Operating Instructions
Liquisys M CLM223/253
Transmitter for Conductivity
About this document
Safety messages
The structure, signal words and safety colors of the signs comply with the specifications of
ANSI Z535.6 ("Product safety information in product manuals, instructions and other
collateral materials").
Symbols
Endress+Hauser
Liquisys M CLM223/253
Table of contents
1 Basic safety instructions . . . . . . . . . . . . . 5 6 Commissioning . . . . . . . . . . . . . . . . . . . 30
1.1 Requirements for the personnel . . . . . . . . . . . . . . . 5 6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
1.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6.2 Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
1.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6.3 Quick start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
1.4 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6.4 System configuration . . . . . . . . . . . . . . . . . . . . . . . 35
1.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 6.4.1 Setup 1 (Conductivity) . . . . . . . . . . . . . . . . 35
1.6 Electrical symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 6.4.2 Setup 2 (Temperature) . . . . . . . . . . . . . . . 36
6.4.3 Current input . . . . . . . . . . . . . . . . . . . . . . . 39
2 Incoming acceptance and product 6.4.4 Current outputs . . . . . . . . . . . . . . . . . . . . . 42
6.4.5 Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
identification . . . . . . . . . . . . . . . . . . . . . . 7 6.4.6 Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . 7 6.4.7 Relay contact configuration . . . . . . . . . . . 49
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.4.8 Temperature compensation with table . . 62
2.3 Product identification . . . . . . . . . . . . . . . . . . . . . . . . 8 6.4.9 Concentration measurement . . . . . . . . . . 63
2.3.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 6.4.10 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
2.3.2 Identifying the product . . . . . . . . . . . . . . . . 8 6.4.11 E+H Service . . . . . . . . . . . . . . . . . . . . . . . . . 67
2.4 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8 6.4.12 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . 68
2.4.1 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 6.5 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
2.4.2 CSA general purpose . . . . . . . . . . . . . . . . . . 8 6.6 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Endress+Hauser 3
Liquisys M CLM223/253
12 Appendix . . . . . . . . . . . . . . . . . . . . . . . .106
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . .110
4 Endress+Hauser
Liquisys M CLM223/253 Basic safety instructions
Endress+Hauser 5
Basic safety instructions Liquisys M CLM223/253
Safety icons
Warning!
# This symbol alerts you to hazards. They can cause serious damage to the instrument or to
persons if ignored.
Caution!
" This symbol alerts you to possible faults which could arise from incorrect operation. They
could cause damage to the instrument if ignored.
! Note!
This symbol indicates important items of information.
Electrical symbols
Ground connecting
) A terminal which, from the user’s point of view, is already grounded using a grounding
system.
Alarm relay
b
Input
Output
DC voltage source
Temperature sensor
6 Endress+Hauser
Liquisys M CLM223/253 Incoming acceptance and product identification
Endress+Hauser 7
Incoming acceptance and product identification Liquisys M CLM223/253
2.3.1 Nameplate
The nameplate contains the following information:
• Manufacturer data
• Order code
• Extended order code
• Serial number
• Operating conditions
• Safety icons
Compare the order code on the nameplate with your order.
2.4.1 CE mark
Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus
complies with the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
Version Approval
8 Endress+Hauser
Liquisys M CLM223/253 Mounting
3 Mounting
1
ENDRESS+HAUSER
LIQUISYS S
REL 1
ALARM
REL 2
CAL REL
REL 1
E REL 2
ENDRESS+HAUSER
LIQUISYS S
6 REL 1
ALARM
REL 2
CAL REL
REL 1
5 E REL 2
a0008937
Endress+Hauser 9
Mounting Liquisys M CLM223/253
Pg 13.5
Pg 7
Pg 16
11/
0.43
M5
157 / 6.18
REL1
ALARM
REL2
CAL REL
REL1
247 / 9.72
E
70 / 2.76
REL2
Ø 0.24
Ø6/
154 / 6.06
170 / 6.69
mm / inch
a0005733
There is a hole in the punching for the cable entry (connection of supply voltage). It
serves as a pressure balance during air freight dispatching. Make sure no moisture
penetrates the inside of the housing before the cable installation. The housing is
completely air-tight after the cable installation.
1 4
a0005734
10 Endress+Hauser
Liquisys M CLM223/253 Mounting
6/
0.24
96 / 3.78
92 / 3.62
90 / 3.54
139 / 5.47
149.5 / 5.89
mm / inch
a0005735
Endress+Hauser 11
Mounting Liquisys M CLM223/253
Ø 0.24
Ø6/
154 / 6.06
mm / inch
a0005736
12 Endress+Hauser
Liquisys M CLM223/253 Mounting
You require a post mounting kit to secure the field device to horizontal and vertical posts
or pipes (max. Ø 60 mm (2.36")). The kit can be acquired as an accessory (see
"Accessories" section).
3
mm / inch
Ø max. 60 /
2.36
1
1
3
a0005737
Endress+Hauser 13
Mounting Liquisys M CLM223/253
You can also secure the field device to a square universal post in conjunction with the
weather protection cover. These can be acquired as accessories, see "Accessories" section.
3
2
1
3
a0005738
Fig. 7: Mounting field device with universal posts and weather protection cover
14 Endress+Hauser
Liquisys M CLM223/253 Mounting
+0.5 +0.02
92 / 3.62
92+0.5 / 3.62+0.02
1
2
3.62
92 /
90 / 3.54
96 / 3.78
3
3
57 / 2.24
1.30
33 /
max. 45 /
approx. 25 / 139 / 5.47 1.77 6/
0.98 0.24 mm / inch
approx. 165 / 6.50*
a0005739
Endress+Hauser 15
Electrical connection Liquisys M CLM223/253
4 Electrical connection
! WARNING
Device is energized
Improper connection can cause injury or death.
‣ The electrical connection must only be carried out by a certified electrician.
‣ Technical personnel must have read and understood the instructions in this manual and
must adhere to them.
‣ Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
4.1 Wiring
16 Endress+Hauser
Liquisys M CLM223/253 Electrical connection
PE
84 85+
15 V
15 H
86 _
A 16
83
41
42
84 I
43
15
B 16
47
83
48
J
49
S
11 57
12 58
C K
13 59
51
31
mA 52
D L
32
Lf 53
33 54
mA 55
E 34 M
56
81 23 +
24 -
F 10-50 V N
82
mA
~ =
93 L1 L+
N L–
G 10-50 V O
94
PE
a0008920
Endress+Hauser 17
Electrical connection Liquisys M CLM223/253
54 55 56 51 52 53 57 58 59 47 48 49 41 42 43
YE Mains
Hilfsenergie
83 13 84 15
NC
NC 31 32 33 34 NC 85 86 93 94 81 82 23 24 NC L1 N (AC)
+ - + - + - + - + - + - + - (DC)
4..20mA
Binary
Binär/
Binary
Binär/
16 11 12 S
GN WH
(opt.)
Temp.
10mA
+15V
Lf
13 1
1
11 12 131095-4F
PE
a0008915
18 Endress+Hauser
Liquisys M CLM223/253 Electrical connection
a0008912
NOTICE
Nonobservance could cause incorrect measurement
‣ Terminals marked NC may not be wired.
‣ Unmarked terminals may not be wired.
Please label the sensor terminal block with the sticker provided.
Inductive sensors CLS50, CLS52 Cable permanently attached Junction box VBM + CLK5
to sensor
Conductivity measurement (inductive) max. 55 m (180.46 ft) with CLK5 (sensor cable incl.)
Endress+Hauser 19
Electrical connection Liquisys M CLM223/253
83
83 84
A
A sc 15
84
16 sc
sc
WH
BK RD
GN 11 GN 11
WH 12 BN(nc)
WH 12
YE 13
YE 13
BN(nc)
CYK71 CLK5
a0008938-de
For further information on the cables and junction boxes, please refer to the
"Accessories" section.
Field instrument measuring cable connection
Proceed as follows to connect a conductivity sensor to the field instrument:
1. Open the housing cover to access the terminal block in the connection compartment.
2. Break the punching of a cable gland from the housing, mount a cable gland and guide
the cable through this cable gland.
3. Connect the cable in accordance with the terminal assignment (see connection
compartment sticker).
4. Tighten the cable gland.
20 Endress+Hauser
Liquisys M CLM223/253 Electrical connection
PE 11 12 13 83 16 15 84
S 11 12 13 83 16 15 84
G N W H YE
GN WH YE BU RD RD BU
BK
RD WH
CYK71/CPK9
CLK5
GN WH YE BK
3 1 2
a0008919 a0008918
Fig. 13: Connection of conductive sensors Fig. 14: Connection of inductive sensors
A B
41 41
42 42
43 43
a0006415
Endress+Hauser 21
Electrical connection Liquisys M CLM223/253
22 Endress+Hauser
Liquisys M CLM223/253 Operability
5 Operability
5.2.1 Display
LED display
REL 1
Indicates the activated relay in the "Manual" mode (red LED)
REL 2
Endress+Hauser 23
Operability Liquisys M CLM223/253
LC display
1 2 3 4 5
mS/cm
12 15 R232
13 6
12 Setpoint 8
11 10 9
a0008922-en
1 7
mS/cm
12.15 R232
Setpoint
2 6
3 5
4
a0008921-en
24 Endress+Hauser
Liquisys M CLM223/253 Operability
CAL key
When you press the CAL key, the device first prompts you for the
calibration access code:
• Code 22 for calibration
• Code 0 or any other code for reading the last calibration data
Use the CAL key to accept the calibration data or to switch from field
to field within the calibration menu.
ENTER key
When you press the ENTER key, the device first prompts you for the
setup mode access code:
• Code 22 for setup and configuration
• Code 0 or any other code for reading all configuration data.
The ENTER key has several functions:
• Calls up the Setup menu from the measuring mode.
• Saves (confirms) data entered in the setup mode.
• Moves on within function groups.
REL key
In the manual mode, you can use the REL key to switch between the
relay and the manual start of cleaning.
REL 1 In the automatic mode, you can use the REL key to read out the
switch-on points (for limit contactor) or set points (for PID controller)
REL 2
assigned to the relay in question.
Press the PLUS key to jump to the settings of the next relay. Use the
REL key to get back to the display mode (automatic return after 30 s).
Endress+Hauser 25
Operability Liquisys M CLM223/253
AUTO key
You can use the AUTO key to switch between automatic mode and
manual mode.
Escape function
If you press the PLUS and MINUS key simultaneously, you return to
the main menu or are taken to the end of calibration if calibrating. If
you press the PLUS and MINUS key again, you return to the
measuring mode.
26 Endress+Hauser
Liquisys M CLM223/253 Operability
3. To enable the manual mode, enter the code 22 via the PLUS and
MINUS keys. The bottom LED beside the AUTO key lights up.
5. Switch the relay. It is switched on with PLUS and switched off with
MINUS.
The relay remains in its switched state until it is switched over
again.
6. Press the AUTO key to return to the measuring mode, i.e. to the
automatic mode. All the relays are triggered again by the
transmitter.
Endress+Hauser 27
Operability Liquisys M CLM223/253
Operating modes
Measuring mode:
standard mode of
operation, displaying
current measured values
Code
If no key is pressed in the setup mode for approx. 15 min, the device automatically
returns to the measuring mode. Any active Hold (Hold during setup) is reset.
Access codes
All device access codes are fixed and cannot be altered. When the device requests the access
code, it distinguishes between different codes.
• Key CAL + Code 22: access to Calibration and Offset menu
• Key ENTER + Code 22: access to the setup menus
• Keys PLUS + ENTER: locks the keyboard
• Keys CAL + MINUS: unlocks the keyboard
• Key CAL or ENTER + any code: access to read mode, i.e. all the settings can be read but
not modified.
The device continues measuring in the read mode. It does not shift to the Hold status. The
current output and the controllers remain active.
28 Endress+Hauser
Liquisys M CLM223/253 Operability
Menu structure
The configuration and calibration functions are arranged in function groups.
• In setup mode, select a function group with the PLUS and MINUS keys.
• In the function group itself, switch from function to function with the ENTER key.
• Within the function, select the desired option with the PLUS and MINUS keys or edit the
settings with these keys. Then confirm with the ENTER key and continue.
• Press the PLUS and MINUS keys simultaneously (Escape function) to exit programming
(return to the main menu).
• Press the PLUS and MINUS simultaneously keys again to switch to the measuring mode.
If a modified setting is not confirmed with ENTER, the old setting is retained.
An overview of the menu structure is provided in the Appendix to these Operating
Instructions.
1
-
E E E
E E
C07-CxM2x3xx-19-06-00-xx-002.eps
Endress+Hauser 29
Commissioning Liquisys M CLM223/253
6 Commissioning
6.2 Switching on
Familiarize yourself with the operation of the transmitter before it is first switched on.
Please refer in particular to the "Safety instructions" and "Operation" sections.
After power-up, the device performs a self-test and then goes to the measuring mode.
Now calibrate the sensor in accordance with the instructions in the "Calibration" section.
During commissioning, the sensor must be calibrated so that the measuring system can
return precise measurement data.
Then perform the first configuration in accordance with the instructions in the "Quick start-
up" section. The values set by the user are kept even in the event of a power failure.
The following function groups are available in the transmitter (the groups only available in
the Plus Package are marked accordingly in the functional description):
Setup mode
• SETUP 1 (A)
• SETUP 2 (B)
• CURRENT INPUT (Z)
• CURRENT OUTPUT (O)
• ALARM (F)
• CHECK (P)
• RELAY (R)
• TEMPERATURE COMPENSATION (T)
• CONCENTRATION MEASUREMENT (K)
• SERVICE (S)
• E+H SERVICE (E)
• INTERFACE (I)
Calibration mode
• CALIBRATION (C)
A detailed explanation of the function groups available in the transmitter can be found
in the "System configuration" section.
Function display:
The displayed code indicates
s the function position in the
2 F3 function group.
C07-CLD132xx-07-06-00-en-003.eps
30 Endress+Hauser
Liquisys M CLM223/253 Commissioning
å 21.
The structure of this coding is given in
C121
The first column indicates the function
C C1 C111 group as a letter (see group designations).
The functions in the individual groups are
C07-CLD132xx-13-06-00-xx-005.eps
counted from the top to the bottom and from
Fig. 21: Function coding
the left to the right.
Factory settings
The first time it is switched on, the device has the factory setting for all functions. The table
below provides an overview of the most important settings.
All other factory settings can be found in the description of the individual function groups in
the "System configuration" section (the factory setting is highlighted in bold).
Measuring range 0 μS/cm to 2000 mS/cm (no measuring ranges for setting).
The setting is flowing and is guided by the connected sensors.
Endress+Hauser 31
Commissioning Liquisys M CLM223/253
Language
NEL = Dutch
ESP = Spanish
cond = conductive
9. In A1, select the desired mode of operation, e.g.
"cond" = conductive.
ind = inductive
MOhm = resistivity
cond A1
Press F to confirm.
conc = concentration
Oper.Mode
%
ppm
10. In A2, press F to confirm the factory settings.
(if A1 = conc, else step 12)
mg/l
TDS = Total Dissolved
% A2
Solids
none
Conc.Unit
XX.xx
11. In A3, press F to confirm the factory settings.
X.xxx
XXX.x
XX.xx A3
XXXX
Format
13. In A5, enter the cell constant of the connected cond: 1.000 cm-1
1/cm
sensor.
Refer to the sensor’s quality certificate for the
ind: 1.98 cm-1
MOhm: 0.01 cm-1
1.000 A5
exact value. 0.0025 to 99.99 cm-1
Cellconst
32 Endress+Hauser
Liquisys M CLM223/253 Commissioning
0
@
14. In A6, enter the resistance of the cable
(conductive sensors only). 0 to 99.99
0 A6
Cable-Res
Pt 100
18. In B1, select your temperature sensor.
Press F to confirm.
Pt 1k = Pt 1000
NTC30
Pt100 B1
fixed
ProcTemp.
none
lin = linear
19. In B2, select the appropriate temperature NaCl = common salt
compensation for your process, e.g. "lin" = linear. (IEC 60746)
Press F to confirm your selection. Pure = ultra pure lin B2
For detailed information on temperature water NaCl
compensation, see chapter "Setup 2" PureH = ultra pure TempComp.
water HCl
Tab = table
Press F to confirm.
2.1 %/K
0.0 ... 20.0 %/K
2.10 B3
Alpha val
Endress+Hauser 33
Commissioning Liquisys M CLM223/253
Sel.Out
For inductive sensors you must perform an airset before installation of the sensor, refer
to the chapter "Calibration"
34 Endress+Hauser
Liquisys M CLM223/253 Commissioning
Function group
A A Basic settings.
SETUP 1
SETUP 1
Display varies depending on instrument
version:
cond = conductive
– cond/resistivity/conc
A1
Select operating
mode
ind = inductive
MOhm = resistivity
cond A1 – ind/conc
Oper.Mode
conc = concentration Any change in operating mode
causes an automatic reset of user
settings.
%
Select concentration
ppm
A2
unit to be displayed
(only with Plus
mg/l % A2 A2 only active if A1 = conc.
TDS = Total Dissolved Solids
Package)
none Conc.Unit
Endress+Hauser 35
Commissioning Liquisys M CLM223/253
The temperature coefficient depends on the chemical composition of the medium and the
temperature itself.
In order to compensate for this dependence, four different compensation types can be
selected in the transmitter:
• Linear temperature compensation
• NaCl compensation
• Ultrapure water compensation NaCl (neutral compensation)
• Ultrapure water compensation HCl (acid compensation)
• Temperature compensation with table
T
a0008945
NaCl compensation
2.7
The NaCl compensation (according to IEC
[%/K]
2.3
2.1
0 48 96 144
T [°C]
a0008939
36 Endress+Hauser
Liquisys M CLM223/253 Commissioning
in relation to temperature. When using the alpha table function for temperature
In case of transmitter with Plus package you can enter a table with temperature coefficients
A B
1
2
(T0)
1
2
T1 T0 T2 T T1 T2 T
a0008944
Calculated values
A Required data
B
Use the following formula to calculate the values for the temperatures occurring in your
process:
Endress+Hauser 37
Commissioning Liquisys M CLM223/253
In the SETUP 2 function group, you can change the settings for temperature measurement.
Basic version does not include functions in italic.
Function group
B B Settings for temperature measurement.
SETUP 2
SETUP 2
If set to "fixed":
Pt100
Manual temperature compensation (MTC),
B1
Select temperature
sensor
Pt1k = Pt 1000
NTC30
Pt100 B1 no temperature measurement if fixed
temperature value is specified in B4. No
fixed
ProcTemp. temperature output if "fixed"!
none
lin = linear
This option is not displayed for
NaCl = common salt
B2
Select temperature
compensation type
(IEC 60746) lin B2 concentration measurement.
"Pure" and "PureH" are only available for
Pure = ultrapure water NaCl
PureH = ultrapure water HCl TempComp. conductive devices.
Tab = table
°C
B4
Enter process
temperature
25 °C
-35.0 to 250.0 °C
25.0 B4 Only if B1 = fixed.
This value can only be specified in °C.
ProcTemp.
°C
B7
Enter reference
temperature
25 °C
-5.0 to 100 °C
25.0 B7
ref temp.
38 Endress+Hauser
Liquisys M CLM223/253 Commissioning
1
A
1
B
Z2 Z3
0
C
0
E
D F
a0008923
Fig. 25: Alarm signalling and dosing switch-off by the main stream
A Flow in main stream F Flow restoration
B Relay contacts of PID controller Z2 Delay for controller switch-off, see field Z2
C Alarm relay Z3 Delay for controller switch-on, see field Z3
D Flow below switch-off limit Z4 or flow failure 0 Off
E Flow alarm 1 On
Endress+Hauser 39
Commissioning Liquisys M CLM223/253
1.5
0.5
0 20 40 Z7 60 80 100 X
a0008942
40 Endress+Hauser
Liquisys M CLM223/253 Commissioning
CURRENT INPUT Z
Z Current input settings.
function group
CUR.INPUT
0 to 100% corresponds to 4 to 20 mA at
Enter the switch-off %
Z4 limit value for the
50% 50 Z4 the current input. Observe measured value
0 to 100% assignment to the current output of the
current input
A.Thresh flowmeter.
Endress+Hauser 41
Commissioning Liquisys M CLM223/253
[mS/cm]
2000
1500
1000
Signal
500
0 4 8 12 16 20
I [mA]
I [mA]
a0008943
42 Endress+Hauser
Liquisys M CLM223/253 Commissioning
1 500 4
2 1000 16 41.66
3 2000 20 250
First enter the desired current output configuration into the following blank table with a
pencil. Calculate the resulting signal distance per mA to observe the necessary minimum
slope. Then enter the values in the device.
Endress+Hauser 43
Commissioning Liquisys M CLM223/253
Sel.Range
cond/ind: 0.00 μS/cm Here you can enter the measured value at
0/4 mA value: µS/cm
O312 Enter corresponding
MOhm: 0.00 kcm
Conc: 0.00 %
0.00 O312
which the min. current value (0/4 mA) is
applied at the transmitter output.
measured value
Temp: 0.00 °C
0/4 mA (Spreading: see Technical data.)
cond/ind: 2000 mS/cm Here you can enter the measured value at
20 mA value: mS/cm
O313 Enter corresponding
MOhm: 500 kcm
Conc: 99.99 %
2000 O313
which the max. current value (20 mA) is
applied at the transmitter output.
measured value
Temp: 150 °C
20 mA (Spreading: see Technical data.)
44 Endress+Hauser
Liquisys M CLM223/253 Commissioning
mA value entered.
Back to O3.
Message as to
O336 whether table status
yes
no
yes O336 If status = no, correct table (all settings
made up until now are retained) or back
is OK
Status ok to measuring mode (table is deleted).
Endress+Hauser 45
Commissioning Liquisys M CLM223/253
6.4.5 Alarm
In the ALARM function group, you can define various alarms and configure output contacts.
Each individual error can be defined to be effective or not (at the contact or as an error
current).
ALARM function
F F Alarm function settings.
group
ALARM
Cont.Type
s
F3 Enter alarm delay
0 s (min)
0 to 2000 s (min)
0 F3 Depending on the option selected in F2,
the alarm delay is entered in s or min.
Err.Delay
Err.Curr
"2.4 mA" may not be used.
46 Endress+Hauser
Liquisys M CLM223/253 Commissioning
F8
Automatic cleaning
function start
no
yes
no F8 This field is not available for certain errors,
see "Trouble-shooting and fault
elimination" section.
CleanTrig
6.4.6 Check
The CHECK function group is only available for devices with a Plus Package.
In the CHECK function group, you can select different monitoring functions for the
measurement:
t
a0006744
Any PCS alarm pending is automatically deleted as soon as the sensor signal changes.
Endress+Hauser 47
Commissioning Liquisys M CLM223/253
CHECK function
P P Settings for sensor and process monitoring
group
CHECK
Off
Alarm signalling optionally with or
Low
without simultaneous controller switch-
High
P2
Set PCS Alarm
(live check)
Lo+Hi Off P2 off.
XXXX = without controller switch-off
Low!
High! SCS Ref XXXX! = with controller switch-off
(error no.: E154, E155)
Lo+Hi!
µS/cm
P4
Enter lower alarm
threshold
0 μS/cm
0 to 9999mS/cm
0.00 P4
LowAlarm
mS/cm
P5
Enter upper alarm
threshold
9999 μS/cm
0 to 9999mS/cm
9999 P5
HighAlarm
Off AC = sensor alternation check (E152)
AC CC = controller check (E156, E157)
Select process CC Alarm signalling optionally with or
P6 monitoring AC+CC Off P6 without simultaneous controller switch-
(PCS alarm) AC! off.
CC! ProcMonit XXXX = without controller switch-off
AC+CC! XXXX! = with controller switch-off
Enter maximum
permissible duration min
P7 for lower CC setpoint
60 min
0 to 2000 min
60 P7 Only when P6 = CC or AC+CC
limit violation (field
P9) Tmax Low
Enter maximum
permissible duration min
P8 for upper CC setpoint
120 min
0 to 2000 min
120 P8 Only when P6 = CC or AC+CC
limit violation (field
P9) Tmax High
48 Endress+Hauser
Liquisys M CLM223/253 Commissioning
Endress+Hauser 49
Commissioning Liquisys M CLM223/253
5
6
1
4
2
A 3
7
t1 t2 t3 t4 t5 t6 t7 t
7
3
2
B 4
6
1 5
t1 t2 t3 t4 t5 t6 t7 t
a0005986
P(ID) controller
You can define various controller functions for the transmitter. On the basis of the PID
controller, P, PI, PD and PID controllers can be implemented. For an optimum control system,
use the controller that best suits your application. Depending on the option selected in the
R 237/R 266 field, the actuating signal can be output via relays or via current output 2 (if
available).
• P controller
Used for simple linear control purposes with small system deviations. Where major
changes are to be controlled, overshooting may occur. In addition, a lasting control
deviation is to be expected.
• PI controller
Is used for control systems where overshooting is to be avoided and no lasting control
deviation should occur.
• PD controller
Is used for processes that require quick changes and where peaks are to be corrected.
• PID controller
Is used for processes where a P, PI or PD controller does not control sufficiently.
Configuration options of the PID controller
The following configuration options are available for a PID controller:
• Change control gain Kp (P influence)
• Set integral action time Tn (I influence)
• Set derivative action time Tv (D influence)
Basic load dosing (Basic)
The basic load dosing (field R231) is used to set a constant dosage (field R2311)
PID controlling plus basic load dosing
If you select this function (PID + Basic) in field R231 the PID controlled dosage will not be
lower than the basic load value entered in field R2311.
50 Endress+Hauser
Liquisys M CLM223/253 Commissioning
A
100 %
100 % C
a0008940
Fig. 30: Control characteristic PID controller with basic load dosing
A PID with basic load
B Basic load
C PID
Commissioning
If you do not yet have any experience for setting the control parameters, set the values that
yield the greatest possible stability in the control circuit. Proceed as follows to optimize the
control circuit further:
• Increase the control gain Kp until the controlled variable just starts to overshoot.
• Reduce Kp slightly and then reduce the integral action time Tn so that the shortest possible
correction time without overshooting is achieved.
• To reduce the response time of the controller, also set the derivative action time Tv.
Control and fine optimization of the set parameters with a recorder
A A
1 2
B B
A A
3 4
B B
B
a0008924
Endress+Hauser 51
Commissioning Liquisys M CLM223/253
A t1 A
t2 t1
1 1
0 0
t [s] t [s]
T T1 T2
a0008928
Fig. 31: Signal of a pulse-length modulated controller contact (left) and of a pulse-frequency modulated controller contact (right)
A Contact 1 = on, 0 = off T Period length
B Time [s] t1 = ton t2 = toff T1 T2 Impulse period length (impulse freq. 1/T1 and 1/
T2)
Constant controller
Via the current output 2, the minimum actuating variable (0 %) of the controller is output
with 0/4 mA and the maximum actuating variable (100 %) of the controller is output with
20 mA.
Control characteristic for direct and inverse control action
You can choose between two control characteristics in the R236 field:
• Direct control action = maximum function
• Inverse control action = minimum function
Y
100 %
– XW XW
Y
100 %
– XW XW
a0008941
Fig. 32: Control characteristic of a proportional controller with direct and inverse control action
A Direct = max. function
B Inverse = min. function
52 Endress+Hauser
Liquisys M CLM223/253 Commissioning
Timer and Chemoclean do not work independently of one another. While one of the
two functions is active, the other cannot be started.
1
A
0
t
1
B
0
t
t0 t1 t2 t3 t4
a0006794
Fig. 33: Correlation of cleaning time, pause time and hold dwell period
A Wiper and/or spray cleaning system t0 Normal mode
B Hold function t1 Cleaning start
0 Inactive t2-t1 Cleaning time
1 Active t3-t2 Clean hold dwell period (0 to 999 s)
t4-t3 Pause time between two cleaning intervals
(1 to 7200 min)
Chemoclean function
Just like the timer function, Chemoclean can also be used to start a cleaning cycle. However,
Chemoclean also gives you the added option of defining different cleaning and rinsing
intervals.
As a result, it is possible to clean irregularly with different repeat cycles and to separately set
the cleaning times with post rinse times.
Pay attention to the following:
• To use the Chemoclean function the transmitter has to be equipped with a designated
relay board (see product structure or chapter "accessories").
• Timer and Chemoclean do not work independently of one another. While one of the two
functions is active, the other cannot be started.
• For the Chemoclean function, the relays 3 (water) and 4 (cleaner) are used.
• If cleaning is prematurely aborted, a post rinse time always follows.
• If the setting is "Economy", cleaning only takes place with water.
Endress+Hauser 53
Commissioning Liquisys M CLM223/253
1
A
0
1
B
0
1
C
0
t0 t1 t2 t3 t4 t5
a0006793
Limit values for pharmaceutical water according to United States Pharmacopeia (USP)
and European Pharmacopoeia (EP) (conductive only)
The transmitter (conductive sensors only) can monitor "Water for Injection" (WFI), "Highly
Purified Water" (HPW) and "Purified Water" (PW) according to the standard United States
Pharmacopeia (USP) part 645 and European Pharmacopoeia (EP).
USP function: For "Water for Injection" (WFI) according to the standard of USP and EP and
for "Highly Purified Water" (HPW) according to the standard of EP the transmitter uses the
temperature depending limit values of the following table:
50/122 1.9
54 Endress+Hauser
Liquisys M CLM223/253 Commissioning
EP-PW function: For "Purified Water" (PW) according to the standard of EP the transmitter
uses the temperature depending limit values of the following table:
50/122 7.1
Endress+Hauser 55
Commissioning Liquisys M CLM223/253
RELAY function
R Relay contact settings.
R group
RELAIS
LC PV = limit contactor
PV = process value
Configuration limit cond (1)
If Rel4 is selected in the R1 field,
contactor for LC °C = limit contactor T (2)
Clean = Chemoclean cannot be selected.
R2 (1)
conductivity,
resistivity or
PID controller (3)
Timer (4)
LC PV R2 By confirming with ENTER, another relay
function already switched on is switched
concentration Clean = Chemoclean (5)
measurement USP (6) Sel.Type off and its settings are reset to the factory
settings.
EP PW (7)
R211
Switch function of
R2 (1) off or on
Off
On
off R211 All the settings are retained.
Function
s
R214 Enter pick-up delay
0s
0 to 2000 s
0 R214
On Delay
s
R215 Enter drop-out delay
0s
0 to 2000 s
0 R215
Off Delay
If the alarm threshold is undershot/
overshot, this triggers an alarm with the
cond/ind: 9999 mS/cm mS/cm error message (E067 to E070) and error
R216
Enter alarm
threshold
MOhm: 200 Mcm 9999 R216 current at the transmitter (note alarm
conc: 9999 % delay in field F3).
A.Thresh If defined as a Min contact, the alarm
threshold must be < switch-off point.
56 Endress+Hauser
Liquisys M CLM223/253 Commissioning
R217
Display status for
limit contactor
MAX
MIN
MAX R217 Display only.
LC State
LC PV = limit contactor cond
(1)
LC °C = limit contactor T
Configure limit By confirming with ENTER, another relay
(2)
R2 (2)
contactor for
temperature
PID controller (3) LC °C R2 function already switched on is switched
off and its settings are reset to the factory
Timer (4)
measurement
Clean = Chemoclean (5) Sel.Type settings.
USP (6)
EP PW (7)
R221
Switch function of
R2 (2) off or on
Off
On
off R221
Function
°C
R222
Enter switch-on
temperature
250.0 C
-35.0 to 250.0 C
250.0 R222 Never set the switch-on point and the
switch-off point to the same value!
On value
Entering a switch-off point selects either a
Max contact (switch-off point < switch-on
°C point) or a Min contact (switch-off point >
R223
Enter switch-off
temperature
250.0 C
-35.0 to 250.0 C
250.0 R223 switch-on point), thereby implementing a
hysteresis that is constantly required (see
Off value "Illustration of the alarm and limit
functions" figure).
0 R224
0s s
R224 Enter pick-up delay
0 to 2000 s
On Delay
0 R225
0s s
R225 Enter drop-out delay
0 to 2000 s
Off Delay
If the alarm threshold is undershot/
overshot, this triggers an alarm with the
Enter alarm °C error message (E067 to E070) and error
R226 threshold (as
250.0 C
-35.0 to 250.0 C
250.0 R226 current at the transmitter (note alarm
absolute value) delay in field F3).
A.Thresh If defined as a Min contact, the alarm
threshold must be < switch-off point.
R227
Display status for
limit contactor
MAX
MIN
MAX R227 Display only.
LC State
Endress+Hauser 57
Commissioning Liquisys M CLM223/253
R2 (3)
Configure P(ID)
controller
PID controller (3)
Timer (4)
PID R2 function already switched on is switched
off and its settings are reset to the factory
Clean = Chemoclean (5) settings.
USP (6) Sel.Type
EP PW (7)
Off
On = PID controller
R231
Switch function of
R2 (3) off or on
On
Basic
off R231 Basic = basic load dosing
PID+B = PID controller + basic load dosing
PID+B
Function
Kp
Enter derivative
min
R235
action time Tv
(0.0 = no D-
0.0 min
0.0 to 999.9 min
0.0 R235 See "P(ID) controller" section.
component)
Time Tv
58 Endress+Hauser
Liquisys M CLM223/253 Commissioning
s
R2310
Enter minimum
switch-on time tON
0.3 s
0.1 to 5.0 s
0.3 R2310 This field only appears if pulse length is
selected in R237.
Min.PTime
When you select the basic load, you enter
the desired dosing quantity.
%
R2311 Enter basic load
0%
0 to 40 %
0 R2311 100% basic load would correspond to:
– Constantly on for R237 = len
BasicLoad – Fmax at R237 = freq (field R239)
– 20 mA at R237 = curr
R241
Switch function of
R2 (4) off or on
Off
On
off R241
Function
s
R242
Enter rinsing/
cleaning time
30 s
0 to 999 s
30 R242 Settings for Hold and relay are active for
this time.
RinseTime
Endress+Hauser 59
Commissioning Liquisys M CLM223/253
R251
Switch function of R2 Off
(5) off or on On
off R251
Function
Int = internal (time-
controlled) The cycle for the "int" function is started by
Ext = external (digital input the end of the pause time (R257).
R252
Select type of start
pulse
2) int R252 No real time clock is available.
I+ext = internal + external External suppression is required for
I+stp = internal, suppressed CleanTrig irregular time intervals (e.g. weekends).
by external
s
R253 Enter pre-rinse time
20 s
0 to 999 s
20 R253 Rinsing with water takes place.
PreRinse
s
R254 Enter cleaning time
10 s
0 to 999 s
10 R254 Cleaning with cleaning agent and water
takes place.
CleanTime
s
R255 Enter post rinse time
20 s
0 to 999 s
20 R255 Rinsing with water takes place.
PostRinse
R256
Enter number of
repeat cycles
0
0 to 5
0 R256 R253 to R255 is repeated.
Rep.Rate
Enter number of
After cleaning with cleaner, up to 9
cleaning cycles
R259 without cleaning
0
0 to 9
0 R259 cleaning sessions can be carried out with
water only until the next cleaning session
agent (economy
function) EconomyCl with cleaner takes place.
60 Endress+Hauser
Liquisys M CLM223/253 Commissioning
R261
Switch function of R2 Off
(6) off or on On
off R261
Function
The pre-alarm effects a contact query.
Should the alarm value be reached (100
Pre-alarm threshold: %
R262 Enter switch-on
80 %
0.0 to 100.0 %
80.0 R262 %) the alarm relay also responds.
Example: At 15 °C and 1.0 μS/cm with the
point
On Value setting 80.0 % an USP-pre-alarm is also
triggered at 0.8 μS/cm.
s
R264
Pre-alarm threshold: 0 s
Enter pick-up delay 0 to 2000 s
0 R264
On Delay
s
R265
Pre-alarm threshold: 0 s
Enter drop-out delay 0 to 2000 s
0 R265
Off Delay
The EP PWcontact can be configured as a
LC PV = limit contactor cond pre-alarm, i.e., it issues an alarm before
(1) the limit is reached.
LC °C = limit contactor T (2) When an alarm is output, the error no.
R2 (7)
Configure EP PW
contact
PID controller (3)
Timer (4)
EP PW R2 E151 is displayed.
By confirming with ENTER, another relay
Clean = Chemoclean (5)
USP (6)
Sel.Type function already switched on is switched
off and its settings are reset to the factory
EP PW (7) settings.
PW acc. to EP
R271
Switch function of R2 Off
(7) off or on On
Off R271
Function
The pre-alarm effects a contact query.
Should the alarm value be reached (100
Pre-alarm threshold: %
R272 Enter switch-on
80 %
0.0 to 100.0 %
80.0 R272 %) the alarm relay also responds.
Example: At 15 °C and 1.0 μS/cm with the
point
On Value setting 80.0 % an EP PW-pre-alarm is
also triggered at 0.8 μS/cm.
s
R274
Pre-alarm threshold: 0 s
Enter pick-up delay 0 to 2000 s
0 R274
On Delay
Endress+Hauser 61
Commissioning Liquisys M CLM223/253
s
R275
Pre-alarm threshold: 0 s
Enter drop-out delay 0 to 2000 s
0 R275
Off Delay
Function group
T T Settings for temperature compensation.
ALPHA TABLE
ALPHA TAB
%/K
T5
coefficient
Enter temperature 2.10 %/K
0.00 to 20.00 %/K
2.10 T5
alpha val
Only display
Message, whether or If status = "no", then set table correctly (all
T6 not the table status is
yes
no
yes T6 previous settings are kept) or back to
ok measurement mode (this makes the table
Status ok invalid)
62 Endress+Hauser
Liquisys M CLM223/253 Commissioning
2
T3 2
1 T3 1
T2
T2
T1
T1
Fig. 35: Measured data for variable process temperatures (example)
Conductivity 1 Measuring point
C Concentration 2 Measuring range
T Temperature
Endress+Hauser 63
Commissioning Liquisys M CLM223/253
1 T 1 T
Fig. 36: Measured data for a constant process temperature (example)
Conductivity T Constant temperature
C Concentration 1 Measuring range
The characteristics received from the
measuring points must be strictly
monotonously increasing or strictly
monotonously decreasing in the range of the
process conditions. Therefore, neither
maxima / minima nor ranges with a
C C
constant behaviour can occur. Curve profiles a0008927
such as those in Fig. 37 are not permitted. Fig. 37: Impermissible curve profiles
Conductivity
C Concentration
Value entry
Enter the three characteristic values for each measured sample in the fields K6 to K8 (value
triplets of uncompensated conductivity, temperature and concentration).
• Variable process temperature:
Enter at least four value triplets.
• Constant process temperature:
Enter at least two value triplets.
• Please make sure that the concentrations and temperatures measured for your samples
correspond to the measuring range of the process. If the measured values of the process
are outside the range of your sample values, this considerably reduces the level of accuracy
and the error message E078 or E079 will be displayed.
If you enter an additional value triplet of 0 μS/cm and 0 % for each temperature used, you
can work from the start of measuring range with sufficient accuracy and without an error
message.
Enter the values in the order of increasing concentration (see the following example).
mS/cm % °C
240 96 60
380 96 90
220 97 60
340 97 90
120 99 60
200 99 90
• The setups for temperature compensation in the menu Setup 2 (fields B2 and B3) are
inoperative for the concentration measurement. The temperature is already processed by
the concentration tables.
64 Endress+Hauser
Liquisys M CLM223/253 Commissioning
Selection of
concentration curve, The curves are independent of each other.
K1 to be used to
1
1 to 4
1 K1 Therefore, four different curves can be
calculate the display defined.
value act.curve
When editing a curve, another curve
should be used to calculate the
K2
Selection of table to
be edited
1
1 to 4
1 K2 corresponding values.
Example: When editing curve 2, either
editCurve curve 1, curve 3 or curve 4 should be active
(see K1).
K4
Enter number of
reference triplets
1
1 to 10
1 K4 Each triplet consists of three numeric
values.
No. Elem.
1
K5 Select triplet 1 to number of triplets in K4 1 K5 Any triplet can be edited.
If "assign", go to K9.
assign
Sel.Elem.
%
K7
Enter concentration
value for K6
0.00 %
0.00 to 99.99 %
0.00 K7 Measuring unit selected as in A2.
Format selected as in A3.
concentr.
°C
K8
Enter temperature
value for K6
0.0 °C
-35.0 to 250.0 °C
0.0 K8
Temp.val.
Only display
Message whether or
K9 not the table status is
yes
no
yes K9 If not, then set table correctly (all previous
settings are kept) or back to measurement
ok
Status ok mode (this makes the table invalid).
Endress+Hauser 65
Commissioning Liquisys M CLM223/253
6.4.10 Service
SERVICE function
S Service function settings.
S group
SERVICE
ENG = English
GER = German
This field has to be configured once during
S1 Select language
FRA = French
ITA = Italian
ENG S1 device configuration. Then you can exit S1
and continue.
Language
NL = Dutch
ESP = Spanish
S3 Manual Hold
Off
On
off S3
The setting is retained even in the event of
a power failure.
Man. HOLD
s
S4
Enter Hold dwell
period
10 s
0 to 999 s
10 S4
Cont.Time
The code is located on the nameplate.
If an incorrect code is entered, you are
Enter SW upgrade taken back to the measurement menu.
S5 release code (Plus
0000
0000 to 9999
0000 S5 The number is edited with the PLUS or
Package) MINUS key and confirmed with the ENTER
PlusCode key.
"1" is displayed if the code is active.
The code is located on the nameplate.
If an incorrect code is entered, you are
Enter SW upgrade taken back to the measurement menu.
S6 release code
0000
0000 to 9999
0000 S6 The number is edited with the PLUS or
Chemoclean MINUS key and confirmed with the ENTER
CleanCode key.
"1" is displayed if the code is active.
S7
Order number is
displayed
order S7
If the device is upgraded, the order code is
not automatically adjusted.
PR0005
S8
Serial number is
displayed
SerNo S8
12345678
66 Endress+Hauser
Liquisys M CLM223/253 Commissioning
Test
E+H SERVICE
E E Information on the device version
function group
E+H SERV
E111
E121 If E1 = contr: instrument software
E131
Software version is
displayed
xx.xx E111 If E1 = trans, main, rel: module firmware
E141 If E1 = sens: sensor software
E151 SW-Vers.
E112
E122
E132
Hardware version is
displayed
xx.xx E112 Only display function
E142
E152 HW-Vers.
E113
E123
E133
Serial number is
displayed
SerNo E113 Only display function
E143
E153 12345678
E114
E124
E134
Module ID is
displayed
LSG E114 Only display function
E144
E154 Modul-ID
Endress+Hauser 67
Commissioning Liquisys M CLM223/253
6.4.12 Interfaces
I2
Display of measuring
point
Tag I2
@@@@@@@@
6.5 Communication
For devices with a communication interface, please also refer to the separate Operating
Instructions BA00208C/07/EN (HART) or BA00209C/07/EN (PROFIBUS).
68 Endress+Hauser
Liquisys M CLM223/253 Commissioning
6.6 Calibration
To access the "Calibration" function group, press the CAL key.
This function group is used to calibrate and adjust the transmitter. Two different types of
calibration are possible:
• Calibration by measurement in a calibration solution of a known conductivity.
• Calibration by entering the exact cell constant of the conductivity sensor.
Pay attention to the following:
• At first start-up of inductive sensors, an airset is absolutely required in order for the
measuring system to be able to generate accurate measuring values.
• If the calibration procedure is aborted by pressing the PLUS and MINUS keys at the same
time (return to C114, C126 or C136) or if the calibration is faulty, then the previous
calibration data are reinstated. A calibration error is indicated by the “ERR” message and
flashing of the sensor symbol on the display.
Repeat calibration!
• The instrument is automatically switched to hold during calibration (factory setting).
• After calibration, the system jumps back to the measuring mode. During the hold dwell
period the hold symbol is displayed.
• For conductive sensors only the fields C121 to C126 are relevant.
Basic version does not include functions in italic.
Function group
C C Calibration settings.
CALIBRATION
CALIBRAT
µS/cm
C112
Residual coupling is
displayed (airset)
-80.0 to 80.0 μS/cm 5.3 C112 Residual coupling of measuring system
(sensor and transmitter).
AirSetVal
Endress+Hauser 69
Commissioning Liquisys M CLM223/253
°C
C121
Enter calibration
temperature (MTC)
25 °C
-35.0 to 250.0 °C
25.0 C121 Only exists if B1 = fixed.
ProcTemp.
This value is specified in the Technical
Information of all E+H calibration
Enter value of
2.10 C122
2.10 %/K %/K solutions. You can also use the printed-on
C122
calibration solution 0.00 to 20.00 %/K table to calculate the value.
Set to 0 for calibration with
alpha val uncompensated values.
1/cm
C124
Calculated cell
constant is displayed
0.0025 to 99.99 cm-1 1.000 C124 The calculated cell constant is displayed
and entered in A5.
Cellconst
70 Endress+Hauser
Liquisys M CLM223/253 Commissioning
°C
C131
Enter process
temperature (MTC)
25 °C
-35.0 to 250.0 °C
25.0 C131 Only exists if B1 = fixed.
MTC temp.
Enter correct
mS/cm
C133
conductivity value of
the calibration
current measured value
0.0 μS/cm to 9999 mS/cm
10.30 C133 You should select a value close to the
application range.
solution
Real val.
On inductive sensors, the distance from
the sensor to the wall of the pipe and the
Calculated material of the pipe (conductive or non-
C134 installation factor is
1
0.10 to 5.00
1 C134 conductive) influence the measured
displayed value. The installation factor shows this
InstFact influence. See the technical information
of the installed sensor.
Endress+Hauser 71
Diagnostics and troubleshooting Liquisys M CLM223/253
Error Display Tests and/or remedial measures Alarm contact Error current Autom. PROFIBUS
no. cleaning start status
E001 EEPROM memory error 1. Switch device off and then on again. Yes No — —2) 0C 0C
2. Load device software compatible
with the hardware (with optoscope,
Instrument not calibrated,
see "Optoscope service tool" section).
calibration data invalid, no
user data, user data 3. Load measurement-parameter
E002 invalid (EEPROM error), specific device software. Yes No — —2 0C 0C
instrument software not 4. If the error persists, send in the
suitable to hardware device for repair to your local
(controller) service organisation or replace the
device.
Limit for Airset offset Repeat Airset (in air) or replace sensor.
E025 No No No 80 80
exceeded Dry sensor.
Calibration range of
E036 Clean sensor and recalibrate; if No No No 80 80
sensor exceeded
necessary, check sensor and
Below calibration range of connections.
E037 No No No 80 80
sensor
72 Endress+Hauser
Liquisys M CLM223/253 Diagnostics and troubleshooting
Error Display Tests and/or remedial measures Alarm contact Error current Autom. PROFIBUS
no. cleaning start status
Calibration range of
E049 installation factor Check pipe diameter, clean sensor and No No — —2 80 80
exceeded recalibrate.
Below calibration range of
E050 No No — —2 80 80
installation factor
Main parameter
E057 Yes No No 44 80
measuring range exceeded
Check measurement and connections;
Below temperature
E059 check device and measuring cable with Yes No No 80 44
measuring range
simulator if necessary.
Temperature measuring
E061 Yes No No 80 44
range exceeded
Temperature outside
E077 Yes No No 44 80
value table range
Clean sensor; check table.
Temperature outside
E078 Yes No No 44 80
concentration table
Conductivity outside
E079 Yes No No 44 80
concentration table
Endress+Hauser 73
Diagnostics and troubleshooting Liquisys M CLM223/253
Error Display Tests and/or remedial measures Alarm contact Error current Autom. PROFIBUS
no. cleaning start status
1) PV = Process variable
2) If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error).
3) Current error messages not applicable via PROFIBUS
74 Endress+Hauser
Liquisys M CLM223/253 Diagnostics and troubleshooting
Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Incorrect temperature measurement Check temperature value on instrument Temperature measuring instrument,
and reference unit. precision thermometer
Incorrect temperature compensation Check compensation method (none / Please note: transmitter has separate
ATC / MTC) and compensation type calibration and operating temperature
(linear/substance/user table). coefficients.
Display deviates from
reference measurement Reference instrument calibration Calibrate reference instrument or use Calibration solution, operating
faulty calibrated instrument. instructions of reference instrument
Incorrect ATC setting on reference Compensation method and Operating instructions of reference
instrument compensation type must be identical on instrument
both instruments.
Incorrect line resistance in field A6 Enter correct value CYK71: 165 /km
Short circuit / moisture in sensor Check sensor. See chapter "Checking inductive
conductivity sensors".
Short circuit in cable or junction box Check cable and junction box. See chapter "Checking extension cable
and junction box".
Transmitter module defective Test with new module. See chapter "Spare parts".
Impermissible instrument operating Switch instrument off and back on. EMC problem: check grounding and line
state (no response to key actuation) routing if problem persists or call E+H
Service to test.
Incorrect sensor connection Verify connections using connection Connection diagram in chapter "Electrical
diagram; three-wire connection connection"
mandatory.
Incorrect temperature
Measuring cable defective Check cable for interruption/short Ohmmeter
value
circuit/ shunt.
Incorrect temperature sensor type Select temperature sensor type on
instrument (field B1).
Endress+Hauser 75
Diagnostics and troubleshooting Liquisys M CLM223/253
Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Polarization effects (only with Use suitable sensor Measuring range tables e.g. in FA
Incorrect conductivity conductive sensors) "Conductivity" or technical data of
• Use larger cell constant
measured value in process conductivity sensors
• Use graphite instead of stainless steel
(check resistance)
Flow rate too high (may cause Reduce flow or choose low turbulence
bubbles) mounting position.
Interference current in medium (only Ground medium close to sensor. Most frequent cause of currents in
when conductive) medium: defective submerged motors
Sensor soiled or coated Clean sensor (see chapter "Cleaning Heavily soiled media: use spray cleaning.
conductivity sensors").
Incorrect line resistance in field A6 Enter correct value. CYK71: 165 /km
Measuring cable interferences Connect cable screen according to See chapter "Electrical connection".
connection diagram.
Signal output line interferences Check line routing, try separate line Separate routing of signal output and
routing. measuring input lines
Measured value
fluctuates Interference currents in medium Eliminate source of interference or
ground medium close to sensor.
With conductive sensors: Use shielded cables and ground the cable
Electromagnetic interferences on shield
signal lines
Controller or timer cannot No relay module available Install module LSR1-2 or LSR1-4.
be activated
Controller in "Manual/On" mode Set controller to "Manual/Off" or "Auto". Keyboard, REL and AUTO keys
Controller/limit contact Dropout delay setting too long Shorten dropout delay. See field R2xx.
works continuously
Control loop interruption Check measured value, current output,
actuators, chemical supply.
Line open or short-circuited Disconnect line and measure directly on mA meter 0–20 mA
No conductivity current instrument.
output signal
Output defective See chapter "Instrument specific errors".
76 Endress+Hauser
Liquisys M CLM223/253 Diagnostics and troubleshooting
Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Current output table not Value interval too small Select practical intervals
accepted
Instrument does not have 2nd Refer to nameplate for variant; change Module LSCH-x2,
No temperature output current output LSCH-x1 module if necessary. see chapter "Spare parts".
signal
Instrument with PROFIBUS PA PA instrument has no current output!
Plus package functions Plus package not enabled (enable – When upgrading instrument with For a detailed description, see chapter
not available with code that depends on serial Plus package: code received from E+H "Replacement of central module".
(Live Check, current curve number and is received from E+H enter.
2 to 4, alpha value curve with order of extension package) – After replacing defective LSCH/LSCP
2 to 4, user conc. curve module: first enter instrument serial
1 to 4) number (s. nameplate) manually,
then enter code.
Line interferences
Several devices set to same address Set addresses correctly. Communication not possible with several
devices set to same address.
Endress+Hauser 77
Diagnostics and troubleshooting Liquisys M CLM223/253
Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
No or incorrect DD/DLL
Incorrect baud rate setting for For further information, see BA00209C/
segment coupler in DPV-1 server 07/EN "PROFIBUS PA/DP - Field
No PROFIBUS®
Incorrect station (master) addressed communication with Liquisys M
communication
or duplicate address CxM223/253".
Line problems
(too long, cross section too small; not
shielded, screen not grounded, wires
not twisted)
78 Endress+Hauser
Liquisys M CLM223/253 Diagnostics and troubleshooting
Error Possible cause Tests and/or remedial measures Execution, tools, spare parts
Device cannot be
Press CAL and MINUS keys
operated, display value Operation locked See "Function of keys" section.
simultaneously.
9999
No line voltage Check whether line voltage is present. Electrical technician/e.g. multimeter
User
Compare actual line voltage and
Supply voltage wrong/too low (data for energy supply company or
nameplate data.
multimeter)
Display is on but Device or module in device not CLM223: reinstall insert. Perform with the aid of the installation
correctly mounted CLM253: remount display module. drawings in the "Spare parts" section.
– No change in display
and/or Poss. EMC problem: if this persists, check
– Device cannot be Operating system in unpermitted
Switch device off and then on again. the installation or have it checked by
mode
operated Endress+Hauser Service.
Compare line voltage and nameplate
Voltage wrong/too high User, electrical technician
data.
Device gets hot
Diagnosis only by Endress+Hauser
Power unit defective Replace power unit.
Service
Endress+Hauser 79
Diagnostics and troubleshooting Liquisys M CLM223/253
Error Possible cause Tests and/or remedial measures Execution, tools, spare parts
No function of additional CLM253: ribbon cable item 320 Check ribbon cable seating, renew cable
See "Spare parts" section.
relay loose or defective if required.
Only 2 additional relays Relay module LSR1-2 installed with
Upgrade to LSR1-4 with 4 relays. User or Endress+Hauser Service
can be triggered 2 relays
Additional functions (Plus Replacement modules LSCH or LSCP In the case of LSCH/LSCP with SNR
package and/or have the device serial number 0000 0000, a device serial number can be
For a detailed description, see "Replacing
Chemoclean) missing when they leave the factory. The entered once in fields E114 to E116.
central module" section.
after LSCH/LSCP module Plus package or Chemoclean are not Then enter the release code for the Plus
replaced enabled on leaving the factory. package and/or Chemoclean.
80 Endress+Hauser
Liquisys M CLM223/253 Maintenance
8 Maintenance
Take all the necessary measures in time to guarantee the operational safety and reliability
of the entire measuring system.
Maintenance work at the transmitter comprises:
• Calibration (see "Calibration" section)
• Cleaning of assembly and sensor
• Cable and connection check
When performing any work on the device, bear in mind any potential impact this may have
on the process control system or on the process itself.
NOTICE
Electrostatic discharge (ESD)
Risk of damage to electronic components
‣ Take personal protective measures to avoid ESD, such as discharging beforehand at PE or
permanent grounding with a wrist strap.
‣ For your own safety, use only genuine spare parts. With genuine spare parts, the
function, accuracy and reliability are also guaranteed after repair.
Endress+Hauser 81
Maintenance Liquisys M CLM223/253
82 Endress+Hauser
Liquisys M CLM223/253 Maintenance
Conductivity
For conductivity, the values in the following table are valid, if the cell constant k is set to the
10
-1
1 cm 100 mS/cm
0.1 cm -1
1 mS/cm 1 k cm
100 1 cm -1
10 mS/cm
0.1 cm -1
0.1 mS/cm 10 k cm
1000 1 cm -1
1 mS/cm
10 cm-1 10 mS/cm
0.01 cm -1
1 μS/cm 1 M cm
0.1 cm -1
10 μS/cm 100 k cm
10 k
1 cm-1 100 μS/cm
-1
10 cm 1 mS/cm
0.01 cm -1
0.1 μS/cm 10 M cm
100 k 0.1 cm -1
1 μS/cm 1 M cm
1 cm-1 10 μS/cm
0.01 cm -1
0.01 μS/cm 100 M cm
1 M 0.1 cm -1
0.1 μS/cm 10 M cm
1 cm-1 1 μS/cm
0.01 cm -1
0.001 μS/cm
100 M cm
10 M
0.1 cm -1
0.01 μS/cm
The M measurement is normally used for pure and ultrapure water and therefore is
only wise for cell constants where k= 0.01 or k = 0.1 cm-1.three-line system.
Endress+Hauser 83
Maintenance Liquisys M CLM223/253
Conductivity simulation:
Pull a cable through the sensor opening and then connect, e.g. to a decade resistor.
84 Endress+Hauser
Liquisys M CLM223/253 Maintenance
CLS54: approx. 1 to 3
– Inductivity approx. 260 to 500 mH (at 2 kHz)
CLS50: approx. 250 to 450 mH
CLS52/54: approx. 180 to 550 mH
(Measure the white and red coaxial cables, between the inner conductor and screen in
both cases.)
• Testing the coil shunt
– A shunt between the two sensor coils is not allowed. The resistance measured should be
>20 M.
Test with ohmmeter between red coaxial cable and white coaxial cable.
• Testing the temperature sensor
Use the table in chapter "Simulation of inductive sensors for device test" to check the
Pt 100 / Pt 1000 in the sensor.
Measure between the green and white wires and between green and yellow. The
resistance values should be identical.
• Testing the temperature sensor shunt
– Shunts between the temperature sensor and the coils are not allowed. Check with
ohmmeter for >20 M.
Measure between the temperature sensor wires (green + white + yellow) and the coils (red
and white coaxial cables ).
Endress+Hauser 85
Maintenance Liquisys M CLM223/253
86 Endress+Hauser
Liquisys M CLM223/253 Repair
9 Repair
Endress+Hauser 87
Repair Liquisys M CLM223/253
420a
345
420b
340
400
240
15/20
40 ... 91
320 330
310
430
290/300
15
230V
B 115V
100V
a0008913
88 Endress+Hauser
Liquisys M CLM223/253 Repair
81 Central module ind. (controller) LSCP-PA PROFIBUS PA/no current output 51501221
91 Central module ind. (controller) LSCP-DP PROFIBUS DP/no current output 51502501
290 Kit CxM2x3 Relay module LSR2-DP Relay module + 2 relays 71134732
PROFIBUS DP Current input and terminals DP
valid of: hardware version 2.10
310, 320, Housing mechanical parts Rear plate, side panel, end frame, 51501076
340, 400 special screw
340 End frame Rear frame for PROFIBUS DP, with D- 51502513
PROFIBUS-DP submin plug connector
345 Grounding terminal strip PE and screening connections 51501086
420a, Terminal strip set Complete terminal strip set, standard 51501203
420b + HART
Endress+Hauser 89
Repair Liquisys M CLM223/253
Incorrect mounting is not possible. Modules inserted in the electronics box incorrectly
are not operable since the ribbon cables cannot be connected.
Make sure the cover seals are intact to guarantee IP 65 ingress protection.
90 Endress+Hauser
Liquisys M CLM223/253 Repair
A
B
Pos. 10
Netzteil 100 – 230 V AC
15
C
Sicherung
410
230 V
370
D 115 V
460b 100 V
40 ... 91
450b 330
450a
310/320
400
460a
240
430
420
15/20
290/300
403
340
450c 480
470
a0008916
Endress+Hauser 91
Repair Liquisys M CLM223/253
81 Central module ind. (controller) LSCP-PA PROFIBUS PA/no current output 51501221
91 Central module ind. (controller) LSCP-DP PROFIBUS DP/no current output 51502501
290 Kit CxM2x3 Relay module LSR2-DP Relay module + 2 relays 71134732
PROFIBUS DP Current input and terminals DP
valid of: hardware version 2.10
330, 340, Internal housing parts Docking assembly, empty electronics 51501073
450 box, small parts
420a, Terminal strip set Terminal strip set complete for 51502493
420c PROFIBUS DP
92 Endress+Hauser
Liquisys M CLM223/253 Repair
Press ENTER to confirm the serial number or cancel the entry to enter the number
again.
7. If available, enter the release codes for the Plus Package and/or Chemoclean in the
"Service" menu.
8. Check the Plus Package release (e.g. by opening the function group CHECK / Code P) or
the Chemoclean function.
9. Make the customer device settings again.
Endress+Hauser 93
Repair Liquisys M CLM223/253
9.5 Return
The device must be returned if repairs or a factory calibration are required, or if the wrong
device has been ordered or delivered. According to legal regulations, Endress+Hauser, as an
ISO-certified company, is required to follow certain procedures when handling returned
products that are in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions on the internet site:
www.services.endress.com/return-material
9.6 Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Please observe local regulations.
94 Endress+Hauser
Liquisys M CLM223/253 Accessories
10 Accessories
10.1 Sensors
Condumax W CLS12
• Conductive conductivity sensor for standard, Ex and high-temperature applications;
• Ordering acc. to product structure, www.products.endress.com/cls12
• Technical Information TI00082C/07/EN
Condumax W CLS13
• Conductive conductivity sensor for standard, Ex and high-temperature applications;
• Ordering acc. to product structure, www.products.endress.com/cls13
• Technical Information TI00083C/07/EN
Condumax W CLS15
• Conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);
• Ordering acc. to product structure, www.products.endress.com/cls15
• Technical Information TI00109C/07/EN
Condumax H CLS16
• Hygienic conductive conductivity sensor for pure and ultra-pure water applications (incl.
Ex);
• Ordering acc. to product structure, www.products.endress.com/cls16
• Technical Information TI00227C/07/EN
Condumax W CLS19
• Conductive conductivity sensor for pure and ultra-pure water applications;
• Ordering acc. to product structure, www.products.endress.com/cls19
• Technical Information TI00110C/07/EN
Condumax W CLS21
• Conductive conductivity sensor for applications with middle to high conductivity (incl. Ex);
• Ordering acc. to product structure, www.products.endress.com/cls21
• Technical Information TI00085C/07/EN
Indumax P CLS50
• Inductive conductivity sensor for standard, Ex and high-temperature applications
• Ordering acc. to product structure, www.products.endress.com/cls50
• Technical Information TI00118C/07/EN
Indumax H CLS52
• Inductive conductivity sensor with short response time for food applications
• Ordering acc. to product structure, www.products.endress.com/cls52
• Technical Information TI00167C/07/EN
Indumax H CLS54
• Inductive conductivity sensor for standard, Ex and in hygienic design for applications in
food, beverages, pharmaceuticals and biotechnology
• Ordering acc. to product structure, www.products.endress.com/cls54
• Technical Information TI00400C/07/EN
Endress+Hauser 95
Accessories Liquisys M CLM223/253
300 (11.8)
270 (10.6)
320 (12.6)
mm (inch)
a0019166
96 Endress+Hauser
Liquisys M CLM223/253 Accessories
1495 / 58.86
150 /
5.91
60 x 60 /
2.36 x 2.36
120 /
4.72
mm/inch
a0005742
90 / 3.54
Ø max. 60 / 2.36
70 / 2.76
1.18
30 /
Ø8/
0.31
M6
70 / 2.76
90 / 3.54
60 / 2.36
Ø8/
0.31
Ø 6 / 0.24
70 / 2.76
90 / 3.54
mm / inch
a0005676
Endress+Hauser 97
Accessories Liquisys M CLM223/253
98 Endress+Hauser
Liquisys M CLM223/253 Technical data
11 Technical data
11.1 Input
Measured variables Conductivity, resistivity, temperature
Cable specification Cable length (conductive): conductivity: max. 100 m (328 ft) (CYK71)
resistivity: max 15 m (49 ft) (CYK71)
Cable length (inductive): max 55 m (180 ft) (CLK5)
Cable resistance CYK71: 165 /km (conductivity measurement)
Endress+Hauser 99
Technical data Liquisys M CLM223/253
11.2 Output
Output signal 0/4 to 20 mA, galvanically separated, active
HART
Signal coding Frequency Shift Keying (FSK) + 0.5 mA via current output
signal
PROFIBUS PA
PROFIBUS DP
Data transfer rate 9.6 kBd, 19.2 kBd, 93.75 kBd, 187.5 kBd, 500 kBd, 1.5 MBd
100 Endress+Hauser
Liquisys M CLM223/253 Technical data
Endress+Hauser 101
Technical data Liquisys M CLM223/253
Manufacturer ID 11h
DD files www.products.endress.com/hart
PROFIBUS PA
Manufacturer ID 11h
PROFIBUS DP
Manufacturer ID 11h
102 Endress+Hauser
Liquisys M CLM223/253 Technical data
PROFIBUS PA
Polarity sensitive no
PROFIBUS DP
Endress+Hauser 103
Technical data Liquisys M CLM223/253
Temperature: 0.1 °C
Maximum Conductivity:
measured error1) Display: max. 0.5 % of measured value ± 4 digits
Conductivity signal output: max. 0.75 % of current output range
Resistivity:
Display: max. 0.5 % of measured value ± 4 digits
Resistivity signal output: max. 0.75 % of current output range
Temperature:
Display: max. 1.0 % of measuring range
Temperature signal output: max. 1.25 % of current output range
11.5 Environment
Ambient temperature -10 to +55 °C (+14 to +131 °F)
Electrical safety according EN/IEC 61010-1:2001, Installation Category II, for use up to 2000 m above sea
level
CSA Apparatus with CSA General Purpose Approval are certified for indoor use.
104 Endress+Hauser
Liquisys M CLM223/253 Technical data
Endress+Hauser 105
Appendix Liquisys M CLM223/253
12 Appendix
Operating matrix
Calibration Calibration temperature Entry of a value of Entry of correct Display of calculated Calibration status
entry (MTC) calibration solution conductivity value of installation factor is displayed
calibration solution
.
InstF 25.0 °C 2.10 %/K 1.0 o.k.;
Current meas. value C133 0.10 ... 5.0
= Installation factor C1 (3) –35.0 ... +250.0 °C 0.00 ... 20.00 %/K E---
C131 C132 0.0 µS/cm ... 9999 mS/cm C134 C135
Calibration temperature Entry of a a value of Entry of correct Display of calculated Calibration status
entry calibration solution conductivity value of cell constant is displayed
Cellc calibration solution
C1 (2) (if B1 = fixed) .
= Cell constant . .
o.k.;
2.10 %/K 0.0025 ... 99.99 1/cm
25.0 °C Current meas. value C123 E---
C121 0.00 ... 20.00 %/K C122 0.0 mS/cm ... 9999 mS/cm C124 C125
–35.0 ... +250.0 °C
Function group
CALIBRATION Residual coupling Display of residual coupling Calibration status Store calibration
Start calibration (Airset) is displayed results
AirS = Airset C1 (1) .
Temperature display Temperature display Measured value display Measured value display Uncompensated
measured value
MEAS. VALUE DISPLAY in °F suppressed Current output in % Current output in mA is displayed
with
1st error is displayed Other errors are displayed
TEMPERATURE DISPLAY
in °C (up to 10 errors)
(if present)
any code
Read mode:
Code 22
Edit mode:
Function group Selection of operation mode Selection of unit Display format selection Selection of unit Entry of cell constant Entry of cable
displayed (if A1 = conc) displayed resistance
SETUP 1 cond / ind / MOhm (if A1 = cond)
cond = conductive
ind = inductive ppm; mg/l; %; TDS; none XX.xx; X.xxx; XXX.x; XXXX auto; µS/cm; mS/cm; S/cm; 1.000 / 1.98 / 0.01 1/cm
MOhm = resistance (% only if A1 = conc) µS/m; mS/m; S/m 0.0025 ... 99.99 1/cm for 0.00ŸW
conc = concentration autoW; kW×cm; MW×cm; kW×m cond; ind; MOhm 0.00 ... 99.99 W
A A1 A2 A3 (omitted if A1 = conc) A4 A5 A6
Function group Selection of temperature Selection of temperature Entry of a value Entry of correct Temperature sensor Enter temperature
SETUP 2 measurement compensation type (if B2 = linear) process temperature calibration (omitted difference (omitted
.none (if B1 = fixed) if B1 = fixed) if B1 = fixed)
Pt100 lin = linear 2.10 %/K
Pt1k (= Pt 1000) NaCI = common salt
Pure = pure water NaCl 0.00 ... 20.00 %/K 25.0 °C Display of actual value Current offset
NTC30 (= NTC 30 kW) PureH = pure water HCl –35.0 °C ... +250.0 °C –35.0 ... +250.0 °C –5.0 ... 5.0 °C
B fixed B1 Tab = table B2 B3
B4 B5 B6
Function group Cont. switch-off Delay for cont. Delay for cont. Switch-off limit value Switch-off direction Feedforward control
CURRENT INPUT by current input switch-off current input switch-on current input for current input for current input to PID controller
Off; Input 0s 0 ... 2000 s 0s 0 ... 2000 s 50% 0 ... 100% Low; High Off; lin = linear
Z Z1 Z2 Z3 Z5 Z6
Z4
Function group Current output selection Select measured variable Current range selection Entry of 0/4 mA value Entry of 20 mA value
CURRENT OUTPUT for 2nd current output
Out1; Out2 lin = linear O3 (1) 4–20 mA; 0–20 mA 0 µS/cm / 0 kW·cm / 2000 mS/cm / 500 kW·cm /
°C; mS/cm; Contr 0Ÿ% / 0 °C 9999 % / 150.0 °C
entire measuring range entire measuring range
O O1 O2 O311 O312 O313
Function group Select contact type Select alarm Alarm delay Error current Error number Set alarm contact
ALARM delay unit setting selection to be effective
Stead = steady contact; 0 s (min)
F Fleet = fleeting contact s; min 0 s ... 2000 s (min) 22 mA 1 yes; no
F1 F2 (depends on F2) F3 2.4 mA F4 1 ... 255 F5 F6
Function group Switch polarisation Set alarm threshold Enter alarm delay Set lower Set upper Select process
CHECK detection on or off alarm threshold alarm threshold monitoring
off; on Off; Low: High; Lo+Hi; 0 s (min) 0 µS/cm 9999 µS/cm Off; AC; CC; AC+CC
P P1 P4 P6
Lo!; Hi!; LoHi! P2 0 … 2000 s (min) P3 0 … 9999 mS/cm 0 … 9999 mS/cm P5 AC!; CC!; ACCC!
C07-CLM2x3xx-13-06-00-en-001.eps
106 Endress+Hauser
Liquisys M CLM223/253 Appendix
Store calibration
results
Store calibration
results
Entry of measured
value damping
1 (no damping)
1 ... 60
A7
Entry of
reference temperature
25 °C
–35 ... 250 °C
B7
Feedforward control = 1
at
50%
0 … 100%
Z7
Set max. perm. period Set max. perm. period Set monitoring
for lower limit exceeded for upper limit exceeded value
60 min 0 ... 2000 min 120 min 0 ... 2000 min 1000 µS/cm
P7 P8 0 ... 9999 mS/cm P9
C07-CLM2x3xx-13-06-00-en-002.eps
Endress+Hauser 107
Appendix Liquisys M CLM223/253
Function of R2 (5) Start pulse selection Entry of pre-rinse time Entry of cleaning time
Switch off or on int = internal
Clean ext = external 20 s 10 s
= Chemoclean R2 (5) Off; On i+ext = internal + external 0 ... 999 s 0 ... 999 s
(only with rel. 3) i+stp = internal,
R251 suppr. by ext R252 R253 R254
Function of R2 (4) Rinse time setting Pause time setting Set minimum
Switch off or on pause time
30 s 360 min
Off; On 0 ... 999 s 1 ... 7200 min 120 min
Timer R2 (4) 1...3600 min
R241 R242 R243 R244
Function of R2 (3) Entry of set point Entry of Entry of integral action time
Switch off or on control gain Kp Tn (0.0 = no I component)
0 µS/cm / 0 kW×cm /
Off; On; Basic; PID+B 0% 1.00 0.0 min
PID controller R2 (3) entire meas. range 0.01 ... 20.00 0.0 ... 999.9 min
R231 R232 R233 R234
Function group Select contact to be Function of R2 (1) Select contact Select contact Pickup delay
RELAY configured Switch off or on switch-on point switch-off point setting
Rel1; Rel2; Rel3; Rel4 LC PV R2 (1) Off; On 9999 mS/cm / 200 MW×cm / 9999 mS/cm / 200 MW×cm / 0s
= cond. limit contactor 9999 % 9999 % 0 ... 2000 s
entire meas. range entire meas. range
R R1 R211 R212 R213 R214
Function group Table option selection Entry of number of Selection of table value Entry of temperature Entry of temperature Table status o.k.
ALPHA TABLE table value pairs value (x value) coefficient a (y value)
read 1 yes; no
edit 1 ... number of 0.0 °C 2.10 %/K
T 1 table value pairs –35.0 ... +250.0 °C 0.00 ... 20.00 %/K
T1 1 ... 10 T2 asign T3 T4 T5 T6
Function group Selection of concentration Selection of table Table option Set number of Select value pair Entry of
curve for calculation of to be edited selection value pairs uncompensated
CONCENTRATION conductivity value
display value 1
1 read 1 1 ... number of value pairs
Curve 1 ... 4 1 ... 4 edit 1 ... 10 in K4 0.0 µS/cm
K K1 K2 K3 K4 K5 0.0 ... 9999 mS/cm K6
Function group Language selection Hold configuration Manual hold Entry of Entry of SW upgrade Entry of SW upgrade
SERVICE - none = no hold hold dwell period release code release code
ENG; GER - s+c = during setup off; on (plus package) Chemoclean
ITA; FRA and calibration 10 s
ESP; NEL - CAL = during calibration 0 ... 999 s 0000 0000
S S1 - Setup = during setup S2 S3 S4 0000 ... 9999 S5 0000 ... 9999 S6
Function group
E + H SERVICE
Software Hardware Serial number Module name
version version is displayed is displayed
C07-CLM2x3xx-13-06-00-en-003.eps
108 Endress+Hauser
Liquisys M CLM223/253 Appendix
Dropout delay
entry
0
0 ... 2000 s
R275
Dropout delay
entry
0
0 ... 2000 s
R265
Entry of post-rinse time Number of repeat cycles Set interval between Set minimum pause Number of cleaning
two cleaning cycles time cycles without
20 s 0 (pause time) cleaning agent
0 ... 999 s 120 min
0 ... 5
360 min 1 ... R357 min 0
R255 R256 1 ... 7200 min 0 ... 9
R257 R258 R259
Entry of derivative Selection of Selection Entry of Entry of max. Entry of min. ON time Enter basic load
action time Tv control characteristic pulse interval pulse frequency tON
(0.0 = no D component) len = pulse length
dir = direct; freq = pulse frequency 10.0 s 120 1/min 0.3 s 40%
0.0 min inv = inverted curr = current input 2 0.5 ... 999.9 s 60 ... 180 1/min 0.1 ... 5.0 s 0 ... 40%
0.0 ... 999.9 min R236 R237
R235 R238 R239 R2310 R2311
0s 250.0 °C MAX
0 ... 2000 s –35.0 ... +250.0 °C MIN
R225 R226 R227
C07-CLM2x3xx-13-06-00-en-004.eps
Endress+Hauser 109
Liquisys M CLM223/253
Index
A E
Access codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 E+H Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Electrical connection. . . . . . . . . . . . . . . . . . . . . . . . . . 16–21
Calibration solutions . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Electrical icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Electrical symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Electromagnetic compatibility. . . . . . . . . . . . . . . . . . . . . . . 5
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46, 101 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Alarm contact. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 EP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . 104 European Pharmacopoeia . . . . . . . . . . . . . . . . . . . . . . . . . 54
Automatic mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Auxiliary voltage output . . . . . . . . . . . . . . . . . . . . . . . . . 101 F
Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
B Field device connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Binary inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Fieldbus connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Freezing of outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
C
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 H
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Hold function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29, 66
CE mark. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cell constant. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 I
Central module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 Icons
Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Checking Electrical symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Conductive sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Ingress protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Inductive sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9–15
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Chemoclean function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Installation instructions . . . . . . . . . . . . . . . . . . . . . . . 12–15
Cleaning Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Isolation voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30–69 K
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Key assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Concentration measurement . . . . . . . . . . . . . . . . . . . . . . 63
Connection diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
L
Contact outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 Limit contactor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49, 101
Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 Linearization transmission behaviour . . . . . . . . . . . . . 100
CSA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104 Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
CSA general purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27–29
Current input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39, 99
Current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
D
Declaration of conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Designated use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Diagnosis code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Dismantling
Field instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Panel-mounted instrument . . . . . . . . . . . . . . . . . . . . 87
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
110 Endress+Hauser
Liquisys M CLM223/253
M Q
Mains protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 Quick . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 Quick commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Measuring point. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 Quick setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 Quick start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Manual mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 R
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . 104 Reference temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Relative humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Measuring cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Relay contact configuration . . . . . . . . . . . . . . . . . . . . . . . . 49
Measuring frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Repeatability1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Replacing the central module . . . . . . . . . . . . . . . . . . . . . . 93
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Requirements for the personnel . . . . . . . . . . . . . . . . . . . . . 5
Mechanical construction. . . . . . . . . . . . . . . . . . . . . . . . . 105 Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100, 104
Menu Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Alpha table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 S
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Safety instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–6
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Concentration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Sensor connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Current input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Setup 1 (conductivity). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
E+H Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Setup 2 (temperature) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Simulation of conductive sensors . . . . . . . . . . . . . . . . . . . 82
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Simulation of inductive sensors. . . . . . . . . . . . . . . . . . . . . 84
Setup 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Setup 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
N System configuration . . . . . . . . . . . . . . . . . . . . . . . . . 35–69
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
T
O Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99–105
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Temperature compensation. . . . . . . . . . . . . . . . . . . . . . 104
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 linear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 NaCl. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 with table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Key assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Temperature compensation with table . . . . . . . . . . . . . . 62
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Temperature offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Temperature sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100–102 Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Timer for cleaning function . . . . . . . . . . . . . . . . . . . . . . . . 53
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . . 101 Troubleshooting
Instrument specific errors . . . . . . . . . . . . . . . . . . . . . . 79
P Process specific errors. . . . . . . . . . . . . . . . . . . . . . . . . . 75
P controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 System error messages . . . . . . . . . . . . . . . . . . . . . . . . . 72
PD controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Troubleshooting instructions. . . . . . . . . . . . . . . . . . . . . . . 72
Performance characteristics . . . . . . . . . . . . . . . . . . . . . . 104
PI controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 U
PID controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 United States Pharmacopeia . . . . . . . . . . . . . . . . . . . . . . . 54
Polarisation detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 USP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Pollution degree. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Post mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 W
Power consumption. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 Wall mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16–22
Protocol specific data . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 Workplace safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Endress+Hauser 111
71212629
www.addresses.endress.com