up to four viscometers
selectable

The ProClean system and micro-dosing


elminate dangerous manual filtration
of the sample

­Advantages
AVS® Pro III
22
Working with the AVS® Pro III is …

The AVS® Pro III is controlled by a PC, connected via RS232- During the entire process, all of the parameters (depending
C interface. The intuitive user interface guides the user on the menu level) and the respective status of the individual
clearly through the program. All data inputs are made using measuring positions, the temperature regulation system and
the computer keyboard and mouse. the sample transfer system are either visible or can be
selected.
A faulty operating status is indicated by acoustic or optical
signals such as ­arrows, icons and other status messages or The operator interface of the AVS® Pro III is available in Ger-
request messages. During the entire work sequence, the man and English. Commercially available printers for which
respective status of the AVS® Pro III is documented on the Windows drivers are available are suitable for documenta-
computer screen. Furthermore, status indicators can be tion purposes. ­
selected for­each i­ndividual measuring position, which pro-
Precision, reproducibility and comparability comply with
vide additional information on operation.
DIN 51 562-1(1999-01), ASTM D 445 and ISO 3105 standards.
For the respective type of measurement, pre-parameterized
The AVS® Pro III is built in accordance with international equip-
sets of para­meters depending on the viscometers, tempera-
ment safety standards: CE symbol (equipment safety, low volt-
ture and other measurement criteria are already provided. In
age safety, emitted interference and interference immunity).
addition, all parameters can be individually adjusted to spe-
cial requirements at a special menu level. All of the standard If requested, the AVS® Pro III automatic sampler can be sup-
calculation methods are available. plied with a manufacturer’s inspection certificate based on
direct comparison with normal viscometers of the first order
The proved and tested AVS® Pro III software also makes it
in accordance with DIN 51 562 - 4: 1999-01.
possible to prepare additional individually selected calcula-
tions, such as:

• mean value,
• standard deviation,
• outlier test (A %),
• Hagenbach correction,
•  
absolute viscosity, dynamic viscosity (density value
required),
• viscosity index (measurement at two temperatures required),
• SUS and SFS,
• relative viscosity,
• specific viscosity,
• reduced viscosity (viscosity number),
• inherent viscosity
• intrinsic viscosity and
›16 sample rack‹
• K-value after Fikentscher

®
The AVS® Pro III allows individual allocation between the

AVS   Pr
characteristics of the sample and the viscometers that are
currently in operation.

22 23
… easy, reliable and safe

This means that it is not only possible to simultaneously test


the characteristics of samples with greatly differing viscosi-
ties, but also to perform measurements in various capillary
sizes and different types of viscometers. This even applies to
a combination of optical and thermal scanning. Therefore,
preliminary sorting of the samples with regard to viscosity
and the size of capillary required for the testing process is
not necessary.

It is possible to “individually” allocate each sample to a capil-


lary viscometer that is currently being used by means of the
conventional MS-Windows® “drag and drop” method.

The allocation between the sample and the viscometer is


shown on the status display. ›options‹

This mode is used to specify the number of mea­sure­ments,


the preliminary temperature regulation period, the allow-
able standard deviation, the maximum allowable tempera-
ture tolerance, the rinsing type and method of the viscome-
ter.

›selection of method‹

This mode is used to specify what monitoring pa-rameters ›dosing parameters‹


are to be activated, e.g. if the temperature control of the
thermostats is supposed to be handled via the PC. This mode is used to specify the filling quantity of the vis-
cometer, the dosing speed depending on the viscosity and
the type of rinsing.

ro III 24
Technical data AVS® Pro III

Sampling system Sample bottles

Sample rack
100 ml screw-type and bottles with standard ground joint (16 pcs per rack)

20 ml round bottom glass pieces (56 pcs. per rack)

for 100 ml screw-type and bottles with standard ground joint


AV
for 100 ml screw-type and bottles with standard ground joint (temperature controlled up to
135°C)

for 20 ml round bottom glass pieces

Measured value recording Method meniscus scanning by means of opto-electronic system or thermal conductivity (TC)

Measuring parameter throughput time in seconds [s]

temperature in degrees Celsius [°C]

Calculated parameters mean value, standard deviation, outlier test (A %), Hagenbach correction, absolute viscosity, dynamic viscosity (knowledge of
density required), viscosity index (measurement at two temperatures required) SUS and SFS, relative viscosity, specific viscosity,
reduced viscosity (viscosity number), inherent viscosity, K-value

Selection parameters by means of PC keyboard, mean value, standard deviation, outlier test (A %), Hagenbach correction, absolute viscosity, dynamic
viscosity (knowledge of density required), viscosity index (measurement at two temperatures required) SUS and SFS, relative
viscosity, specific viscosity, reduced viscosity (viscosity number), inherent viscosity, K-value, rack position, date/time, temperature
regulation period, number of measurements, number of rinsing operations, start, stop/reset

Number of measurements 1 to 99

Temperature regulation period 0 to 99 min., selectable in increments of 1 min.

Number of Viscometer tests 0 to 9 with next sample (observe sample quantity) or with preselected rack position

Data memory by means of PC

Viscosity measurement range 0.35 to 1,200 mm²/s (at room temperature of samples)

Time up to 9999.99 s, resolution = 0.01 s

Vacuum pressure automatically controlled

Viscometers available for use Ubbelohde viscometer in accordance with DIN standards

Ubbelohde viscometer in accordance with ASTM standards

Micro-Ubbelohde viscometer in accordance with DIN standards

Micro-Ostwald viscometerCannon-Fenske-Routine visco

Cannon-Fenske-Routine viscometer

TC Ubbelohde viscometer

TC Micro-Ubbelohde viscometer

25
®
VS   Pro III
Measuring accuracy ± 0.01 s ± 1 digit, but not more precise than 0.01 %

The measuring uncertainty for measurements of absolute kinematic viscosity is also dependent on the uncertainty of the numeric
value for the viscometer constant and on the measuring conditions, especially the measuring temperature.

Evaluations / results Correction Hagenbach correction (HC) for Ubbelohde, Cannon-Fenske-Routine, Micro-Ubbelohde and
Micro-Ostwald viscometers

Statistical evaluation standard deviation, outlier search

Ambient conditions Ambient temperature 10 to + 40 °C

Air humidity max. 85 % relative humidity

Equipment safety CE-symbol in accordance with Guideline 89 / 336 / EEC of the Council (EMC compatibility)

in accordance with Standard EN 50 081, Part 1;

interference immunity in accordance with Standard EN 50 082, Part 2;

in accordance with Guideline 73 / 23 / EEC of the Council (low-voltage guideline)

Housing plastic/stainless steel / aluminium casing with chemically resistant two-component coating of the plastic pieces

Dimensions w = 1,300 mm, h = 1,000 mm, d = 620 mm (~51“ x 43“ x 24“)

Weight dependent on the number of measuring positions

~70 kg

Connections Pneumatic connections screw-type connections for viscometer

Electric connections circular connectors with bayonet lock for measuring stand and TC viscometer

Viscometers up to 8 viscometers connected by individual control units

Temperature via serial interface RS232-C of suspended thermostat,

type: 1 pc. CT 72/4 or up to 2 pcs. CT 72/2

Interfaces control system using PC with 2 x RS232-C interfaces

Safety overfilling safety device of waste bottle and suction hose

Mains connection European built-in plug DIN 49 457 6 with fuse

Data transmission Interface internal bidirectional serial interface in accordance with EIA RS232-C (daisy chain concept)

Interface external via PC, bidirectional serial interface in accordance with EIA RS232-C

Power supply Mains voltage 230 V (AC) or 115 V (AC), 50 to 60 Hz (AC)

26
CT 72 Thermostat Series –
Transparent Thermostats conforming to Standards

CT 72
CT 72/4

As their predecessor CT 52 the new transparent thermostats The SI Analytics transparent thermostats are particularly
CT 72/P, CT 72/2, CT 72/2-TT and CT 72/4 meet DIN 51 562 designed for the determination of the viscosity of newtonian
(part 1), ASTM D 445 and ISO 3105 standards. liquids in glass capillary viscometers. They may be adapted
for manual as well as for automatic measurements. The new
models offer the core features of the CT 52 series incorporat-
ing the new improved immersion thermostats.

26 27
Bright display shows
ongoing process,
at any time.

Programmable set temperatures through inte-


CT 72/2
grated clock with controller.

Display of the momentary and the set


­temperature.

Increased safety from separate operation and


temperature safety sensors.

Over-temperature safety system


on front panel.

CT 72/2 and CT 72/4 can be used up to 150 °C. Automatic fuses on the back panel.
High temperature version is standard.
Output of data formats
Draining valve comes with via RS232 connection.
CT 72/2, CT 72/2-TT and
CT 72/4. New
Advantages immersion
baths thermostats
28
Transparent Thermostats conforming to Standards:
The CT 72 series

CT 72/P
CT 72/4

CT 72/2
CK 310

Suitable temperature control liquids

Liquid Alcohol Water Paraffine oil Silicon oil


Temperature range – 40 °C to + 10 °C + 5 °C to +80 °C + 40 °C to +150 °C + 80 °C to +150 °C

Technical data

Working temperature
Measuring positions for AVS
Measuring positions TC
Measuring positions micro TC
Temperature stability according
DIN 58 966 at 25 °C
Size (W x H x D in mm)
Filling volume
Material
CT 72/P

2
2
2
± 0.01 K
CT 72
+ 10 °C to +60 °C

355 x 370 x 250


18 l
PMMA
CT 72/2-TT
– 40 °C to + 150 °C
2
2
2
± 0.01 K

355 x 370 x 250


15 l
St. steel & glass
CT 72/2
+ 5 °C to +150 °C
2
2
2
± 0.01 K

355 x 370 x 250


15 l
St. steel & glass
CT 72/4
+ 5 °C to + 150 °C
4
4
4
± 0.01 K

605 x 370 x 250


27 l
St. steel & glass
Weight (empty) ~5 kg ~14 kg ~13.5 kg ~28 kg

At applications within normal temperature range (+ 5 °C up to approx. + 40 °C) cooling will be necessary for maintaining temperature stability.

28 29
Ordering information
Type no. Order no. Description
CT 72/P, 230V 285418526 Immersion thermostat 230 V and thermostatic bath (acrylic glass container with two manual gauge slides), basic configura­tion for
the attachment of one flow-through cooler.
CT 72/P, 115V 285418513 Immersion thermostat 115 V and thermostatic bath (acrylic glass container with two manual gauge slides), basic configura­tion for
the attachment of one flow-through cooler.
CT 72/2, 230V 285418547 Immersion thermostat 230 V and thermostatic bath (stainless steel container with one manual gauge slide), basic configura­tion for
the attachment of one flow-through cooler.
CT 72/2, 115V 285418532 Immersion thermostat 115 V and thermostatic bath (stainless steel container with one manual gauge slide), basic configura­tion for
the attachment of one flow-through cooler.
CT 72/2 - M, 230V 285418584 Immersion thermostat 230 V and thermostatic bath (stainless steel container with one manual gauge slide), equipped with two
magnetic stirrer positions. Basic configuration for the attachment of one flow-through cooler.
CT 72/2 - M, 115V 285418593 Immersion thermostat 115 V and thermostatic bath (stainless steel container with one manual gauge slide), equipped with two
magnetic stirrer positions. Basic configuration for the attachment of one flow-through cooler.
CT 72/2 - TT, 230V 285418615 Immersion thermostat 230 V and thermostatic bath (stainless steel container with one manual gauge slide), basic configuration
for the attachment of one flow-through cooler.
CT 72/2 - TT, 115V 285418607 Immersion thermostat 115 V and thermostatic bath (stainless steel container with one manual gauge slide), basic configuration
for the attachment of one flow-through cooler.
CT 72/4, 230V 285418568 Immersion thermostat 230 V and thermostatic bath (stainless steel container with two manual gauge slides), basic
configuration for the attachment of one flow-through cooler.
CT 72/4, 115V 285418554 Immersion thermostat 115 V and thermostatic bath (stainless steel container with two manual gauge slides), basic
configuration for the attachment of one flow-through cooler.
CT 72/E, 230V 285418501 Immersion thermostat 230 V/50 Hz

CT 72/E, 115V 285418495 Immersion thermostat 115 V/60 Hz

More Accessories and spare parts

CK 300, 230V 285414348 Flow through cooler, 230 V

CK 300, 115V 285414331 Flow through cooler, 115 V

CK 310, 230V 285414320 Flow through cooler, 230 V, stainless steel version

CK 310, 115V 285414310 Flow through cooler, 115 V, stainless steel version
VZ 5210 1007628 CT 72 retrofit set for CT 62-thermostatic bath,
contains: Immersion thermostat CT72/E-230 V, adapter plate and cooling devices
VZ 5213 285420397 CT 72 retrofit set for CT 62-thermostatic bath,
contains: Immersion thermostat CT72/E-115 V, adapter plate and cooling devices
VZ 5402 285415171 Manual gauge slide for transparent thermostats
VZ 5403 285420684 3-fold manual gauge slide for transparent
thermostats
VZ 5404 285418573 Dust protection cover for transparent thermostat

VZ 5405 285418620 Transparent thermostatic bath backlight


VZ 7100 285421051 Control thermometer measuring range
+ 19 to + 21 °C
VZ 7101 285421068 Control thermometer measuring range
+ 24 to + 26 °C
VZ 7102 285421076 Control thermometers measuring range
+ 29 to + 31 °C
VZ 7103 285421084 Control thermometers measuring range
+ 39 to + 41 °C
VZ 7104 285421092 Control thermometers measuring range
+ 99 to + 101 °C
VZ 7105 285421105 Control thermometers measuring range
+ 134 to + 136 °C

30
Viscometers and their range of use

Viscometer type

w
flo
e
tin

e
rs
ou

ve
e

-R

re
hd

ke

e
e

sk
o

hd

ns
al
el

en
e

Fe
o
bb
hd

st

-F
el

n-
d

O
U

bb

on
o

al

o
el

ro

ro

nn
w

nn
U
bb

ic

ic
st
Measurement substance property

TC

Ca

Ca
M

M
O
U

Transparent liquids
++ ++ - + + + o
manual measurement

Transparent liquids
++ ++ + - + + -
automatic measurement

Ubbe
Opaque liquids
- - - - - - +
manual measurement

Opaque liquids
- - ++1) - - - -
automatic measurement

Foaming liquids o o o + + + o

Liquid mixture with


o o o + + + o
highly volatile components

Minimum measurement substance


- ++ - - + - -
and/or rinsing agent quantities

High-temperature or low-
+ + + o o o o
temperature measurements

Selection of glass capillary viscometers


++ use preferably 1) up to 30,000 mm²/s
+ highly suitable 2) above 30,000 mm²/s
o less suitable
- unsuitable

31
Ubbelohde viscometers, normal form (DIN)

Viscometers with suspended level for can also be checked by means of man-
determination of absolute and relative ual measurements. The recommended
kinematic viscosity of liquids with Newto- minimum flowthrough time is 200s for
nian flow behavior. The calibrated viscom- absolute measurements of kinematic
eters are delivered with manufacturer’s viscosity. For relative measurements
certificate in accordance with DIN 55 350, (polymer analytics), the minimum flow
Part 18. All viscometers are provided time of 50s is allowed by ISO 1628-1
with ring marks. This ensures that vis- (depending on the capillary size).
cometers for automatic measurements

Ubbelohde-Viskosimeter (DIN)

- in accordance with DIN 51 562 Part 1, ISO/DIS 3105 (BS-IP-SL)


- filling quantity: 15 to 20 ml

elohde
- overall length: approx. 290 mm

calibrated, calibrated ν = K · t
with constant, with constant, K = ν
t
for manual measurements for manual measurements;
automatic measurement t = ν
K
with stand AVS®/SK-HV

Capillary No. Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Type No. Order No. acc. DIN Ø i ± 0,01 [mm] (approx.) (approx.)
501 00 285400004 – – 0 0.36 0.001 0.3 to 1
501 03 285400012 – – 0c 0.47 0.003 0.5 to 3
501 01 285400029 – – 0a 0.53 0.005 0.8 to 5
501 10 285400037 – – I 0.63 0.01 1.2 to 10
501 13 285400045 – – Ic 0.84 0.03 3 to 30
501 11 285400053 – – la 0.95 0.05 5 to 50
501 20 285400061 – – II 1.13 0.1 10 to 100
501 23 285400078 – – IIc 1.50 0.3 30 to 300
501 21 285400086 – – lla 1.69 0.5 50 to 500
501 30 285400094 – – III 2.01 1 100 to 1,000
501 33 285400107 – – IIIc 2.65 3 300 to 3,000
501 31 285400115 – – IIla 3.00 5 500 to 5,000
501 40 285400123 – – IV 3.60 10 1,000 to 10,000
– – 502 43 285400131 IVc 4.70 30 3,000 to 30,000
– – 502 41 285400148 IVa 5.34 50 6,000 to 30,000
– – 502 50 285400156 - 6.30 100 > 10,000

not calibrated, calibrated, ν = K · t ν = kinematic viscosity in mm²/s


without constant; with constant for K = ν K = constant [mm²/s]
t
for determination of automatic measurements t = flow-through time in s
relative viscosity t = ν
K

Capillary No. Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Type No. Order No. acc. DIN Ø i ± 0,01 [mm] (approx.) (approx.)
– – 532 00 285400164 0 0.36 0.001 0.3 to 1
530 03 285400304 532 03 285400201 0c 0.47 0.003 0.5 to 3
530 01 285400312 532 01 285400218 0a 0.53 0.005 0.8 to 5
530 10 285400329 532 10 285400226 I 0.63 0.01 1.2 to 10
530 13 285400337 532 13 285400234 Ic 0.84 0.03 3 to 30
– – 532 11 285400172 la 0.95 0.05 5 to 50
530 20 285400345 532 20 285400242 II 1.13 0.1 10 to 100
530 23 285400353 532 23 285400259 IIc 1.50 0.3 30 to 300
– – 532 21 285400189 lla 1.69 0.5 50 to 500
530 30 285400361 532 30 285400267 III 2.01 1 100 to 1,000
530 33 285400378 532 33 285400275 IIIc 2.65 3 300 to 3,000
– – 532 31 285400197 IIla 3.00 5 500 to 5,000
530 40 285400386 532 40 285400283 IV 3.60 10 1,000 to 10,000

32
Ubbelohde viscometers, normal form (ASTM)

Ubbelohde Viscometer (ASTM)

• in accordance with ISO 3105, ASTM D 2515, ASTM D 446


• filling quantity: 15 to 20 ml
• overall length: approx. 285 mm

calibrated, not calibrated, calibrated,


with constant for manual without constant for with constant for
measurements determination of relative automatic measurements
Viscosity

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Type No. Order No. Type No. Order No. Capillary No. Ø i ± 0,01 [mm] (approx.) (approx.)
525 00 285400501 526 00 285400707 527 00 285401255 0 0.24 0.001 0.35 to 1
525 03 285400518 526 03 285400715 527 03 285401271 0c 0.36 0.003 0.6 to 3
525 01 285400526 526 01 285400723 527 01 285401263 0b 0.46 0.005 1 to 5
525 10 285400534 526 10 285400731 527 10 285401152 I 0.58 0.01 2 to 10
525 13 285400542 526 13 285400748 527 13 285401169 Ic 0.78 0.03 6 to 30

Ubbelo
525 20 285400559 526 20 285400756 527 20 285401177 II 1.03 0.1 20 to 100
525 23 285400567 526 23 285400764 527 23 285401185 IIc 1.36 0.3 60 to 300
525 30 285400575 526 30 285400772 527 30 285401193 III 1.83 1 200 to 1,000
525 33 285400583 526 33 285400789 527 33 285401288 IIIc 2.43 3 600 to 3,000
525 40 285400591 526 40 285400797 527 40 285401296 IV 3.27 10 2,000 to 10,000
525 43 285400604 526 43 285400801 527 43 285401309 IVc 4.32 30 6,000 to 30,000

33
Ubbelohde viscometers,
with additional tube and threads

Viscometers with suspended level for ered with manufacturer’s certificate in


determination of absolute or relative accordance with DIN 55 350, Part 18.
kinematic viscosity. These viscometers The ring marks present serve as auxil-
are preferably used for automatic iary marks in case the viscometers
measurements. The additional filling must be checked by manual measure-
and cleaning tube and the glass ments.
thread ensure safe operational use.
The calibrated viscometers are deliv-

Ubbelohde viscometer (DIN)

• in accordance with ISO 3105, DIN 51 562, Part 1, BS 133, NFT 60-100
• filling quantity: 18 to 22 ml
• overall length: approx. 290 mm

calibrated,
with constant for automatic
measurements

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Capillary No. Ø i [mm] (approx.) (approx.)
541 03 285401925 0c 0.47 0.003 0.5 to 3
541 01 285401917 0a 0.53 0.005 0.8 to 5
541 10 285401933 I 0.63 0.01 1.2 to 10
541 13 285401941 Ic 0.84 0.03 3 to 30
541 11 285401950 Ia 0.95 0.05 5 ... 50
541 20 285401958 II 1.13 0.1 10 to 100
541 23 285401966 IIc 1.50 0.3 30 to 300
541 21 285408719 IIa 1,69 0,5 50 ... 500
541 30 285401974 III 2.01 1 100 to 1,000
541 33 285401982 IIIc 2.65 3 300 to 3,000

ohde
541 40 285401999 IV 3.60 10 1,000 to 6,000

Ubbelohde viscometer (ASTM)

• the technical measurement characteristics are in accordance with ISO 3105,


ASTM D 2515, ASTM D 446
• filling quantity: 15 to 22 ml
• overall length: approx. 290 mm

calibrated,
with constant for automatic measurements

Capillary No. Capillary Constant K Measuring range [mm²/s]


Type No. Order No. acc. DIN Ø i [mm] (approx.) (approx.)
545 00 285402005 0 0.24 0.001 0.35 to 1
545 03 285402021 0c 0.36 0.003 0.6 to 3
545 01 285402013 0b 0.46 0.005 1 to 5
545 10 285402038 I 0.58 0.01 2 to 10
545 13 285402046 Ic 0.78 0.03 6 to 30
545 20 285402054 II 1.03 0.1 20 to 100
545 23 285402062 IIc 1.36 0.3 60 to 300
545 30 285402079 III 1.83 1 200 to 1,000
545 33 285402087 IIIc 2.43 3 600 to 3,000
545 40 285402095 IV 3.27 10 2,000 to 10,000
545 43 285402108 IVc 4.32 30 6,000 to 30,000

34
Micro-Ubbelohde viscometers
Viscometers for dilution viscometry

Viscometers with suspended level for Micro-Ubbelohde viscometers (DIN)


determination of absolute and relative
The calibrated viscometers are deliv-
kinematic viscosity of liquids with
ered with manufacturer’s certificate in
Newtonian flow behavior. Due to their
accordance with DIN 55 350, Part 18.
design, these viscometers are espe-
For measurements with automatic vis-
cially suitable for measurement of
cosity measuring instruments, another
small liquid quantities and for particu-
constant is valid. This constant is deter-
larly short running times. All viscome-
mined by multiplication of the con-
ters are provided with ring marks. This
stant K with the correction factor F.
ensures that viscometers for automatic
measurements can also be checked • in accordance with DIN 51562, Part 2

cro-
by means of manual measurements.
• filling quantity: 3 to 4 ml

• overall length: approx. 290 mm

calibrated, calibrated, not calibrated,


with constant for manual with constant for without constant;
measurement automatic measurement for determination of
relative viscosity

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Type No. Order No. Type No. Order No. Capillary No. Ø i [mm] (approx.) (approx.)
536 10 285401009 537 10 285401103 538 10 285401206 MI 0.40 0.01 0.4 to 6
536 13 285401017 537 13 285401111 538 13 285401214 M Ic 0.53 0.03 1.2 to 18
536 20 285401025 537 20 285401128 538 20 285401222 M II 0.70 0.1 4 to 60
536 23 285401033 537 23 285401136 538 23 285401239 M llc 0.95 0.3 12 to 180
536 30 285401041 537 30 285401144 538 30 285401247 M lll 1.26 1 40 to 800

elohde
Viscometers for dilution viscometry

Viscometers with suspended level our ­piston burettes TITRONIC®  300 or


designed according to the principle of TITRONIC® 500.
the Ubbelohde viscometers for
• filling quantity: 15 to 75 ml
determination of the limit viscosity
­
number of polymers. The limit viscos- • overall length: approx. 290 mm
ity ­number can be determined auto-
matically in combination with one of

calibrated, for automatic measurements,


Model with glass filter on request

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Capillary No. Ø i [mm] (approx.) (approx.)
531 00 285401403 0 0.36 0.001 0.35 to 0.6
531 03 285401428 0c 0.47 0.003 0.5 to 2
531 01 285401411 0a 0.53 0.005 0.8 to 3
531 10 285401436 I 0.64 0.01 1.2 to 6
531 13 285401444 Ic 0.84 0.03 3 to 20
531 20 285401452 II 1.15 0.1 10 to 60

35
Cannon-Fenske viscometers

Cannon-Fenske routine viscometers • the present design has a deepen-


ing in the lower bend. Accord-
comply with standards ISO 3105,
ingly, these viscometers can also
ASTM D 2515, BS 188 with respect to
be used for automatic measure-
technical measuring specifications.
ments.
• are suitable for all Newtonian
• filling quantity: approx. 7 to 10 ml
liquids with a viscosity of 0.35 to
20,000 mm²/s • overall length: approx. 245 mm

Cann
calibrated, with constant
with ring mark, for automatic measurements
for manual measurements

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Type No. Order No. Capillary No. Ø i [mm] (approx.) (approx.)
513 00 285403507 520 00 285403704 25 0.30 0.002 0.4 to 1.6
513 03 285403515 520 03 285403712 50 0.44 0.004 0.8 to 3.2
513 01 285403523 520 01 285403729 75 0.54 0.008 1.6 to 6.4
513 10 285403531 520 10 285403737 100 0.63 0.015 3 to 15
513 13 285403548 520 13 285403745 150 0.78 0.035 7 to 35
513 20 285403556 520 20 285403753 200 1.01 0.1 20 to 100
513 23 285403564 520 23 285403761 300 1.27 0.25 50 to 200
513 21 285403572 520 21 285403778 350 1.52 0.5 100 to 500
513 30 285403589 520 30 285403786 400 1.92 1.2 240 to 1,200
513 33 285403597 520 33 285403794 450 2.35 2.5 500 to 2,500
513 40 285403601 520 40 285403807 500 3.20 8 1,600 to 8,000
513 43 285403618 520 43 285403815 600 4.20 20 4,000 to 20,000

Fen
Cannon-Fenske reverse flow
viscometers

• Comply with standards ISO 3105, • filling quantity: approx. 12 ml


ASTMD 2515, ASTM D 446, NF T
• overall length: approx. 295 mm
60 - 100 with respect to technical
measuring specifications.

calibrated,
with 3 ring marks,
with 2 constants,
only for manual measurement

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Capillary No. Ø i [mm] (approx.) (approx.)
511 00 285403001 25 0,31 0.002 0.4 to 1.6
511 03 285403018 50 0,42 0.004 0.8 to 3.2
511 01 285403026 75 0,54 0.008 1.6 to 6.4
511 10 285403034 100 0,63 0.015 3 to 15
511 13 285403042 150 0,78 0.035 7 to 35
511 20 285403059 200 1,02 0.1 20 to 100
511 23 285403067 300 1,26 0.25 50 to 200
511 21 285403075 350 1,48 0.5 100 to 500
511 30 285403083 400 1,88 1.2 240 to 1,200
511 33 285403091 450 2,20 2.5 500 to 2,500
511 40 285403104 500 3,10 8 1,600 to 8,000
511 43 285403112 600 4,00 20 4,000 to 20,000

36
Cannon-Fenske routine viscometers • are suitable for all Newtonian
with additional tube and threads liquids with a viscosity
of 0.35 to 20,000 mm²/s.
comply with standards ISO 3105, ASTM
D 2515, BS 188 with respect to technical • filling quantity: approx. 7 to 12 ml
measuring specifications. These viscom-
• overall length: approx. 245 mm
eters are preferably used for automatic
measurements. The ­additional filling
and cleaning tube and the glass thread
ensure safe operational use. The cali-
brated viscometers are delivered with

non-
manufacturer’s certificate in accor-
dance with DIN 55 350, Part 18.

nske
calibrated,
with ring marks,
with constant for automatic measurements

Type No.
546 00
Order No.
285402116
Capillary No.
25
Capillary
Ø i [mm]
0.30
Constant K
(approx.)
0.002
Measuring range [mm²/s]
(approx.)
0.4 to 1.6
546 03 285402132 50 0.44 0.004 0.8 to 3.2
546 01 285402124 75 0.54 0.008 1.6 to 6.4
546 10 285402149 100 0.63 0.015 3 to 15
546 13 285402157 150 0.78 0.035 7 to 35
546 20 285402165 200 1.01 0.1 20 to 100
546 23 285402181 300 1.27 0.25 50 to 200
546 21 285402173 350 1.52 0.5 100 to 500
546 30 285402198 400 1.92 1.2 240 to 1,200
546 33 285402202 450 2.35 2.5 500 to 2,500
546 40 285402219 500 3.20 8 1,600 to 8,000
546 43 285402227 600 4.20 20 4,000 to 20,000

37
Ostwald viscometers

Ostwald viscometers

• filling quantity: 3 ml

• overall length: approx. 220 mm

with ring marks,


without constant,
for manual measurements

Capillary Transit time for water Constant K for use from [mm²/s]
Type No. Order No. Ø i [mm] ~ [s] (approx.) (approx.)
509 03 285404006 0.3 250 0.004 0.3
509 04 285404014 0.4 75 0.01 1
509 05 285404022 0.5 30 0.03 2.5

Ostwal
509 06 285404039 0.6 15 0.07 5.5
509 07 285404047 0.7 10 0.1 10

Micro-Ostwald viscometers

• are suitable for measurements of


small liquid quantities even with
excessive foam formation.

• filling quantity: 2 ml

• overall length: approx. 290 mm

calibrated, calibrated, not calibrated,


with ring marks, with ring marks, without constant;
with constant, with constant, for determination of relative viscosity
for manual measurements for automatic measurements

Capillary Constant K Measuring range [mm²/s]


Type No. Order No. Type No. Order No. Typ-Nr. Bestell-Nr. Capillary No. Ø i [mm] (approx.) (approx.)
516 10 285404203 517 10 285404306 518 10 285404409 I 0.43 0.01 0.4 to 6
516 13 285404211 517 13 285404314 518 13 285404417 Ic 0.60 0.03 1.2 to 18
516 20 285404228 517 20 285404322 518 20 285404425 II 0.77 0.1 4 to 60
516 23 285404236 517 23 285404339 518 23 285404433 IIc 1.00 0.3 12 to 180
516 30 285404244 517 30 285404347 518 30 285404441 III 1.36 1 40 to 800

38
Accessories

Brackets and stands stats the viscometers can only be used


with the appropriate stand or bracket.
All brackets and stands are designed
to ensure that the viscometers are For DIN Ubbelohde viscometers that
held vertically. They also protect the are used as reference measuring stan-
viscometers from breakage. The maxi- dard, specifically modified bracket (VZ
mum deviation is < 1°. In conjunction 5840) must be used.
with SI Analytics and other commer-
cially available seethrough thermo-

Brackets made of stainless steel


suitable for use with all Ubbelohde viscometers
for manual and automatic measurements

Type No. Order No.

053 92 285405043

VZ 5840 (accessory for reference measuring standard) 285417201

suitable for use with Ubbelohde viscometers with TC sensors

ld
Type No. Order No.

053 93 285405035

suitable for use with all reverse flow viscometers


(Cannon-Fenske and BS/IP U-tube viscometers) for manual and
automatic measurements (not illustrated)

Type No. Order No.

053 96 285405019

suitable for use with Micro-Ostwald viscometers


for manual and automatic measurements

Type No. Order No.

053 97 285405027

39
Accessories

Brackets for reference measuring The extension set for the test standard
standard (VZ 5840) as supplement to the
bracket guarantees vertical slope with
DIN Ubbelohde viscometers which
a maximum deviation of < 1° and the
are used as testing standard should be
centered positioning of the capillaries.
stored in a specially modified viscom-
eter bracket according to official
inspection / calibration authorities.

Type No. Order No.

VZ 5840 285417201

Control thermometers

Type No. Order No. Measuring range °C Graduation °C

VZ 2801 285415763 – 5 to + 38 1/10

Acce
VZ 2802 285415771 + 33 to + 67 1/10

VZ 2803 285415788 + 66 to + 102 1/10

VZ 2804 285415796 + 95 to + 152 1/10

VZ 2901 285415809 + 20 to + 25 1/100

VZ 2907 285417078 + 22 to + 27 1/100

VZ 2908 285415825 + 37 to + 42 1/100

VZ 2905 285415841 + 45 to + 50 1/100

VZ 2906 285415858 + 97 to + 101 1/100

VZ 2909 285417094 + 132 to + 137 1/100

Control thermometers for CT 72 thermostat series

Type No. Order No. Measuring range °C Graduation °C


VZ 7100 285421051 + 19 to + 21 1/100
VZ 7101 285421068 + 24 to + 26 1/100
VZ 7102 285421076 + 29 to + 31 1/100
VZ 7103 285421084 + 39 to + 41 1/100
VZ 7104 285421092 + 99 to + 101 1/100
VZ 7105 285421105 + 134 to + 136 1/100

40
Accessories

LabPump Since the LabPump VZ 5655 and the


connections are made of PTFE or
The LabPump VZ 5655 (not illustrated)
stainless steel, the pump is suitable for
used with manual and semi-automatic
use with aggressive mediums.
measurements to extract and pump
solutions: The range of use for semi-automatic
processing of samples, e.g. with a vis-
• filling of viscometers
cosity measuring instrument AVS®
• rinsing with the next sample 360, AVS® 370 or AVS® 470, is possi-
ble up to a viscosity of 30,000 mm²/s.
• extract between manual
For semi-automatic processing work,
measurements
the PTFE tube combination with stand
• emptying of viscometers without (see illus-tration) and the waste bottle,
removing them from the type no. VZ 5624, are used.
thermostatic bath

Type No. Order No.

VZ 5655 1040755

essories
Polyamide bracket
for use with Cannon-Fenske routine viscometers,
Cannon-Fenske reverse flow viscometers and all Ostwald viscometers
for manual measurements only

Type No. Order No.

064 99 285405105

PTFE bracket
for use with Cannon-Fenske routine viscometers,
for automatic measurements only (not illustrated)

Type No. Order No.

065 99 285405113

41
Accessories

Temperature stabilization jackets plate depend upon the type of vis-


cometer being used. The support
In the absence of a see-through ther-
plate has been designed to facilitate
mostat the temperature of capillary
changing the viscometer when
viscometers can be stabilized in this
required. An additional hole is pro-
type of jacket using circulation ther-
vided in the support plate so that a
mostats in the temperature range 0 to
control thermometer can be fitted. A
180 °C. The shape of the jacket and
quick-action seal simplifies changing
the number of holes in the support
viscometers.

Temperature stabilization jacket with support plate


for Ubbelohde viscometers

Type No. Order No. Item No. Comment

577 00 285405508 complete, without viscometer

Component parts

577 01 285405516 1 temperature stabilization jacket, straight

238 00 285405524 2 support plate with 4 silicone rings


(d = 4, 6, 8 and 10 mm)

225 34 285405532 3 silicone O-ring, ND 60

072 34 285405549 4 quick-action seal, NW 60

Silicone rings

Type No. Order No. d mm D mm h mm

228 11 285405808 4 10 5

Accesso
228 14 285405816 6 16 5

228 16 285405824 8 16 5

228 17 285405832 10 16 5

42
AVS® measuring stands

AVS® measuring stands The measuring stands or brackets can


be exchanged.
Measuring stands of the series AVS®/S
can be used to measure the flow- The distance between the levels of the
through time in viscometers automati- automatic optoelectronic unloading
cally. system is 40.00 mm ± 0.03 mm. This
results in a standard deviation of
The measuring stands can be con-
VK = 0.05 % for Ubbelohde viscometers
nected to all measuring instruments
if the measuring stand is replaced or
made by SI Analytics for automatic mea-
changed within the process.
surement of viscosity and operate with
all standard viscometers for repetitive For repetitive measurements with vis-
measurements. cosity measuring instruments and
Automatic measurements have the fol- Ubbelohde viscometers with measur­
lowing advantages: ing stands, the standard deviation
•  the repetitive standard deviation is VK = 0.03 %.
less than for manual measurements
Manually calibrated Ubbelohde visco-
•  the measurement is free from

measu
meters can also be used in AVS® mea-
subjec-tive factors of influence
suring stands. If the automatic sensing
• the results can be printed and/or be
levels do not correspond to the ring
automatically documented with a
marks, the superimposed meniscus
data memory system
detection system will provide a higher
• automatic processing of sample
constant. The difference amounts to
series is available.
0.1 % per millimeter of height offset.
The use of different materials ensures
adaptation to existing measurement
temperatures and applications.

stan
43
Measuring stands

AVS®/S AVS®/S-HT AVS®/SK AVS®/S-CF AVS®/SK-V

Available Ubbelohde viscometers in accordance with DIN, Cannon-Fenske- Ubbelohde-


viscometers ASTM, ISO 3105, Micro-Ubbelohde viscometers, routine dilution
Micro-Ostwald viscometers viscometer viscometer

Temperature – 80 to + 100 °C – 80 to + 200 °C 0 to + 60 °C – 80 to + 100 °C 0 to + 60 °C


range other
temperature
ranges available
on request

Suitable 05392 no bracket


brackets 05397 required
(type no.)

Material Aluminium, TiO2-anodized PVDF, Aluminium, PVDF,


stainless steel TiO2-anodized stainless steel

Dimensions 90 x 447 x 90 90 x 496 x 90 90 x 447 x 90 90 x 447 x 90 90 x 447 x 90

uring
(W x H x D) mm

Weight 1.0 1.25 0.8 1.0 0.8


(kg) appr.

Accessories Bracket Type No. 05392 for Ubbelohde viscometers, tube/cable tube/cable
included in tube/cable combination VZ 5505 combination combination
scope of VZ 5505 VZ 5857, mag-
delivery netic stirring
rods,
fastening springs
for viscometer

Note:
When TC viscometers are being used, a bracket type no. 05393, with the necessary tube set is required
only. A measuring stand is not required.

nds
Suitable for use with the measuring units: AVS® 350, AVS® 360, AVS® 370,
AVS® 450, AVS® 470, AVS®Pro

Suitable for use with the thermostatic baths: CT 72/P, CT 72/2-TT, CT 72/2, CT 72/4

Electrical connection: Cable VZ 6225 for all measuring stands to all instruments (is
included in hose sets VZ 5505, VZ 5622 and VZ 5857), con-
trol lamp as function display

Distance between measuring levels: 40.00 mm ± 0.03 mm at 25 °C

Signal transmission: Optically using optical fibres from the measuring level in the
stand head, converted into analogue signal from stand to
measuring instrument

44
TITRONIC® 300 - the new piston burette from SI Analytics

The TITRONIC® 300 is a perfect motor-driven burette for manual titration and
and precise dosing instrument for dispensable liquids, solvents and titrating
agents. However, the TITRONIC® 300 is not only first-rate as a stand-alone
instrument, it also thrives as the heart of a computer-controlled dosing or
titrating system.

Manual Titration
It is true that the automatic titration is gaining ground, but manual titration remains
one of the standard cost effective applications in the lab, whereever high precision
and flexibility are required.
Titration with hand controller (mouse).
Titration rate can be adjusted in five different steps to optimize the titration
speed and accuracy.
Automatic calculation of results in different units and exportation to a printer or
USB memory device.
Automatic weight recording when a balance is connected.

45
Titrating manually, perfectly dosing

Dosing
Beside the titration there are various routine dosing tasks that must be performed in
the lab. A piston burette is the ideal device for precise dosing tasks:

Adjustable dosing and filling rate for each method.


Adjustable filling between each dose step.
The intelligent filling function checks if a dosing step is feasible without filling in
advance. This reduces faulty operations during a serial dosing process.

46
Simple and accurate dosing with the new TITRONIC® 500
piston burette

The TITRONIC® 500 is the ideal piston burette for manual Important features:
titrations, accurate dosing applications as well as the prepa-
Intelligent interchangeable modules with 5, 10, 20
ration of solutions. When used with TitriSoft 3.0, it acts as a
and 50 ml volume capacity.
titration burette or with the TitroLine®7000 and TitriSoft 3.0,
it is an automatic dosing unit perfect to pre-dose a titrant. Connect to a printer and/or an analytical balance.

Remote control access via RS232 or USB interface.

Connect up to 16 devices using the two RS232 ports.


The first burette is connected to one PC RS232 or
USB port.

10
47
Dosing, preparing solutions and performing manual titrations

Manual Titration

It is true that the automatic titration is gaining ground, but


manual titration remains one of the standard cost effective
applications in the lab. Everywhere high precision and flexi-
bility are required; a piston burette with an interchangeable
dosing module is the best choice.

Important features:

• Titration using the manual controller dosing buttons.

• Titration rate can be adjusted to optimize titration


speed and accuracy.

• Programmable automatic calculations, printer ready.

• Automatic weight recording when balance is con-


nected.

Dosing Solutions preparation

Besides titration, there are various routine dosing tasks that A special sample preparation mode is available on the
must be performed in the lab. TITRONIC® 500 where a reagent is dosed into a sample until

Important features: the required concentration is reached. The sample is


weighed, the dosing volume is determined. The volume can
• Control dosing using the manual controller and the then be automatically added to the sample.
dedicated keypad.
Important features:
• Adjustable dosing and filling rates optimize speed and
accuracy. • Adjustable dosing and filling speed.

• Store dosing methods with different parameters. • Dosing volume is automatically calculated without
additional PC software.

• Automatic weight recording when balance is con-


nected.

11
48
PolyVISC® Automatic Viscometer
Instruments for
Automated
Viscosity
Determination

CANNON
INSTRUMENT COMPANY
®

49
Instruments for Automated Viscosity

CANNON® PolyVISC® Automatic Viscometer


ASTM D 2857/D 1628/D 4603 (PET), ASTM D 789, ISO 307 (Nylon), ASTM D 1243/D 3591,
D 4603 (PVC), ASTM D 871/D 1795 (Cellulose), ASTM D 1601 (Ethylene Polymers)

l Automatic glass capillary polymer viscosity


analysis between 0.3 to 20,000 cSt*
l Compatible with most solvents used
for polymer work
l Unique AIRBATH® technology – ultrastable
thermostatic air chamber
l Temperatures from 20 to 100°C
(135°C with High Temperature Option)
l VISCPRO® Windows® software calculates relative
viscosity, inherent viscosity, reduced viscosity, and
intrinsic viscosity
l May also be used for other viscosity applications
* Special viscosity ranges available upon request

Features:
The CANNON® PolyVISC® Automatic Glass Capillary
Viscometer combines automatic sampling, viscosity
measurement, viscometer washing, and polymer solvent
compatibility to provide a convenient benchtop unit ideal for
dilute solution polymer viscosity analysis. Both transparent and
opaque samples can be measured with the accuracy required
by ASTM D 2857. Infrared optical sensors are used for all
transparent samples and some opaque samples. Thermal
sensors are available upon request for heavily opaque CANNON PolyVISC® Automatic Viscometer
samples.
The PolyVISC Viscometer utilizes a unique “air bath”
technology to maintain excellent temperature control. Analyses
are made in an ultra-stable thermostatic air chamber (the After each analysis is completed, the sample flows back into
AIRBATH® Chamber). The AIRBATH Chamber consists of an the sample vial. Residual sample is cleaned from the inside,
outer insulated chamber enclosing a second, inner chamber, outside, and tip of the viscometer by automatic rinsing with a
where the viscometer and sample are continuously bathed in a wash solvent, followed by a drying solvent. The viscometer is
thermally stabilized stream of air. Temperature is controlled then dried. A typical wash requires only 10 to 20 mL of
within 0.01°C with advanced measuring electronics, washing solvent and 10 mL of drying solvent. Wash parameters
thermoelectric cooling, and controlling software. Because both are adjustable by the user.
the sample carousel and the viscometer are completely
contained in the AIRBATH Chamber, only the first sample Software Features
requires a delay for thermal equilibrium. The AIRBATH The operating system is designed to be user friendly. Powerful
Chamber also allows rapid temperature and viscometer VISCPRO® software for Windows automatically calculates
changes. Built-in thermoelectric cooling enables testing below relative, inherent, reduced, and two types of intrinsic viscosity
40°C without the need for an external chiller. (limiting viscosity number), Solomon-Ciuta and Billmeyer
relationships, and the K-value relationship after Fickentscher. In
Overview of Automatic Sample addition, it provides flexible and sophisticated control and
Loading and Analysis database management features. Data can be printed,
Samples are placed in disposable vials which are then loaded displayed on the computer screen, or automatically routed to
into a carousel and inserted into the AIRBATH for analysis. The an existing LIMS data-processing network. Data storage is in a
carousel accepts eleven vials, each containing 10 to 15 mL of Microsoft® Access database.
sample. The extended-range viscometer used in the PolyVISC Solvent Compatibility*
device is similar to that in the CANNON Automatic Viscometer The PolyVISC Viscometer is compatible with most solvents
(CAV). The viscometer contains a lower and an upper timing commonly used for the dilute solution viscosity determination
bulb. First the sample is pulled into the lower chamber of the of polymers. Polymer solutions routinely analyzed in the
viscometer. The time required to fill the lower bulb is measured. standard PolyVISC Viscometer include polyacrylonitrile in DMF,
If the time is short, the sample is pulled into the upper bulb. If nylon in formic acid, nylon in sulfuric acid, polyester in HFIP,
the time is long, the lower bulb is used. A PolyVISC viscometer polyester in phenol-TCE, epoxy resins in methanol, vinyl resins
covers a range of approximately 100-fold viscosity range. This in MIBK, cellulose acetate in acetone and methylene chloride,
means that a single size 1-100 PolyVISC viscometer can and polycarbonate in methylene chloride, and others. We will
analyze a range of samples that would require five manual be happy to answer any questions concerning the compatibility
viscometers. Viscometers can be easily changed in less than 60 of the PolyVISC Viscometer with other solvents.
seconds without the inconvenience of messy bath liquids.

2 CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PA • 16803-1733 • USA
800-676-6232 • 814-353-8000 • Fax 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
50
Instruments for Automated Viscosity

CANNON® PolyVISC® Accessories & Options


ASTM D 2857

PVDF/PTFE component upgrades


A PVDF coating upgrade is available for the PolyVISC
instrument. The AIRBATH Chamber interior is coated
with this durable polyvinylidene fluoride (PVDF)
compound. The upgrade permits testing with highly
aggressive solvents, including concentrated sulfuric
acid. The PVDF coating upgrade also includes a
PTFE wash platform and solid PTFE sample carrier
and wash station.
PolyVISC Volatile Sample Testing
Option
Provides protection from solvent evaporation. The
topless plastic screw cap for standard PolyVISC
sample vials can be modified by the addition of a
scored film and rubber seat.
Ubbelohde
In conjuction with an aluminum foil seal, the rubber
inserts enhance test precision for
highly volatile samples by preventing evaporation that can
affect the concentration of the polymer solution. The design
also helps to reduce toxic and/or corrosive vapors in the
AIRBATH® by resealing the cap aperture following the test,
and the rubber inserts wipe the exterior of the viscometer
tube as it is withdrawn from the vial, further reducing the risk
of cross-contamination when performing multiple sample
tests.
The solvent-resistant rubber inserts are sold in packages of
one dozen each, and can be re-used multiple times with
appropriate cleaning and drying procedures.
PolyVISC viscometer tubes
PVDF interior Two tube types are available for the PolyVISC instrument.
with PTFE carousel The modified Atlantic compound viscometer tube is suitable
for most samples. The modified Ubbelohde tube is
particularly suited for more volatile samples, and may
Extended-Range Viscometers significantly enhance determinability and repeatability.
Size/Range (cSt) Contact CANNON Technical Services for order
0.14 – 14 16 – 1600 information.
0.3 – 30 25 – 2500
0.5 – 50 75 – 7500
* Some solvents and solvent vapors may be corrosive to AIRBATH
1 – 100 300 – 30000 components. The user is responsible for determining chemical
3 – 300 600 – 60000 compatibility with PolyVISC components. Contact CANNON for
consultation and special maintenance guidelines when using potentially
8 – 800 incompatible chemicals. Atlantic

PolyVISC Specifications Computer Requirements


Dimensions: 515 mm wide x 756 mm high x 520 mm deep Please contact CANNON for Computer Specifications.
(20.3 x 29.8 x 20.5")
Viscosity Range: 0.3 to 20,000 cSt (special ranges available on request) Order Information
Solvent Volume: 10 mL for each wash and rinse (application dependent) Catalog # Item Description
Temperature Range: 20 to 100°C 9724-Z18 PolyVISC Automatic Viscosity Measuring
(135°C High-Temperature Option available) System 115V, 50/60 Hz, 1750 watts
Temperature Stability: ± 0.01°C at calibrated temperature 9724-Z20 PolyVISC Automatic Viscosity Measuring
System 230V, 50/60 Hz, 1850 watts
Drop Time Resolution ± 0.01 second
P65.7500 PVDF/PTFE Component Upgrades
Electrical: 115/230V, 50/60 Hz, 1750/1850 watts
(please specify voltage when ordering) P65.3283 PolyVISC Volatile Sample Testing Option
Compliance: CE Mark; EMC directive (89/336/EEC); Low voltage PolyVISC Viscometer Tubes
directive (73/2/EEC); HI-POT (1900 VDC, 60 sec.)

51 CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PA • 16803-1733 • USA
800-676-6232 • 814-353-8000 • Fax 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Instruments for Automated Viscosity

CANNON® PolyVISC® SPS Solution Preparation System

l Prepares polymer solutions for relative viscosity


measurements using weight-to-weight methodology
l Eliminates errors due to variations in solvent density
and cumbersome manual weight and volume
measurements
l Provides concentration accuracy often better than
±0.02%

The PolyVISC® SPS Solution Preparation System from CANNON


Instrument Company is a semi-automated solution preparation system
that uses gravimetric rather than volumetric methodology. The SPS can
be used with the CANNON PolyVISC® Automatic Viscometer to
provide a complete polymer analysis system.
Why Gravimetric Measurement?
When manual volumetric methods of preparing solutions are
employed, variations of solvent density with temperature can lead to
errors larger than the error inherent in the subsequent analytical
procedure. In the PolyVISC SPS Solution Preparation System, the
solvent is weighed rather than measured volumetrically, eliminating
errors due to variations in solvent density. For samples over 100 mL,
concentration accuracy is often better than ± 0.02 percent.
When solutions are prepared using the gravimetric method, the
solution concentration can be expressed in units of volume when the
density of the solvent is known. Thus, it is possible for the PolyVISC SPS
CANNON® PolyVISC® SPS Solution Preparation System Solution Preparation System to prepare solutions for procedures
written for the volumetric method.
PolyVISC® SPS components
A computer-controlled burette pump and proprietary PolyVISC SPS software are at the heart of the Solvent Preparation System.
The software permits convenient user configuration of multiple sample/solvent recipes. Other required system components
(ordered separately) include a Windows-compatible computer and an electronic balance. A number of digital electronic
balances having an RS-232 (COM) output are compatible with the SPS software package.
Solution Preparation
SPS software for Windows® makes solution preparation rapid and simple. No volumetric glassware is used. After the sample ID
information is entered, and the user-configured recipe is selected, the operator places an empty container on the balance pan.
The container is weighed and the operator adds the amount of solute specified by the computer to the sample container. The
precise solute weight is obtained; then the computer calculates the appropriate amount of solvent and adds the solvent to the
sample container on the balance. Solution concentration is calculated by the software, and all sample information is stored on
the computer hard disk drive. A sample label can be printed.

Specifications for the PolyVISC® SPS Solution Preparation System


Pump Module Order Information
Number of Pumps: One Catalog # Item Description
Type of Pumps: Syringe 9724-Z61 PolyVISC® SPS Solution
Flow Rate: 0 to 400 mL/min Preparation System, 115V,
50/60HZ
Solution Types: Weight-to-Weight, Weight-to-Volume, Dilution-to-Volume
9724-Z62 PolyVISC® SPS Solution
Solution Preparation ± 0.02% or better for samples over 100 mL Preparation System, 230V,
Accuracy: 50/60HZ
Solution Preparation Time: 2 minutes (50 mL of solvent)
Compatible Balances: Contact CANNON for updated list
Computer Operating System: Windows® 95/98/XP®/NT®

CANNON
INSTRUMENT COMPANY
®
2139 High Tech Road • State College • PA • 16803 • USA
800 676 6232 • 814 353 8000 • Fax 814 353 8007
e-mail: cannon@cannoninstrument.com • www.cannoninstrument.com

52
© 2009 CANNON Instrument Company
• Accurate and Fast Analysis
• Supports Standard and
Proprietary Methods
• Intuitive and Versatile Software
• Simple Operation
• Robust Performance

Now Available
with 10-place
Auto-Sampler

53
INNOVATIONS IN VISCOSITY MEASUREMENT
CANNON® miniPV® and miniPV-X POLYMER SOLUTION Automatic Viscometers

Essential Automation for Polymer Solution Operation


Viscosity Operation of the miniPV® is simple. First, the sample vial is
The miniPV® from CANNON Instrument Company is an essential charged with about 15 mL of solution. Second, the sample vial
tool for the measurement of polymer solution viscosities. Its is positioned in the vial holder beneath the viscometer. Third,
accuracy and convenience are based on the masterful design the sample platform is raised to the viscometer. This action
of temperature control, fluid handling, sensor technology, and initiates the sample loading procedure and the subsequent
automation. Complete control of the instrument is provided series of automatic processes that were specified by the user
through VISCPRO II software (for Windows® 98/NT®/XP®) which in the VISCPRO II method parameters. In general, the sample
also facilitates procedural tasks such as calibration, data entry, is drawn into the viscometer tube, held until the temperature
method specification, calculation selection, report formatting, has equilibrated, and then released to drain during automatic
and data exporting. All data is securely maintained by VISCPRO timing of the flow rate. Data is transferred via the RS-232 serial
II in a Microsoft Access database. connection to a host computer that stores and maintains the
data in a MS Access® database. The sample is then ejected as
The miniPV® occupies roughly the same footprint as an waste and the sample vial becomes a wash station as solvent
analytical balance, yet is able to provide a high level of is automatically injected into the viscometer tube to initiate
automation which includes sample loading, temperature the cleaning cycle which can include a second solvent for a
equilibration, flow time measurement, and cleaning of the entire subsequent rinse and a final purge of compressed air. Following
system between measurements. Support for a sequence of tube drying, the vial holder is returned to its original position,
samples is available with the optional autosampler. ready to receive the next sample.
VISCPRO II is unique in its ability to perform the Huggins- Auto-Sampler
Kraemer determination of intrinsic viscosity in addition to
single-point methods based on Billmeyer and on Solomon- The user fills the sample vials and then places them into a
Ciuta. Multiple viscosity calculations are also available, leading numbered carousel beneath the viscometer. Carousel locations
to relative, inherent, and reduced viscosity determinations. In are numbered 1 to 10. Sample ID information is entered via the
addition, the software provides flexible and sophisticated control computer. The user initiates the test with a single mouse click.
of data export options and database management.

• Designed to Meet the Unique miniPV® and miniPV-X Features • Thermistor Sensors
Needs of Users in the Polymer
Industry • Peltier Technology Cooling for
efficient operation near ambient
• Calculation of Relative, temperatures
Inherent, Reduced and Intrinsic
Viscosity • Powerful VISCPRO II® Software for
Control, Calculation and Database
• 0.3 to 5,000 centistokes range Management
• (special ranges available on • Compatible with Most Solvents used
request) for Polymer Characterization
• Superb Temperature Control • Applications: Solutions of
from Elastomers, Plastics, TPEs,
Copolymers, Bio-polymers, Blends,
• 20 to 100°C Oligomers, Pre-Polymers, Resins,
• Accurate Flow Times ( to 0.01 Viscosity Modifiers, Pulp and
seconds) Cellulose
For applications using potentially incompatible chemicals,
pleae consult CANNON for special maintentance

miniPV® Automatic Viscometer Specifications miniPV® Order Information


miniPV® Dimensions: 254 mm wide x 437 mm deep x 598 mm high (10 x 17.2 x 23.5”) Catalog # Item Description
Power Supply Dimensions: 330 mm wide x 396 mm deep x 172 mm high (13 x 15.6 x 6.8”) 9725-A91 miniPV® 115V 50/60 hz
Weight: Bath Unit: 27 kg (45 lbs); Power Supply: 12 kg (26 lbs); 9725-A92 miniPV® 100V 50/60 hz
Waste Receiver: 6 kg (13 lbs)
9725-A93 miniPV® 230V 50/60 hz
Shipping Weight: 72 kg (159 lbs) with all units/accessories
9725-C11 miniPV-X 115V 50/60 Hz
Fuse Rating: 115V & 100V Units: M 250V 8A, 1.25 x 0.25”; 230V
Unit M 250V 4A, 1.25 x 0.25” 9725-C12 miniPV-X 230V 50/60 Hz
Compliance: CE Mark P61.5600 Dostmann Thermometer, Calibrated,
Digital Dual Channel Kit
Computer Requirements Computer not included, please contact Cannon for specifications.
P61.5601 Standard Size Smart Probe

2139 High Tech Road • State College • PA • 16803


800-676-6232 • 814-353-8000 • Fax 814-353-8007
email: cannon@cannoninstrument.com • www.cannoninstrument.com
54
©2011 CANNON Instrument Company • Printed in USA
• Chemically Resistant
to Solvent and Acid
• Accurate and Fast
Unattended Analysis
• Minimize Operator
Exposure
• Robust Performance
• Small Foot Print
• On-Board Diagnostics

55
INNOVATIONS IN VISCOSITY MEASUREMENT
CANNON® miniPV®-HX Fully Automated Polymer Solution Viscometer
Automates standard & Proprietary solution viscosity methods • ASTM D2857, D4603, D789, D1243, D1795, USP 911,
D4243, SCAN-CM 15:99 • ISO 1628, ISO 307, ISO 5351, Tappi T230, PAPTAC G.24P, Etc.

Essential Automation for Polymer Solution • Designed to Meet the Unique Needs
Viscosity of Users in the Polymer Industry
The miniPV -HX™from CANNON Instrument Company is
®

an essential tool for the measurement of polymer solution • Calculation of Relative, Inherent,
viscosities. Its accuracy and convenience are based on the Reduced and Intrinsic Viscosity
masterful design of temperature control, fluid handling, sensor • 0.3 to 5,000 centistokes range
technology, and automation. Complete control of the instrument
is provided through VISCPRO II software (for Windows® 98/ • (special ranges available on request)
XP®/Vista®/Windows 7®) which also facilitates procedural tasks
• Superb Temperature Control from
such as calibration, data entry, method specification, calculation
selection, report formatting, and data exporting. • 20 to 100°C (±0.01°C)
The miniPV®-HX™occupies roughly the same footprint as • Accurate Flow Times ( to 0.01
an analytical balance, yet is able to provide a high level of seconds)
automation which includes sample loading, temperature
equilibration, flow time measurement, and cleaning of the entire • Thermistor Sensors
system between measurements. VISCPRO II® is unique in its • Peltier Technology Cooling for
ability to perform an automatic Huggins-Kraemer determination efficient operation near ambient
of intrinsic viscosity in addition to single-point methods based on temperatures
Billmeyer and on Solomon-Ciuta. Multiple viscosity calculations
are also available, leading to relative, inherent, and reduced • Powerful VISCPRO II® Software for
viscosity determinations. In addition, the software provides Control, Calculation and Database
flexible and sophisticated control of data export options and Management
database management.
CANNON® miniPV-HX® • Compatible with Acids and Solvents
Operation Fully Automated Polymer used for Polymer Characterization
Solution Viscometer
Samples are prepared using optional Cannon Instrument SPS
Sample Preparation System in reusable or disposable sample
vial. The samples are inserted into the multiple position auto CANNON® PolyVISC SPS
carousel. The user then selects start. This action initiates the Solution Preparation System
sample loading procedure and the subsequent series of
automatic processes that were specified by the user in the • Prepares polymer solutions for relative viscosity
VISCPRO II® method parameters. In general, the sample is measurements using weight-to-weight methodology
drawn into the viscometer tube, held until the temperature • Eliminates errors due to variations in solvent density and
has equilibrated, and then released to drain during automatic cumbersome manual weight
timing of the flow rate. Data is transferred via the RS-232 serial and volume measurements
connection to a host computer that stores and maintains the
data in a database. The sample is then ejected as waste and the • Provides concentration accuracy
sample vial becomes a wash station as solvent is automatically often better than ±0.02%
injected into the viscometer tube to initiate the cleaning cycle.
The PolyVISC® SPS Solution
The user may select to run a second solvent for a subsequent
Preparation System from CANNON
rinse and a final purge of compressed air during the wash cycle.
Instrument Company is a semi-
Following tube drying, the sample carousel advances to the next
automated solution preparation system
position and repeats the automation process.
that uses gravimetric rather than
volumetric methodology. The SPS can
be used with the CANNON miniPV®-
HX Automatic Viscometer to provide a
miniPV®-HX Automatic Viscometer Specifications complete polymer analysis system.
miniPV®-HX Dimensions: 254 mm wide x 396 mm deep x 787.4 mm
high (10 x 15.6 x 30.5”)
Power Supply Dimensions: 330 mm wide x 396 mm deep x 172 mm high Order Information
(13 x 15.6 x 6.8”) Catalog # Item Description
Weight: Bath Unit: 24 kg (53 lbs); Power Supply: 12 kg 9725-C17 miniPV®-HX 115V 50/60 hz
(26 lbs); Waste Receiver: 6 kg (13 lbs)
9725-C16 miniPV®-HX 100V 50/60 hz
Shipping Weight: 72 kg (159 lbs) with all units/accessories Unit
M 250V 4A, 1.25 x 0.25” 9725-C18 miniPV®-HX 230V 50/60 hz
Compliance: P61.5600 Dostmann Thermometer, Calibrated, Digital Dual
Channel Kit
Computer Requirements: Computer not included, please contact
Cannon for specifications. P61.5601 Standard Size Smart Probe

2139 High Tech Road • State College • PA • 16803


800-676-6232 • 814-353-8000 • Fax 814-353-8007
email: cannon@cannoninstrument.com • www.cannoninstrument.com 56
©2012 CANNON Instrument Company • Printed in USA
CANNON® Commitment to Quality & to You!

CANNON Commitment to Quality & to You!


®

ASTM D 445/ISO 3104

TE-1500 – page 22
The thermoelectrically-cooled TE-1500 provides
rock-steady cooling to –30°C using Peltier cell
technology. Cooling fans and front panel controls are
the only moving parts in this advanced design. A
four-pane frost-free viewing window and twin fluorescent
lamps provide excellent visibility and illumination within
the bath.
TE-3000 – page 23
Our premiere digital bath model for low-temperature
applications, the TE-3000 combines reliable
thermoelectric cooling with convenient keypad
operation. Temperature settings remain stored in
memory even if bath power is cut off. The liquid-crystal
display indicates temperature in Celsius or Fahrenheit. A
self-diagnostic routine and upgradable firmware are just
two of many improvements in our internationally-
respected bath technology.

*Need a CT
bath manifold?
TE-1500 TE-3000 See page 70.

TE-1500 TE-3000 CT-600 CT-2000 CT-500/518 CT-1000 CT-1000HT


200°C Analog Digital Analog Digital Analog Analog Analog
+200°C
150°C +150°C +150°C
+100°C
100°C +100°C

75°C

50°C +30°C
+10°C
+20°C +20°C +25°C
0°C +10°C +10°C*

-10°C

-20°C

-30°C
-30°C -30°C
-40°C

* Requires optional chiller/cooler and/or use of built-in cooling coil for operation near and below ambient.

57
Constant Temperature Bath

CANNON CT-500 Constant Temperature Bath


®

ASTM D 445/ISO 3104

l Accurate Temperature The CT-500 bath chamber is a


Control from 20°C to cylindrical clear glass vessel 305
100°C mm (12 inches) in diameter and
305 mm (12 inches) high. The
l Safe/Reliable Operation CT-518 bath chamber is a full
457 mm (18 inches) high. A
The CANNON CT-500/518 Teflon®-coated stainless steel
baths maintain accurate baffle located in the center of
temperature control of ± 0.01°C the bath provides a plain
within the range of 20°C to reflective background to aid in
100°C, providing the temperature viewing instruments. The top
sensitivity required by ASTM D cover contains seven round
445 for kinematic viscosity holes 51 mm (2 inches) in
measurements with glass capillary diameter for insertion of
viscometers. Two electric heating viscometer holders, allowing up
elements inside the bath rapidly to seven viscosity measurements
heat the medium to any desired to be made simultaneously.
temperature within the range. A Covers are supplied for capping
cooling coil, when connected to unused holes. Two additional
tap water or a cooling system, holes 10 mm (p-inch) in
permits operation below or slightly diameter, are provided for
above ambient temperature. thermometers. Other hole
configurations are available on
Temperature Selection special order. All wetted parts of
A special selector switch on the the CT-500/518 Constant
front panel facilitates rapid setting Temperature Baths are made of
of ten popular temperatures for stainless steel, glass, or Teflon.
kinematic viscosity determinations. The frame is fabricated from
heavy aluminum and coated
Temperatures which may be with a corrosion-resistant epoxy
selected with this switch are 20, finish. Viscometers, holders, bath
25, 30, 40, 50, 60, 70, 80, 90, CANNON CT-500 Constant Temperature Bath oil, and thermometers must be
and 100°C. After the switch has purchased separately.
been set, the bath will equilibrate
to within a fraction of one degree Safety Features
of the desired temperature. A fine A thermistor in the bath senses
tuning control then allows temperature adjustment to within ± any over-temperature fault condition. If such a condition
0.01°C of the target temperature. By switching to the occurs, all power is removed from the bath until an operator
“variable” setting on the front panel, any temperature within resets the over-temperature limit circuit. A second safety
the operating range can be obtained. feature cuts power to the bath heaters if the control thermistor
Description of Bath is disconnected. Finally, operation of the bath is not possible
unless it is filled with liquid to a safe operating level. A
A solid-state control circuit provides proportional control of liquid-level sensor prevents the control circuit from heating the
temperature. The sensing element for the control circuit is a bath until a safe operating level is attained. The bath heaters
stainless steel-encased thermistor. are automatically turned off when the bath liquid drops below
A motor-driven stirrer ensures a uniform temperature minimum safe level.
throughout the bath. The entire electrical control system is
located in a convenient drawer, mounted on glides for easy
access if adjustment or repair is necessary.

CT-500 Constant Temperature Bath Specifications Order Information


Size (CT-500): 407 mm wide x 362 mm deep x Catalog # Item Description
610 mm high (16 x 14.25 x 24 9726-A10 Model CT-500, 120 volts AC, 50/60 Hz, 1000 watts
inches)
9726-A15 Model CT-500F, 240 volts AC, 50/60 Hz, 1000 watts
Capacity CT-500: 17 L (4.5 gal)
9726-A17 Model CT-518, 120 volts AC, 50/60 Hz, 1300 watts
Capacity CT-518: 23.8 L (6.3 gal)
9726-A19 Model CT-518F, 240 volts AC, 50/60 Hz, 1300 watts
Weight (CT-500): 22 kg (48 lbs)
Shipping Weight (CT-500): 34 kg (75 lbs)

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
18 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
58
Constant Temperature Bath

CANNON CT-1000/1000HT Constant Temperature Baths


®

ASTM D 445/ISO 3104

l Select Temperature from The top cover contains seven


20 to 150°C; 200°C for round holes 51 mm (2 inches) in
HT Model diameter. Up to seven glass
capillary viscometers (in holders)
l Multiple Safety Features can be placed in the bath. Other
hole configurations can be
The CANNON CT-1000 Constant supplied on special order.
Temperature Bath maintains the A solid-state control circuit
accurate control required by ASTM equipped with a stainless
D 445 for kinematic viscosity steel-encased thermistor provides
measurements with glass capillary proportional control of
viscometers. Within the range of 20 temperature. A motor-driven
to 100°C temperature is controlled stirrer ensures a uniform
to 0.01°C; above 100°C temperature throughout the bath.
temperature is controlled to 0.03°C. All wetted parts of the bath are
Setting Temperature made of stainless steel, glass, or
Ten of the most popular Teflon®. The bath housing is
temperatures for kinematic viscosity fabricated from heavy aluminum
measurement can be quickly set by and coated with a
rotating a special switch on the corrosion-resistant epoxy finish.
bath’s front panel. This switch allows Viscometers, holders, bath oil,
20, 25, 30, 40, 50, 60, 80, 100, and thermometers must be
135, and 150°C to be selected. purchased separately.
After selection of the temperature, Safety Features
the bath will equilibrate within a There are multiple safety
fraction of one degree of the features. A thermistor in the bath
desired temperature. A fine tuning senses any over-temperature
control then allows temperature fault condition. If such a
adjustment to within 0.01°C of the condition occurs, all power is
target temperature. By switching to CANNON CT-1000 Constant Temperature Bath removed from the bath until an
the “variable” setting on the front operator resets the
panel, any temperature within the over-temperature limit control
operating range can be obtained. circuit. A second safety feature cuts power to the heaters if the
Bath Features control thermistor is disconnected. A liquid-level sensor
The bath chamber is a cylindrical clear Pyrex® vessel 300 mm prevents the control circuit from heating the bath until a safe
(12 inches) in diameter x 300 mm (12 inches) high. A operating level is attained.
stainless steel baffle coated with white Teflon is located in the
center of the bath to provide a good background for viewing
viscometers. Two fluorescent lamps illuminate the interior of CT-1000HT Constant Temperature Bath
the bath brightly and uniformly, without glare. l Precision temperature control to 200°C
Two heating elements inside the bath rapidly heat the bath The CT-1000HT Constant Temperature Bath is identical to
medium to any temperature within the bath range. the CT-1000, except that the CT-1000HT can achieve a
A stainless steel cooling coil is also incorporated. When temperature range from 25°C to 200°C. The temperature
connected to tap water or a recirculating water chiller, the selection dial includes quick-set options for 25, 40, 60,
cooling coil allows operation at temperatures slightly above 80, 100, 135, 150, 165, 185, and 200°C Temperatures
or below ambient. from 150 to 200°C can be controlled to 0.03°C (more
precise than required by ASTM D 445 for high
temperature viscometry).

CT-1000 Constant Temperature Bath Specifications Order Information


Dimensions: 438 mm wide x 464 mm deep x 584 mm high Catalog # Item Description
(17.25 x 18.25 x 23 inches) 9726-A20 CT-1000, 120 volts AC, 50/60 Hz, 1400 watts
Capacity: 17 L (4.5 gal) 9726-A25 CT-1000F, 240 volts AC, 50/60 Hz, 1400 watts
Weight: 43 kg (95 lb) 9726-A27 CT-1000HT, 120 volts AC, 50/60 Hz, 1400 watts
Shipping Weight: 55.8 kg (123 lb) 9726-A29 CT-1000HTF, 240 volts AC, 50/60 Hz, 1400 watts

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
59 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com 19
Constant Temperature Bath

CANNON CT-2000 Constant Temperature Bath


®

ASTM D 445/ISO 3104

l Microprocessor-Controlled The top cover contains seven


Temperature Regulation round holes 51 mm (2 inches)
to 150°C in diameter. Up to seven glass
capillary viscometers (in
l Built-in Platinum Resistance holders) can be placed in the
Thermometer bath. Two additional holes 10
l On-Screen Temperature mm (p-inch) in diameter are
Monitoring provided for thermometers.
A solid-state control circuit
equipped with a stainless
The CANNON CT-2000 Constant steel-encased thermistor
Temperature Bath maintains the provides proportional control
accurate control required by ASTM D of temperature. A motor
445 for kinematic viscosity driven stirrer ensures a
measurements. Within the range of uniform temperature
20 to 100°C temperature is throughout the bath. The
controlled to 0.01°C; above 100°C entire electrical control system
temperature is controlled to 0.03°C. is located in a convenient
Setting Temperature drawer for easy access if
A keypad permits microprocessor- adjustment or repair is
controlled temperature selection to necessary.
three decimal places at any setting All wetted parts of the bath are
within the operational range of the made of stainless steel, glass,
instrument. Twin electric heating or Teflon®. The bath housing is
elements inside the bath rapidly heat fabricated from heavy
the medium to any desired aluminum and coated with a
temperature within the range. A corrosion-resistant epoxy
cooling coil, when connected to tap CANNON CT-2000 Constant Temperature Bath finish. The top cover consists
water or a cooling system, permits of three parts: a stainless steel
operation below or slightly above top surface, an insulating
ambient temperature. The built-in platinum resistance layer, and a bottom stainless steel heat reflector. The bath is
thermometer continuously monitors bath temperature and designed to use water or oil (not supplied). Viscometers,
displays it digitally on a four line by twenty character back-lit holders, bath oil and thermometers must be purchased
liquid-crystal screen. Either Celsius or Fahrenheit display of separately.
temperature may be selected. On power-up the CT-2000
performs an automatic self-test of its circuits. Safety Features
There are multiple safety features. A thermistor in the bath
The CT-2000 also features RS-232 and RS-485 senses any over-temperature fault condition. If such a
communication ports which permit data acquisition and condition occurs, all power is removed from the bath until an
networking via an external computer. Several baths may be operator resets the over-temperature limit control circuit. A
networked to a single computer. second safety feature cuts power to the heaters if the control
Bath Features thermistor is disconnected. A liquid-level sensor prevents the
®
The bath chamber is a cylindrical clear Pyrex vessel 300 mm control circuit from heating the bath until the safe operating
(12 inches) in diameter x 300 mm (12 inches) high. A level is attained. The bath heaters are automatically turned
stainless steel baffle coated with white Teflon is located in the off when the bath liquid drops below the minimum safe level.
center of the bath to provide a good background for viewing

viscometers. Two fluorescent lamps illuminate the interior of


the bath brightly and uniformly, without glare.
CT-2000 Constant Temperature Bath Specifications Order Information
Size: 438 mm wide x 464 mm deep x 584 mm Catalog # Item Description
high (17.3 x 18.3 x 23 inches) 9726-A30 CT-2000, 100-130 VAC, 50/60 Hz, 1400 watts
Bath Volume: 17 liters (4.5 gallons) 9726-A35 CT-2000F, 200-250 VAC, 50/60 Hz, 1400 watts
Weight: 51.4 kg (113 lbs) without bath fluid
Shipping Weight: 60 kg (131 lbs) (total/two boxes)
Electrical: Specify exact line voltage when ordering
Display Resolution: 3 decimal places

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
20 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
60
Constant Temperature Bath

CANNON CT-600 & Bath Accessories


®

ASTM D 445/ISO 3104

CT-600 Constant Temperature Bath CBC-100 Bath Cooler


l Ideal for kinematic viscosity determinations l Thermoelectrically Cools
between 10 and 30°C Baths as much as 15°C Below
l Built-in thermoelectric cooling provides Ambient
“rock-solid” temperature control at ambient l Compact – 1/10th the Size of
and below Standard Refrigeration Units
l Meets ASTM D 445 precision specifications l Improves Temperature
from 10° to 100°C Stability at Ambient and
The new Cannon CT-600 has many of the engineering Below
and control features of the CT-500 bath (see page 18), The CANNON Bath Cooler
but the CT-600 also includes a built-in thermoelectric (CBC-100) is a powerful addition to
cooling unit. The CT-600 controls temperature per the quality line of CANNON
ASTM D 445 precision specifications between 10°C thermoelectric products.
and 100°C, and is ideal for controlled temperature About one-tenth the size of many
applications between 10°C and 30°C. When tests at conventional chillers, the
low temperatures are to be made, the cooling unit is environmentally-friendly CBC-100 is
engaged via a switch on the front panel. When working an efficient and reliable laboratory
at higher temperatures (30°C and above), the switch is workhorse. Solid-state electronics and
turned off. Nothing could be simpler. proven thermoelectric technologies
A special selector switch on the front panel facilitates provide enough cooling power to
rapid setting of ten popular temperatures for kinematic viscosity control accurately at
determinations: 10, 15, 20, 25, 30, 40, 50, 60, 80, and 100°C. After the 15°C below ambient,
switch has been set the bath will equilibrate to within a fraction of a degree even in open shallow
of the desired temperature. A fine-tuning control then allows temperature baths.
adjustment to within ± 0.01°C of the target temperature as measured by
the user’s reference thermometer. By switching to the “variable” setting on
the front panel, any temperature within the operating range can be
obtained.
For a description of bath safety features, see the CT-500 entry on page 18. CBC-100 Bath Cooler

Power Supply (same for both versions) Power Supply and Cooling Column for both versions
Size: 121 mm wide x 241 mm deep x 102 mm high Weight: 5.7 kg (12.5 lbs)
(4.75 x 9.5 x 4 inches) Power 100 to 240 volts, 50/60 Hz, 250 watts
Requirements:
Cooling Column (excluding immersion probe)
Maximum Bath The Cannon Bath Cooler should not be
Size: 108 mm wide x 190 mm deep x 222 mm high Temperature exposed to temperatures above 80°C.
(4.25 x 7.5 x 8.75 inches)
CBC-100R: Total height including immersion probe is Order Information
572 mm (22.5-inches) Catalog # Item Description
*Need a CT bath manifold? See page 70. 9726-A36 CT-600 Constant Temperature Bath, 115V, 50/60 Hz
9726-A37 CT-600F Constant Temperature Bath, 230V, 50/60 Hz
9726-D05 CBC-100R (regular version with US plug)
9726-D10 CBC-100R-F (regular version with special plug)

CT Series Storage Unit Specifications


CT Series Storage Unit Size: 457 mm wide x 457 mm
l Convenient Viscometer Storage deep x 165 mm high
can be used as Bath Support (18 x 18 x 6.5 inches)
This sturdy two-drawer unit fits beneath Weight: 10.9 kg (24 lbs)
all CANNON constant temperature Shipping Weight: 13.6 kg (30 lbs)
baths. It provides elevated viewing of the
bath, and ample room for storage of Order Information
viscometers and other accessories. Catalog # Item Description
CT Series Storage Unit 9726-A50 CT Series Storage Unit

61
Constant Temperature Baths

CANNON TE-1500 Constant Temperature Bath


®

ASTM D 445/ISO 3104

l Temperature Range from Safety/Reliability


+10°C to –30°C Bath safety and reliability are
l Precision and Stability ensured by a number of
Satisfying ASTM D 445 built-in features including an
Specifications internal thermostat which
l Environmentally-Friendly senses any over-temperature
fault condition. If such a
Thermoelectric Cooling condition occurs, all power is
System removed from the bath until
l Compact Footprint; the internal temperature
No Accessory decreases sufficiently to reset
Heating/Cooling the thermostat. If the control
Units Required thermistor is disconnected, all
power is removed from the
bath until the thermistor
The CANNON TE-1500 connection is restored.
Thermoelectric Low Temperature
Bath incorporates many new Description of Bath
advances in science and The bath holds 2.5 liters of
technology, including bath medium (not supplied).
custom-developed Peltier Cells The bath housing is fabricated
and fan-cooled high-density heat from heavy aluminum coated
sinks. The TE-1500 provides the with a corrosion-resistant
precision required by ASTM D 445 epoxy finish. The top cover of
for measurement of kinematic the bath contains two round
viscosity at low temperatures. holes 51 mm (2 inches) in
diameter for insertion of glass
Compact Features capillary viscometers or other
The TE-1500 is a compact, instruments.
tabletop instrument which uses an An additional hole is provided
air-cooled thermoelectric cooling for a reference thermometer
system requiring no accessory or probe. The aluminum bath
heating or cooling units. vessel measures 305 mm
A selector switch on the front high x 121 mm wide x 83
panel allows direct setting of eight mm (12" x 4.75" x 3.25")
temperatures commonly used for CANNON TE-1500 deep. A baffle within the bath
low-temperature measurement: Thermoelectric Low Temperature Bath vessel facilitates circulation of
10, 0, –5, –10, –15, –20, –25, liquid and provides a
and –30°C. After this switch has background for easy viewing
been set, the bath will equilibrate of instruments. A four-pane
to within a fraction of a degree of the desired temperature. A “frost-free” viewing window and internal illumination by twin
fine-tuning control then allows exact temperature adjustment. fluorescent lamps provide excellent visibility within the bath.
By switching to the “variable” setting on the front panel, any Viscometers, holders, appropriate bath medium, and
temperature within the operating range can be obtained. thermometers must be purchased separately.

TE-1500 Specifications Order Information


Bath Size: 375 mm wide x 420 mm deep x 620 mm high Catalog # Item Description
(14.8 x 16.5 x 24.4 inches) 9726-A62 TE-1500, 120 volts,
Bath Capacity: 2.5 liters (0.66 gal) 50/60 Hz, 1000 watts
Bath Dimensions 121 mm wide x 83 mm deep x 305 mm high 9726-A66 TE-1500F, 230 volts,
(Internal): (4.75 x 3.25 x 12 inches) 50/60 Hz, 1000 watts
Viewing Area: 265 mm high x 95 mm wide (10.45 x 3.75 inches) 9726-A68 TE-1500, 100 volts,
Bath Weight: 43.1 kg (95 lbs) 50/60 Hz, 1000 watts
Bath Shipping Weight: 59 kg (130 lbs)
Electrical: Available in 120V, 220V, and 240V models. All models 50/60 Hz
Please specify exact line voltage when ordering.

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
22 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
62
Constant Temperature Baths

CANNON TE-3000 Constant Temperature Bath


®

ASTM D 445/ISO 3104

l Temperature Range from –30°C to +30°C


l Self-Contained Thermoelectric Cooling —
No Accessory Units Required
l Convenient Keypad Data Entry
l Selectable Temperature Scale
(Celsius or Fahrenheit)
l Meets ASTM D 445 Requirements
for Temperature Stability

The CANNON TE-3000 Thermoelectric Constant Temperature Bath


provides accurate temperature control between –30°C and +30°C,
meeting the requirements of ASTM D 445 for measurement of
kinematic viscosity at low temperatures.
The low temperatures are obtained by built-in air-cooled Peltier cells
which provide thermoelectric cooling. No external refrigeration unit
or air/water heat exchanger is used.
Keypad Operation
The desired bath temperature and calibration temperatures are
entered digitally by means of a keypad located on the front panel.
Temperatures remain in memory even if power to the bath is
accidentally cut off. Both the target temperature and the actual
temperature of the bath are displayed continuously on a liquid
crystal display. An on-screen menu permits selection of Celsius or
Fahrenheit operation, along with several diagnostic features.
Durable Construction
The top cover of the bath contains two round holes 51 mm
(2 inches) in diameter for insertion of glass capillary viscometers or
other instruments. An additional hole is provided for a reference
thermometer or probe. A four-pane “frost-free” viewing window and
internal illumination by fluorescent lamps provide excellent visibility
within the bath. CANNON TE-3000 Thermoelectric
Constant Temperature Bath
The bath holds 2.5 liters (0.66 gal) of bath medium (not supplied).
The bath housing is fabricated from heavy aluminum coated with a
corrosion-resistant epoxy finish. Contained in the housing is an
aluminum bath vessel that measures 305 mm high x 121 mm wide
x 83 mm deep. A baffle within the bath vessel facilitates circulation
of the liquid and provides a smooth background for viewing
instruments.
Viscometers, holders, appropriate bath medium, and thermometers
must be purchased separately.

TE-3000 Specifications Order Information


Bath Size: 375 mm wide x 420 mm deep x 620 mm deep Catalog # Item Description
(14.8 x 16.5 x 24.4 inches) 9726-A70 TE-3000, 120 volts, 50/60 Hz, 1000
Bath Capacity: 2.5 liters (0.66 gal) watts
Bath Dimensions 121 mm wide x 83 mm deep x 305 mm high 9726-A75 TE-3000F, 240 volts, 50/60 Hz, 1000
(Internal): (4.75 x 3.25 x 12 inches) watts
Viewing Area: 265 mm high x 95 mm wide (10.45 x 3.75 inches) 9726-A77 TE-3000, 100 volts, 50/60 Hz, 1000
Bath Weight: 43.1 kg (95 lbs) watts
Bath Shipping Weight: 59 kg (130 lbs)
Electrical: Available in 120V, 220V, and 240V models.
All models 50/60 Hz
Please specify exact line voltage when ordering.

63
Glass Capillary Viscometers
CANNON Glass Capillary Viscometers
®

Size, Range, & Expanded Uncertainty

ASTM D445ASTM D 446, ASTM D 2171, ISO 3104, ISO 3105

Specifications for
CANNON® Viscometers
Table 1
Cannon-Fenske,
Cannon-Ubbelohde, and
Cannon-Ubbelohde Dilution
Viscometers
Approx.
constant, Range
Size cSt/s centistokes*
25 0.002 0.5 to 2
50 0.004 0.8 to 4
75 0.008 1.6 to 8
100 0.015 3 to 15
150 0.035 7 to 35
200 0.1 20 to 100
300 0.25 50 to 250
350 0.5 100 to 500
400 1.2 240 to 1200
450 2.5 500 to 2500
500 8 1600 to 8000
600 20 4000 to 20 000
650 45 9000 to 45 000
700 100 20 000 to 100 000
CANNON Glass Capillary Viscometers
* Ranges in centistokes represent flow times
of 250 to 1000 seconds for Size 25, and
200 to 1000 for all other sizes.
Glass Capillary Viscometers
The tables on this page provide size and range information for viscometers on Table 2
pages 26-30. Additional sizing information is available on the individual product Ubbelohde Viscometers
pages for other viscometer types.
Approx.
Why does CANNON offer so many different types of glass capillary viscometers? constant, Range
Primarily because no single capillary viscometer is ideally suited for all kinematic Size cSt/s centistokes*
viscosity determinations. Individual analysts also have their own preferences. The 0 0.001 0.3 to 1
brief descriptions on the following page may assist you in determining which
0C 0.003 0.6 to 3
viscometer is most appropriate for your particular application. Feel free to contact
CANNON with your questions. Our experienced staff will be happy to assist you. 0B 0.005 1 to 5
1 0.01 2 to 10
Expanded Uncertainty 1C 0.03 6 to 30
CANNON Instrument Company provides expanded uncertainty data with all 1B 0.05 10 to 50
of our calibrated viscometers. This is the universally accepted statistic when dealing
with calibration data. In order to maintain A2LA accreditation, we are required to 2 0.1 20 to 100
determine and specify the expanded uncertainty for all our calibration data. 2C 0.3 60 to 300
ISO publications define expanded uncertainty as "quantity defining the interval about 2B 0.5 100 to 500
the result of a measurement within which the values that could reasonably be 3 1.0 200 to 1000
attributed to the measurement may be expected to lie with a high level of 3C 3.0 600 to 3000
confidence." 3B 5.0 1000 to 5000
Expanded uncertainty is a more comprehensive and meaningful term than 4 10 2000 to 10 000
"precision" because it incorporates all variables in the measurement process. When 4C 30 6000 to 30 000
comparing the accuracy of viscometers it is essential that the expanded uncertainty 4B 50 10 000 to 50 000
be specified. Detailed information regarding CANNON measurements and
expanded uncertainty is available for download from the CANNON website, 5 100 20 000 to 100 000
www.cannoninstrument.com. * Ranges in centistokes represent flow times
of 300 to 1000 seconds for Size 0, and 200
CANNON Instrument Company also provides uncertainty data with every bottle of to 1000 seconds for all other sizes.
our certified viscosity standards (see pages 42-49).

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
24 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
64
Glass Capillary Viscometers

Routine & Qpaque Cannon-Fenske Viscometers


ASTM D 445, ASTN D 446, ISO 3104, ISO 3105

Cannon-Fenske Routine Viscometers Calibrated


Uncalibrated CFRC
Calibrated CFRC Series Size CFRU Series Series
For measuring kinematic viscosity of transparent Newtonian 25 9721-A50 9721-B50
liquids, particularly petroleum products or lubricants, according 50 9721-A53 9721-B53
to ASTM D 445 and ISO 3104. Specifications conform to
ASTM D 446 and ISO 3105. 75 9721-A56 9721-B56
100 9721-A59 9721-B59
Provided with certificate of calibration and instruction sheet. See
page 24 for significance of expanded uncertainty. 150 9721-A62 9721-B62
200 9721-A65 9721-B65
l Minimum sample volume 7 mL. Require liquid bath depth of
203 mm (8 inches). 300 9721-A68 9721-B68
l Constants and ranges are as specified in Table 1 on
350 9721-A71 9721-B71
page 24. 400 9721-A74 9721-B74
l Holder is not supplied. For holders, see 9726-M50 and 450 9721-A77 9721-B77
-M53 on page 37. 500 9721-A80 9721-B80
l Viscometers with special constants are available by special 600 9721-A83 9721-B83
order at additional cost. 650 9721-A86 9721-B86
Uncalibrated CFRU Series 700 9721-A89 9721-B89
Similar to 9721-B50 series, but uncalibrated.
l Holder is not supplied. For holders, see 9726-M50, -M53, and -M79 on page 37.

9721-A50 Series
9721-B50 Series Cannon-Fenske Opaque Viscometers
Reverse-flow type for dark liquids Uncalibrated
Calibrated
CFOC
Calibrated CFOC Series Size CFOU Series Series
Reverse-flow type, for measurement of kinematic viscosity of 25 9721-E50 9721-F50
dark Newtonian liquids according to ASTM D 445 and ISO 50 9721-E53 9721-F53
3104. Permit measurements of liquids whose meniscus cannot 75 9721-E56 9721-F56
be seen at the timing marks when using Cannon-Fenske
100 9721-E59 9721-F59
Routine-type viscometer. Used to measure kinematic viscosity of
liquid (cutback) asphalts and road oils at 60°C (140°F) in 150 9721-E62 9721-F62
range of 30 to 6000 centistokes according to ASTM D 2170. 200 9721-E65 9721-F65
Also used to study lubricating oils at low temperatures when 300 9721-E68 9721-F68
investigating the effect of various additives on lubricating and 350 9721-E71 9721-F71
hydraulic oils. External pressure can be applied to vary shear
400 9721-E74 9721-F74
stress and shear rate. Specifications conform to ASTM D 446
and ISO 3105 for single determinations of viscosity for opaque 450 9721-E77 9721-F77
liquids. Upper bulb may be used for determining the degree of 500 9721-E80 9721-F80
non-Newtonian behavior of the sample. 600 9721-E83 9721-F83
Provided with certificate of calibration and instruction sheet. See 650 9721-E86 9721-F86
page 24 for significance of expanded uncertainty. 700 9721-E89 9721-F89
l Minimum sample volume 12 mL. Require liquid bath depth
of 229 mm (9 inches).
l Size 25 has approximate constant 0.002 cSt/s and range of 0.4 to 2 cSt. Constants and ranges for
other sizes are as given in Table 1 on page 24.
l Holder is not supplied. For holders, see 9726-M50 and -M53 on page 37.
l For replacement micro stoppers, see 9726-N98 on page 37.
l Viscometers with special constants are available by special order at additional cost.
Uncalibrated CFOU Series
Similar to 9721-F50 series, but uncalibrated.
9721-E50 Series
9721-F50 Series l Holder is not supplied. For holders, see 9726-M50, -M53, and -M79 on page 37.

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
65 26 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Glass Capillary Viscometers

Suspended Level Ubbelohde Viscometers


ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

Ubbelohde Viscometers Uncalibrated Calibrated Calibrated


Size UBU Series UBC Series UBXC Series
Calibrated UBC Series
0 9721-N50 9721-R50 9721-U50
For measurement of kinematic viscosity of
0C 9721-N53 9721-R53 9721-U53
transparent Newtonian liquids by suspended level
principle as described in ASTM D 445 and D 446, 0B 9721-N56 9721-R56 9721-U56
and ISO 3104 and 3105. Similar to 1 9721-N59 9721-R59 9721-U59
Cannon-Ubbelohde but less rugged in design. 1C 9721-N62 9721-R62 9721-U62
Provided with certificate of calibration and instruction 1B 9721-N65 9721-R65 9721-U65
sheet. See page 24 for significance of expanded 2 9721-N68 9721-R68 9721-U68
uncertainty. 2C 9721-N71 9721-R71 9721-U71
l Minimum sample volume 11 mL. Require liquid 2B 9721-N74 9721-R74 9721-U74
bath depth of 241 mm (9.5 inches). 3 9721-N77 9721-R77 9721-U77
l Constants and ranges are as specified in Table 2 3C 9721-N80 9721-R80 9721-U80
on page 24. 3B 9721-N83 9721-R83 9721-U83
l Holder is not supplied. For holder, see
4 9721-N86 9721-R86 9721-U86
9726-M82 on page 37.
4C 9721-N89 9721-R89 9721-U89
Uncalibrated UBU Series 4B 9721-N92 9721-R92 9721-U92
Similar to 9721-R50 series, but uncalibrated. 5 9721-N95 9721-R95 9721-U95
l Holder is not supplied. For holder, see
9726-M82 on page 37.
Ubbelohde Unity Factor Calibrated UBXC Series
Similar to 9721-R50 series, but with constant within ± 0.2% of values listed in Table 2 on page 24.
Provided with certificate of calibration and instruction sheet.
9721-N50 Series l Holder is not supplied. For holder, see 9726-M82 on page 37.
9721-R50 Series
9721-U50 Series

Cannon-Ubbelohde Viscometers Uncalibrated Calibrated


Size CUU Series CUC Series
Calibrated CUC Series
25 9721-J50 9721-K50
Suspended level viscometer for measurement of kinematic
viscosity of transparent Newtonian liquids according to ASTM D 50 9721-J53 9721-K53
445 and ISO 3104. Also used for evaluating jet and hydraulic 75 9721-J56 9721-K56
lubricants. Especially suited for use at temperatures above 93°C 100 9721-J59 9721-K59
(200°F) or below –18°C (0°F). Require no kinetic energy 150 9721-J62 9721-K62
corrections over 0.5 to 100 000 centistokes range. Viscometer 200 9721-J65 9721-K65
constant is same at all temperatures. Specifications conform to
ASTM D 446 and ISO 3105. 300 9721-J68 9721-K68
350 9721-J71 9721-K71
Provided with certificate of calibration and instruction sheet. See
page 24 for significance of expanded uncertainty. 400 9721-J74 9721-K74
450 9721-J77 9721-K77
l Minimum sample volume 11 mL. Require liquid bath depth
of 254 mm (10-inches); must be submerged to 500 9721-J80 9721-K80
approximately 220 mm (8.5-inches). 600 9721-J83 9721-K83
l Constants and ranges are as specified in Table 1 on 650 9721-J86 9721-K86
page 24. 700 9721-J89 9721-K89
l Holder is not supplied. For holders, see 9726-M70, -M73,
and -M76 on page 37.
Uncalibrated CUU Series
Similar to 9721-K50 series, but uncalibrated.
Holder is not supplied. For holders, see 9726-M70, -M73, and -M76 on page 37.

9721-J50 Series
9721-K50 Series

66
Glass Capillary Viscometers

Semi-Micro & Dilution Cannon-Ubbelohde Viscometers


ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

Cannon-Ubbelohde Uncalibrated Calibrated


Semi-Micro Viscometers Size CUSMU Series CUSMC Series
25 9722-G50 9722-H50
Semi-Micro Calibrated CUSMC Series
50 9722-G53 9722-H53
For measuring kinematic viscosity of small samples of
75 9722-G56 9722-H56
transparent Newtonian liquids according to ASTM D 445
and ISO 3104. Specifications conform to ASTM D 446 100 9722-G59 9722-H59
and ISO 3105. Viscometer constant is same at all 150 9722-G62 9722-H62
temperatures. 200 9722-G65 9722-H65
Provided with certificate of calibration and instruction 300 9722-G68 9722-H68
sheet. See page 24 for significance of expanded 350 9722-G71 9722-H71
uncertainty. 400 9722-G74 9722-H74
l Minimum sample volume approximately 1.0 mL. Sizes 450 9722-G77 9722-H77
400 through 600 may require larger sample volume. 500 9722-G80 9722-H80
Initial sample can be diluted to as much as 20 mL for
600 9722-G83 9722-H83
intrinsic viscosity determinations. Require liquid bath
depth of 240 mm (9.5 inches).
l Size 25 has approximate constant 0.002 cSt/s and a range of 0.4 to 2 cSt. Constants and ranges
for all other sizes are as given in Table 1 on page 37.
l Holder is not supplied. For holders, see 9726-M70, -M73, and -M76 on page 37.
Semi-Micro Uncalibrated CUSMU Series
Similar to 9722-H50 series, but uncalibrated.
l Holder is not supplied. For holders, see 9726-M70, -M73, and -M76 on page 37.

9722-G50 Series
9722-H50 Series

Cannon-Ubbelohde Uncalibrated Calibrated


CUDU CUDC
Dilution Viscometers Size Series Series
Dilution Type Calibrated CUDC Series 25 9722-L50 9722-M50
For measurement of intrinsic viscosity of transparent 50 9722-L53 9722-M53
Newtonian liquids according to ASTM D 445 and ISO 75 9722-L56 9722-M56
3104. Specifications conform to ASTM D 446 and ISO
100 9722-L59 9722-M59
3105. Viscometer constant is same at all temperatures.
150 9722-L62 9722-M62
Provided with certificate of calibration and instruction
200 9722-L65 9722-M65
sheet. See page 24 for significance of expanded
uncertainty. 300 9722-L68 9722-M68
350 9722-L71 9722-M71
l Minimum sample volume of 8 mL can be diluted to 40
mL. Require liquid bath depth of 305 mm (12 inches). 400 9722-L74 9722-M74
The CANNON CT-518 bath is recommended. 450 9722-L77 9722-M77
l Constants and ranges are as specified in Table 1 on 500 9722-L80 9722-M80
page 24. 600 9722-L83 9722-M83
l Holder is not supplied. For holders, see 9726-M70,
-M73, and -M76 on page 37.
Dilution Type Uncalibrated CUDU Series
Similar to 9722-M50 series, but uncalibrated.
l Holder is not supplied. For holders, see 9726-M70, -M73, and -M76 on page 37.

9722-L50 Series
9722-M50 Series

67
Glass Capillary Viscometers

Cannon-Ubbelohde Four-Bulb Shear Dilution Viscometers


ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

Cannon-Ubbelohde Four-Bulb Shear Specifications for


Dilution Viscometers Ten-fold Dilution Type
Approx. Shear rate at
Five-fold dilution type constant *Range wall, reciprocal
Calibrated CUSDC-11 Series Size† , cSt/s cSt seconds
For measurement of intrinsic viscosities extrapolated to 25 0.002 0.4 to 2 82 to 3300
zero shear rate. Provide five-fold range of shear rates. 50 0.004 0.8 to 4 45 to 1800
Design gives extra mechanical strength. Negligible 75 0.008 1.6 to 8 24 to 960
kinetic energy correction. 100 0.015 3 to 15 15 to 600
Provided with certificate of calibration and instruction 150 0.035 7 to 35 8 to 320
sheet. See page 24 for significance of expanded * Ranges in centistokes represent flow times of
uncertainty. 200 to 1000 seconds.
l Minimum sample volume 20 mL. Require liquid bath
depth of 280 mm (11 inches). Uncalibrated Calibrated
l Holder is not supplied. For holders, see 9726-M70, CUSDU-11 CUSDC-11
-M73, and -M76 on page 37. Size† Series Series
Uncalibrated CUSDU-11 Series 25 9723-L50 9723-M50
Similar to 9723-M50 series, but uncalibrated. 50 9723-L53 9723-M53
75 9723-L56 9723-M56
l Holder is not supplied. For holders, see 9726-M70,
-M73, and -M76 on page 37. 100 9723-L59 9723-M59
150 9723-L62 9723-M62

Ten-fold dilution type


Calibrated CUSDC-17 Series
Similar to 9723-M50 series, but provide ten-fold range Specifications
of shear rates. for Five-fold Dilution Type
Shear rate
Provided with certificate of calibration and instruction Approx. at wall,
sheet. See page 24 for significance Constant, *Range reciprocal
of expanded uncertainty. Size† cSt/s cSt seconds
l Minimum sample volume 20 mL. Require liquid bath 25 0.002 0.4 to 2 82 to 1650
depth of 430 mm (17 inches). The CT-518 bath is 50 0.004 0.8 to 4 45 to 900
recommended. 75 0.008 1.6 to 8 24 to 480
l Holder is not supplied. For holders, see 9726-M70,
100 0.015 3 to 15 15 to 300
-M73, and -M76 on page 37.
150 0.035 7 to 35 8 to 160
Uncalibrated CUSDU-17 Series * Ranges in centistokes represent flow times of
Similar to 9723-R50 series, but uncalibrated. 200 to 1000 seconds.
l Holder is not supplied. For holders, see 9726-M70,
-M73, and -M76 on page 37. Uncalibrated Calibrated
CUSDU-17 CUSDC-17
Size† Series Series
25 9723-N50 9723-R50
50 9723-N53 9723-R53
9723-L50 Series 75 9723-N56 9723-R56
9723-M50 Series 100 9723-N59 9723-R59
9723-N50 Series
9723-R50 Series 150 9723-N62 9723-R62

†Additional tube sizes with larger constants


available by request.

68
Glass Capillary Viscometers

Cannon-Manning Semi-Micro Low Charge Viscometers


ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

Cannon-Manning Uncalibrated Calibrated


Semi-Micro Viscometers Size CMSMU Series CMSMC Series
25 9721-X50 9721-Y50
Calibrated CMSMC Series
50 9721-X53 9721-Y53
For measurement of kinematic viscosity of small samples of
75 9721-X56 9721-Y56
transparent Newtonian liquids, including petroleum products
or lubricants, according to ASTM D 445 and ISO 3104. 100 9721-X59 9721-Y59
Specifications conform to ASTM D 446 and ISO 3105. 150 9721-X62 9721-Y62
Provided with certificate of calibration and instruction sheet. 200 9721-X65 9721-Y65
See page 24 for significance of expanded uncertainty. 300 9721-X68 9721-Y68
l Sample volume approximately 1.0 mL. Sizes 400 through 350 9721-X71 9721-Y71
600 may require more sample. Require liquid bath depth 400 9721-X74 9721-Y74
of 200 mm (8 inches). 450 9721-X77 9721-Y77
l Size 25 has approximate constant 0.002 cSt/s and a 500 9721-X80 9721-Y80
range of 0.4 to 2 cSt. Constants and ranges for all other 600 9721-X83 9721-Y83
sizes are as given in Table 1 on page 24.
l Holder is not supplied. For holders, see 9726-M50 and
-M53 on page 37.
Uncalibrated CMSMU Series
Similar to 9721-Y50 series, but uncalibrated.
l Holder is not supplied. For holders, see 9726-M50 and -M53 on page 37.

9721-X50 Series
9721-Y50 Series

Cannon-Manning Semi-Micro Uncalibrated Calibrated


Extra Low Charge Viscometers CMSMEU CMSMEC
Size Series Series
Extra Low Charge Calibrated CMSMEC Series
25 9722-C50 9722-D50
Similar to 9721-Y50 series, but require smaller sample
50 9722-C53 9722-D53
volume of approximately 0.5 mL. Sizes 200 through 600
may require larger volume of sample. 75 9722-C56 9722-D56
100 9722-C59 9722-D59
Provided with certificate of calibration and instruction sheet.
See page 24 for significance of expanded uncertainty. 150 9722-C62 9722-D62
200 9722-C65 9722-D65
l Without holder.
300 9722-C68 9722-D68
Extra Low Charge Uncalibrated CMSMEU Series
350 9722-C71 9722-D71
Similar to 9722-D50 series, but uncalibrated. 400 9722-C74 9722-D74
l Without holder. 450 9722-C77 9722-D77
500 9722-C80 9722-D80
600 9722-C83 9722-D83

Companion Product!
Constant Temperature Baths
(ASTM D 445 compliant) -
see pages 16-23.

9722-C50 Series
9722-D50 Series

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
69 30 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Glass Capillary Viscometers

CANNON Zeitfuchs Transparent Viscometers


®

ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

Specifications for Zeitfuchs®


Zeitfuchs® Transparent Viscometers Transparent* & Zeitfuchs® Cross-Arm
Viscometers
With permanently attached Approximate
round metal holder Size constant, cSt/s Viscosity, cSt†
Calibrated ZTC-RO Series 1 0.003 0.6 to 3
For measurement of kinematic viscosity of transparent 2 0.01 2 to 10
Newtonian liquids according to ASTM D 445 and ISO 3 0.03 6 to 30
3104. Specifications conform to ASTM D 446 and 4 0.1 20 to 100
ISO 3105. Filling, measurement, and cleaning can be
done without removal from constant temperature bath. 5 0.3 60 to 300
6 1.0 200 to 1000
Certificate of calibration and instruction sheet
provided. See page 24 for significance of expanded 7 3.0 600 to 3000
uncertainty. 8 10 2000 to 10 000
l Minimum sample volume 15 mL. Require liquid
9 30 6000 to 30 000
bath depth of 292 mm (11.5 inches). 10 100 20 000 to 100 000
l For constants and ranges, see table at right. * Zeitfuchs Transparent viscometers are*available
only in sizes 1 through 7.
l Manufactured with permanently attached round
† Ranges in centistokes represent flow times of 200
metal holder, for hole 51 mm (2 inches) in to 1000 seconds.
diameter.
Uncalibrated ZTU-RO Series
Similar to 9723-B50 series, but uncalibrated. Uncalibrated Calibrated
Size ZTU-RO Series ZTC-RO Series
1 9723-A50 9723-B50
2 9723-A55 9723-B55
3 9723-A60 9723-B60
9723-A50 Series 4 9723-A65 9723-B65
9723-B50 Series
5 9723-A70 9723-B70
6 9723-A75 9723-B75
With permanently attached 7 9723-A80 9723-B80
rectangular metal holder
Calibrated ZTC-RE Series
For measurement of kinematic viscosity of transparent Uncalibrated Calibrated
Newtonian liquids according to ASTM D 445 and ISO Size ZTU-RE Series ZTC-RE Series
3104. Specifications conform to ASTM D 446 and 1 9723-C50 9723-D50
ISO 3105. Filling, measurement, and cleaning can be 2 9723-C55 9723-D55
done without removal from constant temperature bath. 3 9723-C60 9723-D60
Provided with certificate of calibration and instruction 4 9723-C65 9723-D65
sheet. See page 24 for significance of expanded 5 9723-C70 9723-D70
uncertainty. 6 9723-C75 9723-D75
l Minimum sample volume 15 mL. Require liquid 7 9723-C80 9723-D80
bath depth of 292 mm (11.5 inches).
l For constants and ranges, see table (above right).
l Manufactured with permanently attached rectangular
metal holder which requires a bath opening of 25 x
48 mm (1 x 1.9 inches).
Calibrated ZTU-RE Series
Similar to 9723-D50 series, but uncalibrated.

9723-C50 Series
9723-D50 Series

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com 31 70
Glass Capillary Viscometers

CANNON Zeitfuchs Cross-Arm Viscometers


®

ASTM D 445, ASTM D 446, ASTM D 2170, ISO 3104, ISO 3105

Calibrated Zeitfuchs® ZCAC ZCAXC Series


Cross-Arm Viscometers Size Series Unity Factor
1 9723-S50 9723-U50
For measurement of kinematic viscosity of transparent and
opaque Newtonian liquids according to ASTM D 445 and 2 9723-S53 9723-U53
ISO 3104. Specifications conform to ASTM D 446 and ISO 3 9723-S56 9723-U56
3105. Also used to measure kinematic viscosity of liquid 4 9723-S59 9723-U59
(cutback) asphalts and road oils at 60°C (140°F) and 5 9723-S62 9723-U62
asphalt cements at 135°C (275°F) in the range of 30 to 6 9723-S65 9723-U65
6000 centistokes according to ASTM D 2170. Can be filled
and cleaned while in constant temperature bath. 7 9723-S68 9723-U68
8 9723-S71 9723-U71
l Minimum sample volume 1 to 3 mL. Require liquid bath
depth of 230 mm (9 inches). 9 9723-S74 9723-U74
l Constants and ranges are as described in table on page 10 9723-S77 9723-U77
33.
l All instruments calibrated. Certificate of calibration and
instruction sheet provided. See page 24 for expanded Unity Factor
uncertainty. Offered with and without holder as described ZCAC-RO ZCAXC-RO
below. Size Series Series
1 9723-W50 9723-Y50
ZCAC Series
2 9723-W53 9723-Y53
As described above. Without holder.
3 9723-W56 9723-Y56
l For holder, see 9726-M95 and -M98 on page 37. 4 9723-W59 9723-Y59
Unity Factor ZCAXC Series 5 9723-W62 9723-Y62
Similar to 9723-S50 series, but have constant within ± 6 9723-W65 9723-Y65
0.2% of values listed in table on page33. 7 9723-W68 9723-Y68
9723-S50 Series
l For holder, see 9726-M95 and -M98 on page 37. 8 9723-W71 9723-Y71
9723-U50 Series
9 9723-W74 9723-Y74
With permanently attached 10 9723-W77 9723-Y77
round metal holder
ZCAC-RO Series
As described above. Supplied with permanently Unity Factor
ZCAC-RE ZCAXC-RE
attached round metal holder, for hole 51 mm Size Series Series
(2 inches) in diameter.
1 9724-B50 9724-D50
Unity Factor ZCAXC-RO Series 2 9724-B53 9724-D53
Similar to 9723-W50 series, but have constant 3 9724-B56 9724-D56
which is within ± 0.2% of values listed in table 4 9724-B59 9724-D59
on page 33.
5 9724-B62 9724-D62
6 9724-B65 9724-D65
With permanently attached 7 9724-B68 9724-D68
rectangular metal holder 8 9724-B71 9724-D71
ZCAC-RE Series 9 9724-B74 9724-D74
As described above. Supplied with permanently 10 9724-B77 9724-D77
attached rectangular metal holder which
requires bath opening
25 x 48 mm (1 x 1.9 inches).
Unity Factor ZCAXC-RE Series
Similar to 9724-B50 series, but have constant
within ± 0.2% of value listed in table on page 33.

9723-W50 Series
9723-Y50 Series
9724-B50 Series
9724-D50 Series

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
71 32 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Glass Capillary Viscometers

CANNON BS/U-Tube Viscometers


®

ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

BS/U-Tube Viscometers Specifications for BS/U-Tube


For Transparent Liquids Viscometers
For measurement of kinematic viscosity of transparent Approx.
Constant, Viscosity Range
Newtonian liquids according to ASTM D 445 and ISO Size mm2/s2 (cSt/s) mm2/ sec (cSt)*
3104. Specifications conform to ASTM D 446 and ISO
3105. A 0.003 0.9 to 3
B 0.01 2.0 to 10
Provided with certificate of calibration and instruction sheet.
See page 24 for significance of expanded uncertainty. C 0.03 6 to 30
D 0.1 20 to 100
l Minimum sample volume 7 mL in sizes A to C, 12 mL in
sizes D to F, and 23 mL in sizes G and H. Require liquid E 0.3 60 to 300
bath depth of 280 mm (11 inches). F 1.0 200 to 1000
l Without holder. G 3.0 600 to 3000
H 10.0 2000 to 10 000
* Ranges in centistokes represent flow times
of 300 to 1000 seconds for Size A and 200
to 1000 seconds for all other sizes.

Calibrated
Size BS/U-Tube Series
A 9724-E50
B 9724-E53
C 9724-E56
D 9724-E59
E 9724-E62
F 9724-E65
9724-E50 Series
G 9724-E68
H 9724-E71

BS/U/M Miniature
U-Tube Viscometers
Specifications for BS/U/M Miniature
For measurement of kinematic viscosity of transparent U-Tube Viscometers
Newtonian liquids according to ASTM D 445
and ISO 3104. Specifications conform to ASTM D 446 and Approx.
ISO 3105. Constant, Viscosity Range,
Size mm2/s2 (cSt/s) mm2/ sec (cSt)*
Provided with certificate of calibration and instruction sheet. M1 0.001 0.2 to 1
See page 24 for significance of expanded uncertainty.
M2 0.005 1 to 5
l Minimum sample volume 2 mL. Require liquid bath depth M3 0.015 3 to 15
of 230 mm (9 inches).
M4 0.04 8 to 40
l Without holder.
M5 0.1 20 to 100
* Ranges in centistokes represent flow times
of 200 to 1000 seconds.

Calibrated
Size BS/U/M Series
M1 9724-E80
M2 9724-E83
M3 9724-E86
M4 9724-E89
M5 9724-E92

9724-E80 Series

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com 33 72
Glass Capillary Viscometers

CANNON Suspended Level Viscometers


®

ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

BS/IP/SL Suspended Specifications for BS/IP/SL


Level Viscometers Suspended Level Viscometers
For measurement of kinematic viscosity Approx.
Constant Viscosity Range, Calibrated
of transparent Newtonian liquids Size mm2/s2 (cSt/s) mm2/sec (cSt)* BS/IP/SL Series
according to ASTM D 445 and ISO
3104. Specifications conform to 1 0.01 3.5 to 10 9724-F10
ASTM D 446 and ISO 3105. 1A 0.03 6 to 30 9724-F13
Provided with certificate of calibration 2 0.1 20 to 100 9724-F16
and instruction sheet. 2A 0.3 60 to 300 9724-F19
See page 24 for significance 3 1.0 200 to 1000 9724-F22
of expanded uncertainty. 3A 3.0 600 to 3000 9724-F25
l Minimum sample volume 11 mL. 4 10 2000 to 10 000 9724-F28
Require liquid bath depth of 250 4A 30 6000 to 30 000 9724-F31
mm (10 inches). 5 100 20 000 to 100 000 9724-F34
l Holder is not supplied. For holder,
* Ranges in centistokes represent flow times of 350 to 1000 seconds
see 9726-M99 on page 37. for Size 1 and 200 to 1000 seconds for all other sizes.

9724-F10 Series

BS/IP/SL(S) Viscometers Specifications for Shortened Form BS/IP/SL(S)


For measurement of kinematic viscosity Suspended Level Viscometers
of transparent Newtonian liquids Approx.
according to ASTM D 445 and ISO Constant, Viscosity Range, Calibrated
3104. Specifications conform to Size mm2/s2 (cSt/s) mm2/sec (cSt) BS/IP/SL(S) Series
ASTM D 446 and ISO 3105. 1 0.0008 1.05 minimum* 9724-F50
Provided with certificate of calibration 2 0.003 2.1† to 3 9724-F53
and instruction sheet. See page 24 for 3 0.01 3.8‡ to 10 9724-F56
significance of expanded uncertainty. 4 0.03 6¤ to 30 9724-F59
l Minimum sample volume 10 mL. 5 0.1 20 to 100 9724-F62
Require liquid bath depth of 230 6 0.3 60 to 300 9724-F65
mm (9 inches).
7 1.0 200 to 1000 9724-F68
l Holder is not supplied. For holder,
8 3.0 600 to 3000 9724-F71
see 9726-M99 on page 37.
9 10 2000 to 10 000 9724-F74
* 1 320 seconds minimum flow time.
† 600 seconds minimum flow time to 1000 seconds.
‡ 380 seconds minimum flow time to 1000 seconds.
¤ Ranges in centistokes represent flow times of 200 to 1000 seconds
for all other sizes.

9724-F50 Series

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
73 34 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Glass Capillary Viscometers

Suspended Level & Reverse Flow BS/IP Viscometers


ASTM D 445, ASTM D 446, ISO 3104, ISO 3105

BS/IP/MSL Miniature Specifications for BS/IP/MSL


Suspended Level Miniature Suspended Level Viscometers
Viscometers Approx. Calibrated
Constant, Viscosity Range, Series
For measurement of kinematic viscosity Size mm2/s2 (cSt/s) mm2/sec (cSt)* BS/IP/MSL
of transparent Newtonian liquids 1 0.003 0.6 to 3 9724-G10
according to ASTM D 445 and ISO
3104. Specifications conform to 2 0.01 2 to 10 9724-G13
ASTM D 446 and ISO 3105. 3 0.03 6 to 30 9724-G16
Provided with certificate of calibration 4 0.1 20 to 100 9724-G19
and instruction sheet. 5 0.3 60 to 300 9724-G22
See page 24 for significance 6 1.0 200 to 1000 9724-G25
of expanded uncertainty. 7 3.0 600 to 3000 9724-G28
l Minimum sample volume 4 mL. * Ranges in centistokes represent flow times of 200 to 1000 seconds.
Require liquid bath depth of
305 mm (12 inches).
l Holder is not supplied. For holder,
see 9726-M99 on page 37.

9724-G10 Series

BS/IP/RF U-Tube Reverse Specifications for BS/IP/RF


Flow Viscometers U-Tube Reverse Flow Viscometers
For measurement of kinematic viscosity Approx. Calibrated
of opaque Newtonian liquids according Constant Viscosity Range BS/IP/RF
to ASTM D 445 and ISO 3104. Size mm2/s2 (cSt/s) mm2/sec (cSt)* U-Tube Series
Specifications conform 1 0.003 0.6 to 3 9724-G50
to ASTM D 446 and ISO 3105. 2 0.01 2 to 10 9724-G53
Provided with certificate of calibration 3 0.03 6 to 30 9724-G56
and instruction sheet. See page 24 for 4 0.1 20 to 100 9724-G59
significance of expanded uncertainty. 5 0.3 60 to 300 9724-G62
l Minimum sample volume 7 mL. 6 1.0 200 to 1000 9724-G65
Require liquid bath depth of 280 7 3.0 600 to 3000 9724-G68
mm (11 inches). 8 10 2000 to 10 000 9724-G71
l Holder is not supplied. For holder, 9 30 6000 to 30 000 9724-G74
see 9726-N45 on page 37.
10 100 20 000 to 100 000 9724-G77
11 300 60 000 to 300 000 9724-G80
* Ranges in centistokes represent flow times of 200 to 1000 seconds.

9724-G50 Series

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com 35 74
Glass Capillary Viscometers

CANNON Vacuum Viscometers


®

ASTM D 2171

Cannon-Manning Specifications for Cannon-Manning


Vacuum Viscometers Vacuum Viscometers
Cannon-Manning Vacuum Viscosity Approx. constant poise/second
Size range, poise at 300 mm Hg vacuum
Calibrated CMVC Series
Bulb B Bulb C
For measurement of viscosity of highly 4 0.036 to 0.8 0.002 0.0006
viscous materials such as asphalt cement at
60°C (140°F) according to ASTM D 2171. 5 0.12 to 2.4 0.006 0.002
Applicable to materials with viscosity range of 6 0.36 to 8 0.02 0.006
0.036 to 80 000 poise. Vacuum of 300 mm 7 1.2 to 24 0.06 0.02
Hg is applied to small arm with timing bulbs; 8 3.6 to 80 0.2 0.06
two bulbs are incorporated to extend viscosity 9 12 to 240 0.6 0.2
range of viscometer.
10 36 to 800 2 0.6
l Minimum sample volume 6 mL. Require
11 120 to 2400 6 2
liquid bath depth of 180 mm (7.1 inches).
12 360 to 8000 20 6
l Provided with certificate of calibration and
instruction sheet. See page 24 for 13 1200 to 24 000 60 20
significance of expanded uncertainty. 14 3600 to 80 000 200 60
l Holder is not supplied. For holder, see
9726-N15 on page 37. Calibrated Calibrated
Cannon-Manning Vacuum Size CMVC Series CMVC-RO Series
Calibrated CMVC-RO Series 4 9724-H50 9724-L50
Same as 9724-H50 series, but with 5 9724-H53 9724-L53
permanently attached round metal holder, for 6 9724-H56 9724-L56
hole 51 mm (2 inches) in diameter. 7 9724-H59 9724-L59
8 9724-H62 9724-L62
9 9724-H65 9724-L65
9724-H50 Series
10 9724-H68 9724-L68
11 9724-H71 9724-L71
Asphalt Institute
12 9724-H74 9724-L74
Vacuum Viscometers
13 9724-H77 9724-L77
Asphalt Institute Vacuum 14 9724-H80 9724-L80
Calibrated AIVC Series
Similar to 9724-H50 series, but with
graduated capillary instead of two timing Specifications for
bulbs. Asphalt Institute Vacuum Viscometers
l Minimum sample size 3 mL. Require bath
Approx. constant
Viscosity poise/second at 300
depth of 180 mm (7.1 inches). Size range, poise mm Hg vacuum
l Note: For sizes 400R and 800R, the
at B at C at D
“roofing” design is normally used for
segment B, which extends the viscosity 25 42 to 800 2 1 0.7
range 50 180 to 3200 8 4 3
to 1 400 000 and 5 800 000 poise, 100 600 to 12 800 32 16 10
respectively. See ASTM D 2171, Table X2. 200 2400 to 52 000 128 64 40
l Provided with certificate of calibration and 400R 9600 to 1 400 000 500 250 160
instruction sheet. See page 24 for 800R 38 000 to 5 800 000 2000 1000 640
significance of expanded uncertainty.
l Holder is not supplied. For holder, see
9726-N15 on page 37. Calibrated Calibrated
Size AIVC Series AIVC-RO Series
Asphalt Institute Vacuum 25 9724-R50 9724-T50
Calibrated AIVC-RO Series
50 9724-R55 9724-T55
Same as 9724-R50 series, but with 100 9724-R60 9724-T60
permanently attached round metal holder, for
hole 51 mm (2 inches) in diameter. 200 9724-R65 9724-T65
400R 9724-R70 9724-T70
9724-R50 Series
800R 9724-R75 9724-T75

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
75 36 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Glass Capillary Viscometers

CANNON Vacuum Viscometers & Viscometer Holders


®

ASTM D 2171

Cannon Modified Koppers Specifications for Cannon Modified


Vacuum Viscometers Koppers Vacuum Viscometers
Size Viscosity Approx. constant
Calibrated MKVC Series range, poise poise/second at
For measurement of viscosity of asphalts at 60°C 300 mm Hg vacuum
(140°F) in accordance with ASTM Standard D 2171. at B at C at D
Applicable to materials with viscosity range of 42 to 25 42 to 800 2 1 0.7
200 000 poise. Two-piece design, with sample
reservoir and calibrated capillary. Vacuum of 300 mm 50 180 to 3200 8 4 3
Hg must be applied to top of capillary tube. 100 600 to 12 800 32 16 10
l Minimum sample volume 2 mL. Require liquid bath 200 2400 to 52 000 128 64 40
depth of 178 mm (7 inches). 400 9600 to 200 000 500 250 160
l Provided with certificate of calibration and
instruction sheet. See page 24 for significance of
expanded uncertainty. Size Calibrated MKVC Series
l Holder is not supplied. For holder, see 9726-N30.
25 9724-W50
50 9724-W55
100 9724-W60
200 9724-W65
400 9724-W70

9724-W50 Series Removable Holders for Cannon Viscometers


All CANNON viscometer holders fit holes 51 mm (2 inches) in diameter. Rubber holders are made of
neoprene rubber. Plastic holders are self-aligning and hold the viscometer firmly by means of a spring
clip. Metal holders are also self aligning, and are constructed of chrome-plated machined brass. All
metal holders are provided with a handle.

Viscometer Holder Selection Chart


Catalog Numbers For
Type of Holder Viscometer Holders
Viscometer Plastic Rubber Metal Holder Catalog #
Cannon-Fenske Routine H100 H110 H113 H100 9726-M50
Cannon-Fenske Opaque H100 H110 H113 H101 9726-M70
Cannon-Ubbelohde H101 H120 H131 H110 9726-M53
Ubbelohde — — H130 H113 9726-M79
Cannon-Manning Semi-Micro H100 H110 — H115 9726-M95
Cannon-Ubbelohde Semi-Micro H101 H120 H131 H120 9726-M73
Cannon-Ubbelohde Dilution H101 H120 H131 H130 9726-M82
Cannon-Ubbelohde Four-Bulb H101 H120 H131 H131 9726-M76
Viscometer Holders Shear Dilution H132 9726-M98
Zeitfuchs® Cross-Arm — H115 H132 H133 9726-M99
BS/IP/SL Suspended Level — — H133 H140 9726-N15
BS/IP/SL(S) Suspended Level — — H133 H141 9726-N30
BS/IP/MSL Miniature Suspended Level — — H133 H142 9726-N45
BS/IP/RF U-Tube Reverse Flow — H142 —
Cannon-Manning Vacuum — H140 —
Asphalt Institute Vacuum — H140 —
Modified Koppers Vacuum — H141 —
Cannon-Ubbelohde Filter Stick — H142 —

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com 37 76
Viscosity Standards

CANNON General Purpose Viscosity Standards


®

The Table below provides kinematic viscosity values for the General Purpose Viscosity Standards from CANNON Instrument
Company. CANNON certified standards are useful for calibration and verification procedures for all types of viscometers,
including glass capillary viscometers, rotational viscometers, falling ball viscometers, cup viscometers, and others. Calibration
or verification with a certified viscosity standard helps to ensure the reliability of measurements made with any viscometer.
For information on CANNON's available in-house testing services please see the inside back cover of this catalog.
Table 1 • Kinematic Viscosity in mm2/s (Centistokes)
Catalog Viscosity 20°C 25°C 37.78°C 40°C 50°C 60°C 80°C 98.89°C 100°C SUS SUS
Number Standard 68°F 77°F 100°F 104°F 122°F 140°F 176°F 210°F 212°F 100°F 210°F
9727-C10 N.4† 0.47 0.45 0.41 0.40 — — — — — — —
9727-C15 N.8† 0.74 0.70 0.61 0.60 — — — — — — —
9727-C20 N1.0† 1.3 1.2 1.0 0.97 0.91 — — — — — —
9727-C22 N2 2.9 2.6 2.1 2.0 1.7 — 1.1 0.95 0.93 — —
9727-C25 S3* 4.6 4.0 3.0 2.9 2.4 — 1.5 1.2 1.2 — —
9727-C27 N4 6.7 5.8 4.2 4.0 3.2 — 1.9 1.5 1.5 — —
9727-C30 S6 11 8.9 6.0 5.7 4.4 — 2.4 1.8 1.8 — —
9727-C31 N7.5 14 12 8.0 7.5 5.8 — 3.1 2.3 2.3 — —
9727-C32 N10 21 17 11 10 7.3 — 3.5 2.5 2.4 — —
9727-C34 N14 30 25 15 14 10 — 5.0 3.5 3.4 — —
9727-C35 S20 44 34 20 18 13 — 5.6 3.9 3.8 100 —
9727-C36 N26 57 46 27 25 18 — 7.9 5.3 5.2 130 —
9727-C37 N35 87 66 35 32 21 — 8.5 5.4 5.3 170 —
9727-C38 N44 110 86 48 44 30 — 12 7.7 7.5 220 —
9727-C40 S60 160 120 60 54 35 — 12 7.7 7.5 280 —
9727-C41 N75 200 150 82 75 50 — 19 12 12 380 —
9727-C42 N100 330 230 110 97 60 — 19 11 11 500 —
9727-C43 N140 400 300 160 140 90 — 31 19 18 720 —
9727-C45 S200 550 400 200 180 110 — 35 22 20 925 105
9727-C46 N250 770 570 280 250 160 — 51 30 29 1300 140
9727-C47 N350 1000 720 340 310 190 — 55 34 32 1570 160
9727-C48 N415 1400 990 470 420 250 — 77 43 41 2180 200
9727-C50 S600‡ 1700 1300 590 520 310 — 93 51 50 — 240
9727-C51 N750 2600 1900 850 750 440 — 130 68 66 — —
9727-C52 N1000 3400 2400 — 940 550 350 150 — 80 — —
9727-C53 N1400 5100 3600 — 1400 820 510 220 — 120 — —
9727-C55 S2000 8300 5300 1900 1600 800 — 160 75 72 — 360
9727-C56 N2500 8400 6000 — 2500 1500 950 430 — 230 — —
9727-C57 N4000 20000 12000 — 3400 1600 850 290 — 120 — —
9727-C58 N5100 28000 18000 — 5100 2500 1300 420 — 170 — —
9727-C60 S8000 41000 25000 8000 6700 3200 — 530 — 215 — —
9727-C61 N10200 58000 36000 — 10200 4900 2500 775 — 300 — —
9727-C62 N15000 77000 47000 — 13000 6100 3000 980 — 360 — —
9727-C63 N18000 103000 64000 — 18000 8500 4300 1320 — 500 — —
9727-C65 S30000 — 79000 27000 22000 11000 — 1700 — 630 — —
* cSt value at –40°C/–40°F (70) also provided
† Incurs additional shipping charges due to low flash point
‡ SFS value at 122°C (150) also provided

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
77 44 800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com
Viscosity Standards
CANNON General Purpose Viscosity Standards
®

Dynamic Viscosity Values

Quality/Traceability
The Table below provides dynamic viscosity values for the General Purpose Viscosity Standards from CANNON Instrument
Company. Viscosities for all standards sold by CANNON are based on the National Institute of Standards and Technology
(NIST) value of 1.0016 mPa·s (centipoise) for water at 20°C (68°F).
All CANNON viscosity standards meet the precision specifications of ASTM D 445/446 and ISO 3104/3105, and are
traceable to the NIST (National Institute of Standards and Technology). CANNON viscosity standards are manufactured and/or
certified in CANNON's ISO 9001:2000 registered and A2LA-accredited laboratory. A2LA has accredited CANNON for
technical competence in the field of calibration (certificate number 1262.01). The accreditation includes the determination of
kinematic and dynamic viscosity (including that of viscosity standards and customer samples) in accordance with ISO/IEC
17025-1999. All calibrations are based on the master viscometer procedures detailed in ASTM D 2162.
CANNON has been delegated by NIST to provide U.S. national measurement standards and to issue calibration
and measurement certificates for certified liquid viscosity reference standards.
Table 2 • Dynamic Viscosity in mPa·s (Centipoise)
Catalog Viscosity 20°C 25°C 37.78°C 40°C 50°C 60°C 80°C 98.89°C 100°C SUS SUS
Number Standard 68°F 77°F 100°F 104°F 122°F 140°F 176°F 210°F 212°F 100°F 210°F
9727-C10 N.4† 0.32 0.30 0.28 0.27 — — — — — — —
9727-C15 N.8† 0.64 0.60 0.52 0.51 — — — — — — —
9727-C20 N1.0† 0.93 0.86 0.72 0.70 0.62 — — — — — —
9727-C22 N2 2.2 2.0 1.6 1.5 1.3 — 0.85 0.66 0.65 — —
9727-C25 S3 3.9 3.3 2.5 2.4 1.9 — 1.2 0.90 0.90 — —
9727-C27 N4 5.2 4.5 3.3 3.1 2.5 — 1.4 1.1 1.1 — —
9727-C30 S6 10 8.0 5.3 5.0 3.8 — 2.0 1.5 1.5 — —
9727-C31 N7.5 11 9.5 6.3 5.9 4.5 — 2.4 1.7 1.7 — —
9727-C32 N10 21 16 9.5 8.8 6.3 — 2.9 2.1 2.0 — —
9727-C34 N14 24 20 12 11 8.2 — 3.8 2.7 2.6 — —
9727-C35 S20 37 29 17 15 11 — 4.6 3.0 3.0 100 —
9727-C36 N26 47 37 22 20 14 — 6.2 4.1 4.0 130 —
9727-C37 N35 75 56 30 27 18 — 7.1 4.5 4.4 170 —
9727-C38 N44 92 71 39 36 24 — 9.5 6.0 5.9 220 —
9727-C40 S60 141 104 52 47 30 — 10 6.4 6.2 280 —
9727-C41 N75 160 125 68 61 41 — 16 9.5 9.2 380 —
9727-C42 N100 283 202 95 84 52 — 16 9.5 9.2 500 —
9727-C43 N140 340 250 130 120 74 — 25 15 14 720 —
9727-C45 S200 460 330 160 145 90 — 25 16 14 925 105
9727-C46 N250 650 480 230 210 130 — 41 23 23 1300 140
9727-C47 N350 830 600 280 250 150 — 40 26 24 1570 160
9727-C48 N415 1200 830 390 350 210 — 62 34 33 2180 200
9727-C50 S600‡ 1500 1100 490 440 260 — 75 41 40 — 240
9727-C51 N750 2200 1600 710 620 370 — 100 55 53 — —
9727-C52 N1000 2900 2000 — 800 460 280 120 — 65 — —
9727-C53 N1400 4300 3000 — 1200 680 420 180 — 92 — —
9727-C55 S2000 7300 4700 1700 1400 700 — 140 62 59 — 360
9727-C56 N2500 7100 5100 — 2100 1200 780 350 — 180 — —
9727-C57 N4000 17000 11000 — 2900 1400 730 250 — 100 — —
9727-C58 N5100 25000 16000 — 4500 2100 1100 360 — 140 — —
9727-C60 S8000 33000 20000 7000 5900 2800 — 450 — 200 — —
9727-C61 N10200 52000 32000 — 9000 4200 2100 660 — 250 — —
9727-C62 N15000 68000 41000 — 11000 5400 2600 840 — 320 — —
9727-C63 N18000 92000 57000 — 16000 7500 3800 1140 — 420 — —
9727-C65 S30000 — 71000 23000 20000 9300 — 1400 — 540 — —
† Incurs additional shipping charges due to low flash point.
‡ SFS value at 122°C (150) also provided

CANNON INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PENNSYLVANIA • 16803 • USA
800-676-6232 • 814-353-8000 • Fax: 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com 45 78
Viscosity Standards

High-Viscosity & Low Temperature Viscosity Standards


ASTM D 2170/2171, ASTM D 2983, ASTM D 3829/4684

High-Viscosity Standards
These standards are furnished in a range of viscosities and temperatures which are of particular interest in asphalt viscosity
determinations and the determination of the viscosities of polymers. See ASTM D 2170 and D 2171.
See Table 3 for approximate viscosities. Exact values are supplied with each standard. All data are traceable to the NIST.
Table 3 • High-Viscosity Standards
Approximate Viscosity in mPa·s (cP) Kinematic Viscosity
mm2/s (cSt)
Catalog Viscosity 20°C/ 25°C/ 40°C/ 60°C/ 135°C/ 60°C/ 100°C/
Number Standard 68°F 77°F 104°F 140°F 275°F 140°F 212°F
9727-E10 N600† — 1100 — 160 17 200 —
9727-C52 N1000* 2900 2000 800 280 — 350 80
9727-E15 N2000† — 4900 — 380 21 440 —
9727-C57 N4000* 17 000 11 000 2900 730 — 850 120
9727-E20 N8000 — 20 000 — 1400 — 1600 —
9727-C62 N15000* 68 000 41 000 11 000 2600 — 3000 360
9727-E25 N30000 130 000 80 000 — 4700 — 5400 —
9727-E27 N62000 — 200 000 — 13 000 — — 1600
9727-E29 N150000 — 420 000 — 24 000 — — 3000
9727-E30 N190000 900 000 520 000 140 000 33 000 — — 4000
9727-E35 N450000 — 1 600 000 — 100 000 2350 — —
9727-E40 N2700000 — 5 300 000 — 340 000 — — —
9727-E42 N2700000SP Dynamic Shear Rheometer (DSR) standard at 52, 58, 64, 70, & 76°C
* Additional data provided with these standards are listed in Tables 1 and 2.
† Additional data provided at 135°C.

Low-Temperature Standards
For Low Temperature Gear Oils (SAE 70W, 75W, 80W and 85W). Also for ATF fluids. See ASTM D 2983.
Catalog Viscosity
Number Standard Table 4 • Low-Temperature Viscosity Standards
9727-G10 N27B Viscosities in centistokes/centipoise at –40°F (24,000/20,000), –30°F (12,000/10,000),
–20°F (6100/5200), –15°F ( 4500/4000), –10°F (3500/3000), and 0°F (2000/1700)
9727-G12 N27C Viscosities in centistokes/centipoise at –40°F (45,000/40,000), –30°F (22,000/20,000),
–20°F (12,000/10,000), –15°F (8000/7000), –10°F (6100/5000), and 0°F (3500/3000)
9727-G15 N115B Viscosities in centistokes/centipoise at –20°F (160,000/140,000), –15°F (110,000/95,000),
–10°F (75,000/65,000), 0°F (37,000/30,000), +10°F (20,000/17,000), and +20°F (12,000/9500)
9727-G25 N60B 150,000 mPa·s (cP) at –55°C
9727-G30 N120B 150,000 mPa·s (cP) at –40°C
9727-G35 N480B 150,000 mPa·s (cP) at –26°C
9727-G40 N1400B 150,000 mPa·s (cP) at –12°C
9727-G45 N2B* Jet Fuel standard, 8 mm2/s (cSt) at –20°C
9727-G50 N74B 3400 mPa·s (cP) at 0°C
9727-G55 N38B 3100 mPa·s (cP) at –10°C
9727-G60 N25B 8000 mPa·s (cP) at –25°C
9727-G65 N14B 17,000 mPa·s (cP) at –40°C
* Incurs additional shipping charges due to low flash point
For CANNON Mini-Rotary Viscometer (CMRV) (see ASTM D 3829 and ASTM D 4684).
Catalog Viscosity
Number Standard Table 5 • Approximate Viscosity in mPa·s (Centipoise)
9727-T10 N105B CMRV viscosity standard, 30,000 mPa·s (cP) at –20°C; 56,200 mPa·s (cP) at –25°C
9727-T20 YS-30 Exhibits yield stress of 70 Pa ±35 Pa at –30°C
9727-T25 YS-35 Exhibits yield stress of 70 Pa ±35 Pa at –35°C

79
Viscosity Standards

CANNON CCS Viscosity Standards


®

CANNON CL Standards provide certified dynamic viscosity data (in cP or mPa·s) at temperatures from –5°C to –35°C. These
standards are used to calibrate the CANNON Cold-Cranking Simulator (CCS) oil testing (see ASTM D 5293 and SAE
Specification J300).
Table 6 • Dynamic Viscosity in mPa·s (Centipoise)
Catalog Viscosity
Number Standard* –5°C –10°C –15°C –20°C –25°C –30°C –35°C
9727-N02 CL080 — — — — — — 800
9727-N04 CL090 — — — — — — 1200
9727-N06 CL100 (CL10) — — — — — 900 1700
9727-N07 CL110 — — — — — 1400 2500
9727-N08 CL120 (CL12) — — — — 800 1600 3200
9727-N09 CL130 — — — — 1500 2700 4800
9727-N10 CL140 (CL14) — — — 800 1600 3300 7000
9727-N11 CL150 — — — 1200 2500 4500 8000
9727-N12 CL160 (CL16) — — — 1200 2500 5500 11000
9727-N13 CL170 — — 1100 1800 3500 6200 11000
9727-N14 CL190 (CL19) — — 900 1800 3500 7400 17200
9727-N16 CL200 — — 1200 2400 4300 7400 13000
9727-N18 CL220 (CL22) — — 1300 2500 5000 11000 25000
9727-N20 CL240 — — 1700 3400 6000 10000 20000
9727-N22 CL250 (CL25) — 900 1800 3500 7300 15900 33000
9727-N24 CL260 — 1200 2500 4400 7200 14000 —
9727-N26 CL280 (CL28) — 1300 2500 5000 9300 21000 50000
9727-N28 CL300 — 1700 3500 6000 11000 18000 —
9727-N30 CL320 (CL32) — 1800 3500 7300 15900 32000 —
9727-N31 CL340 — 2500 4200 7000 12000 — —
9727-N32 CL380 (CL38) 1600 2900 5800 13000 26000 — —
9727-N33 CL420 — 5000 8500 15000 — — —
9727-N34 CL480 (CL48) 2300 4500 9500 21000 — — —
9727-N35 CL530 — 5900 10000 18000 — — —
9727-N36 CL600 (CL60) 3700 7300 15600 — — — —
9727-N37 CL680 — 9500 — — — — —
9727-N38 CL740 (CL74) 6000 12000 — — — — —
9727-N43 Set of 14 Set of CCS-2050/2100 Low Temp Standards (includes CL10, CL12, CL14, CL16, CL19, CL22, CL25(2),
CL28, CL32, CL38, CL48, CL60, & CL74)
9727-N42 Set of 8 Set of CCS Low Temp Standards (includes CL14, CL19, CL22, CL25, CL28, CL32, CL48, & CL74)
* CL Standards nomenclature has changed. The previous standard names are indicated in parentheses.

" NIST will delegate responsibility for U.S. national standards for certified liquid
viscosity reference standards to the CANNON Instrument Company.
James R. Whetstone, Chief
Process Measurements Division/Chemical Science and Technology Laboratory "
National Institute of Standards and Technology
January 6, 2003

80
Viscosity Standards
CANNON Cup Viscosity Standards
®

Table 7 • Certified Data at 25°C Viscosity Standards


Catalog Viscosity Kinematic Zahn Cup Shell Cup Ford Cup for Zahn, Shell,
Number Standard viscosity
Size Drain Size Drain Size Drain & Ford Cups
mm2/s Approximate values
(cSt) time, time, time,
sec sec sec are shown in Table 7.
Exact values
9727-W10 C6 8.9 — — 1 52 — — calculated from the
— — 2 20 — — equations in Table VIII
9727-W12 C10 17 1 45 2 35 1 70 are supplied with
— — 2¼ 22 — — each standard.
9727-W15 C20 34 1 60 2 64 2 42
2 24 2½ 40 — — For more information
— — 3 24 — — on cups, contact
9727-W20 C35 66 2 33 2½ 74 2 64
Cannon Instrument
— — 3 46 3 35 Company.
— — 3½ 32 — —
9727-W25 C60 120 2 48 3½ 57 3 58
— — 4 36 4 36
9727-W30 C100 230 3 27 4 68 4 64
4 21 5 36 — —
9727-W35 C200 460 3 47 5 72 5 40
4 36 6 29 — —
9727-W40 C350 850 4 62 6 53 5 72
5 37 — — — —
9727-W45 C600 1300 5 54 — — — —

Table 8 • Ranges of Viscosity Cups Equations


Cup Equations‡ where Drain time Kinematic viscosity for Zahn, Shell,
Number t = Time of flow, seconds seconds centistokes & Ford Cups
n = kinematic viscosity, mm2/s Equations for conversion of
min max min max
kinematic viscosity or viscosity into
Zahn Cups (See ASTM D 4212) other viscometric units are given
1 n = 1.1 (t – 29) 35 80 5 60 here. All data must be at 25°C. Note
2 n = 3.5 (t – 14) 20 80 20 250
that the equations apply only
between the minimum and
3 n = 11.7 (t – 7.5) 20 80 100 800 maximum drain times shown.
4 n = 14.8 (t – 5) 20 80 200 1200
5 n = 23t 20 80 400 1800
Shell Cups (See ASTM D 4212)
1 n = 0.226 (t – 13) 20 80 2 20 Need SAE
2 n = 0.576 (t – 5) 20 80 10 50 Viscosity Check Oils?

3
n = 0.925 (t – 3)
n = 1.51 (t – 2)
20
20
80
80
20
30
80
120
See page 43!
3½ n = 2.17 (t – 1.5) 20 80 40 170
4 n = 3.45 (t – 1) 20 80 70 270
5 n = 6.5 (t – 1) 20 80 125 520
6 n = 16.2 (t – 0.5) 20 80 320 1300
Ford Cups (See ASTM D 1200)
1 n = 0.49 (t – 35.0) 55 100 10 35
2 n = 1.44 (t – 18.0) 40 100 25 120
3 n = 2.31 (t – 6.58) 20 100 49 220
4 n = 3.85 (t – 4.49) 20 100 70 370
5 n = 12.1 (t – 2.00) 20 100 200 1200
‡ While the equations in Table VIII are widely used, the equation n = at - b/t is often more applicable,
especially at low flow times and low kinematic viscosity.

81
Viscosity Standards
CANNON® Viscosity Standards & Flash Point Reference Materials

High-Temperature Standards Catalog


Number
Viscosity
Standard Table 9 • High-Temperature Standards
For High-Temperature High-Shear 9727-U50 HT39 2.0 mPa·s (cP) at 150°C, 4.7 mPa·s (cP) at 100°C
Viscometer measurements at 150°C and 9727-U55 HT75 2.7 mPa·s (cP) at 150°C, 7.0 mPa·s (cP) at 100°C
6 -1
10 s shear rate.
9727-U60 HT150 3.7 mPa·s (cP) at 150°C, 10.5 mPa·s (cP) at 100°C
See ASTM D 5481.
9727-U65 HT240 5.0 mPa·s (cP) at 150°C, 15.4 mPa·s (cP) at 100°C
9727-U70 HT390 7.0 mPa·s (cP) at 150°C, 23.5 mPa·s (cP) at 100°C
9727-U92 HTNN-1 Non-Newtonian standard, 3.6 mPa·s (cP) at 150°C and 106s-1
9727-U94 HTNN-2 Non-Newtonian standard, 3.1 mPa·s (cP) at 150°C and 106s-1

Viscosity Standards Table 10 • Certified Data at 25°C


Catalog Viscosity
for Thomas®-Stormer® Number Standard Viscosity (Centipoise) Krebs Unit (KU)
Viscometer 9727-Y10 S200(KU) 330 61.2
The KREBS UNIT (KU) is often used for 9727-Y15 N350(KU) 600 77
THOMAS-STORMER Viscometers. 9727-Y20 K400(KU) 940 84
9727-Y25 S600(KU) 1060 88
9727-Y30 N1000(KU) 2000 106

Silicone Viscosity Standards Table 11 • Silicone Viscosity Standards


Approximate
for Rotational Viscometers Viscosity mPa·s
CANNON silicone standards are intended for calibration of Brookfield type Catalog Viscosity (Centipoise)
rotational viscometers, such as the 2020 Series Rotary Viscometers (page 51) Number Standard 25°C (77°F)
offered by CANNON. Viscosity data at 20, 23, 24, 25, 26, 27, and 40°C is 9727-Z10 RT5 4.9
provided. CANNON has been granted ISO 9001 registration for the 9727-Z14 RT10 9.4
manufacture and certification of viscosity standards, including the silicone 9727-Z18 RT50 48
standards listed here. 9727-Z22 RT100 96
Viscosities for all silicone standards sold by CANNON are based on the 9727-Z26 RT500 480
National Institute of Standards and Technology (NIST) value of 1.0016 mPa•s 9727-Z30 RT1000 960
(centipoise) for water at 20°C (68°F). All silicone standards are traceable to the 9727-Z34 RT5000 4800
NIST. Silicone standards are sold in 500 mL bottles. 9727-Z38 RT12500 12000
Note: The listed silicone standards are intended for use with rotational viscometers only. 9727-Z42 RT30000 29000
Silicone standards are not recommended for glass capillary viscometers (including vacuum
viscometers), metal cup-type viscometers (such as ISO, Zahn, Ford, Shell, and Fisher cups), 9727-Z46 RT60000 58000
or any other viscometers providing kinematic viscosity – CANNON hydrocarbon-based oil 9727-Z50 RT100000 97000
standards should be used for these types of viscometers.

Flash Point Reference Materials Table 12 • Flash Point Reference Materials


ASTM D 56 ASTM D 92 ASTM D 93
CANNON Instrument Company can now provide Catalog Nominal Flash Nominal Flash Nominal Flash
Flash Point Reference Materials for calibrating Flash Number Product ID Point (°C) Point (°C) Point (°C)
Point Testers. A total of six different standards are 9727-A10 FPRM10* 50 – 55
available for the following ASTM methods: ASTM D 9727-A15 FPRM11* 66 – 72
56, ASTM D 92, and ASTM D 93. The data
9727-A20 FPRM14 – 116 113
provided for each standard includes the average
flash point for the lot and the standard deviation 9727-A25 FPRM16 – 137 132
observed from 20 tests in 10 different laboratories. 9727-A30 FPRM2D – 163 155
Nominal flash points for the various standards are 9727-A35 FPRM4D – 224 218
shown in the accompanying table. The four
standards with lower flash points are made from high *Incurs additional shipping charges due to low flash point
purity alkane materials, while the two standards with
higher flash points are made from narrow
distributions of decane oligomers. Flash Point
Standards are sold in 200 mL bottles (155 g).

82
Asphalt Industry Test Equipment
CANNON DVR 1000 Series Digital Vacuum Regulators
®

ASTM D 2171

l Solid-State Components—Contains No Mercury


l Preset for Vacuum Viscometry per ASTM D 2171
l Horizontal or Vertical Configuration
l Convenient Keypad Controls

The CANNON DVR 1000 Series Digital Vacuum Regulators are designed for precise measurement and control of vacuum at
300 mm Hg below atmospheric pressure. Unlike other vacuum regulators, the solid-state DVR-1000 and DVR-1500 models
use no mercury.
The DVR meter displays the amount of vacuum in mm/Hg or one of nine other units of measurement selected using a keypad
on the meter.
In asphalt laboratories the Digital Vacuum Regulator (DVR) may be used in conjunction
with Cannon-Manning, Asphalt Institute, and Modified Koppers vacuum viscometers for
measurement of highly viscous materials such as asphalt cement at 60°C (140°F) according
to ASTM D 2171. The DVR is also useful in other laboratory systems where accurate
measurement and control of vacuum is required.
Two basic versions of the DVR are available. The DVR-1500 is equipped with an internal
vacuum pump and does not require an external vacuum source. The DVR-1000 uses the
customer’s in-house vacuum system.
The internal set points for the instrument gauge are preset at CANNON to regulate
vacuum at 300 ± 0.5 mm Hg below atmospheric pressure (the vacuum required by
ASTM D 2171). These set points may be altered to fit the user’s specific needs within the
DVR’s operating range of 28 to 410 mm
Hg below atmospheric pressure.
Both the DVR-1000 and DVR-1500 are
available in either horizontal or vertical
configurations (the horizontal version will
support the weight of a constant
temperature bath like the CANNON
CT-1000 or CT-2000).
The instrument housing is made of durable
enameled steel and is supported on non-slip
rubber feet.

The CANNON DVR 1000 Series Digital Vacuum Regulators

DVR 1000 Series Specifications Order Information


Dimensions: 470 mm wide x 458 mm deep x 172 mm Catalog # Item Description
high (18.5 x 18 x 6.75 inches) [in the 9726-T20 DVR-1000H (DVR without vacuum pump,
horizontal position] horizontal orientation, 120V, 100 watts)
Weight: 20 kg (45 lbs) 9726-T23 DVR-1000HF (DVR without vacuum pump,
Range: Preset at 300 mm Hg below atmospheric horizontal orientation, 240V, 100 watts)
pressure, but may be adjusted to accurately 9726-T25 DVR-1000V (DVR without vacuum pump,
control from 28 mm to 410 mm Hg below vertical orientation, 120V, 100 watts)
atmospheric pressure
9726-T27 DVR-1000VF (DVR without vacuum pump,
Reading Accuracy: ± 0.05% of reading ± the least significant vertical orientation, 240V, 100 watts)
digit (includes combined effects of linearity,
repeatability, hysteresis, and temperature) 9726-T60 DVR-1500H (DVR with vacuum pump,
[NIST certification is supplied] horizontal orientation, 120V, 175 watts)
Vacuum Regulation ± 0.5 mm Hg (as required by ASTM D 9726-T63 DVR-1500HF (DVR with vacuum pump,
Accuracy: 2171) horizontal orientation, 240V, 175 watts)
Operating Temperature: 0 to 40°C (32 to 104°F) 9726-T65 DVR-1500V (DVR with vacuum pump,
vertical orientation, 120V, 175 watts)
Upper/Lower Safety Limit: 746 mm Hg below atmospheric pressure
9726-T67 DVR-1500VF (DVR with vacuum pump,
vertical orientation, 240V, 175 watts)

83
Pro-Pak®

CANNON Pro-Pak Distillation Packing


®

l Ideal for Laboratory and Pilot Scale


Distillation Columns
l High Efficiency Design
l Unique Wetting Properties
l Atmospheric and Reduced Pressure
Distillation Applications

Pro-Pak is a high efficiency distillation column packing


with unique wetting properties. Applications range
from laboratory and pilot plant distillation columns to
those used in small manufacturing plants. Pro-Pak is
self wetting, so equilibrium time is reduced —
distillations can be stopped and started without lengthy
delay. The entire surface of the packing is available for
mass transfer, contributing to its high efficiency. The
shape of the packing eliminates interlocking and
makes filling and emptying the column as easy as
pouring water.
Pro-Pak is formed from metal ribbon through which
CANNON Pro-Pak Distillation Packing more than 1000 tiny holes per square inch have been
punched. As the points of the die push through the
metal, jagged burrs are formed on the reverse side of
the ribbon. Pro-Pak’s self-wetting properties derive from a rectangular network of capillaries on the surface of the metal,
produced by the combined effect of the holes in the ribbon and the roughness of the burrs. These properties ensure that
Pro-Pak stays wet in service in the distillation column and retains its high efficiency. Because the surface area of the packing
remains effective under low liquid loadings, high column efficiency is maintained at vapor velocities very much lower than the
flooding velocity.
Pro-Pak can be used for both atmospheric and reduced pressure distillations down to 10 mm Hg condenser pressure. As the
operating pressure decreases, efficiency increases. The HETP (height equal to a theoretical plate) varies from 0.5 inch to 1.4
inch in a 1-inch diameter column, depending on operating pressure and reflux rate.
The large surface area made effective by the wettability of Pro-Pak allows for efficient mass transfer between the liquid and
vapor phases. The surface area is approximately 576 square feet per cubic foot for the 0.16-inch size and 372 square feet per
cubic foot for the 0.24-inch size. The packing factor is 693 for the 0.16-inch size and 420 for the 0.24-inch size.
High free space is especially important in vacuum distillation (for reduction of pressure drop) and in extractive distillation and
absorption (where high liquid or vapor loading is common). Pro-Pak has high free space: 94% for the 0.16-inch size and 96%
for the 0.24-inch size. Because of its high free space, Pro-Pak has a high flood rate.
In addition to type 316 stainless steel packing, listed below, Pro-Pak is also available in Monel, nickel, and other metals on
special order. Discounts are provided for bulk quantities.
CANNON manufactures special interdistributor screens and hold down screens that provide optimum efficiency for Pro-Pak
columns. These screens are manufactured on a custom basis to fit any specified column diameter.
Our Pro-Pak technical bulletin contains additional information. A free sample of Pro-Pak can also be supplied. Please contact
CANNON to receive a free copy of the Pro-Pak bulletin and a sample of the packing.

Order Information
Catalog # Item Description
3947-A20 316 Stainless Steel Packing, 0.16-inch. 1 L
3947-A30 316 Stainless Steel Packing, 0.16-inch. cubic foot
3947-A60 316 Stainless Steel Packing, 0.24-inch. 1 L
3947-A70 316 Stainless Steel Packing, 0.24-inch. cubic foot Interdistributor
& Hold-Down Screens
also Available!

84
®

UVS Viscosity Systems

85
Measuring Instruments

Viscosity Systems UVS® basic

UVS® BASIC is the new introduction viscosity


measurement device which is capable of
determining the intrinsic viscosity of polymers as
well as the kinematic viscosity of mineral oil
products in compliance with DIN51562-FF, ASTM
D 445 and DIN ISO EN 3105 and all polymere
standards (see table typical examples)
The measuring instrument not only saves place of
work, but also measures automatically precise
and reproducible processing times.

The UVS® BASIC consists of the measuring


unit UVS® MD.
The UVS® BASIC is expandable.

It is always possible to use in the case either


another unit of measurement UVS® MD or the
controlled vacuum pump UVS® DRAIN for
automatic sample exhauster. This exhauster can
also be integrated in a sample changer SC8.
Thus, the instrument becomes a fully automatic
machine for polymer applications.

All UVS ® measuring instruments work


independently and are able to operate multiple
devices with a PC.

86
Measuring Instruments

UVS® basic Viscosity Systems

UVS BASIC, Configuration

Cleaning Measuring
Configurations Measuring Sample
solvents
Waste
samples at
Max.
UVS basic points filling bottles positions
per MP same time

Manual
UVS basic MD 1 Manual 0 1 1
X
Manual
UVS basic MD / MD 2 Manual 0 2 2
X
Manual
UVS basic MD + UVS basic MD / MD 3 Manual 0 3 3
X
Manual
2 x UVS basic MD / MD 4 Manual 0 4 4
X
Semiauto
1 x UVS basic MD / Drain 1 Manual 1 1 1
X
Automatic
1 x UVS basic MD / Drain + SC8 1 Automatic 1 1 24
2
1 x UVS basic MD / MD + SC8 + UVS Automatic
2 Automatic 1 ~1.5 24
waste 1
1 x UVS basic MD / MD + 2 x SC8 + Automatic
2 Automatic 1 2 48
UVS waste 2
2 x UVS basic MD / MD + 2 x SC8 + Automatic
4 Automatic 1/2 ~2.5 48
UVS waste 1 (2)
2 x UVS basic MD / MD + 4 x SC8 + Automatic
4 Automatic 1 4 96
UVS waste 4
4 x UVS basic MD / MD + 8 x SC8 + Automatic
8 Automatic 1 8 192
UVS waste 8

other combinations possible

87
Measuring Instruments

Viscosity Systems UVS® MD

UVS MD, the measuring device for flow time


measuring of Newtonian fluids in common
capillary viscometers with precision to the
millisecond. The UVS MD is suitable for all
measurements of polymer solutions and mineral
oils products.
UVS MD, the Meniscus
measuringdetection
device foris effected
flow time
optical through light barriers or with
measuring of Newtonian fluids in common thermistors
by determining
capillary the heat
viscometers withconductance.
precision to The
the
optical measurement
millisecond. The UVS amplifier supports
MD is suitable for the
all
scanning of a third safety light barrier
measurements of polymer solutions and mineral(Triple
Detection)
oils products. Meniscus detection is effected
optical through light barriers or with thermistors
by determining the heat conductance. The
optical measurement amplifier supports the
scanning of a third safety light barrier (Triple
Detection)

UVS Drain, the device is responsible for


automated cleaning. This device controls the
discharging of the sample, rinsing with solvent,
drying and discharging solvent to the waste
bottle. Sensor detection for a full waste bottle
complies safety requirements. The cleaning
process can be optimzed by configuration to
save time and get best results.

88
Measuring Instruments

UVS® MD Viscosity Systems

UVS MD, Technical Data


Measuring range (time)
Time 0.100 to 9999.000 s
Resolution 0.001 s
Accuracy (flow time : 0 - 100s) ± 0.001 s
Accuracy (flow time : > 100 s) <10 ppm
Ambient temp. error of time base (15 - 35°C) <4 ppm
Measuring range (viscosity)
2
Pressure mode 0.35 to 1000 mm /s
2
Suction mode 0.35 to 10000 mm /s
Pumping pressure power + 350 hPa
Pumping suction power - 350 hPa
Configuration Parameters (with PC)
Tempering period 0 to 20 min
Additional quite tempering period 0 to 20 min
No. of measurements 1 to 10 (250)
pump power range 1 to 100 %
Rising time (ramp) of pump power 0.02 to 2 % /s
Connections
UVS LB triple light barrier 5 pin connector with screw lock
SI Analytics light barriers with adapter cable
TC-Viscometers 4 pin connector with screw lock
Viscometer tubes M12x1,25
Communication backplane DIN 41612
Size
Dimensions (W x H x D) 70 x 128 x 170 mm
Weight 2 kg
Materials
Pumps PTFE, FFKM, PPS housing
Valves ETFE, FFKM
Cover anodized aluminium
Internal tubes, seals PTFE, ETFE

89
Sample Changer

Viscosity Systems Sc8

SC8 the sample changer with 8 and respectively 24 positions in


dependance of sample bottle size.
With integration of a SC8 the UVS Basic becomes a fully
automatic machine. The sampler can serve one or two
viscometers. In case of two viscometers the filling and rinsing
procedure works in an alternating operation mode. The
turntable is available for common laboratory 100 ml sample
bottles, 50 ml centrifugal tubes, 30 ml disposable PP beakers
and 20 ml micro tubes. An adaption of individual sample
vessels are possible on request. With the stage of extension -
2x UVS Basic and 4x SC8 - the system have up to 96 sample
positions and can measure 80 samples per hour.
The rinsing of the sample transfer tube and piston pump is
realized with build-in ports for a rinsing and a drying solvent.
Additional the recommended rinsing with next sample is also
configurable.
The used valves are chemical resistant and have a lean dead
volume. Sample filtration with filter backwashing is available as
an option for low viscous samples.

! Sensor for recognition of empty sample flasks or


unused positions
! Inputs for fluid level sensor
! Configurable sample transfer speed
90
Viscometer baths

Chiller Viscosity Systems

Our viscometer baths are in compliance with


DIN 51 562 (Part 1) and ASTM D 445.
Each our viscometer bath models are for use
with temperatures between +10 °C and
+100 °C.
For working temperatures below 35°C the
chiller HMC is recommended to maintain the
temperature constancy.
The models JB18 are able to work at high
temperature conditions till 150°C. The
resolution of 0.01°C is displayed. Other
temperatures and bath sizes are on request
deliverable.

JB 18

TB 2 -2V (4 V) HMC
91
VisKey - The Key to Viscosity

Viscosity Systems Control and Evaluation Software

PCalculation of kinematic, dynamic, relative


and intrinsic viscosity
PDetermination of capillary constant by
system calibration, blank value (t0),
viscosity index, enzyme kinetic, dilution
viscoity
PFormula editor for individual calculations
PNo limitation of method numbers
PEach measuring station can work in a
different method
PViscometer database
PUser level access controlled software with
three levels: administrator, authorized user,
worker
PAccess restrictions can be disabled
PMultitasking operations – each measuring
point works independent
PPossibility to enter sample ID using a
barcode scanner
PLogbook features, CFR21 part 11
compliant
PResult database with screening and sorting
functions
PSample changer functions like priority
sample, subsequent addition or removal of
samples
PSupported operating system Windows XP,
Vista, Win7, Win8
PSupported periphery: UVS Basic, UVS
Tower, UVS Clean, SC8 92
Light Barrier

UVS® LB Viscosity Systems

UVS LB, the Triple Detection light barrier made


of PVDF.
The stand with integrated viscometer fixation is
suited for all common capillaries. Viscometers.
with fixing bracket are also usable. The light
barriers are working with infrared light and are
able to detect certainly the meniscus pass. The
third and topmost light barrier is used as "over
suck" protection. For measurements with Micro-
Ubbelohde viscometers this light barrier can
increases the reproducibility of values through
constant kept start conditions.
Additional fixation holder for Canon-Fenske
Routine, Canon-Fenske Ubbelohde, Witeg-,
Paragon-, Tamson-Ubbelohde available.
93
Viscometers

Viscosity Systems

Viscometer Ubbelohde Micro TC TC Micro Micro Cannon- Cannon-


type Ubbelohde Ubbelohde Ubbelohde Ostwald Fenske Fenske
routine reverse flow
Application
characteristics
Polymer solutions ++ + - - + + -

Polymer solutions + ++ - - + - -
Ultra short flowtimes

Mineraloil ++ ++ ++ ++ + + +
transparent

Mineraloil opaque - - ++ ++ - - ++*


Foaming liquids o o o o + + o

Liquid mixture with


highly volatile + + + + ++ ++ o
components

sample volume (ml) 15-20* 3-4 18-22 3-4 2 7-10 12


Exact volume no no no no yes (yes) yes
necessary

*4-leg re- Short Short * only manual


commended flowtime flowtime measurement
18-22 ml

Ubbelohde with Thermistor Cannon-Fenske Routine Micro-Ubbelohde ASTM-Ubbelohde

94
ASTM Equipment

1.6 ASTM D445 High Temperature Viscosity Baths - TV Series


ASTM D445 - ASTM D2162 - ASTM D2170 - ISO/EN 3104 - IP 71 - DIN 51366

This test method specifies a procedure for the determination of the


kinematic viscosity of liquid petroleum products, both transparent and
opaque, by measuring the time for a volume of liquid to flow under
gravity through a calibrated glass capillary viscometer.

Main characteristics
Bath can be operated from ambient +5°C up to +230°C (41..446°F).
With the use of the built-in cooling coil and an external Tamson cooling
circulator, span lies 5°C above the temperature of the cold flow of the
external cooling circulator. The TV2000, TV4000(DC), and TV7000DC
have a bath volume of 20, 40, and 70 litres respectively. The TV2000
has a cover with three openings and lids. The TV4000(DC) and
TV7000DC have a cover with seven openings and lids. These openings
accommodate glass capillary viscometers in viscometer holders. The
TV4000DC has the same specifications as the TV4000, but with a
unique stability of 0.002°C at 40°C. The TV7000DC is used to calibrate
long thermometers, to calibrate master viscometers or for oxidation
stability tests. A bath overflow outlet protects from expanding bath oil
when the bath filling is expanding at higher temperatures. A bath drain
is standard included. The bath is fitted with a double window of which TV4000
the front pane is detachable for cleaning purposes.

TV2000 TV4000 TV4000DC TV7000DC

Range …230°C/…446°F
Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304, brass bearings
Setting 0.01°C
Offset ± 0.01°C (optional 0.005°C)
0.002°C @40°C
Stability ± 0.01°C 0.01°C 0.005°C @100°C 0.01°C
0.010°C @150°C
Heating 2800 W
Heaters 2
Bath volume 20 L 40 L 40 L 70 L
Number of Ø 51 mm lids 3 (4 optional) 4 + 3 (4 + 4 optional)
Window 140 x 285 mm 270 x 285 mm 270 x 585 mm
Opening bath (without cover) 130 x 165 mm 260 x 240 mm 260 x 240 mm
Depth bath 300 mm 300 mm 630 mm

 TV2000 P/N 00T0782 (230V/50-60Hz) • P/N 00T0784 (115V/60Hz)


 TV4000 P/N 00T0772 (230V/50-60Hz) • P/N 00T0774 (115V/60Hz)
 TV4000DC P/N 00T0802 (230V/50-60Hz) • P/N 00T0804 (115V/60Hz)
 TV7000DC P/N 00T0796 (230V/50-60Hz) • P/N 00T0798 (115V/60Hz)

TV2000 TV4000 TV7000


95
ASTM Equipment

1.7 ASTM D445 High Temperature Viscosity Baths - TV2500


and TV3500
ASTM D445 - ASTM D2162 - ISO/EN 3104 - IP 71 - DIN 51366 - ASTM D1298

Main Characteristics
The TV2500 or TV3500 bath can be operated from ambient +5°C up
to +120°C (41..248°F). With the use of the built-in cooling coil and an
external Tamson cooling circulator, span lies 5°C above the temperature
of the cold flow of the external cooling circulator. The TV2500 has a
bath volume of 25 liters. The TV3500 has a bath volume of 35 liters.
The TV2500 offers two rows of two openings with lids. The TV3500 can
be equipped with a cover with nine openings and lids. These openings
will accommodate glass capillary viscometers in viscometer holders. A
bath overflow outlet protects from expanding bath fluid when the bath
filling is too high. A bath drain is standard included.

TV2500 TV3500
Range …120°C/…248°F
Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304
Setting 0.01°C
Stability ± 0.02°C
Heating 1400 W TV3500

Heaters 1
Bath volume 25 L 35 L
Number of lids 4 9
Opening lid Ø51 mm
Window 230 x 300 mm
Opening bath (without
185 x 155 mm 400 x 120 mm
cover)
Depth bath 300 mm 320 mm

 TV2500 P/N 19T1003 (230V/50-60Hz) • P/N 19T1004 (115V/60Hz)


 TV3500 P/N 19T1005 (230V/50-60Hz) • P/N 19T1006 (115V/60Hz)

 For ASTM D1298 (density by hydrometer) you can use


the TV2500 in combination with a special cover with three
holes (P/N 03T2119) and a TLC10-3 cooling circulator.
Please also see section ‘ASTM D1298 – Density by
Hydrometer’.

TV2500 for ASTM


D1298

TV2500 TV3500
96
ASTM Equipment

1.8 ASTM D445 High Temperature Viscosity Baths - TV12


ASTM D445 - ISO/EN 3104 - IP 71 - DIN 51366

Main Characteristics
The TV12 is a unique viscosity bath with a small footprint. The TV12
can be operated from ambient +5°C up to +120°C (41..248°F). With
the use of the built-in cooling coil and an external Tamson cooling
circulator, span lies 5°C above the temperature of the cold flow of
external cooling circulator. The TV12 has a bath volume of only 12
liters and a unique stability of 0.005°C. The TV12 offers one row of four
openings with lids. These openings will accommodate glass capillary
viscometers in viscometer holders. A permanent LED light is located
in the top plate to supply clear light and guarantees optimal visibility
inside the bath. A bath overflow outlet protects from expanding bath
fluid when the bath filling is too high. A bath drain is standard included.
The bath is fitted with a double window of which the front pane is
detachable for cleaning purposes.

Range …120°C/…248°F
Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304
Setting 0.01°C TV12
Offset ± 0.01°C
Stability ± 0.005°C
Heating 1280 W
Heaters 2
Bath volume 12..15 L
Number of lids 4
Opening lid Ø 51 mm
Window 140 x 285 mm
Opening bath 248 x 73 mm
Depth bath 300 mm
 TV12 P/N 00T0400 (230V/50-60Hz) • P/N 00T0405 (115V/60Hz)

 For spare parts, options and accessories, please see our TV12LT.

PowerPoint with
more details at
www.tamson.com

TV12
97
C
A aS lTi M
b r aEtqi u
o inp m
B aetnhts & M e a s u r e m e n t

1.10 ASTM D445 Low Temperature Viscosity Bath: TV12LT


ASTM D445 - ISO/EN 3104 - IP 71 - DIN 51366 - IP EM PJ - IEC 61868

This test method specifies a procedure for the determination of the


kinematic viscosity of liquid petroleum products, both transparent and
opaque, by measuring the time for a volume of liquid to flow under
gravity through a calibrated glass capillary viscometer.

Main Characteristics
The TV12LT with integrated cooling is a unique viscosity bath
with a small footprint. It can be operated from -42°C up to +20°C
(-40..68°F). Ranges from -42°C up to +80°C and -50°C up to +20°C are
available. The TV12LT has a small bath volume of 15 liters and a unique
temperature stability. The TV12LT offers one row of four openings with
lids. These openings will accommodate glass capillary viscometers in
viscometer holders. A permanent LED light is located in the top plate
to supply clear light and guarantees optimal visibility inside the bath. A
bath overflow outlet protects from expanding bath fluid when the bath
filling is too high. A bath drain is standard included. The bath is fitted
with two windows of which the front panel is detachable for cleaning
purposes. TV12LT is specially designed for kinematic viscosity testing
of aviation fuels at sub zero temperatures. The TV12LT also is suitable
to calibrate thermometers or other sensors. TV12LT

Range -42..+20°C*/-43.6..+68°F
Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304
Setting 0.01°C
Offset ± 0.01°C (optional 0.005°C)
Stability ± 0.01°C
Heating 1200 W
Heaters 2
Bath volume 15....18 L
Number of lids 4
Window 255 x 230 mm
Opening lid Ø 51 mm
Opening bath 248 x 112 mm

 TV12LT P/N 00T0410 (230V/50Hz) • P/N 00T0420 (230V/60Hz) • P/N 00T0415 (115V/60Hz)

* Range from -42°C up to +80°C


 TV12LT P/N 00T0425 (230V/50Hz) • P/N 00T0435 (230V/60Hz) • P/N 00T0430 (115V/60Hz)

* Range from -50°C up to +20°C


 TV12LT P/N 00T0470 (230V/50Hz) • P/N 00T0480 (230V/60Hz) • P/N 00T0475 (115V/60Hz)

TV12LT
98
C
A aS lTi M
b r aEtqi o
unipm
B aetnhts & M e a s u r e m e n t

1.11 ASTM D445 Low Temperature Viscosity Baths 40&70 litres


ASTM D445 - ASTM D2162 - ISO/EN 3104 - IP 71 - DIN 51366

Main Characteristics
The TV4000LT and TV7000LT are unique visibility baths with wide
dimensions. Working temperature ranges from -40°C up to +100°C.
The integrated cooling system is able to lower the temperature of
the bath content down to -40°C within approximately two hours. The
cooling power can be controlled over several stages, so that the heat
removal capacity is trimmed to the desired set point temperature,
saving up to 50% of energy. Due to its wide temperature range, the
bath can be used for multiple purposes varying from low temperature
viscosity measurement to calibration of sensors and thermometers.
The large windows show the bath contents clearly. The windows are
heated preventing built up of condensate. The maximum temperature
lies around +100°C. Overall accuracy is better than ± 0.02°C over the
whole temperature range. Ambient temperature drifts <0.003°C/°C.
The homogeneity is better than 0.02°C. Drift, accuracy and
homogeneousness were measured with methanol as bath medium and
are true min-max values found over eight hours. The TV7000LT can be
used to calibrate long thermometers or to calibrate hydrometers.

TV4000LT

TV4000LT TV7000LT

Range -40..100°C/-40..212°F
Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304
Setting 0.01°C
Offset ± 0.01°C
Stability ± 0.02°C
Heating 2000 W
Heaters 2
Bath volume 40 L 70 L
Number of lids 4 + 3 (4 + 4 optional)
Window 270 * 285 mm 270 * 585 mm
Opening lid Ø 51 mm
Opening bath 260 x 240 mm
Depth bath 300 mm 600 mm
 TV4000LT P/N 00T0460 (230V/50Hz)
 TV7000LT P/N 00T0450 (230V/50Hz)

TV4000LT TV70000LT
99
ASTM Equipment

1.12 ASTM D445 Ultra Low Temperature Viscosity Bath


ASTM D445 - ISO/EN 3104 - IP 71 - DIN 51366 - ASTM D2532

Main Characteristics
The TLV25 system contains a 25 litres Dewar flask. The fluid in this
flask must permanently be cooled by a separate immersion cooler
when using at temperatures below ambient. The temperature set point
is maintained via a microprocessor controlled heating element. When
using the TCC-IC, a minimum working temperatures of -80°C can be
reached. All presented data is measured by using a TLV25 filled with
methanol and a TCC-IC immersion cooler. At the minimum temperature,
the provided heat removal is still enough to maintain stable tem­pera­
ture control, even when glass capillary viscometers are placed in the
bath. The systems accuracy is better than the requirements of ASTM
D445 and ISO 3104. The bath is illuminated by a fluor­escent light built-
in behind the Dewar flask. The top lid has a turn table construction
with three openings and three lids. By turning the lid, the immersed
viscometer can be positioned in front of the window. This window is
heated to keep clear sight at sub zero temperatures. The set point can
be set in steps of 0.1°C from -90°C up to +60°C (-130..140°F). Overall
accuracy is better than ± 0.04K. Set-up of TLV25 and TCC-IC can also
be used for ASTM D2532.

Range -80..60°C/-130…140°F TLV25


Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304
Setting 0.1°C
Offset ± 0.1°C
Stability ± 0.04°C
Heating 1500 W
Heaters 1
Bath volume 25 L
Number of lids 3
Opening lid Ø 51 mm
Opening bath Ø 161 mm
Depth bath 400 mm

 TLV25 P/N 00T0650 (230V/50-60Hz) • P/N 00T0780 (115V/60Hz)

 Please also see our ‘TCC-IC’ where you will find the necessary TCC-IC immersion cooler for the TLV25.

Necessary options TLV25


P/N 00T0300 TCC-IC to reach -80°C (230V/50HZ) •
P/N 00T0301 TCC-IC to reach -80°C (230V/60HZ) •

Accessories TLV25
P/N 10T6066 Viscohanger •
P/N 10T6090 8 Channel stop-watch •
P/N 00T0239 ASTM thermometer holder •
P/N 14T0303 Adapter to insert an E20 thermometer in the bath cover •

 For Viscosity Accessories, see section ‘Viscosity Accessories’.

TLV25
100
ASTM Equipment

1.16 ASTM D1298 Density by Hydrometer


ASTM D287 - ASTM D1298 - ASTM D6822 - IP 160 - ASTM E126 - API 2547 - ISO 3675

This specific method covers following laboratory determinations by


hydrometer method:
-Density,
-Relative density (specific gravity),
-American Petroleum Institute (gravity of crude petroleum
and liquid petroleum products).

Main Characteristics
A Tamson TLC15-5 refrigerator circulates a (cold) flow through the
double walls of the cylinders. The TLC15-5 maintains the bath
temperature within ± 0.25°C of the test temperature required by the
method. The PTFE valves ensure fast drainage and cleaning of the
cylinders. The easy detachable spill tray prevents that portions of the
sample are spilled on the workbench. The rack offers full view on the
hydrometer. Our special ASTM D1298 support eliminates the problem
of mounting brackets or portions of a thermostatic bath blocking the
view. The Tamson D1298 apparatus is a complete set-up existing of
three Jacketed glass cylinders, rack and TLC15-5 circulator. Our system
offers an interesting price advantage compared to other solutions
based on a combination of a thermostatic bath with an external chiller.
The ASTM D1298 set-up can also be used for ASTM D287 or ASTM ASTM D1298
D6822 test methods. Apparatus

Range 5°C....120°C
Setting 0.01°C
Stability ± Better than 0.1°C
TLC15-5 Rack
Bath volume 5L
Opening bath 85 x 150 mm (effective use)
Depth bath 150 mm
Length 420 mm (460 mm incl drain) 360 mm (front to back)
Width 265 320 (left to right)
Height 565 800 mm
Weight 30 kg 9 kg (glass parts included)
Power 1100 W (1 heater)
Ambient
15..28°C
temperature
CE Products conform to CE regulation

 Set-up ASTM D1298 P/N 00T2000 (230V/50Hz) • P/N 00T2001 (230V/60Hz) • P/N 00T2002 (115V/60Hz)
 An additional rack is available to get an apparatus for six positions with one circulator • P/N 00T1260
 Alternative for this setup is a TV2500 in combination with the TLC10-3. For further information, please see section 1.7.

Accessories ASTM D1298


P/N 25T0928BW Thermometer with blue filling similar to ASTM 34C •
P/N 25T2153 Thermometer holder D1298 •

Density by Hydrometer
101
Low Temperature Baths

3.2 Tamson Cooling Circulators - TLC Line TLC10-3


The TLC10-3 has been developed to be used as general purpose
circulator and it can be used in various applications where cooling is
needed. Experiments, refractometers, rotary-evaporators, distillation
units, water baths, viscosity baths, flash point testers are applications
for which the TLC10-3 can be used.

Main Characteristics
The apparatus holds a small bath containing 3.5 litres and the filling
opening measures 40 x 120 mm. Due to its special design the system
is extremely quiet, compact and saves costly laboratory space.
Readout is in degrees °C or in °F. The bath comes standard with
RS232 communication and when using the free software tool Tamcom,
it provides temperature logging or a predefined temperature set point
curve. A pump provides circulation in the bath or via an external
circuit. The pump offers seven litres per minute with a maximum
counter pressure of 300 mBar.

Range -10..60°C* / 14..140°F


Reading Standard °C, °F on request TLC10-3
Used materials inside bath Stainless steel 304, brass bearings
Setting 0.01°C
Stability ± Better than 0.05°C
Heating 1100 W
Bath volume 3.5 L
Opening bath 40 x 120 mm
Depth bath 150 mm
Pump (mBar) Max pressure 300
Pump (L/min) Max flow 7
Dimension LxWxH 360 x 228 x 552 mm
Weight 18 kg
Heat removal See section 3.1

* On request the TLC10-3 can have a working range from -10..120°C

 TLC10-3 P/N 00T0050 (230V/50Hz) • P/N 00T0051 (230V/60Hz) • P/N 00T0052 (115V/60Hz)

 For options and spare parts, please see section `TC15-5 • TC30-5’

Manuals see
www.tamson.com

TLC10-3
102
Low Temperature Baths

3.3 Tamson Cooling Circulators – TLC Line TLC15-5, TLC30-5


The TLC15-5 and TLC30-5 have been developed to be used as general
purpose circulator and they can be used in various applications where
cooling is needed. Examples are experiments, refractometers, rotary-
evaporators, distillation units, water baths, viscosity baths, flash point
testers, foaming tests and ductility tests.

Main Characteristics
The apparatus holds a small bath containing five litres and the filling
opening measures 85 x 150 mm. Due to its special design the system is
extremely quiet, compact and saves costly laboratory space. Readout
is in degrees °C or in °F. A pump provides circulation in the bath or
to an external circuit. The pump offers seven litres per minute with a
maximum counter pressure of 300 mBar.

TLC15-5 TLC30-5
TLC15-5

Range -15..60°C*/-5..140°F -30..60°C**/-5..140°F


Reading Standard °C, °F on request
Used materials inside bath Stainless steel 304, brass bearings
Setting 0.01°C
Stability ± Better than 0.02°C
Heating 1100 W
Bath volume 5L
Opening bath 85 x 150 mm
Depth bath 150 mm
Pump (mBar) Max pressure 300
Pump (L/min) Max flow 7
Dimension LxWxH 420 X 265 X 565 mm
Weight 30 kg
Heat removal See section 3.1
* On request the TLC15-5 can have a working range from -15..120°C
** On request the TLC30-5 can have a working range from -30..120°C

 TLC15-5 P/N 00T0565 (230V/50Hz) • P/N 00T0567 (230V/60Hz) • P/N 00T0570 (115V/60Hz)
 TLC30-5 P/N 00T0555 (230V/50Hz) • P/N 00T0562 (230V/60Hz) • P/N 00T0560 (115V/60Hz)

T a mso n I nst r uments


TLC15-5 TLC30-5 Tel. +31 10 522 43 73
Email sales@tamson.com
Website www.tamson.com
103
Low Temperature Baths

3.4 Tamson Cooling Circulators – TLC Line TLC40, TLC80, TLC90


The TLC40-14 cooling circulator can be used for multiple heat removal
purposes such as replacement of tap water, fast cool down, flash point,
distillation, density or cloud and pour point testing. Also, this unit can
be used to cool down water baths and viscosity baths which have a
large bath volume. TLC80-14 and TLC90-14 cooling circulators can
be used for general low temperature use and shows excellent heat
removal performance in combination with cloud and pour point tests or
cold filter plugging point tests.

Main Characteristics
The apparatus holds a bath containing 14 litres and the filling opening
measures 240 x 170 mm. Due to its special design the system is
extremely quiet, compact and saves costly laboratory space. The
minimum temperature which can be reached with the TLC40-14 is
-40°C. The minimum temperature which can be reached with the
TLC80-14 is -80°C. At -80°C the heat removal capacity is still 300
Watts! The minimum temperature which can be reached with the
TLC90-14 is -90°C. Readout is in degrees °C or in °F. A pump provides
circulation in the bath or pumps the flow to an external circuit. The
pump offers seven litres per minute with a maximum counter pressure TLC40-14
of 300 mBar. Optional is a performance pump which offers 16 litres
per minute.

TLC40-14 TLC80-14 TLC90-14

Range -40..+20°C*/-40..+68 °F -80..+20°C/-112..+68°F -90..+20°C/-130..+68°F


Reading Standard °C, °F on request
Used materials
Stainless steel 304, brass bearings
inside bath
Setting 0.1°C
Stability ± Better than 0.05°C
Heating 1550 W
Bath volume 14 .. 15 L
Opening bath 240 x 170 mm
Depth bath 150 mm
Pump (mBar) Max pressure 300 (optional 1 Bar)
Pump (L/min) Max flow 7 (optional 16) Max flow 10 (optional 16) Max flow 10 (optional 16)
Dimension LxWxH 810 x 460 x 770 mm
Weight 65 kg 80 kg 80 kg
Heat removal See section 3.1
* On request the TLC40-14 can have an optional working range from -45..+20°C.

 TLC40-14 P/N 00T0520 (230V/50Hz) • P/N 00T0522 (230V/60Hz) • P/N 00T0525 (115V/60Hz)
 TLC80-14 P/N 00T0530 (230V/50Hz) • P/N 00T0532 (230V/60Hz) • P/N on request (115V/60Hz)
 TLC90-14 P/N 00T0580 (230V/50Hz) • P/N 00T0585 (230V/60Hz) • P/N on request (115V/60Hz)

 For options and spare parts, please see section TLC80-14DP

T a mso n I nst r uments


74
TLC40TLC80
TLC40 TLC80TLC90
TLC90 Tel. +31 10 522 43 73
Email sales@tamson.com
Website www.tamson.com
104
STIRRING-HEATING BLOCKS WITH EXTERNAL CONTROL UNIT
For viscosity measurements, R&D, combinatorial chemistry, biological applications, testing the solution behavior
of synthetic materials, quality control in food, chemical and waste water technology, simple and easy system integration.

Fast heating-up
STIRRING DRYBATH 15-100 / STIRRING DRYBATH 15-250 Energy-saving
Stirring-heating block combination with 15 bore holes, suitable for 100 ml lab-flasks and for 150 ml beaker Insulated frame
glasses tall form (ø 57.5 mm, depth 48 mm, distance 65 mm) or with 15 bore holes suitable for 250 ml lab-flasks Durable quality
and for 400 ml beaker glasses tall form (ø 71.5 mm, depth 78 mm, distance 84 mm). Developed and made in Germany

Stirring - maintenance-free and wear-free by inductive


2mag-Magnetic-Drive concept, extremely wide speed range
of 100 to 2,000 rpm, 100% synchronized speed, jerk-free
stirring even at low speeds, 10-step power setting of the
stirrer, high stirring power for viscous liquids and reduced
stirring power for long-term operation without any heating
effects caused by the stirrer, clear digital display for settings
of stirrer speed, stirrer power and temperature of the
integrated heater, SoftStart procedure for reliable catching/
centering and safe acceleration of the stirring bar.

Stirring - maintenance-free and wear-free by inductive Heating - Heating plate made of resistant aluminium alloy,
2mag-Magnetic-Drive concept, extremely wide speed range PTFE-coated for increased chemical resistance and easier
of 100 to 2,000 rpm, 100% synchronized speed, jerk-free cleaning, fast heating-up by powerful integrated electrical
stirring even at low speeds, 10-step power setting of the heater, homogenous temperature distribution, maximum
stirrer, high stirring power for viscous liquids and reduced heating temperature +200 °C, PowerDetect for automatic
stirring power for long-term operation without any heating compensation of physically determined loss in stirring
effects caused by the stirrer, clear digital display for settings power which can come up to 40% at high temperatures,
of stirrer speed, stirrer power and temperature of the independent over-temperature protection inside the plate.
integrated heater, SoftStart procedure for reliable catching/
centering and safe acceleration of the stirring bar.

Type DRYBATH 15-100 DRYBATH 15-250


Stirring points 15
Stirring point distance 65 mm 84 mm
Heating
Bore holes Ø/T - Heating plate Ø made
57.5 /of
48resistant
mm aluminium alloy,
Ø 71.5 / 78 mm Inclusive separate control unit for operation of magnetic
PTFE-coated for
Stirring volume/stirring pointincreased chemical
1 - 150 ml resistance and easier
1 - 400 ml stirrer and integrated heater, for long-term protection
Stirring cleaning,
power (max.)fast heating-up by powerful integrated
40 W electrical of the electronics against aggressive liquids and gases
heater, homogenous temperature4distribution,
Power setting - 40 W (10-step)maximum as well as thermal influences, stainless steel housing.
heating
Temperature temperature +200 °C, PowerDetect
(max.) +200 °Cfor automatic
compensation of physically determined
Speed range loss rpm
100 - 2,000 in stirring Accessories
Operatingpower
voltagewhich
(max.) can come up to 40% at 230high temperatures,
V / 1,000 W
Dimensions (WxDxH)
independent 280 x 430 x 120
over-temperature mm
protection 335 x 515
inside thex plate.
150 mm External probe sensor Pt100
Protection category IP20 Stand for heatMIXcontrol
Weight (gross) approx. 23 kg approx. 31 kg Reducing adapters - Inserts for reducing the bore
Order no. 63100 63250 hole diameter

Individual units and dimensions on request.


3 years warranty on material and manufacturing.
We reserve the right to make technical alterations. We do not assume liability for errors in printing or typeing.

105
STIRRING-HEATING
STIRRING-HEATINGBLOCKS
BLOCKSWITH
WITHEXTERNAL
EXTERNALCONTROL
CONTROLUNIT
UNIT
For viscosity measurements, R&D, combinatorial chemistry, biological applications, testing the solution behavior
of synthetic materials, quality control in food, chemical and waste water technology, simple and easy system integration.

Fast heating-up
Energy-saving
STIRRING DRYBATH 8-250 ERL Insulated frame
Stirring-heating block combination with 8 stirring positions (ø 87.2 mm, Durable quality
depth 78 mm, distance 100/140 mm), suitable for 250 ml Erlenmeyer flasks. Developed and made in Germany

Stirring - Maintenance-free and wear-free by inductive Heating - Massive heating block made of resistant aluminium
2mag-Magnetic Drive concept, extremely wide speed range alloy, PTFE-coated for increased chemical resistance and
of 100 to 2,000 rpm, 100% synchronized speed, jerk-free easier cleaning, contact adapters for an optimal heat transfer,
stirring even at low speeds, 10-step power setting of the lowest possible temperature gradient inside the stirring
stirrer, high stirring power for viscous liquids and reduced vessels, integrated electrical heater, maximum heating
stirring power for long-term operation without any heating temperature +200 °C.
effects caused by the stirrer, clear digital display for settings PowerDetect for automatic compensation of physically
of stirrer speed, stirrer power and temperature of the determined loss in stirring power which can come up to
integrated heater, SoftStart procedure for reliable 40% at high temperatures, specially made insulated
catching/centering and safe acceleration of the stirring bar. stainless steel frame for energy- and time saving heating
process, additionally for a homogenous temperature
distribution and protection against accidental contact,
independent over-temperature protection inside the plate.

Type DRYBATH 8-250 ERL


Stirring points 8
Stirring point distance 100 / 140 mm
Bore holes Ø/T Ø 87.2 / 78 mm
Stirring volume/stirring point 1 - 250 ml
Stirring power (max.) 40 W
Power setting 4 - 40 W (10-step)
Temperature (max.) +200 °C
Speed range 100 - 2,000 rpm
Operating voltage (max.) 230 V / 1,000 W
Dimensions (WxDxH) 335 x 515 x 150 mm
Protection category IP20
Weight (gross) approx. 38 kg
Order no. 63825

106