Mx8000 Systems Installation Manual
Electrical Connections
491-781-3501
Route all cables as indicated in Table 6-/ and shown in Figure 6-12 and Figure 6-13.
Table 6-1. Mains and Protective Conductor Connections
Cable | Type of Cable or From To Comments
No. Function
W1_| Mains cable PDC Gantry Left Leg
W2 _ | Protective PDC Gantry Right Leg
conductor
W3 | UDC Power cable PDC Gantry Right Leg | Shielded cable, for the generator,
w4 Mains cable PDC Heat Exchanger
ws _ | Protective PDC Heat Exchanger | Route parallel with W4.
conductor
we Patient table powey Gantry Left Leg Patient Table | Already routed in Gantry channel.
W7 | Patient Table PE | Gantry LeftLeg| Patient Table | Already routed in Gantry channel.
W11 | Protective PDC ‘Computer Rack
conductor
ws0 | Gantry control Gantry Left Leg | Computer Rack | 50 Pin D-Sub connector
cable
ws1 | MCU control cable! Ganty Left Leg PDC 9 Pin D-Sub connector
w52_| Patient Table Gantry LeftLeg | Patient Table | Already routed in Gantry channel.
control cable
Ws3 | WCS control cable| Gantry Left Leg | Heat Exchanger | 9 Pin D-Sub connector
ws54 | PDC control cable PDC ‘Computer Rack | 15 Pin D-Sub connector
Ws55 _ | Data cable Gantry Left Leg | Computer Rack | Fiber optic cable
waes | Hamess between Console ‘Computer Rack | Permanently connected on
Rack and Console console side.
‘waao | Control cable Gantry Left Leg | Computer Rack | GPC
wast | Mains computer PDC ‘Computer Rack
cable
‘W882 | Line voltage cable PDC Local Power | Shorten to needed length.
Supply
was3_| Additional PDC Local Power | Local ground connected to the
protective ground Distribution Panel] building ground bar.
41 Mains cable PDC Radiation ON | On-site cable for connecting the
Indicator | non-referenced relay switch
contact (250V / 5A ~ AC)
2 Control cable PDC Door Switch On-site cable