You are on page 1of 3

© 2008 USPC Official 12/1/07 - 4/30/08 General Chapters: <643> TOTAL ORGANIC C...

Page 1 of 3

643 TOTAL ORGANIC CARBON


Total organic carbon (TOC) is an indirect measure of organic molecules present in
pharmaceutical waters measured as carbon. Organic molecules are introduced into the water
from the source water, from purification and distribution system materials, and from biofilm
growing in the system. TOC can also be used as a process control attribute to monitor the
performance of unit operations comprising the purification and distribution system.

A number of acceptable methods exist for analyzing TOC. This chapter does not limit or
prevent alternative technologies from being used, but provides guidance on how to qualify
these analytical technologies for use as well as guidance on how to interpret instrument
results for use as a limit test. The Standard Solution is a theoretically easy-to-oxidize solution
that gives an instrument response at the attribute limit. The analytical technology is qualified
by challenging the capability of the instrument using a theoretically difficult to oxidize solution
in the system suitability portion of the method.

Analytical technologies utilized to measure TOC share the objective of completely oxidizing
the organic molecules in an aliquot of sample water to carbon dioxide (CO2), measuring the
resultant CO2 levels, and expressing this response as carbon concentration. All technologies
must discriminate between the inorganic carbon, which may be present in the water from
sources such as dissolved CO2 and bicarbonate, and the CO2 generated from the oxidation
of organic molecules in the sample.

Two general approaches are used to measure TOC. One approach determines TOC by
subtracting the measured inorganic carbon (IC) from the measured total carbon (TC), which
is the sum of organic carbon and inorganic carbon:

TOC = TC – IC.

The other approach first purges the IC from the sample before any carbon measurement is
performed. However, this IC purging step also purges some of the organic molecules, which
can be retrapped, oxidized to CO2, and quantitated as purgeable organic carbon (POC). The
remaining organic matter in the sample is also oxidized to CO2 and quantitated as
nonpurgeable organic carbon (NPOC). In this approach, TOC is the sum of POC and NPOC:

TOC = POC + NPOC.

In pharmaceutical waters, the amount of POC is negligible and can be discounted. Therefore,
for the purpose of this methodology, NPOC is equivalent to TOC.

Apparatus Requirements—This test method is performed either as an on-line test or as an

http://www.uspnf.com/uspnf/pub/getPFDocument?usp=30&nf25&supp=2&collection=/db... 2/11/2008
© 2008 USPC Official 12/1/07 - 4/30/08 General Chapters: <643> TOTAL ORGANIC C... Page 2 of 3

off-line laboratory test using a calibrated instrument. The suitability of the apparatus must be
periodically demonstrated as described below. In addition, it must have a manufacturer's
specified limit of detection of 0.05 mg of carbon per L (0.05 ppm of carbon) or lower.

USP Reference Standards 11 —USP 1,4-Benzoquinone RS. USP Sucrose RS.

Reagent Water—Use water having a TOC level of not more than 0.10 mg per L. [NOTE—A
conductivity requirement may be necessary to ensure method reliability.]

Glassware Preparation—Organic contamination of glassware results in higher TOC values.


Therefore, use glassware and sample containers that have been scrupulously cleaned of
organic residues. Any method that is effective in removing organic matter can be used (see
Cleaning Glass Apparatus 1051 ). Use Reagent Water for the final rinse.

Standard Solution—Unless otherwise directed in the individual monograph, dissolve in the


Reagent Water an accurately weighed quantity of USP Sucrose RS, to obtain a solution
having a concentration of about 1.2 mg of sucrose per L (0.50 mg of carbon per liter).

Test Solution—[NOTE—Use extreme caution when obtaining samples for TOC analysis.
Water samples can be easily contaminated during the process of sampling and transportation
to a testing facility.] Collect the Test Solution in a tight container with minimal head space,
and test in a timely manner to minimize the impact of organic contamination from the closure
and container.

System Suitability Solution—Dissolve in Reagent Water an accurately weighed quantity of


USP 1,4-Benzoquinone RS to obtain a solution having a concentration of 0.75 mg per L (0.50
mg of carbon per liter).

Reagent Water Control—Use a suitable quantity of Reagent Water obtained at the same
time as that used in the preparation of the Standard Solution and the System Suitability
Solution.

Other Control Solutions—Prepare appropriate reagent blank solutions or other specified


solutions needed for establishing the apparatus baseline or for calibration adjustments

http://www.uspnf.com/uspnf/pub/getPFDocument?usp=30&nf25&supp=2&collection=/db... 2/11/2008
© 2008 USPC Official 12/1/07 - 4/30/08 General Chapters: <643> TOTAL ORGANIC C... Page 3 of 3

following the manufacturer's instructions, and run the appropriate blanks to zero the
instrument.

System Suitability—Test the Reagent Water Control in the apparatus, and record the
response, rw. Repeat the test using the Standard Solution, and record the response, rS.
Calculate the corrected Standard Solution response, which is also the limit response, by
subtracting the Reagent Water Control response from the response of the Standard Solution.
The theoretical limit of 0.50 mg of carbon per L is equal to the corrected Standard Solution
response, rS–rw. Test the System Suitability Solution in the apparatus, and record the
response, rss. Calculate the corrected System Suitability Solution response by subtracting the
Reagent Water Control response from the response of the System Suitability Solution, rss–rw.
Calculate the response efficiency for the System Suitability Solution by the formula:
100[(rss–rw) / (rS–rw)].
The system is suitable if the response efficiency is not less than 85% and not more than
115% of the theoretical response.

Procedure—Perform the test on the Test Solution, and record the response, rU. The Test
Solution meets the requirements if rU is not more than the limit response, rS– rw. This method
also can be performed alternatively using on-line instrumentation that has been appropriately
calibrated, standardized, and has demonstrated acceptable system suitability. The
acceptability of such on-line instrumentation for quality attribute testing is dependent on its
location(s) in the water system. These instrument location(s) and responses must reflect the
quality of the water used.

Auxiliary Information—Staff Liaison : Gary E. Ritchie, M.Sc., Scientific Fellow


Expert Committee : (PW05) Pharmaceutical Waters 05
USP30–NF25 Page 257
Pharmacopeial Forum : Volume No. 30(5) Page 1700
Phone Number : 1-301-816-8353

http://www.uspnf.com/uspnf/pub/getPFDocument?usp=30&nf25&supp=2&collection=/db... 2/11/2008

You might also like