RADIOGRAPHIC GENERAL PURPOSE

Principal : PT Siemens Indonesia


ADDRESS : Arkadia Office Park, Tower F,
Lvl 18, Jl. T.B.Simatupang kav
88, Pasar Minggu
PHONE/ FAX. / E-MAIL : +6221 2754 3000

NAME OF PRODUCT Digital Floor Mounted


Radiology System 550 mA
BRAND / MANUFACTURER Siemens

MODEL / TYPE Multix Select DR


OUTSOURCE EXTERNAL MANUFACTURE (OEM) no
PRODUCT'S REGISTRATION (AKL/AKD) AKL 21501212379
PRODUCT'S CERTIFICATION
FDA Clearance yes
CE Mark (MDD) yes
IEC Compliance (Diisi daftar standar IEC yang dipenuhi) IEC 60601
Certificate .... ISO 13485
Locally Content (Tingkat Komponen Dalam Negeri)
WHERE MARKETED Worldwide
REFERENCE OF TECHNICAL
CONFIGURATION digital floor-mounted x-ray system
complete with patient table and
bucky wall stand
Over/under table overtable
Remote operation yes
IMAGING TOWER floor mounted
Longitudinal movement, cm (in) 138 cm
Vertical movement, cm (in) 35 cm to 185 cm
Maximum SID, cm (in) 115 cm at table
Controls High-performance PC with flat
panel display and user interface.
Control console.
Table/tower movement Footswitch for the control of 8-
way tabletop float movements.
Manual movement of x-ray tube.
Radiography yes
Removable grid yes
DETECTOR yes, wired Flat Panel Detector
Detector type Amorphous silicon (a-Si)
Detector size, cm (in) 35 cm x 43 cm
Field sizes, cm (in) 35.3 cm x 42.4 cm (active area)
RADIOGRAPHY, SPOT FILMING
Acquisition matrix size 3072 x 2560 pixels
Maximum spatial resolution 3.6 lp/mm
Acquisition bit depth 14 bits with 8-fold oversampling
RADIOGRAPHY, CONVENTIONAL
Digital detector yes, wired Flat Panel Detector
In table yes
On tower yes
Vertical, chest yes
Cross table yes
PATIENT TABLE
Maximum patient weight, kg (lb) 200 kg
Tabletop dimensions, cm (in) 235 cm x 80 cm
Lateral movement, cm (in) ± 10 cm, transverse
Longitudinal movement, cm (in) ± 46 cm, longitudinal
Minimum height, cm (in) 75 cm
Table tilt, ° -
BUCKY WALL STAND
Vertical movement range 35 cm to 185 cm
Grid Stationary grid
Front-cover-to-detector distance < 40 mm
X-RAY PRODUCTION
Generator power, kW 55 kW
Preferred x-ray tubes POLYDOROS/OPTITOP
Radiographic kVp 40 kV to 133 kV
Radiographic mA range 550 mA at 100 kV
Dose area product Yes
IMAGE DISPLAY
Control room monitor size 19" monitor
CONNECTIVITY
DICOM yes
ACCESSORIES
Handgrips yes
REMOTE DIAGNOSTICS yes
POWER REQUIREMENTS 3-phase, 380/400 V ±10%, 50/60
Hz, 3-phase, 440/480 V ±10%,
50/60 Hz wih pre-transformer

OTHER REQUIREMENTS (room size, wall etc) min. 3.5 m x 4.7 m


FUNCTION Digital Floor Mounted
Radiography System
OTHER INFORMATION *)
*) Example : Delivery Time, Capacity of Production, Price, Delivery time 3-4 months
Commitment of Maintenance, Warranty, Discontinue/Life Warranty 1 year
Time Medical Device, etc. Preventive maintenance 2x/ year
Spareparts availability up to 10
years