APPENDIX A: ILLUSTRATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . 27
NJEX Model 6000 Pump Assembled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
NJEX Model 6000 Pump Actuation Assembly, Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
NJEX Model 6000 Pump Check Valve Assemblies, and Seals, Exploded View . . . . . . . . . . . . . . . . . . . . 29
NJEX Gas Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Z-65/3.10 Controller (exploded) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Z-65 Wiring Control Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Page i
SECTION 1: First Things to Know About the 610G
How to Use this Manual Getting Help
The NJEX-610G Operations Manual is a step- This manual provides solutions to typical ques-
by- step guide containing the procedures need- tions about the 610G system. If the answer can
ed to work with the 610G System. not be found within this manual, contact YZ
The NJEX System Series of odorizers imple- Systems at:
ment the most advanced technology available
in the industry. It is recommended that the T: 1.281.362.6513
technicians working with the NJEX Odoriza- T: 1.800.653.9435 (800.NJEX.HELP)
tion Systems study the manual prior to initiating F: 1.281.362.6513
work on the system for the first time. Em: Service@yzhq.com
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 1
NJEX 610G 10-2011
SECTION 1: First Things to Know About the 610G
Operation Specifications
Maximum Odorant Output: .38 gallons/day
(1.44 liters/day)
Maximum Operating Pressure: 1,440 psig
(99.28 Bar (g))
Operating Temp Range: 0 to 140 ºF
(17°C to 60°C)
Power Supply: LBP-14 Battery
(standard)
Gas Flow Input Signal: pulse
(dry contact or
voltage pulse)
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 2 NJEX 610G 10-2011
SECTION 1: First Things to Know About the 610G
Theory of Operation CAUTION:
Proportional-To-Time Operation: In this mode of Excessive tubing lengths should be avoided.
operation, the 610G injects odorant into the pipe- Installation of the NJEX 610 Odorization system
line at regular time intervals. The volume injected should be as close to the point of injection and
per stroke is set by inserting a stroke spacer Odorant Tank as possible. For answers to your
into the actuation plug of the 6000B pump. The questions contact, YZ Systems Technical Ser-
Z-65/3.10 operates as a recycling timer, periodi- vices at; 800.653.9435 or 1.281.362.6500.
cally energizing a low power solenoid valve. En-
ergizing the solenoid valve allows actuation gas
to stroke the 6000B pump. The rate at which this
occurs is set by the operator using two 10-posi-
tion switches on the Z-65/3.10. The number of
times the solenoid output is activated is recorded
by the onboard LCD indicator.
Power Options:
Long-Life Battery: The Z-65/3.10 timer/counter
operates using a replaceable internal battery
pack. As the standard power source, the bat-
tery pack condition is monitored by way of two
indicator LEDs. When the battery pack needs
replacement, the red LED will illuminate when the
solenoid output is activated. If the battery pack
is good, the green LED will illuminate when the
solenoid is activated. Typical battery life is one
year based on a two minute/stroke rate.
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 3
NJEX 610G 10-2011
SECTION 1: First Things to Know About the 610G
System Accessories
• Odorant Injection Probe, includes a 316 stain- • The sight glass should be fitted with a 1/4”
less steel probe and isolation valve for loca- MNPT x 1/8”t bore through fitting (P/N
tion at the pipeline. When ordering, please A1-0283) and 1/8” S.S. tubing extending
specify pipeline connection required, ½″ or through the bore of the sight glass fitting
¾″. until it just protrudes from the 1/4” dripper
point.
• Odorant Injection Probe with Sightglass, (P/N
A1-0238) 1/8” drop adapter includes a 316
stainless steel probe, visual odorant sight
Application Update
indicator, and an isolation valve for location at • When installed and operated properly we
the pipeline. When ordering, please specify can consistently perform at the smallest
pipeline connection required, ½″ or ¾″. pump displacement of 0.02cc/stroke with
as much as 30 minutes between strokes;
• 1/8″ stainless steel discharge tubing In-line therefore time intervals will be approved
Check Valve. For placement in the odorant between 6 sec. and 30 minutes.
discharged tubing line immediately preceding
the probe assembly, (P/N A3-0438). • Keep in mind that with the LVO there is no
internal microprocessor to do calculations,
• ¼″ stainless steel tubing Dielectric Isolator or make adjustments; therefore, the LVO
Union. These should be installed in every must be operated in Proportional to Time
tubing line that attaches the odorizer to the mode (0.1 min, to 30 min.), or in the Flow
pipeline in any manner. For example the sup- mode as a slave to counting “XX” number
ply gas, odorant discharge, and differential of pulses. The pulses cannot come faster
pressure switch connections, (P/N A1-0182). than 1.5 hz or approximately 670 ms be-
tween pulses, and each pulse must have
• NJEX Scrubbers. These filters are designed a minimum dwell time of 21 ms, and you
to scrub the exhaust gas vented from the cannot count more than 99 pulses before
pumps. They are available by ordering P/N stroking the pump.
C4-0024, 15 gallon scrubber.
• Finally remember the battery life on the
A complete line of odorization accessories rang- LVO will be limited by stroke frequency and
ing from pre-odorized gas scrubbers to injec- temperatures. The faster it runs and the
tion probes is available through YZ. Please colder it gets, the shorter the battery life will
contact your local representative or YZ toll free be.
at 800.344.5399. For technical support call
800.653.9435.
Installation Guidelines
• We recommend that the tubing from the 610
LVO be adapted to 1/8” tubing before it reach-
es the sight glass.
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 5
SECTION 2: System Installation
System Flow Schematic
Figure 3
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 6 NJEX 610G 8-2011
SECTION 2: System Installation
Standard System Mounting
Figure 3
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 7
SECTION 2: System Installation
Standard System Connections
Figure 4
Required field connections to place the 610G into
operation are as follows:
CAUTION:
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 8 NJEX 610G 8-2011
SECTION 3: Filling The Bulk Odorant Tank
Filling the Tank for the First Time ing out. Allow this process to continue until
all ambient air from the tank is purged, and
CAUTION: only inert gas or natural gas is emitting from
Odorant has a very strong odor, which if allowed to this valve, then close V11 and V10. The time
escape to the atmosphere, may cause problems in required to accomplish this task will vary with
the local community. Take necessary precautions the tank size.
when filling an odorant storage tank to assure that 4. Vent purge gas by opening V11 partially until
the local community is not disrupted during the tank pressure just reaches zero, and then
filling process. Verify that the entire system has close V11.
no pressure in it before beginning. Additionally, all 5. Attach odorant supply to V10, open V10, and
personnel should wear protective clothing, and use begin transferring odorant to the bulk tank.
equipment as recommended by the chemical manu- • The tank level indicator will not start to
facturer during this time. If you are uncertain about rise until approximately 1 gallon of odor-
any aspect of the odorant itself, you should contact ant is in the tank
the manufacturer of your chemical prior to proceed- • When 2-gallons of odorant is in the tank
ing. The MAXIMUM amount of odorant in the the tank level should show approximately
standard tank should never exceed 4 gallons. 1.5” on the level indicator tape.
• 3-gallons should show approximately 3”
1. Verify correct position of valves before begin- on the level indicator tape.
ning, Figure 5. • 4-gallons should show approximately 4.5”
• Closed: V10, V11, and V13 on the level indicator tape.
2. Attach inert or natural gas supply to V10. 6. Connect a line from V11 to a flare or vapor
recovery device, and open V11.
3. To purge the tank open valve V10 to introduce
inert or natural gas to the tank to begin displac-
CAUTION:
ing any ambient air from the empty tank. Con-
Fill tank to a maximum level of 80% of the tank
tinue until pressure on the gage located directly
capacity (4-gallons).
below V10 is observed, then partially open V11
to allow ambient air from the tank to begin flow-
7. Close V10 and V11, and remove odorant
Figure 5 transfer equipment, and line to flare or vapor
recovery device.
8. Continue through the remaining procedures
in this manual.
LEGEND
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 2-2012 Page 9
SECTION 3: Filling the Bulk Odorant Tank
Refilling the Bulk Odorant Tank
4. Attach odorant supply to V10, open V10, and
CAUTION:
begin transferring odorant to the bulk tank.
Odorant has a very strong odor, which if allowed to
escape to the atmosphere, may cause problems
CAUTION:
in the local community. Take necessary precau-
Fill tank to a maximum level of 80% of the tank
tions when filling an odorant storage tank to assure
capacity (4-gallons total or about 4.5 inches on the
that the local community is not disrupted during
level indicator tape).
the filling process. Verify that the entire system
• When 2-gallons of odorant is in the tank the
has no pressure in it before beginning. Additionally
tank level should show approximately 1.5” on
all personnel should wear protective clothing, and
the level indicator tape.
use equipment as recommended by the chemical
• 3-gallons should show approximately 3” on the
manufacturer during this time. If you are uncertain
level indicator tape.
about any aspect of the odorant itself, you should
• 4-gallons should show approximately 4.5” on
contact the manufacturer of your chemical prior to
the level indicator tape.
proceeding.
5. Close V10 and V11, and remove odorant trans-
1. Place the Z-65 controller in the OFF Mode by
fer equipment, and line to flare or vapor recov-
placing Mode Switch #1 in the Down position.
ery device.
2. Verify correct position of valves before begin-
6. Open V13 until the storage tank pressure is 15
ning, Figure 6.
psig.
• Close: V10, V11, V13 7. Place the Z-65 controller in the ON Mode by
placing Mode Switch #1 in the Up position.
3. Connect a line from V11 to a flare or vapor re-
covery device, and open V11.
Figure 6
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 10 NJEX 610G 2-2012
SECTION 4: System Control & Electronics
Overview Figure 7
SAFETY NOTES
Always use extreme care when performing
maintenance on Sampling Systems. Always take
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 11
SECTION 5: Programming for Proportional-to-Flow Operation
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 12 NJEX 610G 8-2011
SECTION 5: Programming for Proportional-to-Flow Operation
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 13
SECTION 6: Programming for Proportional-to-Time Operation
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 14 NJEX 610G 8-2011
SECTION 7: Mechanical System
Overview
The 610G mechanical system is composed
of the mechanical/electrical enclosure. Indi-
vidual components of the system are shown
below and described in the following pages.
Figure 17
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513NJEX 610G 8 - 2 0 11
NJEX 610G 8-2011 Page 15
SECTION 7: Mechanical System
NJEX Gas Filter Figure 18
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 16 NJEX 610G 8-2011
SECTION 7: Mechanical System
Model 6000 pump Figure 19
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513NJEX 610G 8 - 2 0 11
NJEX 610G 8-2011 Page 17
SECTION 8: System Maintenance
Preventative Maintenance Recommended Spare Parts List
Schedule Part # Description Recommended
A preventative maintenance program serves to Quantity
anticipate maintenance issues prior to waiting A4-0003 3-way solenoid valve 1
until the system requires service. Like changing C4-0133 NJEX gas filter replacement
the oil & filters in an automobile, by choosing to filter element 1
service the various parts and operation in the D3-0140 Model 6000B pump seal
NJEX System at regular intervals, the techni- replacement kit 1
cian can perform the maintenance service when D3-0142 Z-65/200 fuse replacement kit 1
desired, rather than when required, such as in the E3-2001 LBP-14 internal Z-65 battery 1
middle of night.
Recommended Maintenance
Schedule
Weekly Inspection
Annual Inspection
1. Change filters
2. Rebuild pump
3. Replace solenoid
4. Test regulators and service, as neede
5. Condition the odorant, as needed.
6. Test the NJEX System performance
7. Replace the battery (More frequent replace-
ment may be required if the odorizer strokes
more than 5 times per hour)
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 18
NJEX 610G 8-2011
SECTION 9: System Troubleshooting
How to Use This Section Step-by-Step Resolution
The recommendations contained in this sec- Using a step-by-step method to resolve issues
tion should be used as a preliminary information on the NJEX System will reduce maintenance
resource to remedy operational issues with the time and assist in returning the odorization sys-
NJEX System. It is important to read all of the tem to service quicker.
definitions and notes prior to initiating work. The following represent the recommended chro-
Each sub-section contains a description of the nology to resolve issues:
alarm and non-alarm indicators followed by a 1. Re-establish the correct pressures
step-by-step trouble shooting procedure As- a. Bulk Storage Tank, 25psi (2.07-2.41 Bar)
sistance is available 24 hours a day, 7 days a b. Actuation Supply, 50-60 psi (3.45-4.13
week, 365 days a year, via the telephone num- Bar)
bers listed above. 2. Verify that the odorant storage tank has suf-
ficient odorant in it to supply the NJEX 610G
with odorant.
For Additional Help
Any issue that can not be resolved through the
use of this reference, please contact YZ Techni-
cal Service at:
T: 1.800.653.9435 (1.800.NJEX-HELP)
T: 1.281.362.6500, International Calls
F: 1.281.362.6513
Em: Service@yzhq.com
SAFETY NOTES
• Always use extreme care when perform-
ing maintenance on an odorization system.
Check to ensure the removal of liquid odor-
ant and pressure from the portion of the
system on which work will be performed prior
to removing components or fittings.
• Inspect all tube fittings and valve packings
semi-annually to ensure that liquid odorant
remains within the system.
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 19
SECTION 9: 610G System Troubleshooting
Battery Test: Figure 21
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 20 NJEX 610G 8-2011
SECTION 9: System Troubleshooting
Troubleshooting: Timer Mode Figure 23
DPS Test:
Figure 24
A. Set the mode switches as indicated:
1. Mode switches #1 and #2 to the on posi
tion.
2. Mode switches #3 and #4 to the off posi
tion.
B. Set the time dials to 00 to enter the diag
nostic mode.
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 21
SECTION 9: 610G System Troubleshooting
Figure 25
Troubleshooting: Timer Mode
the on position.
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 22 NJEX 610G 8-2011
SECTION 9: System Troubleshooting
Troubleshooting: Timer Mode Figure 27
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 23
SECTION 9: 610G System Troubleshooting
Troubleshooting: Figure 29
Dry contact Ter. #5
Counter
open collector Mode
Dry contact
(20 mSec duration min.) Ter. #3 Ter. #5
open collector
Count Input Setting (20 mSec duration min.) Ter. #3
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 24 NJEX 610G 8-2011
SECTION 9: System Troubleshooting
Troubleshooting:
7. Ascertain whether the pump is properly
Pump Performance
stroking.
1. Verify operating conditions. Correct as a. If pump actuation does not occur as the
needed. Test key is pressed, remove the actuation
gas line at the top of the pump and test
2. Check to see if the pump operation the pump stroke again. As the Test key is
sounds the same as previously. A dis- pressed, there should be a burst of gas at
tinct bottoming of the plunger piston and the open actuation supply connection. If
a return to the top of the plunger piston the gas is released from the open actuation
housing should be audible with supply line skip to subsection 7.e., other-
each actuation. If not, remove the actua- wise continue to 7.b.
tion cylinder. Inspect for a broken return
spring, or a stuck / sticking actuator b. If solenoid discharge to the pump con-
piston or plunger. Replace the spring if nection does not release gas, remove the
broken. Clean and lubricate the actuation pump solenoid wiring cable. Next connect
cylinder and actuation piston assembly. a volt meter to the two parallel posts of the
Manually push the plunger into the seal solenoid cable. As the Test key is pressed,
assembly and ensure it returns com- observe if a 12VDC current becomes pres-
pletely and freely. If sticking continues ent. If a voltage is detected the solenoid
to occur, the seals should be replaced. should be replaced.
Reassemble and see if alarm reoccurs
after a system restart. c. If the above test indicates the solenoid
is functioning, next determine if the signal is
3. If pump actuation is normal: being sent by the controller, or lost in the ca-
bling and connections. Go to the Z-65 Con-
a. Inspect the discharge lines and troller terminal strip and connect your volt
valves for a restriction. meter, to terminals 1 and 2, with 1 being the
positive wire and 10 being the negative wire.
b. Inspect all the check valves on the Now, test fire the pump with the test switch.
discharge side of the pump, from the A momentary voltage pulse, approximately
NJEX System to the pipeline, for proper .2 seconds in duration, should be observed.
operation. This will be a DC voltage pulse, approxi-
mately 12VDC. If the pulse is at the termina-
c. Verify that the operating conditions tion strip, next check the polarity of connec-
remain unchanged and correct as needed. tions of the wiring. Try reversing the terminal
wiring connections for the terminals #1 and
4. Verify that the actuation pressure re #2, and repeat steps 7a and 7b. If this does
mains unchanged from previous setting. not correct the problem the cable assembly
to the solenoid should be replaced.
5. Verify that all valves and check valves
are properly set system operation. d. If no voltage is detected, Check the
Inspect for valve settings that can restrict fuses located under the face plate of the
pump displacement. Z-65, and replace as necessary.
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 26 NJEX 610G 8-2011
APPENDIX A: Illustrations
Figure 32
Actuation Gas
Actuation Gas
Odorant
Odorant Outlet
Odorant Inlet
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 27
APPENDIX A: Illustrations
* O-Ring
P/N A5-1013
Optional
Stroke Spacer
60% P/N B0-1041
40% P/N B0-1043
20% P/N B0-1040
* O-Ring
P/N A5-1127
Actuation Plug
P/N B0-1039
Piston/Plunger
P/N B0-1005 * Actuation Seal
P/N A6-0076
Actuation Spring
P/N C3-0036
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 28 NJEX 610G 8-2011
APPENDIX A: Illustrations
Guide Cartridge
P/N B0-1037
O-Ring
P/N A5-1113
Stop Bushing
P/N B0-1033
INLET ASSEMBLY
CV Inlet Wafer
P/N B0-1021 O-Ring
A5-1010
O-Ring Inlet Fitting
P/N A5-1909 P/N B0-1031
O-Ring
O-Ring P/N A5-1016
P/N A5-1117
O-Ring
P/N A5-1118
Bottom Cartridge
P/N B0-1030
Filter Bowl
PN: C4-0131
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 30 NJEX 610G 8-2011
APPENDIX A: Illustrations
11
8
12
4 2
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 8-2011 Page 31
APPENDIX A: Illustrations
Figure 60
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
Page 32 NJEX 610G 8-2011
Notes: _______________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 10-2011
201 Ivyland Road
Ivyland, PA 18974
P: 800.344.5399
P: 281.362.6500
F: 281.362.6513
Sales@yzhq.com
www.yzsystems.com
YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • F: 281.362.6513
NJEX 610G 2-2012