PulmoTrend
for
VentMeter
Copyright 2008
Magnamed Tecnologia Médica S.A.
Magnamed Tecnologia Médica S.A.
Rua São Paulino, 221 – Vila Mariana
CEP: 04019-040 – São Paulo - SP – Brasil
Tel: +55 (11) 5081-4115 Fax: +55 (11) 5081-4115
E-mail: magnamed@magnamed.com.br
Website: www.magnamed.com.br
2 Information
The PumoTrend software is intended to be used with the VentMeter and the free version,
that accompanies the VentMeter Product is a basic version with limited resources.
The PulmoTrend is basic software that can be used freely with the VentMeter – Ventilator
Analyzer. It can be downloaded from Magnamed’s website and installed in a Microsoft
Windows XP Operating System or newer.
With this basic version it is possible to connect the VentMeter through a serial RS-232
connection (or through a USB – to SERIAL converter cable with appropriate driver) and
see all the parameters once in the screen.
All the eighteen numerical parameters can be showed simultaneously and trend graphics
for Expired Volume, Maximum Pressure and PEEP (it is fixed this way in this version of
software). The real time graphics presented are Pressure x Time, Flow x Time and Volume
x Time curves, selected by the user.
3 Index
2 INFORMATION .......................................................................................................................................4
3 INDEX ......................................................................................................................................................5
4 DESCRIPTION ........................................................................................................................................6
4.1 INTENDED USE ......................................................................................................................................6
4.2 INSTALLATION........................................................................................................................................6
4.3 VENTMETER CONNECTION.....................................................................................................................6
4.4 STARTING PULMOTREND .......................................................................................................................6
5 OPERATION ...........................................................................................................................................7
5.1 IDENTIFICATION AND CONNECTION .........................................................................................................7
5.2 SCALE ADJUSTMENT......................................................................................................................... 111
5.3 TEST SETUP..................................................................................................................................... 133
4 Description
4.2 Installation
2. Run Setup
3. It is ready to be used
3. If you have a USB to SERIAL converter, than you should assure that the drivers are installed;
5 Operation
Monitored Parameters
Target Ventilator and lung
Print Test Simulator Adjustments
Load Results
Save Results
TRENDTREND Graphics
Graphics
Start Current Test
a. Monitored Parameters – Shows all eighteen monitored parameters and is refreshed each respiration
cycle;
b. Target Lung Ventilator and Lung Simulator Adjustments – In this panel it is possible to program all
the tests that should be done to certify a lung ventilator.
c. Trend Graphics – This panel shows a fixed trend graphics parameters (Expired Volume, Maximum
Pressure and PEEP)
d. Real Time Graphics – This panel shows the Pressure x Time, Flow x Time and Volume x Time curves
that can be selected by the user.
b. Fill in the fields, and press OK, when filalize the test and print it, it will be recorded in the report.
b. Start communication by pressing CONNECT (Communication – Connect) – After you press the
communication indicator bar will fill, the real time curves will run on the screen and the parameters
will start to be shown.
b. Clicking over the left panel of the TREND GRAPHICS it will pop-up a trend graphics scale setup
window in which it is possible to select the minimum and maximum scale of each parameter and also
the time scale.
d. By clicking the TOLERANCE button the tolerances window pops-up and all parameters different from
zero will be registered and can be printed after the tests
e. Note that in the above figure only the Tidal Volume, PEEP, Frequency and Inspiratory Time can be
printed in a test result report.
f. Clicking the TEST button after ADD on the right panel TEST SETTINGS you can add a second
parameter testing. Note that the test number switches 2 to 2, meaning second test.
g. Clicking the TEST button then INSERT on the right panel, you can also type tests before and after the
tests have already been developed.
i. You can also insert an information to be confirmed in the test result, the test results, the text is free
and can be written other information they need to present on the report.
j. See below the report generated and printed in pdf of above test.
l. To start registering the parameters press the start test button. You will see a progress bar in green.
Finishing the test a End of Test Window will pop-up. Click OK to finish.
n. Select the next test by clicking the Next Test button and again press Start Test until all the tests are
done.
o. To save the test results click the SAVE TEST RESULT ICON at the upper left corner of the screen.
p. Open the test results. Click the Load Tests Results ICON at the upper left corner of the screen and
select the test result name to see.
q. Printing the Report – Click the printer icon and select the printer output.