You are on page 1of 136

Servo Ventilator 300A

Servo Ventilator 300A System


Technical Training course
Servo Ventilator 300A

Course program
Day 1

! Introduction SV300A
! Principle of Operation
Servo Ventilator 300A

Course program
Day 2

! Ventilation Modes
! Construction
! Disassemble / Reassemble Pneumatic unit and User
Interface
! Calibration
! Functional Checks
Servo Ventilator 300A

Course program
Day 3

! Servo feedback system


! Principle Diagram / Block Diagram
! Cleaning / Maintenance
! Accessories
Servo Family
Servo Ventilator ICU Systems
Clinical Applications
User Interface and Patient Unit
User Interface
Panel Design
Patient Range
Constant flow Max. Tidal Apnea
during insp. volume alarm
exp. time flow
Airway Pressure
Settings
Airway Pressure
Monitoring
Indications shown in the Bargraphs
Servo Trigger System
Trigger sensitivity
PEEP
Respiratory pattern
Settings
Respiratory pattern
Inspiratory rise time

resp. cycle %
Respiratory pattern
Monitoring
I:E – ratio
Volume
Settings
Volume
Monitoring
O2 concentration
Alarms and messages
Pause hold
Set parameter guide, SPG
Patient safety
Airway pressure

! Upper Pressure Limit


! “AIRWAY PRESSURE TOO HIGH”
! When airway pressure is 15 cmH2O above set PEEP level for more
than 15 seconds
! “HIGH CONTINUOUS PRESSURE”
! In PRVC/VS mode when the internal pressure level reaches 5 cmH2O
below set Upper Pressure Limit an alarm is generated
! “PRESSURE LIMITED”
! If airway pressure should rise more than 5 cmH2O above set Upper
Pressure Limit
! Safety valve shall open
O2 concentration

! Preset alarm limits


±6 vol% (digital)
±10 vol% (analog)

! Default setting (min)


18 vol% (digital)
16 vol% (analog)

Alarm memory delayed 45 seconds

When Fuel Cell is disconnected Alarms & Messages display shows:


“O2 SENSOR”

During Oxygen Breath the audible alarm is muted


Apnea

! Apnea delay time


! Adult 20 s
! Pediatric 15 s
! Neonatal 10 s
Gas supply

! Pressure range
2 (-5%) – 6.5 (5%) bar

! Gas failure of one gas


! Remaining gas takes over

! Both gases fail


! Safety valve is activated
Battery

! Fully charged
Approx.24 V
! Alarms - below
+23 V “LIMITED BATTERY CAPACITY LEFT”
+21 V “NO BATTERY CAPACITY LEFT”
+19.5 V “POWER FAILURE ALARM”

! Charging time of empty battery approximately 6 hours


Expiratory Minute volume

Default settings

ADULT 0.3 l/min (digital)


0.2 l/min (analog)

PEDIATRIC 0.3 l/min (digital)


0.2 l/min (analog)

NEONATAL 0.06 l/min (digital)


0.05 l/min (analog)

Analog alarm system is activated when set alarm limits are passed by 10 %
Connectors
Communication
Ventilation Modes
Ventilation modes
Controlled – Supported – Spontaneous
Pressure Control mode
Pressure Controlled ventilation (PC)
Inspiratory rise time with Pressure regulation
Volume Controlled ventilation (VC)
Insp. rise time with Flow regulation
Volume control

Patient initiated flow increase during inspiration

! Patient can increase the insp. flow when he generates


an insp. pressure of 2 cmH2O below set PEEP level

! The insp. time is terminated when the flow drops


below the preset or to 5% of its peak value

! If the preset insp. tidal volume has not been achieved


the ventilator delivers the remaining tidal volume
Pressure Regulated Volume Control
Pressure level regulation – Principle
Volume support

Inspiratory pressure is kept constant

The inspiratory flow is decelerating

In case of apnea there is an automatic back-up with PRVC


SIMV (VC) + Pressure Support
SIMV (PC) + Pressure Support
Pressure Support
CPAP
Servo Ventilator 300A – Automode
Servo Ventilator 300A – Automode
Interactive Ventilation
Servo Ventilator 300A – Automode
Pressure Control ⇔ Pressure Support
Servo Ventilator 300A – Automode
Patient initiated expiration in Pressure Control mode
Servo Ventilator 300A – Automode
Volume Control ⇔ Volume Support
Servo Ventilator 300A – Automode
PRVC ⇔ Volume Support
Construction
Panel section – disassembling
Panel section – Inside view
Panel section – Potentiometers, Selectors
Micro-processor Module (MM) – PC 1588
Micro processor module – PC 1588
Plug-in electronics
Control section – Top lid
Control section –
Computer Interface board PC 1587
Control section – PC boards
Interconnection cable - Disassembling
Patient unit
Basic principle
Gas module
Gas module
Inspiratory flow
Inspiratory pressure transducer
Safety valve
O2 cell
Expiratory flow transducer
Expiratory pressure transducer
Expiratory valve
Expiratory valve solenoid
Expiratory flap valve
Pneumatic section - Disassembling

Caution!
Pneumatic section –
Internal Disassembling
Pneumatic section
Pneumatic section
Power section - Disassembling
Power section
Power section
Calibration

Refer to Operating Manual


Calibration – Trimmer location
Function check

Refer to Operating Manual


Servo Feedback System
General feedback control system
Servo feedback control system
Flow regulation
Servo feedback control system
Pressure regulation
Servo feedback control system
Expiration
Principle Diagram
Principle diagram
Cleaning and Maintenance

Refer to Operating Manual


Cleaning
Servo Ventilator 300
Cleaning
Servo Ventilator 300
Overhaul
1000 hour overhaul kit for SV 300

Part No.: 62 04 296 E380E


3000 hour overhaul kit for SV 300

Order No.: 61 50 929 E380E


Bi-Phasic Ventilation Module
Servo Screen 390
Servo Screen 390
Connectors
Servo Screen 390
Patient-related information
Servo Screen 390
Curve – waveforms and loops
Servo Screen 390
Curve – loop
Servo Screen 390
Curve – waveforms
Servo Screen 390
Numerics
Servo Screen 390
Trend
Servo Screen 390
Interpretive information
Servo Screen 390
System set-up
Servo Screen 390
Service
CO2 Analyzer Module
for Servo Screen 390
Servo Ultra Nebulizer 345
Nebulization principle
Servo Ultra Nebulizer 345
Patient unit
Servo Ultra Nebulizer 345
Patient unit
Servo Ultra Nebulizer 345
Filling buffer water
Servo Ultra Nebulizer 345
Filling medication
Servo Ultra Nebulizer 345
Start/Stop
Servo Ultra Nebulizer 345
Mechanical build-up
Servo Ultra Nebulizer 345
Gas flow through the system
Servo Ultra Nebulizer 345
Block diagram
With Battery Pack
With Compressor Compact
Front panel cover for SV 300
Support Arm 176
System SV 300
System SV 300
with NO/NO2 monitor
NO/NO2 monitor