vagus Prism Calibration Centre ¢ .
Env. Audit E-mail : info @prismcalibration.com, Info@ eindustries in a
Inst. Supply Website : www. prismcalibration.com, www.eindustries.in PRISM
NABL Approved Calibration Lab cacti erie
Stee 200d
FORMAT NO :- PRISM/DIM/VBM/01 ISSUE NO.:-02 |
(Calibration Certificate No:- 20210227-689-01 [Date Of Receipt / JRF No:-27-02-2021 (20210227-689)
Displacement Measurement
Range:- 0.001 to 1.999 mm
a Applied Reading Thru Master | Reading Observed on master
(mm) (mm) (mm
1 0.100 0.0989 0.0011
2 0.200 0.2000 0.0000
3 0.500 0.5001 0.0001
4 1.000 0.9989 0.0011
3 1.500 1.5001 0.0001
6 1,900 1.9004 0.0004
Remarks:
[()The Measurement Confidence Level i 95.43 % with Coverage ctor k=,
(2) These Results Are Obtained At The Time Of Calibration only relate to he em calibrated
[@)The Calibration Certificate Shall not be reproduced except in full without written approval of prism Calibration Cente.
|) UUC = Unit Under Calibration. Above Instrument Calibrated at Lab,
(5) All the measurement is traceable to national & Intemational Standards
(6) Calibration Procedure No PRISMICALSOPIVBMI3S
(7) Environment Condition During Calibration: 20= °C, 40% 10 60% RM.
(@)The Value Have been rounded off as per 1:2-1960 wherever applicable
eds Hot Wire Anemometer Lab Microcal
Details of Instrument Used for |Range 3 10.20 mis 0-280 mm
Calibration ISr.NoV/ID No. PRISM/LIN/HWA/045 30
Due Date 18-06-2021 11-01-2022
LES-CCLIFFP/PT/445 ILES-Centre for Calibration Laboratory
P/3 17190022 [Accurate Eng.Co.Pvi.Ltd
CALIBRATED BY:- ~~ eEROVEDI2)
Page 02 of 02 ~
-| Vatva, A’bad-382 446.
sh Complex, Beside Syndicate Bank, Nr. Jasoaa Nagar Cross Road-Il, GIDC,
£10), Recceotomer Care : 8098062852, 9099062851 Tech. Help Desk 7878991188“Salbration : . s ®
Validation Prism Calibration Centre é
Env. Audit E-mail : Info@ prismcallbration.com, Info@ sindustrie in aw
Inst. Supply Website: www. prismcalibration.com, www.sindustries.in PRIM
NABL Approved libration Lab we! Your ideas”
"Sine" Yoo
LIBRATION CERTIFICATE
(P) LIM CO
Hhreenathji ind. L’state ,Kujud-Bakrol Road Date of Issue 27/02/2021
bakrol(Bujrang), Ta:Daskrol,Dist:Ahmedabad 382430 Date of Calibration :- 27/022021
[_ Suggested Due Date:- 26/02/2022
FORMAT NO :- PRI M/DIM/VBM/01 ISSUE NO.:-02
Calibration Certificate No:- 20210227-689-01 Date OF Receipt / JRF No:-27-02-2021 (20210227-689)
Identificatic bt IN914734,
Details of Unit Under Calibration VUC) peentlestion Pa
Name of Instrument Vibration Meter
Metrix, |Condition of UUC OK
IN914734 Model No. VM 8200+
Acceleration Measurement
Range:- 0.1 to 199.9 m/sec
Applied Reading Thru Master | Reading Observed on master Deviation In Reading
a (as?) (mis) (must)
1 10 10.1 0.1
2 20 20.0 0.0
3 50 50.0 90
7 100 100.1 o1
3 150 149.9 01
é 200 199.9 01
Velocity Measurement
Range:- 0.1 to 199.9 mm/sec a :
Applied Reading Thru Master | Reading Observed on master Deviation In Reading
Sr.No (mm/s) (mm/s) (mm/s)
0.1
i 10 99 a
; aa 199 +
- 30 49.9 ao
3 0.0
ino 100.0 u
‘ 150.0 oo
3 150 ; a
199. 2
i 200
Page 01 OF 02
Nbad-382 445.
Jicate Bank, Nr. Jasoda Nagar Cross Road, GIDC, Vatva,
F/101, Rudraksh Complex-Il, Beside Syndicate Bank, Nr. Jasoda Nagar Cross Road-Il, GIDC, Vatva, Ab