Professional Documents
Culture Documents
The attached
copy is furnished to the author for internal non-commercial research
and education use, including for instruction at the authors institution
and sharing with colleagues.
Other uses, including reproduction and distribution, or selling or
licensing copies, or posting to personal, institutional or third party
websites are prohibited.
In most cases authors are permitted to post their version of the
article (e.g. in Word or Tex form) to their personal website or
institutional repository. Authors requiring further information
regarding Elsevier’s archiving and manuscript policies are
encouraged to visit:
http://www.elsevier.com/authorsrights
Author's personal copy
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324
ScienceDirect
a r t i c l e i n f o a b s t r a c t
Article history: Objective. Comparison of physical and mechanical properties of one conventional and a new
Received 11 April 2013 NiTi wire, which had received an additional thermomechanical treatment.
Received in revised form Methods. Specimens of both conventional (NiTi) and the new type of wire, called M-Wire
30 June 2013 (MW), were subjected to tensile and three-point bending tests, Vickers microhardness mea-
Accepted 7 October 2013 surements, and to rotating-bending fatigue tests at a strain-controlled level of 6%. Fracture
surfaces were observed by scanning electron microscopy and the non-deformed micro-
structures by transmission electron microscopy.
Keywords: Results. The thermomechanical treatment applied to produce the M-Wire apparently
Endodontic instruments increased the tensile strength and Vickers microhardness of the material, but its appar-
M-Wire ent Young modulus was smaller than that of conventionally treated NiTi. The three-point
Nickel–titanium bending tests showed a higher flexibility for MW which also exhibited a significantly higher
Thermomechanical treatment number of cycles to failure.
Significance. M-Wire presented mechanical properties that can render endodontic instru-
ments more flexible and fatigue resistant than those made with conventionally processed
NiTi wires.
© 2013 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
∗
Corresponding author at: Department of Metallurgical and Materials Engineering, Universidade Federal de Minas Gerais, Av. Antonio
Carlos 6627 Bloco 2, Room 2640, 31270-901 Belo Horizonte, MG, Brazil. Tel.: +55 31 3409 1859; fax: +55 31 3409 1815.
E-mail addresses: vbuono@demet.ufmg.br, vicentebuono@gmail.com (V.T.L. Buono).
0109-5641/$ – see front matter © 2013 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.dental.2013.10.004
Author's personal copy
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324 e319
e320 d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324
3. Results
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324 e321
Table 1 – Mean values (and standard deviations) of the apparent modulus of elasticity, E, transformation stress, A-M ,
ultimate tensile strength, UTS , and total elongation, et , of tensile tested NiTi and MW wires.
NiTi wires E (GPa) A-M (MPa) UTS (MPa) et (%)
NiTi 43.9 (1.8) 506.7 (17.8) 1330.0 (6.6) 19.3 (0.4)
MW 24.6 (0.6) 503.3 (2.3) 1547.3 (1.5) 21.7 (0.5)
Fig. 5 – SEM images of fractured MW (a) and (b) and NiTi (c) and (d) surfaces after fatigue testes. Left: smooth regions at the
edges of the cross-section (arrowed) and fibrous region at the center; right: detail of the smooth region, showing fatigue
striations (dark arrows) and secondary cracks (wider dark arrows).
Author's personal copy
e322 d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324
Fig. 6 – TEM micrographs showing the microstructure of (a) MW with 20-nm wide martensite variants and precipitates; (b)
NiTi containing essentially the -phase austenite, with a dislocation-cell substructure, and precipitates.
cycles to fracture can give an estimate of the resistance of of file design mentioned above and in fact the results of mea-
these instruments during clinical use. surements made on endodontic instruments are controversial
The realization that the mechanical properties of near [9,18,19].
equiatomic NiTi alloys are strongly influenced by the stress- The results of the three-point bending tests indicated that
induced phase transformation taking place in this alloy and the force required to bend MW was smaller than that applied
that the microstructural changes introduced by thermome- to deform NiTi by the same amount, until larger strains
chanical treatments can control this phase transformation were achieved (Fig. 4). These results are consistent with the
constitutes the modern trend toward developing new rotary tensile-tests data, but their relevance lies in the fact that
endodontic instruments with improved mechanical proper- bending strains, as opposing to uniaxial tensile strains, are
ties [7,11]. In the present study, a few methods used in the type encountered during the clinical use of endodon-
engineering for probing into the microstructure and mechan- tic instruments to clean and shape curved root canals. The
ical behavior of materials were employed to analyze NiTi observation that smaller forces are required to bend MW at
wires used in the manufacture of endodontic instruments. intermediate vertical displacements is an indication that the
This approach is justified by the absence of standardized structural damage associated with rotating-bending fatigue of
test methods and devices to assess the mechanical behavior MW should be smaller than in conventional NiTi.
of NiTi rotary instruments [17]. By comparing the properties The results of the rotating-bending fatigue tests confirm
of M-Wire with those of a conventionally treated NiTi wire, the prediction just mentioned that the fatigue resistance of
it should be possible gain a better understanding of some MW is greater than that of NiTi. In fact, the mean value of num-
aspects of the clinical performance of the new instruments ber of cycles to failure of MW, 1125 (141), was almost 3 times
made from MW. In addition, the use of wires enables the greater than that of NiTi, 436 (86). This is in agreement with
elimination of any potential effects of file design (size/taper, the high fatigue strength reported in earlier studies performed
cross-section geometry, helical angle, etc.) on the mechanical with endodontic instruments made with MW [7–9,17].
properties. Studies evaluating the flexural fatigue resistance of
The tensile curves shown in Fig. 3 revealed, for both wires, nickel–titanium rotary files employed devices with varying
the four stages associated with the uniaxial deformation of testing set-up such as steel grooves, steel slopes or 3-point
superelastic NiTi. The first stage corresponds to the elastic steel pins to guide the file into the desired curvature [18,20].
deformation of austenite (the crystalline phase present at In addition, the radii and angles of curvature of the testing
room temperature in superelastic NiTi); the second stage is devices were different, despite the fact that these two param-
the plateau associated with the stress-induced martensitic eters were shown to have a significant effect on the fatigue
transformation; the third stage is the elastic deformation of resistance of rotary nickel–titanium instruments [6,21]. This
martensite, while in the fourth stage corresponds to the plas- lack of an internationally accepted standard for a method or
tic deformation of martensite until final rupture takes place device for fatigue testing has resulted in the observed vari-
[1]. As far as the behavior of endodontic instruments made ability of testing results [8]. For the present study, a testing
of the two wires is concerned, the main difference in prop- device (Fig. 2) was constructed on the basis of the designs by
erty is the lower apparent elastic modulus of MW (Table 1). Tobushi et al. [22] and Figueiredo et al. [23], which allowed the
The higher strength of MW, confirmed by the microhardness standardized choice of the tensile-strain amplitude according
results, would lead to more torsional resistant instruments. to Eq. (2). The value chosen for the tests, 6% strain, is of the
These predictions are difficult to confirm due to the effects order of magnitude of the maximum strains generally found
Author's personal copy
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324 e323
for rotary NiTi instruments employed in formatting curved Shape memory materials. 1st ed. Cambridge, UK: Cambridge
root canals [24]. University Press; 1998.
The improved fatigue behavior of MW should be correlated [2] Otsuka K, Ren X. Physical metallurgy of Ti–Ni-based shape
with its fine structure as revealed in the TEM micrographs memory alloys. Prog Mater Sci 2005;50:511–678.
[3] Saburi T. Ti–Ni shape memory alloys. In: Otsuka K, Wayman
(Fig. 6), which showed the presence of thin martensite variants
CM, editors. Shape memory materials. 1st ed. Cambridge,
in non-deformed specimens. The two images presented in UK: Cambridge University Press; 1998.
Fig. 6 are quite similar to those published by Alapati et al. [11], [4] Hülsmann M, Peters OA, Dummer PMH. Mechanical
considering that the original magnification of these images preparation of root canals: shaping goals, techniques and
is two times that of [11]. The presence of these martensite means. Endod Top 2005;10:30–76.
variants, which can be related to the high transformation [5] Peters OA, Paqué F. Current developments in rotary root
canal instrument technology and clinical use: a review.
temperatures found in M-Wire [11,12], should explain the dif-
Quintessence Int 2010;41:479–88.
ferences in mechanical behavior between the two specimens.
[6] Pruett JP, Clement DJ, Carnes Jr DL. Cyclic fatigue testing of
According to Miyazaki and Otsuka [25], martensite variants nickel–itanium endodontic instruments. J Endod
can undergo stress-induced reorientation at relatively lower 1997;23:77–85.
stresses and this can be the reason for the lower apparent elas- [7] Johnson E, Lloyd A, Kuttler S, Namerow K. Comparison
tic modulus of MW and the smaller force required for bending between a novel nickel–titanium alloy and 508 nitinol on the
by the same amount MW in relation to conventional NiTi. The cyclic fatigue life of ProFile 25/.04 rotary instruments. J
Endod 2008;34:1406–9.
higher fatigue resistance found for MW should also be result
[8] Al-Hadlaq SMS, AlJarbou FA, AlThumairy RI. Evaluation of
of these martensites, which would facilitate crack nucleation cyclic flexural fatigue of M-Wire nickel–titanium rotary
but render propagation more difficult because of the large instruments. J Endod 2010;36:305–7.
number of interfaces present. These interfaces are known to [9] Peixoto IFC, Pereira ESJ, Silva JG, Viana ACD, Buono VTL,
cause the formation of complex arrays of secondary cracks, Bahia MGA. Flexural fatigue and torsional resistance of
dissipating the energy required for crack propagation [26]. In ProFile GT and ProFile GT series X instruments. J Endod
fact, lower fatigue crack propagation rates were observed by 2010;36:741–4.
[10] Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB.
McKelvey and Ritchie [27] in martensitic NiTi alloys, in com-
Effects of raw material and rotational speed on the cyclic
parison with stable and superelastic austenite. More recently, fatigue of ProFile Vortex rotary instruments. J Endod
Figueiredo et al. [23] showed that the number of cycles to fail- 2010;36:1205–9.
ure in martensitic NiTi wires can be as much as 100 times [11] Alapati SB, Brantley WA, Iijima M, Clark WAT, Phil D, Kovarik
larger than in stable and superelastic austenitic NiTi. L, et al. Metallurgical characterization of a new
Within the limitations of this study, M-Wire showed supe- nickel–itanium wire for rotary endodontic instruments. J
Endod 2009;35:1589–93.
rior physical and mechanical properties than conventional
[12] Pereira ESJ, Peixoto IFC, Viana ACD, Oliveira II, Gonzalez BM,
NiTi wire for manufacturing rotary endodontic instruments.
Buono VTL, et al. Physical and mechanical properties of a
Future research should focus on new raw materials and ther- thermomechanically treated NiTi wire used in the
momechanical processing procedures which can be used to manufacture of rotary endodontic instruments. Int Endod J
optimize even further the alloy microstructure in order to 2012;45:469–74.
guarantee the reliability and safety of NiTi rotary instruments [13] Leroy AMF, Bahia MGA, Ehrlacher A, Buono VTL. An
in the clinical practice. analytical mechanical model to describe the response of
NiTi rotary endodontic files in a curved root canal. Mater Sci
Eng C 2012;32:1594–600.
5. Conclusions [14] Vieira EP, França EC, Martins RC, Buono VTL, Bahia MGA.
Influence of multiple clinical use on fatigue resistance of
ProTaper rotary nickel–titanium instruments. Int Endod J
The thermomechanically treated NiTi wire termed M-Wire
2008;41:163–72.
was more flexible but harder than the conventional NiTi [15] Gao Y, Gutmann JL, Wilkinson K, Maxwell R, Ammon D.
superelastic wire. It contained martensite in its non-deformed Evaluation of the impact of raw materials on the fatigue and
microstructure and this is probably the reason why M-Wire mechanical properties of ProFile Vortex rotary instruments. J
showed higher fatigue resistance under 6% strain level than Endod 2012;38:398–401.
the NiTi wire. [16] Gekelman D, Ramamurthy R, Mirfarsi S, Paqué F, Peters OA.
Rotary nickel–titanium GT and ProTaper files for root canal
shaping by novice operators: a radiographic and
Acknowledgements micro-computed tomography evaluation. J Endod
2009;35:1584–8.
This work was partially supported by Fundação de Amparo [17] Ye J, Gao Y. Metallurgical characterization of M-Wire
nickel–titanium shape memory alloy used for endodontic
à Pesquisa do Estado de Minas Gerais (FAPEMIG), Belo Hori-
rotary instruments during low-cycle fatigue. J Endod
zonte, MG, Brazil, and Conselho Nacional de Desenvolvimento 2012;38:105–7.
Científico e Tecnológico (CNPq), Brasília, DF, Brazil. [18] Kramkowski TR, Bahcall J. An in vitro comparison of
torsional stress and cyclic fatigue resistance of ProFile GT
references and ProFile GT Series X rotary nickel–itanium files. J Endod
2009;35:404–7.
[19] Kell T, Azarpazhooh A, Peters OA, El-Mowafy O, Tompson B,
Basrani B. Torsional profiles of new and used 20/.06 GT
[1] Otsuka K, Wayman CM. Mechanism of shape memory effect series X and GT rotary endodontic instruments. J Endod
and superelasticity. In: Otsuka K, Wayman CM, editors. 2009;35:1278–81.
Author's personal copy
e324 d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e318–e324
[20] Cheung GSP, Darvell BW. Low-cycle fatigue of rotary NiTi loading on the behaviour of nickel–titanium wires employed
endodontic instruments in hypoclorite solution. Dent Mater in the manufacture of rotary endodontic instruments. Int
2008;24:753–9. Endod J 2005;38:795–801.
[21] Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C. [25] Miyazaki S, Otsuka K. Deformation and transition behaviour
Dynamic and cyclic fatigue of engine-driven rotary associated with the R Phase in Ti–Ni alloys. Metall Trans A
nickel–titanium endodontic instruments. J Endod 1986;17A:53–63.
1999;25:434–40. [26] Hornbogen H. Fatigue of copper-based shape memory alloys.
[22] Tobushi H, Hachisuka T, Yamada S, Lin PH. Rotating-bending In: Duerig TW, Melton KN, Stöckel D, Wayman CM, editors.
fatigue of TiNi shape-memory alloy wire. Mech Mater Engineering aspects of shape memory alloys. London:
1997;26:35–42. Butterworth-Heinemann; 1990. p. 267–82.
[23] Figueiredo AM, Modenesi P, Buono V. Low-cycle fatigue life [27] McKelvey AL, Ritchie RO. Fatigue-crack growth behavior in
of superelastic NiTi wires. Int J Fatigue 2009;31:751–8. the superelastic and shape-memory alloy Nitinol. Metall
[24] Bahia MGA, Martins RC, Gonzalez BM, Buono VTL. Physical Mater Trans A 2001;32 A:731–43.
and mechanical characterization and the influence of cyclic