Professional Documents
Culture Documents
B
A
Column
b. Give and discuss the two configurations for the part labelled A.
Split mode: the split vent remains open during analysis. This allows for samples with
high concentration, usually greater than 1% to be analysed without overloading the
column. In this way, up to 99% of the sample goes to waste.
Splitless mode: the split valve is closed during injection. This allows for a greater
proportion of the sample to enter the column and thus reach the detector. Used for
samples with low analyte concentration (trace analysis, less than 0,1%).
c. The column shown is a capillary column. How is this different from a
packed column?
Solid support is packed into column Solid support is coated on column wall
0.1mm)
Usually constructed from stainless steel Constructed from polymer and glass
Typically used to analyse for permanent Used to analyse most organic compounds
gases
d. If you had to analyse an air sample, would this column be suitable? Give
a reason for your answer.
No. Air consists of permanent gases such as O2 and N2 and a packed column is
required to analyse it. Capillary columns contain a stationary phase which does not
interact with these gases.
2. Based on the working principle of the FID, explain why organic compounds are
more easily detected by this detector.
Based on the working principle with organic compounds such as hydrocarbons, the
gas stream from the column is mixed with hydrogen and burned. Positive ions (e.g.
CH3+ or CHO+) and electrons are produced in the flame when organic substances
are present.
The ions are collected at electrodes and produce a small, measurable current.
The current is proportional to the amount of Carbon in the gas stream. Since the
amount of carbon is being measured, inorganic compounds or permanent gases
cannot be detected.
4. The chromatograms shown are of the same six compounds, but using two
different oven temperature programmes. Explain in detail why there is poor
separation in Figure 1 and similarly why the separation has improved in Figure
2.
Figure 1
Figure 2
Increasing the thickness of the stationary phase allows the separation of larger
sample sizes
Increasing the thickness of the stationary phase reduces efficiency
Increasing the thickness of the stationary phase is better for separation of highly
volatile compounds due to increased retention.