Professional Documents
Culture Documents
Vol. 13,
15, No.
No. 1
3
FOCUS
Digital Radiography in Industry: Digital
Detector Arrays in Radiographic Testing
by Brad Kraai
SRb
(effective IQI hole ø – d
pixel size)
Figure 2. A 3 × 3 pixel matrix in image quality indicator (IQI) hole—calculation for minimum Figure 3. Duplex wire gage.
magnification.
be fully eliminated, and is composed of the law of Poisson exists, where noise will scale of film—that is, 1.50 to 4.00 H&D
signal and system noise. Common noise increase by the square root of the variable (Hurter and Driffield) or ~250 usable
reduction tactics include frame averaging, signal increase (exposure). That being said, radiographic optical densities. The usable
DDA calibration, X-ray beam filtration, higher exposure or signal provides higher range of the DDA will always be less
X-ray beam collimation, and exposure SNR, albeit with an increase in exposure than the full scale or bit depth, but is still
optimization. The averaging of frames time or image acquisition, and effectually significant. This long scale enables the
basically provides better statistics for a higher dose to the DDA. High SNR very low contrast sensitivity mentioned
the raw image data, by adding the pixel results in extremely low contrast sensitivity previously—provided noise is controlled.
signal value for each individual pixel capability for DDA techniques: quite The long scale also provides adequate image
through all frames captured, dividing by often well below 0.5%, particularly for quality through a wider thickness range in
the number of frames, and assigning the thinner areas of interest within test articles. one exposure. Another advantage of DDAs
averaged value to that pixel for the raw This high SNR also contributes to high is this very high dynamic range as compared
image. Averaging more frames impacts contrast-to-noise ratio (CNR), as measured to short scale imagers, mainly film
throughput, increases dose to the DDA, within an image of a conventional IQI. As can be clearly seen, there are
and typically only presents improved Figure 4 illustrates a CNR measurement multiple variables that must be evaluated
statistics at a threshold. DDA calibrations, on a 0.05 IQI, 127 µm (0.005 in.) T, on a and correlated during DDA system
or normalizing, are a necessary function 1.27 mm (0.05 in.) thick base material. As consideration and design. Resolution and
of the DDA imaging process. Most users important technique attributes, SNR and contrast performance for any radiographic
apply calibration techniques for DDA offset CNR measurement can be accomplished imaging system are the primary objectives,
and gain per the integrator or vendor’s with specific tools within most imaging and the requirements for the inspection
recommendation. Offset is basically dark software. High CNR values are not application must be recognized and
or inherent noise within the DDA, and possible without high SNR values, and understood. The characterizations of
gain is inconsistencies in pixel responses these two metrics can be used to validate performance within any DDA system are
(amplitude or gain) during exposure. one another—provided measurements are various, but can be boiled down to system
Without normalization, any DDA would consistent and controlled. SRb, allowable or required geometric
basically be unusable for an application. magnification, Uimg, and contrast sensitivity
It is extremely important that the user through the material type and thickness
establish DDA normalization protocol and range to be inspected. Many other
evaluate its efficacy X-ray beam filtration considerations exist, more or less depending
is often employed, thereby attenuating on the target application, that are beyond
certain portions of the soft radiation the scope of this article. Realistically, at the
spectrum, which in turn reduces noise beginning, middle, and end of the day, the
and improves signal-to-noise ratio (SNR). required performance of the DDA system
Experimentation is key, and SNR can be is just this: to consistently and reliably
evaluated during system validation and detect and accurately portray substandard
technique development. Applying filtration defects for the selected application and its
also reduces dose to the DDA, which technique.
may increase its lifespan. Collimation,
or restricting the X-ray beam to only Process Controls
illuminate the input region of the DDA, Any DDA system’s performance must be
will also lower signal noise and reduce dose monitored for instability and degradation.
Figure 4. Contrast-to-noise ratio Most users apply a duplex plate phantom
to the DDA’s electronics. Besides fixed
measurement on an image quality
collimators, there are also programmable (DPP) and specialized software to enable
indicator of 29.5; this translates to
X-ray beam shuttering mechanisms that 0.8% contrast sensitivity. consistency within the measurements
can be completely closed (for X-ray system required for stability and performance.
warm-up) or opened/closed, as appropriate, The DPP (Figure 5) must be of the
for the imaging technique. Proper exposure The available DDAs are normally 14-bit same material group as the inspection
ensures optimized signal capture within (214) or 16-bit (216), 0 to 16 383 or 0 to application, with a thin and thick section
the DDA, and a translation to low contrast 65 535 possible pixel values, respectively: closely matching the thinnest and
sensitivity. Within any signal amplification, very long scale as compared to the limited thickest sections of the test article(s).
Hydrophilic emulsifier is commonly supplied as a 100% using a refractometer (Figure 1). As an NDT professional
active concentrate, which is diluted in water before use, to working for an aerospace company, I check the concentration
emulsify and remove excess of non-water-based penetrants of our hydrophilic emulsifier every week per ASTM E 1417/E
(method D). The percentage of concentration depends on 1417M-13 using a handheld refractometer (ASTM, 2013b).
the application technique. The general spray concentration I have found that there is not much information available
is up to 5% remover-to-water, and 20% remover-to-water regarding this subject, and so this article presents a brief
for immersion dip applications. Concentration measurement walkthrough on how the process works.
is performed at initial makeup, during additions, and at The use of a refractometer is suitable and practical for
intervals during use. measuring the concentration of hydrophilic emulsifier,
Most nondestructive testing (NDT) standards require a also referred to as a remover, and also for water-based
weekly check of the hydrophilic emulsifier concentration water-washable penetrant.
Method of Operation
Adjustment to zero (null): place two to three drops of distilled or
Figure 4. Through the eyepiece we can see that the scale reads
reverse osmosis water on the refractometer’s main prism (Figure 3).
zero (that is, calibrated to null).
The water must spread across the surface of the prism without any
AUTHOR
Abhishek Wakaskar: ASNT NDT Level III; Aerospace Precision, Inc.,
2851 Evans St., Hollywood, Florida 33020; (954) 923-3213; fax (954)
923-3223.
Figure 5. Through the eyepiece we can see that the scale reads 4%. REFERENCES
ASTM, ASTM D 6304, Standard Test Method for Determination of Water in
Petroleum Products, Lubricating Oils, and Additives by Coulometric Karl Fischer
Now that the refractometer calibration is complete, measure the Titration, ASTM International, West Conshohocken, Pennsylvania, 2007.
emulsifier sample in the same manner as for calibration. Clean the prism ASTM, ASTM D 95, Standard Test Method for Water in Petroleum Products
using a microfiber cloth. Add two to three drops of emulsifier onto the and Bituminous Materials by Distillation, ASTM International, West
prism, again making sure that the emulsifier spreads evenly across the Conshohocken, Pennsylvania, 2013a.
surface without any bubbles and wait for 30 s. Then look through the ASTM, ASTM E 1417/E 1417M-13, Standard Practice for Liquid Penetrant
eyepiece and read the corresponding scale on the light/dark boundary. Testing, ASTM International, West Conshohocken, Pennsylvania, 2013b.
TNT · July 2016 · 11
NDT Professional Connections | Products and Services