You are on page 1of 7

Int. J. Metrol. Qual. Eng.

8, 2 (2017) International Journal of


© S. Abbane et al., published by EDP Sciences, 2017 Metrology and Quality Engineering
DOI: 10.1051/ijmqe/2016026
Available online at:
www.metrology-journal.org

RESEARCH ARTICLE

Study of the system responsivity to measure the blackbody's


temperature by optical pyrometry from 1200 K to 1570 K
Saif Abbane*, Zahra Ben Achour, and Oualid Touayar
University of Carthage, National Institute of Applied Sciences and Technology, INSAT BP 676 Centre Urbain Nord,
1080 Tunis Cedex, Tunisia

Received: 9 October 2016 / Accepted: 1 December 2016

Abstract. This work presents a method that has been recently adopted in our laboratory to determine the
temperatures of blackbody sources in the range of 1200–1570 K. The system uses a Double Monochromator
System (DMS) based on a grating and a prism as dispersion elements. The detection element was a silicon
photodiode (Si-MMA), over which the spectral range from 800 nm to 900 nm has been used. Between the
blackbody source and the DMS was placed an optical system consists of two convergent lenses. The system
responsivity “G ” was determined by the transmission factor of the optical system and the transmission factor of
the DMS and the photodiode responsivity. The obtained results showed that the relative uncertainty of the
system responsivity “G ” varied from 0.3% to 1.12%. This in turn resulted in a corresponding uncertainty in
temperature of about 2.2 K and 4.5 K (k = 1) over the evaluated temperature range. Although this uncertainty
level was significantly high compared to those obtained by many other national metrology institutes, it was
considered as a step forward in our laboratory to measure high temperatures.
Keywords: temperature / pyrometry / system responsivity / detector / optical transmission

1 Introduction (Cu, Au, Ag, Zn, Al, etc.). Above the copper reference
point the optical pyrometry based on Planck's radiation
On the basis of The mise en pratique for the definition of the law using monochromatic radiation is necessary [11].
kelvin (MeP-K) [1–3] and what have been shown by The radiance is the selected reference parameter used to
Yamada et al. in 1999 that is possible to use metal–carbon materialize the thermodynamic temperature unit within
binary eutectic alloys as potential fixed points above the the range of high temperatures (1234.93–3000 K). In
freezing point of copper (TCu = 1357.77 K) to determine the practice, the spectral radiance is used to express the
thermodynamic temperature [4,5]. This has been followed radiance contained in an elementary wavelength window
by the Consultative Committee for Thermometry (CCT) dl including the wavelength l. This radiance is expressed
to develop and improve the thermal and radiometric in W m3 sr1.
performance of metal (carbide)–carbon (M(C)-C) fixed For the blackbody, the spectral radiance density within
points [6,7]. The efficacy of this method is part of a study the range of l and l + ld is expressed in terms of Planck's
that was made as well in the context the European radiation law connecting the temperature T and the
Metrology Research Programme (EMRP) joint project wavelength l of the emitted radiation. The Planck's law of
‘Implementing the new kelvin’ (Ink) [8] was focused on radiation is used in the definition of the thermodynamic
preparing the temperature metrology community for a temperature unit [12].
comprehensive and effective redefinition of the kelvin [9]. The spectral radiance ratio of two blackbodies
A second method is to use a filter radiometer with a can be traced back to the temperature in the ITS 90
known spectral response and defined measurement geome- scale [11].
try can determine the thermodynamic temperature of The spectral radiance density, in a given direction, at a
blackbody sources [5,10]. given point of a surface, is directly defined by the spectral
The temperature references of the International radiant flux transmitted by an elementary beam passing
Temperature Scale ITS-90 below the freezing point of through the given point and propagating in the solid angle
copper (1357.77 K) are based on various fixed points dV containing the given direction [13].
The blackbody temperature accuracy depends heavily
on the flux measurement and dimensional quantities
* Corresponding author: saif.aben@gmail.com accuracies. For the flux measurements, we used silicon

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0),
which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
2 S. Abbane et al.: Int. J. Metrol. Qual. Eng. 8, 2 (2017)

Fig. 1. Schematic diagram of experimental bench.

photodiodes. The determination of the dimensional values, 2 Experimental set up


with high accuracy, essentially depends on the quality of
the bench set up materializing the solid angle and the The bench presented in Figure 1 is intended to calibrate
emitting surface [14]. optical pyrometers. It is composed mainly of a Lantha-
The materialization of the solid angle is done through the num Chromite blackbody furnace CNHT [15] at high
implementation of an optical system based on convergent temperatures which is equipped with a variable temperature
lenses and diaphragms. This system has different reflections, blackbody. The dynamic temperature of this blackbody
absorptions and transmissions levels which are difficult to covers
evaluate, so annexes benches are required. 873.15–1773.15 K. The blackbody furnace provides a
As Planck's radiation law depends strongly on the 7-mm diameter emission diaphragm and with a current
monochromatic wavelength, we used a device based on regulated source with a stability of 105.
double monochromator (DMS) incorporating a grating and
a prism allowing respectively to disperse and to refine the In all cases, the Lanthanum Chromite cavities requiring
selected wavelength. Despite the utilization of that device, following re-verification on a standard blackbody fixed
the imperfect knowledge of the chosen wavelength and the points in the entire spectral band established by the ITS 90
use of the integrated optical introduce errors and in the range from 600 nm to 1000 nm [12].
uncertainties on the temperatures determination. The cavity bottom of the furnace is imaged by the
Consequently, the combination of the double mono- spectroradiometer which consisted of convergent lenses
chromator and the optical system, together define the L1 and L2 having focal distances of 4 cm and 8 cm,
system responsivity “G ” which involves the optical and respectively. By the help of these two lenses the cavity
dimensional effects. This system responsivity provides bottom of the furnace can be focused on the entrance slit
the correspondence between the flux received by the of the double monochromator with a diameter about
photodiode and the flux emitted by the source. Thus, the 25 mm.
received flux would be the image of the emitted one via The flux enters the double monochromator (DMS)
this system responsivity “G ” characterizing the experi- through the rectangular inlet slit (slit 1) with a height of
mental bench. The determination of “G ” is essential in 2 cm. This double monochromator is composed by a grating
order to determine the unknown temperature Tx of any 300 lines/mm blazed at 1000 nm and a silica's prism [16].
blackbody furnace exposed to our system. After undergoing some reflections by spherical mirrors S1,
This paper describes the method used to determine the S2, S3 and S4 located in the two stages of the double
system responsivity “G ” with the required accuracy and monochromator (grating stage and prism stage), the
this within the spectral range from 800 nm to 900 nm and in diffracted light falls on the silicon detector active area
the temperature range between 1200 K and 1570 K. The placed directly behind the exit slit of the monochromator.
choice of these ranges is dictated by the silicon photodiode The width of the batch of light falling on the silicon
operating range. Therefore, the measuring range of detector is 3 mm which is less than the photodiode active
temperature on one hand, covers the last three fixed area dimension estimated about 10 mm allowing thus flux
points of the ITS 90 (Ag, Au and Cu) and in the other hand, measurement.
is limited by the maximum threshold of our blackbody The photodiode is used along with a current to voltage
furnace CNHT. operational amplifier, which converts the light induced
Two methods are applicable for determining G, either photocurrent to a potential difference measured across
through a calibrated fixed-point blackbody or by using a amplifier resistance. The voltage is then measured by a
blackbody furnace whose temperatures are checked using a digital multimeter, a numerical code controls the grating
transfer pyrometer. In our case, we choose the last method and prism motors used to tune the wavelength of the
to achieve our goal. blackbody source.
S. Abbane et al.: Int. J. Metrol. Qual. Eng. 8, 2 (2017) 3

3 Theory calculation G is the system responsivity of our experimental bench.


With:
3.1 System responsivity “G ” determination M0: total radiant emittance; C1 = 3.74  1016 W m3;
C2 = 1.4388  1012 m K; M0 = pL0 W m3; F Se → Sr : DMS
The determination of the system responsivity “G ” that transmission factor; t ðl;T a Þ : optical transmission factor;
characterizes the experimental bench is accomplished via F Se → Sr ⋅tðl;T a Þ : system responsivity; Se: emitter surface;
the expression of Planck's radiation law that considers Sr: receiver surface; fle : spectral flux emitted; flr : radiate
the energetic flux of the emitting surface, the sensor power; Sl: photodiode spectral responsivity; V: voltage
responsivity, the temperature TCNHT measured by a delivered by the photodiode; I: photocurrent; R:
transfer pyrometer and the wavelength l of the blackbody amplification resistance of the converter current/voltage
emitted radiation. The transfer pyrometer we used is a device.
Heitronics KT 19 II type. This is an Infrared Radiation
Pyrometer based on pyroelectric detector with a response 3.2 Temperature determination and measurement
spectral from 2 mm to 4.5 mm; its accuracy is about ±0.5 °C uncertainty
plus 0.7% of the temperature difference between the
housing containing the measuring instruments and the As we said from the beginning, the objective of our
object to be measured [17]. work is to determine the temperature of any blackbody
By applying Planck's radiation law for a blackbody, the with its uncertainty through the use of the system
total flux emitted over the hemisphere per unit area is responsivity.
the emittance. We generally consider df the elementary For a radiant blackbody with an unknown temperature
radiant flux emitted by an area element dS in the various Tx, placed at the entrance of given the optical bench,
directions where it can radiate and we divide the flux by dS Through the use of Planck's radiation law Tx is expressed
[14,18]. as follows:
The total radiant emittance based on the elementary
flux is defined as [14]: C2 1
Tx ¼ : ð9Þ
df l Lnð1 þ ðRS l S e GC 1 l5 Þ=V Þ
M ¼
0
ðW=m2 Þ: ð1Þ
dS The parameters involved have been previously defined.
Furthermore, the spectral flux emitted according to the The uncertainties propagation law associated to the
spectral radiant emittance is: temperature Tx gives the following relative uncertainty
uT x =T x following:
sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
      
fle ¼ M 0T CNHT ðlÞ⋅S e : ð2Þ
uT x ∂T x 2 uR 2 ∂T x 2 uSl 2 ∂T x 2 uV 2
¼ þ þ
The radiate power by the silicon detector is written: Tx ∂R R ∂S l Sl ∂V V
 2    2  
∂T x ul 2 ∂T x uG 2
I V þ þ : ð10Þ
f lr ¼ ¼ : ð3Þ ∂l l ∂G G
Sl Sl R
With:
So
∂T x GS l Se C 1 C 2
flr ¼ F Se → Sr ⋅tðl;T a Þ ⋅fle : ð4Þ ¼
∂R ðlRSl S e GC 1 þ l6 V ÞLn2 ðððRSl S e GC 1 Þ=l5 V Þ þ 1Þ
Then GS l Se C 1
¼ ⋅T x ;
flr ¼ F Se → Sr ⋅t ðl;T a Þ ⋅S e ⋅M 0T CNHT ðlÞ: ð5Þ ðRS l Se GC 1 þ l V ÞLnðððRS l Se GC 1 Þ=l5 V Þ þ 1Þ
5

ð11Þ
Thus
f lr ∂T x RSl S e C 1 C 2
F Se → Sr ⋅tðl;T a Þ ¼ : ð6Þ ¼
M T CNHT ðlÞSe
0 ∂G ðlRSl S e GC 1 þ l V ÞLn2 ðððRSl S e GC 1 Þ=l5 V Þ þ 1Þ
6

RS l Se C 1
The combination between the two equations (3) and (5) ¼ ⋅T x ;
ðRSl Se GC 1 þ l5 V ÞLnðððRSl Se GC 1 Þ=l5 V Þ þ 1Þ
gives:
ð12Þ
V C 1 ⋅l5 ∂T x GRS l S e C 1 C 2
¼ Se ⋅F Se → Sr ⋅t ðl;T a Þ ; ð7Þ ¼
R⋅S l expðC 2 =ðl⋅T CNHT ÞÞ  1 ∂V lV ðRS l S e GC 1 þ l5 V ÞLn2 ðððRS l S e GC 1 Þ=l5 V Þ þ 1Þ
   
V C2 GRS l S e C 1
G ¼ F Se → Sr ⋅tðl;T a Þ ¼ exp 1 : ¼ :T x ;
RSl S e C 1 l5 lT CNHT V ðRS l S e GC 1 þ l V ÞLnðððRSl S e GC 1 Þ=l5 V Þ þ 1Þ
5

ð8Þ ð13Þ
4 S. Abbane et al.: Int. J. Metrol. Qual. Eng. 8, 2 (2017)

Transmission Factor of the opcal system for


different temperatures
0.95
Transmission Factor of the opcal

0.9
0.85
0.8
0.75 τ(20°C)
system

0.7 τ(30°C)
0.65
0.6
0.55
0.5
650 700 750 800 850 900 950
λ (nm)
Fig. 2. Transmission factor of the optical system.

∂T x 5GRSl S e C 1 C 2 temperatures of 20 °C and 30 °C. The ambient temperature


¼ 7 Ta is measured by a thermocouple k type placed close to
∂l l V ðððRSl S e GC 1 Þ=l5 V Þ þ 1Þ the optical system.
Ln2 ðððRS l S e GC 1 Þ=l5 V Þ þ 1Þ The transmission factor t for a given wavelength is
defined as the ratio between the detector signal with the
C2 optical system and the total signal without the optical system
 2
l LnðððRS l S e GC 1 Þ=l5 V Þ þ 1Þ Figure 2 summarizes the experimental results
 obtained from the determination of the optical system
5GRS l S e C 1
¼ transmission factor for both ambient temperatures 20 °C
l6 V ðððRS l S e GC 1 Þ=l5 V Þ þ 1Þ and 30 °C.
1 These values obtained of t (l,Ta) will be used to adjust
LnðððRSl S e GC 1 Þ=l V Þ þ 1Þ  ⋅T x : 5
ð14Þ
l the system responsivities “G ” according to the ambient
temperatures corresponding to blackbodies' temperatures.

3.3 Systematic effects: room temperature effect 4 Results and discussion


Blackbody sources operated at high temperature emits 4.1 Si-MMA detector spectral responsivity
considerable thermal load. It has been observed that when
the blackbody operates the temperature range of interest The calibration of the (Si-MMA) detector used in our
(1223.15 K and 1423.15 K) temperature of the optical spectroradiometer, was done through the use of a standard
system, which composed from the two lenses, changes from trap detector calibrated traceable to the primary labora-
20 °C to 30 °C. This in turns would affect the spectroradi- tory LNE  CNAM (France).
ometer system, thus, its responsivity. In order for this effect The estimated relative uncertainty of the absolute
to be evaluated, the transmittance of the two lenses has spectral responsivity determination is about 3  103 to
been measured at variable temperatures from 20 °C to 30 °C a 1s confidence level. This uncertainty takes into
and the change of transmittance at different wavelengths account the uncertainty of the standard Si-CNAM,
has been evaluated. This has been carried out by using The repeatability, the reproducibility of measurements,
a quartz halogen lamp directly placed in front of the the current-voltage converter and the used voltmeter
monochromator; however, the mirrors were placed behind calibration.
the monochromator and just before the detection element. Figure 3 shows the shape of the absolute spectral
Two spherical mirrors having a 99% reflectance, reflect responsivity of the Si-MMA photodiode used in our bench.
the radiation delivered from the DMS, to the optical In the range between 800 nm and 900 nm, the absolute
system. The transmission of the optical system essentially spectral responsivity of the used silicon photodiode in our
depends on the nature of the glasses, the wavelength l experimental bench is at its maximum value while keeping
and the ambient temperature Ta in which the system is a linear variation and increases according to the wave-
arranged. length and achieves the peak near 950 nm. To use the bench
The detector used is a silicon photodiode. In our study, with the best signal stability, we will choose the range
we are interested to the wavelengths effects between between 800 nm and 900 nm. This choice will therefore
650 nm and 900 nm on optical system transmission for the determine the range of our experimental bench.
S. Abbane et al.: Int. J. Metrol. Qual. Eng. 8, 2 (2017) 5

Photodiode spectral responsivity


0.8000
Spectreal responsivity(A/W)

0.7000
0.6000 Si CNAM
0.5000 Si MMA
0.4000
0.3000
0.2000
0.1000
0.0000
500 600 700 800 900 1000
λ ( nm )
Fig. 3. Spectral responsivities of Si-MMA and Si-CNAM detectors.

Detector's electric signal


0.25

0.2 1223,15K
1234,85K
0.15
Sél (v)

1353,15K
0.1
1381,15K
0.05 1423,15K

0 1527,15K
600 650 700 750 800 850 900 950 1573,15K
λ(nm)
Fig. 4. Electric responses of Si-MMA detector for different temperatures of the blackbody furnace CNHT.

4.2 The electric signal In our case, we consider for each wavelength, G
The signal delivered by the Si–MMA detector is the the average of eight (08) Gi values corresponding to
response of the experimental bench for various temper- eight (08) different temperatures from 1223.15 K to
atures ranging from 1223.15 K to 1573.15 K and in the 1573.15 K.
wavelength range from 650 nm to 900 nm integrating all Table 1 summarizes the values of G for all the used
the parameters involved (Fig. 4). wavelengths. For the uncertainty estimation we took
the worst case and we estimated the uncertainty u as
We noticed that the shape of the Si-MMA detector G
responses according to the wavelength was similar to the maximum deviation recorded for the G values and
Planck's radiation law applied to the wavelength range for each wavelength.
from 650 nm to 900 nm. These responses are of a major Figure 5 shows the variation of the system responsivity
importance in the determination of the unknown tempera- “G ” of the experimental bench depending on the wave-
ture Tx of any blackbody. length, for various temperatures.
We noticed that in the range between 800 nm and
4.3 System responsivity measuring, G (l, TCNHT, Ta) 900 nm, the errors introduced by the system responsivity
on the determination of the temperature Tx are less than
Having identified the responsivity of the system, as the errors between 650 nm and 800 nm. This is due to the
shown earlier, it has then been used to determine the linear variation of the spectral responsivity of the silicon
temperature. photodiode in this range.
6 S. Abbane et al.: Int. J. Metrol. Qual. Eng. 8, 2 (2017)

Table 1. System responsivity average for each wavelength.


u
l (nm) uV ul uSl uR uT CNHT
G G
V l Sl R T CNHT
G
650 0.00001 0.005 0.003 0.00001 0.007 9.4093E-14 0.0334
700 0.00001 0.005 0.003 0.00001 0.007 1.4606E-13 0.0313
750 0.00001 0.005 0.003 0.00001 0.007 1.7532E-13 0.0275
800 0.00001 0.005 0.003 0.00001 0.007 1.9565E-13 0.0032
850 0.00001 0.005 0.003 0.00001 0.007 2.3353E-13 0.0069
900 0.00001 0.005 0.003 0.00001 0.007 2.8743E-13 0.0112

Table 2. Uncertainties associated to different temperatures for three wavelengths selected.

T1223.15 (K) T1353.15 (K) T1573.15 (K)


uV uG ul uS l uR
l (nm) V l Sl R uT 1223:15 ðKÞ uT 1353:15 ðKÞ uT 1573:15 ðKÞ
G
800 0.00001 0.0032 0.005 0.003 0.00001 2.13 2.61 3.58
850 0.00001 0.0069 0.005 0.003 0.00001 2.32 2.84 3.88
900 0.00001 0.0112 0.005 0.003 0.00001 2.60 3.18 4.34

System responsivity variaon


3.5E-13
1223,15K
3E-13
1234,85K
“G” responsivity

2.5E-13 1353,15K
2E-13 1381,15K

1.5E-13 1423,15K
1524,15K
1E-13
1527,15K
5E-14
1573,15K
0
600 650 700 750 800 850 900 950 1000
λ(nm)
Fig. 5. System responsivities for each temperature.

This approach allows us to consider that the relative For a given wavelength, the uncertainty of the tempera-
uncertainty associated to G vary from 0.3% to 1.12%. ture increases according to the temperature. Also for a given
These levels of uncertainty have a major importance in the temperature, the uncertainty uT x increases according to the
uncertainties estimation of the blackbodies' unknown wavelength. This is predictable because in the expression
temperatures Tx. of the uncertainty associated to Tx (Eq. (10)), the only
predominant uncertainties are those related to the average
5 Discussion of the system responsivities G and the wavelength.
For our pyrometer bench, we should select the
wavelength of 800 nm to obtain the lowest uncertainty
Using the equations (9) and (10) and considering the G uT x of the temperature to be determined.
values (Tab. 1) and their corresponding uncertainties, we
are hence able to determine any temperature Tx between
1223.15 K and 1573.15 K and estimate their associated 6 Conclusion
uncertainty.
Table 2 gives practical examples of specific temper- This paper presents a method that has been recently
atures and their uncertainties for three selected wave- adopted in our laboratory to determine the temperatures of
lengths. blackbody sources in the range of 1200–1570 K. The system
S. Abbane et al.: Int. J. Metrol. Qual. Eng. 8, 2 (2017) 7

is made of a Double Monochromator System (DMS) based Yamada, Y. Yamaguchi, H.W. Yoon, Z. Yuan Campos, M.G.
on a grating and a prism as dispersion elements. The Cox, D. del Campo, W. Dong, M.R. Dury, V. Gavrilov, I.
detection element is a silicon photodiode that is calibrated Grigoryeva, M.L. Hernanz, F. Jahan, B. Khlevnoy, V.
traceable to the LNE-CNAM's trap detector. In between Khromchenko, D.H. Lowe, X. Lu, G. Machin, J.M. Mantilla,
the blackbody source and the DMS is placed the optical M.J. Martin, H.C. McEvoy, B. Rougié, M. Sadli, S.G.R.
system consists of two convergent lenses. The crucial Salim, N. Sasajima, D.R. Taubert, A.D.W. Todd, R. Van den
parameter in this study is the system responsivity “G ” that Bossche, E. van der Ham, T. Wang, A. Whittam, B. Wilthan,
reflects the influence of the instrument in the temperature D.J. Woods, J.T. Woodward, Y. Yamada, Y. Yamaguchi, H.
determination. W. Yoon, Z. Yuan, Thermodynamic temperature assignment
to the point of inflection of the melting curve of high
The optical transmission factor combined with the
temperature fixed points, Philos. Trans. R. Soc. A 374,
DMS transmission factor and the spectral sensitivity of the
20150044 (2016)
detector allowed us to determine the system responsivity 4. Y. Yamada, H. Sakate, F. Sakuma, A. Ono, Radiometric
“G ” of the experimental bench. The use of the DMS with a observation of melting and freezing plateaus for a series of
grating and a prism placed in a small area of 50 cm  30 cm metal-carbon eutectic points in the range 1330 °C to 1950 °C,
compared to other comparators used by many laboratories Metrologia 36, 207–209 (1999)
around the world allows our system to have a fast 5. S.G.R. Salim, K. Anhalt, D.R. Taubert, J. Hollandt, Three-
operational readiness, high effectiveness and easy handling. element trap filter radiometer based on large active
Furthermore, we tried to define the responsivity of our area silicon photodiodes, Appl. Opt. 55 (15), 3958–3965
system to stand out from the other works which they use (2016)
the spectral radiance ratio of two blackbodies to trace back 6. E. Woolliams, G. Machin, D. Lowe, R. Winkler, Metal
to the temperature. (carbide) – carbon eutectics for thermometry and radiome-
In this work, we defined the experimental model of the try: a review of the first seven years, Metrologia 43, R11–R25
system responsivity “G ” to be used for the temperature (2006)
range between 1200 K and 1570 K and from 800 nm to 7. G. Machin, Twelve years of high temperature fixed point
900 nm. This model helps us to determine the temperature research: a review, in Temperature: its measurement and
Tx of any blackbody in this range. control in science and industry, Proceedings of the ninth
The performance of the experimental bench has been international temperature symposium, California, pp. 305–315
thoroughly investigated considering all sources of relative (2013)
uncertainty involved of the system responsivity “G ” which 8. G. Machin, M. Sadli, R. Gavioso, J. Engert, E.R. Woolliams,
varied from 0.3% to 1.12%. This in turn resulted in a The Euramet Metrology Research Programme project:
implementing the new kelvin (Ink), Int. J. Thermophys.
corresponding uncertainty in temperature of about 2.2 K
35, 405–416 (2014), doi:10.1007/s10765-014-1606-4
and 4.5 K (k = 1) from 800 nm to 900 nm.
9. J. Fischer, Low uncertainty Boltzmann constant determi-
nations and the kelvin redefinition, Philos. Trans. R. Soc. A
This research and innovation are carried out under the
374, 20150038 (2016), doi:10.1098/rsta.2015.0038
MOBIDOC device funded by the European Union under the
10. B. Khlevnoy, V. Sapritsky, B. Rougie, C. Gibson, H. Yoon, A.
PASRI program administered by the ANPR.
Gaertner, D. Taubert, J. Hartmann, CCPR-S1 supplemen-
The authors wish to thank the National Agency of Metrology
tary comparison for spectral radiance in the range of
Tunisia (ANM) for supporting this research.
220 nm to 2500 nm, Metrologia 46, S174–S180 (2009)
11. BIPM, Supplementary Information for the ITS 90 (2013),
References Chap. 6
12. G. Negro, Le corps noir. Étude et réalisation de cavités
1. D.C. Ripple, R. Davis, B. Fellmuth, J. Fischer, G. Machin, T. étalons de référence pour le rayonnement thermique, PhD
Quinn, P. Steur, O. Tamura, D.R. White, The roles of the mise thesis, École centrale Paris, 1993
en pratique for the definition of the kelvin, Int. J. Thermophys. 13. I.E.C., Vocabulary Electrotechnical International Genève
31, 1795–1808 (2010), doi:10.1007/s10765-010-0837-2 BCCI (2016), Chap. 845, Publi 50
2. B. Fellmuth, J. Fischer, G. Machin, S. Picard, P.P.M. Steur, 14. J. Sacadura, Initiation aux Transferts Thermiques, 4th edn.,
O. Tamura, D.R. White, H. Yoon, The kelvin redefinition Tec Doc (1993)
and its mise en pratique, Philos. Trans. R. Soc. A 374, 15. BOUHNIK SAS Banc d'étalonnage EIT 90 Datasheet
20150037 (2016), doi:10.1098/rsta.2015.0037 16. SOPRA SAS Double monochromator DMS Datasheet
3. E.R. Woolliams, K. Anhalt, M. Ballico, P. Bloembergen, F. 17. Heitronics INFRARED RADIATION PYROMETER KT 19
Bourson, S. Briaudeau, J. Sasajima, D.R. Taubert, A.D.W. II Datasheet
Todd, R. Van den Bossche, E. van der Ham, T. Wang, 18. G. Asch et al., Les capteurs en instrumentation industrielle,
A. Whittam, B. Wilthan, D.J. Woods, J.T. Woodward, Y. 5th edn., Dunod (1999)

Cite this article as: Saif Abbane, Zahra Ben Achour, Oualid Touayar, Study of the system responsivity to measure the
blackbody's temperature by optical pyrometry from 1200 K to 1570 K, Int. J. Metrol. Qual. Eng. 8, 2 (2017)

You might also like