Professional Documents
Culture Documents
Abbott: CELL-DYN 1800 Automated Hematology Analyzer Service & Support Manual
Abbott: CELL-DYN 1800 Automated Hematology Analyzer Service & Support Manual
CELL-DYN 1800
Automated Hematology Analyzer
Service & Support Manual
ABBOTT CD1800 SM
Theory of Operation ............................................................................................................................................................................ 4
System Overview.................................................................................................................................................................... 4
Major Subsystem Descriptions........................................................................................................................................... 6
Circuit Descriptions ............................................................................................................................................................. 15
Signal Processor Module (SPM)....................................................................................................................................... 18
Cell Count Module (CCM) ................................................................................................................................................... 21
Troubleshooting................................................................................................................................................................................. 31
Troubleshooting Charts...................................................................................................................................................... 33
Raw Data Description.......................................................................................................................................................... 40
CCM On-Board Diagnostic LEDs...................................................................................................................................... 41
CPU Hardware/Software Configuration .......................................................................................................................... 44
Service Special Commands............................................................................................................................................... 45
Sample Probe Description ................................................................................................................................................. 49
CELL-DYN 1800 Error Messages...................................................................................................................................... 60
Software Commands and Sequence ............................................................................................................................... 72
Engineering Drawings and Schematics ............................................................................................................................................. 76
CELL-DYN 1800 PCB Reference .................................................................................................................................................... 77
Removal & Replacement .................................................................................................................................................................. 96
Service Equipment Required............................................................................................................................................. 96
Covers (CD18-A1) .................................................................................................................................................................... 97
Flow Panel (CD18-B1)............................................................................................................................................................ 105
Fluid Power Supply (CD18-C1).............................................................................................................................................. 115
Syringe Assembly (CD18-E1)................................................................................................................................................. 119
RR-E1.04 Sample Syringe Driver Assembly .................................................................................................................... 125
RR-E1.06 Lyse Syringe Driver Assembly..................................................................................................................... 127
Electronics / Card Cage (CD18-F1) ........................................................................................................................................ 129
RR-F1.01 PAM (Pre-Amplifier Module) ............................................................................................................................. 129
RR-F1.02 MPM (Motor Processor Module) Board .......................................................................................................... 131
RR-F1.03 CDM (Cable Distribution Module) Board ........................................................................................................ 133
RR-F1.04 Hard Disk Drive................................................................................................................................................... 135
RR-F1.05 Floppy Disk Drive ............................................................................................................................................... 138
RR-F1.06 Signal Processor Module (SPM)...................................................................................................................... 140
RR-F1.07 Cell Count Module (CCM)................................................................................................................................. 142
RR-F1.08 Data Link Adapter (DLA) ................................................................................................................................... 143
RR-F1.09 Single Board Computer (SBC) ......................................................................................................................... 145
RR-F1.10 Card Cage Backplane PCB .............................................................................................................................. 147
LCD/Keyboard (CD18-G1)..................................................................................................................................................... 149
Power Supply (CD18-H1).................................................................................................................................................... 151
Verification Procedures ................................................................................................................................................................... 154
VP-01 Preparation for Alignment and Verification ................................................................................................................. 154
VP-01 Preparation for Alignment and Verification..................................................................................................... 154
Test Equipment and Supplies Required....................................................................................................................... 154
Preparation for Alignment/Verification ......................................................................................................................... 155
Order of Alignment/Verification...................................................................................................................................... 156
VP-02 Decontamination .................................................................................................................................................... 157
VP-03 Vacuum and Pressure Adjustments.................................................................................................................. 158
Regulator Alignment.......................................................................................................................................................... 159
Pressure Adjustment (0.5 psi)......................................................................................................................................... 160
Pressure Verification (High) ............................................................................................................................................ 161
Vacuum Adjustment (8 inch) ........................................................................................................................................... 161
VP-04 Metering System Timing Adjustments - RBC and WBC .............................................................................. 162
Metering Timing Fault Report.......................................................................................................................................... 163
RBC Metering System Timing Adjustment .................................................................................................................. 165
WBC Metering System Timing Adjustment ................................................................................................................. 166
VP-05 CMOS Setup Verification/Adjustment............................................................................................................... 167
VP-06 Card Cage Backplane Test Points ..................................................................................................................... 171
VP-07 Cable Distribution Module Test Points............................................................................................................. 172
VP-08 Pre-Amplifier Module (PAM) Adjustment ......................................................................................................... 173
VP-09 Signal Processor Module (SPM) Verification/Adjustment ........................................................................... 176
RBC Gain .............................................................................................................................................................................. 178
RER Adjustment ................................................................................................................................................................. 179
WBC Gain ............................................................................................................................................................................. 182
PLT Gain ............................................................................................................................................................................... 184
VP-10 Diluent and Sample Verification/Adjustment ........................................................................................................ 187
Diluent Volume Verification ............................................................................................................................................. 188
Sample Volume Verification ............................................................................................................................................ 188
VP-11 Stepper Motor Power Test and Verification .................................................................................................... 189
VP-12 Sample Probe Alignment Check ........................................................................................................................ 190
VP-13 Serial Transmit to LIS Verification..................................................................................................................... 191
VP-14 Hard Disk Drive Setup and Verification............................................................................................................ 194
VP-15 Software Installation/Upgrades .......................................................................................................................... 197
Theory of Operation
System Overview
The CELL-DYN 1800 Automated Hematology Analyzer is a complex system. Analyzer performance depends on several
components that together make up the complete hematology system. The system is comprised of the following components
and subsystems:
Flow Panel
The Flow Panel consists of tubing, solenoid valves, and other hardware components used for sample aspiration, dilution,
measurement and waste removal.
The Fluid Power Supply contains the vacuum and pressure pumps, accumulators, waste bottles, and associated solenoids
and hardware.
The Reagent Inlet Panel provides connections for incoming reagents and outgoing waste. The Lyse solenoid is also mounted
on this panel.
The Syringe Drive Assemblies include the Sample Syringe for aspirating samples, the Diluent Syringe for supplying Diluent
throughout the Flow Panel, and the Lyse Syringe for dispensing Lyse to the WBC transducer for the HGB measurement
process.
The electronics card cage, with associated PCBs, provides command and control signals for the various electronic
components of the instrument. This assembly contains the Backplane PCB, Cell Count Module, Signal Processor Module,
Data Link Adaptor and Single Board Computer.
The LCD Display Screen provides a visual data display and the keyboard provides data input by the operator.
The AC Regulator PCB provides power to the Backplane PCB (for use with the analog circuitry on various PCBs). It also
provides power to the Pump Relay Module.
The +28VDC Switching Power Supply provides power for the stepper motors via the Motor Processor Module. It also
provides power to initially energize solenoids.
To aid in understanding the overall system, the electronic modules are divided into the following major functional
subsystems:
The purpose of this subsystem is to interface the user data, control data, measurement data, and system status data in the
system. This data is connected via four independent data busses:
See the figure below for a diagram showing the data connections.
The CCM functions as the master controller with all system functional commands residing in firmware (PROM). The CCM
sends control data and receives status data from the CDM (Cable Distribution Module).
The CCM provides current control to the von Behrens RBC and WBC Transducers and the two metering PCBs and serves as
the system's analog voltmeter for use in converting the HGB signal. Data is written and read via the CCM/CDM and
CCM/MPM data buses.
The CDM (Cable Distribution Module) acts as a controller for the solenoids and also interfaces data from various system
sensors.
The MPM (Motor Processor Module) acts as a controller for all Stepper Motor Drive PCBs.
Measurement Subsystem
The measurement subsystem provides detection, amplification, and processing of the signals from the von Behrens
RBC/PLT Transducer, von Behrens WBC Transducer, and HGB Flow Cell. RBC/PLT and WBC metering is also included in
this subsystem.
The PAM (Pre-Amplifier Module) supplies constant current for the von Behrens RBC/PLT and WBC Transducers and HGB
LED voltage.
The RBC/PLT and WBC cell pulses are input to the PAM where they are amplified and routed to the SPM (Signal Processor
Module).
When the SPM receives signals from the RBC/PLT and WBC, the following occurs:
• The RBC/PLT signal is amplified (gain) and split into independent RBC and PLT signals.
• The WBC signal is amplified (gain).
• The RBC signal is routed to the cell editing circuitry.
• Cell editing is performed on the RBC signal to eliminate invalid RBC pulses.
The SPM discriminates cell size by converting pulse height to a proportional digital value. The amplitude of each valid pulse
is measured by a fast A/D converter then sent across the data bus to the CCM.
The A/D data for RBC, PLT, and WBC are individually divided by the CCM into 256 discrete size channels. The cell count in
each channel is accumulated in discrete memory locations and is used to generate count data, percentage data, and
histogram data for RBC, PLT, WBC, and other derived parameters.
Signals from the upper and lower detectors on the RBC/PLT and WBC metering PCBs are converted to TTL levels by
comparators on the CDM. The signals are then routed to the CCM where they are used to control RBC/PLT and WBC
sample timing.
The HGB analog signal from the HGB Flow Cell is captured by the PAM where it is amplified and routed to the CCM. The
HGB signal is then measured and converted to a digital format by a voltmeter-A/D converter.
Solenoid control commands reside in firmware on the CCM. These commands are sent to the CDM where they are
multiplexed to the appropriate SDM (Solenoid Drive Module). The SDM then provides the current to open and close
individual drive solenoids.
Stepper Motor commands are handled in much the same manner as described above. However, the final multiplexing of the
Stepper Drive PCBs is controlled by the MPM.
There are two pressure pumps and one vacuum pump in the CELL-DYN 1800 System. These pumps are described as
follows:
• A pressure pump provides air to bubble-mix samples in the Pre-Mixing Cup and the mixing chambers of the von
Behrens RBC/PLT and WBC Transducers. A pressure regulator regulates the 0.5 psi in the pressure accumulator for
this process.
• An unregulated pressure pump provides air to push waste from the waste bottles inside the instrument to the waste
container attached to the instrument and to apply back pressure to clear the apertures in the von Behrens RBC/PLT
and WBC Transducers.
• An 8" Hg vacuum accumulator, vacuum sensor, and vacuum pump supply a constant vacuum to the entire system to
transport Diluent, Detergent, and Lyse throughout the flow system and to maintain a constant vacuum to the
RBC/PLT and WBC metering tubes. A vacuum regulator maintains a constant vacuum source to both metering
tubes.
See Solenoid, Motor Drive and Pump Block Diagram for a diagram of the solenoid and motor drive connections.
The figure below illustrates the major components of the User Interface Computer.
The Single Board Computer (SBC) PCB is a complete Celeron 850 Megahertz PC computer system with 128 Megabytes of
RAM that is self-contained on one board. It utilizes a PC compatible BIOS with DOS capability that directly interfaces with the
LCD Display Screen, Hard Disk Drive, Floppy Disk Drive, PS/2 Keyboard, one (1) Parallel Port and two (2) Serial Ports. It
connects directly into the Backplane PCB along with the Data Link Adapter (DLA) and utilizes the ISA bus and Backplane
PCB to communicate with one another.
The CELL-DYN 1800 contains two (2) Serial I/O Ports for transferring data to other computer systems:
COM2 Spare
The CELL-DYN 1800 contains one parallel I/O port for transferring data to a printer.
PC/2 Keyboard
The external 101 key PC keyboard is used to enter alphanumeric data, such as demographic information, into the instrument.
The touch pad (membrane) keyboard is located below the LCD Display Screen. The keyboard includes a row of eight (8)
unmarked rectangular-shaped keys corresponding to labels displayed at the bottom of the screen. These keys activate the
indicated function or display the indicated submenu.
An LCD Adapter, connected directly on the SBC PCB, drives the LCD Display Screen. The adapter supports SVGA 640 x
480 and 800 x 600 graphics modes.
The hard disk drive stores the User Interface Software program and the Patient Data Log.
Floppy Disk Drive
The 1.44MB 3.5" floppy disk drive is used for program installation and provides the QC (Quality Control) data
upload/download capability.
Printer
The User Interface Software supports Epson ESC-P or PCL-3 languages. The printer has its own buffer and is capable of
printing on 8.5" x 11" (letter size) or A4 paper size. The printer supports alphanumeric and graphics reports from stored data
and screen displays.
Speaker
The PC speaker or buzzer is controlled by software and is mounted on the SBC PCB.
When the system is turned on, the AC line is routed into the AC Regulator PCB and +28VDC Switching Power Supply. The
AC Regulator PCB automatically accommodates line voltages of 90 - 130VAC and 200 - 260VAC by sensing the input
voltage and utilizing an internal comparator bank and power transformer to produce the 120VAC necessary for the
subsystem's function. The Power Supply Assembly then provides an AC and DC voltage source to various components on
the CELL-DYN 1800 System.
The +28VDC Switching Power Supply provides the voltage source to the DC Regulator PCB, which the board then uses to
produce other voltages. The power supply cooling fan is thermistor controlled, which means that when the internal
temperature rises above 70°C, the fan operates at full speed. The fan is then turned off at 50°C and kept off until the
temperature rises above 70°C.
AC Regulator PCB
The AC Regulator PCB provides the 120VAC used by the Pump Relay Module (PRM) for vacuum and pressure pump
operation.
The ±12VDC (analog) is provided to the Backplane PCB, which is used by the Cell Count Module (CCM) and Signal
Processor Module (SPM). The ±12VDC is also provided to the Cable Distribution Module (CDM), Motor Processor Module
(MPM) and Pre-Amplifier Module (PAM) (for its circuitry).
The +100VDC is provided to the PAM for use in its constant current circuitry. In between the AC Regulator PCB and the PAM
is the Pre Amp Filter that is used to filter out noise.
DC Regulator PCB
The DC Regulator PCB provides +5VDC, ±12VDC (digital), +14VDC and +28VDC. The +5VDC is provided to the Backplane
PCB, which is used by the SPM, CCM, Data Link Adapter (DLA) and Single Board Computer (SBC). This voltage is also used
by the CDM, MPM, Hard and Floppy Disk Drives.
The ±12VDC (digital) is provided to the Backplane PCB for use on the CPU fan. The Hard and Floppy Disk Drives receive
+12VDC for their operation.
The +14VDC is provided through the CDM to the Solenoid Driver Modules (SDM) and is used to hold the solenoids closed or
open for normally closed solenoids.
The +28VDC is provided to the MPM for operating stepper motors and through the CDM to the SDMs to initially close
solenoids or open for normally closed solenoids. This voltage is also used for the system's internal cooling fans. The cooling
fans are thermistor controlled, which means that when the ambient temperature inside the instrument drops below 25°C, the
fans operate at half speed. Once the temperature rises above 25°C, the speed is increased linearly until it reaches 35°C,
when the fans operate at full speed.
The DC Regulator PCB also provides +5VDC and +14VDC to the AC Regulator PCB.
Circuit Descriptions
This section contains a description of the circuitry for the following printed circuit boards:
• Pre-Amplifier Module (PAM)
• Signal Processor Module (SPM)
• Cell Count Module (CCM)
• Cable Distribution Module (CDM)
• Solenoid Driver Module (SDM)
• Motor Processor Module (MPM)
• Stepper Drive Printed Circuit Board (SDP)
• Pressure/Vacuum Regulator Module (PVRM)
• Power Supply Assembly
• Pump Relay Module (PRM)
• Single Board Computer (SBC)
• Data Link Adapter (DLA)
Note
Refer to PAM PCB Diagram.
• Provides constant current control to the von Behrens RBC/PLT and WBC Transducers.
• Amplifies the initial RBC/PLT, WBC and HGB signals.
The constant current bias (+100VDC) is switched by U6, then routed to Q2 and Q3, which supplies constant current to the
von Behrens RBC/PLT transducers. The RBC/PLT current is adjusted with R72. Once received, the RBC/PLT signals are
initially amplified by U7, then routed to U5, where they are re-inverted and further amplified. The combined RBC/PLT signal is
then routed to the SPM PCB.
Transistors Q4, Q5 and associated circuitry provide constant current for the von Behrens WBC Transducer. The WBC
current is adjusted via R66. Once received, the WBC signal is initially amplified by U12, then routed to U11 where it is
re-inverted and further amplified. The WBC signal is then routed to the SPM PCB.
The output of the HGB Flow Cell is amplified by U1 and U2 prior to being routed to the SPM PCB. The HGB self test and gain
voltages are adjusted with R14 and R4 respectively.
Note
Refer to SPM Architecture.
SPM Architecture
EPLD (SPM)
The EPLD is an enhanced programmable logic device. The SPM EPLD is used to control data acquisition.
There are three state machines that run in the EPLD. One each is used for controlling data acquisition elements such as
Multiplexers and Peak/Hold Amplifiers. The inputs to the EPLD state machines are the outputs of the various Threshold and
Slope Detectors as well as Bubble and Area Comparators.
There is a data transfer state machine which controls data flow to the CCM.
Signal Processing
There are two signals from the RBC and WBC transducers that are processed through two basic data acquisition circuits.
The Platelet signal is derived from the RBC signal. There are three gain adjustments associated with each of these signals.
In addition to the three gain adjustments there is a integration adjustment for detecting pulses that are too long.
WBC Signal
• The WBC signal is received through a differential amplifier to reduce noise. There is a gain adjustment after the
differential amplifier but before the Test Signal injection point.
• The WBC signal is then "Baseline Restored" to remove DC components and Baseline fluctuations due to varying
duty cycle of the blood cells.
• A threshold detector signals to the EPLD the presence of valid pulses.
• Simultaneously the Bubble detector signals if the pulse is determined to be a bubble. The bubbles are much larger
than pulses from blood cells. The EPLD discards the information from the pulse if it is determined to be a bubble.
• The data from the pulse is processed on the falling edge of the Threshold Detector unless the Slope Detector senses
another rising edge before the falling edge of the Threshold Detector. The data is processed immediately if there is
another rising edge before the falling edge of the Threshold Detector.
• The WBC Held Peak is converted and sent to the CCM on a 15µS cycle. Average pulses from the impedance
transducers are 35µS.
• RBC MCV
• RBC Count/PLT
RBC MCV
RBC MCV is for RBC MCV data only. In this mode, an integrator is enabled to determine if the cell is moving through the
middle of the aperture. If the cell is not moving through the center of the aperture, the data is falsely high and the pulse is
abnormally long. The integrated signal is compared to a proportion of its height. If the pulse is too long for its height, it is
discarded.
RBC Count/PLT
RBC Count mode is for count information. During this run, Platelets are counted simultaneous with RBCs and no integration
is used. The Platelet signal is derived from the RBC signal by an adjustable gain.
• The RBC signal is received through a differential amplifier to reduce noise. There is a gain adjustment after the
differential amplifier but before the Test Signal injection point.
• The RBC signal is then "Baseline Restored" to remove DC components and Baseline fluctuations due to varying duty
cycle of the blood cells. At this point an additional gain stage is added to create the Platelet signal.
• A threshold detector(s) signals to the EPLD the presence of valid pulses. The Platelet Threshold Detector is used if in
RBC Count mode.
• If in Count mode, the Platelet Lower and Upper Threshold detectors are used. If the Upper Threshold Detector is
triggered, the pulse is a RBC and the RBC pulse is accumulated into the RBC Count histogram. If only the Lower
Detector is triggered, the PLT pulse is accumulated into the PLT histogram.
• The data from the pulse is processed on the falling edge of the Threshold Detector unless the Slope Detector(s)
senses another rising edge before the falling edge of the Threshold Detector. If this occurs, the data is processed
immediately.
• The RBC Held Peak or the PLT Held Peak is converted and sent to the CCM on a 15µs cycle.
Note
Refer to PAM PCB Diagram.
CCM Architecture
System Clock and Microprocessor Description
The CCM uses a 16 MHz crystal as a time base. The fundamental cycle time for the MC68HC11KW1 processor is 4 MHz.
The processor has many built in functions such as:
EPLD (CCM)
The EPLD is an enhanced programmable logic device and is used for address decode and histogram building functions. Its
program is used in coordination with the processor software.
LEDs
Note
Refer to CCM On-Board Diagnostic LEDs.
Histogram Memory
This data is the count information for the respective pulse height.
HGB Measurement
A 12 bit ADC with an input multiplexer is used to measure the HGB signal from the PAM. This data is read by the processor.
Along with the HGB measurement, various DC voltages are read by the 12 bit ADC as a diagnostic.
CDM Port
SBC Port
The SBC port is connected directly to microprocessor ports J and part of port A.
Note
Refer to Solenoid, Motor Drive and Pump Block Diagram.
The CDM communicates with the CCM via the CCM/CDM data bus at J2. Analog outputs of the Metering Modules are
converted to TTL levels by comparators (U12) and placed directly on the CCM/CDM data bus. Signals from the Pump Relay
board, Probe Position Switches, and Start Board (Touch Plate) are interfaced by Data Drivers (U5, U10).
Data is interfaced to the Solenoid Driver Modules via J32. This data is then multiplexed by One-of-Eight Decoders (U1, U2)
via J3, J4, J6, J7, and J9.
Vacuum and pressure control data is latched by U14 and routed to the Pump Relay Module via J11. Pump status signals
(Vac On, Pres On) are converted to TTL levels by U3 and placed on the data bus by U5.
LED drive signals are routed to the Start Board (Touch Plate) via J17. The start signal enters at J17 and is placed on the data
bus by U5.
Note
Refer to Solenoid, Motor Drive and Pump Block Diagram.
The purpose of the SDM is to provide drive current to the solenoids. Each SDM has eight Darlington drivers (Q1-Q8) which
are individually controlled by data bits (D0-D7) and data latch (U3).
There are two power modes available for each solenoid - activate (+28V) and hold (+14V). This is controlled by the Hi CLK
signal in conjunction with data bits (D0-D7) and the current control latch (U1).
Note
Refer to Solenoid, Motor Drive and Pump Block Diagram.
The MPM controls drive data to the Stepper Drive printed circuit boards and also provides self-test capability for motor
winding current. The MPM is comprised of the following major circuits:
• Microprocessor
• Program Control EPROM
• I/O Peripheral Interface Adapter (PIA)
• Direct Memory Access (DMA) Control
• Motor Phase Latches
• Motor Current Latches
• Motor Winding Self Test
Data communications between the CCM and MPM are controlled by I/O PIA (U6) and serial data is interfaced via ACIA (U2)
and Data Bus Connector (J1).
Phase data, motor direction, and step rate are stored in RAM (U7). This data is sent to the Motor Phase Latches under
control of the DMA Control circuitry, which consists of U11, U12, U15, U16, U18, U21 and associated circuitry. The data is
strobed into the appropriate Motor Phase Latch by ALG0 through ALG2.
The Motor Phase Latches U23, U26, and U29 provide phase data to the Stepper Drive printed circuit boards. Each is an 8-Bit
Addressable Latch which can control up to four Stepper Drive printed circuit boards and subsequently four Stepper Motors.
Four levels of motor current for each motor is controlled by the Motor Current Latches (U22, U25, and U28). Each latch can
control up to four stepper drive printed circuit boards. Data is strobed into the appropriate latch by WR0 through WR2.
The Feedback- and Feedback+ inputs at J3 through J14 are connected, via resistors on the Stepper Drive printed circuit
board, to the stepper motor windings. This allows the circuitry consisting of U30, U31, and U32 to monitor the winding current
during an internal self-test. These values can be read by the CCM to isolate a defective Stepper Drive or Stepper Motor.
Note
Refer to MPM section, blocks 20, 23, and 24 of Solenoid, Motor Drive and Pump Block Diagram.
The Stepper Drive printed circuit board consists of two PBL 3717 motor drive chips. Each chip drives a winding of the
Stepper Motor. Bits I0 and I1 are used to control four motor current levels:
• P0 - High Current
• P1 - Medium Current
• P2 - Low Current
• P3 - Current Off
Bits PH0 and PH1 control motor phase and, therefore, direction and step-rate (velocity). Feedback+ and Feedback- are used
to generate a motor self-test.
Note
Refer to Solenoid, Motor Drive and Pump Block Diagram.
Pressure (or vacuum) is sensed by a transducer that is internally configured as a Wheatstone Bridge. Transistor Q1and
resistors R4 and R5 are used to generate a stable reference voltage for the Wheatstone Bridge. The output of the
Wheatstone Bridge is partially amplified (U1-7), stabilized against long term drift (voltage follower U1-1) and made
offset-adjustable by R18 and associated resistors.
Maximum transducer sensitivity can only be achieved when the output is zero volts at TP-1 and when there is no pressure
differential across the transducer. To accomplish this, R18 is adjusted for zero volts when both transducer inlet ports (P1 and
P2) are open to atmospheric pressure.
In order to maintain the operating point of comparator U2-14 at the fixed 2-volt trip level, it is necessary to maintain the output
of U2-8 within a relatively narrow range. This is accomplished by making the differential amplifier (whose inputs are U1-10
and U2-10) adjustable by selecting 1 of 4 possible jumper positions. A stable reference point for the DC operating level of
U1-8 and U2-8 is established by U1-14 in conjunction with R6 and resistor network RP1.
Measurement of pressure in the range of approximately 0.5 lbs/sq. inch is accomplished by using transducer inlet port P1 in
conjunction with jumper setting A/B.
Vacuum pressure in the range of 8 inch Hg is accomplished by using inlet port P2 in conjunction with jumper setting C/D. The
regulation point for either vacuum or pressure is established by the setting of potentiometer R16.
When the output of comparator U2-14 goes positive, the collector of Darlington transistor Q2 is pulled to ground, thereby
turning on either the pressure or vacuum pump. When the pumps are running, LED DS1 is lit and stays lit until either the
pressure or vacuum increases past the hysteresis point established by R8 of U2-14.
The output of the pressure/vacuum regulator can be inhibited by a logic low at J1-1. This completes the circuit description.
Note
Refer to major subsystem descriptions of AC and DC Power Distribution Subsystem.
The PRM provides drive to the vacuum and pressure pumps, via three Solid State Relays: K1, K2, and K3.
The Single Board Computer (SBC) is designed around a Celeron 850MHz microprocessor and connects directly into the
Backplane PCB. The SBC receives power from the Power Supply Assembly via the Backplane PCB and receives status and
measurement data from the CCM through the DLA. The board contains EPROM, CMOS, RAM, input/output circuitry for the
interface ports, disk drives (hard and floppy) circuitry, LCD Display Screen Interface, and PS/2 Keyboard connectivity.
The Data Link Adapter provides interfaces from the SBC to both the CCM (Cell Count Module) and the Touch Pad
(membrane) keyboard. These two interfaces function independently under user interface software control. However, some of
the circuits on the DLA board are shared. The DLA uses an 82C55 PPI (Parallel Peripheral Interface) IC. Refer to the figure
below for an illustration of the DLA board.
DLA (Data Link Adapter) Block Diagram
Interface to CCM
A 20-pin ribbon cable connects the DLA to the CCM. This interface is a bi-directional, parallel interface that is
software-controlled at both ends. Data is transferred in 8-bit bytes on 8 data lines (D0 - D7) in one direction at a time. The
SBC (Single Board Computer) initiates an inquiry message handshake on a periodic basis or when it has a command to send.
The CCM responds by sending data or replying to the command sent.
The interface is controlled cooperatively by the UIC and CCM according to the state of the handshake signals. Refer to Cell
Count Module (CCM). The UIC always sends data first. The CCM responds by sending data back (if any is available). Thus,
the 82C55 switches its A-port from input to output and back to input during every message transaction.
The key handshake signals are REQ1 and REQ2. REQ2 going low initiates the communication protocol. The CCM responds
by bringing REQ1 low. After the DLA has sent its data, it brings REQ2 high. Then the CCM sends its data.
Each byte received by the DLA generates an IRQ. The application software responds to the IRQ by putting the byte into a
buffer. When the CCM brings REQ1 high, the communication is complete; the DLA returns to an idle state, and port A is set
to input. Both REQ1 and REQ2 remains high until the next message/data transfer.
The two LEDs on the printed circuit board, DS1 and DS2, indicate the active state of REQ2 and REQ1, respectively. They
should always be flickering when the application software is running because the User Interface Software program is
constantly polling the CCM to check its state. (There are some exceptions to this during power up and certain diagnostics/test
modes.)
The I-O address of the DLA board is assigned by four jumpers. The default I-O address (pins 3 - 6 on S1 hard-wired) is: [off
off on off], 340 hexadecimal. The DLA interrupt level is assigned by a jumper.
W# IRQ #
W1 IRQ 5 *
W2 IRQ 10
W3 IRQ 10
W4 IRQ 11
W5 IRQ 12
W6 IRQ 15
A 14-pin ribbon cable connects the DLA to the key panel. The key panel is also polled by the User Interface Software
program. This interface is not interrupt-controlled.
The keyboard is interfaced as a parallel switch matrix circuit with 4 lines out and 8 lines in. Four scan rows are driven active
one at a time via an LS175 latch. Then the columns of the key panel matrix are read from port B of the 82C55. The software
interprets a low active signal as a key is pressed.
The DLA board must be plugged into a 16-bit slot. There are no adjustments on the DLA board.
Troubleshooting
Diagnostics Menu Usage
Utilization of the DIAGNOSTICS Menu enables the operator and/or service representative to identify and correct both
operator-correctable and service-correctable faults. When the computer senses a fault, the message <NOT READY: SEE
DIAGNOSTICS> displays in the System Status Box. The following keys are available in the DIAGNOSTICS Menu.
Level One
• INITIALIZATION: Used to perform an Initialization cycle: returns movable components to home position and
performs internal self-tests.
• RAW DATA: Used to display raw measurement data for the last specimen.
• COUNT TEST: Used to run specimens without returning to RUN Menu and display Raw Data.
• MORE: Used to display additional functions (levels).
• PRINTER OUTPUT: Used to toggle printer output ON and OFF.
• HELP/ERROR: Used to display help information regarding the diagnostics menu screens. The fault log can also be
accessed through this key function.
• MAIN: Used to return to Main Menu.
Level Two
• WBC HISTOGRAM: Used to display WBC count and histogram data accumulated in each of 256 size channels.
• RBC HISTOGRAM: Used to display RBC histogram data accumulated in each of 256 size channels.
• PLT HISTOGRAM: Used to display PLT count and histogram data accumulated in each of 256 size channels.
• SMOOTHING ON/OFF: Used to toggle histogram display status. With Smoothing Off, only raw counts are displayed.
With Smoothing On, channels are numbered, data is normalized and the number of the peak channel displays.
Smoothing On/Off (example) gives an example of Smoothing On/Off.
• MORE: See Level One description.
• PRINTER OUTPUT: See Level One description.
• HELP/ERROR: See Level One description.
• MAIN: Used to return to Main Menu.
Level Three
• PROBE HOME: Moves Sample Probe up and above RBC Cup. Displays probe assembly sensor status.
• PROBE UP: Moves Sample Probe up. Displays probe assembly sensor status.
• MORE: See Level One description.
• PRINTER OUTPUT: See Level One description.
• HELP/ERROR: See Level One description.
• MAIN: Used to return to Main Menu.
Level Four
• SYSTEM STATUS: Used to display all pending alarms.
• FAULT REPORT: Used to display all pending faults or warnings.
• SERVICE HEX CODES: Hex system codes. Not used for operator or service troubleshooting.
• SERVICE DEC CODE: Used to initiate individual actions in the CELL-DYN 1800 hardware and software.
• MORE: See Level One description.
• PRINTER OUTPUT: See Level One description.
• HELP/ERROR: See Level One description.
• MAIN: Used to return to Main Menu.
A detailed list of all faults generated by the CELL-DYN 1800 System software and hardware is contained in CELL-DYN 1800
Error Messages. The fault classifications reported in the Fault Report primarily contains data pertaining to the last CCM fault.
If a fault occurs, pressing the [HELP/ERROR] key immediately displays the [FAULT LOG] in the DIAGNOSTICS Menu. This
log may contain up to 16 faults, with the most current fault at the top of the list. An alternative procedure is to go to the MAIN
MENU and press [DIAGNOSTICS]. In this case, the [FAULT REPORT], not the Fault Log, is immediately displayed.
The Fault Log can be viewed from any of the menus, except SETUP. To view the Fault Log, enter the desired menu, followed
by [HELP/ERROR] and [FAULT LOG]. The system displays up to 16 past faults. From the MAIN MENU, press
[DIAGNOSTICS] followed by [MORE] three (3) times and [FAULT REPORT] to display the FAULT REPORT screen. A
display of <NO FAULTS OR WARNINGS PENDING> indicates that all faults have been cleared.
A list of symptoms, probable causes, and corrective actions for the most common problems encountered on the
CELL-DYN 1800 System is given in the Troubleshooting Chart. The probable causes and corrective actions for each
symptom are arranged in descending order from most likely to least likely. When troubleshooting a problem, start with the
most likely cause first.
If possible, thoroughly verify that a component is defective before replacement. Some problems can be verified visually, but
other problems may require a measurement tool such as a DVM (Digital Volt Meter).
When troubleshooting DATA PROBLEMS, only the measured parameters RBC, PLT, WBC, HGB, and MCV should be used
for reference. Using the calculated parameters can become confusing when trying to isolate a problem.
When troubleshooting CLOG AND FLOW ERROR PROBLEMS, refer to VP-04 Metering System Timing Adjustments - RBC
and WBC for the MIN and MAX specifications for the RBC and WBC Upper (T1) and Lower (T2) times.
Troubleshooting Charts
Flash BIOS or Connect a computer monitor to the top of the SBC PCB. Power ON
CMOS settings instrument and open CMOS setup screen. Check that Advanced
LCD Display Screen reset to Default Chipset Setup is configured correctly. Refer to VP-05 CMOS Setup
Blank/Solenoids OK Verification/Adjustment.
Note
Defective Backlight Replace Backlight Interconnect PCB
Do not measure voltage
Interconnect PCB
on backlight inverter
PCB. Measuring voltage Defective LCD Replace LCD Interconnect PCB
can damage PCB. Interconnect PCB
Solenoid 3-1, 3-3, 3-4 Tubing Check Solenoid 3-1, 3-3, 3-4
Diluent Empty Restriction Check Lines
Leak 8" Hg Check for air leaks in Fluid Power Supply and
Flow Panel
Leak 0.5 Psi Check for air leaks in Fluid Power Supply and
Flow Panel
All Results Are "0" No ±12VDC (Analog) Replace Power Supply Module
Replace CCM
Reference lower than Check for bubbles in Detergent line and Solenoid 2-6
sample reading
Inadequate Probe Cleaning Check associated wash block tubing and hardware
Imprecision on all Parameters
Perform Sample Volume Verification
Imprecise Sample Aspiration
Check associated sample aspiration tubing and hardware
WBC "R" Codes, Reagents OK WBC Gain Perform VP-09 Signal Processor Module (SPM)
Verification/Adjustment
Replace SPM
Defective PCBs
Replace CCM
Check power
High Electrical Backgrounds Environmental
Check overhead lighting
Metering Tube position Top of Metering Tube (tapered edge) should be visible on
top of upper DET sensor
Miscellaneous Problems
From the MAIN MENU, press [DIAGNOSTICS] followed by [RAW DATA]. The [RAW DATA] key displays raw data obtained
from the last count cycle.
When a single count is done, all data is contained in the first column. When a PLT recount occurs, data from the first cycle
displays in column #2 and data from the recount displays in column #1.
Note
An example of a raw data report displays in Raw Data (example).
The seven LEDs on the CELL-DYN 1800 CCM can reveal much about the fundamental CCM and overall machine state. In
general, the LEDs indicate whether the CCM is in a normal functioning mode or in a fault state, and in either case, help to
characterize the CCM state. Also, LED2 gives some information about the state of the SBC.
The CCM tests itself on power-up. These fundamental tests include MC68HC11KW1, RAM, and SPM interfaces. If any test
fails, the CCM attempts to execute an endless loop routine which flashes the green LED on the board. Also, it places a 4-bit
fault code into the adjacent yellow LEDs (see Power-on LED Patterns - Fault States on Startup).
Note
Refer to CELL-DYN 1800 CCM for location of LEDs on PCB.
• LED7, a green LED, should not be not flashing after the CCM has successfully completed its internal power-on
self-check diagnostics, otherwise if it is flashing slowly (~1Hz.), there is a fundamental CCM fault.
• LED5 and LED6, for CER and CEW, indicate the state of the CCM firmware generated signals CER (Count Enable
Red) and CEW (Count Enable White). These signals enable cell counting. When the associated LED is on, the DMA
cell counting circuitry is active.
• LED3 and LED4, for WCP and RCP, are on during the metered count time intervals, LED3 for the white count time
interval and LED4 for the red count time interval.
• LED2 is driven by the signal NREQ1, and indicates the UIC/CCM communication activity. This LED state directly
relates to the hi/lo state of NREQ1. When LED2 is flickering, which should always be the case during normal system
operation, it indicates that the UIC/CCM communications link is active. (There is a corresponding LED on the DLA
(UIC comm. board) that will also be flickering in sync with LED2; it is controlled by the DLA output signal NREQ2.)
• LED1 is used to indicate that a self-test is in progress. This indicates that the tests for the pulse processing A to D
circuitry on the SPM with the pulse counting circuitry on the CCM are active.
0 ON ON ON ON Power on
1 RAM testing
3 S3 BUFFALLO 10 E2 DGND
5 J1 CDM 12 E3 DGND
6 J3 SPM 13 E1 DGND
7 J5 DLA 14 LED 1
Detailed information on the CELL-DYN 1800 System Interface Specifications is available on the WWCS Intranet website.
CMOS Setup
The CMOS Setup contains all the information needed by the Basic Input/Output System (BIOS) to establish proper
communications between the single board computer (SBC) and the various computer system devices. Refer to VP-05 CMOS
Setup Verification/Adjustment:
Special Function
Probe Check
There are two probe check functions activated by softkeys in the DIAGNOSTICS menu.
• The [PROBE UP] softkey moves the probe up and maintains position without rotational motion. (The [PROBE
DOWN] softkey then displays to restore the probe to the down position.)
• The [PROBE HOME] softkey places the probe over the RBC/PLT transducer. (The [PROBE DOWN] softkey then
displays to restore the probe to the down position.)
Note
Neither procedure puts the probe in the STANDBY position (on the left).
Discussion
Several commands are available to initiate individual actions in the CELL-DYN 1800 System hardware and software. These
commands are used for troubleshooting and/or alignment when a single action is desired or required to be repeated several
times.
The special command mode resides in the DIAGNOSTICS Menu. From the MAIN Menu, press [DIAGNOSTICS] followed by
[MORE] three (3) times, then [SERVICE DEC CODE]. When this softkey is pressed, the message Test Select --- FOR
SERVICE USE ONLY --- displays.
A command can now be entered. Pressing the Enter key on the keyboard initiates the action. Only one command can be
entered at a time and [SERVICE DEC CODE] must be pressed before a command is entered.
Note
Use only the commands listed in DEC Service Commands and always verify that the correct number has been
entered before initiating the action. Use only those numbers listed in DEC Service Commands. Other numbers may
refer to engineering commands which are not used in the field and which may cause damage if used improperly. Be
fully aware of the purpose of any of the DEC Service Commands before using them. This is a direct-activation
method which should be used with caution because the physical state of the CELL-DYN 1800 System may not be in
agreement with the function to be performed. After using service commands, always re-initialize the system by
turning the power OFF then ON again or by pressing the [INITIALIZATION] key in the DIAGNOSTICS Menu to
ensure the instrument is in the proper configuration for normal operation.
DIAGNOSTICS Menu Service Code Function List
When the [SERVICE DEC CODE] key is pressed, the (Enter number (currently, 102):____ prompt displays.
The number above corresponds to the decimal code for the last code entered.
DEC Service Commands lists the decimal-coded (DEC) service commands that can be invoked by pressing the [SERVICE
DEC CODE] key in the DIAGNOSTICS Menu and entering the appropriate number.
07 NOT USED
08 NOT USED
09 NOT USED
11 NOT USED
16 NOT USED
17 NOT USED
18 NOT USED
20 mini-wash
22 NOT USED
23 NOT USED
24 NOT USED
25 NOT USED
26 NOT USED
33 NOT USED
34 NOT USED
36 NOT USED
41 NOT USED
47 platelet recount
48 initialization (homing)
54 daily shutdown
56 gain adjust
61 dispense 10 ml saline
63 clean-for-shipping
69 NOT USED
72 NOT USED
73 NOT USED
77 NOT USED
78 NOT USED
81 NOT USED
88 probe down (when finished, operator should initialize the instrument to place the probe in the home
position)
129 motor power level test (see Service DEC Code 129)
Note
Certain commands are not sent to the CCM when the system is in an interlock state, such as STANDBY or
UNINITIALIZED.
The CELL-DYN 1800 can be pre-set to do a specified number of RUN cycles without user intervention. This capability
applies only to normal RUN Count Test, Pre-Dilute RUN, (PRE-DIL TEST), Gain Adjust (GAIN ADJ), and Electrical
Background (ELEC BKGD). This capability helps reduce test time for the instrument. The following entry screen displays
after entering code 999:
The motors that enable the Sample Probe to move up/down and to rotate are stepper motors which are under direct
computer control. Since there is no direct positional feedback sent to the computer, position switches are employed to verify
critical positions during normal operation. It is important to understand that these switches only verify and do not control the
movement of the Sample Probe.
In the DIAGNOSTICS Menu, Service DEC Codes 128, 129, and 130 allow the service representative to control and exercise
all stepper motors in the CELL-DYN 1800 System. This description focuses on the Probe Up/Down Motor (B/2) and the
Probe Rotate Motor (C/3) which control the movement of the Sample Probe.
These commands reside in the SERVICE DEC CODE screen of the DIAGNOSTICS Menu and are used to test, control, and
exercise CELL-DYN 1800 stepper motors. A description of each of these three commands is given.
The Motor Power Test should be run whenever a problem is suspected with any assembly that is driven by a stepper motor.
To MPM: {I }
To MPM: {pD32}
To MPM: {C1}
inp: 0415
A report (Motor Power Test (example)) automatically displays and can be printed. Refer to VP-11 Stepper Motor Power Test
and Verification.
Note
Press the [INITIALIZE] key before leaving the DIAGNOSTICS Menu.
0) Full Power
1) Medium Power
2) Low Power
3) Off
This code tests mechanical assemblies at various power levels or to remove idle power so the mechanism can be more
easily moved or checked manually. The Motor Power Level Test (example) screen displays after entering code 129 (press
the ENTER key after each entry):
Note
After the entries are made, a message, such as Motor "A" set to running power of 1 and idle power of 3,
displays.
This code allows the direction, speed, and number of steps to be set when exercising a stepper motor.
The Motor Check (example) screen displays after entering code 130:
The table below contains information on the motor designation, command and direction of the motor to be tested. Motor
Speed Commands lists the motor speed commands to determine the speed of the motor being tested. Both tables are
needed to properly test the motor.
0 Down/Aspirate
A/1 Sample Syringe
1 Up/Dispense
0 Up
B/2 Probe Up/Down
1 Down
0 Down/Aspirate
D/4 Diluent Syringe
1 Up/Dispense
0 CCW/Dispense
E/5 Directional Valve 1 CW/Aspirate
F/6 Spare
G/7 Spare
0 Down/Aspirate
H/8 Lyse Syringe
1 Up/Dispense
1 50
2 75
3 283
4 300
5 166
6 200
7 250
8 10
9 151
10 222
11 25
12 182
13 100
14 125
15 91
16 67
17 111
Initialization Mode
The Initialization cycle places mechanical and electrical components in the "home" position, drains any liquid in the tubing,
Pre-Mix Cup, and the Mixing Chamber of the von Behrens RBC Transducer to the Waste System, then places the instrument
in the INITIALIZED state.
Homing a stepper motor is the process of setting up the initial position from which all future movement is referenced. In the
CELL-DYN 1800 System, this is accomplished by forcing the motor to move against a physical stop (Hard Stop). When the
mechanical assembly, driven by the motor, reaches the Hard Stop, the stepper motor electrically slips until it is forced to stop.
This mechanical position then becomes the zero reference position for the motor.
Operation:
1. The Sample Probe moves up at a fast speed until the Upper Switch (#2) is activated. It is then changed to a slow
speed, and homed against the Upper Hard Stop, which is the metal plate at the top of the Sample Probe Assembly.
2. The probe moves down six steps and the Upper Switch (#2) is checked.
3. The probe moves CCW at a fast speed until the Right Switch (#4) is activated. It is then changed to a slow speed,
and homed against the Right Hard Stop, which is the mounting bracket for Right Switch (#4).
4. The probe moves CW to the Pre-Mix Cup and Left Switch (#3) is checked. The probe then moves into the Pre-Mixing
Cup.
5. The probe moves up and Upper Switch (#2) is checked.
6. The probe moves CCW to center and down positions; and the Lower Switch (#1) is checked.
7. This completes the Initialization cycle.
Run Mode
The figure below illustrates the probe's movements during the RUN cycle.
1. When the Start Switch is pressed, 30 µL of sample is aspirated and Lower Switch (#1) is checked.
2. The Sample Probe then moves up to a position six steps from Upper Hard Stop, and Upper Switch (#2) is checked.
3. The probe moves CW to Pre-Mix Cup and Left Switch (#3) is checked.
4. The probe moves down eight steps and into the Pre-Mix Cup, where dispense, probe shake, and aspiration of RBC
sample takes place.
5. The probe then moves up to a position six steps from Upper Hard Stop, and Upper Switch (#2) is checked.
6. The probe moves CCW to the Mixing Chamber of the von Behrens RBC/PLT Transducer, stops three steps from
Right Hard Stop, and Right Switch (#4) is checked.
7. The probe moves down into the RBC/PLT Mixing Chamber and RBC sample is dispensed.
8. The probe moves up to a position six steps from Upper Hard Stop, and Upper Switch (#2) is checked.
9. After completion of the count cycle, the probe moves CW to center position.
10. The probe moves down and Lower Switch (#1) is checked.
11.This completes the RUN cycle.
From the MAIN MENU, press [DIAGNOSTICS]. The screen immediately displays one of the Fault Reports shown in Lower
Switch (#1) Fault Report, Upper Switch (#2) Fault Report, Left Switch (#3) Fault Report, and Right Switch (#4) Fault Report.
The message <SWITCH: 1 CHECK> indicates that Lower Switch (#1) failed when checked. The message <* NOT ON ANY
SWITCH *> indicates that none of the switches were activated when the failure occurred. Refer to Lower Switch (#1) Fault
Report.
The message <SWITCH: 2 CHECK> indicates that Upper Switch (#2) failed when checked. The message <* NOT ON ANY
SWITCH *> indicates that none of the switches were activated when the failure occurred. Refer to Upper Switch (#2) Fault
Report.
The message <SWITCH: 3 CHECK> indicates that Left Switch (#3) failed when checked. The message <ON SWITCH(ES):
2 TOP> indicates that Left Switch (#3) was activated when the failure occurred. Refer to Left Switch (#3) Fault Report.
Left Switch (#3) Fault Report
The message <SWITCH: 4 CHECK> indicates that Right Switch (#4) failed when checked. The message <ON SWITCH(ES):
2 TOP> indicates Right Switch (#4) was activated when the failure occurred. Refer to Right Switch (#4) Fault Report.
Right Switch (#4) Fault Report
The above conditions do not necessarily indicate that a switch has actually failed. They only indicate that the switch was not
read as activated when checked by the computer. A failure could also be caused by improper switch alignment, an electronic
hardware failure, or a mechanical hardware failure.
CELL-DYN 1800 Error Messages
The table below lists error messages on the CELL-DYN 1800 System.
(Status box)
Time-out at N seconds A CCM process initiated by the user took longer to complete than allowed (usually indicating
a failure of the CCM). The process ran approximately N seconds before the time-out
occurred.
Process Aborted A count test was stopped either by the user or because of a fault detected by the CCM.
Process Monitoring Aborted A process was stopped by the user (using the asterisk (*) key.
Printer Time-out The Printer Output option was ON and the printer did not print the requested report in the
expected time.
Code N is Invalid The user has entered a command for the CCM whose numeric value exceeds 127. The
value entered was N.
Cannot do this Function The user has attempted to issue a command to the CCM that cannot be executed because
of a pending fault condition.
WBC Meniscus Detection, During the most recent count, a meniscus was not detected or was detected at an
RBC Meniscus Detection unexpected time.
WBC Count Time-out (clog), During the most recent count, a Clog occurred.
RBC Count Time-out (clog)
CCM Pulse Height Memory During the most recent count, there was an overflow in one of the pulse-height arrays
Saturation Warning (histograms).
External Waste Full The external waste bottle has been filled.
Invalid Alarm Set A bit was set in the fault message from the CCM that has no valid interpretation.
*NOT ON ANY SWITCH* After some mechanical motion, a reading of all the position sensors indicates that none are
activated. (This message does not necessarily mean that a mechanical fault has occurred.)
Waste Overflow Into A reading of the sensor in the Waste Accumulator suggests that there is liquid in the
Accumulators accumulator.
Vacuum There was a vacuum failure during power-up or the instrument is unable to maintain vacuum
level while in the READY state.
Position Fault A mechanical assembly is not in the correct position for the most recent function to be
performed, as indicated by position sensors.
Sensor Fault - Internal Waste A Time-Out fault occurred in draining one of the waste bottles. This error is also associated
Empty with positive pressure.
Canceling Auto-Cycling This message displays when the user cancels Auto-Count Testing.
Invalid UIC Command Sent to The UIC sent a command to the CCM that it cannot interpret.
CCM
Error in Flow System Timing An error in the timing of a flow script has occurred. This occurs during instrument
initialization when the flow script takes more than three minutes to complete.
Histogram Memory Clear The CCM was unable to clear the pulse-height memory.
CCM Program, RAM Memory The CCM detected a failure in its RAM.
CCM/MPM Message Fault
Abnormal Time-out/No MPM attempt to send or receive motor or other command to or from MPM, or the MPM was unable
Incorrect Command to be
Sent to MPM
No Response from CCM The CCM is not functioning or the signal cable connecting the CCM and UIC is faulty or
disconnected. Turn the instrument OFF, check the CCM/UIC cable, then turn the instrument
ON.
Count Test
The [COUNT TEST] key in the DIAGNOSTICS Menu is used to run specimens and display Count Check data without
returning to the RUN Menu. Coded data relating to specific cycle functions, raw measurement, and flow count time are
displayed for use in troubleshooting or service.
Event Messages During Diagnostic Menu Count Test lists the event messages that are displayed during the Diagnostic Menu
Count Test.
Detergent Low Detergent is low as detected by reagent sensor in reagent inlet tube.
Diluent Low Diluent is low as detected by reagent sensor in reagent inlet tube.
Summary of Error Messages gives a complete listing of error messages available on the CELL-DYN 1800 System.
DOS Errors
Arithmetic overflow
Collection overflow
Disk full
Disk is write-protected
Disk read
Disk seek
Division by zero
Hardware failure
Heap overflow
Range check
Stack overflow
Range check
Stack overflow
Unknown command
Unknown unit
General
No data to print
Stopping
System Fault
Uninitialized
Normal Operator-Correctable
Detergent empty
Diluent empty
Lyse empty
Waste full
DLA Faults
Date
Battery failure or system date/time not set
Disk I/O
Ticket Printer
Data err
Inc. asp.
No Ticket Detected
QC Log
Data Log
Cannot do if uninitialized
Count overrange
Initialize
Communications
Break interrupt
Comm error
Framing error
Parity error
Re-transmit started
Re-transmit time-out
Receiver overrun
Transmit started
Unable to re-transmit
Printer Driver
Printer Fault
Printer Off-line
Printer Time-out
QC Log I/O
Help
Reagent Log
Data Log initialize failed, bytes short for creation of Data Log
X-B File
General Faults
Abnormal time-out / no MPM response
Count Overrange
Detergent sensor
Diluent sensor
Incomplete Aspiration
Invalid Data
Press [INITIALIZATION] key to clear fault and re-initialize meniscus detector: "True" sensed at metering start meniscus: not
detected during valid time interval
RBC Clog
WBC Clog
Fault Log
Lab ID Setup
Accessing DOS
There are two ways to access the DOS program from the CD1800 program:
It is recommended that the Cell-Dyn be put in Stand-by before exiting the program. Otherwise, turn the
instrument off after doing any service work while in DOS.
3. Press the Alt and d keys simultaneously. The DOS prompt C:\CD1800> displays.
4. Execute the desired DOS commands. Most commands can be executed. However, computer memory is limited
since the CD1800 program is still loaded.
It is recommended that the Cell-Dyn be put in Stand-by before exiting the program. Otherwise, turn the
instrument off after doing any service work while in DOS.
3. Press the Alt and x keys simultaneously. The CD1800 program is removed from computer memory. The DOS
prompt (C:\1800>) displays.
4. Execute the desired DOS commands.
Exiting DOS
To exit the DOS program and return to the CD1800 program in the case where the CD1800 program remains loaded
(Alt and d was used to access DOS), type exit at the C:\> prompt and press the Enter key. The DIAGNOSTICS Menu
displays.
To exit the DOS program in the case where the CD1800 program is not loaded, (Alt and x was used to exit the CD1800
program), turn the instrument OFF then ON again to reinitialize the system.
Note
Turning the instrument OFF then ON again prepares the system for normal operation.
Change Drives
When accessing DOS from the CD1800 program, the C:\> prompt displays, indicating the C drive is selected. To change to
the A drive (floppy disk), type A: and press Enter. The prompt changes to A:\> indicating the A drive is selected.
To move from a directory to the root directory (C:\) type cd\ and press the Enter key.
To move one level closer to the root directory type cd.. (where .. are two periods) and press the Enter key. For example, to
move from CD1800\HELP to CD1800 subdirectory type cd.. and press the Enter key.
To move one level away from the root directory type cd_[directory name] (where _ indicates a space) and press the Enter
key. For example, to move from the root directory (c:\) to the CD1800 directory type cd cd1800 and press the Enter key.
Copy
To copy a file on the CELL-DYN 1800 hard drive (C drive) to a floppy disk (on the A drive):
1. Exit the CD1800 program procedure above. Change to A drive, if the file to be copied is on A drive.
2. If necessary, use the cd command to access the directory containing the file(s) to be copied.
3. At the C:\> prompt, type Copy_C:\[filename]_A:\ (where [filename] is the complete file name and _ indicates a
space between characters).
Example: to copy file CONFIG.SYS from the C drive to the A drive, type copy c:\config.sys a:\ and press the Enter
key.
To copy this file from the A drive to the C drive, type copy a:\config.sys c:\ and press the Enter key.
Delete a File
To delete a file:
1. Access DOS (Exiting from CD1800 to DOS). Change to A drive if the file to be deleted is on A drive.
2. If necessary, use the cd command to access the directory containing the file(s) to be deleted.
3. At the C:\> prompt, type delete_ [filename] (where [filename] is the complete file name and _ indicates a space
between characters) and press the Enter key. For example, to delete a file called Sample from the C drive type
delete sample and press the Enter key.
Undelete a File
1. At the C:\> prompt, type undelete_ [filename] (where [filename] is the complete file name and _ indicates a space
between characters) and press the Enter key. For example, to undelete a file called Sample from the C drive, type
undelete sample and press the Enter key.
1. Access DOS (Exiting from CD1800 to DOS). Change to A drive if the files to be viewed are on A drive.
2. If necessary, use the cd command to access the directory containing the file(s) to be viewed.
3. At the C:\> prompt type dir and press the Enter key.
Note
If there are many files, the file names scroll down the screen too rapidly for the user to see. The user can prevent this
by adding "switches" to the DIR command using the format: DIR_[switch] (where the underscore ("_") indicates a
space). For example, to display the file names a screen at a time, type dir_/p and press the Enter key. To display the
file names in columns across the width of the page: type dir_/w and press the Enter key.
Engineering Drawings and Schematics
见 CD1800 电路图
CELL-DYN 1800 PCB Reference
AC Regulator PCB
1 AC Regulator PCB 14 J1
2 J2 15 J3
3 TP6 16 J8
6 TP4 (GND) 19 J5
8 J4 21 LD2 (-12V)
9 J7 22 TP2 (GND)
12 R28 25 J6
13 J9
Backlight PCB
2 CN1
6 SPM
4 TP1 (+5V) 10 J9
CCM PCB
1 CCM PCB 8 J6
3 S3 (BUFALLO) 10 E2 (DGND)
4 J2 (MPM) 11 J8
5 J1 (CDM) 12 E3 (DGND)
6 J3 (SPM) 13 E1 (DGND)
7 J5 (DLA) 14 LED1
CDM PCB
1 CDM PCB 18 J25
3 J11 20 J27
4 J12 21 J28
5 J13 22 J29
6 J14 23 J30
7 J15 24 J31
8 J16 25 J10
9 J17 26 J9
10 J18 27 J8
11 J19 28 J7
12 J32 29 J6
13 J20 30 J5
14 J21 31 J4
15 J22 32 J3
16 J23 33 J2
17 J24 34 J1
DC Regulator PCB
1 DC Regulator PCB 9 J1
3 J5 11 LD3 (+12V)
4 J7 12 J8
5 J6 13 J9
6 J4 14 J10
7 J2 15 LD1 (+5V)
8 J3
DLA PCB
1 DLA PCB 5 DS2 LED
2 J1 6 J2
3 TP1 7 TP2
4 DS1 LED
Indicator PCB
1 Indicator PCB 3 LD2 (yellow)
2 E2 5 J2
3 E1
2 J8 7 J3
3 J7 8 JP1B
4 J5 9 JP1A
MPM PCB
4 TP3 (RESET) 10 J2
5 J3 thru J14 11 J1
6 TP7 (+5V)
PAM PCB
2 J3 14 JMP3
3 J4 15 R14
4 JMP1 16 J5
5 J1 17 R72
6 TP7 18 LD3
7 R4 19 TP5 (GND)
8 TP1 20 R66
11 J2 23 LD2 (+12V)
3 GND
PRM PCB
1 PRM PCB 6 J1
2 J5 7 J2
3 J8 8 J3
4 J7 9 J4
5 J6 10 F1
2 J1 5 TP3
3 TP1
2 E1 6 E6
3 E2 7 E5
4 E3 8 E4
SBC PCB
2 Fan 7 COM1
3 Power 8 COM2
4 Hard Disk Drive 9 PC Keyboard
SDM PCB
3 J9 7 J10
SPM PCB
1 SPM PCB 6 E2 (AGND)
2 R22 (WBC) 7 J1
3 R27 (RBC) 8 J4
Touchpad PCB
1 Touchpad PCB 3 J2
2 J1
3 P1 8 R18
4 TP3 9 TP2
5 TP1 10 E1
3 TP2
Removal & Replacement
Service Equipment Required
Every system subassembly and module can be removed and replaced using the tools listed below. Substitute equipment
having the same characteristics as those listed can also be used. Whenever possible, choose non-magnetic tools to avoid
damaging sensitive printed circuit board mounted components and magnetic media.
Knife Utility
Wrench Adjustable
Covers (CD18-A1)
Removal
Note
Be careful when removing the C clip.
The hinge contains a plastic washer [7]
and a spring [8] which can be lost.
Replacement
Note
During the cycle, make sure that cover does not interfere with
instrument operation.
Removal
Check Caution
Sample Be sure the sample probe is out of the
Probe way before attempting to remove the
Position lower front cover to prevent damage to
the probe.
Note
It may be necessary to lift the right side of
the cover to clear the bezel indentation
while sliding the cover to the left.
Replacement
Verification
Note
During the cycle, make sure that cover does not interfere with
instrument operation.
Replacement
Verification
Note
The Phillips screw secures the bezel to
the instrument.
Note
Do not pull bezel too far away from
instrument. Several cables must be
removed before completely removing the
bezel from the instrument.
Note
Mark all cable connectors and note pin 1
(red stripe on ribbon) on the floppy drive
data cable.
Replacement
Verification
Action Steps Reference
Removal
Replacement
Verification
Action Steps Reference
Verification of Proper Cover 1. Verify that cutouts on front and rear portion of cover are placed
Installation under locking nuts.
2. Be sure cover is secure against right side of instrument.
Removal
Replacement
Verification
Removal
Note
Note the location of the cutout notch
and WBC etched in the Aperture
Plate.
Replacement
Verification
Verify Background 1. Run a normal background count cycle and verify results are within acceptable
Counts limits. Refer to Specifications & Checklist.
Verify Quality 1. Run three (3) levels of CELL-DYN controls and verify results are within
Controls acceptable assay limits.
Both the von Behrens RBC and WBC transducers are removed according to the following procedure:
Removal
Replacement
Verification
Verify Gain Settings 1. Perform VP-09 Signal Processor Module (SPM) Verification/Adjustment.
Verify Background 1. Run a normal background count cycle and verify results are within acceptable
Counts limits. Refer to Specifications & Checklist.
Verify Quality 1. Run three (3) levels of CELL-DYN controls and verify results are within
Controls acceptable assay limits.
Removal
Caution
Take extreme care not to allow the
liquid to touch the printed circuit
boards; it can short them out when
power is turned on. If the board does
get wet, make sure it is dry before
turning power back ON.
Removal
Caution
Take extreme care not to allow the
liquid to touch the printed circuit
boards; it can short them out when
power is turned on. If the board does
get wet, make sure it is dry before
turning power back ON.
Note
The RBC Metering PCB [1] is installed
upside down. Push the metering tube
all the way down to access the top
screw securing the metering board to
the flow panel.
Replacement
Note
Verify that tubing is not pinched when installing the plastic shield.
Verification
Verify Count 1. Verify appropriate count times (see VP-04 Metering System Timing Adjustments -
Times RBC and WBC:
o Verify RBC/PLT count times after installing RBC/PLT metering PCB.
o Verify WBC count times after installing WBC metering PCB.
Removal
Note
This cover is hinged and stands in a
folded back position without support.
Note
Use absorbent tissue to capture any
excess liquid in the HGB Flow Cell.
Replacement
Verification
Verify HGB Offset, Gain, 1. Perform HGB Self Test and Adjust HGB Gain (see Pre-Amp Alignments).
Precision 2. Verify HGB precision.
Verify Background 1. Run a normal background count cycle and verify results are within
Counts acceptable limits. Refer to Specifications & Checklist.
Verify Quality Controls 1. Run three (3) levels of CELL-DYN controls and verify results are within
acceptable assay limits.
Removal
Remove Note
Tubing Cut any cable ties on cables
surrounding the fluid power supply
assembly as needed.
Note
Verify that all cable connectors and
tubing are disconnected.
Replacement
Verification
Removal
Note
Before lifting the Sample Probe Driver
Assembly completely out of the
instrument, check to make sure that all
cables are disconnected. Disconnect
any cables that you find still attached,
making sure to mark them before
unplugging.
Replacement
Verification
Verify Normal Operation 1. Run three (3) cycles and verify normal operation, e.g. binding, liquid
leaks, etc.
Hazard
Removal
Note
Screws are secured by a nut. Verify
that ALL nuts are removed prior to
removing fan.
Note
If necessary, rotate the split pin
installed on the directional valve
drive linkage of the Diluent Syringe
to allow the fan to be removed.
Note
Use absorbent tissue to capture any
excess liquid.
Replacement
Verification
Verify Quality 1. Run three (3) levels of CELL-DYN controls and verify results are within
Controls acceptable assay limits.
Removal
Verification
Verify Sample 1. Verify sample volume (VP-10 Diluent and Sample Verification/Adjustment).
Volume
Verify Quality 1. Run three (3) levels of CELL-DYN controls and verify results are within acceptable
Controls assay limits.
Removal
Replacement
Verification
Verify Quality 1. Run three (3) levels of CELL-DYN controls and verify results are within
Controls acceptable assay limits.
Removal
Note
The J3 connector is for WBC and the
J4 connector is for RBC/PLT.
Replacement
Verification
Verify PAM Printed Circuit Board 1. Perform VP-08 Pre-Amplifier Module (PAM) Adjustment.
Adjustments
Verify SPM Printed Circuit Board 1. Perform VP-09 Signal Processor Module (SPM)
Adjustments Verification/Adjustment.
Note
Repeat this procedure for the snap-in
pin at the center of the PCB.
Verification
Verify Motor Operation 1. Perform VP-11 Stepper Motor Power Test and Verification.
Removal
Note
Repeat this procedure for the snap-in
pin at the center of the PCB.
Replacement
Verification
Verify Normal Operation 1. Run three (3) cycles and verify normal operation, e.g. binding, liquid leaks,
etc.
Removal
Caution
Once the screws are removed, the
card cage drops down. Excessive
movement on the card cage can
damage connectors on the
Backplane PCB.
Note
Verify that all four (4) screws
securing the hard disk drive to the
card cage are accessible.
Note
The same alignment must be
maintained during replacement.
Replacement
Verification
Verify Hard Disk Drive Setup 1. Perform VP-14 Hard Disk Drive Setup and Verification.
Removal
Verification
Removal
Note
Pulls the board slightly out of the card
cage.
Replacement
Verification
Verify SPM PCB 1. Perform VP-09 Signal Processor Module (SPM) Verification/Adjustment.
Adjustments
Note
Make SPM adjustments in the following order:
o RBC
o RER
o WBC
o PLT
Verify Background 1. Run a normal background count cycle and verify results are within acceptable
Counts limits. Refer to Specifications & Checklist.
Verify Quality Controls 1. Run three (3) levels of CELL-DYN controls and verify results are within
acceptable assay limits.
Removal
Note
Pulls the board slightly out of the card
cage.
Replacement
Verification
Verify Background 1. Run a normal background count cycle and verify results are within acceptable
Counts limits. Refer to Specifications & Checklist.
Verify Quality 1. Run three (3) levels of CELL-DYN controls and verify results are within
Controls acceptable assay limits.
Removal
Action Steps Reference
Verification
Verify Background 1. Run a normal background count cycle and verify results are within acceptable
Counts limits. Refer to Specifications & Checklist.
Removal
Replacement
Verification
Removal
Replacement
Verification
Verify Card Cage Backplane 1. Perform VP-06 Card Cage Backplane Test Points.
PCB Voltages
Verify Background Counts 1. Run a normal background count cycle and verify results are within
acceptable limits. Refer to Specifications & Checklist.
Verify Quality Controls 1. Run three (3) levels of CELL-DYN Controls and verify results are within
acceptable assay limits.
LCD/Keyboard (CD18-G1)
Removal
Note
When disconnecting the LVDS cable
connector, pull the connector, NOT
the wires.
Replacement
Note
Failure to properly route and secure
the LVDS cable may result in
damage to the wires and connector.
Removal
Note
Never touch a PCB when the power
is ON.
Note
The power switch contains two (2)
black and two (2) white cable
connectors.
Caution
The power switch is vulnerable to
damage during this process.
Replacement
Verification
Verify System Voltages 1. Perform VP-06 Card Cage Backplane Test Points.
Verification Procedures
Materials Required See Test Equipment and Supplies Required. Time 01:00 hr
4 Stopwatch 1
10 Hemostats 3
12 Silicone Tubing AR
13 Standard Tool Kit 1
18 10 mL Graduated Cylinder 1
19 40 µL Micropipettes AR
20 100 µL Micropipettes AR
21 50 mL Counting Cups AR
22 25 mL Graduated Cylinder 1
24 Oscilloscope (Optional) 1
AR = As Required
Perform the following procedure to prepare the CELL-DYN 1800 System for alignment/verification:
Check Reagents 1. Verify all reagents are correct and are 1/3 full.
Remove Upper 1. Remove Upper Front Cover (RR-A1.01 Upper Front Cover).
Front Cover
Remove Lower 1. Remove Lower Front Cover (RR-A1.02 Lower Front Cover).
Front Cover
Clean Aperture 1. Remove and clean both RBC/PLT and WBC aperture plates following the
Plates procedure in Section 9: Service and Maintenance, Subsection:
Cleaning/Replacing Aperture Plates of the CELL-DYN 1800 Operator's Manual.
Clean HGB Flow 1. Clean the HGB Flow Cell following the procedure in Section 9, Subsection:
Cell Cleaning the HGB Flow Cell of the CELL-DYN 1800 Operator's Manual.
Re-initialize 1. Re-initialize the instrument by turning the system OFF then ON again.
Instrument
Check Tubing, 1. When Initialization is complete, press [PRIME/RUN] to prime the instrument.
Look For Leaks Observe the flow system for leaks, tubing placement, pinched tubing, etc.
Run Background 1. Run a Background count. Verify all background values are within the following
Count specifications:
Record Dilution 1. Type 94043 to display the DILUTION FACTORS screen. Record all Dilution
Factors Factors.
Enter Service 1. In the MAIN MENU, type 999 for Operator ID and press Enter on the PC keyboard
Personnel ID or membrane keypad. This is to identify all runs performed by service personnel.
Order of Alignment/Verification
The following procedures are presented to ensure proper alignment of the CELL-DYN 1800 System.
Sequence Procedure
• At the completion of each Vacuum and Pressure Adjustment procedure, verify the WBC and RBC count times and
normal system operation
• Prior to starting any electronic alignment, verify the power supply voltages in VP-06 Card Cage Backplane Test
Points and VP-07 Cable Distribution Module Test Points.
• Whenever an alignment adjustment is performed on the Pre-Amplifier Module (PAM), verify all parameters on the
Signal Processor Module (SPM).
VP-02 Decontamination
Decontamination 1. The OSHA Bloodborne Pathogen Rule (29 CFR Part 1910.1030) requires the
Ratio decontamination of laboratory equipment prior to servicing or shipment:
o Decontaminate the instrument by performing the Auto-Clean cycle. This
cycle flushes all of the fluid pathways with reagents to purge any waste from
the fluid pathways. The Open Mode Sample Probe is automatically rinsed
after every cycle. The surfaces of the instrument should be wiped with a
nonabrasive detergent solution to remove any soiling, then wiped with a
tuberculocidal disinfectant, such as a 0.5% sodium hypochlorite solution.
2. To calculate the percent (%) sodium hypochlorite concentration desired, use the
following formula:
a. A = Percent (%) of sodium hypochlorite solution desired
b. B = Percent (%) of sodium hypochlorite stock solution (as purchased)
c. X = Parts of water to be mixed with one part of the sodium hypochlorite stock
solution:
e. Add 9.5 parts deionized water to 1 part bleach to obtain a 0.5% sodium
hypochlorite solution, or 9.5 mL of deionized water to 0.5 mL of bleach
(5.25% sodium hypochlorite) to obtain 10 mL of a 0.5% solution of sodium
hypochlorite.
3. If the instrument is to be shipped, it must be decontaminated prior to shipment. This
is accomplished by pressing the [CLEAN FOR SHIPPING] key in the SPECIAL
PROTOCOLS menu. Instructions for this procedure are given in the
CELL-DYN 1800 Operator's Manual, (Section 9 Preparing the Instrument for
Extended Periods of Non-Use or Shipping).
Purpose Align and adjust pressure and vacuum. Module CELL-DYN 1800
Regulator Alignment
The CELL-DYN 1800 System utilizes one vacuum and two pressure levels to accomplish the following tasks: moving sample,
reagents, and waste, bubble mixing of sample, and back flushing RBC and WBC apertures. The vacuum and bubble mix
pressures are adjustable by a solid-state regulator. The backflush pressure is not critical and the pump is under direct
computer control.
The solid state regulator has two input ports: P1 for pressure and P2 for vacuum. It also has jumper terminals which
accommodate all desired vacuum and pressure ranges. The jumper positions are:
Regulator Alignment
1. Disconnect the appropriate pumps on the Pump Relay Board and bleed off
pressure from the Accumulators.
o J4 for low pressure
o J2 for vacuum on new-style pumps
2. Remove the pressure or vacuum line from the top of the regulator.
Measure Voltage 1. On the Pump Relay Board the voltage should be 5.0 ± 0.15 volts at J7 pin 7 for
vacuum, and J8 pin 5 for pressure.
Note
You must remove the board mounting screws and free the rest of the board to
make the remaining adjustments. Disconnect J6 or J7 while relocating the
appropriate printed circuit board to avoid shorting out components.
2. Note the current position of jumper E1 and set E1 to the C-D position.
3. Connect the DVM (Digitial Volt Meter) negative lead to TP3 (GND) on regulator
board. Connect the positive lead to TP2 (REF). The voltage should be
1.00V ± 0.14 volts.
o If voltage is not correct, re-check Step 1.
o If voltage is still not correct, replace the regulator.
4. Connect the DVM negative lead to TP1 and the positive lead to TP2, and read the
voltage. Adjust R18 for a voltage of 0.000 ± 0.005.
Note
If voltage is negative, turn R18 (offset) clockwise; if the voltage is positive, turn R18
counterclockwise.
Reconnect 1. Reconnect the pressure or vacuum line to the top of the regulator and reconnect
Pressure / Vacuum the cable(s) on the pump relay module.
Line
Connect 1. Locate the small silicone tubing connected to the in-line fitting at top of the 0.5 psi
Pressure Gauge Pressure Accumulator (located closest to the rear of the instrument).
2. Connect a 0-5 psi gauge in-line with the silicone tubing and fitting (the Pressure
Accumulator is located on the part of the Fluid Power Supply which is closest to the
rear of the instrument).
Adjust 1. Adjust R16 on the Regulator board for 0.55 psi ± 0.02.
Regulator
Note
Clockwise adjustment increases pressure and counterclockwise adjustment
decreases pressure.
2. Remove the gauge and reconnect the silicone tubing to in-line connector.
Locate In-line 1. Locate solenoid valve 1-6 and trace the tubing back through the flow panel to the
Connector in-line connector.
Verify Correct 1. When the pump activates, verify a pressure of no less than 4 psi.
Pressure o If the pressure is less than 4 psi, check for leaks. Replace the Pressure
Pump if necessary. (Refer to RR-C1.01 Fluid Power Supply.)
1. Locate the solid state vacuum regulator on the rear side of the Fluid Power Supply
(closest to the flow panel).
Adjust R16 1. Remove the TYGON tubing from the top of the Vacuum Accumulator (located on the
back of the Fluid Power Supply closest to the flow panel) and connect a 0-30" Hg
gauge in-line between the tubing and the Accumulator.
2. Adjust R16 (accessible only from the front of the Fluid Power Supply) for
8.25" Hg ± 0.25".
Note
Clockwise adjustment increases vacuum and counterclockwise adjustment decreases
vacuum.
Reconnect 1. Once R16 is adjusted, remove the vacuum gauge and reconnect the tubing line to the
Tubing Line regulator.
Purpose Adjust Metering System to regulate count times. Module CELL-DYN 1800
Links
Detector Masked Flow Error Time Valid Meniscus Time Clog Time
WBC Lower 0 - 3.5 3.5 - 4.0 4.0 - 6.5 6.5+ or greater than the moving average
RBC/PLT Lower 0 - 5.5 5.5 - 6.0 6.0 - 7.5 7.5+ or greater than the moving average
This figure illustrates the timing relationships for WBC and RBC measurements.
Follow the steps below for adjusting the timing of the RBC metering system:
Verify 1. Ensure that the RBC aperture plate has been removed, cleaned, and reinstalled following
Count the procedure in Section 9: Service and Maintenance, Subsection: Cleaning/Replacing
Time Aperture Plates of the CELL-DYN 1800 Operator's Manual.
2. Run a background count and verify that the RBC displayed count time (T2) is 6.7 seconds
± 0.2 seconds.
o If the specification is met, go to Step 4.
o If the specification is not met, go to Step 3.
3. Readjust the vacuum to correct the count time if the time is out of specification (refer to
Adjust R16, Step 2 of the Vacuum Adjustment (8-inch) procedure previously discussed).
4. From the MAIN Menu, press [DIAGNOSTICS] followed by [RAW DATA].
5. Verify the time displayed for RBC UPTIME is 4.8 to 5.2 seconds. (This time, displayed in
milliseconds, represents T1, which is the time when the RBC valve 1-2 opens until the
meniscus reaches the upper detector.)
o If the specification is met, the procedure is complete.
o If the specification is not met, go to Step 6.
6. Adjust the metering tube.
o Raise the metering tube to increase the upper time.
o Lower the metering tube to decrease the upper time.
7. Repeat Step 2 through Step 5 until the RBC Up Time/Count Time is within specification.
Follow the steps below for adjusting the timing of the WBC metering system:
Prerequisite Note
The RBC count times (T1 & T2) must be within specification before performing this
procedure.
Verify Count 1. Remove and clean the WBC aperture plate as described in Section 9: Service and
Time Maintenance, Subsection: Cleaning/Replacing Aperture Plates of the
CELL-DYN 1800 Operator's Manual.
2. Run a background count and verify the count time displayed to the right of the WBC
histogram is 5.0 ± 0.5 seconds.
Note
There is no count time adjustment procedure.
Purpose Provide instructions to verify and/or modify CMOS settings. Module CELL-DYN 1800
Access CMOS Setup 1. Power ON the analyzer and press the Delete key, when prompted, to display
Utility Screen the CMOS Setup Utility Screen.
1. Use the up/down arrow keys to select the STANDARD CMOS SETUP.
2. Press the Enter key.
3. Refer to the figure (H2_6006) and make any necessary adjustments to the
Standard CMOS Setup screen.
Note
Verify/Modify
Press the up/down arrow keys to select a setting and then press PgUp/PgDn
Standard CMOS
keys to modify.
Setup
Note
Press the up/down arrow keys to select a setting and then press PgUp/PgDn
keys to modify.
Verify/Modify
Advanced CMOS
Setup
1. Use the up/down arrow keys to select the ADVANCED CHIPSET SETUP.
2. Press the Enter key.
3. Refer to the figure and make any necessary adjustments to the Advanced
Chipset Setup screen.
Note
Press the up/down arrow keys to select a setting and then press PgUp/PgDn
keys to modify.
Verify/Modify
Advanced Chipset
Setup
Note
The SAMSUNG LCD assembly is Type 8, while the TOSHIBA LCD assembly is Type 7.
1. Use the up/down arrow keys to select the POWER MANAGEMENT SETUP.
2. Press the Enter key.
3. Refer to the figure and make any necessary adjustments to the Power
Management Setup screen.
Note
Verify/Modify Power Press the up/down arrow keys to select a setting and then press PgUp/PgDn
Management Setup keys to modify.
Note
Press the up/down arrow keys to select a setting and then press PgUp/PgDn
keys to modify.
Verify/Modify
Peripheral Setup
Save CMOS Settings 1. Press F10.
and Exit CMOS Setup
The message Save current settings and exit (Y/N)? displays.
Note
Saves the CMOS settings, exits the CMOS Setup Utility and restarts the
computer.
Verification 1. Verify that the system boots up properly to the application software.
The following table lists the test points located on the backplane board and the figure illustrates the test points on the
backplane board.
Note
LEDs indicate voltages supplied to the board.
The following table lists the test points located on the Cable Distribution Module (CDM).
Location
Pin Function Voltage Ripple
(J5 and J8)
Note
On the CDM, +28V is used to initially energize solenoids and +14V is used to keep them closed.
Purpose Adjust HGB self test and gain. Module CELL-DYN 1800
Adjust RBC/PLT and WBC aperture currents.
The Pre-Amplifier Module (PAM) is located on the upper right corner of the Flow Panel. Hemoglobin circuitry, and the
RBC/PLT and WBC aperture currents require verification and/or adjustment on this module. The HGB flow cell should be
cleaned before performing hemoglobin alignments (refer to Section 9, Subsection: Cleaning the HGB Flow Cell in the
CELL-DYN 1800 Operator's Manual).
The following table lists the test points located on the PAM.
Test points for the PAM are shown in the figure below.
Remove Lower Front Cover 1. Remove Lower Front Cover (RR-A1.02 Lower Front Cover).
Remove Pre-amp Cover 1. Remove the pre-amp cover (RR-F1.01 PAM (Pre-Amplifier Module)).
Note
If voltage remains unchanged, the jumper may be in the wrong position.
Adjust HGB Gain 1. Press the touch plate to cycle the instrument and fill the HGB flow cell
with fresh reagent.
2. Verify the voltage at TP7.
3. Adjust R4 (HGB Gain Adjust) for 5.2 volts ± 0.2 volts.
4. Press the touch plate to cycle the instrument and fill the HGB flow cell
with fresh reagent.
5. Verify the voltage at TP7.
Note
If voltage does not meet specification, repeat Step 2 and Step 3.
Verify +100 Volts 1. Connect the positive lead of the DVM to TP4, and the negative lead to
TP5 (Gnd).
2. Verify 100 volts ±0.5 volts.
Adjust RBC/PLT Aperture 1. Disconnect the RBC signal cable from J4.
Current 2. Connect a 20K Ohm 1/4 watt resistor across pins 4 and 5 of J4.
3. Connect the DVM across the resistor.
4. Run a background count.
5. Adjust R72 (RBC/PLT aperture current adjust) for 11.0 volts ± 0.05 volts.
6. Remove the resistor and reconnect cable to J4.
Adjust WBC Aperture 1. Disconnect the WBC signal cable from J3.
Current 2. Connect a 15K Ohm 1/4 watt resistor across pins 4 and 5 of J3.
3. Connect the DVM across resistor.
4. Run a background count.
5. Adjust R66 (WBC aperture current adjust) for 12.0 volts ± 0.05 volts.
6. Remove the resistor and reconnect cable to J3.
Install Covers 1. Reinstall the top cover, upper and lower front covers, bezel and pre-amp
cover.
The following table contains the specifications for the Pre-Amp Module alignment.
Pre-Amp Alignments
Purpose Adjust WBC, RBC, PLT Gain. and RBC Edit Ratio (RER) percentage (%). Module CELL-DYN 1800
Note
The Signal Processor Module (SPM) is located in the main electronics card cage. Alignment of the WBC, RBC, and
PLT gains are critical adjustments that must be verified and/or adjusted before instrument accuracy can be
established. It is important to maintain the adjustment order for this verification procedure. The adjustment order is:
Sequence Activity
1 RBC Gain
2 RER Adjustment
3 WBC Gain
4 PLT Gain
Latex particles are used to perform these adjustments. The particles must be mixed vigorously before diluting to obtain
accurate results.
The Gain and RBC Cell Editing adjustments are performed in the Gain Adjust Mode, which allows multiple counts to be run
on the same sample. When Gain Adjust Mode is entered, whatever is in the Pre-Mix Cup is transferred to the WBC Cup and
5 mL of diluent is left undisturbed in the RBC Cup.
When performing RBC or PLT adjustments, 10 mL of diluent must be placed in Pre-Mix Cup before entering Gain Adjust
Mode to prevent air from being pulled into the von Behrens WBC Transducer during a count cycle.
Test points for the SPM are shown in the figure below:
Prepare Latex 1. Pour 10 mL of diluent into the Pre-Mix Cup and add 1 drop of well-mixed 5.0 latex
Dilution particle solution to the 5 mL of diluent already in the RBC Mixing Chamber.
2. From the RUN screen, press [SPECIMEN TYPE].
3. Press Shift and # (simultaneously) to activate the Gain Adjust Mode.
Note
The message <PREPARING FOR GAIN ADJUST.> displays and WBC dilution is
transferred to the WBC Mixing Chamber. When the preparation cycle is complete,
the RUN screen displays.
Note
Always return to the RUN screen to run a cycle.
Note
When the procedure is completed, or if there is a need to exit the Gain Adjust Mode
while performing this procedure, such as constant clogs or improper dilution ratio,
press [SPECIMEN TYPE] followed by [PATIENT SPECIMEN].
Note
The RBC count should be between 3.5 and 4.5. If the count is outside this range,
modify the dilution ratio of the latex particles, exit Gain Adjust Mode and try again.
Note
The key changes to [SMOOTHING ON] and is highlighted, indicating the smoothing
function is ON.
Note
Turning R27 clockwise decreases the channel number; turning R27
counterclockwise increases the channel number.
6. Return to the RUN screen and run another cycle.
7. Observe the channel number for peak count of 100.
o If the gain is still outside of specification, repeat this process until the peak
count meets the specification (100 is in Channel 98 ± 1 channel). Refer to
RBC Gain Adjustments (Example)
Note
Turning R27 clockwise decreases the channel number; turning R27
counterclockwise increases the channel number.
RER Adjustment
Action Steps Reference
Note
The message <PREPARING FOR GAIN ADJ.>
displays and diluent in the Pre-Mix Cup is
transferred to the WBC Mixing Chamber. When
the preparation cycle is complete, the RUN
screen displays.
Note
Example: A count of 4 million relates to an edit
ratio of 28.0% ± 1.3.
Note
Turning R51 clockwise decreases the RER
percentage; turning R51 counterclockwise
increases the RER percentage.
WBC Gain
Note
Each 10 mL dispense requires key to be pressed twice.
6. Add 1 drop of well mixed 5.0 µm latex particle solution into the 20 mL of diluent and mix
well.
7. Before entering the Gain Adjust Mode, use the prepared WBC latex dilution to fill the
Pre-Mix cup up to the level of the diluent inlet port.
8. From the MAIN Menu, press [RUN] followed by [SPECIMEN TYPE].
9. Press Shift and # (simultaneously) to activate the Gain Adjust Mode.
Note
The message <PREPARING FOR GAIN ADJ.> displays and WBC dilution is
transferred to the WBC Mixing Chamber. When the preparation cycle is complete, the
RUN screen displays.
Note
The key changes to [SMOOTHING ON] and is highlighted, indicating the smoothing
function is ON.
Note
Turning R22 clockwise decreases the channel number; turning R22 counterclockwise
increases the channel number.
PLT Gain
Note
Each 10 mL dispense requires key to be pressed
twice.
Enter Gain 1. Pour 10 mL of diluent into the Pre-Mix Cup (do not add
Adjust any latex particle solution).
Mode 2. From the RUN screen, press [SPECIMEN TYPE].
3. Press Shift and # (simultaneously) to activate the
Gain Adjust Mode.
Note
The message <PREPARING FOR GAIN ADJ.>
displays and WBC dilution is transferred to the WBC
Mixing Chamber. When the preparation cycle is
complete, the RUN screen displays.
Note
When running PLT counts, the lower front cover must
be on the instrument to prevent noise from interfering
with PLT results.
Note
The key changes to [SMOOTHING ON] and is
highlighted, indicating the smoothing function is ON.
Note
Turning R41 clockwise decreases the channel
number; turning R41 counterclockwise increases the
channel number.
Purpose Verify diluent dispense to maintain proper dilution ratios. Module CELL-DYN 1800
WARNING
Potential Biohazard
This procedure consist of both Diluent and Sample adjustment that can be accessed with the following links:
Diluent Volume Verification
To minimize problems like coincidence passage, the CELL-DYN 1800 uses two different dilution ratios of whole blood to
diluent. The ratio for WBC/HGB is 1:285; the ratio for RBC/MCV/PLT is 1:12,801. The following procedure is used to verify
the diluent dispense to maintain proper dilution ratios and thereby optimize instrument performance.
Verify 1. From the MAIN Menu, press [SPECIAL PROTOCOLS] followed by [MORE] twice.
Diluent 2. Place an empty 10 mL graduated cylinder under the Sample Probe and press [10 mL
Volume DISPENSE] twice to dispense 10 mL of diluent.
3. Verify a volume of 10 mL ± 0.2 mL.
4. Place an empty 10 mL graduated cylinder under Sample Probe and press [1/50
DILUTION]. Once the probe has returned to the aspirate position, press
[1/50 DISPENSE] to dispense.
5. Verify a volume of 5 mL ± 0.1 mL.
Note
The volume dispensed is under direct computer control. If the volume is out-of-range, the
Dispenser and Stepper Motor drive circuitry must be repaired. (Refer to RR-E1.02
Diluent Syringe Driver Assembly.)
Sample Volume 1. Remove the 1/32" silicone tubing attached to the top of the sample probe.
Verification 2. Attach a 100 microliter pipette to the silicone tubing.
3. Place the tip of the pipette on the bottom of a small container and press [10 mL
DISPENSE] twice.
Note
Keep the tip of the pipette submerged when dispensing.
4. Take the pipette out of the container and wipe any drop from the end of the pipette,
being careful not to wick any liquid from the end.
5. Press [1/50 DILUTION].
6. Verify that the column of liquid is no more than 1/16 inch above or below 100
microliter mark on the pipette.
7. Replace the 100 microliter pipette with a 40 microliter pipette.
8. Place the tip of the 40 microliter pipette on the bottom of a small container and press
[1/50 DISPENSE].
9. Take the pipette out of the container and wipe any drop from the end of the pipette,
being careful not to wick any liquid from the end.
10. Press [1/250 DILUTION].
11.Verify that the column of liquid is no more than 1/16" above or below 40 microliter
mark on the pipette.
Remove Pipette 1. Place a waste container under the tip of the pipette and press [1/250 DISPENSE].
2. Remove the pipette and re-attach the silicone tubing to the sample probe.
Note
The volume aspirated is under direct computer control. If the volume is out-of-range,
the Sample Syringe and Stepper Motor Drive must be repaired. (Refer to RR-E1.04
Sample Syringe Driver Assembly.)
Test Stepper 1. From the MAIN Menu, press [DIAGNOSTICS] followed by [MORE] three times.
Motor Power Then press [SERVICE DEC CODE].
2. Type in 128 from the keyboard and press Enter.
3. The test runs for approximately 30 seconds.
4. When the test is complete, the results are displayed on the screen.
5. Compare the results displayed on the screen with the nominal values listed in Motor
Power Specifications. Ensure that all values fall within the specified ranges.
Note
Proper probe tip location is essential for
proper cleaning of the probe, and
prevents carryover and CV problems.
This verification procedure is used to verify the operation of the serial port (COM 1) that is used for communication with a
Laboratory Information System (LIS). This procedure only verifies the transmit function.
Note
The screens shown in this procedure are given for example only. Each model instrument has slightly different field
structure and content to the count data message that is downloaded. Consult the applicable interface specification
document for details.
Note
All the other boxes are blocked (gray).
Note
If the message "Unable to open COM1.
Please check your port settings"
displays, there may be another device
assigned to the COM1 port, such as the
modem. Contact the local area IS
administrator for assistance with resolving
device interrupt conflicts.
Note
The screen below is shown for example
only.
Purpose Provide instructions to verify/modify CMOS Settings, partition and format Module CELL-DYN 1800
replacement hard disk drives
Preparation Note
If replacing an existing customer hard disk drive, print copies of all customer
settings, quality control data files, and the data log if the drive is accessible. It
may not be possible to access this data if the hard disk drive is corrupt or
defective.
Install Hard Drive 1. Install the new hard drive using the existing hardware and cables (refer to
RR-F1.04 Hard Disk Drive).
Access CMOS Setup 1. Power ON the analyzer and press the Delete key, when prompted, to display
Utility Screen the CMOS Setup Utility Screen.
Verify CMOS Settings: 1. Use the up/down arrow keys to select the STANDARD CMOS SETUP.
Standard CMOS Setup 2. Press the Enter key.
3. Set Pri Master to Auto.
4. Set 32 Bit Mode to ON.
Note
Pri Slave, Sec Master and Sec Slave show Not Installed.
Verify CMOS Settings: 1. Use the up/down arrow keys to select the ADVANCED CMOS SETUP.
Advanced CMOS Setup 2. Press the Enter key.
3. Set Quick Boot to Enabled.
4. Set 1st Boot Device to Floppy.
5. Set 2nd Boot Device to IDE-0.
6. Set 3rd Boot Device to Disabled.
7. Set 4th Boot Device to Disabled.
8. Press ESC to exit the screen.
Verify CMOS Settings: 1. Use the up/down arrow keys to select the ADVANCED CHIPSET SETUP.
Advanced Chipset 2. Press the Enter key.
Setup 3. Set LCD CRT Selection to Simultaneous.
4. Set LCD Type to Type 8.
5. Press ESC to exit the screen.
Verify CMOS Settings: 1. Use the up/down arrow keys to select the POWER MANAGEMENT SETUP.
Power Management 2. Press the Enter key.
Setup 3. Set Power Management/APM to Disabled.
4. Set CPU Critical Temperature to 60°C/140°F.
5. Press ESC to exit the screen.
Verify CMOS Settings: 1. Use the up/down arrow keys to select the PERIPHERAL SETUP.
Peripheral Setup 2. Press the Enter key.
3. Set OnBoard FDC to Enabled.
4. Press ESC to exit the screen.
Save CMOS Settings 1. Place the DKS (C) SRV ABBOTT CD3200 (8938070001) (service disk) into
and Exit CMOS Setup the floppy drive.
2. Press F10.
Note
Saves the CMOS settings, exits the CMOS Setup Utility and restarts the
computer.
Create DOS Partition 1. At the A:\> prompt, type FDISK and press Enter.
2. At the FDISK Options menu, enter choice 1 and press Enter.
3. At the Create DOS Partition or Logical DOS Drive menu enter choice 1 and
press Enter.
Note
If you get the message Primary DOS Partition already exists, press Esc to
exit. Then go to Format Hard Drive, Step 1.
The message Do you wish to use the maximum available size for a
Primary DOS Partition and make the partition active (Y/N)? displays.
Note
Uses the maximum size for the hard drive DOS partition and makes it active.
5. When the screen displays System will now restart. Insert DOS system
diskette in drive A:. Press any key when ready, press Enter.
Format Hard Drive 1. At the A:\> prompt, type FORMAT C: /S and press Enter.
The message Volume label (11 characters, ENTER for none)? displays.
Install DOS and System 1. At the A:\> prompt, type INSTALL and press Enter.
Software
Note
Installs DOS and system files.
2. When the system installation is complete, remove the service disk and insert
the CELL-DYN 1800 Installation disk into the floppy drive.
3. Type Install and press Enter.
Note
When the installation is complete, the message "Power off the instrument
and remove diskette. Power on the instrument." displays.
4. Remove the CELL-DYN 1800 Installation disk and power OFF the instrument.
Note
During initial power up after a hard disk drive installation, DO NOT turn off the
instrument while the Data Log is being created. This corrupts the Data Log.
3. If printouts are available, re-enter the data from the printouts. If not, perform
instrument calibration procedure and verify control recovery.
Note
When the installation is complete, the message "Power off the instrument
and remove diskette. Power on the instrument." displays.
Note
During initial power up after a hard disk drive installation, DO NOT turn off the
instrument while the Data Log is being created. This corrupts the Data Log.