Födisch AG :: Specifications

4.2.3 Gas Analysers

MGA 23 MGA 05 MCA 04 MCA 04 Analysenschrank O2MD 04

4.2.3.1 Measuring principle MGA 23


By means of the gas analyser MGA 23 up to 4 gas components can be measured at the same time:
maximum 3 infrared-active gases as e.g. CO, CO2, NO, SO2, CH4, R22 (Frigen CHCIF2) as well as O2 with
an electrochemical oxygen measuring cell.

Basic versions of MGA 23 for:


1 Infrared-gas component with/without oxygen measurement
2 Infrared-gas components with/without oxygen measurement
3 Infrared-gas components with/without oxygen measurement

Smallest TÜV-approved and permitted measuring ranges:


1- and 2-component-analyser
CO: 0 to 150 mg/m³
NO: 0 to 250 mg/m³
SO2: 0 to 400 mg/m³

3-component-analyser
CO: 0 to 250 mg/m³
NO: 0 to 400 mg/m³
SO2: 0 to 400 mg/m³
All higher measuring ranges are also allowed.

Special configurations:
The MGA 23 with 2 IR-components without pump and with or without oxygen is also available with 2
separate gas ways. This allows the measurement of 2 measuring points or e.g. at the NOx-measurement the
state before and after converter.

2. Edition, Conditions May 2007 37