Professional Documents
Culture Documents
A n n a Fijalkowska
Department of Anaesthesia a n d Intensive Therapy, University School of M e d i c i n e , 2 0 - 0 9 0 L u b l i n , J a c z e w s k i e g o Street 8 , Poland
Abstract body where they can accumulate, for example, in the adipose
tissue (5). The accumulated drugs can be released from such
The use of intravenous agents to maintain anesthesia has become tissues in the postoperative period, which puts not only the pa
increasingly popular since the introduction of propofol. Its tient's life at risk but also all those people and things that de
popularity results from the fact that in 95% of patients given 2.5 pend on the psychophysical functioning of the patient. This
mg propofol per kilogram of body weight, the induction of the drug problem is especially important for short-acting anaesthetic
is rapid and smooth and is followed by rapid recovery with a low drugs frequently used in one-day surgery (6). Such anesthetic
incidence of postoperative side effects. In order to study the agents are often applied for short medical interventions after
physiological effects of an intravenous anesthetic agent, it may be
which the patient leaves the hospital immediately.
necessary to determine its concentration in blood. This paper deals
Propofol (Diprivan) (2,6-diisopropylphenol) is one of the in
with the high-performance liquid chromatographic analysis of
propofol in blood. The concentration values are calculated from
travenous anesthetic drugs used for anesthetic induction and its
the application of extraction and precipitation procedures. maintenance (7,8). It is also employed as a sedative (9). The pa
Observed differences resulting from the two analytical methods are tient loses consciousness 30-50 s after receiving the drug and
discussed and interpreted with reference to investigations of blood, remains asleep for about 4-6 min (6-8). For this reason, physi
plasma, and water solutions of propofol and the precipitation of cians readily use propofol for one-day surgery.
solid elements in blood. Despite its rapid action in the body, propofol is not metabo
lized completely in a few minutes. It is present in the blood, the
adipose tissue, the liver, and the brain for a few hours, its con
centration decreasing with time (8-10). Of course, its pres
Introduction ence influences the patient's psychophysical efficiency, which is
confirmed by the changes in the electroencephalographic
Significant progress in the field of biochemical and biomed records of the brain waves (6).
ical analysis has opened numerous possibilities for measuring Analyzing and monitoring the concentration of propofol in
the concentration of various drugs in blood (1). Presently, drug blood are important for determining when a patient can safely
monitoring helps to decrease the risk of drug-induced compli leave the hospital. In addition, in the case of the continuous in
cations, which is especially important when strong therapeutic travenous infusion of propofol, the estimation of drug concen
agents are used. tration in blood helps to find the minimal dose sufficient for an
Individual sensitivity to medicines varies because of differ individual patient to maintain sleep.
ences in metabolism and renal elimination, which are deter Two methods are described in the literature for the analysis
mined by genetic factors and the physical condition of an or of propofol in blood (11,12). According to the first procedure
ganism or an individual. Consequently, the same drug dose (11), propofol was extracted from blood using cyclohexane.
can lead to different effects in different patients. Based on phar After solvent evaporation, the remaining sample, which was
macokinetic developments and modern analysis techniques, it diluted in the mobile phase, was injected into the column. In
may be possible to optimize individual doses by measuring the the second method (12), the supernatant obtained from blood
therapeutic agent concentration in blood (2). after protein precipitation and centrifugation was directly in
Currently, anesthesiologists employ many groups of anes jected into the chromatographic column. In preliminary in
thetic agents (3). Regardless of the administration technique, all vestigations, different concentrations were found in the same
anesthetic drugs penetrate the cells of the central nervous blood sample depending on which of the two methods was ap
system (4). They are also transported to other tissues of the plied. Both calibration curves for propofol in blood, plasma, and
* Author to whom correspondence should be addressed. water samples and concentration values of propofol in blood
samples differing in hematocrit values (i.e., in the amount of trations: 0.5,1.0,2.0,4.0,7.0, and 10.0 μg/mL. By using these
solid elements in blood) were considered in order to explain this three sets of solutions and following either the extraction (11)
difference. or precipitation (12) procedure, five calibration curves were
plotted: propofol blood solutions, extraction method; propofol
water solutions, extraction method; propofol blood solutions,
Experimental precipitation method; propofol plasma solutions, precipita
tion method; and propofol water solutions, precipitation
Equipment
method. To construct these curves, the ratios of the peak
A Gilson solvent delivery system (Middleton, WI) composed heights from propofol and the internal standard were used.
of two high-pressure pumps (Models 305 and 306), a Model 805 Following the methodologies described in the literature
manometric module, and a Model 811 C dynamic mixer was (11,12), thymol was employed as an internal standard in the ex
used. The ultraviolet-visible variable wavelength detector was traction procedure, and n-dibutylphthalate was used in the
a Model 308 from MIM (Budapest, Hungary). precipitation procedure. Thymol was added to the blood,
Chromatographic separations were carried out using a 250- plasma, or water samples as a single methanol solution,
x 4-mm i.d. column packed with a laboratory-made reversed- whereas n-dibutylphthalate was added to the samples as a
phase octadecyl silane sorbent. The samples were injected into component of the precipitating solution.
the column by a Model 7125 injection valve from Rheodyne
(Cotati, CA).
Reagents
A mixture composed of 67% acetonitrile (gradient grade for
All chemicals, except for those mentioned previously, were
chromatography, LiChrosolv series, Merck; Germany) and 33%
obtained from the Polish Factory of Chemical Reagents-POCh
double-distilled water (pH 4.0) were used as the mobile phase.
(Gliwice, Poland) and were of analytical grade. The blood used
The desired pH was achieved by the addition of acetic acid (ap
for the preparation of standards and the blood samples used for
proximately 0.3-0.5 mL depending on the pH of the water).
analysis were collected in tubes containing sodium citrate and
Propofol was obtained from ICI-Pharma (Goteborg, Sweden).
stored at 4°C for no longer than 5 days. A stock solution of
Methods thymol in methanol (1 mg/mL) was diluted with methanol to
Calibration standards the appropriate concentration. Tetraethylammonium hydroxide
The samples for the calibration curves were prepared from (TEAH) (25% in ethanol) was diluted with ethanol (3:37). The
a solution containing 10 mg propofol in 100 mL acetonitrile. precipitating solution, which contained an internal standard,
The solution was appropriately diluted using water (to prepare was prepared by diluting 2 mL n-dibutylphthalate solution (1
water standards), blood (to obtain blood standards), or plasma mg in 1 mL acetonitrile) in 100 mL acetonitrile-65% per
(to prepare plasma standards) to obtain the following concen chloric acid (67:33, v/v).
Extraction procedure
The internal standard solution (thymol
dissolved in methanol) (20 pL), phosphate
buffer (1 mL of 0.1M NaH PO ), and cyclo-
2 4
378
Journal of Chromatographic Science, Vol. 33, July 1995
pending on the blood hematocrit) was subjected to the extrac The amount of bonded ODS was calculated on the basis of an
tion and concentration procedures previously described and fi equation from reference 15, and the data was determined by a
nally to chromatographic separation. Hewlett-Packard Model 185 CHN analyzer.
Controlled porosity glass was prepared from Vycor glass ac
Precipitation procedure cording to a procedure described elsewhere (16,17).
A known amount of blood was centrifuged in order to sepa
rate plasma from the solid elements of blood. The precipi
tating solution containing n-dibutylphthalate as the internal
standard (0.5 mL) was added to the plasma (0.5 mL) and mixed Results and Discussion
for 2 min with a vortex mixer. The precipitated substances
were separated by centrifugation (1200 ×g for 5 min), and the The aim of this paper was to compare the results of propofol
supernatant was injected into the chromatographic column. analysis carried out by means of the extraction and precipita
The same procedure was used for plasma and water standard tion methods (11,12). The curves obtained by plotting propofol
solutions, except without preliminary centrifugation. concentration in blood as a function of time (from the injection
moment of a single propofol intravenous bolus dose, which is
Sorbent 2.5 mg of propofol per 1 kg of the patient's body weight) are
The sorbent for chromatographic separation was obtained by shown in Figure 1. The values for these curves were quanti
the chemical modification of a laboratory-made controlled tated using the regression parameters obtained from appro
porosity glass (CPG) (mean pore diameter, 128 A; specific surface priate calibration graphs. Significant differences between the
2 3
area, 318 m /g; pore volume, 1.17 cm /g; particle size, 7 μm) corresponding magnitudes are seen. The propofol concentra
with ODS according to a procedure described elsewhere (13,14). tions obtained from the precipitation analysis (dashed line)
The coverage density of the support surface by octadecyl radicals are about 30% lower than those obtained from the extraction
2
was 4.01 μmol/m (21.94% carbon content). procedure.
The specific surface area measurement of the support (nec On the basis of this comparison, at least one procedure
essary for coverage density calculation) was carried out by the seems to include either a wrong preparative assumption or
Brunauer-Emmet-Teller (BET) method using a Model 1800 ni incorrectly chosen internal standards. Systematic or manual
trogen sorptomat from Carlo Erba (Milan, Italy). errors made during sample preparation are also possibilities.
The mean pore diameter and pore volume of the CPG were The calibration curves relating to the extraction and the
calculated from the porosimetric data. Porosimetric measure- precipitation procedures are plotted in Figures 2 and 3, re
ments were CARRIED out with a Model 4000 Mercury Porosimeter(Carlo Erba). spectively. Solid lines correspond to the blood standards. The
porosimeter (Carlo Erba).VVV1VVU11 calibration curves for the water series are drawn in dashed
lines. The dashed-dotted line (Figure 3)
refers to the plasma solutions.
By analyzing the two sets of calibration
curves independently, one for the extrac
tion method (Figure 2) and one for the pre
cipitation method (Figure 3), small differ
ences are seen. The slopes of the water
calibration curves (dashed lines) are a little
higher in comparison with the calibration
line constructed for blood and plasma stan
dards. In the precipitation method, the dif
ference is more explicit. These small differ
ences can result from the different behavior
of the internal standards during the sample
preparation procedures. This behavior is
probably connected with the use of two dif
ferent systems: one that contains blood pro
teins (blood or plasma standards) and one
that is protein-free (water standards). This
conclusion is additionally confirmed by a
Figure 2. Calibration curves obtained using the extraction method. The height ratios of propofol and very slight difference between the calibra
the internal standard peaks are plotted against the propofol concentrations. The solid line with rings tion curves constructed in the precipitation
(slope, 0.5649) corresponds to blood solutions. The standard deviation values (n = 7) for concentra
procedure when blood and plasma standard
tions of 0.5,1.0, 2.0, 4.0, 7.0, and 10.0μg/mLwere 0.007, 0.046,0.061, 0.130, 0.271, and 0.353,
solutions were used (compare the solid and
respectively. The dashed line with triangles (slope, 0.5811) corresponds to water solutions. The stan
dashed-dotted lines in Figure 3).
dard deviation values (n = 7) for the same concentrations used for the blood solution were 0.006,
0.042,0.057,0.123,0.259, and 0.341.
The chemical structure of thymol (5-
methyl-2-isopropyl-l-phenol) is very sim-
379
Journal of Chromatographic Science, Vol. 33, July 1995
ilar to that of propofol (2,6-diisopropylphenol). Both the parti greater difference between the structures of propofol and n-
tion coefficients of the substances in the extraction process dibutylphthalate leads to a somewhat greater difference be
and the runs of the calibration curves are very similar. The tween the calibration curves (compare standards for blood,
plasma, and water series) obtained in the pro
cedure involving the precipitation process.
The small differences within Figures 2 and
3 suggest the influence of protein presence
on the behavior of propofol and the internal
standards, but their presence does not ex
plain the significant differences (exceeding
30%) seen in Figure 1.
In Figures 4 and 5, the chromatograms of
the extracts from water (4A) or blood (4B)
containing propofol and the chromatograms
of the supernatants obtained from the precip
itation procedure of plasma (5A), blood (5B),
or water (5C) solutions of propofol are shown.
All these chromatograms correspond to the
samples containing the same amount of
propofol: 4 μg in 1 mL blood, plasma, or water.
As shown in Figure 4, in the extraction
Figure 3. Calibration curves obtained using the precipitation procedure. The height ratios of propofol procedure, the heights of the propofol peaks
and the internal standard peaks are plotted against the propofol concentrations. n-Dibutylphthalate was are almost equal regardless of the type of sol
used as the internal standard. The solid line with rings (slope, 0.4711) corresponds to the blood solu
vent used (blood or water). This proves that
tions. The standard deviation values (n = 7) for concentrations of 0.5,1.0,2.0,4.0, 7.0, and 10.0 μg/mL
almost the same amount of propofol is ex
were 0.009,0.058,0.072,0.148,0.299, and 0.401, respectively. The dashed-dotted line with squares
tracted from blood as from water standards.
(slope, 0.4789) corresponds to plasma solutions. The standard deviation values (n = 7) for analogous
standards were 0.008,0.055,0.067,0.141,0.283, and 0.374, respectively. The dashed line with tri
The case of the precipitation procedure is
angles (slope, 0.5026) corresponds to the water standards. The standard deviation values (n = 7) for different (see Figure 5). As in Figure 4, the
the same concentrations were 0.007, 0.044,0.064,0.132,0.272, and 0.356. similarity of the height ratio of the propofol
peak to the internal standard peak is notice
able and confirms the conclusions made on
the basis of the plots presented in Figures 2 and 3 about the sim
ilar behavior of propofol and the internal standards. The chro
matograms in Figure 5 indicate that the propofol concentration
in the supernatant (after a partial protein precipitation from
blood or plasma solution) is lower than that in the mixture ob
tained by adding the precipitating solution to the water solution
(compare the heights of propofol peaks marked P). The differ
ence is even more explicit between chromatograms 5A (plasma
solution) and 5C (water solution). Because of a small diminution
of plasma sample volume (the result of removing denatured
proteins from plasma after precipitation and centrifugation),
an inverse situation can be expected. Co-precipitation of a por
tion of propofol together with denatured proteins and a probable
propofol decrease due to its adsorption on precipitated proteins
seem to be reasonable explanations of the observed phenomena
(Figure 5) and explain the differences shown in Figure 1.
When chromatograms 5A (plasma solution) and 5B (blood
solution) are compared, the difference in the heights of the
propofol peaks is also observed. This case is of a more complex
nature. Assuming co-precipitation of propofol with denatured
proteins, it should also be taken into account that, in agreement
with the procedure, the solid elements of blood were separated
during the preliminary centrifugation process, decreasing the
sample volume significantly (about 40%). Consequently, despite
Figure 4. Chromatograms of extracts from the water (A) and blood (B)
the propofol concentration diminution (as a result of its co-
solutions containing propofol (4 μg/mL). Peaks P and T are propofol and
precipitation and/or adsorption), the propofol peak observed in
thymol, respectively.
Figure 5B is higher than that in 5A.
380
Journal of Chromatographic Science, Vol. 33, July 1995
Conclusion
381
Journal of Chromatographic Science, Vol. 33, July 1995
8. P. Dailland, I.D. Cockshott, J.D. Lirzin, P. Jacquinot, J.C. Jorrot, J. 13. B. Buszewski, A . Jurasek, J . Garaj, L. Nondek, and D. Berek. The
Devery, J . L . Harmey, and C . Conseiller. Intravenous propofol effect of the reaction medium on the coverage density of C 1 8
during cesarean section: Placental transfer, concentrations in chemically bonded phases. J. Liq. Chromatogr. 10:2325 (1987).
breast milk, and neonatal effects. A preliminary study. Anesthe 14. H. Engelhardt and H. Muller. Optimal conditions for the reversed-
siology 71: 827 (1989). phase chromatography of proteins. Chromatographia 19: 77
9. J . Albanese, C . Martin, B. Lacarelle, P. Saux, A . Durand, and (1984).
F. Gouin. Pharmacokinetics of long-term propofol infusion used for 15. G . E . Berendsen and L. de G a l a n . Preparation and chromato
sedation in I C U patients. Anesthesiology 7 3 : 2 1 4 (1990). graphic properties of some chemical bonded phases for reversed-
10. B. Marsh, M. White, N. Morton, and G . N . C . Kenny. Pharmacoki phase liquid chromatography. J. Liq. Chromatogr. 1:561 (1978).
netic model driven infusion of propofol in children. Br. J. Anaesth. 16. A . L . D a w i d o w i c z , J. Matusewicz, and J. Wysocka-Lisek. Deter
67:41 (1991). mination of residual boron in thermally treated controlled-porosity
11. G.F. Plummer. Improved method for the determination of propofol glasses, by colorimetry, spectrography and isotachophoresis.
in blood by high performance liquid chromatography with fluo Talanta 36:581 (1989).
rescence detection. J. Chromatogr. 421:171 (1987). 17. A.L. Dawidowicz, A. Waksmundzki, and A. Derylo. Column pack
12. I. Pavan, E. Buglione, M. Massiccio, C . Gregoretti, L. Burbi, and ings for liquid permeation chromatography obtained by crushing
M. Berardino. Monitoring propofol serum levels by rapid and coarse fractions of porous glasses. I. Evaluation of applicability of
sensitive reversed-phase high-performance liquid chromatography glass packings. Chem. Anal. (Warsaw) 24:811 (1979).
during prolonged sedation in I C U patients.j. Chromatogr. Sci. 30:
164-66(1992). Manuscript accepted October 1 3 , 1 9 9 4 .
382