Professional Documents
Culture Documents
REFERENCE OF TECHNICAL
Azurion 7 M12/M20 with Clarity IQ
Gantry
Gantry Mounting system Ceiling Mounted
Longitudinal movement motorized and manual range of 260 cm
The gantry positions can be recalled based on the stored Yes
image information
The sequence of projections must be physician Yes
dependent
Collision protection Yes, With BodyGuard
Patient support
Floor‐mounted radio translucent floating top. It shall be Yes
able to cover full head to toe examination without
repositioning the patient
The patient can have CPR at any position of the tabletop Yes
Table Length 319 cm
Table width 50 cm
Pivot Table ‐90°/+180° or ‐180°/90°
Generator
Voltage range 40‐125 kV
Maximum current 1000 mA
Different exposure protocols can be customized Yes
Detector
Detector size Available in two options, diagonal size 20 inch (M20) for
large detector or diagonal size 12 inch for small detector.
*Pembeli wajib menyebutkan ukuran 1etector yang ingin dibeli pada saat proses pembelian melalui situs e‐catalogue
Monitors
Monitor Type Medical Grade
Monitor in Examination Room 1 x 58‐inch monitor
System functionality
QA packages for Coronary analysis. Yes
QA packages for Vascular analysis. Yes
QA packages for Left Ventricular analysis. Yes
Acquisition mode: Shall be possible for any body part to
Rotational Angiography (single Axis Rotation Scan) provide 3D impression of the complex vasculature and
coronary arteries including lower limb.
Rotation Swing / Simultaneous Dual Axis Rotation Scan: Yes, With CardiacSwing
Yes, Interventional workstation for reconstructing images
from Rotational Angiography to provide 3D images of brain
3D Rotational Angiography:
& peripheral vessels to facilitate endovascular interventions
in neuro & vascular radiology.
Pre‐ and post‐Stent Deployment Tool: Yes, With StentBoost
Digital Subtracted Angiography Yes
Subtracted Bolus Chase and Bolus Chase Reconstruction Yes
Automatic creation and storage of a dynamic roadmap Yes, Dynamic Coronary Roadmap
from each acquired coronary angiogram (Roadmap on
coronary angiogram)
Capability to visualize of balloons and stents in coronary Yes, with StentBoost Live
arteries on‐the‐fly to clarify the situation at any moment
during an intervention (Live images)
3D post processing access on patient table side Yes, with SmartCT
Vessel analysis based on pre‐acquired 3D volume data Yes, with SmartCT Vessel Analysis
with access on patient table side
Dose‐saving measures
Zero Dose Positioning Yes
The anti‐scatter grid shall be removable, e.g., for pediatric Yes
acquisitions
Accumulated AK indication for different patient body Yes
zones, with warning in case 2Gy has been reached for
every individual zone
On screen display for stand position, tilt position, tube Yes
load status, fluoro mode detector field size, copper
filtration. Both rate and accumulated dose shall be
available in the exam and control room.
Hemodynamic System
Flexibility mounting at the table side to improve access to Yes
the patient
Equipped with touch screen monitor on the hemo system Yes
Autorotate view on the hemo system Yes
Minimum monitoring capabilities:
• NIBP Yes
• SpO2 Yes
• Invasive Pressures Yes
• 12 Lead Surface ECG Yes
Accessories
Intercom Yes
Peripheral X‐ray Filter Yes
Arm Support Yes
Head Support Yes
Table mounted radiation shield Yes
Ceiling mounted radiation shield Yes
Exam Light Yes
Pan Handle Yes
Injector Yes
Full System UPS Yes